SAR dipole certificate

FCC ID: V65S1360

RF Exposure Info

Download: PDF
FCCID_2006307

                                                              Applicant Kyocera
                                                               FCC ID:: V65S1360
                                                              Report #: CT-S1360-9D-0513-R0

EXHIBIT 9 APPENDIX D: SAR DIPOLE CALIBRATION CERTIFICATE

Total pages including cover page = 9

# 5d016




© 2013 Comptest Services LLC                          Page 1 of 1                                SAR Dipole Cert
    This report shall not be reproduced except in full, without the written consent of CompTest Services LLC.


Calibration Laboratory of                                                                                   S       Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                    Service suisse d‘étalonnage
  Engineering AG                                                                                            C       Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                S swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       Kyocera USA                                                                        Certificate No: D1900V2—5d016_Sep12

[CALIBRATION CERTIFICATE
 Object                                D1900V2 — SN: 54016


 Calibration procedure(s)              QA CAL—05.v8
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     September 14, 2012




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                       ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power mater EPM—442A                    GB37480704                05—Oct—11 (No. 217—01451)                             Oct—12
 Power sensor HP 8481A                   US37202783                05—Oct—11 (No. 217—01451)                             Oct—12
 Reference 20 dB Attenuator              SN: 5058 (20k)            27—Mar—12 (No. 217—01530)                             Apr—13
 Type—N mismatch combination             SN: 5047.2 / 06327        27—Mar—12 (No. 217—01533)                             Apr—13
 Reference Probe ES3DV3                  SN: 3205                  80—Dec—11 (No. ES3—3205_Dec11)                        Dec—12
  DAE4                                   SN: 601                   27—Jun—12 (No. DAE4—601_Jun12)                        Jun—18

  Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
  Power sensor HP 8481A                  MY41092317                18—0ct—02 (in house check Oct—11)                     In house check: Oct—13
  RF generator R&S SMT—06                100005                    04—Aug—99 (in house check Oct—11)                     In house check: Oct—13
  Network Analyzer HP 8753E              US37390585 $4206          18—Oct—01 (in house check Oct—11)                     In house check: Oct—12




                                                                                                                          o (ULEE
                                         Name                                   Function                                  Signature
  Calibrated by:                        Jeton Kastrati                          Laboratory Technician



  Approved by:                           KatJa Pokovie                          Technical Manager



                                                                                                                         Issued: September 14, 2012
  This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


 Certificate No: D1900V2—5d016_Sep12                                   Page 1 of 8


Calibration
      s     Laboratory of                                     sllly95
                                                            a \Q\/ /s          §   Schweizerischer Kalibrierdienst
Schmid & Partner                                            i4                 c   Service suisse d‘étalonnage
  Engineering AG                                            PCR                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 */7/_\"\\\\@“     S   swiss Calibration Service
                                                                tlhil ls®


Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", December 2003
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
   c) Federal Communications Commission Office of Engineering & Technology (FCC OET),
       "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
       Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
       Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
       Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
     d)   DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
          Measurement Conditions: Further details are available from the Validation Report at the end
          of the certificate. All figures stated in the certificate are valid at the frequency indicated.
          Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          paralle! to the body axis.
          Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
          Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
          SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
          SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.



 Certificate No: D1900V2—5d016_Sep12                             Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                               V52.8.2
     Extrapolation                                   Advanced Extrapolation
     Phantom                                         Modular Flat Phantom

     Distance Dipole Center— TSL                               10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         1900 MHz & 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C           40.0                1.40 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   40.6 £ 6 %          1.37 mho/m x 6 %
     Head TSL temperature change during test                      «0.5 °C            ———


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

     SAR measured                                        250 mW input power                   9.79 mW /g
     SAR for nominal Head TSL parameters                     normalized to 1W        39.8 mW /g + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                   5.19 mW /g
     SAR for nominal Head TSL parameters                     normalized to 1W        21.0 mW /g # 16.5 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Body TSL parameters                                  22.0 °C            53.3               1.52 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C    52.5 +6 %          1.54 mho/m £ 6 %
     Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition

     SAR measured                                        250 mW input power                     10.4 mW / g
     SAR for nominal Body TSL parameters                     normalized to 1W        411 mW / g # 17.0 %(k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

     SAR measured                                        250 mW input power                     5.48 mW / g
     SAR for nominal Body TSL parameters                     normalized to 1W        21.8 mW / g # 16.5 % (k=2)




Certificate No: D1900V2—5d016_Sep12                    Page 3 of 8


Appendix

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           47.2 9 + 4.5 |Q
      Return Loss                                                                       —25.2 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           46.8 0 + 4.7 |Q
      Return Loss                                                                      —24.6 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.196 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph, The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 June 04, 2002




Certificate No: D1900V2—5d016_Sep12                       Page 4 of 8


DASY5 Validation Report for Head TSL
                                                                                   Date: 14.09.2012

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 5d016
Communication System: CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.37 mho/m; &, = 40.6; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2007)

DASY52 Configuration:

    e   Probe: ES3DV3 — SN3205; ConvF(5.01, 5.01, 5.01); Calibrated: 30.12.2011;

    e   Sensor—Surface: 3mm (Mechanical Surface Detection)

    e   Electronics: DAEB4 Sn601; Calibrated: 27.06.2012

    +   Phantom: Flat Phantom 5.0 (front); Type: QDOOOPS5OAA; Serial: 1001

    e   DASY52 52.8.2(969); SEMCAD X 14.6.6(6824)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.784 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 17.342 mW/g
SAR(1 g) = 9.79 mW/g; SAR(10 g) = 5.19 mW/g
Maximum value of SAR (measured) = 12.0 W/kg
               dB
                0



               —3.40



               —6.80



               ~10.20



                13.60



                ~17.00
                           0 dB = 12.0 W/kg =21.58 dB W/kg




 Certificate No: D1900V2—5d016_Sep12             Page 5 of 8


Impedance Measurement Plot for Head TSL



                                                                14 Sep 2012     @8:33144
         CH® sii     a U rs           1: 47.150 a   4.4863 n 375.80 pH         1 900.000 000 MHz
                                                    —~——r .



         De 1




         hy
         16"

         H1d

         CH2




         Cor




         Ay
         16


         H1d

                ~~_START 1 700.000 o0 Mz                                 SToP 2 108.000 0BB MHz




Certificate No: D1900V2—5d016_Sep12                   Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                   Date: 14.09.2012

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50016

Communication System: CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.54 mho/m; & = 52.5; p= 1000 kg/mJ
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)

DASY52 Configuration:

   e    Probe: ES3DV3 — SN3205; ConvF(4.62, 4.62, 4.62); Calibrated: 30.12.2011;

   e    Sensor—Surface: 3mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601; Calibrated: 27.06.2012

   *    Phantom: Flat Phantom 5.0 (back); Type: QDOOOPS50AA; Serial: 1002

   «_   DASY52 52.8.2(969); SEMCAD X 14.6.6(6824)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.784 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 18.122 mW/g
SAR(I g) = 10.4 mW/g; SAR(10 g) = 5.48 mW/g
Maximum value of SAR (measured) = 13.0 W/kg




                           339




                           1016



                            —13.54



                            416.93

                                     0 dB = 13.0 W/kg = 22.28 dB W/kg




Certificate No: D1900V2—5d016_Sep12                 Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                  14 Sep 2012    ©B:44:25
         CH] sai     i uo ors         41 46.771 a   4.6992 0. 392.69 pH         1.900,000 000 MHz

         *                                   *            L           c
         Ded                                                      .
         Cor




         Av
         165


         Hid

         CH2




         Cor




         Ay
         16


         H1d                          +
                   START 1 708.000 000 MHz          m                     STOP 2 109.000 aao MHz




Certificate No: D1900V2—5d016_Sep12                     Page 8 of 8