User Manual

FCC ID: W9Z-ARANET4

Users Manual

Download: PDF
FCCID_4420171

                                              ARANET4

 Aranet4 Datasheet
 Measurements                          CO2 (carbon dioxide), temperature, relative humidity, atmospheric pressure

 Line of Sight Range                   Aranet radio up to 3 km/ 1.9mi
                                       BLE up to 10m/33ft

 Operating environment                 Indoor use

 Transmitter power                     14 dBm

 Frequency                             Depends on base station instructions

 Measurement range                     CO2                           0-9999 ppm
                                       temperature                   0°C to 60°C (14°F to 140°F)*
                                       relative humidity             0% to 85% RH
                                       atmospheric pressure          300-1100 hPa

 Measurement accuracy:**               CO2                           0-2000 ppm:        ±50 ppm or 3% of reading
                                                                     2001-9999 ppm:     ±10% of reading
                                       temperature                   ±0.4°C (0.72°F)
                                       relative humidity             ±3%
                                       atmospheric pressure          ±1 hPa

 CO2 measurement calibration           Automatic or manual (at 400ppm)

 Data Transmission                     1, 2, 5, 10 minutes

 Data Protection                       Data encryption

 Power options                         2 AA Alkaline batteries (Zn/Mn02)

 Battery life @20°C / 68°F             Up to 2 years

 Operating temperature                 0°C to 50°C (32°F to 122°F)

 Operating humidity                    0% to 85% non-condensing

 Dimensions                            70 x 70 x 24 mm / 2.716” x 2.76” x 0.94”

 Weight                                104g (3.7oz)

 Construction                          Polycarbonate

 Protection class                      IP20

 Marking                               CE

 Compatible base stations              Aranet PRO (from v1.3.2)

 Included                              2 AA Alkaline batteries

 Part number                           TDSPC003(EU)


* Please note the operating temperature range.
** For best accuracy, recommended operating range is 10°C to 40°C (50°F to 104°F) and 20% to 60% RH (non-condensing).
Prolonged operation beyond these ranges may result in a shift of sensor reading, with slow recovery time.


FCC Compliance statement                                                The information presented in this guide is the property of SAF
                                                                        Tehnika, JSC. No part of this document may be reproduced or
This equipment has been tested and found to comply with                 transmitted without proper permission from SAF Tehnika, JSC.
the limits for a Class B digital device, pursuant to part 15
of the FCC Rules. These limits are designed to provide                  The specifications or information contained in this document are
reasonable protection against harmful interference in a                 subject to change without notice due to continuing introduction
residential installation. This equipment generates, uses and            of design improvements. If there is any conflict between this
can radiate radio frequency energy and, if not installed and            document and compliance statements, the latter will supersede
used in accordance with the instructions, may cause harmful             this document.
interference to radio communications. However, there is
                                                                        SAF Tehnika, JSC has no liability for typing errors in this
no guarantee that interference will not occur in a particular
                                                                        document or damages of any kind that result from the use of this
installation. If this equipment does cause harmful interference
                                                                        document.
to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to             To get up to date information about accessories and their
try to correct the interference by one or more of the following         availability, please contact sales representative.
measures:
                                                                        Copyright Notice
TT Reorient or relocate the receiving antenna.                          Copyright © 2019 SAF Tehnika, JSC. All rights reserved.

TT Increase the separation between the equipment and receiver.
                                                                        Industry Canada Regulatory Statement
TT Connect the equipment into an outlet on a circuit different
   from that to which the receiver is connected.                        This device complies with Industry Canada’s licence-exempt
                                                                        RSSs. Operation is subject to the following two conditions: (1)
TT Consult the dealer or an experienced radio/TV technician for help.   This device may not cause interference; and (2) This device must
                                                                        accept any interference, including interference that may cause
                 Changes or modifications not expressly
                                                                        undesired operation of the device.
                 approved by the party responsible for
                 compliance could void the user’s authority to          Le présent appareil est conforme aux CNR d’Industrie Canada
                 operate the equipment.                                 applicables aux appareils radio exempts de licence. L’exploitation
                                                                        est autorisée aux deux conditions suivantes : (1) l’appareil ne
This device complies with part 15 of the FCC Rules. Operation is        doit pas produire de brouillage, et (2) l’utilisateur de l’appareil
subject to the following two conditions: (1) This device may not        doit accepter tout brouillage radioélectrique subi, même si le
cause harmful interference, and (2) this device must accept any         brouillage est susceptible d’en compromettre le fonctionnement.
interference received, including interference that may cause
undesired operation.


What is the Aranet4 device?                                           6. Bluetooth status. If the Bluetooth symbol is visible, then
                                                                         Bluetooth connectivity function is enabled. Refer to switch
The Aranet4 sensor is a battery powered CO2 meter with                   positions of the Aranet4 in chapter Switch positions explained.
additional measurements of temperature, relative humidity and            Make sure Bluetooth connectivity is enabled on your smart
atmospheric pressure*. The device is suitable for monitoring             device if you would like to connect to the Aranet4 device.
the CO2 level in your school, office or home environment.
                                                                      7. The Aranet radio status. If the Aranet radio symbol is visible,
Measurement data are displayed on power efficient e-ink                  then the radio connectivity with Aranet PRO base station is
screen allowing for long battery life. Additionally the device can       enabled. Refer to switch positions of the Aranet4 in chapter
provide a visual and sound notifications in case the CO2 level           Switch positions explained.
has exceeded high concentration level above 1400 ppm (level
                                                                      8. CO2 indication mode. If the leaf symbol is visible, then CO2
associated typically with complaints of drowsiness and poor air
                                                                         level indication is set to plant mode (plants require higher
quality).
                                                                         CO2 level than humans). If the leaf symbol is not visible, then
                                                                         CO2 level indication is set to human mode. Selection of the
                                                                         mode can be done using the Aranet4 app.

                                                                      9. CO2 threshold level indication.

                                                                           In case the CO2 indication human mode is selected:

                                                                             • Green represents normal level (below 1000 ppm),
Uses the nondispersive infrared (NDIR) sensor to measure CO2.
The CO2 gas in the chamber absorbs infrared light and this                   • Yellow represents average level (1000 to 1400 ppm),
absorption is measured by the sensor. The less light passes
                                                                             • Red represents high level (above 1400 ppm).
through - the higher the CO2 level. The CO2 gas absorbs only a
specific wavelength of the light, thus an optical filter is needed.        In case the plant mode is selected:

Using the Aranet4 app on your smart devices it is possible to                • Green represents normal level (above 500 ppm),
wirelessly access the current and up to one week historical
                                                                             • Yellow represents average level (500 to 300 ppm),
measurement data and adjust the configuration of the Aranet4
device. The Aranet4 app allows connecting several Aranet4                    • Red represents high level (below 300 ppm).
sensors to monitor their parameters remotely from one smart
                                                                      10. CO2 calibration mode indication. Symbol “A” is       displayed
device.
                                                                          when calibration is set to automatic. No letter is   displayed
Additionally, up to 100 Aranet4 devices can be connected to the           when calibration is set to manual mode. Refer        to switch
Aranet PRO base station. Refer to chapter Using Aranet4 device            positions of the Aranet4 in chapter Switch            positions
with the Aranet PRO base station.                                         explained.
*Atmospheric pressure data are available only on the Aranet app.
                                                                      Switch positions explained
The Aranet4 sensor screen explained

                2             3               4

         10
                                                    5                                   AUTO °C


          1
                                                                                     MANUAL °F 0     0




                                                    6
                     9             8      7

1. CO2 concentration level in ppm (parts per million)

2. Temperature of the air in Celsius or Fahrenheit. Refer to the      The Aranet4 device has switches that allow the user to adjust
   switch positions of the Aranet4 in chapter Switch positions        the preferences of device operation.
   explained.
                                                                      To access the switches, open the lid of the battery compartment.
3. Relative humidity of the air (RH%).                                While batteries remain inserted, adjust the pin position (up
                                                                      or down) using the pin tool that has been supplied with the
4. Battery level.
                                                                      Aranet4 device (or any other suitable thin tool).
5. Buzzer status. The speaker symbol is visible when buzzer
                                                                      Switches can be adjusted while batteries are removed as
   function is activated. The conditions when buzzer should
                                                                      well. Note that if you remove the batteries, the measurement
   sound is adjusted using the Aranet4 application.


history will be erased. The changed preference setting will be           Using Aranet4 device with the Aranet PRO
enabled once the batteries are reinserted.
                                                                         base station
The position of the switches has the following meaning:
                                                                         The Aranet PRO base station serves as a device for collecting,
AUTO / MANUAL - switch the CO2 calibration mode to either                storing and maintaining data from all types of Aranet sensors,
manual (default position) or automatic mode.                             including Aranet4. Each Aranet PRO base station can connect
The Aranet4 device is calibrated at the factory. However,                up to 100 sensors and stores the measurement data in its local
the user can perform CO2 calibration manually if required.               memory for up to 10 years.
To initiate the manual CO2 calibration, change the switch                Find out more about the Aranet PRO base station at
position from MANUAL to AUTO and back to MANUAL                          www.aranet.com and follow the User Guide on how to pair
(maintain maximum 1 sec between each movement of                         Aranet sensors to the Aranet PRO base station.
switch positions). Indication about started calibration
and its progress will be displayed on the screen. Do not                 To connect the Aranet4 device to the Aranet PRO base station,
interrupt the calibration process once it has started.                   initiate the sensor pairing mode on the Aranet PRO base station
During the manual calibration the Aranet4 device must be                 and initiate the pairing on the Aranet4 device by changing
exposed to fresh air (about 400 ppm of CO2) and the environment          the switch position from disable (0) to enable (radio symbol).
should be stable (not changing). Maintain a distance of at least         Alternatively, while switch position is in enable position (radio
1 meter from the device during the calibration process. In case          symbol on screen is visible), remove and reinsert the batteries.
of calibration failure, ensure that the environment fits the             Please note, that removing battery, the measurement history
requirements and repeat the process.                                     will be erased from the Aranet4 memory.

Manual calibration can be initiated using the Aranet4 app as
well.                                                                    Return and Warranty
In case of automatic calibration mode, the Aranet4 device needs
                                                                         In case of return or warranty claim, please contact your
to be exposed to a fresh air at around 400 ppm (for instance,
                                                                         sales representative. For Terms and Conditions refer to
outdoors or room with good air exchange) daily.
                                                                         www.aranet.com/support/
Use MANUAL calibration mode in case you are not certain
which mode to use.

°C / °F - switch to either Celsius or Fahrenheit degrees.

Bluetooth / 0 - enable or disable (0) the Bluetooth connectivity.

Aranet radio / 0 - enable or disable (0) the Aranet radio
connectivity. Refer to chapter Using Aranet4 device with the
Aranet PRO base station.


How to pair the Aranet4 to my smart
device
The Aranet4 device can be connected to a smart device using
the Bluetooth connection and the Aranet4 app (iPhone and
Android based). Make sure Bluetooth connectivity is enabled
on your smart device and on your Aranet4 device.

To connect your Aranet4 device to your smart device follow the
below steps:

  • Launch the Aranet4 app and on the main page (My devices)
    add a new device by clicking on the (+) symbol.

  • Select your Aranet4 device from the list.

  • When prompted, confirm the start of pairing.

  • Type in the 6 digit passcode that is shown on the display of
    your Aranet4 device.