SAR Caldata

FCC ID: TVU-C300H

RF Exposure Info

Download: PDF
FCCID_1277144

                                 TEST REPORT
                          Test Report No.: 1-2034-01-04/10-A


                                   Testing Laboratory
                   CETECOM ICT Services GmbH
                   Untertürkheimer Straße 6 – 10
                   66117 Saarbrücken/Germany
                   Phone:    + 49 681 5 98 - 0
                   Fax:      + 49 681 5 98 - 9075
                   Internet: http://www.cetecom-ict.de
                   e-mail:   info@ict.cetecom.de


                   Accredited Test Laboratory:
                   The test laboratory (area of testing) is accredited
                   according to DIN EN ISO/IEC 17025

                   DAR registration number: DAT-P-176/94-D1


     Appendix with Calibration data, Phantom certificate and system
                         validation information
                               2010-04-23




2010-04-23                                                               Page 1 of 31


                                                Test report no.: 1-2034-01-04/10-A




1     Table of contents

1   Table of contents.......................................................................................................................................2

2   Calibration report “Probe ET3DV6”.........................................................................................................3

3   Calibration report “1900 MHz System validation dipole”....................................................................14

4   Calibration certificate of Data Acquisition Unit (DAE) ........................................................................24

5   Certificate of “SAM Twin Phantom V4.0/V4.0C’’ .................................................................................25

6   Application Note System Performance Check.....................................................................................26
    6.1      Purpose of system performance check .....................................................................................26
    6.2      System Performance check procedure......................................................................................26
    6.3      Uncertainty Budget ......................................................................................................................27
    6.4      Power set-up for validation .........................................................................................................30
    6.5      Laboratory reflection ...................................................................................................................31
    6.6      Additional system checks ...........................................................................................................31




2010-04-23                                                                                                                               Page 2 of 31


                          Test report no.: 1-2034-01-04/10-A




2    Calibration report “Probe ET3DV6”




2010-04-23                                                     Page 3 of 31


Calibration Laboratory of                                                                   Schweiserinchor Kaliverdonst
Schmid & Partner                                                                            Sarvie suirse cetatome
  Enginearing AG                                                                            dorvio svizzere ditarsturs
zoughainetrause 43, 8004 Zurch, Swazeriand                                                  3riss Callbrtion Borvice

Amcredtes byihe Swise Acsredtation Service cBt)                                          Accreditaton to: SCS 108
The Buiiss Accreditation Service is one ofthe signatonis to tho EA
MoltIstoralAgrowmant fothe recogniion at o lbreioneertfcates

Glossary:
ts1                        lissue simuleting iquid
NORMiuy:e                  senalliity in froo apace
ComE                       sonsitiiy in TSL/ NORMxyz
nar                        ciode compression point
or                         crest fector (Liduty_eycle}ofh RF signal
A.B,C                      modulstion degendent Iincarzatian parometers
Polarization o             q rotation around probe axis
Polarization 8             $ rotalon around an axis that is in the plane normal to probe axis t measurement centar),
                           ie., B = 0 is normal to probe axis
Calibration is Performed According to the Following Standards:
      a). IEEE Sid 1028—2003, "IFRE Recommended Practice for De:ermining the Peak Spattal—Averaged Spo: ic
          Absorplion Rete (SAR) in the Humen Head ffom Wiroless Communications Devicas: Meaguremont
          Techniques‘. December 2003
      b) 120 62200—1, Procedure lo messure the Sperific Absarption Rate (3AR)for hand.hcld devions used in close
          proximily to he ear (frequancy range of 300 MHz to 5 GHeY\, February 2008
Methnds Applied and Interpretation of Parameters:
     NORM: y.z: Auscasod for E—feld polarizstion & = 0 (f s 900 MHz in TER—oolk £> 1800 MHz HZZwangmdn]
     NORMsy z are onl intormediacevalues, ie., ie uncertaintion af NORMxy,z does not effect the E—feld
           uncertainty inside TSL (see be ow CanvF)
      *    NORM(s,y,z = NORyz * fequency_response (see Frequency Rosponse Chai), This Iinsarizntion in
           implomonted in BASY4 software versions laler than 4.2. Ihe uncertainty of the Irequency response is included
           in the stazed uncertaintyof Cony
      +    DGPx y,z: DGP are numerical fnearization peramelers assossod based on the data of powerswoop with CW
           signal (no uncertainty required). BCP doss not dopand an frequency nor media
      +0   Argie) Biopie; Cagiz, VBeyA, 8. 2 are numerical Inearization parametrs assessed based on the deta of
           power swaep      specifie moduleliansignial. The parametors do not depend on Treauency nor modia, WRis the
           maximum calibration range expressed in RMS valtage ncross the dade
      +    Convé and Bcundary Effect Parameters. Asassed in fhitphantam using E—feld (or Temperalure Transfer
           Standard for f : 800 MHe) and insit waveguo using analyteal feld oistribut ons basos on power
           measurements for I> 800 MHz. Thesame sctuns are uses for assessment of the paramatore applied for
           boundary compenantion (alphn,depth) of whichtypieal uncerlainly values are givan, These paramelers are
           used in DAGYA software to improve probe accuracy close to the boundary. The sensiy ty in TSL corresponcs
           to NORMc.y.z * ConyF wherehy the uncertairty corresponds to tnat given for CoryF. A frequency dependent
           GanvF is used in DASY version 4.4 and higher which allows extending the validty from # 50 Mto + 100
           whiz
      *    Spharical isoltopy (3D deviation fromisotropy} in fela of low gradients realized using a flxt phantomn
           exposed by a peich antenna
      *    Sensor Offsel. The sensor offset corresponds to the offset of virlual measurement canter from the probe tip
           (on probe axis). No tolerance required




Certfieato No: E73—1880.                                    Page 2t 11


                                                                               CETECOM*


ET3DVG SN:1559                                                                 January 20, 2010




                           Probe ET3DV6

                                           SN:1559
                           Manufactured:                    December 1, 2000
                           Last calibrated                  January 14, 2009
                           Repaired:                        January 11, 2010
                           Recalibrated:                    January 20, 2010

                                   Calibrated for DASY Systems
                                    (Note: non.comeatble with OASYZ system!)




  Ceifsate No: ET3 toos_Jonnd                 Page s or11


ET3DV6 SN:1559                                                                                            January 20, 2010


 DASY — Parameters of Probe: ET3DV6 SN:1559

Basic Calibration Parameters
                                                     .            Sensor X|   SensorY|     Sensor Z nc (ke2)

Norm (uvi(xtim)*)"                                                  187         153          170          +10.1%
boP imyj"                                                           90.8        820         ar2


Modulation Calibration Parameters
un          (communication System Name             PaR               A           a            c            ve       Une®
                                                                     e         abuv                        my      en
 10000      ow                                     aco|    x           200         0.00           soo      an0     £n5%
                                                           v           a.00        a.00)          soul     300
                                                           2           200         a.co|          1.00|    00




 The reporied uncertainly of measurement is stated as the standard uncertainty of measurement muliplied
 by the coverage factor k=2, which for a normal cistribution corresponds to a coverage probabilty of
 approximately 95%

* The ce ainin ofhlonY Zeonatea ror® c           o Ts use Rages mas
* Norercalinanon promeer uncenainyn on
  snsmnc is seaminad usths eacetan on fo Inawrespore snn ns ecalo oi #iduonent i croresee forthesquare o neic



     Gomicate No t13—1388 amnt0                   Pagedot11


ETSDVG SN: 1859                                                                                             January 20, 2010




DASY — Parameters of Probe: ET3DV6 SN: 1559

Calibration Parameter Dotormined in Head Tissue Simulating Medi


ukc         WalictyItie]®    Pormiubiy           Conductviy ComFX GomFY ComFZ                       Algha      Dosth Uns the?)
aso         602100           aa5=9%              oara se       Tes    res   r2s                      ore        165 +33
8ab         502100           #1059%              0802 5%       840    a4o   nan                      oce        235 t10%
s00         £60 2100         41.628%             Oar s 9%      a21    a2:   a21                      oss        245 s110%
16«0        «s0/«100         ao315%              128=5%        s49    sao   sae                      oa0        ar an0%
1750        £50/2100         a0t25%              137 285%      s2s    520   520                      oar        a2ss anio%
1850        £50/2100         a005%               1401 5%       «87    a8r   aar                      oé8        247 +110%
2160        a 5042 101       sa.7 a5%            1534 8%       aas    433   18                       099        17 1108
2450        5072100          sh220%              1802 5%       aa42   40    a42                      aso        ar0 1104


    wo a tuo a           ow in for BASY v1 4 and nahar (sen 2e           ty is is RBE e she ConP unsarainy at alucon recumcy
e t uneanereforine nsostes nsuensy nans




   Cortticats No: 73—1080 Janto                            Pasas ol 11


ET3DVs SN:1559                                                                             January 20, 2010




DASY — Parameters of Probe: ET3DV6 SN:1559

Calibration Parameter Determined in Body Tissue Simulating Media


tm         Valigty [ke]®     Pormitiviy   Conducthrty ComsFX ComEY ComFZ           Aigha     Bepth Un en
aso        £50/2 100         5o7 =5%      0.90 2 5%       iss    Tes   mss          020       ra7 cr30%
s5         £507% 100         5622 0%      a97 4 5w             820   820   a20     oas        258   =11.0%
900        #80 /z 100        ss028%       1.05 1 5%            o0«   sou   6e      oan        298   =11.0%
16e0       a50r1 109         538 +5%      n40a5%               sas   sas   538     oss        aas   £11.0%
175          a r+ 10         s3425%       nasa sn              «83   «43   «3      660        dit   a110%
1950       =5072 100         sasase       182485%              167   167   «87     on         asn   a11.0%
2150       =50/+ 100         530 13%      175« 5%              «42   «22   a42     ose        220   »11.0%
2«so                         sa7a         195 46%              a04   «04   04      69e        are   ad0%



Thvl in 100 i onl ns for DABTY se nc Mavo Page . T—wuncntl ie h RSS t t Csccranty t sniriten foquency
and ve unesrtamy ‘or ie inaested neseney bens.




   Serticate No: ET9—1550.Jonto                   RageGof 11


                                                                                                                    CETECOM*


ET3DVs SN:1550                                                                                                      January 20, 2010


                                                         Frequency Response of E—Field
                                                              {TEM—Collif10 EXX, Waveguide: R22)




                                                             .
                        Frequency response (nomalizd)




                                                        a*




                                                                                     doco     2oo
                                                                                   (tina
                                                                                             Teore         |
                                                        Uncertsinty of Frequency Response of E—fild: 2 6.3% (k=2)




  Carthecia No: 2T3—18580_.ento                                           Page 7 n


                                                                                             CETECOM*


ET3DV6 SN:1569                                                                                January 20, 2010



                                  Receiving Pattern (¢), 8 = 0°

                    1= 600 MHz, TeM if10EXX                       1= 1800 MHz, wG ra2                ‘




                                                                                      | ~o—soue
                                                                                      | ~m—st ie
                                                                                   *\ co—mcune
                                                                                        ~m—rmoome2
                                                                                        <a—280 M

                o          «o         120        iro         2i         a»
                                                 +0



                                Uncenainty of Aial Isatragy Assossment: + 0.5% (ket)




  Geriicts NoCT0—1580..laord                    Pege®of11


                                                      .                                         CETECOM*



ET3DVG SN: 1559                                                                                 January 20, 2010


                                               Dynamic Range F(SARpsq)
                                                   (Wavequide R22, != 1800 MHz)

                                    1Eme


                                    «es


                                    sero
                  tnpur Signat u)




                                     sros


                                    reooe


                                    mwn


                                    wal.|
                                      noom     can        om           en       1       10
                                                                   SAfnwice‘1
                                                ~#~—rotcomponesid               ~o—censersoie




                                     som
                                             TT                o            ;
                                                                   SA fnWem‘]


                                             Uncertainty of Linoanty Assessment 2 0.0% (ke2)




  Certfnts No: ET355_lanto                                 Page 0t 11


                                                                                                        CETECOM*


ET3DV6 SN:1550                                                                                           January 20, 2010



                                        Conversion Factor Assessment

                     1= 900 Mtte, WGLS R0 (head)                             £= 1750 Miz, WGLS Rz2 fread)
                                                                  wo                      |
santmwen‘ w




              oo
                                 ztm                                                   #trim
                   comnsniy in       o—Nemimmants                           ~0—»ralnics          ——Messumine



                                        Deviation from Isotropy in HSL
                                                     Error (6, 8), £= 800 MHz
                                                                                     Etor (oB)




                                       im ow memow mousa murtan i  e
                                       becoam momce micom mosea umc


                                      Unceriainly of Sphorical Isotropy Assossmont: & 2.6% (K=2)




      Centzate No: CT9—1980,.Janto                      2age 10 ot«


                                                          CETECOM*


ETSDVS SN:1559                                                January 20, 2010




Other Probe Parameters

Sensor Arrangement                                                    Triangular|
(Connector Angle (*)                                              Not applicable
Mecranical Surface Detect on Mode                         _             enabled
Optical Surlace Detection Mode                                          enabled
Probe Overall Longth                                                    337 mm
Probe Body Diemater                                                      10 mm

Tip Lengh                                                                10 mm
Tip Diemater                                                            8.8 mm
Probe Tip to Sensor X Cllbration Point                                  27 mm
Probe Tip to Songor Y Calbration Point                                  2.7 mm
Probe Tio to Sensor 2 Callbration Point                                 2.7 mm
Recommended Measurement Distancs from Surtace                             4 mm




   Gonbsats No: 13—1580 centa               Page 1 ott1


                          Test report no.: 1-2034-01-04/10-A




3    Calibration report “1900 MHz System validation dipole”




2010-04-23                                                     Page 14 of 31


                                                                                    CETECOM*


 Calibration Laboratory of                                                   Schweizoriseir Kaherdienst
 Schmid & Partner                                                            Service susse détnlonnage
   Engineering AG                                                            Berveto avizeoro d taratura
 Zounhovastrease 49, 004 Zurich, Sultzodand                                  Suiss Caltbration Sarvice

 Aceredtag by t$s AccreotatonSomviee h)                                   Accrediiation No.: SCS 108
 The Siies Accroditation Serviceis one of the #ignsteriew o the EA
 Muifeteral Agreement fo: the racagiton of cairetion centlstos

 Glossary:
 TSL                       fissue simulating liquid
 ConvF                     sensitvity in LSL / NORMx.y.2
 NA                        not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Sid 1528—2003, IEEE Recommended Practica for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measuremert Techniques®, December 2003
   b) CENELEC EN 50361, "Basic standard for the measurement of Specific Absorption Rate
      related to hunan exposureto electromegnstic fields from mobile phones (300 MHz — 3
      GHz), July 2001
   c Federal Communications Commission Office of Engineering & Technology (FCC OET)
      ‘Evaluating Compliance with FCC Guidelinas for Human Exposure to Radiofrequency
      Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
      Portable Devices with FGC Limits for Human Exposure to Radiofrequency Emissions",
      Supplement C (Edition 01—01) to Bullstin 65

Additional Documentation:
    d) DASY4/S System Handbook

Methads Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipcle is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         paralle! to the body axis.
    *    Feed Point impedance and Return Loss: These paramaters are measured with the dipole
         positioned under the liquid filed phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensuros low
         reflected power. No uncertainty required.
    «.   Electrical Delay: One—way delay between the SMA connector and the anterina feed point
         No uncertainty required.

    *    SAR measured: SAR measured at the stated antenna input power.
    *    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna.
         connector
    *    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.



Gertficats No: D1900V2—5d008_Augo®                          Page 2 ot 9


                                                                                                    cErecom"

Measuremont Conditions
    DASY system conflguration, as furas not given on page1
     DASY Veraion                                            nasys                                    vso
     Extrapolation                                   Advanced Emrapolation
     Phantom                                       Modular Fla: Phantom ¥S 0
     Distance Dipole Conter — TSL                            10 mm                                 with Spacer
     Zoom Scan Resolution                               oh.ddz ~5 mm
     Frequency                                          1900 Mz a 1 lkie

Head TSL parameters
    The following parametore and caiouiations were appliec.
                                                              Tomporature           Permittivity        Conductivity
      Nominal Head TSL paramaters                                220°C                  «00              140 mhaim
      moasured Hoad TSL paramotors                            ir20 =02)°C            «o8 a8%          145 mhoima 6 *
      Head TSL temperature during test                        20 02)°C


SAR result with Head TSL

      SAR averaged over 1 em" (1 u) of Hoad TSL                conation                                             —]
      SAR measured                                        250 mW inpat powor                  10.1 mw/g
      SAR normatized                                       normaliend to TW                  40.4 mW/a
      $AR for nomina‘ Haed TSL paremeters ‘                normalized to 1W            89.7 mW / g » 17.0.% (ke2)

      SAR averaged over 10 em‘ (10 u) of Hoad TSL              Condition
      S4R measurad                                        250 mW input pomer                 528 mW q
      SAR normalized                                       normalized io TW                  21. mW 1g
    | SAR for nominal Haad TSL parameters                  normalizea to1w             21.0 m¥ 7 g 16.5 % (k=2)




* Correction to nominal TSL paramelers according to d}, chapter "8AR Sunaithiton®
Ganiicats No: D1960V2—50009_Augoo                      Page 3 of 0


             Test report no.: 1-2034-01-04/10-A




2010-04-23                                        Page 17 of 31


             Test report no.: 1-2034-01-04/10-A




2010-04-23                                        Page 18 of 31


             Test report no.: 1-2034-01-04/10-A




2010-04-23                                        Page 19 of 31


             Test report no.: 1-2034-01-04/10-A




2010-04-23                                        Page 20 of 31


             Test report no.: 1-2034-01-04/10-A




2010-04-23                                        Page 21 of 31


             Test report no.: 1-2034-01-04/10-A




2010-04-23                                        Page 22 of 31


 Impedance Measurement Plot for Body TSL



                                                       20 ney cans cercorer
     ERD sn      a uis             massose   somine srirlpa       1 se.000 oon hike

     x
                                                                                      UE Harkers
     on                                                                                arauues a
                                                                                        RrEREEE]
     se                                                                               1.50000 ols


     m
     is°

      +
     cie                                              2—2207; an a ore.o00 ooo iz

                                                                                      tH2 Harkers
     Cor                                                                              weaness
                                                                                      Toatbd beoe


     0y
     es
      a
              Staet 1 cpa.oon pee iz                         Stor 2 100.a0a on0 mix




Gortiicato No: D1000V2—531_Meyc0                 Page 0 o0


                           Test report no.: 1-2034-01-04/10-A




4    Calibration certificate of Data Acquisition Unit (DAE)




2010-04-23                                                      Page 24 of 31


                          Test report no.: 1-2034-01-04/10-A




5    Certificate of “SAM Twin Phantom V4.0/V4.0C’’




2010-04-23                                                     Page 25 of 31


                                    Test report no.: 1-2034-01-04/10-A




6     Application Note System Performance Check

6.1   Purpose of system performance check
The system performance check verifies that the system operates within its specifications. System and
operator errors can be detected and corrected. It is recommended that the system performance check is
performed prior to any usage of the system in order to guarantee reproducible results.
The measurement of the Specific Absorption Rate (SAR) is a complicated task and the result depends on
the proper functioning of many components and the correct settings of many parameters. Faulty results due
to drift, failures or incorrect parameters might not be recognized, since they often look similar in distribution
to the correct ones. The Dosimetric Assessment System DASY4 incorporates a system performance check
procedure to test the proper functioning of the system. The system performance check uses normal SAR
measurements in a simplified setup (the flat section of the SAM Twin Phantom) with a well characterized
source (a matched dipole at a specified distance). This setup was selected to give a high sensitivity to all
parameters that might fail or vary over time (e.g., probe, liquid parameters, and software settings) and a low
sensitivity to external effects inherent in the system (e.g., positioning uncertainty of the device holder). The
system performance check does not replace the calibration of the components. The accuracy of the system
performance check is not sufficient for calibration purposes. It is possible to calculate the field quite
accurately in this simple setup; however, due to the open field situation some factors (e.g., laboratory
reflections) cannot be accounted for. Calibrations in the flat phantom are possible with transfer calibration
methods, using either temperature probes or calibrated E-field probes. The system performance check also
does not test the system performance for arbitrary field situations encountered during real measurements of
mobile phones. These checks are performed at SPEAG by testing the components under various conditions
(e.g., spherical isotropy measurements in liquid, linearity measurements, temperature variations, etc.), the
results of which are used for an error estimation of the system. The system performance check will indicate
situations where the system uncertainty is exceeded due to drift or failure.

6.2   System Performance check procedure
Preparation
The conductivity should be measured before the validation and the measured liquid parameters must be
entered in the software. If the measured values differ from targeted values in the dipole document, the liquid
composition should be adjusted. If the validation is performed with slightly different (measured) liquid
parameters, the expected SAR will also be different. See the application note about SAR sensitivities for an
estimate of possible SAR deviations. Note that the liquid parameters are temperature dependent with
approximately – 0.5% decrease in permittivity and + 1% increase in conductivity for a temperature decrease
of 1° C. The dipole must be placed beneath the flat phantom section of the Generic Twin Phantom with the
correct distance holder in place. The distance holder should touch the phantom surface with a light pressure
at the reference marking (little hole) and be oriented parallel to the long side of the phantom. Accurate
positioning is not necessary, since the system will search for the peak SAR location, except that the dipole
arms should be parallel to the surface. The device holder for mobile phones can be left in place but should
be rotated away from the dipole. The forward power into the dipole at the dipole SMA connector should be
determined as accurately as possible. See section 4 for a description of the recommended setup to measure
the dipole input power. The actual dipole input power level can be between 20mW and several watts. The
result can later be normalized to any power level. It is strongly recommended to note the actually used power
level in the „comment“-window of the measurement file; otherwise you loose this crucial information for later
reference.




2010-04-23                                                                                       Page 26 of 31


                                      Test report no.: 1-2034-01-04/10-A


System Performance Check
The DASY4 installation includes predefined files with recommended procedures for measurements and
validation. They are read-only document files and destined as fully defined but unmeasured masks, so you
must save the finished validation under a different name. The validation document requires the Generic Twin
Phantom, so this phantom must be properly installed in your system. (You can create your own
measurement procedures by opening a new document or editing an existing document file). Before you start
the validation, you just have to tell the system with which components (probe, medium, and device) you are
performing the validation; the system will take care of all parameters. After the validation, which will take
about 20 minutes, the results of each task are displayed in the document window. Selecting all measured
tasks and opening the predefined “validation” graphic format displays all necessary information for validation.
A description of the different measurement tasks in the predefined document is given below, together with
the information that can be deduced from their results:
•     The „reference“ and „drift“ measurements are located at the beginning and end of the batch process.
      They measure the field drift at one single point in the liquid over the complete procedure. The indicated
      drift is mainly the variation of the amplifier output power. If it is too high (above ± 0.1dB) the validation
      should be repeated; some amplifiers have very high drift during warm-up. A stable amplifier gives drift
      results in the DASY4 system below ± 0.02 dB.
•     The „surface check“ measurement tests the optical surface detection system of the DASY4 system by
      repeatedly detecting the surface with the optical and mechanical surface detector and comparing the
      results. The output gives the detecting heights of both systems, the difference between the two systems
      and the standard deviation of the detection repeatability. Air bubbles or refraction in the liquid due to
      separation of the sugar-water mixture gives poor repeatability (above ± 0.1mm). In that case it is better
      to abort the validation and stir the liquid. The difference between the optical surface detection and the
      actual surface depends on the probe and is specified with each probe. (It does not depend on the
      surface reflectivity or the probe angle to the surface within ± 30°.) However, varying breaking indices of
      different liquid compositions might also influence the distance. If the indicated difference varies from the
      actual setting, the probe parameter „optical surface distance“ should be changed in the probe settings
      (see manual). For more information see the application note about SAR evaluation.
•     The „area scan“ measures the SAR above the dipole on a parallel plane to the surface. It is used to
      locate the approximate location of the peak SAR with 2D spline interpolation. The proposed scan uses
      large grid spacing for faster measurement; due to the symmetric field the peak detection is reliable. If a
      finer graphic is desired, the grid spacing can be reduced. Grid spacing and orientation have no influence
      on the SAR result.
•     The zoom scan job measures the field in a volume around the peak SAR value assessed in the previous
      „area“ scan (for more information see the application note on SAR evaluation).
If the validation measurements give reasonable results, the peak 1g and 10g spatial SAR values averaged
between the two cubes and normalized to 1W dipole input power give the reference data for comparisons.
The next section analyzes the expected uncertainties of these values. Section 6 describes some additional
checks for further information or troubleshooting.

6.3     Uncertainty Budget
Please note that in the following Tables, the tolerance of the following uncertainty components depends on
the actual equipment and setup at the user location and need to be either assessed or verified on-site by the
end user of the DASY4 system:
• RF ambient conditions
• Dipole Axis to Liquid Distance
• Input power and SAR drift measurement
• Liquid permittivity - measurement uncertainty
• Liquid conductivity - measurement uncertainty
Note: All errors are given in percent of SAR, so 0.1 dB corresponds to 2.3%. The field error would be half of
that. The liquid parameter assessment give the targeted values from the dipole document. All errors are
given in percent of SAR, so 0.1dB corresponds to 2.3%. The field error would be half of that.




2010-04-23                                                                                         Page 27 of 31


                                     Test report no.: 1-2034-01-04/10-A


System validation

In the table below, the system validation uncertainty with respect to the analytically assessed SAR
value of a dipole source as given in the P1528 standard is given. This uncertainty is smaller than
the expected uncertainty for mobile phone measurements due to the simplified setup and the
symmetric field distribution.

Error Sources                  Uncertainty   Probability    Divi- ci    ci     Standard      Standard      vi2
                               Value         Distribution   sor 1g      10g    Uncertainty   Uncertainty   or
                                                                               1g            10g           veff

Measurement System
Probe calibration              ± 4.8%        Normal         1    1      1      ± 4.8%        ± 4.8%        ∞
Axial isotropy                 ± 4.7%        Rectangular    √3   0.7    0.7    ± 1.9%        ± 1.9%        ∞
Hemispherical isotropy         ± 0.0%        Rectangular    √3   0.7    0.7    ± 0.0%        ± 3.9%        ∞
Boundary effects               ± 1.0%        Rectangular    √3   1      1      ± 0.6%        ± 0.6%        ∞
Probe linearity                ± 4.7%        Rectangular    √3   1      1      ± 2.7%        ± 2.7%        ∞
System detection limits        ± 1.0%        Rectangular    √3   1      1      ± 0.6%        ± 0.6%        ∞
Readout electronics            ± 1.0%        Normal         1    1      1      ± 1.0%        ± 1.0%        ∞
Response time                  ± 0.0%        Rectangular    √3   1      1      ± 0.0%        ± 0.0%        ∞
Integration time               ± 0.0%        Rectangular    √3   1      1      ± 0.0%        ± 0.0%        ∞
RF ambient conditions          ± 3.0%        Rectangular    √3   1      1      ± 1.7%        ± 1.7%        ∞
Probe positioner               ± 0.4%        Rectangular    √3   1      1      ± 0.2%        ± 0.2%        ∞
Probe positioning              ± 2.9%        Rectangular    √3   1      1      ± 1.7%        ± 1.7%        ∞
Max. SAR evaluation            ± 1.0%        Rectangular    √3   1      1      ± 0.6%        ± 0.6%        ∞
Test Sample Related
Dipole axis to liquid          ± 2.0%        Normal         1    1      1      ± 1.2%        ± 1.2%        ∞
distance
Power drift                    ± 4.7%        Rectangular    √3   1      1      ± 2.7%        ± 2.7%        ∞
Phantom and Set-up
Phantom uncertainty            ± 4.0%        Rectangular    √3   1      1      ± 2.3%        ± 2.3%        ∞
Liquid conductivity (target)   ± 5.0%        Rectangular    √3   0.64   0.43   ± 1.8%        ± 1.2%        ∞
Liquid conductivity (meas.)    ± 2.5%        Normal         1    0.64   0.43   ± 1.6%        ± 1.1%        ∞
Liquid permittivity (target)   ± 5.0%        Rectangular    √3   0.6    0.49   ± 1.7%        ± 1.4%        ∞
Liquid permittivity (meas.)    ± 2.5%        Normal         1    0.6    0.49   ± 1.5%        ± 1.2%        ∞
Combined Uncertainty                                                           ± 8.4%        ± 8.1%
Expanded Std.                                                                  ± 16.8%       ± 16.2%
Uncertainty




2010-04-23                                                                                      Page 28 of 31


                                     Test report no.: 1-2034-01-04/10-A


Performance check repeatability
The repeatability check of the validation is insensitive to external effects and gives an indication of the
variations in the DASY4 measurement system, provided that the same power reading setup is used for all
validations. The repeatability estimate is given in the following table:

Error Sources                  Uncertainty   Probability    Divi- ci     ci     Standard      Standard        vi2
                               Value         Distribution   sor 1g       10g    Uncertainty   Uncertainty     or
                                                                                1g            10g             veff

Measurement System
Probe calibration              ± 4.8%        Normal         1     1      1      0             0               ∞
Axial isotropy                 ± 4.7%        Rectangular    √3    0.7    0.7    0             0               ∞
Hemispherical isotropy         ± 0.0%        Rectangular    √3    0.7    0.7    0             0               ∞
Boundary effects               ± 1.0%        Rectangular    √3    1      1      0             0               ∞
Probe linearity                ± 4.7%        Rectangular    √3    1      1      0             0               ∞
System detection limits        ± 1.0%        Rectangular    √3    1      1      0             0               ∞
Readout electronics            ± 1.0%        Normal         1     1      1      0             0               ∞
Response time                  ± 0.0%        Rectangular    √3    1      1      0             0               ∞
Integration time               ± 0.0%        Rectangular    √3    1      1      0             0               ∞
RF ambient conditions          ± 3.0%        Rectangular    √3    1      1      0             0               ∞
Probe positioner               ± 0.4%        Rectangular    √3    1      1      0             0               ∞
Probe positioning              ± 2.9%        Rectangular    √3    1      1      0             0               ∞
Max. SAR evaluation            ± 1.0%        Rectangular    √3    1      1      0             0               ∞
Test Sample Related
Dipole axis to liquid          ± 2.0%        Normal         1     1      1      ± 1.2%        ± 1.2%          ∞
distance
Power drift                    ± 4.7%        Rectangular    √3    1      1      ± 2.7%        ± 2.7%          ∞
Phantom and Set-up
Phantom uncertainty            ± 4.0%        Rectangular    √3    1      1      ± 2.3%        ± 2.3%          ∞
Liquid conductivity (target)   ± 5.0%        Rectangular    √3    0.64   0.43   ± 1.8%        ± 1.2%          ∞
Liquid conductivity (meas.)    ± 2.5%        Normal         1     0.64   0.43   ± 1.6%        ± 1.1%          ∞
Liquid permittivity (target)   ± 5.0%        Rectangular    √3    0.6    0.49   ± 1.7%        ± 1.4%          ∞
Liquid permittivity (meas.)    ± 2.5%        Normal         1     0.6    0.49   ± 1.5%        ± 1.2%          ∞
Combined Uncertainty                                                            ± 5.3%        ± 4.9%
Expanded Std.                                                                   ± 10.6%       ± 9.7%
Uncertainty

The expected repeatability deviation is low. Excessive drift (e.g., drift in liquid parameters), partial system
failures or incorrect parameter settings (e.g., wrong probe or device settings) will lead to unexpectedly high
repeatability deviations. The repeatability gives an indication that the system operates within its initial
specifications. Excessive drift, system failure and operator errors are easily detected.




2010-04-23                                                                                        Page 29 of 31


                                     Test report no.: 1-2034-01-04/10-A



6.4     Power set-up for validation
The uncertainty of the dipole input power is a significant contribution to the absolute uncertainty and the
expected deviation in interlaboratory comparisons. The values in Section 2 for a typical and a sophisticated
setup are just average values. Refer to the manual of the power meter and the detector head for the
evaluation of the uncertainty in your system. The uncertainty also depends on the source matching and the
general setup. Below follows the description of a recommended setup and procedures to increase the
accuracy of the power reading:




The figure shows the recommended setup. The PM1 (incl. Att1) measures the forward power at the location
of the validation dipole connector. The signal generator is adjusted for the desired forward power at the
dipole connector and the power meter PM2 is read at that level. After connecting the cable to the dipole, the
signal generator is readjusted for the same reading at power meter PM2. If the signal generator does not
allow a setting in 0.01dB steps, the remaining difference at PM2 must be noted and considered in the
normalization of the validation results. The requirements for the components are:
•     The signal generator and amplifier should be stable (after warm-up). The forward power to the dipole
      should be above 10mW to avoid the influence of measurement noise. If the signal generator can deliver
      15dBm or more, an amplifier is not necessary. Some high power amplifiers should not be operated at a
      level far below their maximum output power level (e.g. a 100W power amplifier operated at 250mW
      output can be quite noisy). An attenuator between the signal generator and amplifier is recommended to
      protect the amplifier input.
•     The low pass filter after the amplifier reduces the effect of harmonics and noise from the amplifier. For
      most amplifiers in normal operation the filter is not necessary.
•     The attenuator after the amplifier improves the source matching and the accuracy of the power head.
      (See power meter manual.) It can also be used also to make the amplifier operate at its optimal output
      level for noise and stability. In a setup without directional coupler, this attenuator should be at least
      10dB.
•     The directional coupler (recommended ³ 20dB) is used to monitor the forward power and adjust the
      signal generator output for constant forward power. A medium quality coupler is sufficient because the
      loads (dipole and power head) are well matched. (If the setup is used for reflective loads, a high quality
      coupler with respect to directivity and output matching is necessary to avoid additional errors.)
•     The power meter PM2 should have a low drift and a resolution of 0.01dBm, but otherwise its accuracy
      has no impact on the power setting. Calibration is not required.
•     The cable between the coupler and dipole must be of high quality, without large attenuation and phase
      changes when it is moved. Otherwise, the power meter head PM1 should be brought to the location of
      the dipole for measuring.
•     The power meter PM1 and attenuator Att1 must be high quality components. They should be calibrated,
      preferably together. The attenuator (³10dB) improves the accuracy of the power reading. (Some higher
      power heads come with a built-in calibrated attenuator.) The exact attenuation of the attenuator at the
      frequency used must be known; many attenuators are up to 0.2dB off from the specified value.




2010-04-23                                                                                       Page 30 of 31


                                    Test report no.: 1-2034-01-04/10-A


•     Use the same power level for the power setup with power meter PM1 as for the actual measurement to
      avoid linearity and range switching errors in the power meter PM2. If the validation is performed at
      various power levels, do the power setting procedure at each level.
•     The dipole must be connected directly to the cable at location “X”. If the power meter has a different
      connector system, use high quality couplers. Preferably, use the couplers at the attenuator Att1 and
      calibrate the attenuator with the coupler.
•     Always remember: We are measuring power, so 1% is equivalent to 0.04dB.

6.5     Laboratory reflection
In near-field situations, the absorption is predominantly caused by induction effects from the magnetic near-
field. The absorption from reflected fields in the laboratory is negligible. On the other hand, the magnetic field
around the dipole depends on the currents and therefore on the feed point impedance. The feed point
impedance of the dipole is mainly determined from the proximity of the absorbing phantom, but reflections in
the laboratory can change the impedance slightly. A 1% increase in the real part of the feed point impedance
will produce approximately a 1% decrease in the SAR for the same forward power. The possible influence of
laboratory reflections should be investigated during installation. The validation setup is suitable for this
check, since the validation is sensitive to laboratory reflections. The same tests can be performed with a
mobile phone, but most phones are less sensitive to reflections due to the shorter distance to the phantom.
The fastest way to check for reflection effects is to position the probe in the phantom above the feed point
and start a continuous field measurement in the DASY4 multi-meter window. Placing absorbers in front of
possible reflectors (e.g. on the ground near the dipole or in front of a metallic robot socket) will reveal their
influence immediately. A 10dB absorber (e.g. ferrite tiles or flat absorber mats) is probably sufficient, as the
influence of the reflections is small anyway. If you place the absorber too near the dipole, the absorber itself
will interact with the reactive near-field. Instead of measuring the SAR, it is also possible to monitor the
dipole impedance with a network analyzer for reflection effects. The network analyzer must be calibrated at
the SMA connector and the electrical delay (two times the forward delay in the dipole document) must be set
in the NWA for comparisons with the reflection data in the dipole document. If the absorber has a significant
influence on the results, the absorber should be left in place for validation or measurements. The reference
data in the dipole document are produced in a low reflection environment.

6.6     Additional system checks
While the validation gives a good check of the DASY4 system components, it does not include all
parameters necessary for real phone measurements (e.g. device modulation or device positioning). For
system validation (repeatability) or comparisons between laboratories a reference device can be useful. This
can be any mobile phone with a stable output power (preferably a device whose output power can be set
through the keyboard). For comparisons, the same device should be sent around, since the SAR variations
between samples can be large. Several measurement possibilities in the DASY software allow additional
tests of the performance of the DASY system and components. These tests can be useful to localize
component failures:
•     The validation can be performed at different power levels to check the noise level or the correct
      compensation of the diode compression in the probe.
•     If a pulsed signal with high peak power levels is fed to the dipole, the performance of the diode
      compression compensation can be tested. The correct crest factor parameter in the DASY software must
      be set (see manual). The system should give the same SAR output for the same averaged input power.
•     The probe isotropy can be checked with a 1D-probe rotation scan above the feed point. The automatic
      probe alignment procedure must be passed through for accurate probe rotation movements (optional
      DASY4 feature with a robot-mounted light beam unit). Otherwise the probe tip might move on a small
      circle during rotation, producing some additional isotropy errors in gradient fields.




2010-04-23                                                                                        Page 31 of 31



Document Created: 2010-05-03 10:19:56
Document Modified: 2010-05-03 10:19:56

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC