03 FCC SAR Report-2

FCC ID: W2Z-02000004

Test Report

Download: PDF
FCCID_3547708

                                                                                                          Test report No.     : 11884511S-A
                                                                                                          Page                : 15 of 48
                                                                                                          Issued date         : September 1, 2017

                                                                                                          FCC ID              : W2Z-02000004


APPENDIX 2: SAR Measurement data
Appendix 2-1:           Evaluation procedure

The SAR evaluation was performed with the following procedure:
Step 1: Measurement of the E-field at a fixed location above the central position of flat phantom was used as a reference value
        for assessing the power drop.
Step 2: The SAR distribution at the exposed side of head or body position was measured at a distance of each device from the
        inner surface of the shell. The area covered the entire dimension of the antenna of EUT and suitable horizontal grid
        spacing of EUT. Based on these data, the area of the maximum absorption was determined by splines interpolation.
Step 3: Around this point found in the Step 2 (area scan), a volume of more than or equal to 30mm(X axis)30mm(Y
        axis)30mm(Z axis) was assessed by measuring 777 points (or more) under 3GHz and a volume of more than or
        equal to 28mm(X axis)28mm(Y axis)24mm (Z axis) was assessed by measuring 887 (ratio step method (*1))
        points (or more) for 3-6GHz frequency band.
        Any additional peaks found in the Step2 which are within 2dB of limit are repeated with this Step3 (Zoom scan).
        On the basis of this data set, the spatial peak SAR value was evaluated under the following procedure:
            (1)    The data at the surface were extrapolated, since the center of the dipoles is 1mm away from the tip of the probe
                   and the distance between the surface and the lowest measuring point is 2mm. The extrapolation was based on a
                   least square algorithm. A polynomial of the fourth order was calculated through the points in z-axes. This
                   polynomial was then used to evaluate the points between the surface and the probe tip.
            (2)    The maximum interpolated value was searched with a straightforward algorithm. Around this maximum the
                   SAR values averaged over the spatial volumes (1g or 10g) were computed by the 3D-Spline interpolation
                   algorithm. The 3D-Spline is composed of three one-dimensional splines with the “Not a knot"-condition (in x, y
                   and z-directions). The volume was integrated with the trapezoidal-algorithm. One thousand points (101010)
                   were interpolated to calculate the average.
            (3)    All neighboring volumes were evaluated until no neighboring volume with a higher average value was found.
Step 4: Re-measurement of the E-field at the same location as in Step 1 for the assessment of the power drift.
Step 5: Repeat Step 1-Step 4 with other condition or/and setup of EUT.


*1. Ratio step method parameters used; the first measurement point: “1.4mm” from the phantom surface, the initial grid separation: “1.4mm”, subsequent graded grid
    ratio: “1.4”. These parameters comply with the requirement of the KDB 865664 D01 (v01r04) and recommended by Schmid & Partner Engineering AG (DASY5
    manual).




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 16 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2:           SAR measurement data
Worst Reported SAR(1g) Plots
Plot 1-1: Wi-Fi: Front-right & touch, 11b(1Mbps), 2462 MHz->Highest reported SAR(1g) of Wi-Fi
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0, 161.0  -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
body-touch/b8,wlan,2462,dsss;front-right&touch/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.27 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.29 W/kg
Z Scan:160,5 (1x1x33): Measurement grid: dx=20mm, dy=20mm, dz=5mm; Maximum value of SAR (measured) = 1.38 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 26.60 V/m; Power Drift = -0.03 dB; Maximum value of SAR (measured) = 1.38 W/kg; Peak SAR (extrapolated) = 1.92 W/kg
SAR(1 g) = 0.845 W/kg; SAR(10 g) = 0.319 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.2(start)/22.2(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 17 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data (cont'd)
Other SAR test plots
Plot 2-1: Wi-Fi: Rear & touch, 11b(1Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b2,wlan,2462,dsss;rear&touch/
Area Scan:96x144,stp12 (9x13x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 0.0929 W/kg
Area Scan:96x144,stp12 (81x121x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 0.105 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 7.948 V/m; Power Drift = 0.10 dB; Maximum value of SAR (measured) = 0.124 W/kg; Peak SAR (extrapolated) = 0.225 W/kg
SAR(1 g) = 0.077 W/kg; SAR(10 g) = 0.029 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.1(start)/22.1(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)

Plot 2-2: Wi-Fi: Bottom & touch, 11b(1Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0         -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
body-touch/b21,wlan,2462,dsss;bottom&touch,b(1m)/
Area Scan:72x144,stp12 (7x13x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 0.0853 W/kg
Area Scan:72x144,stp12 (61x121x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 0.0869 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 6.688 V/m; Power Drift = -0.02 dB; Maximum value of SAR (measured) = 0.0877 W/kg; Peak SAR (extrapolated) = 0.120 W/kg
SAR(1 g) = 0.058 W/kg; SAR(10 g) = 0.030 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 18 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data / Other SAR test plots (cont'd)

Plot 2-3: Wi-Fi: Right-front & touch, 11b(1Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0         -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
body-touch/b6,wlan,2462,dsss;right-front(1)&touch/
Area Scan:84x84,stp12 (8x8x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.21 W/kg
Area Scan:84x84,stp12 (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.22 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 25.20 V/m; Power Drift = -0.04 dB; Maximum value of SAR (measured) = 1.23 W/kg; Peak SAR (extrapolated) = 1.69 W/kg
SAR(1 g) = 0.775 W/kg; SAR(10 g) = 0.306 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.1(start)/22.1(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)

Plot 2-4: Wi-Fi: Front-lens & touch, 11b(1Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0         -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
body-touch/b5,wlan,2462,dsss;front-lens&touch/
Area Scan:108x144,stp12 (10x13x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 0.978 W/kg
Area Scan:108x144,stp12 (91x121x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 0.997 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 22.24 V/m; Power Drift = -0.02 dB; Maximum value of SAR (measured) = 0.962 W/kg; Peak SAR (extrapolated) = 1.29 W/kg
SAR(1 g) = 0.616 W/kg; SAR(10 g) = 0.254 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.1(start)/22.1(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 19 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data / Other SAR test plots (cont'd)

Plot 2-5: Wi-Fi: Front & touch, 11b(1Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0         -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
body-touch/b9,wlan,2462,dsss;front&touch/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.26 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.26 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 25.06 V/m; Power Drift = -0.01 dB; Maximum value of SAR (measured) = 1.23 W/kg; Peak SAR (extrapolated) = 1.72 W/kg
SAR(1 g) = 0.763 W/kg; SAR(10 g) = 0.290 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.2(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)

Plot 2-6: Wi-Fi: Front-right & touch, 11b(1Mbps), 2412 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2412 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2412 MHz; σ = 1.966 S/m; εr = 50.86; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b12,wlan,ch;2412,dsss;front-right&touch,b(1m)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.16 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.17 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 25.15 V/m; Power Drift = -0.06 dB; Maximum value of SAR (measured) = 1.19 W/kg; Peak SAR (extrapolated) = 1.63 W/kg
SAR(1 g) = 0.721 W/kg; SAR(10 g) = 0.277 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 20 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data / Other SAR test plots (cont'd)

Plot 2-7: Wi-Fi: Front-right & touch, 11b(1Mbps), 2437 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2437 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2437 MHz; σ = 2.002 S/m; εr = 50.79; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b11,wlan,ch;2437,dsss;front-right&touch,b(1m)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.21 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.22 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 25.50 V/m; Power Drift = 0.20 dB; Maximum value of SAR (measured) = 1.33 W/kg; Peak SAR (extrapolated) = 1.83 W/kg
SAR(1 g) = 0.814 W/kg; SAR(10 g) = 0.309 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)

Plot 2-8: Wi-Fi: Front-right & touch, 11g(6Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11g(6Mbps, BPSK/OFDM)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b13,wlan,ofdm1;2462;front-right&touch,g(6m)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.02 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.03 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 23.13 V/m; Power Drift = -0.00 dB; Maximum value of SAR (measured) = 1.05 W/kg; Peak SAR (extrapolated) = 1.47 W/kg
SAR(1 g) = 0.641 W/kg; SAR(10 g) = 0.242 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 21 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data / Other SAR test plots (cont'd)

Plot 2-9: Wi-Fi: Front-right & touch, 11g(6Mbps), 2412 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11g(6Mbps, BPSK/OFDM)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2412 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2412 MHz; σ = 1.966 S/m; εr = 50.86; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b15,wlan,ofdm1;2412(>0.8);front-right&touch,g(6m)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 0.887 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 0.892 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 21.99 V/m; Power Drift = -0.02 dB; Maximum value of SAR (measured) = 0.913 W/kg; Peak SAR (extrapolated) = 1.25 W/kg
SAR(1 g) = 0.554 W/kg; SAR(10 g) = 0.212 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)

Plot 2-10: Wi-Fi: Front-right & touch, 11g(6Mbps), 2437 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11g(6Mbps, BPSK/OFDM)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2437 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2437 MHz; σ = 2.002 S/m; εr = 50.79; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b17,wlan,ofdm1;2437(>0.8);front-right&touch,g(6m)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 0.931 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 0.938 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 22.29 V/m; Power Drift = -0.05 dB; Maximum value of SAR (measured) = 0.955 W/kg; Peak SAR (extrapolated) = 1.32 W/kg
SAR(1 g) = 0.580 W/kg; SAR(10 g) = 0.221 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 22 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data / Other SAR test plots (cont'd)

Plot 2-11: Wi-Fi: Front-right & touch, 11n(20HT)(MCS0), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: n20(MCS0, BPSK/OFDM)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b14,wlan,ofdm2;2462;front-right&touch,n20(m0)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.03 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.04 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 23.24 V/m; Power Drift = 0.00 dB; Maximum value of SAR (measured) = 1.06 W/kg; Peak SAR (extrapolated) = 1.48 W/kg
SAR(1 g) = 0.648 W/kg; SAR(10 g) = 0.244 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)

Plot 2-12: Wi-Fi: Front-right & touch, 11n(20HT)(MCS0), 2412 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: n20(MCS0, BPSK/OFDM)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2412 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2412 MHz; σ = 1.966 S/m; εr = 50.86; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b16,wlan,ofdm2;2412(>0.8);front-right&touch,n20(m0)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 0.909 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 0.920 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 22.29 V/m; Power Drift = 0.01 dB; Maximum value of SAR (measured) = 0.940 W/kg; Peak SAR (extrapolated) = 1.29 W/kg
SAR(1 g) = 0.579 W/kg; SAR(10 g) = 0.223 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 23 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data / Other SAR test plots (cont'd)

Plot 2-13: Wi-Fi: Front-right & touch, 11n(20HT)(MCS0), 2437 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: n20(MCS0, BPSK/OFDM)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2437 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2437 MHz; σ = 2.002 S/m; εr = 50.79; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
-Sensor-Surface: 2mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0
-Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section                    -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
body-touch/b18,wlan,ofdm2,2437(>0.8);front-right&touch,n20(m0)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 0.954 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 0.967 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 22.61 V/m; Power Drift = 0.20 dB; Maximum value of SAR (measured) = 0.988 W/kg; Peak SAR (extrapolated) = 1.39 W/kg
SAR(1 g) = 0.623 W/kg; SAR(10 g) = 0.239 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.    : 11884511S-A
                                                                                                        Page               : 24 of 48
                                                                                                        Issued date        : September 1, 2017

                                                                                                        FCC ID             : W2Z-02000004

Appendix 2-2: SAR measurement data (cont'd)
SAR test plot for the "SAR Measurement Variability"
Plot 3-1: (Repeat) Wi-Fi: Front-right & touch, 11b(1Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0, 161.0  -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
body-touch/b10(rep#8),wlan,2462,dsss;front-right&touch,b(1m)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.34 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.36 W/kg
Z Scan:160,5 (1x1x33): Measurement grid: dx=20mm, dy=20mm, dz=5mm; Maximum value of SAR (measured) = 1.38 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 26.49 V/m; Power Drift = 0.01 dB; Maximum value of SAR (measured) = 1.38 W/kg; Peak SAR (extrapolated) = 1.92 W/kg
SAR(1 g) = 0.844 W/kg; SAR(10 g) = 0.319 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)

SAR test plot for the " Device holder perturbation verification "
Plot 3-2: (No device-holder) Wi-Fi: Front-right & touch, 11b(1Mbps), 2462 MHz
EUT: Digital camera; Type: FF170001; Serial: P1000006 (with Wireless LAN module: FZ00010935-800 / Bluetooth module: M-BN-BM-26_A)
Mode: 11b(1Mbps, DBPSK/DSSS)(UID 0, Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2462 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2462 MHz; σ = 2.035 S/m; εr = 50.64; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0, 161.0  -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
body-touch/b24(rep#8),no-device-holde,rwlan,2462,dsss;front-right&touch,b(1m)/
Area Scan:72x72,stp12 (7x7x1): Measurement grid: dx=12mm, dy=12mm; Maximum value of SAR (measured) = 1.11 W/kg
Area Scan:72x72,stp12 (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm; Maximum value of SAR (interpolated) = 1.15 W/kg
Z Scan:160,5 (1x1x33): Measurement grid: dx=20mm, dy=20mm, dz=5mm; Maximum value of SAR (measured) = 1.30 W/kg
Zoom Scan:30x30x30,stp5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 25.67 V/m; Power Drift = 0.03 dB; Maximum value of SAR (measured) = 1.30 W/kg; Peak SAR (extrapolated) = 1.82 W/kg
SAR(1 g) = 0.805 W/kg; SAR(10 g) = 0.309 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of EUT to phantom: 0 mm (2 mm to liquid); ambient: (22.5~24.0) deg.C. / (45~55) %RH,
         *. liquid temperature: 22.3(start)/22.3(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)



UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                          Test report No.        : 11884511S-A
                                                                                                          Page                   : 25 of 48
                                                                                                          Issued date            : September 1, 2017

                                                                                                          FCC ID                 : W2Z-02000004


APPENDIX 3: Test instruments
Appendix 3-1:           Equipment used




       *.   AT (antenna terminal conducted power measurement) was measured July 27, 2017. (Refer to Section 6 in this report.)




The expiration date of calibration is the end of the expired month.
As for some calibrations performed after the tested dates, those test equipment have been controlled by means of an unbroken chains of calibrations.
All equipment is calibrated with valid calibrations. Each measurement data is traceable to the national or international standards.
[Test Item] SAR: Specific Absorption Rate, AT: Antenna terminal conducted power




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                   Test report No.    : 11884511S-A
                                                                                                   Page               : 26 of 48
                                                                                                   Issued date        : September 1, 2017

                                                                                                   FCC ID             : W2Z-02000004

Appendix 3-2:          Configuration and peripherals

These measurements were performed with the automated near-field scanning system DASY5 from Schmid & Partner Engineering
AG (SPEAG). The system is based on a high precision robot), which positions the probes with a positional repeatability of better
than  0.02 mm. Special E- and H-field probes have been developed for measurements close to material discontinuity, the sensors of
which are directly loaded with a Schottky diode and connected via highly resistive lines to the data acquisition unit. The SAR
measurements were conducted with the dosimetry probes EX3DV4 (manufactured by SPEAG), designed in the classical triangular
configuration and optimized for dosimetric evaluation.

                                                                                             Shielded room




                                                                                                                                    Flat phantom

                                                                                                                                    Tissue
                                                                                                                                    simulating
                                                                                                                                    liquid
                                                                                                                                    Device holder




The DASY5 system for performing compliance tests consist of the following items:
     A standard high precision 6-axis robot (Stäubli TX/RX family) with controller, teach pendant and software.
 1
     An arm extension for accommodating the data acquisition electronics (DAE).
 2   An isotropic field probe optimized and calibrated for the targeted measurement.
     A data acquisition electronics (DAE) which performs the signal amplification, signal multiplexing, AD-conversion, offset measurements,
 3   mechanical surface detection, collision detection, etc. The unit is battery powered with standard or rechargeable batteries. The signal is optically
     transmitted to the EOC.
     The Electro-optical converter (EOC) performs the conversion from optical to electrical signals for the digital communication to the DAE. To use
 4
     optical surface detection, a special version of the EOC is required. The EOC signal is transmitted to the measurement server.
     The function of the measurement server is to perform the time critical tasks such as signal filtering, control of the robot operation and fast
 5
     movement interrupts.
 6   The Light Beam used is for probe alignment. This improves the (absolute) accuracy of the probe positioning.
 7   A computer running Win7 professional operating system and the DASY5 software.
 8   R Remote control and teach pendant as well as additional circuitry for robot safety such as warning lamps, etc.
 9   The phantom.
10   The device holder for EUT. (low-loss dielectric palette) (*. when it was used.)
11   Tissue simulating liquid mixed according to the given recipes.
12   Validation dipole kits allowing to validate the proper functioning of the system.




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                        Test report No.   : 11884511S-A
                                                                                                        Page              : 27 of 48
                                                                                                        Issued date       : September 1, 2017

                                                                                                        FCC ID            : W2Z-02000004

Appendix 3-3:          Test system specification

TX60 Lspeag robot/CS8Cspeag-TX60 robot controller
Number of Axes    : 6                         Repeatability                       : 0.02mm                                                EOC
Manufacture       : Stäubli Unimation Corp.
DASY5 Measurement server
Features        : The DASY5 measurement server is based on a PC/104 CPU board with a                                     DAE
                    400MHz intel ULV Celeron, 128MB chip-disk and 128MB RAM. The
                    necessary circuits for communication with the DAE4 electronics box, as well                                           TX60L
                    as the 16 bit AD converter system for optical detection and digital I/O
                    interface are contained on the DASY5 I/O board, which is directly connected
                    to the PC/104 bus of the CPU board.                                                                   Probe
Calibration     : No calibration required.
Manufacture     : Schmid & Partner Engineering AG
Data Acquisition Electronic (DAE)
Features             : Signal amplifier, multiplexer, A/D converter and control logic.
                        Serial optical link for communication with DASY5 embedded system (fully                             Light beam switch
                        remote controlled). 2 step probe touch detector for mechanical surface
                        detection and emergency robot stop (not in -R version)
Measurement Range : 1µV to > 200mV (16bit resolution and 2 range settings: 4mV, 400mV)
Input Offset voltage : < 1µV (with auto zero)
Input Resistance     : 200M
Battery Power        : > 10hr of operation (with two 9V battery)                                                               DASY5 Server
Manufacture          : Schmid & Partner Engineering AG
Electro-Optical Converter (EOC61)
Manufacture         : Schmid & Partner Engineering AG
Light Beam Switch (LB5/80)
Manufacture        : Schmid & Partner Engineering AG
                                                                                                                                     Robot controller
SAR measurement software
Item              : Dosimetric Assessment System DASY5
Software version  : DASY52, V8.2 B969
Manufacture       : Schmid & Partner Engineering AG
E-Field Probe
Model                : EX3DV4 (serial number: 3907)                                                                            EX3DV4 E-field Probe
Construction         : Symmetrical design with triangular core.
                        Built-in shielding against static charges.
                        PEEK enclosure material (resistant to organic solvents, e.g., DGBE).
Frequency            : 10MHz to 6GHz, Linearity: 0.2 dB (30MHz to 6GHz)
Conversion Factors : 2.45, 5.2, 5.25, 5.5, 5.6, 5.75, 5.8 GHz (Head)
                      : 2.45, 5.25, 5.6, 5.75 GHz (Body)
Directivity          : 0.3 dB in HSL (rotation around probe axis)
                        0.5 dB in tissue material (rotation normal to probe axis)
Dynamic Range        : 10W/g to > 100 mW/g; Linearity: 0.2 dB (noise: typically < 1W/g)
Dimension        : Overall length: 330mm (Tip: 20mm)
                    Tip diameter: 2.5mm (Body: 12mm)
                    Typical distance from probe tip to dipole centers: 1mm
Application       : High precision dosimetric measurement in any exposure scenario (e.g., very strong gradient
                     fields). Only probe which enables compliance testing for frequencies up to 6GHz with precision
                     of better 30%.
Manufacture       : Schmid & Partner Engineering AG
Phantom
Type              :   ELI 4.0 oval flat phantom
Shell Material    :   Fiberglass               Shell Thickness : Bottom plate: 2 0.2mm
Dimensions        :   Bottom elliptical: 600400mm, Depth: 190mm (Volume: Approx. 30 liters)                                       ELI 4.0 flat phantom
Manufacture       :   Schmid & Partner Engineering AG
Device Holder
 Urethane foam
 KSDH-01: In combination with the ELI4, the Mounting Device enables the rotation of the
   mounted transmitter device in spherical coordinates. Transmitter devices can be easily and
   accurately positioned. The low-loss dielectric urethane foam was used for the mounting section of
   device holder.
   Material : POM               Manufacture : Schmid & Partner Engineering AG
                                                                                                                                          Device holder


UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                           Test report No.     : 11884511S-A
                                                                                                           Page                : 28 of 48
                                                                                                           Issued date         : September 1, 2017

                                                                                                           FCC ID              : W2Z-02000004

Appendix 3-4:            Simulated tissue composition and parameter confirmation
       Liquid type                                              Body
       Control No.                                           KSLM245-01
  Model No. / Product No.                           MSL2450V2 / SL AAM 245 BA
  Ingredient: Mixture(%)                      Water: 52-75%, DGBE: 25-48%, NaCl: <1.0%
       Manufacture                                Schmid & Partner Engineering AG

*. The dielectric parameters were checked prior to assessment using the DAK3.5 dielectric probe kit.
                                                                                            Liquid parameters (*a)                           ΔSAR
      Measured           Freq. Liquid Ambient Liquid Liquid               Permittivity (r) [-]                Conductivity [S/m]
        date
                                       [deg.C.] temp. Depth                                                                               (1g)    (10g)
                        [MHz] type /[%RH] [deg.C.] [mm]                      Measured                              Measured
                                                                   Target                       Limit Target                      Limit [%] (*b) [%] (*b)
                                                                            Meas. Δr[%]                         Meas. Δ[%]
    August 4, 2017       2450 Body 23.9/42           22.2    (152) 52.7 50.75          -3.7 5% 1.95 2.012 +3.2 5% +2.35 +1.41
*a. The target value is a parameter defined in Appendix A of KDB865664 D01 (v01r04), the dielectric parameters are given at 2000, 2450 and 3000MHz. Parameters
    for the frequencies between 2000-3000MHz were obtained using linear interpolation.
                                                             Standard
           f       Head Tissue Body Tissue         f      Head Tissue Body Tissue        f     Head Tissue Body Tissue
        (MHz)     r [-]  [S/m] r [-]  [S/m] (MHz) r [-]  [S/m] r [-]  [S/m] (MHz) r [-]  [S/m] r [-]  [S/m]
     (1800-)2000 40.0 1.40 53.3 1.52 2450 39.2 1.80                   52.7 1.95 3000 38.5 2.40 52.0 2.73
*b. The coefficients are parameters defined in IEEE Std. 1528-2013.
    ΔSAR(1g)= Cεr Δεr + C Δ, Cεr=-7.854E-4f3+9.402E-3f2-2.742E-2f-0.2026 / C =9.804E-3f3-8.661E-2f2+2.981E-2f+0.7829
    ΔSAR(10g)= Cεr Δεr + C Δ, Cεr=3.45610-3f3-3.53110-2f2+7.67510-2f-0.1860 / C =4.47910-3f3-1.58610-2f2-0.1972f+0.7717

Appendix 3-5: Daily check results
Prior to the SAR assessment of EUT, the Daily check was performed to test whether the SAR system was operating within its
target of 10%. The Daily check results are in the table below.
                                                                       Daily check results
                                                                                   Daily check target & measured
                      Freq. Liquid
                                                        SAR (1g) [W/kg] (*d)                                             SAR (10g) [W/kg] (*d)
       Date                                              Target    Deviation                                   Target                       Deviation
                     [MHz] Type Meas. ΔSAR- 1W                               Limit Pass Meas. ΔSAR- 1W                                                   Limit Pass
                                   (*c) correct scaled Cal. STD Cal. STD [%]        ?    (*c) correct scaled Cal. STD                     Cal. STD [%]          ?
                                                       (*e)  (*f) [%]    [%]                                 (*e)  (*f)                    [%]     [%]
  August 4, 2017      2450 Body 13 12.69 50.76 49.6 n/a +2.3 n/a 10 Pass 6.07 5.98 23.92 23.3 n/a                                         +2.7 n/a 10 Pass
*. Calculating formula: SAR corrected SAR (1g,10g) (W/kg) = (Measured SAR(1g,10g) (W/kg))  (100 - (SAR(%)) / 100
*c. The "Measured" SAR value is obtained at 250 mW for 2450MHz, and at 100 mW for 5GHz band.
*d. The measured SAR value of Daily check was compensated for tissue dielectric deviations (SAR) and scaled to 1W of output power in order to compare with the
    manufacture's calibration target value which was normalized.
*e. The target value is a parameter defined in the calibration data sheet of D2450V2 (sn:822) dipole calibrated by Schmid & Partner Engineering AG (Certification No.
    D2450V2-822_Jan17, the data sheet was filed in this report).
*f. The target value (normalized to 1W) is defined in IEEE Std.1528.

                                                                                                                                          Liquid depth: 152 mm
                                                                                                                                            (August 4, 2017)




                                                                                                  Dipole
                                                                                                (2.45GHz)




                                                Test setup for the system performance check




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                                           Test report No.      : 11884511S-A
                                                                                                                           Page                 : 29 of 48
                                                                                                                           Issued date          : September 1, 2017

                                                                                                                           FCC ID               : W2Z-02000004

Appendix 3-6:               Daily check measurement data

EUT: Dipole(2.45GHz)(sn822); Type: D2450V2; Serial: 822; Forward conducted power: 250mW
Communication System: CW (*. UID:0; Frame Length in ms: 0; PAR: 0; PMF: 1); Frequency: 2450 MHz; Crest Factor: 1.0
Medium: M2450(1708); Medium parameters used: f = 2450 MHz; σ = 2.012 S/m; εr = 50.75; ρ = 1000 kg/m3
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration: -Probe: EX3DV4 - SN3907; ConvF(7.38, 7.38, 7.38); Calibrated: 2017/02/27; -DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)
                        -Sensor-Surface: 2mm (Mechanical Surface Detection), z = 1.0, 31.0, 161.0  -Electronics: DAE4 Sn626; Calibrated: 2016/10/13
                        -Phantom: ELI v4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
Area Scan:60x60,stp15 (5x5x1): Measurement grid: dx=15mm, dy=15mm; Maximum value of SAR (measured) = 19.9 W/kg
Area Scan:60x60,stp15 (41x41x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm; Maximum value of SAR (interpolated) = 19.9 W/kg
Z Scan;160,stp5(mm) (1x1x33): Measurement grid: dx=20mm, dy=20mm, dz=5mm; Maximum value of SAR (measured) = 19.8 W/kg
Zoom Scan:30x30x30,stp5,5,5 (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm;
Reference Value = 100.7 V/m; Power Drift = 0.06 dB; Maximum value of SAR (measured) = 19.8 W/kg; Peak SAR (extrapolated) = 26.3 W/kg
SAR(1 g) = 13 W/kg; SAR(10 g) = 6.07 W/kg




Remarks: *. Date tested: 2017/08/04; Tested by: Hiroshi Naka; Tested place:No.7 shielded room,
         *. liquid depth: 152 mm; Position: distance of dipole to phantom: 8mm (10mm to liquid); ambient: 23.2 deg.C. / 51 %RH,
         *. liquid temperature: 22.2(start)/22.2(end)/22.2(in check) deg.C.; *.White cubic: zoom scan area, Red cubic: big=SAR(10g )/small=SAR(1g)


Appendix 3-7:               Daily check uncertainty

                 Uncertainty of daily check (2.4~6GHz) (*.& tolerance:  5%, DAK3.5, CW) (v08)                                          1g SAR               10g SAR
                       Combined measurement uncertainty of the measurement system (k=1)                                                    ± 11.0 %             ± 10.9 %
                                        Expanded uncertainty (k=2)                                                                         ± 22.1 %             ± 21.8 %
                                                                     Uncertainty      Probability
                      Error Description (v08)                          Value          distribution
                                                                                                       Divisor     ci (1g) ci (10g)          ui (1g)              ui (10g)          Vi, veff
  A    Measurement System (DASY5)                                                                                                        (std. uncertainty)    (std. uncertainty)
  1    Probe Calibration Error                                         ±6.55 %          Normal            1          1          1            ±6.55 %              ±6.55 %              ∞
  2    Axial isotropy error                                            ±4.7 %          Rectangular        √3        √0.5       √0.5          ±1.9 %               ±1.9 %               ∞
  3    Hemispherical isotropy error                                    ±9.6 %          Rectangular        √3         0          0              0%                   0%                 ∞
  4    Probe linearity                                                 ±4.7 %          Rectangular        √3         1          1            ±2.7 %               ±2.7 %               ∞
  5    Probe modulation response (CW)                                  ±0.0 %          Rectangular        √3         1          1              0%                   0%                 ∞
  6    System detection limit                                          ±1.0 %          Rectangular        √3         1          1            ±0.6 %               ±0.6 %               ∞
  7    Boundary effects                                                ±4.8 %          Rectangular        √3         1          1            ±2.8 %               ±2.8 %               ∞
  8    System readout electronics (DAE)                                ±0.3 %           Normal            1          1          1            ±0.3 %               ±0.3 %               ∞
  9    Response Time Error (<5ms/100ms wait)                           ±0.0 %          Rectangular        √3         1          1              0%                   0%                 ∞
 10    Integration Time Error (CW)                                     ±0.0 %          Rectangular        √3         1          1              0%                   0%                 ∞
 11    RF ambient conditions-noise                                     ±3.0 %          Rectangular        √3         1          1            ±1.7 %               ±1.7 %               ∞
 12    RF ambient conditions-reflections                               ±3.0 %          Rectangular        √3         1          1            ±1.7 %               ±1.7 %               ∞
 13    Probe positioner mechanical tolerance                           ±3.3 %          Rectangular        √3         1          1            ±1.9 %               ±1.9 %               ∞
 14    Probe positioning with respect to phantom shell                 ±6.7 %          Rectangular        √3         1          1            ±3.9 %               ±3.9 %               ∞
 15    Max. SAR evaluation (Post-processing)                           ±4.0 %          Rectangular        √3         1          1            ±2.3 %               ±2.3 %               ∞
  B    Test Sample Related
 16    Deviation of the experimental source                             ±3.5 %          Normal            1           1          1           ±3.5 %               ±3.5 %               ∞
 17    Dipole to liquid distance (10mm±0.2mm,<2deg.)                    ±2.0 %         Rectangular        √3          1          1           ±1.2 %               ±1.2 %               ∞
 18    Drift of output power (measured, <0.2dB)                         ±2.3 %         Rectangular        √3          1          1           ±1.3 %               ±1.3 %               ∞
  C    Phantom and Setup
 19    Phantom uncertainty                                              ±2.0 %         Rectangular        √3          1          1           ±1.2 %                ±1.2%               ∞
 20    Algorithm for correcting SAR (e',σ: 5%)                         ±1.2 %          Normal             1          1        0.84          ±1.2 %               ±0.97 %              ∞
 21    Liquid conductivity (meas.) (DAK3.5)                             ±3.0 %          Normal             1        0.78       0.71          ±2.3 %               ±2.1 %               ∞
 22    Liquid permittivity (meas.) (DAK3.5)                             ±3.1 %          Normal             1        0.23       0.26          ±0.7 %               ±0.8 %               ∞
 23    Liquid Conductivity-temp.uncertainty (2deg.C.)                  ±5.3 %         Rectangular        √3        0.78       0.71          ±2.4 %               ±2.2 %               ∞
 24    Liquid Permittivity-temp.uncertainty (2deg.C.)                  ±0.9 %         Rectangular        √3        0.23       0.26          ±0.1 %               ±0.1 %               ∞
       Combined Standard Uncertainty                                                                                                         ±11.0 %              ±10.9 %
       Expanded Uncertainty (k=2)                                                                                                            ±22.1 %              ±21.8 %
*.    This measurement uncertainty budget is suggested by IEEE Std. 1528(2013) and determined by Schmid & Partner Engineering AG (DASY5 Uncertainty Budget).




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


Galibration Laboratory of                                                       . Schwoiischor Katprnsionst




                                                ,
Schmid & Partner                                                                g. Servee suisse éalonnage
  Engineering AG                                                                   Sonvite svicar o tarsura
                                                                                8 Suiss Caltraton Service




                                              E1
Zecanivsstrase d2004 zuic eitedand                   oys
                                                     hn
Accrted by e Snss Acerectafon Sevce ($A9)                                       sestesaion No.: SCS 0108
‘The Suiss Aceredtation Sarvice is oneofthesigntarntothEA
Mottitera Agrecnanfo tharecogntn o catbraton coriictes
ciex        U Japan (Vites)                            '               Cariteat No: EX3—3007_Feb17             C

|CALIBRATION CERTIFICATE
onea                           Exsovs —Snago7
Cattratonprecesint)         | OA CAL—018, OACAL—14.v4, OA CAL—23.u5, OA CAL—25.v6
                               Calibration procedure for dosimetric E—feld probes

Cattraton ce
Thiscaltratoceriical docnentsthe raceity tonolonl ind‘ uhh reate thoians o mensements 0
"Te masssremants and ho unceinies wconidence esare sive on the ftouna pagesand arpar ofthe confea.
   cattrston have been conduciedinthe dosed snatay hoi envronment empeate(2 3]°G and humidty <70%
Calatin Eqipmentusnd (MBTE citel o catratn}

 Emmemes                    o                         Emnere=s                       Sceiued Gateain
 Pore mewchsp               Extan                     seveet6 io 2rreemenzen         reur
 Pove serse N:201                                     C6z0u18 io. zircea)            Aoear
 Pove sersor N291           se mems                   ce018 io. zirezz)              Apeit
 Relonce 20 on atensmee     sc sears @00              sevects ho. rrreens            rour
 Reornce rove essove        att sors                  sromts tcssolsSe®)             beeir
 brea                       steaso                    ose1s (ie bretst1Deetp         Decar
 Sesondan Sndas             ©                         Check Date in use)             Screduled hok
 Pover moucEserg8           en esmur                  o820018 in house cediB         n rause ctesc Junts
 Poversersoration           Enmamen                   5620e1G in hovse cesin_        n reuse cteac ans
 Poverconsor Extiat         o ooortezro               96016 in hoowe chie0)          Inrowechede u18
 h goneutor > ssitc         sn usssrauorico           0+94209 in hoise ceck ini0)    Inrousectesk Jint
 1etore weaize n arsse      onussrsoosts              180901 (ntoe sns a—15)         inrowse ec.ootr
                              m—                          Furdin                       Spave
Catbrat br:                    "ason Kamtal     ~~          Tabortoy Tecicin Q? _                (/Q/
sovord br:

                                                                                       Iesuess Pxbruoy 22017
"hiettralon caifsts stal nc e rproduend esn l uitout witon aporonle h ty


Corifcate No: EX3—3007_Fobt7                         Page tor 11




 an, Inc.


                                                                              1                   ras
                                                                                                         on
                                                                                                         s



Galibration Laboratory of                                                         i Setwotnscheratbrirtonst
Schmid & Partner                                                                  .   Seviensulesadlamnage
  Engineering AG                                                                  § senide mimre a ormum
eaghacnstrasse 9, 6004zarich, Suitenrond                                              Sniss Calbraton Serice

Aeoredtd by e Suiss ccrestaton Sevice GA8)                                        Aecredtaion No: 5CS 0108
"TSutss Accredtation Serviceis oneofthesigntarie to hEA
MoltiteratAgrccmanfoh reeogniton o cxtiraton conitents
Glossary:
Tst                     tssue simulating iquid
NORMicyz                senstivty in free space
Come                    sensituty n TSL /NORMxy.z
bor                     diode compression point
or                      erestfactr(1ily_cycle) of e RE signal
As.c.0                  modulation dependontincarization parametors
Pobrization o           o rotation around probe axis
Polrization 0           8 reteion around an ax‘s that i in the plane normalto probe axis(at measurement center,
                        te, 8 = 0is nomalto probe axis
Connector Angle         Information used in DASY system to align probe sensor X o trobot coordinate system
Calibration is Performed According to the Following Standards:
   a). IEEE Std 1828—2013, "EEE Recommended Practicn for Datermining the Peak Spatia—Averaged Specifo
       Absorption Rate (SAR)in the Human Head from Wereless Communications Devices: Measurement
       Techoiques®,.line 2012
   b)_ 12G 62209—1, "Procedure to measure the Specifc Absorpton Rete (SAR) for hand—held devices used in close
       provimity to the ear (Requency range of 300 MFizt 3 GHz, February 2005
   ): IEC 62209—2, Procedure to determine the Specifc Absorption Rate (SAR)for wireless communication devices
       used in close proximityto the human body (frequency range of 30 Miiz to 6 GHz]¢, March 2010
    )   KOB 865084, ‘SAR Measurement Requirements for 100 Mz to 6 GHz"

Methods Applied and Interpretation of Parameters:
   + NORkky.a: Assessed for E—feld polatzation 8 =0 (f= 900 Mz in TE—call(> 1800 MHz: R2? waveguide).
     NORMxy;z are only ntormodiate valucs, Lo., the uncentaintos of NORVxy,z does notaffect the Eold
     uncertainty insde  TSL (seo below Conv£).
   + NORM(Mayiz NORMxy;z * frequency.respanse (see Frequency Respanse Char}.This incarizaion is
     implemented in DASY4 software versions late than 4.2.The uncertaity ofthe frequency responseis ncluded
     in the stated uncortainy of ConvF.
   + DOPxy:a: DCP are numerical Inearization parameters assessed based on the data of power sweep wih CW
     signal(ho uncertainty required). DCP does not depend on frequency nor meia.
   + PAR: PAR is the Peak to Average Ratio thatis not calbrated but determined based on the signal
     charactoritics
   + Auyic BCopar Doy.e: VRky2: A, B, C, D are numericalInearization paremeters assessed based on
     the data of power sweep for specifo modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum caltralon range expressed in RVS voliage accoss thciode.
   + ConvF and Boundary Effect Parameters: Assossed in flatphantom using E—feld (orTemporature Transter
     Standard forf 800 MHz) and inside waveguide using analyical feld dstibutions based on power
     measurementsfor1> 800 NMHz. The same setups are used for assessment of the parameters appled for
     boundary compensation(alha, depth)of which ypical uncertaity values are given. These perameters are
     used in DAGY4 software to improve probe accuracy close tothe boundary. The sensitviy n TSL coesponds
     to NORMy;z * ConvE whereby the uncertaity corresponds to that gven for Conv. Afrequency dependent
     ConvF is used in DASY version 4.¢ and higher which alous extendingthe vaiity from + 50 MHz to : 100
     Miiz
   * Spherica!Isotropy (30 deviationfrom sotropy): i a feld of ow gredients realzed using a fat phantom
     exposed by a palch antenna.
   + Sensor Offset: The sensr offset correspondsto the offset ovruzal measurement center rom the probe tp
     (on probe axis. No tolerance required.
   * Connsctor Angle: The angle is assessed using the ifrmation galned by determining the NORA (no
     uncertainty required).

Certfeate No: EX3—3007.Feot?                       Page 20011


                                                                                       isa
                                                                                          "an7




  exoove — awasor                                                              Fobrmry27, 2017




                       Probe EX3DV4

                                           SN:3907

                                   Manufactured:           September 4, 2012
                                   Repaired:               February 15, 2017
                                   Calibrated:             February 27, 2017




                                 Calibrated for DASY/EASY Systems
                                     (Note: non—compatble with DASY2 systam)




  Coriieate No: Ex3.3007_Fobtr                   Page sot 11




_ . an,Inc.

                                 .                   ran
                             simile: 181 463 50 6401


                                                                                                    isa
                                                                                              ras
                                                                                                       w17




exsovs— snacor                                                                          remmary2r, 2017


DASY/EASY — Parameters of Probe: EX3DV4 — SN:3907
Basic Calibration Parameters
                                  Semorx                  Sersery             Semere          weres
 Nom (VVimP"                        045                      asr               ose            2101%
 5CP (m)"                           99                       1000              seo

Modulation Calibration Parameters
 vo        Communication System Name                 a           s      c       o      w            Tnst
                                                     as        aaviv           a8      my           (on
 o         ow                                  x     se         —|_os   19|    om      mss      ware
                                               Y     o0         oo      10             1500
                                               2     EM         o0      10             was



 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 mufiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probabiity of approximately 95%.

1 theurceiiniaso Norm x¥ 6on aect ieEfld urceaty ns TSL e Pages S ond )
? Mamerea inantcaton parenat uncotainy nat ronured
© Ureatay in ternined usog e mas. cvatonm Inowresponsearphinretongrcavnatonand is oxrosne o h ssmc of he
fitvibe.




Coifcate No: EX3—3807.Fobt?                    Page dof 11


                                                                                                        sa
                                                                                                           27

                                                                                                           L




©aove— sN00r                                                                                    Fobwary 27, 2017




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3907

Calibration Parameter Determined in Head Tissue Simulating Media
    1 uty       pi.f.!ia“&'.,s   c.:gmyu        ComFX     ComEY       ConyEZ     Apha®     D(:fl')       (Hg)
      2450          2o2             180          137         n3r        t3r       035        oe       £120%

      5200          360             480          550        550         5.50      0.30       180      £13.1%

      5250          350             4n           545         545        545       030        180      a19%

      5500          356             496          503         503        508        035       180      a13%

      5eo0          355             sor          RES        459         469        040       180      £131%

      srso          364             522          480        4.80        480        0.40      160      £131%

      5e00          353             sar          481        481         481        o.t0      180      a131%


© rrenueney vayze 0 io 10Mzony applos s OASY t nc haner se Page 2ts s resticlt 50 itc The
rcetany is theRSS of heGont urceralnyatcatoraten Favancy ons thvncerwntyfohe nc hacuency and,Frecuercy rtaty
Befow 20 hit iss 10.25,1, 50ant 70 ie o onE assensnani t 30,6¢,128,190 and 220 Mrespecna, Abov 5Oe roovirey
yaidy canbe onentes ta 110 is
* Athenuendas bdew 3 GHtz. tw valiyoftsueparantos (on) can brelant 10% 1 vitcompersaton ommi apntedto
rmeosres SR vaies.A raguecios aove 3 Ge, he valtty oIsou araneter (and o)is renvitet s 5%. The nctainy i the RSS of
{re Comwrcatand t inuted tagetue pronates.
* AprafQsnn aredternines dvigcalbraton SPEAG wararst he rerning dston ue o te bundan efectater canpensston is
aligeess han s 1% o Requencas ow Gi and uim s1 o rcuancln bntioon 5.GH t ary amance lage ranhatth rooe tp
darmtetom o boundan:




Corlfcate No: 3—3007_Fobi                           Page Sof11




_ an, Inc.

                                .                   n
                            simile: +81 463 50 6401


                                                                                    2                     tds
                                                                                                      —           2

                                                                                                                  s


exsovs— snissor                                                                                   reowary2r, 2017


DASY/EASY — Parameters of Probe: EX3DV4 — SN:3907

Calibration Parameter Determined in Body Tissue Simulating Media
                  Relatve       Conductly                                                      Dep®             Une
    rouity©     pormitiity®       (sim)"        Convex     ComvEY      |_ConvEz_|   Alpha®      (mm)            te2)
      2450          sa7             195          138         T3          738            056    PES          £120%

      5250          489             536          465         485         465            0.40    180         £18.1%

      5600          485             s77          370         378         378            0.50    180         £19.1%

      srso          483             50«          413         418         419            0.50    180         £18.1 %



©requency vadyanove 300 itc t 100 Mitcotyannte o BASY ve 4 ans Ngte (se Rage2l n esn o 50. The
yscetainyis the RSS ow Gor uncetanty t cttraton fequancyand e urcesany oin ndcte fesuency hnd Fequercy vity
eton 00Mz is a 10254980 ond 70 e o Comansosmonts t 20, 04 12. 150 ard 220 Mc eapecival. Above 5 Oe rsuercy
yaldyco be ecendadt 110 itz
* atRequentos blow3 G, valdy t twue paranatar (un a)cab alaned t 10%iecomponsatin fcencta is appled to
wreosired GAR valies l tequencas atove 2 Gbt e valot t fsie aranates (eod ) s resticind o# 53. Theurcotany s tRSS of
(he Gorne srcetiny for nceales arsetsive pinoto.
*Noraloan o diiernine aving catiraton SPERG waratstat h remaing davaton ue to h bsnday efec atar cononsstonis
aaysessraa1% torrequenis l 3 GHt and bfon 229 forFequendes oabvean 20 GHt atany atrce brge ho ht e rote tp
darmter on re bosnday




Certfcate No: ©0—3007_Febtr                         Page Sot 11


mxoovi—sniaoor                                                                                                                February 27 2017

                                                                   Frequency Response of E—Field
                                                                          (TEM—Cell:ift10 EXX, Waveguide: R22)

                         s
                                  rrmrvadrmionenbrrmbrmrton
                         &      5 8
        Frequency response (normazed)

                                                 F
                                                              f
       w    a § 5     3      a




                                                                     i                                                  \
                                                                    soo           1000       '5‘30      ano          2 Soo
                                                                                             TIMHz]
                                                                            Ted                                  3
                                                                                                                 L

                                                                  Uncertalnty of Frequency Response of E—feld: # 6.3% (x=2)




ConticaNo: EX3— 2007 Fobtr                                                               Page 7 of 11




                                                                           onmene ccecss PAN
                                                              . ..simle: +81 463 50 6101


                                                                                                    isa
                                                                                                        on




    exsovi— sncsoor                                                                          Fobrury 27, 2017



                                      Receiving Pattern (¢), 9 = 0°
                        £=600 MHz, TEM                                      f=1800 MHz,R22
        ner o)




                                   Uncertainty of Axial Isotropy Assessment: + 0.5% (ke2)




    Centicate No: Ex3.3007.Fobi7                        Page 8of 11




       an, Inc.

                                      .                     ran
.                                  simile:+81 463 50 6401


excovsnasor                                                                                                    retbrumy 27, 2017


                                     Dynamic Range f(SARneaq)
                                              (TEM cell , far= 1900 MHz)




                    109.


                goa0l=
                &

                $ «*              lsfpgiifinde               hm                                  £


                    10e


                          io            dot       io       io      o         ie                           10
                                            3x        Sat imwiena)
                                            —
                                       ratcompensatat                 a
                                                                   conpensated

                          #
                          44—                                            JuBl       _1   L   HBW__LHL
                   El
                   § 00|               ‘    {
                                       ithneegaet                                                    ||
                   §       .F           f                            .                                |
                              .                                                 t                     |
                       a4_L UHL NB
                                oo      e          to        foo     to        too  to
                                                       SAM [iWema)
                                            E4)
                                       retcomgansand                  conpensaes
                                   Uncertainty of Linearity Assessmont:# 0.6% (ke2)



Certoas No: EX3—a007.Foot?                              page sot 1




                                       nnmmnes coens,     CAN
                          —simile: +81 463 50 6401


                                                                                                    .                    Sa
                                                                                                                   tas
                                                                                                               .          Laor




      exoove— stacor                                                                                           Febriay 27,2017

                                       Conversion Factor Assessment
                        1= 2150 MizWGLS Ree (1cont)                                   1+ 2150 MieWGLS Ree (_come)




            $.,         >                                          i.
            i                                                      &



                3
                    i       oT   t      & k Siuties
                                      s M      % i  ael                 sfucrie
                                                                                  C    snyon    w       "yor   db)
                                                                                                               we

                                      Deviation from Isotropy in Liquid
                                                     Error (4, 9), f=900 MHz




                                     m sos os cos oz co or on os o8                                       is
                                      Uncertainty of Spherical Isotropy Assossmont: 2 26% (ke2)



      Conticat No: Exe—a007. robt                           Page toort1




        an Ino


                                        ogan ncorrrecuarPAN
womproinron mvo ce www 1 «siimier +81 463 50810 1


                                                                     sa
                                                                        moi7




Excove— sawor                                                Febrary 27,2017



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3907

Other Probe Parameters
 Sensor Avangement                                               Triangular
 Comector Angle ()                                                     iss
 Mechanical Surface Detecion Vode                                  enabied
 Optical Surace Detection Mode                                     isabled
 Probe Overal Lengh                                                357 mm
 Probe Body Diameter                                                10 mm
 Tip Lengh                                                           $mm
 Tip Diameter                                                       25 mm
 Probe Tip to Sensor X Galbralon Point                               1 mm
 Probe Tipto Sensor V Calbration Point                               T mm
 Probe Tisto Sensor Z Galbraion Point                                Tmm
 Recommondad Massurement bistance from Surface                      14 mm




Certfcate No: X33007_Fobi?                    Page 11 0(11




  an, Inc.

                                                   maN
                         simile: +81 463 50 6401


Calibration Laboratory of                                                           . Sotwoirischor Kaertinst
Schmid & Pariner                                                                       Sorvie auimse ditalomage
  Engineering AG                                                                    ©. servido mvizero alorsturn
Zooghavestresse43,0004 zurich, Sutzntan                                             S swss Gatbraton Service

pestestodby th Suiss Accrtaton Sarvce GAG)                                           Aecredtatontie: SCS 0108
"Te Sniss Accrediation Seiceis one ofthesignatoes tothe EA
Moltliteral Agreementfor th recognition of caltvaton corttstes
im         UL Japan Shonan (Vitec)                                           cerifcate ns: D2450V2—822_Jan17

CALIBRATION CERTIFICATE
 Object                        Dedsov2 — SN:e22

 calbralen procedunte)         GA CAL—O5M9
                               Calforation procedure for dipole validation kits above 700 Mz



 Cattraton da:                 January 11, 2017

 hicaibraon curtfenle docamontsha ce toratorasandart,which reatco h siyicl unts o messuroments 3
 The measiamonts and t uncetainies wih condace probssy aregven on th folonng pages ant ar pat ofhe cotfcae.
 Alcaltratons hav beenconducte in haclosns aborseryfcity: envrcomen tanporiixe (2 s 2C and hrniy «70%.
 Calbraton Easementueed (VATE rtenlfocattraton)
 prman sursus                   o+                   caiDats Gortiew No)                   sctecuied Cattaten
 Poner mote nr                  sroms                ceto18 io. etrcezsateens              Roar
 Ponersensor \RB.2ot            onroson              ceeets ho. 217.cze00)                 pad
 Penersensor NR291              on tosess            sercets ho. 2rrezess)                 Roar
 Reterance 20 68 Aternmtor      it soss 0)           cefort6 o 217.czr00)                  Roar
 Tope1 mianatch comnaton        on sorra oeser       ocveeto o. 2t7.ze08)                  Aoar
 Reternce Prove Ece             orow                 stove.1s(io. E3700Decre)              Beoar
 oaee                           onon                 dolint (io. DActa0tdam)               Jns
 secontay sundarts              B                    ctek Date (nouse)                     seredes cteck
 Power mater EP42A              stt aosreaorer       80c15 m house ces Oct—1)              InhousecheckOct18
 Poversonsorp e1sta             stt ussrmsares       orouts (ntowecreck O                  InrousectectOc18
 Poversorsorhp eacta                mroosorr         oromisnhocectexcon#)                  Intousecteckoc—18
 AF genter mas sireos           sn roour2            16u15 tn house check Oc)              Intouse ctecOc—18
 Netwot: Anabizer hP arsse      sn ussracosss        1001 (nhousecteccOc15)                Inouse teatOot7
                                Nime                             Funcion                    Snatie
 Catmtesty:                     seronKaste                       tatoratoy TesictnC)“[    > [//

 soormnedbe:                    Kat rosore                       Tectnicaianager         ///g

                                                                                           Issuetdariany 13, 2017
 Ts cataton cortfcnsshalnot breprodicadaxcapti ul itoct writen approni o he labortoy:

Corifcate No: D2e50V7.822..Jant?                         Page 1 o8




. an, Inc.

                                                          ‘AN
                                    : 181 463 50 6401


                                                                                              sa
                                                                                                207




Calibration Laboratory of                                                  Schwelzerscher Kaltitestonst
Schmid & Partner                                                           Senvce suisse attlonnage
  Engineering AG                                                           Sonvicl sviezero d tartin
‘teughausstrasse43, 6004 zurich, Surtzetand                                uiss Caliration Serice

Resrudtn by he Svis Accudtaton Sevice (GAS)                             Aecrtation: SCS 0108
‘The Sss Accredtation Servieis one of the signatories o thEA
Mulateral Agreemantfor threcogntion of calaton corifcatss
Glossary:
TSL                       tissue simulating liquid
ConvE                     sensitivity in TSL / NORM xy,z
N/A                       not applicable or not measured
Calibration is Performed According to the Following Standards:
    a) IEEE Sid 1528—2013, IEEE Recommended Practice for Determining the Peak Spatia—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques‘, June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
        devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)*,
         February 2005
    c) 12C 62209—2, ‘Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximily to the human body (frequency range of 30
       MHz to 6 GHz)®, March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    * Measurement Conditions: Further details are available from the Validation Report at the end
      of the certiicate. All figures stated in the certificate are valid at the frequency indicated.
    * Antenna Paramoters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
        parallelto the body axis.
    *   Feed Point Impedance and Return Loss: These parameters are measured with the dipole
        positioned under the liquid filed phantom. The impedance stated is transformed from the
        measurement at the SMA connector to the feed point. The Return Loss ensures low
        reflected power. No uncertainty required.
    *   Electiical Delay: One—way delay betweenthe SMA connector and the antenna feed point;
        No uncertainty required.
        SAR measured: SAR measured at the stated antenna input power.
        SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
        connector.
    +   SAR for nominal TSL parameters: The measured TSL parameters are used to calculete the
        nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Corifcate No: Dass0v2.022..Jant?                       Page2 o8




   an, Inc.

                                                          N
                               simile: +81 463 50 6401


                                                                                                    . —cudpy
                                                                                                    nlw cce
                                                                                      se a ow008
                                                                                           mm frodue n




                                            sj0g abr                    Livermeze—enomed on ernumeo



     Cod s set + um ez            mt o peareasou               sietewered 131 foog vwou 10 uv$
           Sum ons               zamod indu mut ose                                peissaw uy$
                                      Torpucs          T51 fbou o (6 01) uo 01 1on0 pabeiare uy$

     (cou x ou * Sum e            mt oi perjeaou               sierewered 151 Aboa feuwou 10 y$
           Sumoet                zomod indumu osg                                   poinseon uy$
                                     veupueo            Ts1 Apog 1o(6 1) jwo 1 10n0 poberane uy$

                                                                       781 Apog uim 1insoi uy$
       —                —            5.50>                To Sutinp abueyo aimieradum 161 Apog
  w or upuzo¢        worces        o. Gozoz                        siotownied 131 Apog poinsvany
   woun se‘t           Ees          o.0%e                           mrownied 161 Spon umon
   Angonpig          Aumpmog       Cmm
                                                   ‘Feidde ow sromoms poe siewnred umojo) eQL
                                                                          sicjoumied 131 Apog

     (c=d s set + Sam ove         wnt o1 poryeaiou             storewered 131 peot| rumiou o) y$
           um ere                Jenod indu mw ose                                   pansean uw5
                                      voupuce           Ts1 peoiiio (601) .wwo o ono pabeiane uys

     oD s on * um ore             mt o perteazou               stetetreied 181 peor| utwou1j y
           Sumver                sawod indu mose                                   pemsvow uy$
                                     vourneo              71 proj 1(6 1) juo 1 Jmo pobrone uys

                                                                       T61 prop tiim 1inser uy$
       —                —            5.50                 o Buuinp adueyo ampereduim 181 prok
  » o wus eo‘t       woroee       o. Gorom)                        simounied T61 proti painseony
    wrouw o.           zse           o.0%                           siomumed 151 peos feulwwon
   Apgonpuog         Apmung        ainjeeduoy
                                                   ‘PeidGe en sromnaes pue srowsied Buropo) s1
                                                                            sicroumied 151 peok


                                   zin 1 rerinoere                                    Aovenbaig
                                   wus— 2 %o %p                        vonjossy usog wooz
        Jeoods uim                     w or                      751 — satia9 arodig sourieia
                                 wownya retsrempont                                  woweya
                                 vomnpdenc peoverpy                             vorodersa
           a#esn                       sheva                                   vorien asva
                                             "Tatsdue uand o se 19 sn WpenSjuce uanle aava
                                                                        suogipuon jowaunseajy

              ons exnraie meumenees
mmz
 czom c aroos


      vorsman +      onpmereu,


                                                                                  Arrrmane         >« nesssriea
                                                                                  reem             : vaze
            Callbration certiicate: Dinole   D2450Y2) condi
     Appendix (Add           mal assessments outside the scope of SCS 0108)
     Antenna Parameters with Head TSL
           Impedance, ranstormed to eed point                                       ssaaes10
           Return Loss                                                                —22000
     Antenna Parameters             with Body TSL
           Impedance, ransformed to fed point                                       soen+sojn
           Retum Loss                                                                —24508
     General Antenna Parameters and Design
           Electrieal Delay (one drecton)                                              1.156ns
     Aftelong term use with 19W radiated power, only a sight warming ofthe cipole near the feedpoint can bo maasured.
     The dipole is made of standard seriigd coaxial able.The conter conductor ofthefeeding tis dreclly comnected to the
     second arm ofthe diole. The antenna is therefor shortieuted for DCsignals. On some ofthe dipoles, small end caps
     are added to he dipole arms in ordet improvo matching when loaded accorting t the positon as explained n the
     "Measurement Conditons‘ paragraph, The SAR data are notalfected by this change. The overal cipole lengthis stil
     eccortingto the Standard.
     No excessive foroe must be applied to the dipale ams,because they mightbend or the soldered connections near the
     feedpoint may be damaged.
     Additional EUT Data
           Marutactured by                                                             sreae
           Manufectured on                                                        December 11, 2008
     Cartfcate No: D2«s0v2—022.lant?                      Pago 4 ot8
UL pupass, anc.
Sho——2 mhant an


                                                                                       sa
                                                                                          w7




DASYS Validation Report for Head TSL
                                                                                   Date: 10.01.2017
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz: Type: D2450V2; Serial: D2450V2 — SN:822

Communication System: UID 0 — CW Frequeney: 2450 MHz
Medium parameters use             2450 MHz: a= 1.88 S/m; 6 =38; p   1000 ke/m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/TEC/ANSI C63.19—2011)
DASY52 Configuration:
    *   Probe: EX3DV4 — SN7349; ConvB(7.72, 7.72, 7.72); Calibrated: 31.12.2016;

    +   Sensor—Surface: 1 Amm (Mechanical Surface Detection)
    +   Electronics: DAEA S601; Calibrated: 04.01.2017
    +   Phantom: Flat Phantom 5.0 (front); Type: QD 000 PSO AA; Serial: 1001
    +   DASY52 52.4.8(1258); SEMCAD X 14.6.1007372)

Dipole Calibration for Head Tissue/Pin=250 mW d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: d«=3mm, dy=Smm, de=3mm
Reference Value= 112.9 V/m; Power Drift =—0.06 dB
Peak SAR (extrapolated) =26.9 W/ke
SAR(T g) = 13.1 W/kg: SAR(LO g) = 6.09 We
Maximum value of SAR (measured) =21.6 Wike
        «8
        a
        a.00
        —8.00
        412.00

        a6.00
        20.00
                  0 dB =21.6 Whe=          13.34 dBWike




Gorifeate No: Dads0¥2.002.Jani?                    Page S ot 8




   an, Inc.

                                                      maN
                           simile: +81 463 50 6401


                                                                                  isa
                                                                                    "an7




Impedance Measurement Plot for Head TSL




                                                  10 amn anir acsemice
          gen sa      s ue          assace stoes ons         2 asa.000 o0 ez
                                          t

          o1
          s


          be
          5
          ma
          ce




          ma
                   Srria s5e.oon co ie                     Sroe 2 csamee vee ie




Corticate No: DadS0V2—622..lant?                Page ote




. an, Inc.

                               .                   ran
                           simile: 181 463 50 6401


                                                                                         i8A
                                                                                   ras
                                                                                           "an7




DASY5 Validation Report for Body TSL
                                                                                  Date: 11.01.2017
Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:822
Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: {= 2450 MHz; o =2.02 S/m; c,= 52.7; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEEAEC/ANST C63.19—2011)
DASY52 Configuration:

    *    Probe: EX3DV— SN7349; ConvF(7.79, 7.79, 7.79); Calibrated: 31.12.2016;

    *    Sensor—Surface: 1 Amm (Mechanical Surface Detection)

    *    Electronics: DAE4 Sn601; Calibrated: 04.01.2017
    *    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Scrial: 1002

    +.   DASY5252.8.8(1258); SEMCAD X 14.6.10(7372)

Dipole Calibration for Body Tissue/Pin=250 mW d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: d=Smm, dy=Smm, de
Reference Value= 104.9 V/m; Power Drift
Peak SAR (extrapolated) =25.5 Wikg
SARCI g=12.6 W/kg: SAR(IO g) = 5.88 Wke
Maximum value of SAR (measured) =20.1 We

         a8
          0
         a.00
         .00

         1200
         16.00
         2000
                                       13.03 dBWhkz




Certicate No: Daés0v2.022.Jant7                  Page o 8




 _an. Inc.

                                &        .          ranr
                        . —‘simile: 181 463 50 6101


                                                                   Test report No.   : 11884511S-A
                                                                   Page              : 48 of 48       (End of Report)
                                                                   Issued date       : September 1, 2017

                                                                   FCC ID            : W2Z-02000004

Appendix 3-9: Calibration certificate: Dipole (D2450V2) (cont’d)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401



Document Created: 2017-09-01 17:19:19
Document Modified: 2017-09-01 17:19:19

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC