SAR report part 2

FCC ID: V5PD200V4

RF Exposure Info

Download: PDF
FCCID_3223792

Calibration Laboratory of                                    \\‘\Q//"/            Schweizerischer Kalibrierdienst
Schmid & Partner                                            iia?-%m&              Service suisse d‘étalonnage
  Engineering AG                                                           +
                                                                           x
                                                                                  Servizio svizzero di taratura
                                                                                     x      s    x       a
Zeughausstrasse 43, 8004 Zurich, Switzerland                 xL—Soy$
                                                                   w              Swiss Calibration Service
                                                                ulutuls®


Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates


Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
    e     DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
         comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    e    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
               DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
               Common mode sensitivity: Influence of a positive or negative common mode voltage on
               the differential measurement.
               Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
               AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
               Input Offset Measurement. Output voltage and statistical results over a large number of
               zero voltage measurements.
               Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.
               Input resistance: Typical value for information: DAE input resistance at the connector,
               during internal auto—zeroing and during measurement.
               Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
               Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE4—1386_Jul16                                 Page 2 of 5


DC Voltage Measurement
    A/D — Converter Resolution —nominal
        High Range:          1LSB =     6.4uV ,       full range =. —100...+300 mV
        Low Range:           1LSB =      6inV ,       full range = —1....... +3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors             KX                           ¥                   Z
      High Range             404.512 £.0.02% (k=2)   404.603 + 0.02% (k=2)   ‘404.126 + 0.02% (k=2)

      Low Range              4.02176 + 1.50% (k=2)   4.01405 + 1.50% (k=2)   4.01423 £ 1.50% (k=2)



Connector Angle

      Connector Angle to be used in DASY system                                   203.5 °+1 °




Certificate No: DAE4—1386_Jul16                      Page 3 of 5


Appendix (Additional assessments outside the scope of SCS0108)
1. DC Voltage Linearity
      High Range                                 Reading (uV)         Difference (uV)             Error (%)
      Channel X        +Input                     199996.01                   —0.60               —0.00
      Channel X        + Input                     20002.59                   0.57                0.00
      Channel X        — Input                    —19998.67                   1.71                >0.01
      Channel Y¥       + Input                    199996.62                   ~0.22               —0.00
      Channel Y¥       +input                      20000.89                   —1.13               —0.01
      Channel Y        —Input                     —20000.90                   —0.54               0.00
      Channel Z        + Input                    199997.61                   0.99                0.00
      Channel Z        + Input                     19999.51                   —2.37               —0.01
      Channel Z        —Input                     —20001.96                   —1.40               0.01


      Low Range                                  Reading (uV)         Difference (uV)             Error (%)
      Channel X        + Input                     2002.92                    0.82                0.04
      Channel X        + Input                      202.39                    0.07                0.03
      Channel X        — Input                     —197.72                    —0.16               0.08
      Channel Y¥       + input                     2002.74                    0.95                0.05
      Channel Y        + Input                      201.78                    —0.26               —0.13
      Channel Y        —Input                      —198.29                    —0.48               0.24
      Channel Z        + Iinput                    2002.23                    0.55                0.03
      Channel Z        + Input                      200.66                    —~1.40              —0.69

      Channel Z        — Input                     —199.39                    ~1.51               0.76



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                         Common mode                        High Range                         Low Range
                         Input Voltage (mV)             Average Reading (uV)              Average Reading (uV)

     Channel X                            200                      ~15.11                      —17.01
                                         — 200                     18.64                        16.95
     Channel Y                            200                      —8.25                        —9.19
                                         — 200                      8.39                        8.13
     Channel Z                            200                      —7.20                        —6.34
                                         — 200                      3.33                        3.76



3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                         Input Voltage (mV)         Channel X (uV)          Channel Y (uV)    Channel Z (uV)

     Channel X                     200                        —                  5.13              —3.37
     Channel Y                     200                   8.77                         —            5.77
     Channel Z                     200                   8.45                    7.08                   —


Certificate No: DAE4—1386_Juli16                              Page 4 of 5


4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                    High Range (LSB)                    Low Range (LSB)

     Channel X                                              16014                            14906

     Channel Y                                              16071                            16958

    Channel Z                                                 16059                          12850



5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10MQ

                            Average (uV)        min. Offset (uV)       max. Offset (uV)      Std. l?:‘\l/;ation

     Channel X                    0.02                —0.91                 0.66                    0.27

    Channel Y                     —0.17               ~2.00                 0.82                     0.39
     Channel Z                    —3.10               —3.84                 —1.49                   0.38


6. Input Offset Current
     Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                  Zeroing (kOhm)                       Measuring {(MOhm)
     Channel X                                             200                               200

     Channel Y                                             200                               200
     Channel Z                                             200                               200



8. Low Battery Alarm Voltage (Typical values for information)
    Typical values                                 Alarm Level (VDC)
     Supply (+ Veoc)                                                            +7.9
     Supply (— Veo)                                                             —7.6



9. Power Consumption (Typical values for information)
     Typical values                         Switched off (mA)         Stand by (mA)       Transmitting (mA)

     Supply {(+ Vee)                               +0.01                   +6                      +14
     Supply (— Veoc)                               —0.01                   —8                      —9




Certificate No: DAE4—1386_Jult6                            Page 5 of 5


                                                                                             6(9»,
                                                  in Cotaberaton nin                       «_ d wBwet
                                                                                           st
                                 TTL 2.3
                                     5 —p—e %cm«s"“
                                                  causmanon Lasoratory                                                  cxusmaon
                        Add NoiS1 Xuevizn Rod, in Diavic Helin, 10019 Ching                   ..m.        eP Sisltin
                        Thcersenomnart           recasioentinam
                        Enal: ctl@dimtlcon            Uite/ineihintin
                 Clent             Auden                                          Cartificate No: 216—97054

            CALIBRATION CERTIFICATE

|           ovieet                                  exsove — siasre
|           Gattration Procecure(s)                 meenien
                                                    Callbration Procedures for Dosimetrc E—feld Probes
            Callbration date                        May 11, 2016
            This callratin Gertficate dacuments the traceabiity to natlonal standards, which realize the physical uns of
            measurements(S). The measurements and the uncertaintes with confidence probabilty are given on the folowing
            pages and are part of the certficate
            All caltbratons have been conducted in the clased laboratory faciiy: environment temperatire@zs3)e and
            humidity=70%

            Galioraion Equiament used (M&TE crtical fr calbration)
            Prmary Sandaras                   D#             Gal Date(Galbraled byGenticate No)            Scheduled Gatbraton
             Power Meter NRPZ              101010              0i—Ju—15 (CTTL, No JisXOi250)               Jun—t5
             Power sensor      NRP—201.    101547              O1—Ju—15 (GTTL, No J16X04258)               Jun—16
    |        Power sensor      NRP—201      101548             01—Ju—15 (CTTL, No JtsX0ez5e)               Jun—t6
        |    Referencet0dBAttenator 18N60W—1008 13—Mar—16(CTTLNo J1éx01547)                                Marts
             Reference20dBAtenuator 1BN5OW—204B 13—Mar—18(CTTL, NoJ1GX01548)                               Marto
             Reference Probe EX3DVé        SN 3617             26—Aug—18(GPEAGNo EXG—3017_Aug1)            Aug—t
             ones                          sN 13               21—Jan—16(SPEAG, NoDAEA—1381_Jant) Jan 17
             Secondary Standards            ID#                Cal Date(Calibrated by, Gertfate No.)      Scheduled Calbration
             SignalGeneratorMG3700A        6201052605          01—Ju18 (CTTL, No.J1$X04258)                Jun—16
             Network Analyzer ESO71G | MY410G73                26—an—16 (CTTL, NoJ1EX00804)                dan 17
                                     Name                          Function                                 Sgnaure
            Galibrated by                  va zZongying                SAR Test Engineer                   éfifie)
            Reviewed by                    l Dianyuen              SAR Project Leader                     o
                                                                                                          m i
            Aporoved by                    Lu Bingsong                 Depuly Director o the laboratory    i’g My
                                                                                        Issued: May 193, ois
            This calibration certicate shall not be reproduced except in fullwihout writen approval of the laboratory

                Certiicate No: 216—97054                           Page t of n


                             causrmmon tasoratory
       Add: No$1 Xoevan Road,tnianDiic, eling, 100191 China
             enemomnant           rcoreloemoisman
             ctactiantson        ifhowtintion
Glossary:
Ts                  tissue simulating liguid
NORtfcyz            senstily in free space
ComyF               sensitviy in TSL / NORMxx.e
bor                 lode compression point
or                  crestfactor(1/duty_cycle) of the RF signal
ABco                modulation dependent Incarization parameters
Polariation ®       @ rotation around probe axis
Polariation 0       8 rotation around an axisthat is i the plane normal to probe axis (at measurement center), i
                    8=0 is normal to probe axis
Connector Angle information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Sid 1828—2013, ‘IEEE Recommended Practice for Determining the Peak Spatial—Averaged
   Specitc Absorption Rate (SAR) in the Human Head from Wreless Communications Devices:
   Measurement Techniques®, June 2013
) IEC 62209—1, ‘Procedure o measure the Specifc Absorption Rate (SAR) for hand—held devices used
   in close proximiy to the ear (requency range of 300Mz to 3GHz), February 2005
c) 1EC 62200—2, "Procedure to determine the Specifc Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (fequency range of 30 Miz to 6 GHz)°, March
   2010
) KDB 8656B4, ‘SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
+ NORMicy;z: Assessed for E—feld polarization 0=0 (fS900MHz in TEN—cell f> 1800MFz: waveguide}.
     NORMxy;z are onl intermediate values, e.,the uncertainties of NORNy.z does not effect the
      E"—ffeld uncertainty inside TSL (see below ConvF)
+ NORMMxy.z NORM.y.z" frequency_response (see Frequency Response Chart. This
     Inearization is implemented in DASY4 Software versions laterthan 4.2.The uncertainy ofthe
     frequency response i included in the stated uncertainty of Conve.
* DOPryz: DCP are numericalIinearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor madia.
+ PAR: PAR is the Peak to Average Ratio that is not callbrated but determined based on the signal
     characteristics.
* Auiz Beyce; Coyce:VRy:2:A,8.0 are numericalInearization parameters assessed based on the
     data of power sweep for specifc modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum callration range expressed in RMS vaitage across the diode
* ConvE and Boundary Effect Paramefers: Assessed in flat phantom using E—field (or Temporature
     Transfer Standard for 800MH) and inside waveguide using analytcalfeld distrbutions based on
     power measurements forf »800MFiz. The same setups are used for assessment of the parameters
     applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given
     These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
     The sensitity in TSL corresponds to NORMy,z" Conv: whereby the uncertainty cortesponds to
     that given for ConvF. A fequency dependent GonvE is used in DASY version 4.4 and higher which
     allows extending the valdty from:50VHz tos100Miz.
+ Spherical sotropy (3D deviation from isotrapy): n feld oflow gradients realized using a fat
     phantom exposed by a patch antenna.
+ Sensor Offset:The sensor offset corresponds to the offset of viual measurement center from the
      probe ti(on probe axis) No tolerance required
+ ConnectorAngle: The angle is assessed using the information gained by determining the NORM«
     (no uncertainty required)

Centeate No: 210—97084                       Page2of 11


                                  In C Ileb<! atio wi

         - TTL!....J?                            e        a      9_ _
                                  CAUBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86­10­62304633­2218      Fax: +86­10­62304633­2209
       E­mail: cttl@chinattl.com      Http://www.chinattl.cn




                      Probe EX3DV4


                                            SN: 3578


                                        Calibrated: May 11, 2016

                              Calibrated for DASY/EASY Systems
                                      (Note: non­compatible with DASY2 system!)




Certificate No: Z16­97054                               Page 3 of 11


                             in Cotaberatonw
              TTL a
                             causrnon tasoratory
        Add No.S1 uyon Rond, HaidanDi Reling, 100191 China
        ece doamoinantd           recoveimidieisante
        E—malictladhinadcon       Hinlicstinatles

     DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3578

Basic Calibration Parameters
                              SensorX           [ Sensor Y          SensorZ         Une (ke2)
| Norm(uV¥im})®               oat                o37                o44             +108%
 bCP(mV)®                     1054              | 1000              123


Modulation Calibration Parameters
 UI        | Communication            *A           Is         ¢           o     Ive             Une®
            System Name                  aB         aBipy                 a8      mV            (ke2)
 o         \ cw                  x       loo        Joo      |io          000     _|1837   _ 126%
                                 y       loo      loo        _li0                 1690     _
                                 z      joo       00          10                  1896     _\


     The reported uncertainty of measurement is stated as the standard uncertainly u«(
     Measurement mulipled by the coverage factor k=2, which for a normal distrbution
     Corresponds to a coverage probabiity of approximately 95%.

AThe unsertainties oNorm X, Y Z do not affect the E—feld uncertaintyInside TSL (see Page 5 and Page 6)
* Numerical Inearzation parameter uncertanty not required
* Uncertainy is determined using the max. devition from Iinear response agplying rectangular cistibution
and is expressed forthe sauare of tfeld value




Cenifiate No: 16—97084                         Pagecofnn


                                     In Collabo atio • ,

             =rTL!....J'                             e      a       9_ _
                                     CALIBRATION LABORATORY

           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86­10­62304633­2218      Fax: +86­10­62304633­2209
           E­mail: cttl@chinattl.com      Hllp:l/www.chinattl.cn



      DASY/EASY - Parameters of Probe: EX3DV4 - SN: 3578

Calibration Parameter Determined in Head Tissue Simulating Media
                  Relative        Conductivity                                                     G   Depth G   Unct.
 f [MHzf                                               ConvF X       ConvF Y     ConvF Z   Alpha
                Permittivity F        (81m) F                                                          (mm)      (k=2)
   750              41.9                0.89               9.65           9.65    9.65      0.40        0.80     ±12%
   835              41.5                0.90               9.48           9.48    9.48      0.13        1.68     ±12%
    900             41.5                0.97               9.40           9.40    9.40      0.20        1.28     ±12%
   1750             40.1                1.37               8.35           8.35    8.35      0.18        1.49     ±12%
   1900             40.0                1.40               7.99           7.99    7.99      0.18        1.50     ±12%
   2000             40.0                1.40               7.99           7.99    7.99      0.17        1.58     ±12%
   2300             39.5                1.67               7.77           7.77    7.77      0.40        0.82     ±12%
   2450             39.2                1.80               7.28           7.28    7.28      0.46        0.83     ±12%
   2600             39.0                1.96               7.21           7.21    7.21      0.49        0.81     ±12%
   3500             37.9                2.91               7.07           7.07    7.07      0.40        1.10     ±13%
   5200             36.0                4.66               5.42           5.42    5.42      0.40        1.50     ±13%
   5300             35.9                4.76               5.22           5.22    5.22      0.39        1.40     ±13%
   5500             35.6                4.96               4.99           4.99    4.99      0.40        1.40     ±13%
   5600             35.5                5.07               4.75           4.75    4.75      0.42        1.45     ±13%
   5800             35.3                5.27               4.82           4.82    4.82      0.45        1.45     ±13%



c Frequency validity of ±100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to ±50MHz. The
uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band.
FAt frequency below 3 GHz, the validity of tissue parameters (E and 0) can be relaxed to ±10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (E and 0) is
restricted to ±5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
G Alpha/Depth   are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than ± 1% for frequencies below 3 GHz and below ± 2% for the frequencies
between 3­6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z16­97054                               Page 5 of 11


                                    in Cotaberatonwn
                     TTL a
                         Causrimon tasoratom
              Add:NoS1 Xuuan Road,Haidin Disict Belin, 10019 Chia
              To oo nemienaerit        raeioenoienare
              Emil:otlachinatl on          Iiwncatine

          DASY/EASY —— Parameters of Probe: EX3DV4 — SN: 3578

Calibration Parameter Determined in Body Tissue Simulating Media
[                                 ‘Conductvity |                        T.                     Deptn® Unct.     |
     1 uk c                           (Sim)"       ConvF X ConvFY ConvE 2 Alpha e              {mem) _| (ee2)   )
       750                            o.96          osi |         sst   |    s           o4o    os5     412% |
       sas                            o7            ass           ass        ass | ons          is1 _| £12%
[_s00                                 105           947           oa7        o4r     | ons      144| £12%
       1750                           140           s02           s02        aor | ons          167|    +12% )
       1900                           182           7.78          778        77e     |   016    178     412% |
  2000                                152          70         770            770         ois 230 +12%
  2300                                ts1       _| Tss      | 7e             Tes         oss os0 £12%
[_2450                                ies       _| 7a1        7a1            7a1_|       on | osr a12%
  2600                                216          7i7        717            7iz         oso | aso +12% )
   3600                               a31          sar        |_6a7          ear         osi   110 | £13% )
  s200                                530          ass        _|_ass         a9s         oso   iso a13%
  saco                                sa2          a7s        a7s            a7s         oso   141| £13%
  5500                                ss           a40        440            440         oss   135 |£19%
  5eoo                                s77          430        a30            130|        oso   185 a19%
(_seoo                                600          a42        a42            «42         oss   146 a13% |

    © Frequency vaidty of +100z only applies or DAGY vi.4 and higher (Page 2), else i is resticted to £50MHz. The
    uncertiny is the RSS of GonvE uncertaityatcalbration frequency and the uncertaityfor th incicated requency band
    "Atfrequency below 3 GHz, the valdty f ssue parameters (and 0) can be relaxed to +10% i quid compensation
    formula is appled to measured SAR values. A frequencies above 3 GHiz,the valdty of ssue parameters (c and 0)is
    restrted to +5%. The uncertainy is tRSS ofthe Conv® uncertintyfrinicated targa touue parameters
    ®AlphaiDepth are determined durig caltration. SPEAG warrantsthatthe remaining deviation due t the baundary
    affect attr compensaton is alrays les than + 19 fo requencias below 3 GHiz and below + 2% for the feauencies
    between 3.6 GHiz at any ditance largerthan halfthe probe tp ciamete rom the boundary




       Certifeate No: Z16—97054                      Pagesof i1


                @ in Co»ubormnewm 5

                         causemon asoratomy
      Add NS Xuevin Rod, itan Divis Being,100191 Chima
      o iosasieenaais       reconioamoieisante
      Eonlachinatlcom       Horisnwsenhinenisn

                                                 Frequency Response of E—Field
                                                 (TEM—Cell: ifi110 EXX, Waveguide: R22)

                                         15
                                         14
       Frequency response (normalized)




                                         13
                                         12
                                         11
                                         10
                                         oe
                                         0.8

                                         0.7

                                         as

                                         a54            i            t          r        r           i       t
                                            o          soo         1000       1500     2000        2500    3000
                                                             E3              f [MHz]          z.
                                                             Tew               fHe]           *
                                               Uncertainty of Frequency Response of E—field: £7.5% (k=2)




Cenifiene No: 21697084                                                Pagerof i1


             infospomniznn
        <TTL a
             iemonastecon
      Add: NoS1 Xueymn Rod,Laitan Ditic, Being,100191 China
      Te ueocmosmant            reloenomnare
      E—mal:n@stinaticon        Hinthinectintion

                          Receiving Pattern (®), 0=0°

           1=600 MHz, TEM                                     f=1800 MHz, R22




      s«
  &$ un f.,.., iovrtT H omm n g—pugopiret                             eersemicg hivragcgss
   &
     «s
      ts                    4
               T            T             &            I          &          m          %
                                              .    Roll
                             100Mz            so0liz          1800btz    +   2800ubt!
                         Uncertainty of Axial Isotropy Assessment: £0.9% (k=2)




Cenifcate No: 16—97054                        Pagesof11


                            in Cotaboraton win
                  TTL s           p     e        a    q
                            causramon Lasorarofy
      Add: No1 Xveimn Road,Haitan Disc. Btin, 10019 China
      receteoamounani             recersnamoimane
      Enail ctl@dtiartcon         Hmihnosinats

                         Dynamic Range f(SARneaq)
                                    (TEM cell, £= 900 MHz)




             3
             $
             §
             ®
             %&
             £




                                                     SARImW/om‘]
                                       Eliinotcompensated           ~~ compensted



              1

        $ ow a uie nnt t geg t 0n the e e yeese6n"
        &&                                                * %
              +                                                 ®
                                                                     %
                                                                         ®
                              w                  "                   %              to     e
                                                 nrsAva\W/cm 1
                              ‘® —not compensated                            compensated
                         Uncertainty of Linearity Assessment: £0.9% (k=2)
CenifcatNo: 216—97034                        Pageoof11


         r\"L in cummemm §


      Add No S1 Xuevim Rod, aidan Divics Reing, 100191 China
      Tiaeioemoimart            receeivenbienarh
      Enallctl@chinalcon        tihnchination

                    Conversion Factor Assessment
    1=900 MHz, WGLS R9(H_convF)                              f=1750 MHz, WGLS R22(H_convF)
                                                        1|     aw




                        12
                        os
                        as
                         aa
                         or




                                to ow          m on o on on
                        Uncertainty of Spherical Isotropy Assessmer

Cenifea No: 216—97054                        Page ioi1


                                  in eonaoor!it"')
         _TTL L..P e a 9_ _
          ­­                     CAUBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86­10­62304633­2218      Fax: +86­10­62304633­2209
       E­mail: cttl@chinattl.com      Http://www.chinattl.cn



   DASY/EASY - Parameters of Probe: EX3DV4 - SN: 3578


    Other Probe Parameters
    Sensor Arrangement                                                     Triangular

    Connector Angle (0)                                                        165.9

    Mechanical Surface Detection Mode                                        enabled

    Optical Surface Detection Mode                                           disable

    Probe Overall Length                                                     337mm

    Probe Body Diameter                                                        10mm

    Tip Length                                                                  9mm

    Tip Diameter                                                              2.5mm

    Probe Tip to Sensor X Calibration Point                                     1mm

    Probe Tip to Sensor Y Calibration Point                                     1mm

    Probe Tip to Sensor Z Calibration Point                                     1mm

    Recommended Measurement Distance from Surface                             1.4mm




Certificate No: Z16­97054                             Page 11 of 11



Document Created: 2017-12-19 04:28:51
Document Modified: 2017-12-19 04:28:51

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC