sar data

FCC ID: V65OASY1

RF Exposure Info

Download: PDF
FCCID_1594461

         The Testcenter facility ‘Dosimetric Test Lab’ within IMST GmbH is accredited by the German National ‘Deutsche
Akkreditierungsstelle GmbH (DAkkS)’ for testing according to the scope as listed in the accreditation certificate: D-PL-12139-01-01.




                                                            Appendix for the Report
                                          Dosimetric Assessment of the
                                           Portable Device Kyocera F42
                                                   (FCC ID: V65OASY1)
                        According to the FCC Requirements

                                                                                    Calibration Data
                                                                                            November 25, 2011


                                                                                                          IMST GmbH
                                                                               Carl-Friedrich-Gauß-Str. 2
                                                                                      D-47475 Kamp-Lintfort

                                                                                                               Customer
                                                                                     KYOCERA Corporation
                                                                           2-1-1 Kagahara, Tsuzuki-ku,
                                                                                  224-8502 Yokohama-shi
                                                                                                                        Japan




SAR_Report_7layers_6620_872_FCC_CalData_F42
























          The Testcenter facility ‘Dosimetric Test Lab’ within IMST GmbH is accredited by the German National ‘Deutsche
 Akkreditierungsstelle GmbH (DAkkS)’ for testing according to the scope as listed in the accreditation certificate: D-PL-12139-01-01.


                                              Calibration Certificate

    Certificate No:                           Cal_D1900V2_SN5d051_Sep2011

    Object:                                   D1900V2 SN: 5d051

    Date of Calibration:                      September 30, 2011

    Next Calibration:                         September 2013

    Object Condition:                         In Tolerance

    Calibration Equipment used:
                                                                           Last                                          Next
    Test Equipment                           Serial Number                                    Calibrated by
                                                                        calibration                                   calibration
                                                                                               Agilent Techn.
   Powermeter E4416A                          GB41050414                   Dec 10            (ISO/IEC 17025,             Dec 12
                                                                                             1-1784162174-1)
                                                                                               Agilent Techn.
   Power Sensor E9301H                        US40010212                   Dec 10            (ISO/IEC 17025,             Dec 12
                                                                                             1-1784041195-1)
                                                                                               Agilent Techn.
   Powermeter E4417A                          GB41050441                   Dec 10            (ISO/IEC 17025,             Dec 12
                                                                                             1-1674038198-1)
                                                                                               Agilent Techn.
   Power Sensor E9301A                        MY41495584                   Dec 10            (ISO/IEC 17025,             Dec 12
                                                                                             1-1784041307-1)
                                                                                             Rohde& Schwarz
   Network Analyzer E5071C                    MY46103220                   Aug 11          (14967-DKD-00201-             Aug 12
                                                                                                 2009-08)
                                                                                             SPEAG, No ET3-
   Reference Probe ET3DV6                        SN 1669                   Feb 11               1669_Feb09
                                                                                                                         Feb 12

                                                                                            SPEAG, No DAE3-
   DAE3                                           SN 335                   Feb 11              335_Feb09
                                                                                                                         Feb 12




Cal_D1900V2_SN5d051_Sep2011                                                                                             Page 1 of 8


Cal_D1900V2_SN5d051_Sep2011                                                            Page 2 of 8

Calibration is performed according the following standards:

                                        IEEE 1528-2003
"IEEE Recommended Practice for Determining the Peak Spatial - Averaged Specific Absorption
Rate (SAR) in the Human Head from Wireless Communication Devices: Measurement
Technique", December 2003
                                          EN 62209-1
"Procedure to measure the Specific Absorption Rate (SAR) for hand - held devices used in close
proximity to the ear (frequency range of 300 MHz to 3GHz)", March 2007
   Federal Communications Commission Office of Engineering & Technologies (FCCOET)
"Evaluating Compliance wit FCC Guidelines for Human Exposure to Radiofrequency
Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and Portable
Devices with FCC Limits for Human Exposure to Radiofrequency Emissions", Supplement C
(Edition 01-01) to Bulletin 65
Additional Documentation:      DASY 4 System Handbook



prepared by:                                         reviewed by:
                 Alexander Rahn                                      André van den Bosch
                 test engineer                                       quality assurance engineer

08/ABo


Cal_D1900V2_SN5d051_Sep2011                                                           Page 3 of 8

                                     Measurement Conditions

DASY Version:                       Dasy 4;                   V4.7

Phantom:                            SAM Phantom               1340

Distance Dipole Center – TSL:       10mm                      With spacer

Zoom Scan res.                      dx, dy, dz = 5mm

Frequency:                          1900 MHz ± 1MHz




                                       Head TSL Parameters
                                       Temperature          Permittivity       Conductivity
Nominal Head TSL Parameters               22.0                  40.0                1.40
Measured Head TSL Parameters              22.8               39.4 ± 6%         1.43 S/m ± 6%




                                    SAR Result with Head TSL
            SAR measured                          250mW input power          9.53 mW/g
 Averaged
  over 1g




            SAR normalized                         normalized to 1W          38.12 mW/g

            SAR for nominal Head TSL parameters    normalized to 1W   37.51 mW/g ± 16.5 % (k=2)

            SAR measured                          250mW input power          4.93 mW/g
 Averaged
 over 10g




            SAR normalized                         normalized to 1W          19.72 mW/g

            SAR for nominal Head TSL parameters    normalized to 1W   19.52 mW/g ± 16.5 % (k=2)


Cal_D1900V2_SN5d051_Sep2011                                                                Page 4 of 8

                                         Body TSL Parameters
                                          Temperature           Permittivity        Conductivity
Nominal Body TSL Parameters                  22.0                  53.30                 1.52
Measured Body TSL Parameters                 23.3               52.10 ± 6%          1.54 S/m ± 6%




                                       SAR Result with Body TSL
            SAR measured                            250mW input power              9.66 mW/g
 Averaged
  over 1g




            SAR normalized                           normalized to 1W             38.64 mW/g

            SAR for nominal Body TSL parameters      normalized to 1W    38.14 mW/g ± 16.5 % (k=2)

            SAR measured                            250mW input power              5.10 mW/g
 Averaged
 over 10g




            SAR normalized                           normalized to 1W             20.40 mW/g

            SAR for nominal Body TSL parameters      normalized to 1W    20.23 mW/g ± 16.5 % (k=2)




                                       General Antenna Parmeters
                                           Impedance, transformed to feed point     48.4 jΩ - 0.7 jΩ
Antenna Parameters with Head TSL
                                                       Return Loss                      -35.2 dB
                                           Impedance, transformed to feed point     52.8 jΩ - 1.8 jΩ
 Antenna Parameter with Body TSL
                                                       Return Loss                      -29.9 dB

After long term use with 100W radiated power, only a slight warming of the dipole near the
feedpoint can be measured. The dipole is made of standard semigrid coaxial cable. The center
conductor of the feeding line is directly connected to the second arm of the dipole. The antenna is
therefore short-circuited for DC signals.


                                          Additional EUT Data
                    Manufactured by:                                      SPEAG
                    Manufactured on:                                 January 15, 1998


Cal_D1900V2_SN5d051_Sep2011                                                         Page 5 of 8

                                    SAR Result with Head TSL


Test Laboratory: Imst GmbH, DASY Yellow (II); File Name: 300911_y_1669.da4

DUT: Dipole 1900 MHz SN: 5d051; Type: D1900V2; Serial: D1900V2 - SN5d051
Program Name: System Performance Check at 1900 MHz

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.43 mho/m; εr = 39.4; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:
- Probe: ET3DV6R - SN1669; ConvF(5.12, 5.12, 5.12); Calibrated: 21.02.2011
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn335; Calibrated: 22.02.2011
- Phantom: SAM Glycol 1340; Type: QD 000 P40 CB; Serial: TP-1340
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 186

d=10mm, Pin=250mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 10.4 mW/g
d=10mm, Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 90.3 V/m; Power Drift = -0.025 dB
Peak SAR (extrapolated) = 17.0 W/kg
SAR(1 g) = 9.53 mW/g; SAR(10 g) = 4.93 mW/g
Maximum value of SAR (measured) = 10.7 mW/g


Cal_D1900V2_SN5d051_Sep2011                                                         Page 6 of 8

                                     SAR Result with Body TSL

Test Laboratory: Imst GmbH, DASY Yellow (II); File Name: 300911_y_1669.da4

DUT: Dipole 1900 MHz SN: 5d051; Type: D1900V2; Serial: D1900V2 - SN5d051
Program Name: System Performance Check at 1900 MHz

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium parameters used: f = 1900 MHz; σ = 1.54 mho/m; εr = 52.1; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:
- Probe: ET3DV6R - SN1669; ConvF(4.54, 4.54, 4.54); Calibrated: 21.02.2011
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn335; Calibrated: 22.02.2011
- Phantom: SAM Glycol 1340; Type: QD 000 P40 CB; Serial: TP-1340
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 186

d=10mm, Pin=250mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 10.6 mW/g
d=10mm, Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 89.6 V/m; Power Drift = 0.004 dB
Peak SAR (extrapolated) = 16.3 W/kg
SAR(1 g) = 9.66 mW/g; SAR(10 g) = 5.1 mW/g
Maximum value of SAR (measured) = 11.0 mW/g


Cal_D1900V2_SN5d051_Sep2011                                 Page 7 of 8

                 Impedance Measurements Plot for Head TSL


Cal_D1900V2_SN5d051_Sep2011                                  Page 8 of 8

                  Impedance Measurements Plot for Body TSL


                                                       DAT-P-152/98-01

                                                                                    Calibration Certificate

                                                          Certificate No:           Cal_D2450V2_SN709_1209

                                                          Object:                   D2450V2 SN: 709

                                                          Date of Calibration:      December 09, 2009

                                                          Next Calibration:         December 2011

                                                          Object Condition:         In Tolerance

                                                          Calibration Equipment used:
                                                                                                         Last                               Next
                                                          Test Equipment           Serial Number                      Calibrated by
                                                                                                      calibration                        calibration
                                                                                                                        Agilent Techn.
                                                         Powermeter E4416A          GB41050414          Dec 08        (ISO/IEC 17025,      Dec 10
                                                                                                                      1-1784162174-1)
                                                                                                                        Agilent Techn.
                                                         Power Sensor E9301H        US40010212          Dec 08        (ISO/IEC 17025,      Dec 10
                                                                                                                      1-1784041195-1)
                                                                                                                        Agilent Techn.
                                                         Powermeter E4417A          GB41050441          Dec 08        (ISO/IEC 17025,      Dec 10
                                                                                                                      1-1674038198-1)
                                                                                                                        Agilent Techn.
                                                         Power Sensor E9301A        MY41495584          Dec 08        (ISO/IEC 17025,      Dec 10
                                                                                                                      1-1784041307-1)
                                                                                                                      Rohde& Schwarz
                                                         Network Analyzer E5071C    MY46103220          Aug 09      (14967-DKD-00201-      Aug 10
                                                                                                                          2009-08)
                                                                                                                      SPEAG, No EX-
                                                         Reference Probe EX3DV4         SN 3536         Sep 09           3536_Sep09
                                                                                                                                           Sep 10

                                                                                                                    SPEAG, No DAE4-
                                                         DAE4                           SN 661          Sep 09                             Sep 10
_ Calibration_Certificate_Dipole_V1.0/01.08.2009/ABo




                                                                                                                       661_Sep09




                                                       Cal_D2450V2_SN709_1209                                                             Page 1 of 8


Cal_D2450V2_SN709_1209                                                                 Page 2 of 8



Calibration is performed according the following standards:

                                     IEEE 1528-2003
"IEEE Recommended Practice for Determining the Peak Spatial - Averaged Specific Absorption
Rate (SAR) in the Human Head from Wireless Communication Devices: Measurement
Technique", December 2003
                                          IEC 62209-1
"Procedure to measure the Specific Absorption Rate (SAR) for hand - held devices used in close
proximity to the ear (frequency range of 300 MHz to 3GHz)", February 2005
   Federal Communications Commission Office of Engineering & Technologies (FCCOET)
"Evaluating Compliance wit FCC Guidelines for Human Exposure to Radiofrequency
Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and Portable
Devices with FCC Limits for Human Exposure to Radiofrequency Emissions", Supplement C
(Edition 01-01) to Bulletin 65
Additional Documentation:      DASY 4 System Handbook



prepared by:                                         reviewed by:
                 Alexander Rahn                                      André van den Bosch
                 test engineer                                       quality assurance engineer

08/ABo


Cal_D2450V2_SN709_1209                                                                      Page 3 of 8



                                          Measurement Conditions

DASY Version:                          Dasy 4;                    V4.7

Phantom:                               SAM Phantom                1341

Distance Dipole Center – TSL: 10mm                                With spacer

Zoom Scan res.                         dx, dy, dz = 5mm

Frequency:                             2450 MHz ± 1MHz




                                            Head TSL Parameters
                                             Temperature             Permittivity       Conductivity
Nominal Head TSL Parameters                     22.0                    39.20              1.80
Measured Head TSL
                                                 22.0                 40.2 ± 6%         1.84 S/m ± 6%
Parameters




                                          SAR result with Head TSL
                 SAR measured                             250mW input power           13.60 mW/g
 Averaged over




                 SAR normalized                            normalized to 1W           54.40 mW/g
      1g




                 SAR for nominal Head TSL parameters       normalized to 1W       54.58 mW/g ± 16.5 %
                                                                                         k=2)
                 SAR measured                             250mW input power           6.16 mW/g
 Averaged over




                 SAR normalized                            normalized to 1W           24.64 mW/g
      10g




                 SAR for nominal Head TSL parameters       normalized to 1W       24.78 mW/g ± 16.5 %
                                                                                         (k=2)


Cal_D2450V2_SN709_1209                                                                           Page 4 of 8



                                              Body TSL Parameters
                                               Temperature              Permittivity         Conductivity
Nominal Body TSL Parameters                       22.0                     52.70                1.95
Measured Body TSL
                                                   22.0                  51.70 ± 6%          2.00 S/m ± 6%
Parameters




                                            SAR result with Body TSL
                 SAR measured                                 250mW input power            13.20 mW/g
 Averaged over




                 SAR normalized                                normalized to 1W            52.80 mW/g
      1g




                                                                                   51.76 mW/g ± 16.5 %
                 SAR for nominal Body TSL parameters           normalized to 1W
                                                                                          (k=2)
                 SAR measured                                 250mW input power             6.01 mW/g
 Averaged over




                 SAR normalized                                normalized to 1W            24.04 mW/g
      10g




                                                                                   23.81 mW/g ± 16.5 %
                 SAR for nominal Body TSL parameters           normalized to 1W
                                                                                          (k=2)



                                            General Antenna Parmeters
Antenna Parameters with Head                Impedance, transformed to feed point       49.7 jΩ - 1.23 jΩ
TSL                                                       Return Loss                       -37.97 dB
Antenna Parameter with Body                 Impedance, transformed to feed point       50.8 jΩ - 1.27 jΩ
TSL                                                       Return Loss                       -36.55 dB
After long term use with 100W radiated power, only a slight warming of the dipole near the
feedpoint can be measured. The dipole is made of standard semigrid coaxial cable. The center
conductor of the feeding line is directly connected to the second arm of the dipole. The antenna is
therefore short-circuited for DC signals.



                                               Additional EUT Data
                         Manufactured by:                                     SPEAG
                         Manufactured on:                                   July 5, 2002


Cal_D2450V2_SN709_1209                                                              Page 5 of 8



                                    SAR result with Head TSL

Test Laboratory: Imst GmbH, DASY Yellow (II); File Name: 091209_y_3536.da4

DUT: Dipole 2450 MHz SN: 709; Type: D2450V2; Serial: D2450V2 - SN:709
Program Name: System Performance Check at 2450 MHz

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium parameters used: f = 2450 MHz; σ = 1.84 mho/m; εr = 40.2; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:
- Probe: EX3DV4 - SN3536; ConvF(7.59, 7.59, 7.59); Calibrated: 18.09.2009
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn631; Calibrated: 14.09.2009
- Phantom: SAM Glycol 1340; Type: QD 000 P40 CB; Serial: TP-1340
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 186

d=10mm, Pin=250mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 15.3 mW/g
d=10mm, Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 91.4 V/m; Power Drift = 0.016 dB
Peak SAR (extrapolated) = 29.6 W/kg
SAR(1 g) = 13.6 mW/g; SAR(10 g) = 6.16 mW/g
Maximum value of SAR (measured) = 15.2 mW/g


Cal_D2450V2_SN709_1209                                                              Page 6 of 8

                                     SAR result with Body TSL

Test Laboratory: Imst GmbH, DASY Yellow (II); File Name: 081209_y_3536.da4

DUT: Dipole 2450 MHz SN: 709; Type: D2450V2; Serial: D2450V2 - SN:709
Program Name: System Performance Check at 2450 MHz

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium parameters used: f = 2450 MHz; σ = 2 mho/m; εr = 51.7; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY4 Configuration:
- Probe: EX3DV4 - SN3536; ConvF(7.57, 7.57, 7.57); Calibrated: 18.09.2009
- Sensor-Surface: 4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn631; Calibrated: 14.09.2009
- Phantom: SAM Glycol 1340; Type: QD 000 P40 CB; Serial: TP-1340
- Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 186

d=10mm, Pin=250mW/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 15.0 mW/g
d=10mm, Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 86.9 V/m; Power Drift = 0.016 dB
Peak SAR (extrapolated) = 28.0 W/kg
SAR(1 g) = 13.2 mW/g; SAR(10 g) = 6.01 mW/g
Maximum value of SAR (measured) = 15.1 mW/g


Cal_D2450V2_SN709_1209                                      Page 7 of 8

                 Impedance Measurements Plot for Head TSL


Cal_D2450V2_SN709_1209                                       Page 8 of 8

                  Impedance Measurements Plot for Body TSL