SAR Report Appendix 1

FCC ID: SS3-MR1SS51905

RF Exposure Info

Download: PDF
FCCID_4470142

PROBE CALIBRATION CERTIFICATES


                                     in Collaboration with

          x7"7"J,    a
          =—


          Wimazeee>~ CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                6—10—62304633—2512        Fax: +86—10—62304633—2504
          —mail: cttl@chinatt.com         Hitp:/www.chinattl.en


Glossary:
TSL                   tissue simulating liquid
NORMx,y,z             sensitivity in free space
ConvF                 sensitivity in TSL / NORMx,y,z
DCP                   diode compression point
CF                    crest factor (1/duty_cycle) of the RF signal
A,B,C,D               modulation dependent linearization parameters
Polarization ®        ® rotation around probe axis
Polarization 0        9 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                      8=0 is normal to probe axis
Connector Angle       information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
  Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
    Measurement Techniques®, June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)®,
    July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
e    NORMx,y,z: Assessed for E—field polarization 8=0 (fs900MHz in TEM—cell; f>1800MHz: waveguide).
     NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
     E" —field uncertainty inside TSL (see below ConvF).
e    NORM(Mx,y,2 = NORMx,y,2* frequency_response (see Frequency Response Chart). This
     linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
     frequency response is included in the stated uncertainty of ConvF.
e    DCPxX,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor media.
e    PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
     characteristics.
e    Axy.2; Bxyz; Coy2iVRxy,z:A,B,C are numerical linearization parameters assessed based on the
     data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum calibration range expressed in RMS voltage across the diode.
e    ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
     Transfer Standard for fs800MHz) and inside waveguide using analytical field distributions based on
     power measurements for f >800MHz. The same setups are used for assessment of the parameters
     applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
     These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
     The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
     that given for ConvF.A frequency dependent ConvF is used in DASY version 4.4 and higher which
     allows extending the validity from+50MHz to+100MHz.
e    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
     phantom exposed by a patch antenna.
e    Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
     probetip (on probe axis). No tolerance required.
e    Connector Angle: The angleis assessed using the information gained by determining the NORMx
     (no uncertainty required).

Certificate No: Z18—60353                              Page 2 of11


               ’T]\le in Collsboration with
                                      s     p      e      a       q
               e                      CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing. 100191, China
        Tel:   +86—10—62304633—2512          x: +86—10—62304633—2504
                 : ettl@chinatl.com        Hip://wwachinattl.en




                        Probe EX3DV4


                                                SN: 7329


                                      Calibrated: September 30, 2018

                               Calibrated for DASY/EASY Systems
                                          (Nete: non—compatible with DASY2 system!)




Certificate No: Z18—60353                              Page 3 of11




                                      in Collaboration with




           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
           E—mail: cttl@chinatt.com       Uittpy//www.chinattl.en


      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7329

Calibration Parameter Determined in Head Tissue Simulating Media
                       —                      —                                                           3
 f MHz]° P::i‘::i"‘l’;y € C°"‘:;/°:1")":y ConvF X ConvF Y ConvF Z Alpha$ E:f::)                               :‘:(:‘;t)
    750            419                 0.89              1001           i1001   1001     040       0.80       £121%
    900            415                 0.97              9.66           9.66    966      o16       139        £121%
   1750            40.1                1.37              8.35           8.35    835      024       1.05       +121%
   1900            40.0                1.40              8.10           8.10    810      0.30      0.95       +121%
   2450            39.2                1.80              7.62           7.62    762      o64       0.70       £121%
   2600            39.0                1.96              7.38           7.38    738      o60       072        £121%
   5200            36.0                4.66              5.52           5.52    552      045       120        +133%
   5300            35.9                4.76              5.28           5.28    528      045       120        +133%
   5600            355                 5.07              am             47      a7i      050       1.20       £133%
   5800            35.3                5.27              4.68           4.68    468      0.50      1.50       £13.3%

© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
" At frequency below 3 GHz, thevalidity of tissue parameters(e and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
©Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probetip diameter from the boundary.




  Certificate No: Z18—60353                              Page 5 of 11


              se® |n Collsboration with
                    TTL a
              \\Wimzzzee~"               CALIBRATION LABORATORY
           Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
           Tel:   +86—10—62304633—2512      Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com        Hup://wwa.chinattl.en



       DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7329

Calibration Parameter Determined in Body Tissue Simulating Media
                  Relative         Conductivi                                                              i6             .
 PMEA         pormagvity ®                (om) "" convF X ConvF Y ConvF Z Alpha® E;;":)                         (li':;)
    750            555                   0.96            1023           1023   1023      040        0.80        £121%
    900            55.0                  1.05            a.79           9.79   979       023        125         £121%
   1750            53.4                  1.49            8.05           8.05   8.05      025        1.05        +121%
   1900            CFF                   1.52            7.70           7.70   770       o2         115         £121%
   2450            52.7                  1.95            T47            7.47   747       056        0.83        £121%
   2600            825                   216             712            7.12   T12       065        072         €121%
   5200            49.0                  5.30            4.92           4.92   492       050        1.50        £133%
   5300            48.9                  5.42            4.79           4.79   479       0.50       1.50        £13.3%
   5600            48.5                  5.77            4.14           4714   414       060        140         £133%
   5800            48.2                  6.00            4.37           4.37   437       0.60       1.35        £133%

© Frequency validity above 300 MHz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
+50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
F At frequency below 3 GHz, the validity of tissue parameters (e and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (€ and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than x 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probetip diameter from the boundary.




  Certificate No: Z18—60353                              Page 6 of11


                                     se© .( Collsboration with
                                x7"7"J,      a
                                ~Wimzeeee>"" CALIBRATION LABORATORY
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
              E—mail: cttl@chinatt.com                           Hitp:/www.chinattl.en



                                                      Frequency Response of E—Field
                                                      (TEM—Cell: ifi110 EXX, Waveguide: R22)
       Frequency response (normalized)




                                         0.5                T              T               T
                                               0           500           1000            1500      20‘00    —    25‘00   —   30‘00

                                                                 E#3                     f [MHz]           ie
                                                                 TEM                                       R22

                                                   Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z18—60353                                                      Page 7 of 11


              owz® in Collsboration with
              77‘ s                                       e a
               imzzee>"~                   CALIEBRATION LABORATORY
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
             Tel: +86—10—62304633—25 12         Faxc: +86—10—62304633—2504
            E—mail: cttl@chinattl.com           Hitpy//www.chinattl.en



                                    Receiving Pattern (®), 0=0°

                  f=600 MHz, TEM                                                 f=1800 MHz, R22
Error[dB]




                                                                                                    T
                    -1[50               -1:)0            —50                 0      50    1(‘70    150
                                                   Rollf®1
                              [Z——I100MHz__———600MHz__——— 1800MHz__——— 2500MHz]
                                 Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




    Certificate No: Z18—60353                                  Page 8 of11


           eC (n Collsboration with
          @77"J,       a
          ~\\Wimzzee~" CALIBRATION LABORATORY
        Add: No.51 Xucyuan Road, Haidian District, B  g, 100191, China
        Tel: +86—10—62304633—2512           +86—10—62304633—2504
                                    i              Hitp//www.chinattl.en



                                          Dynamic Range f(SARneaq)
                                                      (TEM cell, f= 900 MHz)
           Input Signal[uV]




                                                                                             I

                                                      *                  10°              10‘           10°            10°
                                                                         SAR[mW/cm‘]
                                                          CIIX}not compensated      —@— compensated




                       K4                 HHH—————HH                           i———HPRH          i     {—          HAHH
                              10°            10°                   10°               , 10‘            10°             10°
                                                                      SAR[mW/cm‘]
                                        [___C=—Inot compensated             —#— compensated                   _]
                                          Uncertainty of Linearity Assessment: £0.9% (k=2)
Certificate No: Z18—60353                                       Page 9 of 11


           s© 1 Collsboration with
          @7"[‘I,    a
          \CWimzzem" CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512                               Fax: +86—10—62304633—2504
        E—mail: cttl@chinatt.com                                    Hitp:/fwww.chinattl.en



                                   Conversion Factor Assessment
       f=750 MHz, WGLS R9(H_convF)                                                               f=1750 MHz, WGLS R22(H_convF)



                                            SS—S                                                      *E2800
               350                              ‘         m                  7         ‘              3000               I                          ‘


               300                              h                                                                        |                          |
                                                |
            B
                         —*7




                                                                                                   6
                                                                                                                  k***
            §




           550                 ..3‘                                                               g 1500                 [
               1
           3                                i                                                     3
                                                                         |         |                  1000               \
               100                      |                                j         |                                     ‘                      |
               se                                                                  |                  500
               im                       ‘                                %__‘                         oso                            \g_’
                     a                  a                «              co so 10                             o0          n    m     s 4n m e            o
                                                               a{mm}                                                                    #{mm]
                                        _—anaytical


                               Deviation from Isotropy in Liquid
                               2z Ius




                                                              —10     —o80 —o%0 .040 .020    0    020        o40     080     080   10
                                                    Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z18—60353                                                         Page 10 of11


          w® (n Collsboration with
          ©7"7‘][, s p e a g
             Wimazeee~"             CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512       Fax:   +86—10—62304633—2504
        E—mail: ettl@chinatt.com        Hitp://www.chinattl.en


    DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7329


     Other Probe Parameters
     Sensor Arrangement                                                     Triangular

    Connector Angle (°)                                                          44.7

     Mechanical Surface Detection Mode                                        enabled

    Optical Surface Detection Mode                                            disable

     Probe Overall Length                                                     337mm

     Probe Body Diameter                                                        10mm

    Tip Length                                                                   9mm

    Tip Diameter                                                               2.5mm

    Probe Tip to Sensor X Calibration Point                                      1mm

    Probe Tip to Sensor Y Calibration Point                                      1mm

    Probe Tip to Sensor Z Calibration Point                                      1mm

    Recommended Measurement Distance from Surface                              1.4mm




Certificate No: Z18—60353                              Page 11 of11


DIPOLE CALIBRATION CERTIFICATES



























Document Created: 2019-10-08 05:03:44
Document Modified: 2019-10-08 05:03:44

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC