RF Exposure Appendix A

FCC ID: UZ7MC3300R

RF Exposure Info

Download: PDF
FCCID_3998147

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/7/27


System Check_Body_900MHz

DUT: D900V2—043

Communication System: CW ; Frequency: 900 MHz;Duty Cycle: 1:1
Medium: MSL_900_180727 Medium parameters used: £ = 900 MHz; 0 = 1.054 S/m; s = 56.231; p = 1000
kg/m*
Ambient Temperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN3925; ConvF(10.05, 10.05, 10.05); Calibrated: 2018/5/31
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 $n917; Calibrated: 2017/12/14
— Phantom: SAM LEFT; Type: QDO0O0P40CD; Serial: TP:1718
— Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)



Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.70 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 61.84 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 4.14 W/kg
SAR(I g) = 2.79 W/kg; SAR(10 g) = 1.84 W/ikeg
Maximum value of SAR (measured) = 3.70 W/kg

        dB
         o


        —2.17


        —~4.34


        —6.50


        —8.67


        —10.84
                    0 dB = 3.70 W/kg = 5.68 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/8/18


System Check_Body_2450MHz

DUT: D2450V2—929

Communication System: CW ; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: MSL_2450_180818 Medium parameters used : f= 2450 MHz; 0 = 2.001 S/m; & = 52.397; p =
1000 kg/Am*
Ambient Temperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(7.75, 7.75, 7.75); Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn853; Calibrated: 2018/7/24
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)



Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 21.4 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 86.91 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 26.0 W/kg
SAR(I g) = 12.4 W/kg; SAR(10 g) = 5.65 W/ike
Maximum value of SAR (measured) = 21.0 W/kg

        dB
         o


        —4.50


        —8.99


        —13.49


        —17.98


        —22.48
                    0 dB = 21.0 W/kg = 13.22 dBW/Akg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/8/25


System Check_Body_2450MHz

DUT: D2450V2—929

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1: 1
Medium: MSL_2450_180825 Medium parameters used : £= 2450 MHz; 0 = 2.003 S/m; g, = 51.38; p = 1000
kg/m>
Arabient Temperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASYS Configuration
— Probe: EX3DV4 — SN3931;ConvF(7.69, 7.69, 7.69);Calibrated: 2017/9/29
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1399; Calibrated: 2017/11/16
— Phantom: ELI v5.0; Type: QDOVAOO2AA; Serial: TP:1227
— Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) =21.7 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=>mm
Reference Value = 106.5 V/m; Power Drift =—0.08 dB
Peak SAR (extrapolated) = 25.8 W/kg
SAR(I g) =12.7 W/kg; SAR(10 g) = 5.95 W/kg
Maximum value of SAR (measured) = 21.0 W/kg

        dB
         0


        —4,49


        —8.97


        —13.46


        —17.94


        —22. 43       6.
                    0 dB =21.0 W/kg = 13.22 dBW/kzg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/8/16


System Check_Body_5250MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: MSL_5G_180816 Medium parameters used: £ = 5250 MHz; 0 = 5.248 S/m; s, = 50.941; p = 1000
kg/m*
Ambient Temperature : 23.2 °C ; Liquid Temperature : 22.2 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(4.8, 4.8, 4.8); Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn853; Calibrated: 2018/7/24
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)



Pin=100mW/Area Scan (71x71x1); Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 16.6 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 64.76 V/m; Power Drift = 0.19 dB
Peak SAR (extrapolated) = 27.4 W/kg
SAR(I g) =7.34 W/kg; SAR(10 g) = 2.08 W/ikeg
Maximum value of SAR (measured) = 18.2 W/kg

        dB
         o


        —f .24


        —14.48


        —21.13


        —28.97


        —36.21
                    0 dB = 18.2 W/kg = 12.60 dBW/Akg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/8/18


System Check_Body_5250MHz
DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: MSL_5G_180818 Medium parameters used: f = 5250 MHz; σ = 5.386 S/m; εr = 47.823; ρ = 1000
kg/m3
Ambient Temperature:23.3 ℃; Liquid Temperature:22.3 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN7306; ConvF(4.8, 4.8, 4.8); Calibrated: 2018/7/26
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: ELI v5.0; Type: QDOVA002AA; Serial: 1131
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 17.0 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 64.76 V/m; Power Drift = 0.19 dB
Peak SAR (extrapolated) = 28.1 W/kg
SAR(1 g) = 7.53 W/kg; SAR(10 g) = 2.14 W/kg
Maximum value of SAR (measured) = 18.6 W/kg




                    0 dB = 18.6 W/kg = 12.70 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/8/16


System Check_Body_5600MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: MSL_5G_180816 Medium parameters used: £ = 5600 MHz; 0 = 5.724 S/m; s, = 50.357; p = 1000
kg/m*
Ambient Temperature : 23.2 °C ; Liquid Temperature : 22.2 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(4.03, 4.03, 4.03); Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn853; Calibrated: 2018/7/24
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)



Pin=100mW/Area Scan (71x71x1); Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 18.5 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 61.41 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) =32.1 W/kg
SAR(1 g) =7.48 W/kg; SAR(10 g) = 2.07 W/ikeg
Maximum value of SAR (measured) = 19.4 W/kg

        dB
         o


        —f .71


        —15.412


        —23.13


        —30.84


        —38.55
                    0 dB = 19.4 W/kg = 12.88 dBW/Akg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/8/16


System Check_Body_5750MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: MSL_5G_180816 Medium parameters used: £ = 5750 MHz; 0 = 5.943 S/m; s, = 50.117; p = 1000
kg/m*
Ambient Temperature : 23.2 °C ; Liquid Temperature : 22.2 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(4.37, 4.37, 4.37); Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn853; Calibrated: 2018/7/24
— Phantom: ELI v5.0; Type: QDOVAOOZAA; Serial: 1131
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)



Pin=100mW/Area Scan (71x71x1); Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 16.3 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 59.11 V/m; Power Drift =0.10 dB
Peak SAR (extrapolated) = 30.4 W/kg
SAR(I g) = 6.77 W/kg; SAR(10 g) = 1.89 W/ikeg
Maximum value of SAR (measured) = 17.8 W/kg

        dB
         o


        —9.23


        —18.417


        —27.70


        —36.94


        —46.17
                    0 dB = 17.8 W/kg = 12.50 dBW/Akg