SAR Dipole Calibration II

FCC ID: T5MMC7355

RF Exposure Info

Download: PDF
FCCID_2103109

                                                                    w10
Calibration Laboratory of                                          S\ /z                                          Schweizerischer Kalibrierdienst




                                                              §
                                                               wO




                                                                            3
Schmid & Partner                                                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                                                  Servizio svizzero di taratura




                                                              wl


                                                                          Lgf
Zeughausstrasse 43, 8004 Zurich, Switzerland                       o T                                            Swiss Calibration Service
                                                                    Zulitabs®

Accredited by the Swiss Accreditation Service (SAS)                                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       CCS —TW (Auden)                                                                   Certificate No: D1800V2—2d062_Feb13

[CALIBRATION CERTIFICATE
 Object                               D1800V2 — SN: 204062


 Calibration procedure(s)             QA CAL—05.v9
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    February 12, 2013




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                     ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter EPM—442A                  GB37480704                O1—Nov—12 (No. 217—01640)                             Oct—13
 Power sensor HP 8481A                 US37292783                O1—Nov—12 (No. 217—01640)                             Oct—13
 Reference 20 dB Attenuator            SN: 5058 (20k)            27—Mar—12 (No. 217—01530)                             Apr—18
 Type—N mismatch combination           SN: 5047.3 / 06327        27—Mar—12 (No. 217—01533)                             Apr—13
 Reference Probe ES3DV3                SN: 8205                  28—Dec—12 (No. ES3—3205_Dec12)                        Dec—13
 DAE4                                  SN: 601                   27—Jun—12 (No. DAE4—601_Jun12)                        Jun—13


 Secondary Standards                   ID #                      Check Date (in house)                                 Scheduled Check
 Power sensor HP 8481A                 MY41092317                18—Oct—02 (in house check Oct—11)                     In house check: Oct—13
 RF generator R&S SMT—06               100005                    04—Aug—99 (in house check Oct—11)                     In house check: Oct—13
 Network Analyzer HP 8753E             US37390585 S4206          18—Oct—01 (in house check Oct—12)                     In house check: Oct—13



                                       Name                                     Function                               Signature
 Calibrated by:                        Israe El—Naouq                           Laboratory Technician
                                                                                                                   O/an ECf,\gaw.w\
 Approved by:                          KatJa Pokovic                            Technical Manager
                                                                                                                    es
                                                                                                                      Issued: February 12, 2013
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D1800V2—2d062_Feb13                                  Page 1 of 8


Calibration Laboratory of                                       ith
                                                            Si\/%                 Schweizerischer Kalibrierdienst
        :                                                  $s              z
Schmid & Partner                                           SIBE'/{&               Service suisse d‘étalonnage
  Engineering AG                                           zel                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland               7z ///'—\\\ w          Swiss Calibration Service
                                                                Tlilabs®

Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", December 2003
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
            February 2005
    c) Federal Communications Commission Office of Engineering & Technology (FCC OET),
       "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
       Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
       Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
       Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   e Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   e Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
            No uncertainty required.
    e       SAR measured: SAR measured at the stated antenna input power.
    e       SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
            connector.
    e       SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
            nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.



Certificate No: D1800V2—2d062_Feb13                              Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                               V52.8.5
      Extrapolation                                   Advanced Extrapolation
      Phantom                                         Modular Flat Phantom
      Distance Dipole Center — TSL                              10 mm                             with Spacer

     Zoom Scan Resolution                               dx, dy, dz =5 mm
      Frequency                                         1800 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                   22.0 °C           40.0               1.40 mho/m

     Measured Head TSL parameters                               (22.0 + 0.2) °C   39.0 +6 %          1.38 mho/m + 6 %
     Head TSL temperature change during test                       <0.5 °C            ~———                    ———


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
     SAR measured                                         250 mW input power                     9.64 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W          38.7 Wikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                  condition
     SAR measured                                         250 mW input power                  5.06 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W          20.3 Wikg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity

     Nominal Body TSL parameters                                   22.0 °C            53.3              1.52 mho/m
     Measured Body TSL parameters                               (22.0 + 0.2) °C   51.5 26 %           1.52 mho/m + 6 %
     Body TSL temperature change during test                       «08 °C                                     ———


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                    Condition
     SAR measured                                         250 mW input power                     9.69 W/kg

     SAR for nominal Body TSL parameters                     normalized to 1W          38.5 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                  condition

     SAR measured                                         250 mW input power                     5.10 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W          20.3 Wikg * 16.5 % (k=2)



Certificate No: D1800V2—2d062_Feb13                    Page 3 of 8


Appendix

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           49.3 Q — 2.0 jQ
      Return Loss                                                                       — 33.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           45.3 Q — 2.4 jQ
      Return Loss                                                                      —25.2 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.209 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                               January 28, 2003




Certificate No: D1800V2—2d062_Feb13                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 12.02.2013

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1800 MHz; Type: D1800V2; Serial: D1800V2 — SN: 24062

Communication System: CW; Frequency: 1800 MHz
Medium parameters used: f = 1800 MHz; 0 = 1.38 S/m; & = 39; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C€63.19—2007)

DASY52 Configuration:

        Probe: ES3DV3 — SN3205; ConvF(5.04, 5.04, 5.04); Calibrated: 28.12.2012;

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 Sn601; Calibrated: 27.06.2012

        Phantom: Flat Phantom 5.0 (front); Type: QDOOOPS5OAA; Serial: 1001

        DASY52 52.8.5(1059); SEMCAD X 14.6.8(7028)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.713 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 17.6 W/kg
SAR(1 g) = 9.64 W/kg; SAR(10 g) = 5.06 W/kg
Maximum value of SAR (measured) = 11.9 W/kg




                       —3.60




                       —1.20




                       —10.80




                       14.40




                       18.00

                                0 dB = 11.9 W/kg = 10.76 dBW/kg




Certificate No: D1800V2—2d062_Feb13               Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                 12 Feb 2013 10:50:41
          CHI) sii     1 U FS          1:49.314 a   —1.9648 n 45.001 pF      1 800.000 000 MHz




          Del

          Cor




          fAve
          16


          H1d

          CH2




          Cor




          Av
          16 *


          H1d

                     START 1 600.000 000 MHz                           STOP 2 000.000 000 MHz




Certificate No: D1800V2—2d062_Feb13                   Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                   Date: 12.02.2013

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1800 MHz; Type: D1800V2; Serial: D1800V2 — SN: 20062

Communication System: CW; Frequency: 1800 MHz
Medium parameters used: f = 1800 MHz; 0 = 1.52 S/m; & = 51.5; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)

DASY52 Configuration:

    e   Probe: ES3DV3 — SN3205; ConvF(4.73, 4.73, 4.73); Calibrated: 28.12.2012;

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 27.06.2012

        Phantom: Flat Phantom 5.0 (back); Type: QDOOOPS50OAA; Serial: 1002

        DASY52 52.8.5(1059); SEMCAD X 14.6.8(7028)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 95.713 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 17.4 W/kg
SAR(1 g) = 9.69 W/kg; SAR(10 g) = 5.1 W/kg
Maximum value of SAR (measured) = 12.2 W/kg




                         —3.60




                         —7.20




                         —10.80




                         —14.40




                         —18.00

                                  0 dB = 12.2 W/kg = 10.86 dBW/kg




Certificate No: D1800V2—2d062_Feb13                Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                     12 Feb 2013    10:50:17
          CHI sii     1 U e            1:145.344 a   —23613 0   37.445 pF          1 800.000 000 MHz
                                                     zy




          Del

          Cor




          Av
          169



          H1d

          CH2




          Cor




          Av
          16°

          H1d

                    START 1 600.000 000 MHz                                 STOP 2 000.000 000 MHz




Certificate No: D1800V2—2d062_Feb13                    Page 8 of 8


Calibration
      .     Laboratory of                                          all/»
                                                                  dst"/‘s                                   S      Schweizerischer Kalibrierdienst
Schmid & Partner                                                 ifi:\—/E—;':                   @3           fa     Service suisse d‘étalonnage
  Engineering AG                                                 Hgppan l                   (-; T                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                     ‘«/,l/7/‘\\\\\\\‘}          (/QRP‘S\       S      swiss Calibration Service
                                                                     ritabe

Accredited by the Swiss Accreditation Service (SAS)                                              Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       CCS—TW (Auden)                                                                      Certificate No: D1900V2—5d056_Feb13

ICALIBRATION CERTIFICATE
 Object                                D1900V2 — SN: 50056


 Calibration procedure(s)              QA CAL—05.v9
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     February 13, 2013




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
 Power meter EPM—442A                   GB37480704                O1—Nov—12 (No. 217—01640)                            Oct—13
 Power sensor HP 8481A                  US37202783                O1—Nov—12 (No. 217—01640)                            Oct—13
 Reference 20 dB Attenuator             SN: 5058 (20k)            27—Mar—12 (No. 217—01530)                            Apr—13
 Type—N mismatch combination            SN: 5047.3 / 06827        27—Mar—12 (No. 217—01533)                            Apr—18
 Reference Probe ES3DV3                 SN: 3205                  28—Dec—12 (No. ES3—3205_Dec12)                        Dec—13
 DAE4                                   SN: 601                   27—Jun—12 (No. DAE4—601_Jun12)                       Jun—13


 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Power sensor HP 8481A                  MY41092317                18—0ct—02 (in house check Oct—11)                     In house check: Oct—13
 RF generator R&S SMT—06                100005                    04—Aug—99 (in house check Oct—11)                     In house check: Oct—13
 Network Analyzer HP 8753E              US37390585 S4206          18—Oct—01 (in house check Oct—12)                     In house check: Oct—13



                                        Name                                     Function                               Signature
 Calibrated by:                         Leif Klysner                             Laboratory Technician                   WZ



 Approved by:                           KatJa Pokovic                           Technical Manager                  /g? /; +



                                                                                                                       Issued: February 13, 2013
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D1900V2—5d056_Feb13                                   Page 1 of 8


Calibration Laboratory of                                                         Schweizerischer Kalibrierdienst
Schmid & Partner                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                      Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", December 2003
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
      February 2005
   c) Federal Communications Commission Office of Engineering & Technology (FCC OET},
         "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
         Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
         Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
         Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   e Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
         SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
        connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.



Certificate No: D1900V2—5d056_Feb13                             Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not g iven on page 1.
      DASY Version                                               DASYS                               V52.8.5
      Extrapolation                                   Advanced Extrapolation
      Phantom                                          Modular Flat Phantom
      Distance Dipole Center — TSL                               10 mm                             with Spacer

     Zoom Scan Resolution                                dx, dy, dz =5 mm
      Frequency                                          1900 MHz & 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                 Temperature       Permittivity         Conductivity
      Nominal Head TSL parameters                                   22.0 °C           40.0               1.40 mho/m
     Measured Head TSL parameters                                (22.0 + 0.2) °C   39.4 + 6 %         1.38 mho/m + 6 %

     Head TSL temperature change during test                        <0.5 °C            «———                    se+«


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                    Condition
     SAR measured                                          250 mW input power                     9.98 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W          40.1 W/kg + 17.0 % (k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                   condition
     SAR measured                                          250 mW input power                     5.24 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W           21.0 W/kg * 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                 Temperature       Permittivity          Conductivity

     Nominal Body TSL parameters                                    22.0 °C            53.5              1.52 mho/m
     Measured Body TSL parameters                                (22.0 + 0.2) °C   51.9 + 6 %          1.53 mho/m + 6 %

     Body TSL temperature change during test                        x6.5 °C


SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                     Condition
     SAR measured                                          250 mW input power                     10.2 W/kg

     SAR for nominal Body TSL parameters                     normalized to 1W           40.4 Wikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                   condition

     SAR measured                                          250 mW input power                  5.41 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W           21.5 W/kg + 16.5 % (k=2)



Certificate No: D1900V2—5d056_Feb13                     Page 3 of 8


Appendix

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           52.8 Q + 6.2 jQ

      Return Loss                                                                       —23.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.4 Q + 6.7 jQ
      Return Loss                                                                      —23.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.205 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend orthe soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                March 19, 2004




Certificate No: D1900V2—5d056_Feb13                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 13.02.2013

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50056

Communication System: CW; Frequency: 1900 MHz
Medium parameters used: £ = 1900 MHz; 0 = 1.38 S/m; & = 39.4; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2007)

DASY52 Configuration:

    e   Probe: ES3DV3 — SN3205; ConvF(4.98, 4.98, 4.98); Calibrated: 28.12.2012;

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 27.06.2012

        Phantom: Flat Phantom 5.0 (front); Type: QDOOOP50AA; Serial: 1001

        DASY52 52.8.5(1059); SEMCAD X 14.6.8(7028)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 96.002 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 18.1 W/kg
SAR(1 g) = 9.98 W/kg; SAR(10 g) = 5.24 W/kg
Maximum value of SAR (measured) = 12.4 W/kg




        —3.60

        —1.20

        —10.80

        —14.40

        —18.00
                  0 dB = 12.4 W/kg = 10.93 dBW/kg




Certificate No: D1900V2—5d056_Feb13            Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                   13 Feb 2013  ©9:34:49
          CHI sii     1 U Fs          1: 52820 a     6.2012 0   519.45 pH      1 900.000 000 MHz
                                                   meazrges:

          Del

          Ca




          Av
          153



          HId

          CH2




          Ca




          Av
          16 9



          H1d

                    START 1 700.000 000 MHz                              STOP 2 100.000 000 MHz




Certificate No: D1900V2—5d056_Feb13                    Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                Date: 13.02.2013

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 54056

Communication System: CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.53 S/m; & = 51.9; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)

DASY52 Configuration:

        Probe: ES3DV3 — SN3205; ConvF(4.6, 4.6, 4.6); Calibrated: 28.12.2012;

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 Sn601; Calibrated: 27.06.2012

        Phantom: Flat Phantom 5.0 (back); Type: QDOOOP50AA; Serial: 1002

        DASY52 52.8.5(1059); SEMCAD X 14.6.8(7028)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 96.002 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 17.8 W/kg
SAR(1 g) = 10.2 W/kg; SAR(10 g) = 5.41 W/kg
Maximum value of SAR (measured) = 12.9 W/kg




                            —10.20



                            —13.60



                            —17.00

                                     0 dB = 12.9 W/kg = 11.11 dBW/kg




Certificate No: D1900V2—5d056_Feb13                 Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                  13 Feb 2013  09:34:09
          CHI sii     1 U FS           1: 48.398 a    6.7422 0 564.76 pH      1 900.000 000 MHz
                                                     ——p——




          Del

          Ca




          Av
          189


          H1d

          CH2




          Ca




          Av
          16 2



          H1d

                    START 1 700.000 000 MHz                            STOP 2 100.000 000 MHz




Certificate No: D1900V2—5d056_Feb13                    Page 8 of 8



Document Created: 2013-10-08 16:05:19
Document Modified: 2013-10-08 16:05:19

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC