SAR Caldata

FCC ID: TVU-A510H

RF Exposure Info

Download: PDF
FCCID_1581054

                                 TEST REPORT
                            Test Report No.: 1-3741-01-05/11


                                   Testing Laboratory
                   CETECOM ICT Services GmbH
                   Untertürkheimer Straße 6 – 10
                   66117 Saarbrücken/Germany
                   Phone:    + 49 681 5 98 - 0
                   Fax:      + 49 681 5 98 - 9075
                   Internet: http://www.cetecom.com
                   e-mail:   ict@cetecom.com


                   Accredited Test Laboratory:
                   The test laboratory (area of testing) is accredited
                   according to DIN EN ISO/IEC 17025

                   DAR registration number: DAT-P-176/94-D1


     Appendix with Calibration data, Phantom certificate and system
                         validation information
                               2011-09-09




2011-09-09                                                               Page 1 of 29


                                                  Test report no.: 1-3741-01-05/11




1     Table of contents

1   Table of contents .......................................................................................................................................2

2   Calibration report “Probe ET3DV6”.........................................................................................................3

3   Calibration report “1900 MHz System validation dipole”....................................................................14

4   Calibration certificate of Data Acquisition Unit (DAE) ........................................................................22

5   Certificate of “SAM Twin Phantom V4.0/V4.0C’’ .................................................................................23

6   Application Note System Performance Check ..................................................................................... 24
    6.1       Purpose of system performance check ..................................................................................... 24
    6.2       System Performance check procedure ...................................................................................... 24
    6.3       Uncertainty Budget ...................................................................................................................... 25
    6.4       Power set-up for validation .........................................................................................................28
    6.5       Laboratory reflection ................................................................................................................... 29
    6.6       Additional system checks ...........................................................................................................29




2011-09-09                                                                                                                                Page 2 of 29


                          Test report no.: 1-3741-01-05/11




2    Calibration report “Probe ET3DV6”




2011-09-09                                                   Page 3 of 29


                                                                                                          CETECOM


Calibration Laboratory of                                                                    Schweizerischer Kalibrierdienst
Schmid & Partner                                                                             Service suisse d‘étalonnage
  Engineering AG                                                                             Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                 Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                       Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Muitilateral Agreement for the recognition of calibration certificates


Glossary:
TSL                          tissue simulating liquid
NORMx,y,z                    sensitivity in free space
ConvF                        sensitivity in TSL / NORMx,y,z
DCP                          diode compression point
CF                           crest factor (1/duty_cycle) of the RF signal
A, B, C                      modulation dependent linearization parameters
Polarization                 q rotation around probe axis
Polarization 8               8 rotation around an axis thatis in the plane normal to probe axis (at measurement center),
                             .e., 8 = 0 is normal to probe axis

Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged Specific
       Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
       Techniques", December 2003
    b) 1EC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held devices used in close
       proximity to the ear (frequency range of 300 MHz to 3 GHz)", February 2005

Methods Applied and Interpretation of Parameters:
    *    NORMx,y,z: Assessed for E—field polarization 8 = 0 (f s 900 MHz in TEM—cell; f> 1800 MHz: R22 waveguide).
         NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the E*—field
         uncertainty inside TSL (see below ConvF).
    *    NORM(Ax,y,2 = NORMx,y,z * frequency_response (see Frequency Response Chart). This linearization is
         implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
         in the stated uncertainty of ConvF.
    «_   DCPx,y,z; DCP are numerical linearization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
    «0   Asyz Bxuy,z; Coy,2, VRxy,2z: A, B, C are numerical linearization parameters assessed based on the data of
         power sweep for specific modulation signal. The parameters do not depend on frequency nor media. VR is the
         maximum calibration range expressed in RMS voltage across the diode.
    *    ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature Transfer
         Standard for f < 800 MHz) and inside waveguide using analytical field distributions based on power
         measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
         boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
         used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
         to NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
         ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to * 100
         MHz.
    «_   Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
         exposed by a patch antenna.
    «_   Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
         (on probe axis). No tolerance required.




Certificate No: ET3—1559_Jan11                                  Page 2 of 11


                                            .000 _                                 Crscow
ET3DV6 SN:1559                                                                     January 19, 2011




                              Probe ET3DVC6

                                             SN:1559
                              Manufactured:                     December 1, 2000
                              Last calibrated:                  January 20, 2010
                              Recalibrated:                     January 19, 2011

                                   Calibrated for DASY/EASY Systems
                                       (Note: non—compatible with DASY2 system!)




  Certificate No: ET3—1559_Jan11                 Page 3 of 11


                                                                                                                                             CETECOM


 ET3DV6 SN:1559                                                                                                                              January 19, 2011


  DASY/EASY — Parameters of Probe: ET3DV6 SN:1559

 Basic Calibration Parameters
                                                                                              Sensor X Sensor Y Sensor Z |Unc (k=2)
 Norm (uV/(V/im)")"                                                                              1.79           1.59           1.64          _|+10.1%
 DCP ({mV)"                                                                                      96.9           97.6           96.3


 Modulation Calibration Parameters
 UID             Communication System Name                               PAR                       A              B              C             VR          Unc®
                                                                                                  dB           dBuV                            mV         (k=2)
 10000           Cw                                                      0.00|       x              0.00          0.00               1.00|    1321       £+29%
                                                                                     ¥              0.00          0.00               1.00|    137.8
                                                                                     Z              0.00          0.00               1.00|    128.3




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement multiplied
 by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of
 approximately 95%.


A The uncertainties of NormX,Y,2 do not affect the E—feld uncertainty inside TSL (see Pages 5 and 6).
® Numerical linearization parameter. uncertainty not required.
& Uncertainty is determined using the maximum deviation from linear response applying recatangular distribution and is expressed for the square of the field value.




     Certificate No: ET3—1559_Jan11                                     Page 4 of 11


                                                                                                                                        CETECOM


 ET3DV6 SN:1559                                                                                                                         January 19, 2011




 DASY/EASY — Parameters of Probe: ET3DV6 SN:1559

Calibration Parameter Determined in Head Tissue Simulating Media


f [MHz]         Validity [MHz]c        Permittivity        Conductivity       ConvF X        ConvF Y       ConvF 2             Alpha        Depth Unc (k=2)

450             +50 /+ 100             43.5 £ 5%           0.87 £5%                   7.39          7.39          7.39           0.18         2.07 £13.3%
835             * 50 / £ 100           41.5 £5%            0.90 £5%                   6.33          6.33          6.33           0.25         3.00 #11.0%
900             * 50 / # 100           41.5 £5%            0.97 £5%                   6.20          6.20          6.20           0.26         3.00 #11.0%
1750            £50 / 100              40.1 £5%             1.37 45%                  5.28          5.28          5.28           0.79         1.69 £11.0%
1900            £ 50/ £ 100            40.0 £5%             140 + 5%                  5.02          5.02          5.02           0.79         1.60 £11.0%
2450            +50 /# 100             39.2 4 5%            1.80 £ 5%                4.38           4.38          4.38           0.79         2.02 £11.0%




© The validity of + 100 MHz only applies for DASY vd.d and higher (see Page 2). The uncertainty is the RSS of the ConvF uncertainty at calibration frequency

and the uncertainty for the indicated frequency band.




      Certificate No: ET3—1559_Jani11                                  Page 5 of 11


                                                                                                                                        CETECOM


 ET3DVE SN:1559                                                                                                                         January 19, 2011




 DASY/EASY — Parameters of Probe: ET3DV6 SN:1559

Calibration Parameter Determined in Body Tissue Simulating Media


f [MHz]         Validity [MHz]c        Permittivity         Conductivity      ConvFX         ConvFY        ConvFZ              Alpha        Depth Unc (k=2)

450             + 50 /# 100            56.7 #5%             0.94 £ 5%                 7.73          7.73          7.73           0.12         2.07 £13.3%

835             + 50/ £ 100            55.2 £ 5%            0.97 #5%                  6.22          6.22          6.22           0.25         2.98 #11.0%

900             + 50/ £ 100            55.0 £ 5%            1.05 £5%                  6.10          6.10          6.10           0.29         2.87 #11.0%

1750            * 50 /+ 100            53.4 £5%             149 £5%                   4.68          4.68          4.68           0.79         2.39 £11.0%

1900            £50 /+ 100             53.3 £5%             1.52 + 5%                 4.40          4.40          4.40           0.79         2.32 £11.0%

2450            £ 50 / £ 100           52.7 + 5%            1.95 x 5%                 3.91          3.91          3.91           0.70         3.00 £11.0%




° The validity of £ 100 MHz only applies for DASY v4.4 and higher (see Page 2). The uncertainty is the RSS of the ConvF uncertainty at calibration frequency

and the uncertainty for the indicated frequency band.




      Certificate No: ET3—1559_Jan11                                   Page 6 of 11


                                                                                                                                CETECOM


ET3DV6 SN:1559                                                                                                                  January 19, 2011



                                                                Frequency Response of E—Field
                                                                  (TEM—Cell:if110 EXX, Waveguide: R22)
                          Frequency response (normalized)




                                                            0        500       1000        1500     2000     2500        2000
                                                                                        f [MHz]

                                                                           —o—TeM                 —o—nez      E



                                                            Uncertainty of Frequency Response of E—field: & 6.3% (k=2)




  Certificate No: ET3—1559_Jan11                                               Page 7 of 11


                                                                                              cerscom"
ET3DVG SN:1559                                                                                January 19, 2011



                                   Receiving Pattern (¢), 8 = 0°

                  £= 600 MHz, TEM ift10EXX                             800 MHz, We R22




                                                              cow e <op coom|




                                                                                        ~+—some      |
                                                                                        —a—rome
                                                                                        —+—scome
                                                                                        cocnssoue |
                                                                                        —am—zmoime




                                 Uncertainty of AxialIsotropy Assessment: £ 0.5% (=2)




  Certcate No: ET3—1580.Jant 1                   PageSof 11


                                                               |                                               cerscom"
ET3DVG SN:1550                                                                                                 January 19, 2011



                                                      Dynamic Range f(SARneaq)
                                                                (TEM cell,£= 900 MHz)
                                        1105


                                        mer0s
                   Sensor Voltage [M)




                                              0001      001         01              1       10           100
                                                                 sat mwents
                                              neok    ome kear <8—¥ <B—¥ear <#—2. <Bi—Zeor

                                        200

                                        100
                           s
                   Error 1B)




                                          coo          oo          o1           1          10            100
                                                                    SAR ImWiom‘1

                                                     Uncertainty of Linearity Assessments + 0.6% (k=2)




  Certfcale No: ET3—1889_Jantt                                     PageBof 11


                                                                                                        cerscom
ET3DV6 SN:1559                                                                                           January 19, 2011



                                   Conversion Factor Assessment

                (= 00 Miz, WGLS RS (head)                                 1=1750 Mtz, WGLS R22 (head)
                                                                soo
                                                                250
                                                              2200
                                                             <E 150
                                                             #E10
                                                             &a   so

                                                                  oo
        o      ro     zn #tmm
                          uen o              5o     co                                   ghing
             ~o—muen             —o—teumce |                                ~ow                    +nasnren


                                    Deviation from Isotropy in HSL
                                                  Error(§, 9}, £ = 900 MHz
                                                                       es eeseo co.

                                                                                      Error (dB)
                                                                       Abbe




                                 {m—100—380 m a50—080m.a60—010 .0 10—0.20 m.020000
                                 luswow momos momose mowor mos1s


                                  Uncertainty of Spherical Isotropy Assessmonts + 2.6% (ke2)




  Ceriicate No: ET3—1550.Jant1                       Page 10 of 11


                                                            CETECOM


ET3DV6 SN:1559                                              January 19, 2011




Other Probe Parameters

Sensor Arrangement                                                  Triangular
Connector Angle (°)                                             Not applicable
Mechanical Surface Detection Mode                                     enabled
Optical Surface Detection Mode                                        enabled
Probe Overall Length                                                 337 mm
Probe Body Diameter                                                    10 mm
Tip Length                                                             10 mm
Tip Diameter                                                          6.8 mm
Probe Tip to Sensor X Calibration Point                               2.7 mm
Probe Tip to Sensor Y Calibration Point                               2.7 mm
Probe Tip to Sensor Z Calibration Point                               2.7 mm
Recommended Measurement Distance from Surface                         3.7 mm




   Certificate No: ET3—1559_Jan11           Page 11 of 11


                           Test report no.: 1-3741-01-05/11




3    Calibration report “1900 MHz System validation dipole”




2011-09-09                                                    Page 14 of 29


             Test report no.: 1-3741-01-05/11




2011-09-09                                      Page 15 of 29


62 10 9. abeq                                                                                                           60—60— 108




                                                          sio n etee                     1 ifiny snop6—2n00610 on esd



                 (2oals sls 6/ mu oiz           W1 or parpeuuon                   siejourered 151 Apog murtou 701 HyS
                        6/ msrc                amed indu mw ase                                        painsnou gyg
                                                    verpuse            4151 Apost i0 (B 01) uo o1 Jono padeione gys

                 td n wl + 5 rmu sar            M o) parteunou                    s isweie 151 Apea ruwou .0 ies
                        6/ ul you              mmod tndu: Jb 082                                       panseou gye
                                                    veupuoy                751 Apeg 10 (6 1) ..wo 1 10no poberone uys

                                                                                           751 4pog yim 1nsoi S
                                                   a.so>                    1so1 Bupnp obveyp aimereduiey 151. Spon
            q + wouw 2591           %a>@es       o.(ac—bea                            sisroumied 151 Apog posnsety
                wyouu ge,            #ee            9.0%                               sisiewrered 151 Apog feumoy
                Runonpuog        Aamuieg         aimorocuray                           eene
                                                                  seigce aiam suopepoms por seivied buojor 94L
                                                                                            sierewened 151 Apog

                 CSw en = ty murcros            11. 5 peatsuion                  smatueied 51 peok; ouluiou 10) By$
                        Bm is                  Towod indur wyu ns                                    ramssow uve
                                                    voupues             T91 proHiJ0 (6 Di) we o1 Jore pebciore uy$

                 td s ol # E/ m oov             mersoue                        spreurmied 15. peop jeuwou 10) UY$
                     8 m vo.                   rmod indur w oge                                        pememuuve
                                                   vouno5                   T1 peok 10 (6 1) ;we 1 jon0 poBerone uys

                                                                                           791 peek injm 3inser uy$
                   «w                              0. 90>                   jo0; Burinp aBuuyo aimmradwe;791. pooy
            weruoyire               maren       a. earoc                             siorowered Ts1 peok peinsnay
              woul oy 1                        _ pve                                  srzowned 131. peok routmon;
             Ananonpuo5                         cimeisduel,
                                                                  varddo      won suuenalss pus simrelieied fumaoy SlL,
                                                                                                sigpowmied 731 peok


                                                 ahiw 1+ zeoost                                           fousnbasg
                                                  w e~ zn p m                                 uohniosey veaswoo7
                     imoeds um                         wo                               T61 —z9ws9 sodia souaimg
           >                                    wieleae terg ammson                                        wopug
                                               voreodempesumspy                                        vonrioderng
                        zveen                         #sva                      s                     uosion Asva
                                                               1 aBed to uanb ow se ese vomnnliuen menin ASve
                                                                                       suoppuo3 juswsinseayy




  Wm,                                      LSo—Lo—Lp2g—1 ou uodei ise1.


Appendix
Antenna Parameters with Head TSL

      Impedance, transformed to feed poin:                                        sn1ceer in
      Relurn Loss                                                                   —s0 0 8

Antenna Parameters with Body TSL

      Impedarce, transformed to feed point                                        s48 t +280
      Hetum Less                                                                    —seaou



General Antenna Parameters and Design

      Electnoal Delay (one direction}                                                 1188 ns

Alter ong termuse with100Wradiates nower. only a slight waring of Ihe cipo‘e near the feedpoint can be measurad
Te dioole is made of standard semigid coaxial cable. The center conductor of the feecing lirc s direely eennected to the
secand srm ofthe cigole. The antenns is tharelore shor—ireulted fer DC—signals.
No exssaiva torce must ba apl ed o the dipola arms, oocause thay might band ortha solserad carinections neer the
feedpoint may o damagec.

Additional EUT Data
    manutacturee oy                                                                 sreas
     Manulacturee on                                      |                     Febuary 22. 2002




       e No: Tiianava—sacns_Aug1 1                     Pages o‘ a


             Test report no.: 1-3741-01-05/11




2011-09-09                                      Page 18 of 29


             Test report no.: 1-3741-01-05/11




2011-09-09                                      Page 19 of 29


                                                                                   CETECOM*

DASY5 Validation Report for Body TSL
                                                                                   Dae: 17.0%.2011
Test Lahoratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: DLO0OV2; Serial: 119002 — S : 50009
Communication System: CW; Frequeney: 1900 MH7
Medium parameters ased: f= 1900 MHz: a = 1,57 riho/m; 6 = $3.9; p = 1000 kwin‘
Phancom section: Flat Section
Measurement Standard: DASYS (IEEETEC/ANSI C63,19—2007)

DASY52 Configuration
    +   Probe: ES3DV3 — SN3205; ConvR(4.62, 4.62, 4.62); Calibrated: 29.04.201 1

    +   Sensor—Surface: 3mm (Mechanical Surface Detection)
    +   Electronies: DAES $n601: Calibrated: 04.07.201 1

    *   Phantom: Mlat Phantom 5.0 (back); Type: QDOOOPSOAA; Serial: 1002

    +   DASY32 52.6.2(482); SEMCAD X 14.4.5(3634)


Dipole Calibration for Bady Tissue/Pin=250 mW d=10mm/Zoom Scan (7x7x7)/(Cube 0;
Mensurement grid: dx=3mm, dy=Smm, de=
Reference Value = 95.260 V/m; Power Drift = 0.02 4B
Peak SAR (extrapolated) = 18.       Whke
SARCH g)= 104 mWig: SAR(1O g) 5.43 mWig
Maximum value of SAR (measured) = 13.11 1 mWre




                         Hoas

                         sane

                          ie
                                  0 dB = 13.1 10mW/g




Ceriffcate Ne: Dro002sd009.Aug!                  "ago t af n


             Test report no.: 1-3741-01-05/11




2011-09-09                                      Page 21 of 29


                            Test report no.: 1-3741-01-05/11




4    Calibration certificate of Data Acquisition Unit (DAE)




2011-09-09                                                     Page 22 of 29


                           Test report no.: 1-3741-01-05/11




5    Certificate of “SAM Twin Phantom V4.0/V4.0C’’




2011-09-09                                                    Page 23 of 29


                                     Test report no.: 1-3741-01-05/11




6     Application Note System Performance Check

6.1   Purpose of system performance check
The system performance check verifies that the system operates within its specifications. System and
operator errors can be detected and corrected. It is recommended that the system performance check is
performed prior to any usage of the system in order to guarantee reproducible results.
The measurement of the Specific Absorption Rate (SAR) is a complicated task and the result depends on
the proper functioning of many components and the correct settings of many parameters. Faulty results due
to drift, failures or incorrect parameters might not be recognized, since they often look similar in distribution
to the correct ones. The Dosimetric Assessment System DASY4 incorporates a system performance check
procedure to test the proper functioning of the system. The system performance check uses normal SAR
measurements in a simplified setup (the flat section of the SAM Twin Phantom) with a well characterized
source (a matched dipole at a specified distance). This setup was selected to give a high sensitivity to all
parameters that might fail or vary over time (e.g., probe, liquid parameters, and software settings) and a low
sensitivity to external effects inherent in the system (e.g., positioning uncertainty of the device holder). The
system performance check does not replace the calibration of the components. The accuracy of the system
performance check is not sufficient for calibration purposes. It is possible to calculate the field quite
accurately in this simple setup; however, due to the open field situation some factors (e.g., laboratory
reflections) cannot be accounted for. Calibrations in the flat phantom are possible with transfer calibration
methods, using either temperature probes or calibrated E-field probes. The system performance check also
does not test the system performance for arbitrary field situations encountered during real measurements of
mobile phones. These checks are performed at SPEAG by testing the components under various conditions
(e.g., spherical isotropy measurements in liquid, linearity measurements, temperature variations, etc.), the
results of which are used for an error estimation of the system. The system performance check will indicate
situations where the system uncertainty is exceeded due to drift or failure.

6.2   System Performance check procedure
Preparation
The conductivity should be measured before the validation and the measured liquid parameters must be
entered in the software. If the measured values differ from targeted values in the dipole document, the liquid
composition should be adjusted. If the validation is performed with slightly different (measured) liquid
parameters, the expected SAR will also be different. See the application note about SAR sensitivities for an
estimate of possible SAR deviations. Note that the liquid parameters are temperature dependent with
approximately – 0.5% decrease in permittivity and + 1% increase in conductivity for a temperature decrease
of 1° C. The dipole must be placed beneath the flat phantom section of the Generic Twin Phantom with the
correct distance holder in place. The distance holder should touch the phantom surface with a light pressure
at the reference marking (little hole) and be oriented parallel to the long side of the phantom. Accurate
positioning is not necessary, since the system will search for the peak SAR location, except that the dipole
arms should be parallel to the surface. The device holder for mobile phones can be left in place but should
be rotated away from the dipole. The forward power into the dipole at the dipole SMA connector should be
determined as accurately as possible. See section 4 for a description of the recommended setup to measure
the dipole input power. The actual dipole input power level can be between 20mW and several watts. The
result can later be normalized to any power level. It is strongly recommended to note the actually used power
level in the „comment“-window of the measurement file; otherwise you loose this crucial information for later
reference.




2011-09-09                                                                                        Page 24 of 29


                                       Test report no.: 1-3741-01-05/11


System Performance Check
The DASY4 installation includes predefined files with recommended procedures for measurements and
validation. They are read-only document files and destined as fully defined but unmeasured masks, so you
must save the finished validation under a different name. The validation document requires the Generic Twin
Phantom, so this phantom must be properly installed in your system. (You can create your own
measurement procedures by opening a new document or editing an existing document file). Before you start
the validation, you just have to tell the system with which components (probe, medium, and device) you are
performing the validation; the system will take care of all parameters. After the validation, which will take
about 20 minutes, the results of each task are displayed in the document window. Selecting all measured
tasks and opening the predefined “validation” graphic format displays all necessary information for validation.
A description of the different measurement tasks in the predefined document is given below, together with
the information that can be deduced from their results:
•     The „reference“ and „drift“ measurements are located at the beginning and end of the batch process.
      They measure the field drift at one single point in the liquid over the complete procedure. The indicated
      drift is mainly the variation of the amplifier output power. If it is too high (above ± 0.1dB) the validation
      should be repeated; some amplifiers have very high drift during warm-up. A stable amplifier gives drift
      results in the DASY4 system below ± 0.02 dB.
•     The „surface check“ measurement tests the optical surface detection system of the DASY4 system by
      repeatedly detecting the surface with the optical and mechanical surface detector and comparing the
      results. The output gives the detecting heights of both systems, the difference between the two systems
      and the standard deviation of the detection repeatability. Air bubbles or refraction in the liquid due to
      separation of the sugar-water mixture gives poor repeatability (above ± 0.1mm). In that case it is better
      to abort the validation and stir the liquid. The difference between the optical surface detection and the
      actual surface depends on the probe and is specified with each probe. (It does not depend on the
      surface reflectivity or the probe angle to the surface within ± 30°.) However, varying breaking indices of
      different liquid compositions might also influence the distance. If the indicated difference varies from the
      actual setting, the probe parameter „optical surface distance“ should be changed in the probe settings
      (see manual). For more information see the application note about SAR evaluation.
•     The „area scan“ measures the SAR above the dipole on a parallel plane to the surface. It is used to
      locate the approximate location of the peak SAR with 2D spline interpolation. The proposed scan uses
      large grid spacing for faster measurement; due to the symmetric field the peak detection is reliable. If a
      finer graphic is desired, the grid spacing can be reduced. Grid spacing and orientation have no influence
      on the SAR result.
•     The zoom scan job measures the field in a volume around the peak SAR value assessed in the previous
      „area“ scan (for more information see the application note on SAR evaluation).
If the validation measurements give reasonable results, the peak 1g and 10g spatial SAR values averaged
between the two cubes and normalized to 1W dipole input power give the reference data for comparisons.
The next section analyzes the expected uncertainties of these values. Section 6 describes some additional
checks for further information or troubleshooting.

6.3     Uncertainty Budget
Please note that in the following Tables, the tolerance of the following uncertainty components depends on
the actual equipment and setup at the user location and need to be either assessed or verified on-site by the
end user of the DASY4 system:
• RF ambient conditions
• Dipole Axis to Liquid Distance
• Input power and SAR drift measurement
• Liquid permittivity - measurement uncertainty
• Liquid conductivity - measurement uncertainty
Note: All errors are given in percent of SAR, so 0.1 dB corresponds to 2.3%. The field error would be half of
that. The liquid parameter assessment give the targeted values from the dipole document. All errors are
given in percent of SAR, so 0.1dB corresponds to 2.3%. The field error would be half of that.




2011-09-09                                                                                         Page 25 of 29


                                        Test report no.: 1-3741-01-05/11


System validation

In the table below, the system validation uncertainty with respect to the analytically assessed SAR
value of a dipole source as given in the P1528 standard is given. This uncertainty is smaller than
the expected uncertainty for mobile phone measurements due to the simplified setup and the
symmetric field distribution.

Error Sources                  Uncertainty   Probability    Divi- ci    ci     Standard      Standard      vi2
                               Value         Distribution   sor 1g      10g    Uncertainty   Uncertainty   or
                                                                               1g            10g           veff

Measurement System
Probe calibration              ± 4.8%        Normal         1    1      1      ± 4.8%        ± 4.8%        ∞
Axial isotropy                 ± 4.7%        Rectangular    √3   0.7    0.7    ± 1.9%        ± 1.9%        ∞
Hemispherical isotropy         ± 0.0%        Rectangular    √3   0.7    0.7    ± 0.0%        ± 3.9%        ∞
Boundary effects               ± 1.0%        Rectangular    √3   1      1      ± 0.6%        ± 0.6%        ∞
Probe linearity                ± 4.7%        Rectangular    √3   1      1      ± 2.7%        ± 2.7%        ∞
System detection limits        ± 1.0%        Rectangular    √3   1      1      ± 0.6%        ± 0.6%        ∞
Readout electronics            ± 1.0%        Normal         1    1      1      ± 1.0%        ± 1.0%        ∞
Response time                  ± 0.0%        Rectangular    √3   1      1      ± 0.0%        ± 0.0%        ∞
Integration time               ± 0.0%        Rectangular    √3   1      1      ± 0.0%        ± 0.0%        ∞
RF ambient conditions          ± 3.0%        Rectangular    √3   1      1      ± 1.7%        ± 1.7%        ∞
Probe positioner               ± 0.4%        Rectangular    √3   1      1      ± 0.2%        ± 0.2%        ∞
Probe positioning              ± 2.9%        Rectangular    √3   1      1      ± 1.7%        ± 1.7%        ∞
Max. SAR evaluation            ± 1.0%        Rectangular    √3   1      1      ± 0.6%        ± 0.6%        ∞
Test Sample Related
Dipole axis to liquid          ± 2.0%        Normal         1    1      1      ± 1.2%        ± 1.2%        ∞
distance
Power drift                    ± 4.7%        Rectangular    √3   1      1      ± 2.7%        ± 2.7%        ∞
Phantom and Set-up
Phantom uncertainty            ± 4.0%        Rectangular    √3   1      1      ± 2.3%        ± 2.3%        ∞
Liquid conductivity (target)   ± 5.0%        Rectangular    √3   0.64   0.43   ± 1.8%        ± 1.2%        ∞
Liquid conductivity (meas.)    ± 2.5%        Normal         1    0.64   0.43   ± 1.6%        ± 1.1%        ∞
Liquid permittivity (target)   ± 5.0%        Rectangular    √3   0.6    0.49   ± 1.7%        ± 1.4%        ∞
Liquid permittivity (meas.)    ± 2.5%        Normal         1    0.6    0.49   ± 1.5%        ± 1.2%        ∞
Combined Uncertainty                                                           ± 8.4%        ± 8.1%
Expanded Std.                                                                  ± 16.8%       ± 16.2%
Uncertainty




2011-09-09                                                                                      Page 26 of 29


                                        Test report no.: 1-3741-01-05/11


Performance check repeatability
The repeatability check of the validation is insensitive to external effects and gives an indication of the
variations in the DASY4 measurement system, provided that the same power reading setup is used for all
validations. The repeatability estimate is given in the following table:

Error Sources                  Uncertainty   Probability    Divi- ci     ci     Standard      Standard        vi2
                               Value         Distribution   sor 1g       10g    Uncertainty   Uncertainty     or
                                                                                1g            10g             veff

Measurement System
Probe calibration              ± 4.8%        Normal         1     1      1      0             0               ∞
Axial isotropy                 ± 4.7%        Rectangular    √3    0.7    0.7    0             0               ∞
Hemispherical isotropy         ± 0.0%        Rectangular    √3    0.7    0.7    0             0               ∞
Boundary effects               ± 1.0%        Rectangular    √3    1      1      0             0               ∞
Probe linearity                ± 4.7%        Rectangular    √3    1      1      0             0               ∞
System detection limits        ± 1.0%        Rectangular    √3    1      1      0             0               ∞
Readout electronics            ± 1.0%        Normal         1     1      1      0             0               ∞
Response time                  ± 0.0%        Rectangular    √3    1      1      0             0               ∞
Integration time               ± 0.0%        Rectangular    √3    1      1      0             0               ∞
RF ambient conditions          ± 3.0%        Rectangular    √3    1      1      0             0               ∞
Probe positioner               ± 0.4%        Rectangular    √3    1      1      0             0               ∞
Probe positioning              ± 2.9%        Rectangular    √3    1      1      0             0               ∞
Max. SAR evaluation            ± 1.0%        Rectangular    √3    1      1      0             0               ∞
Test Sample Related
Dipole axis to liquid          ± 2.0%        Normal         1     1      1      ± 1.2%        ± 1.2%          ∞
distance
Power drift                    ± 4.7%        Rectangular    √3    1      1      ± 2.7%        ± 2.7%          ∞
Phantom and Set-up
Phantom uncertainty            ± 4.0%        Rectangular    √3    1      1      ± 2.3%        ± 2.3%          ∞
Liquid conductivity (target)   ± 5.0%        Rectangular    √3    0.64   0.43   ± 1.8%        ± 1.2%          ∞
Liquid conductivity (meas.)    ± 2.5%        Normal         1     0.64   0.43   ± 1.6%        ± 1.1%          ∞
Liquid permittivity (target)   ± 5.0%        Rectangular    √3    0.6    0.49   ± 1.7%        ± 1.4%          ∞
Liquid permittivity (meas.)    ± 2.5%        Normal         1     0.6    0.49   ± 1.5%        ± 1.2%          ∞
Combined Uncertainty                                                            ± 5.3%        ± 4.9%
Expanded Std.                                                                   ± 10.6%       ± 9.7%
Uncertainty

The expected repeatability deviation is low. Excessive drift (e.g., drift in liquid parameters), partial system
failures or incorrect parameter settings (e.g., wrong probe or device settings) will lead to unexpectedly high
repeatability deviations. The repeatability gives an indication that the system operates within its initial
specifications. Excessive drift, system failure and operator errors are easily detected.




2011-09-09                                                                                        Page 27 of 29


                                      Test report no.: 1-3741-01-05/11



6.4     Power set-up for validation
The uncertainty of the dipole input power is a significant contribution to the absolute uncertainty and the
expected deviation in interlaboratory comparisons. The values in Section 2 for a typical and a sophisticated
setup are just average values. Refer to the manual of the power meter and the detector head for the
evaluation of the uncertainty in your system. The uncertainty also depends on the source matching and the
general setup. Below follows the description of a recommended setup and procedures to increase the
accuracy of the power reading:




The figure shows the recommended setup. The PM1 (incl. Att1) measures the forward power at the location
of the validation dipole connector. The signal generator is adjusted for the desired forward power at the
dipole connector and the power meter PM2 is read at that level. After connecting the cable to the dipole, the
signal generator is readjusted for the same reading at power meter PM2. If the signal generator does not
allow a setting in 0.01dB steps, the remaining difference at PM2 must be noted and considered in the
normalization of the validation results. The requirements for the components are:
•     The signal generator and amplifier should be stable (after warm-up). The forward power to the dipole
      should be above 10mW to avoid the influence of measurement noise. If the signal generator can deliver
      15dBm or more, an amplifier is not necessary. Some high power amplifiers should not be operated at a
      level far below their maximum output power level (e.g. a 100W power amplifier operated at 250mW
      output can be quite noisy). An attenuator between the signal generator and amplifier is recommended to
      protect the amplifier input.
•     The low pass filter after the amplifier reduces the effect of harmonics and noise from the amplifier. For
      most amplifiers in normal operation the filter is not necessary.
•     The attenuator after the amplifier improves the source matching and the accuracy of the power head.
      (See power meter manual.) It can also be used also to make the amplifier operate at its optimal output
      level for noise and stability. In a setup without directional coupler, this attenuator should be at least
      10dB.
•     The directional coupler (recommended ³ 20dB) is used to monitor the forward power and adjust the
      signal generator output for constant forward power. A medium quality coupler is sufficient because the
      loads (dipole and power head) are well matched. (If the setup is used for reflective loads, a high quality
      coupler with respect to directivity and output matching is necessary to avoid additional errors.)
•     The power meter PM2 should have a low drift and a resolution of 0.01dBm, but otherwise its accuracy
      has no impact on the power setting. Calibration is not required.
•     The cable between the coupler and dipole must be of high quality, without large attenuation and phase
      changes when it is moved. Otherwise, the power meter head PM1 should be brought to the location of
      the dipole for measuring.
•     The power meter PM1 and attenuator Att1 must be high quality components. They should be calibrated,
      preferably together. The attenuator (³10dB) improves the accuracy of the power reading. (Some higher
      power heads come with a built-in calibrated attenuator.) The exact attenuation of the attenuator at the
      frequency used must be known; many attenuators are up to 0.2dB off from the specified value.




2011-09-09                                                                                       Page 28 of 29


                                     Test report no.: 1-3741-01-05/11


•     Use the same power level for the power setup with power meter PM1 as for the actual measurement to
      avoid linearity and range switching errors in the power meter PM2. If the validation is performed at
      various power levels, do the power setting procedure at each level.
•     The dipole must be connected directly to the cable at location “X”. If the power meter has a different
      connector system, use high quality couplers. Preferably, use the couplers at the attenuator Att1 and
      calibrate the attenuator with the coupler.
•     Always remember: We are measuring power, so 1% is equivalent to 0.04dB.

6.5     Laboratory reflection
In near-field situations, the absorption is predominantly caused by induction effects from the magnetic near-
field. The absorption from reflected fields in the laboratory is negligible. On the other hand, the magnetic field
around the dipole depends on the currents and therefore on the feed point impedance. The feed point
impedance of the dipole is mainly determined from the proximity of the absorbing phantom, but reflections in
the laboratory can change the impedance slightly. A 1% increase in the real part of the feed point impedance
will produce approximately a 1% decrease in the SAR for the same forward power. The possible influence of
laboratory reflections should be investigated during installation. The validation setup is suitable for this
check, since the validation is sensitive to laboratory reflections. The same tests can be performed with a
mobile phone, but most phones are less sensitive to reflections due to the shorter distance to the phantom.
The fastest way to check for reflection effects is to position the probe in the phantom above the feed point
and start a continuous field measurement in the DASY4 multi-meter window. Placing absorbers in front of
possible reflectors (e.g. on the ground near the dipole or in front of a metallic robot socket) will reveal their
influence immediately. A 10dB absorber (e.g. ferrite tiles or flat absorber mats) is probably sufficient, as the
influence of the reflections is small anyway. If you place the absorber too near the dipole, the absorber itself
will interact with the reactive near-field. Instead of measuring the SAR, it is also possible to monitor the
dipole impedance with a network analyzer for reflection effects. The network analyzer must be calibrated at
the SMA connector and the electrical delay (two times the forward delay in the dipole document) must be set
in the NWA for comparisons with the reflection data in the dipole document. If the absorber has a significant
influence on the results, the absorber should be left in place for validation or measurements. The reference
data in the dipole document are produced in a low reflection environment.

6.6     Additional system checks
While the validation gives a good check of the DASY4 system components, it does not include all
parameters necessary for real phone measurements (e.g. device modulation or device positioning). For
system validation (repeatability) or comparisons between laboratories a reference device can be useful. This
can be any mobile phone with a stable output power (preferably a device whose output power can be set
through the keyboard). For comparisons, the same device should be sent around, since the SAR variations
between samples can be large. Several measurement possibilities in the DASY software allow additional
tests of the performance of the DASY system and components. These tests can be useful to localize
component failures:
•     The validation can be performed at different power levels to check the noise level or the correct
      compensation of the diode compression in the probe.
•     If a pulsed signal with high peak power levels is fed to the dipole, the performance of the diode
      compression compensation can be tested. The correct crest factor parameter in the DASY software must
      be set (see manual). The system should give the same SAR output for the same averaged input power.
•     The probe isotropy can be checked with a 1D-probe rotation scan above the feed point. The automatic
      probe alignment procedure must be passed through for accurate probe rotation movements (optional
      DASY4 feature with a robot-mounted light beam unit). Otherwise the probe tip might move on a small
      circle during rotation, producing some additional isotropy errors in gradient fields.




2011-09-09                                                                                        Page 29 of 29



Document Created: 2011-11-14 12:54:59
Document Modified: 2011-11-14 12:54:59

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC