Cal cert Mixer FSZ60

FCC ID: TA8AKRD901079

Test Report

Download: PDF
FCCID_4158050

                                             Farran Technology Ltd.,
                                             Unit 1 Airport East Technology Park.
                                             Farmers Cross, Co. Cork.
                                             Ireland.

                                Certificate of Calibration
                           Certificate No: FCC 10300

Model#: FS-Z60              S/N: 100147             Cal Freq: 40-60GHz
Calibration Date : 30/11/2017
As shipped Condition: On completion of testing, measurement points within
specification.
Comments: Unit calibrated to spec, results on file.
Measurement Uncertainty: +/-1.5dB.
Customer : Rohde & Schwarz.
Location of Calibration: Farran Technology
Certified By: Michael Crowley     Date 30/11/2017


Signature


Date 30/11/2017

This calibration certificate documents, that the named item is tested and measured
against defined specifications.
Measurement results are located usually in the corresponding interval with a
probability of approx 95% (coverage factor k=2).
Calibration is performed with test equipment and standards directly or indirectly
traceable by means of approved calibration techniques to the PTB/DKD or other
national / international standards, which realize the physical units of measurement
according to the International System of Units (SI).
Principals and methods of calibration correspond and are conformant with EN
ISO/IEC 17025 and ANSI/NCSL Z540-1-1994. The applied quality system is certified
to EN ISO 9001.
This calibration certificate may not be reproduced other than in full. Calibration
certificates without signatures are not valid.
The user is obliged to have the object recalibrated at appropriate intervals.




                               Page 1 of 3                        FTL FORM/189/A


Calibration Procedure No: STP 27
Calibration Equipment used:
Description:        Model No:    S/N:         Certificate No.    Cal Due

Agilent Spectrum   8563E         3728A07479   1-7002610057-21    July 2019
Analyzer
Agilent Power      E4418A        GB39512672   1-7002610057-17    July 2019
Meter
Agilent Power      8481A         US37293022   1-7002610057-9     July 2019
Sensor
Agilent Power      V8486A        MY44420202   437543-1545636-1   Oct 2019
Sensor
Agilent Power      Q8486A        3318A01897   1-5446996787-1     Nov 2019
Sensor
Agilent            E8247B        MY43320714   1-7002610057-22    July 2019
Synthesized
Generator
Agilent            83712B        US37100716   1-7002610057-3     July 2019
Synthesized
Generator




                             Page 2 of 3                 FTL FORM/189/A


Conversion Loss Data for Harmonic Mixer FS-Z60 (40 ..
                      60 GHz)
                      (Order No. 8800000291/53RA/L6)
                                     Revision 1.30




Ser. No.                                                 100147


Conversion loss data valid for following operating conditions:

IF                                           741,4 MHz (valid for 404,4 MHz and 729,9 MHz too)
Order of
Harmonic                                                       4
LO frequency range                                   10 to 15GHz
                                                       +15.5dBm
LO level                                                  nominal
Bias                                                        0 mA


                     Conversion Loss /                              Conversion Loss /
Frequency / GHz            dB                Frequency / GHz              dB
     40.00                -16.36                     50.20               -14.92
     40.41                -15.72                     50.61               -15.55
     40.82                -15.10                     51.02               -15.50
     41.22                -15.61                     51.43               -14.95
     41.63                -15.73                     51.84               -15.28
     42.04                -15.68                     52.25               -15.97
     42.45                -14.99                     52.65               -15.72
     42.86                -14.82                     53.06               -15.38
     43.27                -15.66                     53.47               -15.68
     43.67                -15.77                     53.88               -16.10
     44.08                -15.17                     54.29               -15.96
     44.49                -14.81                     54.69               -15.58
     44.90                -15.33                     55.10               -15.39
     45.31                -15.50                     55.51               -15.37
     45.71                -15.19                     55.92               -15.03
     46.12                -14.81                     56.33               -14.92
     46.53                -15.31                     56.74               -15.21
     46.94                -15.87                     57.14               -15.46
     47.35                -15.84                     57.55               -15.37
     47.76                -15.75                     57.96               -15.74
     48.16                -15.53                     58.37               -16.43
     48.57                -15.78                     58.78               -16.79
     48.98                -16.26                     59.18               -16.41
     49.39                -15.79                     59.59               -16.07
     49.80                -15.34                     60.00               -15.38



                                   Page 3 of 3                               FTL FORM/189/A



Document Created: 2017-11-30 12:11:22