                                                                                  Test report No.   :   33KE0019-HO-A-R3
                                                                                  Page              :   49 of 111
                                                                                  FCC ID            :   TZ5XT
                                                                                  Issued date       :   August 27, 2013
                                                                                  Revised date      :   October 17, 2013



APPENDIX 2 : System Validation
1.            System validation result

System validation result for Head 1640MHz
Simulated Tissue Liquid Parameter confirmation
                            DIELECTRIC PARAMETERS MEASUREMENT RESULTS
                Ambient Relative             Liquid    Measured
                                    Liquid                                     Target                    Deviation    Limit
      Date       Temp. Humidity              Temp.     Frequency Parameters              Measured
                                     type                                     Value*1                      [%]         [%]
                [deg.c]   [%]                [deg.c]     [MHz]
                                 HSL180                       εr               40.3        40.6             0.8        +/-5
     10-Jul       24.0      56               23.5  1640
                                    0                     σ [mho/m]            1.31        1.30             -0.7       +/-5
                                 HSL180                       εr               40.3        39.3             -2.4       +/-5
  11-Jul     24.0        54                  23.5  1640
                                    0                     σ [mho/m]            1.31        1.25             -5.0       +/-5
 εr: Relative Permittivity / σ : Coductivity
*1 The Target value is a parameter defined in FCC OET 65.


System validation result (for calibration by manufacture)
                                                 SYSTEM VALIDATION
                                                    SAR 1g [W/kg]
                Frequency
     Date                Forward Power 250mW          Conversion 1W           Target 1W *2       Deviation     Limit
                  [MHz]        Measured                 Calculation                                 [%]         [%]
 10-Jul     1640.00               8.08                    32.32                  33.30              -2.9       +/-10
 11-Jul     1640.00               7.71                    30.84                  33.30              -7.4       +/-10
*2 The taget value is the parameter defined in 1g SAR (normalizes to 1W) in manufacturer calibrated dipole (D1640V2 SN




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                         Test report No.   :   33KE0019-HO-A-R3
                                                                         Page              :   50 of 111
                                                                         FCC ID            :   TZ5XT
                                                                         Issued date       :   August 27, 2013
                                                                         Revised date      :   October 17, 2013



HEAD 1640MHz System Validation DATA / Dipole 1640MHz / Forward Conducted Power : 250mW

Communication System: UID 0, CW; Communication System Band: D1640 (1640.0 MHz); Frequency: 1640 MHz;Duty
Cycle: 1:1
Medium parameters used: f = 1640 MHz; σ = 1.301 S/m; εr = 40.618; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY5 Configuration
Probe: ET3DV6 - SN1705; ConvF(5.45, 5.45, 5.45); Calibrated: 2013/06/17
Sensor-Surface: 4mm (Mechanical Surface Detection)
Electronics: DAE4 Sn1372; Calibrated: 2013/06/03
Phantom: SAM Twin TP1762; Type: QD000P40CD; Serial: TP:1762
Measurement SW: DASY52, Version 52.8 (7);

Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 9.69 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 90.822 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 12.0 W/kg
SAR(1 g) = 8.08 W/kg; SAR(10 g) = 4.65 W/kg
Maximum value of SAR (measured) = 9.11 W/kg
Date: 2013/07/10
Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C.




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                         Test report No.   :   33KE0019-HO-A-R3
                                                                         Page              :   51 of 111
                                                                         FCC ID            :   TZ5XT
                                                                         Issued date       :   August 27, 2013
                                                                         Revised date      :   October 17, 2013



HEAD 1640MHz System Validation DATA / Dipole 1640MHz / Forward Conducted Power : 250mW

Communication System: UID 0, CW; Communication System Band: D1640 (1640.0 MHz); Frequency: 1640 MHz;Duty
Cycle: 1:1
Medium parameters used: f = 1640 MHz; σ = 1.246 S/m; εr = 39.327; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY5 Configuration
Probe: ET3DV6 - SN1705; ConvF(5.45, 5.45, 5.45); Calibrated: 2013/06/17
Sensor-Surface: 4mm (Mechanical Surface Detection)
Electronics: DAE4 Sn1372; Calibrated: 2013/06/03
Phantom: SAM Twin TP1762; Type: QD000P40CD; Serial: TP:1762
Measurement SW: DASY52, Version 52.8 (7);

Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 9.33 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 90.799 V/m; Power Drift = -0.05 dB
Peak SAR (extrapolated) = 11.4 W/kg
SAR(1 g) = 7.71 W/kg; SAR(10 g) = 4.44 W/kg
Maximum value of SAR (measured) = 8.72 W/kg
Date: 2013/07/11
Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C.




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                               Test report No.   :   33KE0019-HO-A-R3
                                                                               Page              :   52 of 111
                                                                               FCC ID            :   TZ5XT
                                                                               Issued date       :   August 27, 2013
                                                                               Revised date      :   October 17, 2013



System validation result for Body 1640MHz
Simulated Tissue Liquid Parameter confirmation
                        DIELECTRIC PARAMETERS MEASUREMENT RESULTS
           Ambient Relative               Liquid    Measured
                                Liquid                                      Target                    Deviation    Limit
   Date     Temp. Humidity                Temp.     Frequency Parameters              Measured
                                 type                                      Value*1                      [%]         [%]
           [deg.c]   [%]                  [deg.c]     [MHz]
                                                                  εr        53.7        53.1             -1.2       +/-5
 20-Aug     24.0       50     MSL1800 23.5            1640
                                                              σ [mho/m]     1.42        1.41             -0.5       +/-5
                                                                  εr        53.7        52.5             -2.2       +/-5
  2-Oct     24.0       54     MSL1800 23.5            1640
                                                              σ [mho/m]     1.42        1.38             -2.7       +/-5
 εr: Relative Permittivity / σ : Coductivity
*1 The Target value is a parameter defined in FCC OET 65.


System validation result (for calibration by manufacture)
                                              SYSTEM VALIDATION
                                                    SAR 1g [W/kg]
           Frequency
  Date                   Forward Power 250mW          Conversion 1W           Target 1W *2       Deviation     Limit
             [MHz]             Measured                 Calculation                                 [%]         [%]
 20-Aug     1640.00               7.83                    31.32                  33.50              -6.5       +/-10
  2-Oct     1640.00               8.15                    32.60                  33.50              -2.7       +/-10
*2 The taget value is the parameter defined in 1g SAR (normalizes to 1W) in manufacturer calibrated dipole (D1640V2 SN




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                         Test report No.   :   33KE0019-HO-A-R3
                                                                         Page              :   53 of 111
                                                                         FCC ID            :   TZ5XT
                                                                         Issued date       :   August 27, 2013
                                                                         Revised date      :   October 17, 2013



BODY 1640MHz System Validation DATA / Dipole 1640MHz / Forward Conducted Power : 250mW

Communication System: UID 0, CW; Communication System Band: D1640 (1640.0 MHz); Frequency: 1640 MHz;Duty
Cycle: 1:1
Medium parameters used: f = 1640 MHz; σ = 1.413 S/m; εr = 53.092; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY5 Configuration
Probe: ET3DV6 - SN1685; ConvF(5.05, 5.05, 5.05); Calibrated: 2013/07/19
Sensor-Surface: 4mm (Mechanical Surface Detection)
Electronics: DAE4 Sn509; Calibrated: 2013/07/16
Phantom: ELI 4.0; Type: QDOVA001BA; Serial: xxxx
Measurement SW: DASY52, Version 52.8 (7);

Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 9.50 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 83.948 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 11.5 W/kg
SAR(1 g) = 7.83 W/kg; SAR(10 g) = 4.53 W/kg
Maximum value of SAR (measured) = 8.81 W/kg
Date: 2013/08/20
Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C.




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                         Test report No.   :   33KE0019-HO-A-R3
                                                                         Page              :   54 of 111
                                                                         FCC ID            :   TZ5XT
                                                                         Issued date       :   August 27, 2013
                                                                         Revised date      :   October 17, 2013



BODY 1640MHz System Validation DATA / Dipole 1640MHz / Forward Conducted Power : 250mW

Communication System: UID 0, CW; Communication System Band: D1640 (1640.0 MHz); Frequency: 1640 MHz;Duty
Cycle: 1:1
Medium parameters used: f = 1640 MHz; σ = 1.382 S/m; εr = 52.524; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY5 Configuration
Probe: ET3DV6 - SN1685; ConvF(5.05, 5.05, 5.05); Calibrated: 2013/07/19
Sensor-Surface: 4mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 4mm
(Mechanical Surface Detection)
Electronics: DAE4 Sn1372; Calibrated: 2013/06/03
Phantom: ELI v5.0 TP1207; Type: QDOVA001BB; Serial: TP:1207
Measurement SW: DASY52, Version 52.8 (7); SEMCAD X Version 14.6.10 (7164)

Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 10.2 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 86.647 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 11.7 W/kg
SAR(1 g) = 8.15 W/kg; SAR(10 g) = 4.76 W/kg
Maximum value of SAR (measured) = 9.17 W/kg
Date: 2013/10/02
Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C.




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                               Test report No.   :   33KE0019-HO-A-R3
                                                                               Page              :   55 of 111
                                                                               FCC ID            :   TZ5XT
                                                                               Issued date       :   August 27, 2013
                                                                               Revised date      :   October 17, 2013



System validation result for Body-worn 1640MHz
Simulated Tissue Liquid Parameter confirmation
                        DIELECTRIC PARAMETERS MEASUREMENT RESULTS
           Ambient Relative               Liquid    Measured
                                Liquid                                      Target                    Deviation    Limit
   Date     Temp. Humidity                Temp.     Frequency Parameters              Measured
                                 type                                      Value*1                      [%]         [%]
           [deg.c]   [%]                  [deg.c]     [MHz]
                                                                  εr        53.7        53.1             -1.2       +/-5
 10-Oct     24.0       55     MSL1800 23.5            1640
                                                              σ [mho/m]     1.42        1.38             -3.0       +/-5
 εr: Relative Permittivity / σ : Coductivity
*1 The Target value is a parameter defined in FCC OET 65.


System validation result (for calibration by manufacture)
                                              SYSTEM VALIDATION
                                                    SAR 1g [W/kg]
           Frequency
  Date                   Forward Power 250mW          Conversion 1W           Target 1W *2       Deviation     Limit
             [MHz]             Measured                 Calculation                                 [%]         [%]
 10-Oct     1640.00               7.65                    30.60                  33.50              -8.7       +/-10
*2 The taget value is the parameter defined in 1g SAR (normalizes to 1W) in manufacturer calibrated dipole (D1640V2 SN




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                         Test report No.   :   33KE0019-HO-A-R3
                                                                         Page              :   56 of 111
                                                                         FCC ID            :   TZ5XT
                                                                         Issued date       :   August 27, 2013
                                                                         Revised date      :   October 17, 2013



BODY 1640MHz System Validation DATA / Dipole 1640MHz / Forward Conducted Power : 250mW

Communication System: UID 0, CW; Communication System Band: D1640 (1640.0 MHz); Frequency: 1640 MHz;Duty
Cycle: 1:1
Medium parameters used: f = 1640 MHz; σ = 1.378 S/m; εr = 53.066; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY5 Configuration
Probe: ET3DV6 - SN1685; ConvF(5.05, 5.05, 5.05); Calibrated: 2013/07/19; ${Probe: Calibration Date}
Sensor-Surface: 4mm (Mechanical Surface Detection)
Electronics: DAE4 Sn509; Calibrated: 2013/07/16
Phantom: ELI 4.0; Type: QDOVA001BA; Serial: xxxx
Measurement SW: DASY52, Version 52.8 (7);

Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 9.58 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 85.689 V/m; Power Drift = -0.06 dB
Peak SAR (extrapolated) = 11.4 W/kg
SAR(1 g) = 7.65 W/kg; SAR(10 g) = 4.36 W/kg
Maximum value of SAR (measured) = 8.61 W/kg
Date: 2013/10/10
Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C.




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   57 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013



2.      System Validation Dipole (D1640V2,S/N: 320)




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   58 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
    DASY Version                                              asys                                vees.t
    Extrapolation                                    Advanced Extrapolation
    Phantom                                           Modular Fiat Phantom
    Distance Dipole Center — TSL                              10 mm                              wih Spacer
    Zoom Scan Resolution                                obx,dy, dz =5 mm
    Frequency                                           1640 Mz « 1 Nz

Head TSL parameters
   ‘The following parameters and calculations were applled,
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  220°C              «02               1.31 mhoim
     Measured Head TSL parameters                              (220202)°C         41016%            1.38 mhoim +6%
     Head TSL temperature change during test                     <a5°0                —                    —
SAR result with Head TSL

     SAR averaged over 1 om" (1 g) of Head TSL                 Condiion
     SAR measured                                         250 mW input power               8.48 mW / a
     SAR for nominal Head TSL parameters                      normatized to 1W       33.3 mW /9 2 17.0 % (=2)

     SAR averaged over 10 em° (10 g) of Head TSL                 condiion
     SAR measured                                         250 mW input power               450 mW / a
     SAR for nominal Head TSL parameters                      normalized to 1W       18.1 mW i9 2 16.% (kx2)

Body TSL parameters
   The folowing parameters and calculations were applied
                                                                Temperature      Permittivity         Conductivity
     Nominal Boy TSL parameters                                    220°C             sar               1.42 mhoim
     Measured Body TSL parameters                               (22.0202)°C       se+6%             1.44 mhoim +8 %
     Body TSL temperature change during test                       <o5°c                                   —
SAR result with Body TSL

     SAR averaged over 1 cm" (1 g) of Body TSL                 Condiion
     SAR measured                                         250 riW input power              8.49 mW /a
     SAR for nominal Body TSL parameters                      normalized to 1W       30.5 mW / g 2 17.0 % (ke2)

     SAR averaged over 10 em* (10 g) of Body TSL               condiion
     SAR measured                                         250 mW input power                    463 mW /o
     SAR for nominal Body TSL parametrs                    normalized to 1W          18.4 mW / g # 16.5 % (ke2)



Conticate No: D1640v2320_Junt2                         Page 3 of 8


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   60 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   61 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   62 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   63 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   64 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                                         Test report No.   :   33KE0019-HO-A-R3
                                                                                         Page              :   65 of 111
                                                                                         FCC ID            :   TZ5XT
                                                                                         Issued date       :   August 27, 2013
                                                                                         Revised date      :   October 17, 2013



D1640V2 Calibration for Impedance and Return-loss


 Equipment        Dipole Antenna                            Model                D1640V2
 Manufacture      Schmid&Partner Engineering AG             Serial               320
 Tested by        Tomotsugu Koyama
1. Test environment
 Date                     June 21, 2013
 Ambient Temperature          24.0 deg.C                   Relative humidity            40%RH

2. Equipment used

Control No.      Instrument           Manufacturer           Model No            Serial No     Test Item        Calibration
                                                                                                                Date *
                                                                                                                Interval(month)
MNA-01           Network Analyzer     Agilent/HP             E8358A              US41080381 SAR                 2012/09/14 * 12
MNCK-01          Type N Calibration   Agilent                85032F              MY41495257 SAR                 2012/09/18 * 12
                 Kit
EST-46           3.5mm ECONOMY        Agilent                85052D              MY43252869 SAR                 2012/08/13 * 12
                 CALIBRATION KIT
MDA-11           Dipole Antenna       Schmid&Partner         D1640V2             320            SAR             2012/06/22 * 12
                                      Engineering AG
MPSAM-04         SAM Phantom          Schmid&Partner         SAM Twin            1764           SAR             2013/06/11 * 12
                                      Engineering AG         Phantom V4.0
                                                             and V5.0
MPF-03           2mmOval Flat         Schmid&Partner         QD OVA 002A         1207           SAR             2013/06/18 * 12
                 Phantom ERI 5.0      Engineering AG         (ERI5.0)
MOS-26           Thermo-Hygrometer    Custom                 CTH-201             -              SAR             2013/05/16 * 12
HSL1800                                                                                                         Daily check
MSL1800                                                                                                         Daily check
SAR room1                                                                                                       Daily check

3. Test Result
Impeadance,Transformed to feed point            Head          Deviation      Tolerance                          Result
Calibration (SPEAG) 2012/06/22                  52.6 Ω+2.8jΩ -               -                                  -
Calibration(ULJ)2013/6/21                       55.43Ω+1.824jΩ+2.83Ω-0.976jΩ +/-5Ω+/-5jΩ                        Complied

Return loss                                     Head                 Deviation          Tolerance               Result
Calibration (SPEAG) 2012/06/22                  -28.5dB              -                  -                       -
Calibration(ULJ)2013/6/21                       -25.30dB             +3.20dB            -28.5 *+/-20%           Complied

Impeadance,Transformed to feed point            Body          Deviation     Tolerance                           Result
Calibration (SPEAG) 2012/06/22                  48.9 Ω+2.8jΩ -              -                                   -
Calibration(ULJ)2013/6/21                       49.62Ω+3.801jΩ0.72Ω+1.001jΩ +/-5Ω+/-5jΩ                         Complied

Return loss                                     Body                 Deviation          Tolerance               Result
Calibration (SPEAG) 2012/06/22                  -30.2dB              -                  -                       -
Calibration(ULJ)2013/6/21                       -28.33dB             1.87dB             -30.2 *+/-20%           Complied
*Tolerance : According to the KDB450824D02




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                      Test report No.   :   33KE0019-HO-A-R3
                                                      Page              :   66 of 111
                                                      FCC ID            :   TZ5XT
                                                      Issued date       :   August 27, 2013
                                                      Revised date      :   October 17, 2013



Measurement Plots

<Head Liquid>




<Body Liquid>




UL Japan, Inc.
Head Office EMC Lab.
4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN
Telephone: +81 596 24 8999
Facsimile: +81 596 24 8124


                                                                                     Test report No.   :   33KE0019-HO-A-R3
                                                                                     Page              :   67 of 111
                                                                                     FCC ID            :   TZ5XT
                                                                                     Issued date       :   August 27, 2013
                                                                                     Revised date      :   October 17, 2013



3.      Validation uncertainty

The uncertainty budget has been determined for the DASY5 measurement system according to the SPEAG documents[2]
and is given in the following Table.


                                           Uncertai Probability               (ci)                Standard         vi
Error Description                          value ± distribution     divisor   1g                  (1g)             or
                                                                                                                   veff
Measurement System
Probe calibration                          ± 6.00     Normal        1         1                   ± 6.00           ∞
Axial isotropy of the probe                ± 4.7      Rectangular   √3        1                   ± 2.7            ∞
Spherical isotropy of the probe            ±   9.6    Rectangular   √3        0                   ±    0.0         ∞
Boundary effects                           ±   1.0    Rectangular   √3        1                   ±    0.6         ∞
Probe linearity                            ±   4.7    Rectangular   √3        1                   ±    2.7         ∞
Detection limit                            ±   1.0    Rectangular   √3        1                   ±    0.6         ∞
Modulation response                        ±   0.0    Rectangular   √3        1                   ±    0.0         ∞
Readout electronics                        ±   0.3    Normal        1         1                   ±    0.3         ∞
Response time                              ±   0.0    Rectangular   √3        1                   ±    0.0         ∞
Integration time                           ±   0.0    Rectangular   √3        1                   ±    0.0         ∞
RF ambient Noise                           ±   1.0    Rectangular   √3        1                   ±    0.6         ∞
RF ambient Reflections                     ±   1.0    Rectangular   √3        1                   ±    0.6         ∞
Probe Positioner                           ±   0.8    Rectangular   √3        1                   ±    0.5         ∞
Probe positioning                          ±   6.7    Rectangular   √3        1                   ±    3.9         ∞
Max.SAR Eval.                              ±   2.0    Rectangular   √3        1                   ±    1.2         ∞
Dipole Related
Deviation of exp.dipole                    ± 5.5      Rectangular   √3        1                   ± 3.2            ∞
Dipole Axis to Liquid Distance             ± 2.0      Rectangular   √3        1                   ± 1.2            ∞
Input power and SAR drift meas.            ± 3.4      Rectangular   √3        1                   ± 2.0            ∞
Phantom and Setup
Phantom uncertainty                        ±   4.0    Rectangular   √3        1                   ±    2.3         ∞
Liquid conductivity (target)               ±   5.0    Rectangular   √3        0.78                ±    2.3         ∞
Liquid conductivity (meas.)                +   5.0    Normal        1         0.26                +    1.3         ∞
Liquid permittivity (target)               ±   5.0    Rectangular   √3        0.78                ±    2.3         ∞
Liquid permittivity (meas.)                -   5.0    Normal        1         0.23                -    1.2         ∞
Liquid conductivity
                                           ± 1.7      Rectangular   √3        0.78                ± 0.8            ∞
- temp.unc (below 2deg.C.)
Liquid permittivity
                                           ± 0.3      Rectangular   √3        0.23                ± 0.0            ∞
- temp.unc (below 2deg.C.)

Combined Standard Uncertainty                                                                     ± 10.157
Expanded Uncertainty (k=2)                                                                        ± 20.3


Note: This uncertainty budget for validation is worst-case.