Operation Description

FCC ID: S9NMHBGW1

Operational Description

Download: PDF
FCCID_1733239

                                                                             SPMHBGW1

                                                      Body Gateway Electronic patch
                                                                                          Data brief 1.1
Features

      Single Lead AFE for ECG with self calibration
      circuitry
      Bio-impedance
      STM32F M3 Cortex ARM cortex
      BT radio class 2: STLC 2584
      Digital 3 axis Accelerometer: LIS 331 DLH
      External parallel NAND FLASH 2Gbit
      Battery Level sensing
      Rechargeable battery LiPo 380 mAh
      3 LEDs
      Mechanical ON/OFF button




Functions                                              Applications
  •     Heart rate detection                              •   Elderly people home monitoring
  •     One lead ECG recording and                        •   Chronic cardiac disease monitoring
        transmission                                      •   Event monitoring
  •     Physical activity estimation                      •   Single lead holter
  •     Breathing rate measurement
  •     Body position detection




                                                                                                   1/12


                                                                                                                                                       SPMHBGW1




Contents
1      Body Gateway device description ..................................................................................................................... 3
2      Body Gateway device architecture .................................................................................................................... 4
    2.1     Hardware: ................................................................................................................................................... 4
    2.2     Embedded SW Framework: ........................................................................................................................ 4
3      Embedded SW Framework ............................................................................................................................... 5
    3.1     Connectivity: ............................................................................................................................................... 5
    3.2     Embedded Data Aggregator........................................................................................................................ 5
       3.2.1        ECG: extraction of .............................................................................................................................. 5
       3.2.2        Bio-impedance: extraction of .............................................................................................................. 5
       3.2.3        Accelerometer: extraction of ............................................................................................................... 5
4      Operating modes: ............................................................................................................................................. 6
5      Physiological parameters & personalization features ........................................................................................ 7
    5.1     Parameters: ................................................................................................................................................ 7
    5.2     Personalization: .......................................................................................................................................... 7
6      Body Gateway mechanical dimension .............................................................................................................. 8
    6.1     Body Gateway device ................................................................................................................................. 8
7      Power consumption......................................................................................................................................... 10
    7.1     Usage model example 1: streaming mode ................................................................................................ 10
    7.2     Usage model example 2: monitoring mode ............................................................................................... 10
8      ECG EC38 signal snapshot ............................................................................................................................ 11
9      Regulatory Compliance ................................................................................................................................... 12
    Safety and EMI Approval, ................................................................................................................................... 12
10          Revision history ......................................................................................................................................... 12




2/12


SPMHBGW1

RoHS compliance
ST modules are RoHS compliant and comply with ECOPACK® norms.



 1 Body Gateway device description

The Body Gateway (BGW) Devices is a wearable electronic, battery operated device that is worn on the
chest for the acquisition, recording and transmission of physiological parameters to external devices which
can analyze or forward the data to additional storage elements or system.

The BGW device is also capable to record symptomatic and asymptomatic events and is indicated for
ambulatory monitoring of non lethal cardiac arrhythmias.

Additionally, resident in the Body Gateway device is a heart rate, respiration rate and activity level
calculation algorithm, which allows the system to manage information messages from/to the Server
according to specific settings defined by the physicians/operators.

The device is a part of a Multi-parameter Analysis System, Body Gateway System (BGW system) and
communicates via a BT radio link with the external device. Specification of BGW System is beyond the
scope of this document.

At its heart is a 32-bit ARM Cortex microcontroller with 768 kByte Flash, chosen for its flexible architecture
and low power processing capability. Bluetooth radio was selected for connectivity because its availability in
most commercial mobile phones ensures proper coverage and patient access.




                                                                                                         3/12


                                                                                              SPMHBGW1




 2 Body Gateway device architecture


2.1 Hardware:
The hardware architecture for the BGW device is shown here below.




2.2 Embedded SW Framework:
The embedded SW Frameworks includes a RT OS, board support package, proprietary data aggregation
algorithms, embedded memory management system, a BT stack and proprietary data exchange protocol.
This SW is modular and provides an application interface to allow integration in customer specific
environment including the possibility to plug in proprietary algorithms for physiological data elaboration.




4/12


SPMHBGW1




3 Embedded SW Framework


3.1 Connectivity:

              Bluetooth Radio class 2
              BT supported profiles: SPP
              Point to point link to a PC or to any BT enabled device, supporting SPP service, both as a
              client and as a server
              Proprietary data exchange protocol on top of SPP designed to optimize power
              consumption minimizing packets overhead.



3.2 Embedded Data Aggregator


3.2.1 ECG: extraction of

              · Heart rate
              · RR peak variability
              · Heart Rate variability
              · Electrode detachment alert


3.2.2 Bio-impedance: extraction of

              · Breathing rate


3.2.3 Accelerometer: extraction of

              · Activity level index
              · Body posture


      Baselines and thresholds are programmable for 3.2.1., 3.2.2., 3.2.3. with hysteresis to avoid false
      alerts.




                                                                                                      5/12


                                                                                                     SPMHBGW1




 4 Operating modes:

Platform can operate in 4 modes:

       a)   Power down mode : BGW device is completely switched off

       b)   Idle mode: the BGW device is visible and connectable to the remote unit through the BT link

       c)   Streaming Mode: is when the Body Gateway is powered ON and sampling the data as specified in
            the Configuration Settings, sending them directly to the External Device with periodicity specified by
            the last configuration commands received from the External Device.

       d)   Monitoring mode: is when the Body Gateway is powered ON, gathering and storing in the internal
            memory data at the frequency specified in the Configuration Settings, and sending the data as
            requested by the Associated Device.




6/12


SPMHBGW1




 5 Physiological parameters & personalization features


5.1 Parameters:
             Heart rate
             Heart rate reliability
             RR interval variability
             Breathing rate
             Activity level
             Body position

        Other available data:

             ECG raw
             Z0 bio impedance
             Electrode reliability
             Battery level



5.2 Personalization:
Available data and parameters depends on the selected operative mode


Streaming mode
                                                       Monitoring mode

             Raw Data:                                             Raw Data
                     ·   ECG (128 or 256 Hz)                           o ECG (128 or 256 Hz)
                     ·   3-axis      accelerometer                 Derived values
                         (32Hz or 64Hz or 50 Hz)                       o Activity level
                    ·    dZ bioimpedance (32                           o Body position values
                         Hz)                                           o Breathing rate values
                    ·    Z0 bioimpedance (32Hz)                        o Heart rate values
             Derived values:                                           o Heart rate reliability
                    ·    Heart rate values,                            o    RR interval variability
                    ·    Heart rate reliability,                       o    Battery level
                    ·    Breathing rate values,
                    ·    Activity level,
                    ·    RR interval variability,
                    ·    Body position values,
                    ·    Battery level




                                                                                                      7/12


                                                                                    SPMHBGW1


Signals and Parameters                Streaming Mode              Monitoring mode

ECG raw                               128/256 Hz                  128/256 Hz

Heart Rate                            1 each 10-15-30-60 secs     1 each 10-15-30-60 secs

Heart Rate Reliability                1 each 10 secs              1 each 10 secs

R-R Variability                       1 each 10 secs              1 each 10 secs



XYZ raw                               50.0 Hz                     -

Activity level                        1 each 5-10-15-30-60 secs   1 each 5-10-15-30-60 secs



BIOIMP-Z0 raw                         32 Hz                       -

BIOIMP-DZ raw                         32 Hz                       -

Breathing Rate                        1 each 15-30-60 secs        1 each 15-30-60 secs



Battery Level                         1 each 10-15-30-60 secs     1 each 10-15-30-60 secs




 6 Body Gateway mechanical dimension


6.1 Body Gateway device
       BGW device is water resistant IPx4

       Mechanical dimension:

           • Weight: 35g
           • Size: 59 x 50 x 16.4mm




8/12


SPMHBGW1




           9/12


                                                                                                SPMHBGW1



 7 Power consumption


7.1 Usage model example 1: streaming mode
   Continuous data acquisition, processing and real time transmission (ECG raw signal, Bio-impedance
   signal, accelerometer data raw signal, HR, BR, Activity level, RR variability, HR reliability and battery
   level)

   Using a 3.7V 380 mAh rechargeable battery, battery will last in 16Hours




7.2 Usage model example 2: monitoring mode
   Continuous data acquisition, processing and storage (ECG raw signal, Bio-impedance signal, HR, BR,
   Activity level, RR variability, HR reliability and battery level), wireless uploading of all physiological
   parameters plus 5 minutes ECG and bio-impedance raw data per hour

   Using a 3.7V 380 mAh rechargeable battery, battery will last in 40 hours




10/12


SPMHBGW1



8 ECG EC38 signal snapshot




                             11/12


                                                                                       SPMHBGW1

 9 Regulatory Compliance

FCC
FCC ID: S9NMHBGW1

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.


CE
Safety and EMI Approval,
Compliance with
IEC 60601-1
IEC 60601-1-1
IEC 60601-1-2
IEC 60601-1-4
IEC 60601-1-6
IEC 60601-2-47
IEC 60601-2-49
IEC 60529
EC38
EC57
UL 60601-1
CAN/CSA C22.2




 10 Revision history

            Date       Revision    Description of changes
         10/07/2011       1.0      First issue.




12/12


SPMHBGW1




                                              Please Read Carefully:

Information in this document is provided solely in connection with ST products. STMicroelectronics NV and its
subsidiaries (“ST”) reserve the right to make changes, corrections, modifications or improvements, to this
document, and the products and services described herein at any time, without notice.

All ST products are sold pursuant to ST’s terms and conditions of sale.

Purchasers are solely responsible for choice, selection and use of the ST products and services described herein,
and ST assumes no liability whatsoever relating to the choice, selection or use of the ST products and services
described herein.

No license, express or implied, by estoppels or otherwise, to any intellectual property rights is granted under this
document. If any part of this document refers to any third party products or services it shall not be deemed a
license grant by ST for the use of such a third party products or services, or any intellectual property contained
therein or considered as a warranty covering the use in any manner whatsoever of such third party products or
services or any intellectual property contained therein.

UNLESS OTHERWISE SET FORTH IN ST’S TERMS AND CONDITIONS OF SALE ST DISCLAIMS ANY
EXPRESS OR IMPLIED WARRANTY WITH RESPECT TO THE USE AND/OR SALE OF ST PRODUCTS
INCLUDING WITHOUT LIMITATION IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A
PARTICULAR PURPOSE (AND THEIR EQUIVALENTS UNDER THE LAWS OF ANY JURISDICTION), OR
INFRINGEMENT OF ANY PATENT, COPYRIGHT OR OTHER INTELLECTUAL PROPERTY RIGHT.

UNLESS EXPRESSLY APPROVED IN WRITING BY AN AUTHORIZED ST REPRESENTATIVE, ST
PRODUCTS ARE NOT RECOMMENDED, AUTHORIZED OR WARRANTED FOR USE IN MILITARY, AIR
CRAFT, SPACE, LIFE SAVING, OR LIFE SUSTAINING APPLICATIONS, NOR IN PRODUCTS OR SYSTEMS
WHERE FAILURE OR MALFUNCTION MAY RESULT IN PERSONAL INJURY, DEATH, OR SEVERE
PROPERTY OR ENVIRONMENTAL DAMAGE. ST PRODUCTS WHICH ARE NOT SPECIFIED AS
“AUTOMOTIVE GRADE” MAY ONLY BE USED IN AUTOMOTIVE APPLICATIONS AT USER’S OWN RISK.


Resale of ST products with provisions different from the statements and/or technical features set forth in this
document shall immediately void any warranty granted by ST for the ST product or service described herein and
shall not create or extend in any manner whatsoever, any liability of ST.


              St and the ST logo are trademarks or registered trademarks of ST in various countries.

            Information in this document supersedes and replaces all information previously supplied.

The ST logo is a registered trademark of STMicroelectronics. All other names are the property of their respective
                                                   owners.

                                 © 2011 STMicroelectronics - All rights reserved

                                     STMicroelectronics group of companies

Australia – Belgium - Brazil – Canada - China – Czech Republic - Finland - France - Germany - Hong Kong - India –
Israel - Italy - Japan - Malaysia - Malta – Morocco - Singapore - Spain - Sweden - Switzerland - United Kingdom –
United States of America

                                                http://www.st.com




                                                                                                             13/12



Document Created: 2012-02-28 09:44:02
Document Modified: 2012-02-28 09:44:02