SAR Test Report Part 4of4

FCC ID: V3H-B8

RF Exposure Info

Download: PDF
FCCID_913595

NCL Calibration Laboratories
Division of APREL Laboratories.




Sensitivity in Body Tissue

Frequency:                                                  1900 MHz

Epsilon:            53.3 (+/—5%)                            Sigma:              1.52 S/im (+/—5%)


ConvF

Channel X:          5.75

Channel Y:          5.75

Channel Z:          5.75

Tissue sensitivity values were calculated using the load impedance of the APREL
Laboratories Dag—Paq.


Boundary Effect:

Uncertainty resulting from the boundary effect is less than 2% for the distance
between the tip of the probe and the tissue boundary, when less than 2.44mm.


Spatial Resolution:

The measured probe tip diameter is 5 mm (+/— 0.01 mm) and therefore meets the
requirements of SSI/DRB—TP—D01—032 for spatial resolution.




Page 4 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Receiving Pattern 1900 MHz (Air)



                                   Receiving Pattern Probe s/n 264 1900MHz




                                  ==#— Civt —Bi—Ch2     —i—ChE —6—Tot




Page 5 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Isotropy Error 1900 MHz (Air)

’7                                                                             Isotropy Error
                                                                           Probe s/n 264 1900MHz
                              34664644 bittocne auovo 08
            Error [dB]




                                                           0   100                  200                    300




                                                                     Isotropy Plot Probe s/n 264 1900MHz

                         os                                                                                      zis
     Voltage (mV)




                                                                                  Degrees (x10)

                                                                                  om




Isotropicity in Tissue:                                                                               0.10 dB


Page 6 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Dynamic Range

                                                          Probe 264 Dynamic Range


                     10000




                     1000 +——                                                                                —




                       100      /                                                                      —_——]
  Voltage [microv]




                                                                                                                            —4&— Uncompensated
                       8




                                                                                                                            —t— Compensated




                       0.1                            «




                      0.01
                             1.00   10.00   100.00   1,000.00       10,000.00   100,000.00   1,000,000.00   10,000,000.00
                                                     Electric Field [(Vim)®2)




Page 7 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Video Bandwidth


                                     Probe Frequency Characteristics


   73     .
                                            T11                     TT
                                            nsc                     :‘                            =SA6p
   1 °|~*~                                    =                     7                                           |
   $.]                                           .                                                  |          |
   $& 21—                                    —                               ~
                                                                                 \

      4                                                                          |


      5                                                  f      s                         \

      s                     L    }                   I                   |            1       I         1 |    J
          1                     10                            100                    1000                     10000

                                                                                                  Frequency (Hz)




Video Bandwidth at 500 Hz                                    1 dB
Video Bandwidth at 1000 Hz                                   3 dB




Page 8 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Conversion Factor Uncertainty Assessment


Frequency:                                                 1900MHz

Epsilon:            53.3 (+/—5%)                           Sigma:              1.52 S/im (+/—5%)

ConvF

Channel X:          5.75                                   7%(K=2)

Channel Y:          5.75                                   7%(K=2)

Channel Z:          5.75                                   7%(K=2)

To minimize the uncertainty calculation all tissue sensitivity values were calculated
using a load impedance of 5 MS2.

Boundary Effect:

For a distance of 2.4mm the evaluated uncertainty (increase in the probe sensitivity)
is less than 2%.




Page 9 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Test Equipment

The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server
R:ANCL\Calibration Equipment\Instrument List May 2007.




Page 10 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


Appendix E. Dipole Calibration Data

Validation Dipole 835MHz
P/N: ALS-D-835-S-2
S/N: QTK-315




















Appendix E. Dipole Calibration Data

Validation Dipole 1900MHz
P/N: ALS-D-1900-S-2
S/N: QTK-318


           NCL CALIBRATION LABORATORIES


                       Calibration File No: DC—408—1
                 Project Number: QTKB—Dipole Cal—5230


CERTIFICATE OF CALIBRATION
     it is certified that the equipment idenlified below has been calibrated in the
  NCL CALIBRATION LABORATORIES by qualified personnel following recognized
         procedures and using transfer standards traceable to NRC/NIST.



                               Quietek Validation Dipole


                         Manufacturer: APREL Laboratories
                           Part number: ALS—D—1900—§—2
                                  Frequency: 1.9 GHz
                                  Serial No: OTK—318


                                   Customer: Quictek



                                Calibrated: 15 June 2006
                               Released on: 15 June 2006




                   Catal aeee,
                               w             »*"*"    yit"


                       hx us           .A   4
  Released By:                 C/,!c.




                  51 SPECTRUM WAY                    Bivision ol APREL Lab
                   NEPEAN, ONTARIO                    TEL: (613) 620—400a
                   CANADA K2R 1E6                     FAX: (613) 820—4161


NCL Calibration Laboratories
Division of APREL Laboratories.




Calibration Results Summary
The following results relate the Calibrated Dipole and should be used as a quick
reference for the user.

Mechanical Dimensions

Length:                  70.0 mm
Height:                  39.5 mm

Electrical Specification

SWR:                       11 U
Return Loss:            —25.7 4B
Impedance:               46.8 Q

System Validation Results

| Frequency       1 Gram      10 Gram    Peak
   18 GHz           360 | 20.78 | 67.7




Page 1 of9                              Calibrated by   je:        Approved by:   +2(




                                                                 Sn n ie      ies‘ Mieemnmm s en


NCL Calibration Laboratories
Division of APREL Laboratories.



Introduction

This Calibration Report has been produced in line with the SSI Dipole Calibration
Procedure SSI—TP—018—ALSAS. The results contained within this report are for
Validation Dipole QTK—318. The calibration routine consisted of a three—step
process. Step 1 was a mechanical verification of the dipole to ensure that it meets |
the |IEEE/APREL mechanical specifications. Step 2 was an Electrical Calibration for —
the Validation Dipole, where the SWR, Impedance, and the Return loss were
assessed. Step 3 involved a System Validation using the ALSAS—10U, along with
QTK E—020 130 MHz to 26 GHz E—Field Probe Serial Number 212.


References

SSI—TP—018—ALSAS Dipole Calibration Procedure
SSI—TP—016 Tissue Calibration Procedure
IEEE 1528 "Recommended Practice for Determining the Peak Spatial—Average _
Specific Absorption Rate (SAR) in the Human Body Due to Wireless
Communications Devices: Experimental Techniques"

Conditions

Dipole QTK—318 was received for calibration.

Ambient Temperature of the Laboratory:            22 °C +/— 0.5°C
Temperature of the Tissue:                        20 °C +/— 0.5°C




                                                   us
Page 3 of 9                       Calibrated by    /;//   ‘z‘__a    Approved by: _


NCL Calibration Laboratories
Division of APREL Laboratories.




Dipole Calibration Results

Mechanical Verification

IEEE Length          IEEE Height       Measured               Measured
                                         Length                 Height
   68.0 mm              39.5 mm         70.0 mm                39.5 mm


Tissue Validation

 Head Tissue 1900 MHz             Measured

 Dielectric constant, s;            39.9
 Conductivity, & [S/m]              1.42




                                                                                     &~
                                                           yer   (
                                                                                            C
Page 4 of 9                                Calibrated by   _ Z» -’fi     Approved by: _ C’: 6


     NCL Calibration Laboratories
     Division of APREL Laboratories.




     Electrical Calibration

              Test          |_ _   Result         __
            $11 NL                 —25.7 dB
            SWR                     11 U          ]
|_        Impedance                46.8 0          _
     The Following Graphs are the results as displayed on the Vector Network Analyzer.

     $11 Parameter Return Loss


       §22 REVERSE REFLECTION                                                     th 45— g22
                                                                                  REFERENCE PLANE
       LOG MAGNITUDE               »REF=~—280.000 dB             5.000 d8/01V        0.0000 an
                                                                                 PMARKER 1
                                                                                    1.898384 oHz
                                                                                     "25.748 dP

                                                                                  NARKER T0 MAX
                                                                                  MARKER TO MIN




                     :                 >                   :                      MARKER READOUT
       1. 250614                           GHz                        2.050376    FUNCTIONS




                                                                     hi.                                   x
                                                                 i      x                            fauk. c
     Page 5 of 9                                 Calibrated by       L                Approved hy: _,..gfe =>


NCL Calibration Laboratories
Division of APRET T ahoratories.




SWR

  399 REVERSE AEFLECTION
                                                                           cH 4 — s99                   .
  SHR                            »REF—1.200 1           P00 .000 mU/DIV    REFERENCE PLanE             &
                                                                              9.8000 in
          '                                     i         |               »MARKER 1
                                                                             1.898384 oHz
                                                                                1.186 i1

                                                                           MARKER TN MAX
                                                                           MARKER TO MLN




        :        3           .         .    &       :     &     :         MARKER READOUT
  1.750614                           BHz                      2.050376    FUNCTIONS




                                                        es Sas                 j             Co    4
Page 6 of 9                                Calibrated by ,:-E,                 Approvedby: _6‘ _
                                                                                                   A
                                                                                                   V


—_._      —               ~——_—._>ma_—




       NCL Calibration Laboratories
       Division of APREL Laboratories.


       Smith Chart Dipole Impedance

         $22 REVERSE REFLECTION     1
                                  nfusts s                           CH 4 — §22
         INPEDANCE              m          *                         REFERENCE PLANE
                                                                        0.9000 mn

                                                                    RMARKER 1          —
                                                                       1.898384 GHz
                                                                         146.767 9           >
                                                                         ~3.770 j2

                                                                     MARKER TO MAX
                                                                     MARKER TO HIN




                                                                    MARKER READOUT
                                 1.250614 — 2.050376 GHz            FUNCTIONS




       Page 7 of0                                          h   —
                                             Calibrated by _/ _C_
                                                                                       5
                                                                        Approved by: C «>>


    NCL Calibration Laboratories
    Division of APREL Laboratories.




    System Validation Results Using the Electrically Calibrated Dipole


] FrequencyI 1 Gram | 10 Gram | ~Posk —
                                         _Above Faed Point _.
|      19 GHz |        36.0     | 20.78 | _     67.7




Puge 8 of 9                              Calibrated by   .7 /   C    Approved by:


NCL Calibration l.sborfitnrics
Division of APREL Laboratories.


Test Equipment

The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server .
R:ANCL\Calibration Equipment\instrument List
                        |




                                                  §1%
                                                   mell ce                    5y
Page 9 of9                        Calibrated by   _/_ "*          Approved by: <~ C