SAR report part 1

FCC ID: SERCWFB124

RF Exposure Info

Download: PDF
FCCID_4389367

         FCC SAR TEST REPORT                                                                 Report No. : FA971705




                       FCC SAR TEST REPORT

    FCC ID                : SERCWFB124
    Equipment             : Wi-Fi Module
    Model Name            : CWFB-124
    Applicant             : Sintai Optical(Shenzhen)Co., Ltd.
                            Qiwei Ind Sec,1st, 2nd, & 3RD Bldg, Lisonglang Village,
                            Gongming Town, Bao'an District, Shenzhen, Guangdong,
                            China
    Manufacturer          : Sintai Optical(Shenzhen)Co., Ltd.
                            Qiwei Ind Sec,1st, 2nd, & 3RD Bldg, Lisonglang Village,
                            Gongming Town, Bao'an District, Shenzhen, Guangdong,
                            China
    Standard              : FCC 47 CFR Part 2 (2.1093)
                            ANSI/IEEE C95.1-1992
                            IEEE 1528-2013

      The product was installed into Digital Camera (Brand Name: KODAK, Model Name: KODAK
      PIXPRO WPZ2 Digital Camera) during test.

      The product was received on Jul. 20, 2019 and testing was started from Jul. 24, 2019 and
      completed on Jul. 24, 2019. We, SPORTON INTERNATIONAL INC., would like to declare that
      the tested sample has been evaluated in accordance with the test procedures and has been
      in compliance with the applicable technical standards.

      The report must not be used by the client to claim product certification, approval, or
      endorsement by TAF or any agency of government.

      The test results in this report apply exclusively to the tested model / sample. Without
      written approval of SPORTON INTERNATIONAL INC. EMC & Wireless Communications
      Laboratory, the test report shall not be reproduced except in full.




      Approved by: Cona Huang / Deputy Manager



SPORTON INTERNATIONAL INC. EMC & Wireless Communications Laboratory
                       No. 52, Huaya 1st Rd., Guishan Dist., Taoyuan City, Taiwan (R.O.C.)



TEL : 886-3-327-3456                                                                     Page 1 of 20
FAX : 886-3-328-4978                                                               Issued Date : Aug. 01, 2019
Form version: 181113


              FCC SAR TEST REPORT                                                                                                                  Report No. : FA971705

                                                                Table of Contents
1. Statement of Compliance ............................................................................................................................................. 4
2. Guidance Applied .......................................................................................................................................................... 4
3. Equipment Under Test (EUT) Information ................................................................................................................... 4
     3.1 General Information ............................................................................................................................................... 4
4. RF Exposure Limits....................................................................................................................................................... 5
     4.1 Uncontrolled Environment ...................................................................................................................................... 5
     4.2 Controlled Environment.......................................................................................................................................... 5
5. Specific Absorption Rate (SAR) ................................................................................................................................... 6
     5.1 Introduction ............................................................................................................................................................ 6
     5.2 SAR Definition ........................................................................................................................................................ 6
6. System Description and Setup .................................................................................................................................... 7
     6.1 E-Field Probe ......................................................................................................................................................... 8
     6.2 Data Acquisition Electronics (DAE) ........................................................................................................................ 8
     6.3 Phantom ................................................................................................................................................................. 9
     6.4 Device Holder........................................................................................................................................................10
7. Measurement Procedures ...........................................................................................................................................11
     7.1 Spatial Peak SAR Evaluation ................................................................................................................................ 11
     7.2 Power Reference Measurement............................................................................................................................12
     7.3 Area Scan .............................................................................................................................................................12
     7.4 Zoom Scan ............................................................................................................................................................13
     7.5 Volume Scan Procedures ......................................................................................................................................14
     7.6 Power Drift Monitoring...........................................................................................................................................14
8. Test Equipment List .....................................................................................................................................................14
9. System Verification ......................................................................................................................................................15
     9.1 Tissue Simulating Liquids ......................................................................................................................................15
     9.2 Tissue Verification .................................................................................................................................................16
     9.3 System Performance Check Results .....................................................................................................................17
10. Conducted RF Output Power (Unit: dBm) ................................................................................................................18
11. SAR Test Results .......................................................................................................................................................19
     11.1 Head SAR ...........................................................................................................................................................19
     11.2 Hand SAR ...........................................................................................................................................................19
12. Uncertainty Assessment ...........................................................................................................................................20
13. References ..................................................................................................................................................................20
Appendix A. Plots of System Performance Check
Appendix B. Plots of High SAR Measurement
Appendix C. DASY Calibration Certificate
Appendix D. Test Setup Photos




TEL : 886-3-327-3456                                                                                                                          Page 2 of 20
FAX : 886-3-328-4978                                                                                                                    Issued Date : Aug. 01, 2019
Form version: 181113


         FCC SAR TEST REPORT                                                    Report No. : FA971705



                                 History of this test report
    Report No.         Version                             Description              Issued Date

     FA971705            01      Initial issue of report                           Aug. 01, 2019




TEL : 886-3-327-3456                                                           Page 3 of 20
FAX : 886-3-328-4978                                                     Issued Date : Aug. 01, 2019
Form version: 181113


seorronias. FCC SAR TEST REPORT                                                                          Report No. : FA971705
    1. Statement of Compliance
 The       maximum      results    of   Specific   Absorption     Rate   (SAR)    found    during       testing   for   Sintai
 Optical(Shenzhen)Co., Ltd., Wi—Fi Module, CWFB—124, are as follows.


                                                                                       Highest SAR Summary
           Equipment                           es                               Head                        Hand
             Class                              eCls                      (Separation Omm)            (Separation Omm)
                                                                            irercitte                     10g SAR (Whkg)
             DTS                                    2.4GHz WLAN                0.06                            0.18
                            Date of Testing:                                                2019/7/24


 Sporton Lab is accredited to ISO 17025 by Taiwan Accreditation Foundation (TAF code: 1190) and the FCC
 designation No. TW1190 under the FCC 2.948(e) by Mutual Recognition Agreement (MRA) in FCC test. This
 device is in compliance with Specific Absorption Rate (SAR) for general population/uncontrolled exposure
  limits (1.6 W/kg for Head 1g SAR, 4.0 W/kg for Hand 10g SAR) specified in FCC 47 CFR part 2 (2.1093) and
 ANSI/IEEE €95.1—1992, and had been tested in accordance with the measurement methods and procedures
 specified in IEEE 1528—2013 and FCC KDB publications


  Reviewed by: Jason Wang
  Report Producer: Daisy Peng


 2. Guidance Applied
       The Specific Absorption Rate (SAR) testing specification, method, and procedure for this device is in accordance with
       the following standards:
       >      FCC 47 CFR Part 2 (2.1093)
              ANSVIEEE C95.1—1992
              IEEE 1528—2013
              FCC KDB 865664 D01 SAR Measurement 100 MHz to 6 GHz v01r04
              FCC KDB 865664 D02 SAR Reporting v01r02
              FCC KDB 447498 D01 General RF Exposure Guidance v06
              FCC KDB 248227 D01 802.11 Wi—Fi SAR v02r02



  3. Equipment Under Test (EUT) Information

  3.1 General Information

                                  Wi—Fi Module
                                  CWEFB—124
                                  SERCWEFB124
                                  WLAN 2.4GHz Band: 2412 MHz ~ 2462 MHz
                                  WLAN: 802.11b/g/n HT20 / HT4O
                                  Pre—production




 TEL : 886—3—327—3456                                                                                Page 4 0f 20
 FAX : 886—3—328—4978                                                                          Issued Date : Aug. 01, 2019
 Form version: 181113


          FCC SAR TEST REPORT                                                                      Report No. : FA971705
4. RF Exposure Limits
4.1 Uncontrolled Environment
Uncontrolled Environments are defined as locations where there is the exposure of individuals who have no knowledge or
control of their exposure. The general population/uncontrolled exposure limits are applicable to situations in which the
general public may be exposed or in which persons who are exposed as a consequence of their employment may not be
made fully aware of the potential for exposure or cannot exercise control over their exposure. Members of the general
public would come under this category when exposure is not employment-related; for example, in the case of a wireless
transmitter that exposes persons in its vicinity.


4.2 Controlled Environment
Controlled Environments are defined as locations where there is exposure that may be incurred by persons who are
aware of the potential for exposure, (i.e. as a result of employment or occupation). In general, occupational/controlled
exposure limits are applicable to situations in which persons are exposed as a consequence of their employment, who
have been made fully aware of the potential for exposure and can exercise control over their exposure. The exposure
category is also applicable when the exposure is of a transient nature due to incidental passage through a location where
the exposure levels may be higher than the general population/uncontrolled limits, but the exposed person is fully aware
of the potential for exposure and can exercise control over his or her exposure by leaving the area or by some other
appropriate means.


                             Limits for Occupational/Controlled Exposure (W/kg)




                        Limits for General Population/Uncontrolled Exposure (W/kg)




     1.   Whole-Body SAR is averaged over the entire body, partial-body SAR is averaged over any 1gram of tissue
          defined as a tissue volume in the shape of a cube. SAR for hands, wrists, feet and ankles is averaged over any
          10 grams of tissue defined as a tissue volume in the shape of a cube.




TEL : 886-3-327-3456                                                                              Page 5 of 20
FAX : 886-3-328-4978                                                                        Issued Date : Aug. 01, 2019
Form version: 181113


         FCC SAR TEST REPORT                                                                            Report No. : FA971705
5. Specific Absorption Rate (SAR)
5.1 Introduction
    SAR is related to the rate at which energy is absorbed per unit mass in an object exposed to a radio field. The SAR
    distribution in a biological body is complicated and is usually carried out by experimental techniques or numerical
    modeling. The standard recommends limits for two tiers of groups, occupational/controlled and general
    population/uncontrolled, based on a person’s awareness and ability to exercise control over his or her exposure. In
    general, occupational/controlled exposure limits are higher than the limits for general population/uncontrolled.


5.2 SAR Definition
    The SAR definition is the time derivative (rate) of the incremental energy (dW) absorbed by (dissipated in) an
    incremental mass (dm) contained in a volume element (dv) of a given density (ρ). The equation description is as
    below:

                                                        𝐝 𝐝𝐖    𝐝 𝐝𝐖
                                              𝐒𝐀𝐑 =       (   )= (    )
                                                        𝐝𝐭 𝐝𝐦   𝐝𝐭 𝛒𝐝𝐯
    SAR is expressed in units of Watts per kilogram (W/kg)

                                                              𝛔|𝐄|𝟐
                                                        𝐒𝐀𝐑 =
                                                                𝛒
    Where: σ is the conductivity of the tissue, ρ is the mass density of the tissue and E is the RMS electrical field strength.




TEL : 886-3-327-3456                                                                                  Page 6 of 20
FAX : 886-3-328-4978                                                                            Issued Date : Aug. 01, 2019
Form version: 181113


         FCC SAR TEST REPORT                                                              Report No. : FA971705
6. System Description and Setup
The DASY system used for performing compliance tests consists of the following items:




     A standard high precision 6-axis robot with controller, teach pendant and software. An arm extension for
      accommodating the data acquisition electronics (DAE).
     An isotropic Field probe optimized and calibrated for the targeted measurement.
     A data acquisition electronics (DAE) which performs the signal amplification, signal multiplexing,
      AD-conversion, offset measurements, mechanical surface detection, collision detection, etc. The unit is
      battery powered with standard or rechargeable batteries. The signal is optically transmitted to the EOC.
     The Electro-optical converter (EOC) performs the conversion from optical to electrical signals for the
      digital communication to the DAE. To use optical surface detection, a special version of the EOC is
      required. The EOC signal is transmitted to the measurement server.
     The function of the measurement server is to perform the time critical tasks such as signal filtering,
      control of the robot operation and fast movement interrupts.
     The Light Beam used is for probe alignment. This improves the (absolute) accuracy of the probe
      positioning.
     A computer running WinXP or Win7 and the DASY5 software.
     Remote control and teach pendant as well as additional circuitry for robot safety such as warning lamps,
      etc.
     The phantom, the device holder and other accessories according to the targeted measurement.




TEL : 886-3-327-3456                                                                     Page 7 of 20
FAX : 886-3-328-4978                                                               Issued Date : Aug. 01, 2019
Form version: 181113


spontonm Las.   FCC SAR TEST REPORT                                                                     Report No. : FA971705

    6.1 E—Field Probe
   The SAR measurement is conducted with the dosimetric probe (manufactured by SPEAG).The probe is specially
   designed and calibrated for use in liquid with high permittivity. The dosimetric probe has special calibration in liquid at
   different frequency. This probe has a built in optical surface detection system to prevent from collision with phantom.


   <ES3DV3 Probe>
   Construction              Symmetric design with triangular core
                             Interleaved sensors
                             Built—in shielding against static charges
                             PEEK enclosure material (resistant to organic
                             solvents, e.g., DGBE)
   Frequency                 10 MHz — 4 GHz;
                             Linearity: +0.2 dB (30 MHz —4 GHz)
   Directivity               +0.2 dB in TSL (rotation around probe axis)
                             +0.3 dB in TSL (rotation normal to probe axis)
   Dynamic Range             5 uW/g —>100 mWig;
                             Linearity: +0.2 dB
   Dimensions                Overall length: 337 mm (tip: 20 mm)
                             Tip diameter: 3.9 mm (body: 12 mm)
                             Distance from probe tip to dipole centers: 3.0 mm


   <EX3DV4 Probe>
   Construction              Symmetric design with triangular core
                             Built—in shielding against static charges
                             PEEK enclosure material (resistant to organic
                             solvents, e.g., DGBE)
   Frequency                 10 MHz —>6 GHz
                             Linearity: +0.2 dB (30 MHz —6 GHz)
   Directivity               +0.3 dB in TSL (rotation around probe axis)
                             +0.5 dB in TSL (rotation normal to probe axis)
   Dynamic Range             10 uW/g—>100 mW/g
                             Linearity: +0.2 dB (noise: typically <1 uW/g)
   Dimensions                Overall length: 337 mm (tip: 20 mm)
                             Tip diameter: 2.5 mm (body: 12 mm)
                             Typical distance from probe tip to dipole centers: 1
                             mm



    6.2 Data Acquisition Electronics (DAE)

   The data acquisition electronics (DAE) consists of a highly sensitive
   electrometer—grade     preamplifier   with   auto—zeroing,   a   channel   and
   gain—switching multiplexer, a fast 16 bit AD—converter and a command
   decoder and control logic unit. Transmission to the measurement server is
   accomplished through an optical downlink for data and status information as
   well as an optical uplink for commands and the clock.
   The input impedance of the DAE is 200 MOhm; the inputs are symmetrical
   and floating. Common mode rejection is above 80 dB.

                                                                                           Fig 5.1     Photo of DAE




   TEL : 886—3—327—3456                                                                                Page   8 of 20
   FAX : 886—3—328—4978                                                                         Issued Date ; Aug. 01, 2019
   Form version: 181113


      FCC SAR TEST REPORT                                                                            Report No. : FA971705
6.3 Phantom
<SAM Twin Phantom>
Shell Thickness              2 ± 0.2 mm;
                             Center ear point: 6 ± 0.2 mm
Filling Volume               Approx. 25 liters
Dimensions                   Length: 1000 mm; Width: 500 mm; Height:
                             adjustable feet
Measurement Areas            Left Hand, Right Hand, Flat Phantom




The bottom plate contains three pair of bolts for locking the device holder. The device holder positions are adjusted to the
standard measurement positions in the three sections. A white cover is provided to tap the phantom during off-periods to
prevent water evaporation and changes in the liquid parameters. On the phantom top, three reference markers are
provided to identify the phantom position with respect to the robot.



<ELI Phantom>
Shell Thickness          2 ± 0.2 mm (sagging: <1%)
Filling Volume           Approx. 30 liters
Dimensions               Major ellipse axis: 600 mm
                         Minor axis: 400 mm




The ELI phantom is intended for compliance testing of handheld and body-mounted wireless devices in the frequency
range of 30 MHz to 6 GHz. ELI4 is fully compatible with standard and all known tissue simulating liquids.




TEL : 886-3-327-3456                                                                                Page 9 of 20
FAX : 886-3-328-4978                                                                          Issued Date : Aug. 01, 2019
Form version: 181113


      FCC SAR TEST REPORT                                                                            Report No. : FA971705
6.4 Device Holder
<Mounting Device for Hand-Held Transmitter>
In combination with the Twin SAM V5.0/V5.0c or ELI phantoms, the Mounting Device for Hand-Held Transmitters enables
rotation of the mounted transmitter device to specified spherical coordinates. At the heads, the rotation axis is at the ear
opening. Transmitter devices can be easily and accurately positioned according to IEC 62209-1, IEEE 1528, FCC, or
other specifications. The device holder can be locked for positioning at different phantom sections (left head, right head,
flat). And upgrade kit to Mounting Device to enable easy mounting of wider devices like big smart-phones, e-books, small
tablets, etc. It holds devices with width up to 140 mm.




                          Mounting Device for Hand-Held
                                                               Mounting Device Adaptor for Wide-Phones
                                  Transmitters




<Mounting Device for Laptops and other Body-Worn Transmitters>
The extension is lightweight and made of POM, acrylic glass and foam. It fits easily on the upper part of the mounting
device in place of the phone positioned. The extension is fully compatible with the SAM Twin and ELI phantoms.




                                                Mounting Device for Laptops




TEL : 886-3-327-3456                                                                                Page 10 of 20
FAX : 886-3-328-4978                                                                          Issued Date : Aug. 01, 2019
Form version: 181113


           FCC SAR TEST REPORT                                                                         Report No. : FA971705
7. Measurement Procedures
   The measurement procedures are as follows:

   <Conducted power measurement>
   (a) For WWAN power measurement, use base station simulator to configure EUT WWAN transmission in conducted
       connection with RF cable, at maximum power in each supported wireless interface and frequency band.
   (b) Read the WWAN RF power level from the base station simulator.
   (c) For WLAN/BT power measurement, use engineering software to configure EUT WLAN/BT continuously
       transmission, at maximum RF power in each supported wireless interface and frequency band
   (d) Connect EUT RF port through RF cable to the power meter, and measure WLAN/BT output power

   <SAR measurement>
   (a) Use base station simulator to configure EUT WWAN transmission in radiated connection, and engineering
       software to configure EUT WLAN/BT continuously transmission, at maximum RF power, in the highest power
       channel.
   (b) Place the EUT in the positions as Appendix D demonstrates.
   (c) Set scan area, grid size and other setting on the DASY software.
   (d) Measure SAR results for the highest power channel on each testing position.
   (e) Find out the largest SAR result on these testing positions of each band
   (f) Measure SAR results for other channels in worst SAR testing position if the reported SAR of highest power
       channel is larger than 0.8 W/kg

         According to the test standard, the recommended procedure for assessing the peak spatial-average SAR value
         consists of the following steps:
   (a)     Power reference measurement
   (b)     Area scan
   (c)     Zoom scan
   (d)     Power drift measurement


7.1 Spatial Peak SAR Evaluation
     The procedure for spatial peak SAR evaluation has been implemented according to the test standard. It can be
     conducted for 1g and 10g, as well as for user-specific masses. The DASY software includes all numerical
     procedures necessary to evaluate the spatial peak SAR value.
     The base for the evaluation is a "cube" measurement. The measured volume must include the 1g and 10g cubes
     with the highest averaged SAR values. For that purpose, the center of the measured volume is aligned to the
     interpolated peak SAR value of a previously performed area scan.
     The entire evaluation of the spatial peak values is performed within the post-processing engine (SEMCAD). The
     system always gives the maximum values for the 1g and 10g cubes. The algorithm to find the cube with highest
     averaged SAR is divided into the following stages:
     (a)    Extraction of the measured data (grid and values) from the Zoom Scan
     (b)    Calculation of the SAR value at every measurement point based on all stored data (A/D values and
            measurement parameters)
     (c)    Generation of a high-resolution mesh within the measured volume
     (d)    Interpolation of all measured values form the measurement grid to the high-resolution grid
     (e)    Extrapolation of the entire 3-D field distribution to the phantom surface over the distance from sensor to surface
     (f)    Calculation of the averaged SAR within masses of 1g and 10g




TEL : 886-3-327-3456                                                                                  Page 11 of 20
FAX : 886-3-328-4978                                                                            Issued Date : Aug. 01, 2019
Form version: 181113


      FCC SAR TEST REPORT                                                                           Report No. : FA971705
7.2 Power Reference Measurement
The Power Reference Measurement and Power Drift Measurements are for monitoring the power drift of the device under
test in the batch process. The minimum distance of probe sensors to surface determines the closest measurement point
to phantom surface. This distance cannot be smaller than the distance of sensor calibration points to probe tip as defined
in the probe properties.


7.3 Area Scan
The area scan is used as a fast scan in two dimensions to find the area of high field values, before doing a fine
measurement around the hot spot. The sophisticated interpolation routines implemented in DASY software can find the
maximum found in the scanned area, within a range of the global maximum. The range (in dB0 is specified in the
standards for compliance testing. For example, a 2 dB range is required in IEEE standard 1528 and IEC 62209 standards,
whereby 3 dB is a requirement when compliance is assessed in accordance with the ARIB standard (Japan), if only one
zoom scan follows the area scan, then only the absolute maximum will be taken as reference. For cases where multiple
maximums are detected, the number of zoom scans has to be increased accordingly.
Area scan parameters extracted from FCC KDB 865664 D01v01r04 SAR measurement 100 MHz to 6 GHz.




TEL : 886-3-327-3456                                                                              Page 12 of 20
FAX : 886-3-328-4978                                                                        Issued Date : Aug. 01, 2019
Form version: 181113


       FCC SAR TEST REPORT                                                                   Report No. : FA971705
7.4 Zoom Scan
Zoom scans are used assess the peak spatial SAR values within a cubic averaging volume containing 1 gram and 10
gram of simulated tissue. The zoom scan measures points (refer to table below) within a cube shoes base faces are
centered on the maxima found in a preceding area scan job within the same procedure. When the measurement is done,
the zoom scan evaluates the averaged SAR for 1 gram and 10 gram and displays these values next to the job’s label.
Zoom scan parameters extracted from FCC KDB 865664 D01v01r04 SAR measurement 100 MHz to 6 GHz.




TEL : 886-3-327-3456                                                                        Page 13 of 20
FAX : 886-3-328-4978                                                                  Issued Date : Aug. 01, 2019
Form version: 181113


       FCC SAR TEST REPORT                                                                          Report No. : FA971705
7.5 Volume Scan Procedures
The volume scan is used for assess overlapping SAR distributions for antennas transmitting in different frequency bands.
It is equivalent to an oversized zoom scan used in standalone measurements. The measurement volume will be used to
enclose all the simultaneous transmitting antennas. For antennas transmitting simultaneously in different frequency bands,
the volume scan is measured separately in each frequency band. In order to sum correctly to compute the 1g aggregate
SAR, the EUT remain in the same test position for all measurements and all volume scan use the same spatial resolution
and grid spacing. When all volume scan were completed, the software, SEMCAD postprocessor can combine and
subsequently superpose these measurement data to calculating the multiband SAR.


7.6 Power Drift Monitoring
All SAR testing is under the EUT install full charged battery and transmit maximum output power. In DASY measurement
software, the power reference measurement and power drift measurement procedures are used for monitoring the power
drift of EUT during SAR test. Both these procedures measure the field at a specified reference position before and after
the SAR testing. The software will calculate the field difference in dB. If the power drifts more than 5%, the SAR will be
retested.



8. Test Equipment List
                                                                                                       Calibration
Manufacturer            Name of Equipment            Type/Model         Serial Number
                                                                                              Last Cal.            Due Date
   SPEAG          2450MHz System Validation Kit        D2450V2               736            Aug. 31, 2018     Aug. 30, 2019
   SPEAG           Data Acquisition Electronics         DAE4                 914            Jun. 20, 2019     Jun. 19, 2020
   SPEAG               Dosimetric E-Field Probe        ES3DV3                3170           Nov. 02, 2018     Nov. 01, 2019
  RCPTWN                    Thermometer                 HTC-1              TM560-2          Nov. 12, 2018     Nov. 11, 2019
   SPEAG                    Device Holder                N/A                 N/A                 N/A                 N/A
   Anritsu                Signal Generator            MG3710A            6201502524         Dec. 11, 2018     Dec. 10, 2019
   Agilent              ENA Network Analyzer           E5071C            MY46104758         Sep. 19, 2018     Sep. 18, 2019
   SPEAG                 Dielectric Probe Kit          DAK-3.5               1126           Sep. 19, 2018     Sep. 18, 2019
 LINE SEIKI              Digital Thermometer       DTM3000-spezial           3169           Sep. 11, 2018     Sep. 10, 2019
   Anritsu                  Power Meter               ML2495A              1218006           Oct. 08, 2018    Oct. 07, 2019
   Anritsu                  Power Sensor              MA2411B              1207363           Oct. 08, 2018    Oct. 07, 2019
   Anritsu                  Power Meter               ML2495A              1419002          May. 29, 2019     May. 28, 2020
   Anritsu                  Power Sensor              MA2411B              1339124          May. 29, 2019     May. 28, 2020
   Agilent               Spectrum Analyzer             E4408B            MY44211028         Aug. 28, 2018     Aug. 27, 2019
   Anritsu               Spectrum Analyzer            MS2830A            6201396378         Jun. 27, 2019     Jun. 26, 2020
 Mini-Circuits             Power Amplifier             ZVE-8G+            070501814          Oct. 08, 2018    Oct. 07, 2019
 Mini-Circuits             Power Amplifier             ZVE-8G+               6382           Aug. 09, 2018     Aug. 08, 2019
     ATM               Dual Directional Coupler       C122H-10           P610410z-02                      Note 1
   Woken                     Attenuator 1            WK0602-XX               N/A                          Note 1
     PE                      Attenuator 2             PE7005-10              N/A                          Note 1
     PE                      Attenuator 3             PE7005- 3              N/A                          Note 1
General Note:
 1.  Prior to system verification and validation, the path loss from the signal generator to the system check source and
     the power meter, which includes the amplifier, cable, attenuator and directional coupler, was measured by the
     network analyzer. The reading of the power meter was offset by the path loss difference between the path to the
     power meter and the path to the system check source to monitor the actual power level fed to the system check
     source.




TEL : 886-3-327-3456                                                                               Page 14 of 20
FAX : 886-3-328-4978                                                                         Issued Date : Aug. 01, 2019
Form version: 181113


         FCC SAR TEST REPORT                                                               Report No. : FA971705
9. System Verification
9.1 Tissue Simulating Liquids
    For the measurement of the field distribution inside the SAM phantom with DASY, the phantom must be
    filled with around 25 liters of homogeneous body tissue simulating liquid. For head SAR testing, the liquid
    height from the ear reference point (ERP) of the phantom to the liquid top surface is larger than 15 cm,
    which is shown in Fig. 10.1. For body SAR testing, the liquid height from the center of the flat phantom to
    the liquid top surface is larger than 15 cm, which is shown in Fig. 10.2.




      Fig 10.1Photo of Liquid Height for Head SAR           Fig 10.2 Photo of Liquid Height for Body SAR




TEL : 886-3-327-3456                                                                      Page 15 of 20
FAX : 886-3-328-4978                                                                Issued Date : Aug. 01, 2019
Form version: 181113


seorrow cas. _FCC SAR TEST REPORT                                                                                    Report No. : FA971705
  9.2 Tissue Verification
  The following tissue formulations are provided for reference only as some of the parameters have not been
  thoroughly verified. The composition of ingredients may be modified accordingly to achieve the desired target
  tissue parameters required for routine SAR evaluation.




           750                   41.1                      .           1.4
           835                   40.3                      .           1.4
           900                   40.3                      .           1.4
    1800, 1900, 2000             55.2                                  0.3
           2450                  55.0                                   0
           2600                  54.8                                  0.1


           750                   51.7                                  0.9          0.1
           835                   50.8                                  0.9          0.1
           900                   50.8                                  0.9
    1800, 1900, 2000             70.2          0                       0.4           0
           2450                  68.6          0                        0            0
           2600                  68.1          0                       0.1           0


  Simulating Liquid for 5GHz, Manufactured by SPEAG

                        Water                                    64~78%
                   Mineral oil                                    11~18%
                   Emulsifiers                                    9~15%
             Additives and Sait                                   2~8%



  <Tissue Dielectric Parameter Check Results>
  Frequency Tissue         -I;qur:;j    Conductivity Permittivity Conductivity Permittivity Delta (0)   Delta (€:)   ooo         Date
    (MHz)    Type            frol] i         [(          (€)       Target (0) Target (s))      Eo)         o            h

    2450          HSL       22.5           1.793        39.212        1.80        39.20       —0.39       0.03         +5     2019/7/24




  TEL : 886—3—327—3456                                                                                           Page 16 of 20
  FAX : 886—3—328—4978                                                                                     Issued Date : Aug. 01, 2019
  Form version: 181113


       FCC SAR TEST REPORT                                                                        Report No. : FA971705
9.3 System Performance Check Results
Comparing to the original SAR value provided by SPEAG, the verification data should be within its specification of 10 %.
Below table shows the target SAR and measured SAR after normalized to 1W input power. The table below indicates the
system performance check can meet the variation criterion and the plots can be referred to Appendix A of this report.
                           Input                                      Measured Targeted Normalized
          Frequency Tissue        Dipole        Probe         DAE                                  Deviation
   Date                    Power                                       1g SAR 1g SAR     1g SAR
            (MHz)    Type           S/N          S/N           S/N                                    (%)
                           (mW)                                        (W/kg)   (W/kg)    (W/kg)
2019/7/24    2450    HSL    250 D2450V2-736 ES3DV3 - SN3170 DAE4 Sn914 12.30     52.70     49.2      -6.64



                           Input                                       Measured Targeted Normalized
          Frequency Tissue        Dipole        Probe         DAE                                   Deviation
  Date                     Power                                       10g SAR 10g SAR 10g SAR
            (MHz)    Type           S/N          S/N           S/N                                     (%)
                           (mW)                                         (W/kg)   (W/kg)    (W/kg)
2019/7/24    2450    HSL    250 D2450V2-736 ES3DV3 - SN3170 DAE4 Sn914   5.72     24.60     22.88     -6.99




          Fig 8.3.1 System Performance Check Setup                                      Fig 8.3.2 Setup Photo




TEL : 886-3-327-3456                                                                             Page 17 of 20
FAX : 886-3-328-4978                                                                       Issued Date : Aug. 01, 2019
Form version: 181113


          FCC SAR TEST REPORT                                                                                    Report No. : FA971705
10. Conducted RF Output Power (Unit: dBm)
<WLAN Conducted Power>
General Note:
1.  Per KDB 248227 D01v02r02, SAR test reduction is determined according to 802.11 transmission mode configurations and
      certain exposure conditions with multiple test positions. In the 2.4 GHz band, separate SAR procedures are applied to
      DSSS and OFDM configurations to simplify DSSS test requirements. For OFDM, in both 2.4 and 5 GHz bands, an initial
      test configuration must be determined for each standalone and aggregated frequency band, according to the transmission
      mode configuration with the highest maximum output power specified for production units to perform SAR measurements.
      If the same highest maximum output power applies to different combinations of channel bandwidths, modulations and data
      rates, additional procedures are applied to determine which test configurations require SAR measurement. When
      applicable, an initial test position may be applied to reduce the number of SAR measurements required for next to the ear,
      UMPC mini-tablet or hotspot mode configurations with multiple test positions.
2.    For 2.4 GHz 802.11b DSSS, either the initial test position procedure for multiple exposure test positions or the DSSS
      procedure for fixed exposure position is applied; these are mutually exclusive. For 2.4 GHz and 5 GHz OFDM
      configurations, the initial test configuration is applied to measure SAR using either the initial test position procedure for
      multiple exposure test position configurations or the initial test configuration procedures for fixed exposure test conditions.
      Based on the reported SAR of the measured configurations and maximum output power of the transmission mode
      configurations that are not included in the initial test configuration, the subsequent test configuration and initial test position
      procedures are applied to determine if SAR measurements are required for the remaining OFDM transmission
      configurations. In general, the number of test channels that require SAR measurement is minimized based on maximum
      output power measured for the test sample(s).
3.    For OFDM transmission configurations in the 2.4 GHz and 5 GHz bands, When the same maximum power is specified for
      multiple transmission modes in a frequency band, the largest channel bandwidth, lowest order modulation, lowest data rate
      and lowest order 802.11a/g/n/ac mode is used for SAR measurement, on the highest measured output power channel for
      each frequency band.
4.    DSSS and OFDM configurations are considered separately according to the required SAR procedures. SAR is measured
      in the initial test position using the 802.11 transmission mode configuration required by the DSSS procedure or initial test
      configuration and subsequent test configuration(s) according to the OFDM procedures.18 The initial test position
      procedure is described in the following:
      a. When the reported SAR of the initial test position is ≤ 0.4 W/kg, further SAR measurement is not required for the other
            test positions in that exposure configuration and 802.11 transmission mode combinations within the frequency band or
            aggregated band.
      b. When the reported SAR of the test position is > 0.4 W/kg, SAR is repeated for the 802.11 transmission mode
            configuration tested in the initial test position to measure the subsequent next closet/smallest test separation distance
            and maximum coupling test position on the highest maximum output power channel, until the report SAR is ≤ 0.8 W/kg
            or all required test position are tested.
      c. For all positions/configurations, when the reported SAR is > 0.8 W/kg, SAR is measured for these test
            positions/configurations on the subsequent next highest measured output power channel(s) until the reported SAR is
            ≤ 1.2 W/kg or all required channels are tested.

<2.4GHz WLAN>

                                                               Frequency        Average power          Tune-Up
                         Mode               Channel                                                                     Duty Cycle %
                                                                 (MHz)              (dBm)               Limit

                                                1                 2412               15.10               16.00
                    802.11b 1Mbps               6                 2437               17.00               17.50              100.00
                                               11                 2462               18.30               18.50
                                                1                 2412               15.50               16.50
 2.4GHz WLAN        802.11g 6Mbps               6                 2437               16.00               16.50              100.00
                                               11                 2462               16.10               16.50
                                                1                 2412               15.50               16.50
                     802.11n-HT20
                                                6                 2437               15.90               16.50              100.00
                         MCS0
                                               11                 2462               16.00               16.50
                                                3                 2422               14.50               15.50
                     802.11n-HT40
                                                6                 2437               15.00               15.50              100.00
                         MCS0
                                                9                 2452               14.60               15.50



TEL : 886-3-327-3456                                                                                          Page 18 of 20
FAX : 886-3-328-4978                                                                                    Issued Date : Aug. 01, 2019
Form version: 181113


seorron ias._FCC SAR TEST REPORT                                                                                       Report No. : FA971705
     11. SAR Test Results
     General Note:
     1.      Per KDB 447498 DO1v06, the reported SAR is the measured SAR value adjusted for maximum tune—up tolerance.
           a. Tune—up scaling Factor = tune—up limit power (mW) / EUT RF power (mW), where tune—up limit is the maximum rated
               power among all production units.
           b. For SAR testing of WLAN signal with non—100% duty cycle, the measured SAR is scaled—up by the duty cycle scaling
               factor which is equal to "1/(duty cycle)"
           c.. For WLAN: Reported SAR(W/kg)= Measured SAR(W/kg)* Duty Cycle scaling factor * Tune—up scaling factor
     2.     Per KDB 447498 DO1v06, for each exposure position, testing of other required channels within the operating mode of a
            frequency band is not required when the reported 1—g or 10—g SAR for the mid—band or highest output power channel is:
             *    <0.8 Wikg or 2.0 Wikg, for 1—g or 10—g respectively, when the transmission band is < 100 MHz
             >    <0.6 Wikg or 1.5 Wikg, for 1—g or 10—g respectively, when the transmission band is between 100 MHz and 200 MHz
             >    <0.4 Wikg or 1.0 Wikg, for 1—g or 10—g respectively, when the transmission band is > 200 MHz
     3.     Per KDB 865664 DO1v01r04, for each frequency band, repeated SAR measurement is required only when the measured
            SAR is 20.8W/kq.

     WLAN Note:
     1. Per KDB 248227 DQ1v02r02, for 2.4GHz 802.11g/n SAR testing is not required when the highest reported SAR for DSSS is
            adjusted by the ratio of OFDM to DSSS specified maximum output power and the adjusted SAR is < 1.2 Wkg
     2.     When the reported SAR of the test position is > 0.4 W/kg, SAR is repeated for the 802.11 transmission mode configuration
            tested in the initial test position to measure the subsequent next closet‘smallest test separation distance and maximum
            coupling test position on the highest maximum output power channel, until the report SAR is < 0.8 W/kg or all required test
            position are tested.
     3.     For all positions / configurations, when the reported SAR is > 0.8 W/kg, SAR is measured for these test positions /
            configurations on the subsequent next highest measured output power channel(s) until the reported SAR is S 1.2 W/kg or all
            required channels are tested.
     4.     During SAR testing the WLAN transmission was verified using a spectrum analyzer.


     11.1 Head SAR
     <WLAN SAR>
                                                                             Average Tune—Up Tune—up Duty guctlye Power Measured Reported
                                             Test       Gap ch. uce
                                                                              eolEuiss                      t      Drift 1g SAR 1g SAR
                                            LeECo MR    Cc          (MHz)                                 EeeIU)
                                                                             (dBm)    (dBm)   Factor   %     "CR,00 (dB)          (Wikg)   (Wkg)
01        |WLANZ.4GHz 802.11b 1Mbps     Front Face |Omm| 11         2462      18.30   18.50   1.047    100   1.000     —0.11      0.053    0.055
          WLANZ.4GHz| 802.11b 1Mbps     Front Face |Omm|        1   [2412]    15.10   16.00   1.230    100   1.000     0.06       0.028    0.034
          WLANZ.4GHz| 802.11b 1Mbps     Front Face |Omm|        6   |2437|    17.00   17.50   1.122    100   1.000     0.14       0.041    0.046


     11.2 Hand SAR
     <WLAN SAR>




     WLANZ.4GHz         802.11b 1Mbps       |Bottom Face|Omm                      .       .      .             1.000     0.11
     WLANZ.4GHz         802.11b 1Mbps         Edge 1     Omm                      .       .      .             1.000     —0.04
     WLANZ.4GHz         802.11b 1Mbps         Edge 2|    Omm                      .       .      .             1.000     0.05
     WLANZ.4GHz         802.11b 1Mbps         Edge 2     |Omm                     .       —      —             1.000     |—0.19
     WLANZ.4GHz         802.11b 1Mbps         Edge 2     |Omm                     .       .      —             1.000     |—0.09
     WLANZ.4GHz         802.11b 1Mbps         Edge 3     |Omm                     .       .      —             1.000       0
     WLANZ.4GHz         802.11b 1Mbps         Edge 4     |Omm                     .       .      .             1.000     0.19

              Test Engineer :       Nick Yu
     TEL : 886—3—327—3456                                                                                           Page 19 of 20
     FAX : 886—3—328—4978                                                                                     Issued Date : Aug. 01, 2019
     Form version: 181113


         FCC SAR TEST REPORT                                                                   Report No. : FA971705
12. Uncertainty Assessment
    Per KDB 865664 D01 SAR measurement 100MHz to 6GHz, when the highest measured 1-g SAR within a frequency
    band is < 1.5 W/kg and the measured 10-g SAR within a frequency band is < 3.75 W/kg. The expanded SAR
    measurement uncertainty must be ≤ 30%, for a confidence interval of k = 2. If these conditions are met, extensive
    SAR measurement uncertainty analysis described in IEEE Std 1528-2013 is not required in SAR reports submitted
    for equipment approval. For this device, the highest measured 1-g SAR is less 1.5W/kg and highest measured 10-g
    SAR is less 3.75W/kg. Therefore, the measurement uncertainty table is not required in this report.



13. References

       [1]   FCC 47 CFR Part 2 “Frequency Allocations and Radio Treaty Matters; General Rules and
             Regulations”
       [2]   ANSI/IEEE Std. C95.1-1992, “IEEE Standard for Safety Levels with Respect to Human Exposure
             to Radio Frequency Electromagnetic Fields, 3 kHz to 300 GHz”, September 1992
       [3]   IEEE Std. 1528-2013, “IEEE Recommended Practice for Determining the Peak Spatial-Average
             Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
             Measurement Techniques”, Sep 2013
       [4]   SPEAG DASY System Handbook
       [5]   FCC KDB 248227 D01 v02r02, “SAR Guidance for IEEE 802.11 (WiFi) Transmitters”, Oct 2015.
       [6]   FCC KDB 447498 D01 v06, “Mobile and Portable Device RF Exposure Procedures and
             Equipment Authorization Policies”, Oct 2015
       [7]   FCC KDB 865664 D01 v01r04, "SAR Measurement Requirements for 100 MHz to 6 GHz", Aug
             2015.
       [8]   FCC KDB 865664 D02 v01r02, “RF Exposure Compliance Reporting and Documentation
             Considerations” Oct 2015.




TEL : 886-3-327-3456                                                                          Page 20 of 20
FAX : 886-3-328-4978                                                                    Issued Date : Aug. 01, 2019
Form version: 181113



Document Created: 2019-08-07 10:41:47
Document Modified: 2019-08-07 10:41:47

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC