Test report HAC part 2

FCC ID: SRQ-A2020UPRO

Test Report

Download: PDF
FCCID_4401843

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab             Date: 2019/6/28


HAC_E_Dipole_835

DUT: HAC—Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: 6 =0 S/m, s, = 1; p = 0 kg/m>
Ambient Temperature : 23.6 °C

DASYS5 Configuration:
—   Probe: EF3DV3 — SN4047; ConvF(1, 1, 1) @ 835 MHz; Calibrated: 2019/1/30
—   Sensor—Surface: (Fix Surface)
—   Electronics: DAF4 Sn§53; Calibrated: 2018/7/24
—   Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
—   Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan — measurement distance from the probe sensor center to CD835 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 133.9 V/m; Power Drift = —0.05 dB
PMR not calibrated. PMF = 1.000 is applied.
E—field emissions = 112.0 V/m
Average value of Total=(108.4+112.0) / 2 = 110.2 V/m

                                               PMF scaled E—field

                                                Grid 1 M4 Grid 2 M4 Grid 3 M4
                                                108.1 V/m 108.4 V/m 100.8 V/m
                                                Grid 4 M4 Grid 5 M4 Grid 6 M4
                                                60.89 V/m 60.93 V/m 59.37 V/m
                                                Grid 7 M4 Grid 8 M4 Grid 9 M4
                                                108.1 V/m 112.0 V/m 110.5 V/m
                                                                  Cursor:
                                                          Total = 112.0 V/m
                                                           E Category: M4
                                                     Location: —0.5, 73.5, 9.7 mm




               —2.02


               —4.04


               —6.07


               —8.09


               —10.11
                           —
                          0 dB = 112.0 V/m = 40.98 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab             Date: 2019/7/18


HAC_E_Dipole_835

DUT: HAC—Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: 0 =0 S/m, s = 1; p=0 kg/m>
Ambient Temperature : 23.5 °C

DASYS5 Configuration:
— Probe: EF3DV3 — SN4047; ConvF(1, 1, 1) @ 835 MHz; Calibrated: 2019/1/30
— Sensor—Surface: (Fix Surface)
— Electronics: DAF4 Sn914; Calibrated: 2019/6/20
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (2),SEMCAD X Version 14.6.12 (7470)


E Scan — measurement distance from the probe sensor center to CD835 = 10mm & 15mm/Hearing
Aid Compatibility Test at 1 5mm distance (41x361x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 134.3 V/m; Power Drift = —0.05 dB
PMR not calibrated. PMF = 1.000 is applied.
E—field emissions = 112.4 V/m
Average value of Total=(108.7+112.4) / 2 = 110.55 V/m

                                               PMF scaled E—field

                                                Grid 1 M4         Grid 2 M4     Grid 3 M4
                                                108.7 V/m 108.7 V/m 101.1 V/m
                                                Grid 4 M4         Grid 5 M4     Grid 6 M4
                                                61.06 V/m 61.11 V/m 59.54 V/im
                                                Grid 7 M4         Grid 8 M4     Grid 9 M4
                                                108.4 V/m 112.4 V/m 110.9 V/m
                                                                  Cursor:
                                                          Total = 112.4 V/m
                                                           E Category: M4
                                                     Location: —0.5, 73.5, 9.7 mm




               —2.02


               —4.04


               —6.06


               —8.08

                           —
               —10.10
                          0 dB = 112.4 V/m = 41.02 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab             Date: 2019/6/28


HAC_E_Dipole_1880
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency : 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: 0 = 0 S/m, s, = 1; p =0 kg/m>
Ambient Temperature : 23.6 C

DASYS5 Configuration:
— Probe: EF3DV3 — SN4047; ConvF(1, 1, 1) @ 1880 MHz; Calibrated: 2019/1/30
— Sensor—Surface: (Fix Surface)
— Electronics: DAF4 Sn§53; Calibrated: 2018/7/24
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan — measurement distance from the probe sensor center to CD1880 =10mm & 15mm/Hearing
Aid Compatibility Test at 1 5mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 157.8 V/m; Power Drift = 0.00 dB
PMR not calibrated. PMF = 1.000 is applied.
E—field emissions = 87.21 V/m
Average value of Total=(86.74+87.21) / 2 = 86.975 V/m

                                               PMF scaled E—field

                                                Grid 1 M3         Grid 2 M3     Grid 3 M3
                                                86.01 V/m 86.74 V/m 83.53 V/m
                                                Grid 4 M3         Grid 5 M3     Grid 6 M3
                                                64.44 V/m 64.69 V/m 63.35 V/m
                                                Grid 7 M3         Grid 8 M3     Grid 9 M3
                                                86.61 V/m 87.21 V/m 83.78 V/m
                                                                  Cursor:
                                                           Total =87.21 V/m
                                                            E Category: M3
                                                        Location: 1, 34, 9.7 mm




               —0.86


               —1.73


               —2.59


               —3.416


               —4.32
                           —
                          0 dB =87.21 V/m =38.81 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019/7/23


HAC_E_Dipole_1880
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 1880 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn914; Calibrated: 2019/6/20
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 156.6 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 88.21 V/m
Average value of Total=(88.21+87.01) / 2 = 87.61 V/m

                                               PMF scaled E-field
                                                Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                87.89 V/m         88.21 V/m   84.78 V/m
                                                Grid 4 M3         Grid 5 M3   Grid 6 M3
                                                65.54 V/m         65.64 V/m   64.19 V/m
                                                Grid 7 M3         Grid 8 M3   Grid 9 M3
                                                86.29 V/m         87.01 V/m   84.10 V/m
                                                              Cursor:
                                                         Total = 88.21 V/m
                                                           E Category: M3
                                                      Location: 1.5, -33, 9.7 mm




                          0 dB = 88.21 V/m = 38.91 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab             Date: 2019/6/28


HAC_E_Dipole_2450
DUT: HAC Dipole 2450 MHz

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: 0 = 0 S/m, s, = 1; p =0 kg/m>
Ambient Temperature : 23.6 C

DASYS5 Configuration:
— Probe: EF3DV3 — SN4047; ConvF(1, 1, 1) @ 2450 MHz; Calibrated: 2019/1/30
— Sensor—Surface: (Fix Surface)
— Electronics: DAF4 Sn§53; Calibrated: 2018/7/24
— Phantom: HAC Test Arch with AMCC; Type: SD HAC POI BA;
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan — measurement distance from the probe sensor center to CD2450 =10mm & 15mm
MHearing Aid Compatibility Test at 1 5mm distance (41x181x1): Interpolated grid: dx=0.5000 mm,
dy=0.5000 mm
Device Reference Point: 0, 0, —6.3 mm
Reference Value = 82.73 V/m; Power Drift =0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E—field emissions = 87.54 V/m
Average value of Total=(85.82+87.54) / 2 = 86.68 V/m

                                               PMF scaled E—field

                                                Grid 1 M3         Grid 2 M3     Grid 3 M3
                                                85.12 V/m 85.82 V/m 82.61 V/m
                                                Grid 4 M3         Grid 5 M3     Grid 6 M3
                                                76.51 V/m 76.98 V/m 75.06 V/im
                                                Grid 7 M3         Grid 8 M3     Grid 9 M3
                                                8$5.80 V/m 8$7.54 V/im 85.56 V/m
                                                                  Cursor:
                                                           Total = 87.54 V/m
                                                            E Category: M3
                                                       Location: 0, 24.5, 9.7 mm




               —1.36


               —2.12


               —4.08


               —5.44


               —6.80
                          0 dB = §7.54 V/m = 38.84 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019/6/28


HAC_E_Dipole_2600
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.6 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 74.06 V/m; Power Drift = 0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 90.25 V/m
Average value of Total=(87.71+90.25) / 2 = 88.98 V/m

                                               PMF scaled E-field
                                                Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                86.54 V/m         87.71 V/m   85.20 V/m
                                                Grid 4 M3         Grid 5 M3   Grid 6 M3
                                                81.18 V/m         81.87 V/m   80.10 V/m
                                                Grid 7 M3         Grid 8 M3   Grid 9 M3
                                                89.21 V/m         90.25 V/m   87.70 V/m
                                                              Cursor:
                                                         Total = 90.25 V/m
                                                          E Category: M3
                                                      Location: 0.5, 23, 9.7 mm




                          0 dB = 90.25 V/m = 39.11 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab         Date: 2019/7/18


HAC_E_Dipole_2600
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2019/1/30
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn914; Calibrated: 2019/6/20
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm & 15mm/Hearing
Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 74.38 V/m; Power Drift = 0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 90.52 V/m
Average value of Total=(87.97+90.52) / 2 = 89.25 V/m

                                               PMF scaled E-field
                                                Grid 1 M3 Grid 2 M3 Grid 3 M3
                                                86.79 V/m         87.97 V/m   85.45 V/m
                                                Grid 4 M3         Grid 5 M3   Grid 6 M3
                                                81.42 V/m         82.11 V/m   80.34 V/m
                                                Grid 7 M3         Grid 8 M3   Grid 9 M3
                                                89.48 V/m         90.52 V/m   87.96 V/m
                                                              Cursor:
                                                         Total = 90.52 V/m
                                                          E Category: M3
                                                      Location: 0.5, 23, 9.7 mm




                          0 dB = 90.52 V/m = 39.13 dBV/m



Document Created: 2019-08-13 19:39:05
Document Modified: 2019-08-13 19:39:05