SAR Test Report-Appendix C-1

FCC ID: WIYVEGA3000-LTE

RF Exposure Info

Download: PDF
FCCID_4029346

FCC SAR Test Report

Appendix C. Calibration Certificate for Probe and Dipole

The SPEAG calibration certificates are shown as follows.




Report Format Version 5.0.0                                Issued Date   :   Sep. 19, 2018
Report No. : SA180802C04


Calibration
      l
            Laboratory of                                           \e\‘&'qu,,
                                                                  St >                                     S       Schweizerischer Kalibrierdienst
SChm!d & Partner                                                 %                                         fa      Service suisse d‘étalonnage
  Engineering AG                                                  Pd                                               Servizio svizzero di taratura
                            s
Zeughausstrasse 43, 8004 Zurich,    &                             * —a‘e
                                 Switzerland                         //A\\o‘                               S       sgwiss Calibration Service
                                                                      Alul ult


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       Auden                                                                             Certificate No: D750V3—1132_Dec17


CALIBRATION CERTIFICATE
 Object                                D750V3 — SN:1132


 Calibration procedure(s)              QA CAL—05.v9
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     December 18, 2017


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—17 (No. 217—02521/02522)                       Apr—18
 Power sensor NRP—Z91                   SN: 103244                04—Apr—17 (No. 217—02521)                             Apr—18
 Power sensor NRP—Z91                   SN: 103245                04—Apr—17 (No. 217—02522)                             Apr—18
 Reference 20 dB Attenuator             SN: 5058 (20k)            07—Apr—17 (No. 217—02528)                             Apr—18
 Type—N mismatch combination            SN: 5047.2 / 06327        07—Apr—17 (No. 217—02529)                             Apr—18
 Reference Probe EX3DV4                 SN: 7349                  31—May—17 (No. EX3—7349_May17)                        May—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18

 Secondary Standards                     ID #                     Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
  Power sensor HP 8481A                 SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
  RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                    In house check: Oct—18
  Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—17)                    In house check: Oct—18


                                         Name                                    Function                                Signature
 Calibrated by:                         Michael Weber                            Laboratory Technician                  lk&r



  Approved by:                          Katja Pokovie                            Technical Manager                   /m


                                                                                                                        Issued: December 18, 2017
  This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


Certificate No: D750V3—1132_Dec17                                      Page 1 of 8


Calibration Laboratory of                                    is                    Schweizerischer Kalibrierdienst
      i
Schmlld    & I_’artner
                                                            s            x     S
                                                           il;‘
                                                              \gEMEf’          c   Service suisse d‘étalonnage
  Englneerlng AG                                           2 zy                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                {v,,///-_\\\\:\?   S   Swiss Calibration Service
                                                                hi us

Accredited by the Swiss Accreditation Service (SAS)                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured
Calibration is Performed According to the Following Standards:
    a)    IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
          MHz to 6 GHz)", March 2010
     d)   KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
     e) DASY4/5 System Handbook
Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
          of the certificate. All figures stated in the certificate are valid at the frequency indicated.
          Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          parallel to the body axis.
          Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
          Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
          SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
          SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D750V3—1132_Dec17                               Page 2 of 8


Measurement Conditions
   DASY system contfiguration, as far as not given onpage 1.
     DASY Version                                              DASY5                              V52.10.0
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom

     Distance Dipole Center— TSL                               15 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                          750 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           41.9               0.89 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   41.2 26 %          0.89 mho/m + 6 %

     Head TSL temperature change during test                      <0.5 °C

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     2.11 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        8.41 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                  1.37 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         5.47 Wikg * 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C            55.5              0.96 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   54.2 +6 %          0.98 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C            ~>                      ~——~


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        250 mW input power                     2.23 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         8.74 Wikg + 17.0 %(k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

     SAR measured                                        250 mW input power                  146 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         5.75 Wikg + 16.5 % (k=2)



Certificate No: D750V3—1132_Dec17                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           55.0 Q — 0.4 jQ
      Return Loss                                                                       — 26.5 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           51.3 Q —2.2 jQ

      Return Loss                                                                       —32.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.035 ns

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG

      Manufactured on                                                               October 20, 2014




Certificate No: D750V3—1132_Dec17                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                     Date: 18.12.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN: 1132

Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: £ = 750 MHz; 0 = 0.89 $/m; & =41.2; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(10.49, 10.49, 10.49); Calibrated: 31.05.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   &   Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

   &   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 59.63 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 3.31 W/kg
SAR(I g) = 2.11 W/kg; SAR(10 g) = 1.37 W/kg
Maximum value of SAR (measured) = 2.89 W/kg




       ~4.80

       —1.20

       —9.60

       —12.00
                  0 dB = 2.89 W/kg =4.61 dBW/kg




Certificate No: D750V3—1132_Dec17              Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                 18 Dec 2017    13:38:38
         CHI) sii     14 Fs           4: 54.971 a   —365.23 me 581.04 pF       750.000 000 MHz




         Del

         Ca




         Av
         169


         H1d


          CH2




         Ca




         Av
         16"

         H1d

                    START 550.000 000 MHz                             STOP 950.000 000 MHz




Certificate No: D750V3—1132_Dec17                     Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                     Date: 18.12.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN: 1132

Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: f = 750 MHz; 0 = 0.98 S/m:; & = 54.2; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C€63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(10.35, 10.35, 10.35); Calibrated: 31.05.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   &   Phantom: Flat Phantom 4.9 (Back); Type: QD OOR P49 AA; Serial: 1005

   &   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 58.19 V/m; Power Drift = —0.00 dB
Peak SAR (extrapolated) = 3.40 W/kg
SAR(I g) = 2.23 W/kg; SAR(10 g) = 1.46 W/kg
Maximum value of SAR (measured) = 2.99 W/kg


       dB
        0

       —2.40

       ~4.80

       —7.20

       —9.60

       —12.00
                  0 dB = 2.99 W/kg = 4.76 dBW/kg




Certificate No: D750V3—1132_Dec17              Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                              18 Dec 2017     15:45:21
         (CHT sii    i U rs          4: 51.260 0   —246020 98.237 pF         750.000 000 MHz
                                                   _""*~=~,

         Del

         Cor




         five
         16


         H1d

          CH2




         Cor




         Av
         159


         H1d

                    START 550.000 000 MHz                              STOP 950.000 000 MHz




Certificate No: D750V3—1132_Dec17                   Page 8 of 8


Calibration
      ‘     Laboratory of                                         a Ct
                                                                    o                                       S      Schweizerischer Kalibrierdienst
Schmid & Partner                                                 iia\\éu*%&                                 c      Service suisse d‘étalonnage
  Englneerlng AG                                                  prrol                                            Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       *///\\\\o‘                               S      gwiss Calibration Service
                                                                      rlita®s

Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       Auden                                                                                Certificate No: D835V2—40120_Jun18

|CALIBRATION CERTIFICATE
 Object                                D835V2 — SN:4d120


 Calibration procedure(s)             QA CAL—O5.v10
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    June 20, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                       Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                             Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                             Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                             Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                             Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN:    GB37480704         07—Oct—15     (in   house check Oct—16)               In   house   check:   Oct—18
 Power sensor HP 8481A                  SN:    US37202783         07—Oct—15     (in   house check Oct—16)               In   house   check:   Oct—18
 Power sensor HP 8481A                  SN:    MY41092317         07—Oct—15     (in   house check Oct—16)               In   house   check:   Oct—18
 RF generator R&S SMT—06                SN:    100972             15—Jun—15     (in   house check Oct—16)               In   house   check:   Oct—18
 Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—17)                     In house check: Oct—18


                                        Name                                      Function
 Calibrated by:                         Claudio Leubler                          Laboratory Technician



 Approved by:                           Katia Pokovic                            Technical Manager



                                                                                                                       Issued: June 21, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D835V2—4d120_Jun18                                    Page 1 of 8


Calibration Laboratory of                                                         Schweizerischer Kalibrierdienst
Schmid & Partner                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                      Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
         (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
         300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
         communication devices used in close proximity to the human body (frequency range of 30
         MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    e    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D835V2—40120_Jun18                              Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not iven on page 1.
     DASY Version                                              DASY5                              V52.10.1

     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              15 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                          835 MHz &1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  220 °C            41.5               0.90 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   40.7 + 6 %         0.93 mho/m + 6 %

     Head TSL temperature change during test                      <0.5 °C            «es+                    wes«

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

     SAR measured                                        250 mW input power                     241 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        9.37 Wikg + 17.0 %(k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition

     SAR measured                                        250 mW input power                 1.55 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        6.06 W/kg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C            55.2              0.97 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   55.0 + 6 %         0.99 mho/m + 6 %
     Body TSL temperature change during test                      «05 °C             ————                    ————


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
     SAR measured                                        250 mW input power                     2.46 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         9.68 W/kg + 17.0 % (k=2)


     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

      SAR measured                                       250 mW input power                  1.61 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W         6.36 W/kg + 16.5 %(k=2)



Certificate No: D835V2—4d120_Jun18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           51.5 Q — 3.1 jQ
      Return Loss                                                                       — 29.3 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.3 Q — 5.8 jQ

      Return Loss                                                                       — 24.3 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.398 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in orderto improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG

      Manufactured on                                                                 June 29, 2010




Certificate No: DB35V2—4d120_Jun18                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                    Date: 20.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN:4d120

Communication System: UID 0 — CW; Frequency: 835 MHz
Medium parameters used: f = 835 MHz; 0 = 0.93 S/m; & = 40.7; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(9.9, 9.9, 9.9) @ 835 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 S$n601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

        DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 62.60 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 3.71 W/kg
SAR(1 g) = 2.41 W/kg; SAR(10 g) = 1.55 W/kg
Maximum value of SAR (measured) = 3.27 W/kg




        2.20


        —4.40


        —6.60


        —8.80


        —11.00
                   0 dB = 3.27 W/ikg = 5.15 dBW/kg




Certificate No: D835V2—4d120_Jun18              Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                   20 Jun 2018    12:52:08
         CHZ sii      i U FS         4: 51.529 a   —34484 n   60.539 pF          835.000 000 MHz




         De l


         Ca




         Av
         169


         H1d


          CH2




         Ca




          Av
          16*

          H1d

                   START 635.000 000 MHz                                  STOP 1 035.000 000 MHz




Certificate No: DB35V2—4d120_Jun18                   Page 6 of 8



Document Created: 2018-10-03 15:36:02
Document Modified: 2018-10-03 15:36:02

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC