ANNEX A   Calibration Documents




                 Test Report No.: G0M-1407-3954-TFC093S-V06
                               Eurofins Product Service GmbH
                 Storkower Str. 38c, D-15526 Reichenwalde, Germany
                                                                     Page 43 of 104


Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       Eurofins                                                                           Certificate No: DAE3—522_Sep13

CALIBRATION CERTIFICATE
 Object                                DAE3 — SD 000 DO3 AA — SN: 522


 Calibration procedure(s)              QA CAL—06.v26
                                       Calibration procedure for the data acquisition electronics (DAE)




 Calibration date:                     September 11, 2013




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               02—Oct—12 (No:12728)                                  Oct—13


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Auto DAE Calibration Unit              SE UWS 053 AA 1001        07—Jan—13 (in house check)                            In house check: Jan—14
 Calibrator Box V2.1                    SE UMS 006 AA 1002        07—Jan—13 (in house check)                            In house check: Jan—14




                                        Name                                   Function                                  Signature
 Calibrated by:                         R.Mayoraz                              Technician
                                                                                                                            A iAz_$
 Approved by:                           Fin Bomholt                             Deputy Technical Manager
                                                                                                                         NEl
                                                                                                                        Issued: September11, 2013
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE3—522_Sep13                                        Page 1 of 5


Calibration Laboratory of                                    aNl/4                Schweizerischer Kalibrierdienst
Schmid & Partner                                            iia%;//fi              Service suisse d‘étalonnage
  Engineering AG                                            o ie                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 2 //r_\\\ w          Swiss Calibration Service
                                                               Ti abe®®

Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates


Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
   e  DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
     comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    e    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
               DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
               Common mode sensitivity: Influence of a positive or negative common mode voltage on
               the differential measurement.
               Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
               AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
               Input Offset Measurement: Output voltage and statistical results over a large number of
               zero voltage measurements.

               Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.
               Input resistance: Typical value for information: DAE input resistance at the connector,
               during internal auto—zeroing and during measurement.
               Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
               Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE3—522_Sep13                                  Page 2 of 5


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          1LSB=      6.1uV ,       full range =  —100...+300 mV
       Low Range:           1LSB=       61nV ,       full range =   ~1....... +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

     Calibration Factors             X                            Y¥                    Z
     High Range             404.235 + 0.02% (k=2)   403.905 + 0.02% (k=2)   404.733 + 0.02% (k=2)
     Low Range              3.96415 + 1.50% (k=2)   3.95903 + 1.50% (k=2)   3.97308 + 1.50% (k=2)


Connector Angle

     Connector Angle to be used in DASY system                                       §70°4 1°




Certificate No: DAE3—522_Sep13                      Page 3 of 5


Appendix

1. DC Voltage Linearity
     High Range                               Reading (uV)         Difference (uV)            Error (%)
     Channel X       + Input                   199994.84                   —1.61               —0.00
     Channel X       + Input                    20002.41                   2M                  0.01
     Channel X       — Input                   ~19996.48                   4.99                —0.02
     Channel Y       + Input                   199996.01                   —0.32               —0.00
     Channel Y       + Input                    20000.12                   0.00                0.00
     Channel Y       — Input                   ~19996.80                   4.74                —0.02
     Channel Z       + Input                   199996.36                   —0.14               —0.00
     Channel Z       + Input                    19999.47                   —0.63               —0.00
     Channel Z       — Input                   —20000.32                   1.29                —0.01

     Low Range                                Reading (uV)         Difference (uV)             Error (%)
     Channel X       + Input                    2000.73                    0.37                0.02
     Channel X       + Input                     202.12                    1.52                0.76
     Channel X       — Input                    —198.07                    1.15                —0.58
     Channel Y       + Input                    2000.27                    —0.12               —0.01
     Channel Y       + Input                     200.27                    —0.32               —0.16
     Channel Y       — Input                    ~199.53                    —0.32               0.16
     Channel Z       + Input                    2000.40                    0.21                0.01
     Channel Z       + Input                     199.46                    ~1.06               —0.53
     Channel Z2      — Input                    —200.31                    —0.98               0.49



2. Common mode sensitivity
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                       Common mode                       High Range                         Low Range
                       Input Voltage (mV)            Average Reading (uV)              Average Reading (uV)

    Channel X                           200                     —3.46                        5.15
                                       —200                     6.33                         5.03
    Channel Y                           200                     —0.62                        —0.84
                                       —200                     ~1.40                        ~1.23
    Channel Z                           200                     17.02                        16.63
                                       —200                     —16.85                      ~17.55


3. Channel separation
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                       Input Voltage (mV)        Channel X (uV)          Channel Y (uV)    Channel Z (uV)

    Channel X                    200                       —                  —2.55             —1.42
    Channel Y                    200                   7.66                        —            —0.32
    Channel Z                    200                   6.37                   4.74                   —


Certificate No: DAE3—522_Sep13                             Page 4 of 5


4. AD—Converter Values with inputs shorted
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                   High Range (LSB)                    Low Range (LSB)

    Channel X                                                15741                          16839

    Channel Y                                                15715                          14908

    Channel Z                                                16057                          16580



5. Input Offset Measurement
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
   Input 10MQ

                           Average (uV)        min. Offset (uV)       max. Offset (uV)      Std: |(3:\\I/;ation

    Channel X                    1.14                —0.51                 2.54                    0.62
    Channel Y                    —0.99               —2.51                 0.31                     0.52

    Channel Z                    0.68                —0.45                 2.23                     0.54



6. Input Offset Current
    Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                 Zeroing (kOhm)                       Measuring (MOhm)
    Channel X                                             200                               200

    Channel Y                                             200                               200
    Channel Z                                             200                               200



8. Low Battery Alarm Voltage (Typical values for information)
    Typical values                                Alarm Level (VDC)
    Supply (+ Vec)                                                             +7.9
    Supply (— Vec)                                                             —7.6



9. Power Consumption (Typical values for information)
    Typical values                         Switched off (mA)         Stand by (mA)       Transmitting (mA)
    Supply (+ Vec)                                +0.01                   +6                      +14
    Supply (— Vec)                                —0.01                   —8                      —9




Certificate No: DAE3—522_Sep13                            Page 5 of 5