SAR Caldata

FCC ID: TVU-SL78H

RF Exposure Info

Download: PDF
FCCID_1069991

CETECOM ICT Services GmbH
Untertuerkheimer Str. 6-10, 66117 Saarbruecken, Germany   Phone: +49 (0) 681 598-0      Fax: -8475
SAR-Laboratory                                            Phone: +49 (0) 681 598-8454             .




                                            Accredited testing laboratory

                              DAR registration number: DAT-P-176/94-D1

                               Federal Motor Transport Authority (KBA)
                               DAR registration number: KBA-P 00070-97




                          Appendix to test report 1-0876-01-04/08
                          Calibration data, Phantom certificate
                          and detail information of the DASY4 System




As of 2008-11-13                                                                                      Page 1 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08




Table of Content

1    Calibration report “Probe ET3DV6”.................................................................................................................................... 3
2    Calibration report “1900 MHz System validation dipole” ................................................................................................. 12
3    Calibration certificate of Data Aquisition Unit (DAE) ...................................................................................................... 21
4    Certificate of “SAM Twin Phantom V4.0/V4.0C” ........................................................................................................... 22
5    Application Note System Performance Check ................................................................................................................... 24



2008-11-13




As of 2008-11-13                                                                                                                                      Page 2 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08

1 Calibration report “Probe ET3DV6”




As of 2008-11-13                                                               Page 3 of 29


                                                                                                      CETECOM*


Calibration Laboratory of                                                         Schwatzerisctr Kaibrientienst
Schmid & Partner                                                                  Service sulsse sitelonnage
  Engineering AG                                                                  Servalo ustzzone d tarmturs
Zoughausstratse 43, 8004 zurien, Suzarand                                         Swizs Caliration Servicn

Acorediec y the Sitss Acerediaion Servics (BA8)                                Acerediation No.: SGS 108
The Swise Aceredtafion Sorvieo is ono ofthe signatories to the EA
WutilsteraAgreement for the recognlton of cabratlon certleates


Glossary:
TSL                        tissue simulating liquid
NORMixy,z                  sensitivity in free space
ConvF                      sensitivity in TSL / NORMy,z
DCP                       diode compression point
Polarization q            @ ratation around probe axis
Polarization 8            8 rotation around an axis that is in the plane normal to probe axis (at
                           measurement center), Le., 5 = 0 is normal to probe axis

Calibration is Performed According to the Following Standards:
   a) IEEE Sid 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial~
         Averaged Specific Absarption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques®, December 2003
    b} 1EC 62209—1, *Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximily to the ear (frequency range of 300 MHz to 3 GHz)‘,
       February 2005

Methods Applied and Interpretation of Paramsters:
   * NORMxy.z: Assessed for E—field polarizsion 8 0 (f < 900 MHz in TEM—cell; f> 1800 MHz:
         R22 waveguide}. NORMx.y.z are only intermediate values, L.e., the uncertainties of
         NORKx,y,z does not effect the E"—field uncertainty inside TSL (see below ConvF).
    *    NORM(fxy.z = NORMx.y,z * frequency_response (see Frequency Response Chart). This
         linearization is implemented in DASY4 software versions later than 4.2, The uncertainty of
         the frequency response is included in the stated uncertainty of ConvF,
    «—   DCPxy,z: DCP are numerical linearization parameters assessed based on the data of
         power sweep (no uncertainty required). DCP does not depend on frequency nor media.
    «_   ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or
         Temperature Transfer Standard for f < 800 MHz) and inside waveguide using analytical field
         distributions based on power measurements forf> 800 MHz. The same setups are used for
         assessment of the parameters applied for boundary compansation (alpha, depth) of which
         typical uncertainty values are given. These parameters are used in DASY4 software to
         improve probe accuracy clase to the boundary. The sensitivity in TSL corresponds to
         NORMxy,z " ConvF whereby the uncertainty corresponds to that given for ConvF, A
         frequency dependent Convi~ is used in DASY version 4.4 and higher which allows extending
         the validity from : 50 MHz to x 100 MHz.
   *     Spherical isotropy (3D deviation from isofropy); in a field of low gradients reallzed using a
         fat phantom exposed by a patch anterina.
   =—    Sensor Offset. The sensor offset correspondsto the offset of virtual measurement center
         from the probe tip (on probe axis}. No tole—ance required


Gertate No: ET3—1580_Augo8                                 Page 2 or9


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08




As of 2008-11-13                                                               Page 5 of 29


                                                                                                          CETECOM*

    ET3DV6 SN:1558                                                                      August 15, 2008


    DASY — Parameters of Probe: ET3DV6 SN:1558

    Sensilivity in Free Space®                                                Diede Compression®
             NormX                 203 s 10.1%         wViVim)®               DCP x         93 m
             Norm¥Y                188 a101%           wV/tVim®               per y         92 mV
             NormZ                 170 at0t%           pVi(imY®               PCP Z         95 m

    Sensitivity in Tissue Simulating Liquid (Conversion Factors)
    Please seo Page 8.


    Boundary Effect

    TsL                  900 NHz        Typlcal SAR gradient5% per mm

             Sensor Center to Phentorn Suface Dislance                        3.7 mm 4. tim
             $ARe              Without Correction Algorithm                     108    B4
             SAR[%)            With Corection Algorithm                         o8     04

   TSE                   1750 mz        Typleal SA gradient: 10 % per rim



             Sensor Canter to Phantom Surtace Distance                        37 mm 4.7 mm
             SARs, (%)         Without Correction Algorthm                      10.   61
             SAR, (%)          With Correction Algorithm                        oa    o6


    Sensor Offset

             Probe Tipto Sensor Center                                   2.7 mm



    (The raported uncertainty of measurement is stated as the standard uncertainty of
    measurementmultiplied by the coverage factor k=2, which for a normal distribution
    (corresponds to a coverage probability of approximately 95%.

   A Thuncariinia o Nerm2CY.2 d t afat the E—hld unceralny ieade T9L {seo Page 0)
   * Nameres!Insaieatonpammetor unsedainy not romures



Gerifcate No: Er3—1658_Auso8                       Poge 4 o9


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08




As of 2008-11-13                                                               Page 7 of 29


                                                                                                                                  CETECOM*


    ETSDVS SN:155§                                                                                            August 15, 2008



                                     Receiving Pattern (¢), 8 = 0°
                                                                                                                  s
                                                                                           T= 1800 MHz, WG R22                    |




                                                                                 pososs
                                                                                  ce—x ~e—y              <e—z    —e—fat




         fZTLEL
         M    |
                                _E           a|I                                       51—
                                                                                        mm
         06                     +    ‘                                                               +          —3— 30 MHz
     g ®8                                                      |                                ofei~<] |—m— 100 We
     & 4*                 11                             gpe                           t                —+— 0o k
        g‘g       "HMes                            a               |             ag                             | m 1800 hi)
     5& ne                           |                                           |                              | —a—2500 hiz |
        «6          |                lt                                                          7       1
        3                       A4                                           1                       —H
              —         ul               !                              —.                                :


              0          s                         120                 180       240           a0o




                                Uncertainty of Axtal Isotropy Assessment: £ 0.5% (k=Z)




Gortficate No: ET9—1058_Augos                                      Page 6 ato


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08




As of 2008-11-13                                                               Page 9 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08




As of 2008-11-13                                                               Page 10 of 29


                                                                                                 CETECOM*



    ETSDV6 SN:1558                                                             August 15, 2008


                             Deviation from Isotropy in HSL
                                        Error (4, 8), f= 900 MHz




                                                                                    Ervor IdB]




                              |m—1ac—080moso—oms mant.s40 maac—o20 m—s 20003
                               momaz memoso me«ooso mossoso mem it


                         Uncertainty of Spherical Isotrapy Assessment:#2.6% (K=2)




Conicata No: ET3—165¢ Aug0                  Page 0 of0


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08

2 Calibration report “1900 MHz System validation dipole”




As of 2008-11-13                                                               Page 12 of 29


                                                                                                          CETECOM*


Calibration Laboratory of                                                     § SottwelzerisehorKaliberdionst
Schmid & Partner                                                              g Sorvie sulese @itslonnage
   Engineering AG                                                                Serviio suirzero i tereture
Zoughausstrasse 43 4004 2urich, Suizarand                                     5. sutss Caltration Sarvice

Jormdted by the Sutes Acstacitaion Sanice (EAS)                                Accredtation no: SCS 108
The Swrise Accredtatlon Service Is one of the signatories o the EA
Multiaterat Agreementfor the resounition of ealiration cerihcate
Glossary:
TSL                        tissue simulating liquid
ConvE                                     n TSL / NORM xy.z
N/A                                   icable or not measured
Callbration is Performed Accerding to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Poak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
        Communications Devices: Measurement Techniques®, December 2003
    6 CENELEC &N 50361, "Basic standard for the measurement of Specific Absorption Rate
        related to human exposure to electromagnetic fields from mobile phones (300 MHz — 3 GHz),
        July 2001
    c) Federal Communications Commission Office of Engineering & Technology (FCC OET),
        "Evaluating Compliance with ECC Guidelines for Human Exposure to Rediofrequency
        Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
        Portable Devices with FCC Limits for Human Exposure to Rediafrequency Emissions‘,
        Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
   d) DASY4/5 System Handbook

Methods Applied and Interpretation of Paramoters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
        of the certificate. All figures stated in the cortificate are valid at the frequency indicated.
    *   Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
        point exactly below the center marking of the flat phantom section, with the arms oriented
        parallel to the body axis.
    *   Feed Point Impedance and Return Loss: These parameters are measured with thedipole
        positioned under the liquid filled phantom. The impedance stated is transformed from the
        measurement at the SMA connector to the feed point. The Return Loss ensures low
        reflected power. No uncertainty required.
   «_   Electrical Defay: One—way delay between the SMA connector and the antenna feed point. No
        uncertainty required.
   _    SAR measured: SAR measured at the stated antenna input power

   *    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
        connector
   *    SAR for nominal TSL parameters: The measured TSL paramefers are used to calculate the
        nominal SAR result.



Gertficate No: D1200V7—531_May0A                            Page 2 of 9


                                                                                                                                 CETECOM*


Measurement Conditions
    DASY system confiquretion, as faras naot gven on page 1.
      DASY version                                           provs                                      var
      Extrapolation                                  Advanced Extragolation
      Phantom                                       Noduler Flat Phantom ¥5.0
      Distance Dipole Genter» TSL                             10 m                                   wih Spacer
      Zoom Scan Resolution                              dsoy, oz = 8 mm
     [ Frequency                                        1900 MHz+ 1 NKz

Head TSL parameters
    The falowing gerematers and oaloulaions ware appled.
                                                               Tempornture             Permitivity        Conductivity
      Nominal Hoad TSL p-ramct;;          —                         220°C                 40.0             140 mhoim
      Measured Haed TSL prrametars                             (22.020.2) °C           30426 %         _’\:b mho/l;\ i-fi %
      Hoad TSL temporatura during tast                         @rczo2)‘C                                       —


SAR result with Hoad TSL

      SAR averaged over 1 om(1 5)of Head TBL                   condiion                                                      ]
      SAR mensured                                        250 mW inout power                     847 mWw‘ g
      SAR nommalized                                       nomalzed to 1W                        357 mW g
      SAR for nominal Head TSL paremeters ‘                normalied to 1W         —      37.8 MWV a247.0 % (ke2)


      SAR averaged over 10 em(10 5) of Head ToL.                Conditon
      SAR measured                                        250 mW inout power                     5.00 mw7g
      SAR nommalized                                       normalied o 1W                        20.0 mW!q
      SAR for nominal Head TSL paremators                  no;mahzed to 1W                1927 mW 1 q 2 16.5 % (ke2)




! Comection to nominal TBL parameters according to d}. chapler ‘SAR Sensiivites"
Cerifeate No: D1900V2—531_Meyo®                        Page 3 oft


                                                                                                                  cErecom"

 Body TSL parameters

     The following parameters and calculations were appled
                                                                Temperature       Permittivity      Gondiustivity
       Nominal Body TSL parameters                                220°C               sas            182 mhoim
       Measured Body TSL. narameters                            @20202)°C         52026 %         1.654 mhoim#6 %
       Body TSL temparnture during test                         cosso2)°C                                  —

SAR result with Body TSL

       SAR averaged over 1 em" (1 g) of Body TSL                  Gondition
       SAR measured                                          250 mW input power              9.78 mW / a
       SAR nommalzed                                          normalized to 1W               304 mW g
     SAR for nminal Body TL parameters *           _______| _ normalized to 1W       38.3 mW ! o # 17.0 % (         _
       SAR avaraged over 10 om" (10 ) of Bocy TSL.               conition
      SAR messured                                          250 n Input power                5.13 mW/g
      SAR nomnalized                                          nomalized to 1W                20.5 mW/3
      SAR for nominal Body TEL peremeters *                   nomalized to 1W        20.2 mW [ g 2 16.5 % (k=2)




* Correion to nomina! TSL paramaters acoording to 0}, chapter BAR Sensitvties®
Cortficata No: D1900v2—531_Maya                         Page 4 o19


                                                                                                                      CETECOM*

Appondix

Anterna Parameters with Head TSL

     [ impedance, transformed to feed point                                       seon+s7in
     | Return Loss                                                                   +24.6 cB

Antenna Parameters              with Body TSL

      impedance, transformed to feeel point                                       «s2n+s3im
      Rowm Lose                                                                     —24.0 48

General Antenna Parameters and Design

      Eleetrcal Delay (one direction)                                                1201 ns
After long term use with 100W radiated power, only a slight warming of Ihe dipole nearthe taedpoint can be measured
The dipols is mads of standard semirigid coxial cable. The center conductor of the fosing ini direcily connected to the
second arm of the dipale. The antenna is therefare short—sroulted for DC—signols.
No excessive force mus be applied to the dipole arms, because they might bend or the soidered connections near the
feecpoint may be camaged

Additional EUT Data

      Manufactured by                                                                 sreao
      Manufactured on                                                            January 24, 2001




Certficate No: D1900V7—531_Mayo6                       Page 5 ofa


C                                                                                                      CETECOM*


DASY4 Validation Report for Head TSL
                                                                               Date/Time: 08.05.2008 14:26:07

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial; D1900V2 — SN:631

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL U10 BB;
Medium parameters used: f= 1900 MHz: 0 = 1.44 mhorm; £, = 39.4; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessmant)

DASY4 Configuration:
    *    Probe: ESSDVZ — SN3025; ComvR(4.0, 4.9, 4.9}; Calbraled; 28.04.2006
    +.   Sensor—Surtace: 3.4mm (Mechanical Surtace Detection)
    +    Electronics: DAE4 Sn601; Calbrated: 14.09.2008
    *    Phantom: Flat Phantom 5.0 (front); Type: GDO0ORSOAA;;
    +    Measuremant SW: DASY4, ¥4.7 Build 55; Postprocessing SW SEMCAD, V1.8 Buld 172

Pin = 250 mW; d = 10 mm/Zoom Scan (7x7x7)/Cube 0:
Measurement giid: dx=5mm, dy=5mm, dz=5mm
Reference Value =90.1 Vim; Power Drift =0.030 d8
Peak SAR (extrapalated) = 17.9 Wikg
SAR(1 g) = 9.67 mWig; SAR(10 g) = 5 mWig
Maximum value of SAR (measured) =11 4 mWig

         da
         0.000




         +10.9



         "14.5


         "18.1

                   0dB= 14mWlg




Corfcale No: D1900V2—531_MayO®                       Page 6 of$


'''''                                                                                                 CETECOM*


   Impedance Measurement Plot for Head TSL


                                                          a my awes seraorce
       EBJ sn      a uie             ®sasize   assire ssussph      1.800,000 so0 rets

                                                                                        t Hankers
                                                                                         asaus
                                                                                          ces a
                                                                                        1.s080e on




                                                                                        C#2 nankers
                                                                                        amasses ce
                                                                                         ae0000 ane




  Certiicate No: D1900V2.—531_May8                  Page 7 08


                                                                                                   CETECOM*


DASY4 Validation Report for Bodly TSL
                                                                             Date/Time: 14.05.2008 16:36:27
Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:531

Communication System: CW; Frequency: 1800 MHz;Duty Cycie: 1:1
Medium: MSL U1O BB;
Medium parameters used: f = 1900 MHz; 0 = 1.54 mhoim; £. = 52.2; p = 1000 kg/m®
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessment)

DASY4 Configuration:
    +   Probe: ES3DV2— SN3026; ComF(4.5, 4.5, 4.5}; Callbrated: 28.08.2008
    *   Sensor—Surfacs: 3.4mm (Machanical Surface Detection)
    *   Electronics: DAE4 Sn801; Callbrated: 14.09.2008
    *   Fhantom: Flat Phaniam 5.0 (back}; Type: QDOOOPSOAA;;
    +   Measurement SW: DASY4, V4.7 Buld 71; Postprocessing SW: SEMCAD, V1.8 Buld 184

Pin = 250 mW;           = 10 mmiZoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=&mm, dy=&mm, dz=&mm
Reference Value = 90.1 V/m; PowerDrift=0.011 dB
Peak SAR (extrapolated) = 17.4 W/kg
SAR(1 g) = 9.78 mWig; SAR(10 g} = 5.13 mWig
Maximum value of SAR (measured) = 11.8 mWig

          «n
           0.008




          420




          240




          ana




          186

                     00B =11.8mWiq



Gertficate No: D1900V2—531_May08                    Page 8 or0


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08




As of 2008-11-13                                                               Page 20 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08

3 Calibration certificate of Data Aquisition Unit (DAE)




As of 2008-11-13                                                               Page 21 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08

4       Certificate of “SAM Twin Phantom V4.0/V4.0C”




As of 2008-11-13                                                               Page 22 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08




As of 2008-11-13                                                               Page 23 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08

5 Application Note System Performance Check
5.1.1.1   Purpose of system performance check
The system performance check verifies that the system operates within its specifica-tions. System and operator
errors can be detected and corrected. It is recommended that the system performance check is performed prior to
any usage of the system in order to guarantee reproducible results.
The measurement of the Specific Absorption Rate (SAR) is a complicated task and the result depends on the proper
functioning of many components and the correct settings of many param-eters. Faulty results due to drift, failures
or incorrect parameters might not be recognized, since they often look similar in distribution to the correct ones.
The Dosimetric Assessment System DASY4 incorporates a system performance check procedure to test the proper
functioning of the system. The system performance check uses normal SAR measurements in a simplified setup
(the at section of the SAM Twin Phantom) with a well characterized source (a matched dipole at a specified
distance). This setup was selected to give a high sensitivity to all parameters that might fail or vary over time (e.g.,
probe, liquid parameters, and software settings) and a low sensitivity to external effects inherent in the system (e.g.,
positioning uncertainty of the device holder). The system performance check does not replace the calibration of the
components. The accuracy of the system performance check is not sufficient for calibration purposes. It is possible
to calculate the field quite accurately in this simple setup; however, due to the open field situation some factors
(e.g., laboratory re ections) cannot be accounted for. Calibrations in the at phantom are possible
with transfer calibration methods, using either temperature probes or calibrated E-field probes. The system
performance check also does not test the system performance for arbitrary field sit-uations encountered during real
measurements of mobile phones. These checks are performed at SPEAG by testing the components under various
conditions (e.g., spherical isotropy measurements in liquid, linearity measurements, temperature variations, etc.),
the results of which are used for an error estimation of the system. The system performance check will indicate
situations where the system uncertainty is exceeded due to drift or failure.

5.1.1.2   System Performance check procedure
Preparation
The conductivity should be measured before the validation and the measured liquid parameters must be entered in
the software. If the measured values differ from targeted values in the dipole document, the liquid composition
should be adjusted. If the validation is performed with slightly different (measured) liquid parameters, the expected
SAR will also be different. See the application note about SAR sensitivities for an estimate of possible SAR
deviations. Note that the liquid parameters are temperature dependent with approximately – 0.5% decrease in
permitivity and + 1% increase in conductivity for a temperature decrease of 1° C. The dipole must be placed
beneath the flat phantom section of the Generic Twin Phantom with the correct distance holder in place. The
distance holder should touch the phantom surface with a light pressure at the reference marking (little hole) and be
oriented parallel to the long side of the phantom. Accurate positioning is not necessary, since the system will search
for the peak SAR location, except that the dipole arms should be parallel to the surface. The device holder for
mobile phones can be left in place but should be rotated away from the dipole. The forward power into the dipole at
the dipole SMA connector should be determined as accurately as possible. See section 4 for a description of the
recommended setup to measure the dipole input power. The actual dipole input power level can be between 20mW
and several watts. The result can later be normalized to any power level. It is strongly recommended to note the
actually used power level in the „comment“-window of the measurement file; otherwise you loose this crucial
information for later reference.
System Performance Check
The DASY4 installation includes predefined files with recommended procedures for measurements and validation.
They are read-only document files and destined as fully defined but unmeasured masks, so you must save the
finished validation under a different name. The validation document requires the Generic Twin Phantom, so this
phantom must be properly installed in your system. (You can create your own measurement procedures by opening
a new document or editing an existing document file). Before you start the validation, you just have to tell the
system with which components (probe, medium, and device) you are performing the validation; the system will
take care of all parameters. After the validation, which will take about 20 minutes, the results of each task are


As of 2008-11-13                                                                                       Page 24 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08

displayed in the document window. Selecting all measured tasks and opening the predefined “validation” graphic
format displays all necessary information for validation. A description of the different measurement tasks in the
predefined document is given below, together with the information that can be deduced from their results:
•   The „reference“ and „drift“ measurements are located at the beginning and end of the batch process. They
    measure the field drift at one single point in the liquid over the complete procedure. The indicated drift is
    mainly the variation of the amplifier output power. If it is too high (above ± 0.1dB) the validation should be
    repeated; some amplifiers have very high drift during warm-up. A stable amplifier gives drift results in the
    DASY4 system below ± 0.02 dB.
•   The „surface check“ measurement tests the optical surface detection system of the DASY4 system by
    repeatedly detecting the surface with the optical and mechanical surface detector and comparing the results.
    The output gives the detecting heights of both systems, the difference between the two systems and the
    standard deviation of the detection repeatability. Air bubbles or refraction in the liquid due to separation of the
    sugar-water mixture gives poor repeatability (above ± 0.1mm). In that case it is better to abort the validation
    and stir the liquid. The difference between the optical surface detection and the actual surface depends on the
    probe and is specified with each probe. (It does not depend on the surface reflectivity or the probe angle to the
    surface within ± 30°.) However, varying breaking indices of different liquid compositions might also influence
    the distance. If the indicated difference varies from the actual setting, the probe parameter „optical surface
    distance“ should be changed in the probe settings (see manual). For more information see the application note
    about SAR evaluation.
•   The „area scan“ measures the SAR above the dipole on a parallel plane to the surface. It is used to locate the
    approximate location of the peak SAR with 2D spline interpolation. The proposed scan uses large grid spacing
    for faster measurement; due to the symmetric field the peak detection is reliable. If a finer graphic is desired,
    the grid spacing can be reduced. Grid spacing and orientation have no influence on the SAR result.
•   The zoom scan job measures the field in a volume around the peak SAR value assessed in the previous „area“
    scan (for more information see the application note on SAR evaluation).
If the validation measurements give reasonable results, the peak 1g and 10g spatial SAR values averaged between
the two cubes and normalized to 1W dipole input power give the reference data for comparisons. The next section
analyzes the expected uncertainties of these values. Section 6 describes some additional checks for further
information or troubleshooting.

5.1.1.3   Uncertainty Budget

Please note that in the following Tables, the tolerance of the following uncertainty components depends on the
actual equipment and setup at the user location and need to be either assessed or verified on-site by the end user of
the DASY4 system:

• RF ambient conditions
• Dipole Axis to Liquid Distance
• Input power and SAR drift measurement
• Liquid permittivity - measurement uncertainty
• Liquid conductivity - measurement uncertainty

Note: All errors are given in percent of SAR, so 0.1 dB corresponds to 2.3%. The field error would be half of that.
the liquid parameter assessment give the targeted values from the dipole document. All errors are given in percent
of SAR, so 0.1dB corresponds to 2.3%. The field error would be half of that.




As of 2008-11-13                                                                                       Page 25 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08


System validation

In the table below, the system validation uncertainty with respect to the analytically assessed SAR value of a dipole
source as given in the P1528 standard is given. This uncertainty is smaller than the expected uncertainty for mobile
phone measurements due to the simplified setup and the symmetric field distribution.


Error Sources                   Uncertainty Probability Divi-     ci     ci     Standard    Standard     vi2 or
                                  Value     Distribution sor                   Uncertainty Uncertainty    veff
                                                                 1g     10g
Measurement System
Probe calibration                ± 4.8%       Normal       1      1      1      ± 4.8%       ± 4.8%        ∞
Axial isotropy                   ± 4.7%     Rectangular    √3    0.7    0.7     ± 1.9%       ± 1.9%        ∞
Hemispherical isotropy           ± 0.0%     Rectangular    √3    0.7    0.7     ± 0.0%       ± 0.0%        ∞
Boundary effects                 ± 1.0%     Rectangular    √3     1      1      ± 0.6%       ± 0.6%        ∞
Probe linearity                  ± 4.7%     Rectangular    √3     1      1      ± 2.7%       ± 2.7%        ∞
System detection limits          ± 1.0%     Rectangular    √3     1      1      ± 0.6%       ± 0.6%        ∞
Readout electronics              ± 1.0%       Normal       1      1      1      ± 1.0%       ± 1.0%        ∞
Response time                    ± 0.0%     Rectangular    √3     1      1      ± 0.0%       ± 0.0%        ∞
Integration time                 ± 0.0%     Rectangular    √3     1      1      ± 0.0%       ± 0.0%        ∞
RF ambient conditions            ± 3.0%     Rectangular    √3     1      1      ± 1.7%       ± 1.7%        ∞
Probe positioner                 ± 0.4%     Rectangular    √3     1      1      ± 0.2%       ± 0.2%        ∞
Probe positioning                ± 2.9%     Rectangular    √3     1      1      ± 1.7%       ± 1.7%        ∞
Max. SAR evaluation              ± 1.0%     Rectangular    √3     1      1      ± 0.6%       ± 0.6%        ∞
Dipole
Dipole axis toliquid distance    ± 2.0%       Normal       1      1      1      ± 1.2%       ± 1.2%        ∞
Input power and power drift      ± 4.7%     Rectangular    √3     1      1      ± 2.7%       ± 2.7%        ∞
Phantom and Set-up
Phantom uncertainty              ± 4.0%     Rectangular    √3     1      1      ± 2.3%       ± 2.3%        ∞
Liquid conductivity (target)     ± 5.0%     Rectangular    √3    0.64   0.43    ± 1.8%       ± 1.2%        ∞
Liquid conductivity (meas.)      ± 2.5%     Rectangular    1     0.64   0.43    ± 1.6%       ± 1.1%        ∞
Liquid permittivity (target)     ± 5.0%     Rectangular    √3    0.6    0.49    ± 1.7%       ± 1.4%        ∞
Liquid permittivity (meas.)      ± 2.5%     Rectangular    1     0.6    0.49    ± 1.5%       ± 1.2%        ∞
Combined Uncertainty                                                            ± 8.4%       ± 8.1%       ∞
Expanded Std. Uncertainty                                                       ± 16.8%      ± 1.2%




As of 2008-11-13                                                                                         Page 26 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08


Performance check repeatability
The repeatability check of the validation is insensitive to external effects and gives an indication of the variations in
the DASY4 measurement system, provided that the same power reading setup is used for all validations. The
repeatability estimate is given in the following table:

Error Sources                   Uncertainty Probability Divi-     ci     ci     Standard    Standard     vi2 or
                                  Value     Distribution sor                   Uncertainty Uncertainty    veff
                                                                 1g     10g
Measurement System
Probe calibration                ± 4.8%       Normal       1      1      1         0            0          ∞
Axial isotropy                   ± 4.7%     Rectangular    √3    0.7    0.7        0            0          ∞
Hemispherical isotropy           ± 0.0%     Rectangular    √3    0.7    0.7        0            0          ∞
Boundary effects                 ± 1.0%     Rectangular    √3     1      1         0            0          ∞
Probe linearity                  ± 4.7%     Rectangular    √3     1      1         0            0          ∞
System detection limits          ± 1.0%     Rectangular    √3     1      1         0            0          ∞
Readout electronics              ± 1.0%       Normal       1      1      1         0            0          ∞
Response time                    ± 0.0%     Rectangular    √3     1      1         0            0          ∞
Integration time                 ± 0.0%     Rectangular    √3     1      1         0            0          ∞
RF ambient conditions            ± 3.0%     Rectangular    √3     1      1         0            0          ∞
Probe positioner                 ± 0.4%     Rectangular    √3     1      1         0            0          ∞
Probe positioning                ± 2.9%     Rectangular    √3     1      1         0            0          ∞
Max. SAR evaluation              ± 1.0%     Rectangular    √3     1      1         0            0          ∞
Dipole
Dipole axis toliquid distance    ± 2.0%       Normal       1      1      1      ± 1.2%       ± 1.2%        ∞
Input power and power drift      ± 4.7%     Rectangular    √3     1      1      ± 2.7%       ± 2.7%        ∞
Phantom and Set-up
Phantom uncertainty              ± 4.0%     Rectangular    √3     1      1      ± 2.3%       ± 2.3%        ∞
Liquid conductivity (target)     ± 5.0%     Rectangular    √3    0.64   0.43    ± 1.8%       ± 1.2%        ∞
Liquid conductivity (meas.)      ± 2.5%     Rectangular    1     0.64   0.43    ± 1.6%       ± 1.1%        ∞
Liquid permittivity (target)     ± 5.0%     Rectangular    √3    0.6    0.49    ± 1.7%       ± 1.4%        ∞
Liquid permittivity (meas.)      ± 2.5%     Rectangular    1     0.6    0.49    ± 1.5%       ± 1.2%        ∞
Combined Uncertainty                                                            ± 5.3%       ± 4.9%       ∞
Expanded Std. Uncertainty                                                       ± 10.6%      ± 9.7%

The expected repeatability deviation is low. Excessive drift (e.g., drift in liquid parameters), partial system failures
or incorrect parameter settings (e.g., wrong probe or device settings) will lead to unexpectedly high repeatability
deviations. The repeatability gives an indication that the system operates within its initial specifications. Excessive
drift, system failure and operator errors are easily detected.




As of 2008-11-13                                                                                         Page 27 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08


5.1.1.4   Power set-up for validation
The uncertainty of the dipole input power is a significant contribution to the absolute uncertainty and the expected
deviation in interlaboratory comparisons. The values in Section 2 for a typical and a sophisticated setup are just
average values. Refer to the manual of the power meter and the detector head for the evaluation of the uncertainty
in your system. The uncertainty also depends on the source matching and the general setup. Below follows the
description of a recommended setup and procedures to increase the accuracy of the power reading:




The figure shows the recommended setup. The PM1 (incl. Att1) measures the forward power at the location of the
validation dipole connector. The signal generator is adjusted for the desired forward power at the dipole connector
and the power meter PM2 is read at that level. After connecting the cable to the dipole, the signal generator is
readjusted for the same reading at power meter PM2. If the signal generator does not allow a setting in 0.01dB
steps, the remaining difference at PM2 must be noted and considered in the normalization of the validation results.
The requirements for the components are:

•   The signal generator and amplifier should be stable (after warm-up). The forward power to the dipole should be
    above 10mW to avoid the influence of measurement noise. If the signal generator can deliver 15dBm or more,
    an amplifier is not necessary. Some high power amplifiers should not be operated at a level far below their
    maximum output power level (e.g. a 100W power amplifier operated at 250mW output can be quite noisy). An
    attenuator between the signal generator and amplifier is recommended to protect the amplifier input.
•   The low pass filter after the amplifier reduces the effect of harmonics and noise from the amplifier. For most
    amplifiers in normal operation the filter is not necessary.
•   The attenuator after the amplifier improves the source matching and the accuracy of the power head. (See
    power meter manual.) It can also be used also to make the amplifier operate at its optimal output level for noise
    and stability. In a setup without directional coupler, this attenuator should be at least 10dB.
•   The directional coupler (recommended ³ 20dB) is used to monitor the forward power and adjust the signal
    generator output for constant forward power. A medium quality coupler is sufficient because the loads (dipole
    and power head) are well matched. (If the setup is used for reflective loads, a high quality coupler with respect
    to directivity and output matching is necessary to avoid additional errors.)
•   The power meter PM2 should have a low drift and a resolution of 0.01dBm, but otherwise its accuracy has no
    impact on the power setting. Calibration is not required.
•   The cable between the coupler and dipole must be of high quality, without large attenuation and phase changes
    when it is moved. Otherwise, the power meter head PM1 should be brought to the location of the dipole for
    measuring.
•   The power meter PM1 and attenuator Att1 must be high quality components. They should be calibrated,
    preferably together. The attenuator (³10dB) improves the accuracy of the power reading. (Some higher power


As of 2008-11-13                                                                                      Page 28 of 29


CETECOM ICT Services GmbH
Calibration Data and Phantom Information to test report no.: 1-0876-01-04/08

    heads come with a built-in calibrated attenuator.) The exact attenuation of the attenuator at the frequency used
    must be known; many attenuators are up to 0.2dB off from the specified value.
•   Use the same power level for the power setup with power meter PM1 as for the actual measurement to avoid
    linearity and range switching errors in the power meter PM2. If the validation is performed at various power
    levels, do the power setting procedure at each level.
•   The dipole must be connected directly to the cable at location “X”. If the power meter has a different connector
    system, use high quality couplers. Preferably, use the couplers at the attenuator Att1 and calibrate the attenuator
    with the coupler.
•   Always remember: We are measuring power, so 1% is equivalent to 0.04dB.

5.1.1.5   Laboratory reflections
In near-field situations, the absorption is predominantly caused by induction effects from the magnetic near-field.
The absorption from reflected fields in the laboratory is negligible. On the other hand, the magnetic field around the
dipole depends on the currents and therefore on the feedpoint impedance. The feedpoint impedance of the dipole is
mainly determined from the proximity of the absorbing phantom, but reflections in the laboratory can change the
impedance slightly. A 1% increase in the real part of the feedpoint impedance will produce approximately a 1%
decrease in the SAR for the same forward power. The possible influence of laboratory reflections should be
investigated during installation. The validation setup is suitable for this check, since the validation is sensitive to
laboratory reflections. The same tests can be performed with a mobile phone, but most phones are less sensitive to
reflections due to the shorter distance to the phantom. The fastest way to check for reflection effects is to position
the probe in the phantom above the feedpoint and start a continuous field measurement in the DASY4 multimeter
window. Placing absorbers in front of possible reflectors (e.g. on the ground near the dipole or in front of a metallic
robot socket) will reveal their influence immediately. A 10dB absorber (e.g. ferrite tiles or flat absorber mats) is
probably sufficient, as the influence of the reflections is small anyway. If you place the absorber too near the
dipole, the absorber itself will interact with the reactive near-field. Instead of measuring the SAR, it is also possible
to monitor the dipole impedance with a network analyzer for reflection effects. The network analyzer must be
calibrated at the SMA connector and the electrical delay (two times the forward delay in the dipole document) must
be set in the NWA for comparisons with the reflection data in the dipole document. If the absorber has a significant
influence on the results, the absorber should be left in place for validation or measurements. The reference data in
the dipole document are produced in a low reflection environment.

5.1.1.6   Additional system checks
While the validation gives a good check of the DASY4 system components, it does not include all parameters
necessary for real phone measurements (e.g. device modulation or device positioning). For system validation
(repeatability) or comparisons between laboratories a reference device can be useful. This can be any mobile phone
with a stable output power (preferably a device whose output power can be set through the keyboard). For
comparisons, the same device should be sent around, since the SAR variations between samples can be large.
Several measurement possibilities in the DASY software allow additional tests of the performance of the DASY
system and components. These tests can be useful to localize component failures:
•   The validation can be performed at different power levels to check the noise level or the correct compensation
    of the diode compression in the probe.
•   If a pulsed signal with high peak power levels is fed to the dipole, the performance of the diode compression
    compensation can be tested. The correct crest factor parameter in the DASY software must be set (see manual).
    The system should give the same SAR output for the same averaged input power.
•   The probe isotropy can be checked with a 1D-probe rotation scan above the feedpoint. The automatic probe
    alignment procedure must be passed through for accurate probe rotation movements (optional DASY4 feature
    with a robot-mounted light beam unit). Otherwise the probe tip might move on a small circle during rotation,
    producing some additional isotropy errors in gradient fields.



As of 2008-11-13                                                                                        Page 29 of 29



Document Created: 2009-01-07 12:15:07
Document Modified: 2009-01-07 12:15:07

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC