RF Exposure Report Appendix DAE4

FCC ID: WIYMP200

RF Exposure Info

Download: PDF
FCCID_2566292

Schmid & Partner Engineering AG                               s       p        e       a       g

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 44 245 9700, Fax +41 44 245 9779
info@speag.com, http://www.speag.com




                                        IMPORTANT NOTICE


USAGE OF THE DAE 4
The DAE unit is a delicate, high precision instrument and requires careful treatment by the user. There are no
serviceable parts inside the DAE. Special attention shall be given to the following points:

Battery Exchange: The battery cover of the DAE4 unit is closed using a screw, over tightening the screw may
cause the threads inside the DAE to wear out.

Shipping of the DAE: Before shipping the DAE to SPEAG for calibration, remove the batteries and pack the
DAE in an antistatic bag. This antistatic bag shall then be packed into a larger box or container which protects the
DAE from impacts during transportation. The package shall be marked to indicate that a fragile instrument is
inside.

E—Stop Failures: Touch detection may be malfunctioning due to broken magnets in the E—stop. Rough handling
of the E—stop may lead to damage of these magnets. Touch and collision errors are often caused by dust and dirt
accumulated in the E—stop. To prevent E—stop failure, the customer shall always mount the probe to the DAE
carefully and keep the DAE unit in a non—dusty environmentif not used for measurements.

Repair: Minor repairs are performed at no extra cost during the annual calibration. However, SPEAG reserves
the right to charge for any repair especially if rough unprofessional handling caused the defect.

DASY Configuration Files: Since the exact values of the DAE input resistances, as measured during the
calibration procedure of a DAE unit, are not used by the DASY software, a nominal value of 200 MOhm is given
in the corresponding configuration file.

Important Note:
Warranty and calibration is void if the DAE unit is disassembled partly or fully by the
Customer.


Important Note:
Never attempt to grease or oil the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by certified SPEAG personnel only and is part of the annual
calibration procedure.


Important Note:
To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
while turning the locking nut of the connector. The same care shall be used when
disconnecting the probe from the DAE.



Schmid & Partner Engineering


TN_BRO40315AD DAE4.doc                                                                                 11.12.2009


Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
   Engineering AG                                                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       CCS—TW (Auden)                                                                     Certificate No: DAE4—558_Jul14

[CALIBRATION CERTIFICATE                                                                                                                             |
 Object                                DAE4 — SD 000 DO4 BM — SN: 558


 Calibration procedure(s)              QA CAL—06.v26
                                       Calibration procedure for the data acquisition electronics (DAE)




 Calibration date:                     July 22, 2014




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               01—Oct—13 (No:13976)                                 Oct—14


 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Auto DAE Calibration Unit              SE UWS 053 AA 1001        07—Jan—14 (in house check)                            In house check: Jan—15
 Calibrator Box V2.1                    SE UMS 006 AA 1002        07—Jan—14 (in house check)                            In house check: Jan—15




                                        Name                                   Function                                 Signature
              y
 Calibrated by:                                y
                                        R.Mayoraz                              Technician                                  eW

 Approved by:
           y                            Fin Bomholt                             Deputy Technical Manager



                                                                                                                        Issued: July 22, 2014
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE4—558_Jul14                                        Page 1 of 5


Calibration Laboratory of                                          N.3                Schweizerischer Kalibrierdienst




                                                           "/o,/




                                                         K                   3
Schmid & Partner                                                   Az‘                Service suisse d‘étalonnage
  Engineering AG                                                   zm                 Servizio svizzero di taratura




                                                          gwl



                                                                             *h
Zeughausstrasse 43, 8004 Zurich, Switzerland                   en                     Swiss Calibration Service




                                                                            in
                                                                   Aabsidutst®

Accredited by the Swiss Accreditation Service (SAS)                                Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates



Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
   e DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
         comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    *    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
         *     DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
         *    Common mode sensitivity: Influence of a positive or negative common mode voltage on
              the differential measurement.
         e     Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
         e    AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
         *     Input Offset Measurement. Output voltage and statistical results over a large number of
               zero voltage measurements.
         *     Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.
         *     Input resistance: Typical value for information: DAE input resistance at the connector,
               during internal auto—zeroing and during measurement.
         e     Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
         *     Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE4—558_Jul14                                       Page 2 of 5


DC Voltage Measurement
    A/D — Converter Resolution nominal
        High Range:          1LSB=      6.AuV ,       full range =    ~100...+300 mV
        Low Range:           1LSB=       6inV ,       full range =   ~1.......+3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors             X                            Y                   Z
      High Range             404.844 £ 0.02% (k=2)   404.729 + 0.02% (k=2)    404.894 £ 0.02% (k=2)
      Low Range              3.96194 + 1.50% (k=2)   8.93880 £1.50% (k=2)     3.98844 £1.50% (k=2)


Connector Angle

      Connector Angle to be used in DASY system                                    227.0°#1°




Certificate No: DAE4—558_Jul14                       Page 3 of 5


Appendix (Additional assessments outside the scope of SCS108)

1. DC Voltage Linearity
      High Range                               Reading (uV)            Difference (uV)                Error (%)
      Channel X       + Input                   199998.17                       147                   0.00
      Channel X       + Input                    20005.83                      5.44                   0.03
     Channel X        — Input                   —19994.54                      6.77                   —0.03
     Channel Y        + Input                   199993.64                      —3.03                  —0.00
     Channel Y        + Input                    20004.82                      4.A7                   0.02
     Channel Y        — Input                   ~19995.59                      5.82                   —0.03
     Channel Z        + input                   199994.63                      —2.09                  —0.00
     Channel Z        + Input                    20001.00                      0.48                   0.00
     Channel Z        — Input                   ~20002.27                      —0.76                  0.00


     Low Range                                 Reading (V)             Difference (1V)                Error (%)
     Channel X        + Input                    1999.88                       —0.37                  —0.02
     Channel X        + Input                     200.42                       —0.24                  —0.12
     Channel X        — Input                    —199.48                       —0.16                  0.08
     Channel Y        + Input                    2000.46                       0.14                   0.01
     Channel Y        + Input                     200.54                       —0.12                  —0.06
     Channel Y        =— Input                   —199.21                       0.16                   —0.08
     Channel Z        + Input                    2000.16                       —0.06                  —0.00
     Channel Z2       + Input                     199.84                       —0.65                  —0.33

     Channel Z        — Input                    —200.34                       —0.98                  0.49



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                         Common mode                      High Range                               Low Range
                         Input Voltage (mV)           Average Reading (1V)                    Average Reading (LV)

     Channel X                          200                          0.76                           —0.85

                                       — 200                         1.30                           —0.50
     Channel Y                          200                          8.73                           8.42

                                       —200                         —9.88                           —9.94
    Channel Z                           200                          4.49                           4.13
                                       — 200                        —5.44                           —5.85


3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                        Input Voltage (mV)        Channel X (uV)             Channel Y (uV)       Channel Z (1V)

    Channel X                    200                       —                      5.00                 —0.39
    Channel Y                    200                   9.66                        —                   5.56
    Channel Z                    200                   6.90                       7.73                      —

Certificate No: DAE4—558_Jul14                                 Page 4 of 5


4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                    High Range (LSB)                        Low Range (LSB)

     Channel X                                              16230                                15605

     Channel Y                                              15714                                16041

     Channel Z                                              16060                                17151



5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10MQ

                            Average (1V)        min. Offset (uV)         max. Offset (uV)        Std. ?:“,’;atlon

     Channel X                   0.49                 —0.80                       1.82                   0.48
     Channel Y                   —0.21                ~1.16                       1.15                   0.46
     Channel Z                   —0.87                ~1.49                   0.15                       0.32



6. Input Offset Current
     Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                  Zeroing (kOhm)                           Measuring (MOhm)

     Channel X                                             200                                   200

     Channel Y                                             200                                   200
     Channel Z                                             200                                   200



8. Low Battery Alarm Voltage (Typical values for information)
    Typical values                                Alarm Level (VDC)
     Supply (+ Vec)                                                                +7.9

     Supply (— Vee)                                                                 —7.6



9. Power Consumption (Typical values for information)
    Typical values                          Switched off (mA)        Stand by (mA)            Transmitting (mA)
    Supply (+ Vec)                                 +0.01                     +6                        +14
    Supply (— Vee)                                 —0.01                     —8                        —9




Certificate No: DAE4—558_Jul14                             Page 5 of 5