SAR Dipole Performance Measurement Report

FCC ID: TE7C7LITEV1

RF Exposure Info

Download: PDF
FCCID_4209786

 <Justification of the extended calibration>
 Referring to KDB 865664 D01, if dipoles are verified in return loss<-20dB, (within 20% of
 prior calibration),and in impedance (within 5 ohm of prior calibration), the annual calibration
 is not necessary and the calibration interval can be extended.

                                         Head 750 MHz

      Date of
                     Return Loss (dB)       Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -25.42                -                 55.1                 -

    2018.08.15            -23.45              -7.75              55.84                0.74

 The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
 prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Head 750 MHz


                                         Body 750 MHz

      Date of
                     Return Loss (dB)      Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -27.21               -                51.3                 -

    2018.08.15            -26.44             -2.82              51.50               0.2


The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Body 750 MHz


                                        Head 835 MHz

      Date of
                     Return Loss (dB)      Delta (%)         Impedance          Delta(ohm)
   Measurement

    2017.08.15            -28.11               -                51.6                 -

    2018.08.15            -27.05             -3.77              54.69              3.09

 The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
 prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Head 835MHz


                                         Body 835 MHz

      Date of
                     Return Loss (dB)      Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -23.87               -                49.0                 -

    2018.08.15            -22.21             -6.95              52.59               3.59


The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Body 835MHz


                                        Head 900 MHz

      Date of
                     Return Loss (dB)      Delta (%)         Impedance          Delta(ohm)
   Measurement

    2017.08.15            -27.32               -                53.9                 -

    2018.08.15            -26.28              3.8               52.5                -1.4

 The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
 prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Head 900MHz


                                         Body 900 MHz

      Date of
                     Return Loss (dB)      Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -23.53               -                56.4                 -

    2018.08.15            -22.49              4.4               55.9                -0.5


The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Body 900MHz


                                        Head 1800 MHz

      Date of
                     Return Loss (dB)      Delta (%)         Impedance          Delta(ohm)
   Measurement

    2017.08.15            -29.51               -                46.7                 -

    2018.08.16            -25.32             -14.2              45.24              -1.46

 The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
 prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Head 1800 MHz


                                        Body 1800 MHz

      Date of
                     Return Loss (dB)      Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -24.59               -                48.8                 -

    2018.08.16            -24.95             1.46               45.42              -3.38
The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Body 1800 MHz


                                        Head 1900 MHz

      Date of
                     Return Loss (dB)      Delta (%)         Impedance          Delta(ohm)
   Measurement

    2017.08.15            -23.68               -                51.2                 -

    2018.08.16            -25.00             5.57               47.12              -4.08

 The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
 prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Head 1900 MHz


                                        Body 1900 MHz

      Date of
                     Return Loss (dB)      Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -20.22               -                48.8                 -

    2018.08.16            -23.26             15.03              47.11              -1.79
The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Body 1900 MHz


                                        Head 2450 MHz

      Date of
                     Return Loss (dB)      Delta (%)         Impedance          Delta(ohm)
   Measurement

    2017.08.15            -26.00               -                46.1                 -

    2018.08.16            -29.56            13.69               47.53              1.43

 The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
 prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Head 2450 MHz


                                        Body 2450 MHz

      Date of
                     Return Loss (dB)      Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -32.75               -                48.8                 -

    2018.08.16            -31.06             -5.16              45.61              -3.19
The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Body 2450 MHz


                                        Head 2600 MHz

      Date of
                     Return Loss (dB)      Delta (%)         Impedance          Delta(ohm)
   Measurement

    2017.08.15            -34.35               -                50.2                 -

    2018.08.16            -31.25             -9.02              53.76              3.56

 The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
 prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Head 2600 MHz


                                        Body 2600 MHz

      Date of
                     Return Loss (dB)      Delta (%)          Impedance          Delta(ohm)
   Measurement

    2017.08.15            -24.18               -                45.7                 -

    2018.08.16            -23.36             -7.52              46.79               1.09
The return loss is <-20dB, within 20% of prior calibration; the impedance is within 5 ohm of
prior calibration. Therefore the verification result should support extended calibration.

<Dipole Verification Data>

Body 2600 MHz


                                         Body 5000 MHz

      Date of
                     Return Loss (dB)        Delta (%)          Impedance           Delta(ohm)
   Measurement

    2017.08.15            < 13.94                -                  -                   -

    2018.08.16            < 13.52                -                  -                   -
The return loss is <-8dB, within 20% of prior calibration; Therefore the verification result
should support extended calibration.

<Dipole Verification Data>

Body 5000MHz



Document Created: 2019-03-19 17:43:40
Document Modified: 2019-03-19 17:43:40

© 2024 FCC.report