SAR Test Data&Instruments(1)

FCC ID: W2Z-01000001

RF Exposure Info

Download: PDF
FCCID_1413640

                                                                                   Test report No. : 31CE0283-HO-01-C
                                                                                   Page            : 17 of 59
                                                                                   Issued date     : December 22, 2010
                                                                                   FCC ID         : W2Z-01000001

APPENDIX 2: SAR Measurement data
Appendix 2-1: Evaluation procedure

The SAR evaluation was performed with the following procedure:
Step 1: Measurement of the E-field at a fixed location above the central position of flat phantom was used as a reference
        value for assessing the power drop.
Step 2: The SAR distribution at the exposed side of body position was measured at a distance of each device from the
        inner surface of the shell. The area covered the entire dimension of the antenna of EUT and suitable horizontal
        grid spacing of EUT. Based on these data, the area of the maximum absorption was determined by splines
        interpolation.
Step 3: Around this point found in the Step 2 (area scan), a volume of 28mm(X axis)28mm(Y axis)22.5mm(Z axis)
        was assessed by measuring 8810 points. And for any secondary peaks found in the Step2 which are within
        2dB of maximum peak and not with this Step3 (Zoom scan) is repeated.
        On the basis of this data set, the spatial peak SAR value was evaluated under the following procedure:
         (1) The data at the surface were extrapolated, since the center of the dipoles is 1mm away from the tip of the
             probe and the distance between the surface and the lowest measuring point is 2mm. The extrapolation was
             based on a least square algorithm [4]. A polynomial of the fourth order was calculated through the points in
             z-axes. This polynomial was then used to evaluate the points between the surface and the probe tip.
         (2) The maximum interpolated value was searched with a straightforward algorithm. Around this maximum
             the SAR values averaged over the spatial volumes (1g or 10g) were computed by the 3D-Spline
             interpolation algorithm. The 3D-Spline is composed of three one-dimensional splines with the “Not a
             knot"-condition (in x, y and z-directions) [4], [5]. The volume was integrated with the trapezoidal-algorithm.
             One thousand points (101010) were interpolated to calculate the average.
         (3) All neighboring volumes were evaluated until no neighboring volume with a higher average value was
             found.
Step 4: Re-measurement of the E-field at the same location as in Step 1 for the assessment of the power drift.
Step 5: Repeat Step 1-Step 4 with other condition or/and setup of EUT.




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                               Test report No. : 31CE0283-HO-01-C
                                                                                               Page            : 18 of 59
                                                                                               Issued date     : December 22, 2010
                                                                                               FCC ID        : W2Z-01000001

Appendix 2-2:         Measurement data
Step A1: Worst setup position search (Antenna#: Top)
Step A1-1 Top-touch (separation=0mm)
(Worst SAR for antenna-Top)
Date tested: 2010/11/30              Tested place: No.7 Shielded room                      Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.36 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant3,top,d0mm,11a-6m/
Area Scan: (51x91x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.183 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 3.35 V/m; Power Drift = -0.169 dB, Peak SAR (extrapolated) = 0.296 W/kg
SAR(1 g) = 0.085 mW/g ; SAR(10 g) = 0.025 mW/g
Maximum value of SAR (measured) = 0.173 mW/g
*.   Z Scan (1x1x12): Measurement grid: dx=20mm, dy=20mm, dz=1mm,
     Maximum value of SAR (measured) = 0.172 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 22.6 deg.C. /(after) 22.6 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)


UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                               Test report No. : 31CE0283-HO-01-C
                                                                                               Page            : 19 of 59
                                                                                               Issued date     : December 22, 2010
                                                                                               FCC ID        : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)
Step A1: Worst setup position search (Antenna#: Top) (cont’d)

Step A1-2 Front-touch (Patient side, separation=0mm)
Date tested: 2010/11/30              Tested place: No.7 Shielded room                      Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.36 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant3,frt(patient side),d0mm,11a-6m
Area Scan: (101x51x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.030 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 1.08 V/m; Power Drift = 0.20 dB, Peak SAR (extrapolated) = 0.069 W/kg
SAR(1 g) = 0.017 mW/g; SAR(10 g) = 0.014 mW/g
Maximum value of SAR (measured) = 0.028 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 23.0 deg.C. /(after) 23.0 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                Test report No. : 31CE0283-HO-01-C
                                                                                                Page            : 20 of 59
                                                                                                Issued date     : December 22, 2010
                                                                                                FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)
Step A1: Worst setup position search (Antenna#: Top) (cont’d)

Step A1-3 Rear-touch (Operator side, separation=0mm)
Date tested: 2010/11/30              Tested place: No.7 Shielded room                          Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.36 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant3,rear(operator side),d0mm,11a-6m
Area Scan: Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.057 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 1.68 V/m; Power Drift = 0.21 dB, Peak SAR (extrapolated) = 0.124 W/kg
SAR(1 g) = 0.031 mW/g; SAR(10 g) = 0.017 mW/g
Maximum value of SAR (measured) = 0.060 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 22.9 deg.C. /(after) 23.0 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                Test report No. : 31CE0283-HO-01-C
                                                                                                Page            : 21 of 59
                                                                                                Issued date     : December 22, 2010
                                                                                                FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)

Step A2: Change mode (Antenna#: Top)
Step A2-1 11n(20HT)-MCS0, Top-touch (separation=0mm)
Date tested: 2010/11/30              Tested place: No.7 Shielded room                          Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11n(20HT) (MCS0); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.36 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant3side,top,d0mm,11n(20ht)-mcs0
Area Scan: (51x221x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.076 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 2.02 V/m; Power Drift = 0.166 dB, Peak SAR (extrapolated) = 0.146 W/kg
SAR(1 g) = 0.043 mW/g; SAR(10 g) = 0.012 mW/g
Maximum value of SAR (measured) = 0.085 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 22.6 deg.C. /(after) 22.6 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                               Test report No. : 31CE0283-HO-01-C
                                                                                               Page            : 22 of 59
                                                                                               Issued date     : December 22, 2010
                                                                                               FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)

Step B1: Worst setup position search (Antenna#: Side-top)
Step B1-1 Right-touch (separation=0mm)
(Worst SAR for antenna-Side-top)
Date tested: 2010/11/30              Tested place: No.7 Shielded room                         Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.36 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant1,right,d0mm,11a-6m
Area Scan: (91x51x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.137 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 3.99 V/m; Power Drift = -0.085 dB, Peak SAR (extrapolated) = 0.193 W/kg
SAR(1 g) = 0.052 mW/g; SAR(10 g) = 0.017 mW/g
Maximum value of SAR (measured) = 0.119 mW/g
*.   Z Scan (1x1x12): Measurement grid: dx=20mm, dy=20mm, dz=1mm
     Maximum value of SAR (measured) = 0.112 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 22.9 deg.C. /(after) 22.9deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)



UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                Test report No. : 31CE0283-HO-01-C
                                                                                                Page            : 23 of 59
                                                                                                Issued date     : December 22, 2010
                                                                                                FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)
Step B1: Worst setup position search (Antenna#: Side-top) (cont’d)

Step B1-2 Front-touch (patient side, separation=0mm)

Date tested: 2010/12/2               Tested place: No.7 Shielded room                         Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.32 S/m; εr = 46.7; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant1,front(patient side),d0mm,11a-6m/
Area Scan: (251x161x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.051 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 0.183 V/m; Power Drift = 0.21 dB, Peak SAR (extrapolated) = 0.045 W/kg
SAR(1 g) = 0.0063 mW/g; SAR(10 g) = 0.00263 mW/g
Maximum value of SAR (measured) = 0.057 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 38.0 %; liquid temeprature: (before) 23.3 deg.C. /(after) 233 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                               Test report No. : 31CE0283-HO-01-C
                                                                                               Page            : 24 of 59
                                                                                               Issued date     : December 22, 2010
                                                                                               FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)
Step B1: Worst setup position search (Antenna#: Side-top) (cont’d)

Step B1-3 Rear-touch (operator side, separation=0mm)
Date tested: 2010/12/2               Tested place: No.7 Shielded room                         Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.32 S/m; εr = 46.7; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant1,rear(optr),d0mm,11a-6m
Area Scan: (81x151x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.062 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 1.37 V/m; Power Drift = -0.20 dB, Peak SAR (extrapolated) = 0.087 W/kg
SAR(1 g) = 0.034 mW/g; SAR(10 g) = 0.020 mW/g
Maximum value of SAR (measured) = 0.069 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 38.0 %; liquid temeprature: (before) 23.2 deg.C. /(after) 232 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                Test report No. : 31CE0283-HO-01-C
                                                                                                Page            : 25 of 59
                                                                                                Issued date     : December 22, 2010
                                                                                                FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)

Step B2: Change mode (Antenna#: Side-top)
Step B2-1 11n(20HT)-MCS0, Right-touch (separation=0mm)
Date tested: 2010/12/2               Tested place: No.7 Shielded room                          Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11n(20HT) (MCS0); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.32 S/m; εr = 46.7; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant1(Simultaneously-Tx1&2&3),right,d0mm,11n-20ht-mcs0
Area Scan:(221x51x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.045 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 2.44 V/m; Power Drift = 0.21 dB, Peak SAR (extrapolated) = 0.164 W/kg
SAR(1 g) = 0.024 mW/g; SAR(10 g) = 0.00647 mW/g
Maximum value of SAR (measured) = 0.055 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 38.0 %; liquid temeprature: (before) 23.2 deg.C. /(after) 23.2 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                Test report No. : 31CE0283-HO-01-C
                                                                                                Page            : 26 of 59
                                                                                                Issued date     : December 22, 2010
                                                                                                FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)

Step C1: Worst setup position search (Antenna#: Side-bottom)
Step C1-1 Right-touch (separation=0mm)
(Worst SAR for antenna-Side-bottom)
Date tested: 2010/11/30              Tested place: No.7 Shielded room                          Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.36 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant2,right,d0mm,11a-6m
Area Scan: (91x51x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.084 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 2.74 V/m; Power Drift = -0.046 dB, Peak SAR (extrapolated) = 0.171 W/kg
SAR(1 g) = 0.032 mW/g; SAR(10 g) = 0.00899 mW/g
Maximum value of SAR (measured) = 0.076 mW/g
*.   Z Scan (1x1x12): Measurement grid: dx=20mm, dy=20mm, dz=1mm,
     Maximum value of SAR (measured) = 0.075 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 22.7 deg.C. /(after) 22.8 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)


UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                 Test report No. : 31CE0283-HO-01-C
                                                                                                 Page            : 27 of 59
                                                                                                 Issued date     : December 22, 2010
                                                                                                 FCC ID           : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)

Step C1: Worst setup position search (Antenna#: Side-bottom)
Step C1-2 Front-touch (patient side, separation=0mm)
Date tested: 2010/12/1              Tested place: No.7 Shielded room                          Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.38 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant2,front(patient side),d0mm,11a-6m
Area Scan: (171x61x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (interpolated) = 0.032 mW/g




                                                 Max.SAR(intrpl.): 0.032mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 23.0 deg.C. /(after) 23.0 deg.C.
*. The zoom scan was not applied. The SAR value measured in area scan was very small. Therefore, for the zoom scan procedure, the correct
   interpolation process could not be achieved, because the target transmission level was very small.




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                 Test report No. : 31CE0283-HO-01-C
                                                                                                 Page            : 28 of 59
                                                                                                 Issued date     : December 22, 2010
                                                                                                 FCC ID           : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)
Step C1: Worst setup position search (Antenna#: Side-bottom) (cont’d)

Step C1-3 Rear-touch (operator side, separation=0mm)
Date tested: 2010/12/1               Tested place: No.7 Shielded room                         Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11a (6Mbps); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.38 S/m; εr = 47.2; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant2,rear(optr),d0mm,11a-6m
Area Scan:120x60,step=10 (121x61x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (interpolated) = 0.025 mW/g




                                     Max.SAR(intrpl.): 0.025mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 34.0 %; liquid temeprature: (before) 23.0 deg.C. /(after) 23.0 deg.C.
*. The zoom scan was not applied. The SAR value measured in area scan was very small. Therefore, for the zoom scan procedure, the correct
   interpolation process could not be achieved, because the target transmission level was very small.




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                Test report No. : 31CE0283-HO-01-C
                                                                                                Page            : 29 of 59
                                                                                                Issued date     : December 22, 2010
                                                                                                FCC ID          : W2Z-01000001

Appendix 2-2: Measurement data (cont’d)

Step C2: Change mode (Antenna#: Side-bottom)
Step C2-1 11n(20HT)-MCS0, Right-touch (separation=0mm)
Date tested: 2010/12/2               Tested place: No.7 Shielded room                          Tested by: Hiroshi Naka

DUT: DR-ID 601SE; Type: X-ray imaging system with wireless LAN; Serial: A121013
Communication System: 11n(20HT) (MCS0); Frequency: 5180 MHz(default-low); Crest Factor: 1.0
Medium: MSL5800; Medium parameters used: f = 5180 MHz; σ = 5.32 S/m; εr = 46.7; ρ = 1000 kg/m3
DASY4 Configuration:
- Probe: EX3DV4 - SN3679; ConvF(4.04, 4.04, 4.04); Calibrated: 2010/04/23
- Sensor-Surface: 2mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn626; Calibrated: 2010/02/10
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1059; Phantom section: Flat Section
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 184
m5180,ant-2(1&2&3sousin),right,d0mm,11n-20ht-mcs0/
Area Scan: (221x51x1): Measurement grid: dx=10mm, dy=10mm, Maximum value of SAR (interpolated) = 0.044 mW/g
Zoom Scan:28x28x22.5,step=4&2.5 (8x8x10)/Cube 0: measurement grid: dx=4mm, dy=4mm, dz=2.5mm
Reference Value = 2.65 V/m; Power Drift = 0.054 dB, Peak SAR (extrapolated) = 0.157 W/kg
SAR(1 g) = 0.010 mW/g; SAR(10 g) = 0.00149 mW/g
Maximum value of SAR (measured) = 0.033 mW/g




Additional information:
*.position of distance of DUT to phantom: 0mm (2mm to liquid), liquid depth: 144mm
*.ambient: 24.0deg.C / 38.0 %; liquid temeprature: (before) 23.2 deg.C. /(after) 23.2 deg.C.
*.white cubic: zoom scan area, red big cubic: SAR(10g ), red small cubic: SAR(1g)




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                              Test report No. : 31CE0283-HO-01-C
                                                                                                              Page            : 30 of 59
                                                                                                              Issued date     : December 22, 2010
                                                                                                              FCC ID             : W2Z-01000001

APPENDIX 3: Test instruments
Appendix 3-1: Equipment used




                                                                                                                              (Daily check)
                                                                                                                              Target value 5%




 No.7              SAR shielded room             TDK                     -                    -                SAR            (Daily check)
 Shielded room     (2.76m(W)x3.76m(D)x2.4m(H))                                                                                Ambient noise:
                                                                                                                              < 12mW/kg


The expiration date of calibration is the end of the expired month.
As for some calibrations performed after the tested dates, those test equipment have been controlled by means of an unbroken chains of calibrations.
All equipment is calibrated with traceable calibrations. Each calibration is traceable to the national or international standards.
[Test Item] SAR: Specific Absorption Rate, Ant.pwr: Antenna terminal conducted power




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                 Test report No. : 31CE0283-HO-01-C
                                                                                                 Page            : 31 of 59
                                                                                                 Issued date     : December 22, 2010
                                                                                                 FCC ID           : W2Z-01000001

Appendix 3-2: Dosimetry assessment setup

These measurements were performed with the automated near-field scanning system DASY4 from Schmid & Partner
Engineering AG (SPEAG). The system is based on a high precision robot (working range greater than 0.9 m), which
positions the probes with a positional repeatability of better than +/- 0.02 mm. Special E- and H-field probes have been
developed for measurements close to material discontinuity, the sensors of which are directly loaded with a Schottky diode
and connected via highly resistive lines to the data acquisition unit. The SAR measurements were conducted with the
dosimetry probes EX3DV4, SN: 3679 (manufactured by SPEAG), designed in the classical triangular configuration and
optimized for dosimetric evaluation. The probe has been calibrated according to the procedure described in [2] with accuracy
of better than +/-10%. The spherical isotropy was evaluated with the procedure described in [3] and found to be better than
+/-0.25 dB.



Appendix 3-3: Configuration and peripherals




The DASY4 system for performing compliance tests consist of the following items:
     A standard high precision 6-axis robot (Stäubli RX family) with controller and software.
 1
     An arm extension for accommodating the data acquisition electronics (DAE).
     A dosimetric probe, i.e., an isotropic E-field probe optimized and calibrated for usage in tissue simulating liquid. The probe is
 2
     equipped with an optical surface detector system.
     A data acquisition electronic (DAE), which performs the signal amplification, signal multiplexing, AD-conversion, offset
 3   measurements, mechanical surface detection, collision detection, etc. The unit is battery powered with standard or rechargeable
     batteries. The signal is optically transmitted to the EOC.
     The Electro-optical converter (EOC) performs the conversion between optical and electrical of the signals for the digital
 4   communication to the DAE and for the analog signal from the optical surface detection.
     The EOC is connected to the measurement server.
     The function of the measurement server is to perform the time critical tasks such as signal filtering, control of the robot operation and
 5
     fast movement interrupts.
 6   A probe alignment unit which improves the (absolute) accuracy of the probe positioning.
 7   A computer operating Windows XP.
 8   DASY4 software.
 9   Remote control with teaches pendant and additional circuitry for robot safety such as warning lamps, etc.
10   The phantom.
11   The device holder for EUT. (low-loss dielectric palette)
12   Tissue simulating liquid mixed according to the given recipes.
13   Validation dipole kits allowing to validate the proper functioning of the system.




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                  Test report No. : 31CE0283-HO-01-C
                                                                                                  Page            : 32 of 59
                                                                                                  Issued date     : December 22, 2010
                                                                                                  FCC ID             : W2Z-01000001

Appendix 3-4: System components
1) EX3DV4 Probe Specification
Construction:
 Symmetrical design with triangular core.
 Built-in shielding against static charges.
 PEEK enclosure material (resistant to organic solvents, e.g., DGBE).
Calibration (S/N 3679):
Basic broad band calibration in air.
Conversion Factors(Head and Body): 2450, 5200, 5300, 5500, 5600, 5800MHz
Frequency:
10 MHz to > 6GHz, Linearity: 0.2 dB (30MHz to 6GHz)
Directivity:
0.3 dB in HSL (rotation around probe axis)
0.5 dB in tissue material (rotation normal to probe axis)
Dynamic Range:
10W/g to > 100 mW/g; Linearity: 0.2 dB (noise: typically < 1W/g)
                                                                                                                         EX3DV4 E-field Probe
Dimensions:
Overall length: 330mm (Tip: 20mm)
Tip diameter: 2.5mm (Body: 12mm)
Typical distance from probe tip to dipole centers: 1mm
Application:
High precision dosimetric measurement in any exposure scenario (e.g., very strong gradient fields).Only probe
which enables compliance testing for frequencies up to 6GHz with precision of better 30%.

2) Phantom (Flat type)
Construction:
A cover prevents evaporation of the liquid. Reference markings on the phantom
allow the complete setup of all predefined phantom position and measurement grids
by manually teaching three points with the robot.
Shell Thickness:
Bottom plate: 2 0.2mm
Dimensions:
Bottom elliptical: 600400mm, Depth: 190mm
Filling Volume:                                                                               ELI.4.0 flat phantom
Approx. 30 liters

2) Phantom (SAM twin type)
Construction:
The shell corresponds to the specifications of the Specific Anthropomorphic
Mannequin (SAM) phantom defined in IEEE 1528-200X, CENELEC EN
50361 and IEC 62209. It enables the dosimetric evaluation of left and right
hand phone usage as well as body mounted usage at the flat phantom region.
A cover prevents evaporation of the liquid. Reference markings on the
phantom allow the complete setup of all predefined phantom positions and
measurement grids by manually teaching three points with the robot.
Shell Thickness:
Flat section: 2  0.2mm, ERP: 6  0.2mm                                                       SAM twin phantom
Filling Volume:
Approx. 25 liters
Dimensions:                                                                                              Device holder
8101000500mm (HLW)

3) Device Holder
For this measurement, the urethane foam was used as device holder.
In combination with the Twin SAM Phantom V4.0/V4.0c or ELI4, the Mounting Device enables
the rotation of the mounted transmitter device in spherical coordinates. Transmitter devices can be
easily and accurately positioned.
The low-loss dielectric urethane foam was used for the mounting section of device holder.



UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                          Test report No. : 31CE0283-HO-01-C
                                                                                          Page            : 33 of 59
                                                                                          Issued date     : December 22, 2010
                                                                                          FCC ID          : W2Z-01000001

Appendix 3-5: Test system specification

RX60L Robot
     Number of Axes        :   6                             Payload                      :    1.6 kg
     Reach                 :   800mm                         Repeatability                :    0.025mm
     Control Unit          :   CS7M                          Programming Language         :    V+
     Manufacture           :   Stäubli Unimation Corp. Robot Model: RX60
DASY4 Measurement server
     Features    : 166MHz low power Pentium MMX.
                      32MB chipdisk and 64MB RAM Serial link to DAE (with watchdog supervision) 16 Bit A/D converter for
                      surface detection system. Two serial links to robot (one for real-time communication which is supervised
                      by watchdog) Ethernet link to PC (with watchdog supervision).
                      Emergency stop relay for robot safety chain. Two expansion slots for future applications.
     Manufacture : Schmid & Partner Engineering AG
Data Acquisition Electronic (DAE)
       Features              : Signal amplifier, multiplexer, A/D converter and control logic.
                                  Serial optical link for communication with DASY4 embedded system (fully remote controlled).
                                  2 step probe touch detector for mechanical surface detection and emergency robot stop (not in -R
                                  version)
       Measurement Range : 1µV to > 200mV (16bit resolution and two range settings: 4mV, 400mV)
       Input Offset voltage : < 1µV (with auto zero)
       Input Resistance      : 200M                          Battery Power      : > 10hr of operation (with two 9V battery)
       Dimension             : 606068mm                     Manufacture        : Schmid & Partner Engineering AG
Software
       Item                 :      Dosimetric Assessment System DASY4
       Software version No. :      DASY4, V4.7 B80           Manufacture / Origin :       Schmid & Partner Engineering AG
E-Field Probe
       Model               :   EX3DV4 (sn: 3679)             Construction         :   Symmetrical design with triangular core
       Frequency           :   10MHz to 6GHz                 Linearity            :   0.2dB (30MHz to 3GHz)
       Manufacture         :   Schmid & Partner Engineering AG
Phantom (1)
      Type                 :   ELI 4.0 oval flat phantom     Shell Material :     Fiberglass
      Shell Thickness      :   Bottom plate: 2 0.2mm        Dimensions :         Bottom elliptical: 600400mm, Depth: 190mm
      Manufacture          :   Schmid & Partner Engineering AG
Phantom (2)
      Type                 :   Flat phantom                    Shell Material :   Fiberglass
      Shell Thickness      :   flat section: 2 0.2mm, ERP: 6 0.2mm
      Manufacture          :   Schmid & Partner Engineering AG


Appendix 3-6: Simulated tissue composition

                                    Mixture (%)
      Ingredient
                           Body 5800MHz (type: SL AAM 580)
        Water                        60-78 %
    Mineral Oil                      11-36 %
    Emulsifiers                      0.5-15 %
  Additives and Salt                  0.4-3 %
    Manufacture            Schmid&Partner Engineering AG




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                                             Test report No.        :   31CE0283-HO-01-C
                                                                                                                             Page                   :   34 of 59
                                                                                                                             Issued date            :   December 22, 2010
                                                                                                                             Revised date           :   January 31, 2011
                                                                                                                             FCC ID                 : W2Z-01000001

Appendix 3-7: Simulated tissue parameter confirmation
The dielectric parameters were checked prior to assessment using the 85070E dielectric probe kit.
The dielectric parameters measurement is reported in each correspondent section.
                                                             Dielectric parameter measurement results
                                Ambient         Liq.Temp.[deg.C.] Liquid                                            Target value                        Deviation   Deviation
                  Freq.                                                                                                                                                         Limit
    Date                    Temp Humidity                               Depth           Parameters                #1:Std.   #2:Cal.     Measured          for #1      for #2
                 [MHz]                              Before    After                                                                                     (Std.)[%]   (Cal.)[%]
                                                                                                                                                                                [%]
                           [deg.C.] [%]                                 [mm]                                       (*1)       (*2)
 November 30,                                                                   Relative permittivity: r [-]     49.0       47.7           47.10        -3.9        -1.3        5
                  5200      23.3          34        23.4.      23.4     (144)
    2010                                                                         Conductivity:  [S/m]            5.30       5.30           5.369        +1.3        +1.3        5
 December 1,                                                                    Relative permittivity: r [-]     49.0       47.7           47.15        -3.8        -1.2        5
                  5200      23.6          35        23.5       23.5     (144)
    2010                                                                         Conductivity:  [S/m]            5.30       5.30           5.384        +1.6        +1.6        5
 December 2,                                                                    Relative permittivity: r [-]     49.0       47.7           46.83        -4.5        -1.8        5
                  5200      23.6          35        23.7       23.7     (144)
    2010                                                                         Conductivity:  [S/m]            5.30       5.30           5.352        +1.0        +1.0        5
*1. The target value is a parameter defined in OET65, Supplement C.
*2 The target values are the calibrated dipole MSL parameters. (D5GHzV2(SN:1070), page 42 in this report)

*. Decision on Simulated Tissues of 5200MHz
    In the current standards (e.g., IEEE 1528, OET 65 Supplement C), the dielectric parameters suggested for head and body tissue simulating
    liquid are given at 3000MHz and 5800MHz. As an intermediate solution, dielectric parameters for the frequencies between 5000 to 5800
    MHz were obtained using linear interpolation. Therefore the dielectric parameters of 5200MHz (the frequency for the validation) and
    5180MHzwere decided as following.
    Standard and interpolated dielectric parameters for head and body tissue simulating liquid in the frequency range 3000 to 5800MHz.
                                           Head Tissue                                  Body Tissue
            f (MHz)                                                                                                              Reference
                                     r               [S/m]                      r               [S/m]
                3000                38.5               2.40                      52.0               2.73                          Standard
                5800                35.3               5.27                      48.2               6.00                          Standard
                5180                  -                  -                       49.0               5.28                         Interpolated
                5200                  -                  -                       49.0               5.30                         Interpolated
                5240                  -                  -                       49.0               5.35                         Interpolated



Appendix 3-8: System validation data
Prior to the SAR assessment of EUT, the system validation kit was used to test whether the system was operating within its
specifications of 10%. The validation results are in the table below.
                                                                          System validation results
                                                                                                                                      System dipole validation target & measured
                                     Ambient   Liquid Temp. [deg.C.] Liquid Permittivity Conductivity Power
             Freq.     Liquid                                                                                    SAR(1g) [W/kg] (at 1W)
  Date                                                               Depth measured measured           drift                                                         Deviation Limit
            [MHz]       Type     Temp Humidity                                                                Target
                                               Check Before After [mm]         r [-]      [S/m]]     [dB]               Measured (*3)                                [%]      [%]
                                [deg.C.] [%]                                                                 value (*2)
November                                                                                                                                                80.1
 30, 2010
            5200       Body        23.5        34       23.4     22.9    22.9   144         47.1           5.37          0.016       75.9                              +5.5     10
                                                                                                                                                (8.01 (at 100mW))
December                                                                                                                                                74.2
 1, 2010
            5200       Body        23.9        38       23.5     23.6    23.4   144         47.2           5.38        -0.099        75.9                              -2.2     10
                                                                                                                                                (7.42 (at 100mW))
December                                                                                                                                                79.5
 2, 2010
            5200       Body        24.0        38       24.7     23.7    23.4   144         46.8           5.35          0.055       75.9                              +4.7     10
                                                                                                                                                (7.95 (at 100mW))
Note: Refer to Appendix 3-10 Validation measurement data for the above result representation in plot data.
*2. The target values are the calibrated dipole parameters. (D5GHzV2(SN:1070), page 42 in this report)
*3. The measurement value was normalized to 1W forward power.




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401


                                                                                                         Test report No. : 31CE0283-HO-01-C
                                                                                                         Page            : 35 of 59
                                                                                                         Issued date     : December 22, 2010
                                                                                                         FCC ID                : W2Z-01000001

Appendix 3-8: System validation data (cont’d)




                                                                                                                                          Liquid depth: 144mm


                                                                                                        Dipole (D5GHzV2)


                                             Test setup for the system performance check


Appendix 3-9: Validation uncertainty


              Uncertainty of SAR measurement system                                        56 GHz
                             /Validation                                             1g SAR     10g SAR
     combined measurement uncertainty of the measurement system (k=1)                ± 12.1%    ± 11.9%
                      expanded uncertainty (k=2)                                     ± 24.2%    ± 23.7%
[56GHz]
                                                       Uncertainty    Probability                 ci         ci          ui                  ui
                     Error Description                                               Divisor                                                              vi, veff
                                                         Value        distribution               (1g)      (10g)        (1g)               (10g)
  A    Measurement System                                                                                          (std. uncertainty) (std.uncertainty)
  1    Probe calibration                                   ±6.8 %       Normal          1         1          1        ±6.8 %             ±6.8 %             ∞
  2    Axial isotropy                                      ±4.7 %     Rectangular      √3        0.7        0.7       ±1.9 %             ±1.9 %             ∞
  3    Hemispherical isotropy (*flat phantom, <5)         ±2.6 %     Rectangular      √3        0.7        0.7       ±1.1 %             ±1.1 %             ∞
  4    Boundary effects                                    ±2.0 %     Rectangular      √3         1          1        ±1.2 %             ±1.2 %             ∞
  5    Probe linearity                                     ±4.7 %     Rectangular      √3         1          1        ±2.7 %             ±2.7 %             ∞
  6    System detection limit                              ±1.0 %     Rectangular      √3         1          1        ±0.6 %             ±0.6 %             ∞
  7    System readout electronics                          ±0.3 %       Normal          1         1          1        ±0.3 %             ±0.3 %             ∞
  8    Response time                                       ±0.8 %     Rectangular      √3         1          1         ±0 %               ±0 %              ∞
  9    Integration time                                    ±2.6 %     Rectangular      √3         1          1         ±0 %               ±0 %              ∞
 10    RF ambient - noise                                  ±3.0 %     Rectangular      √3         1          1        ±1.7 %             ±1.7 %             ∞
 11    RF ambient - reflections                            ±3.0 %     Rectangular      √3         1          1        ±1.7 %             ±1.7 %             ∞
 12    Probe positioner mechanical tolerance               ±0.8 %     Rectangular      √3         1          1        ±0.5 %             ±0.5 %             ∞
 13    Probe positioning with respect to phantom shell     ±9.9 %     Rectangular      √3         1          1        ±5.7 %             ±5.7 %             ∞
 14    Max.SAR evaluation                                  ±4.0 %     Rectangular      √3         1          1        ±2.3 %             ±2.3 %             ∞
  B    Dipole
 15    Dipole axis to liquid distance                      ±2.0 %     Rectangular      √3         1          1        ±1.2 %             ±1.2%              ∞
 16    Input power and SAR drift measurement               ±4.7 %       Normal          1         1          1        ±4.7 %             ±4.7 %             ∞
  C    Phantom and Setup
 17    Phantom uncertainty                                 ±4.0 %     Rectangular      √3          1          1      ±2.3 %              ±2.3 %             ∞
 18    Liquid conductivity (target)                        ±5.0 %     Rectangular      √3        0.64       0.43     ±1.8 %              ±1.2 %             ∞
 19    Liquid conductivity (meas.)                         ±3.0 %       Normal          1        0.64       0.43     ±1.9 %              ±1.3 %             ∞
 20    Liquid permittivity (target)                        ±5.0 %     Rectangular      √3         0.6       0.49     ±1.7 %              ±1.4 %             ∞
 21    Liquid permittivity (meas.)                         ±3.2 %       Normal          1         0.6       0.49     ±1.9 %              ±1.6 %             ∞
       Combined Standard Uncertainty                                                                                 ±12.1 %            ±11.9 %             ∞
       Expanded Uncertainty (k=2)                                                                                    ±24.2 %            ±23.7 %
*.    This measurement uncertainty budget is suggested by Schmid & Partner Engineering AG. [6]




UL Japan, Inc.
Shonan EMC Lab.
1-22-3 Megumigaoka, Hiratsuka-shi, Kanagawa-ken, 259-1220 JAPAN
Telephone: +81 463 50 6400 / Facsimile: +81 463 50 6401



Document Created: 2011-01-31 11:54:14
Document Modified: 2011-01-31 11:54:14

© 2025 FCC.report
This site is not affiliated with or endorsed by the FCC