SAR Report

FCC ID: RPNE5KTX10046340

RF Exposure Info

Download: PDF
FCCID_391858

                                                             TTI-P-G 158




                                                                                                                                         Report
                                                             Dosimetric Assessment of the Portable Device
                                                                   MSA EVOLUTION 5000 Transmitter
                                                                          (FCC ID: RPNE5KTX1004634)
                                                                     According to the FCC Requirements
                                                                                                                         January 22, 2004
                                                                                                                            IMST GmbH
                                                                                                              Carl-Friedrich-Gauß-Str. 2
                                                                                                                   D-47475 Kamp-Lintfort



                                                                                                                                   Customer
                                                                                                                                 MSA AUER
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                                                           Thielemannstraße 1
                                                                                                                                     12059-Berlin
                                                                                                                                        Germany



                                                                                                       The test results only relate to the items tested.
                                                                               This report shall not be reproduced except in full without the written
                                                                                                                   approval of the testing laboratory.
                                                       DASY Test-Report, Phoenix_6575_367                                                 Page 1 of 28


                                                       DASY Test-Report, Phoenix_6575_367                                                                          Page 2 of 28

                                                       Executive Summary
                                                       The EVOLUTION 5000 Transmitter is a Wireless Video Transmission System (Portable Device)
                                                       from MSA operating in the 2450 MHz frequency range. The device has an patch antenna and
                                                       was tested in combination with the Evolution 5000 Thermal Imaging System.
                                                       The examinations have been carried out with the dosimetric assessment system „DASY4“.
                                                       The measurements were made according to the Supplement C to OET Bulletin 65 of the Federal
                                                       Communications Commission (FCC) Guidelines [FCC 2001] for evaluating compliance of
                                                       mobile and portable devices with FCC limits for human exposure (general population) to
                                                       radiofrequency emissions. All measurements have been performed in accordance to the
                                                       recommendations given by SPEAG.
                                                       The MSA EVOLUTION 5000 Transmitter (FCC ID: RPNE5KTX1004634) tested in
                                                       combination with the Evolution 5000 Thermal Imaging System is in compliance with the
                                                       Federal Communications Commission (FCC) Guidelines [FCC 2001] for uncontrolled
                                                       exposure.




                                                       prepared by: ...........................................   reviewed by:   ..............................................

                                                                          André van den Bosch                                    Dipl.-Ing. Christoph Hennes
                                                                          test engineer                                          quality assurance engineer




                                                                                                                                                      IMST GmbH
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                                                                       Carl-Friedrich-Gauß-Straße 2
                                                                                                                                            D-47475 Kamp-Lintfort

                                                                                                                                                Tel. +49- 2842-981 373
                                                                                                                                                Fax +49- 2842-981 399
                                                                                                                                                email: hennes@imst.de


                                                       DASY Test-Report, Phoenix_6575_367                                                                                                                     Page 3 of 28

                                                       Table of Contents

                                                       1     SUBJECT OF INVESTIGATION ................................................................................................................... 4

                                                       2     THE IEEE STANDARD C95.1 AND THE FCC EXPOSURE CRITERIA................................................. 4

                                                           2.1       DISTINCTION BETWEEN EXPOSED POPULATION, DURATION OF EXPOSURE AND FREQUENCIES ...................... 4

                                                           2.2       DISTINCTION BETWEEN MAXIMUM PERMISSIBLE EXPOSURE AND SAR LIMITS ................................................ 5

                                                           2.3       SAR LIMIT ................................................................................................................................................... 5

                                                       3     THE FCC MEASUREMENT PROCEDURE ................................................................................................ 6

                                                           3.1       GENERAL REQUIREMENTS ............................................................................................................................ 6

                                                           3.2       DEVICE OPERATING NEXT TO A PERSON’S EAR ............................................................................................. 6

                                                           3.3       BODY-WORN AND OTHER CONFIGURATIONS ................................................................................................. 9

                                                       4     THE MEASUREMENT SYSTEM ................................................................................................................ 10

                                                           4.1       PHANTOM .................................................................................................................................................. 12

                                                           4.2       PROBE ....................................................................................................................................................... 12

                                                           4.3       MEASUREMENT PROCEDURE ...................................................................................................................... 13

                                                           4.4       UNCERTAINTY ASSESSMENT ........................................................................................................................ 14

                                                       5     SAR RESULTS................................................................................................................................................ 15

                                                       6     EVALUATION ................................................................................................................................................ 16

                                                       7     APPENDIX ...................................................................................................................................................... 17

                                                           7.1       ADMINISTRATIVE DATA .............................................................................................................................. 17

                                                           7.2       DEVICE UNDER TEST AND TEST CONDITIONS .............................................................................................. 17

                                                           7.3       TISSUE RECIPES ......................................................................................................................................... 18

                                                           7.4       MATERIAL PARAMETERS ............................................................................................................................. 18

                                                           7.5       SIMPLIFIED PERFORMANCE CHECKING ...................................................................................................... 18
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                           7.6       ENVIRONMENT ........................................................................................................................................... 21

                                                           7.7       TEST EQUIPMENT....................................................................................................................................... 21

                                                           7.8       CERTIFICATES OF CONFORMITY .................................................................................................................. 22

                                                           7.9       PICTURES OF THE DEVICE UNDER TEST ....................................................................................................... 23

                                                           7.10      TEST POSITIONS FOR THE DEVICE UNDER TEST ........................................................................................... 24

                                                           7.11      PICTURES TO DEMONSTRATE THE REQUIRED LIQUID DEPTH ........................................................................ 28

                                                       8     REFERENCES ................................................................................................................................................ 28


                                                       DASY Test-Report, Phoenix_6575_367                                                   Page 4 of 28

                                                       1 Subject of Investigation
                                                       The EVOLUTION 5000 Transmitter is a Wireless Video Transmission System (Portable Device)
                                                       from MSA operating in the 2450 MHz frequency range. The device has an patch antenna.




                                                                                    Fig. 1:Picture of the device under test.

                                                       The device was tested in combination with the Evolution 5000 Thermal Imaging System. The
                                                       examinations have been carried out with the dosimetric assessment system „DASY4“ described
                                                       below.


                                                       2 The IEEE Standard C95.1 and the FCC Exposure Criteria
                                                       In the USA the recent FCC exposure criteria [FCC 2001] are based upon the IEEE Standard
                                                       C95.1 [IEEE 1999]. The IEEE standard C95.1 sets limits for human exposure to radio frequency
                                                       electromagnetic fields in the frequency range 3 kHz to 300 GHz.

                                                       2.1 Distinction Between Exposed Population, Duration of Exposure and Frequencies
                                                       The American Standard [IEEE 1999] distinguishes between controlled and uncontrolled
                                                       environment. Controlled environments are locations where there is exposure that may be
                                                       incurred by persons who are aware of the potential for exposure as a concomitant of employment
                                                       or by other cognizant persons. Uncontrolled environments are locations where there is the
                                                       exposure of individuals who have no knowledge or control of their exposure. The exposures may
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                       occur in living quarters or workplaces. For exposure in controlled environments higher field
                                                       strengths are admissible. In addition the duration of exposure is considered.
                                                       Due to the influence of frequency on important parameters, as the penetration depth of the
                                                       electromagnetic fields into the human body and the absorption capability of different tissues, the
                                                       limits in general vary with frequency.


                                                       DASY Test-Report, Phoenix_6575_367                                                      Page 5 of 28

                                                       2.2 Distinction between Maximum Permissible Exposure and SAR Limits
                                                       The biological relevant parameter describing the effects of electromagnetic fields in the
                                                       frequency range of interest is the specific absorption rate SAR (dimension: power/mass). It is a
                                                       measure of the power absorbed per unit mass. The SAR may be spatially averaged over the total
                                                       mass of an exposed body or its parts. The SAR is calculated from the r.m.s. electric field strength
                                                       E inside the human body, the conductivity σ and the mass density ρ of the biological tissue:
                                                                                             E2        ∂T
                                                                                   SAR = σ        =c                 .                                   (1)
                                                                                             ρ         ∂t   t →0 +


                                                       The specific absorption rate describes the initial rate of temperature rise ∂ T / ∂ t as a function of
                                                       the specific heat capacity c of the tissue. A limitation of the specific absorption rate prevents an
                                                       excessive heating of the human body by electromagnetic energy.

                                                       As it is sometimes difficult to determine the SAR directly by measurement (e.g. whole body
                                                       averaged SAR), the standard specifies more readily measurable maximum permissible exposures
                                                       in terms of external electric E and magnetic field strength H and power density S, derived from
                                                       the SAR limits. The limits for E, H and S have been fixed so that even under worst case
                                                       conditions, the limits for the specific absorption rate SAR are not exceeded.
                                                       For the relevant frequency range the maximum permissible exposure may be exceeded if the
                                                       exposure can be shown by appropriate techniques to produce SAR values below the
                                                       corresponding limits.

                                                       2.3 SAR Limit
                                                       In this report the comparison between the American exposure limits and the measured data is
                                                       made using the spatial peak SAR; the power level of the device under test guarantees that the
                                                       whole body averaged SAR is not exceeded.
                                                       Having in mind a worst case consideration, the SAR limit is valid for uncontrolled environment
                                                       and mobile respectively portable transmitters. According to Table 1 the SAR values have to be
                                                       averaged over a mass of 1 g (SAR1g) with the shape of a cube.

                                                                        Standard                               Status    SAR limit [W/kg]

                                                                        IEEE C95.1                            In force         1.6
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                    Table 1: Relevant spatial peak SAR limit averaged over a mass of 1 g.


                                                       DASY Test-Report, Phoenix_6575_367                                                     Page 6 of 28

                                                       3 The FCC Measurement Procedure
                                                       The Federal Communications Commission (FCC) has published a report and order on the 1st of
                                                       August 1996 [FCC 1996], which requires routine dosimetric assessment of mobile telecom-
                                                       munications devices, either by laboratory measurement techniques or by computational
                                                       modeling, prior to equipment authorization or use. In 2001 the Commission’s Office of
                                                       Engineering and Technology has released Edition 01-01 of Supplement C to OET Bulletin 65.
                                                       This revised edition, which replaces Edition 97-01, provides additional guidance and information
                                                       for evaluating compliance of mobile and portable devices with FCC limits for human exposure
                                                       to radiofrequency emissions [FCC 2001].

                                                       3.1 General Requirements
                                                       The test shall be performed in a laboratory with an environment which avoids influence on SAR
                                                       measurements by ambient EM sources and any reflection from the environment itself. The
                                                       ambient temperature shall be in the range of 20°C to 26°C and 30-70% humidity.

                                                       3.2 Device Operating Next to a Person’s Ear

                                                       3.2.1 Phantom Requirements
                                                       The phantom is a simplified representation of the human anatomy and comprised of material
                                                       with electrical properties similar to the corresponding tissues. The physical characteristics of the
                                                       phantom model shall resemble the head and the neck of a user since the shape is a dominant
                                                       parameter for exposure.

                                                       3.2.2 Test Positions
                                                       As it cannot be expected that the user will hold the mobile phone exactly in one well defined
                                                       position, different operational conditions shall be tested. The Supplement C to OET Bulletin 65
                                                       requires two test positions. For an exact description helpful geometrical definitions are
                                                       introduced and shown in Fig. 2 - 3.
                                                       There are two imaginary lines on the mobile, the vertical centerline and the horizontal line. The
                                                       vertical centerline passes through two points on the front side of the handset: the midpoint of the
                                                       width wt of the handset at the level of the acoustic output (point A on Fig. 2), and the midpoint of
                                                       the width wb of the bottom of the handset (point B). The horizontal line is perpendicular to the
                                                       vertical centerline and passes through the center of the acoustic output (see Fig. 2). The two lines
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                       intersect at point A.


                                                       DASY Test-Report, Phoenix_6575_367                                                  Page 7 of 28




                                                                           Fig. 2: Handset vertical and horizontal reference lines.




                                                                                      Fig. 3: Phantom reference points.
                                                       According to Fig. 3 the human head position is given by means of the following three reference
                                                       points: auditory canal opening of both ears (RE and LE) and the center of the closed mouth (M).
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                       The ear reference points are 15-17 mm above the entrance to the ear canal along the BM line
                                                       (back-mouth), as shown in Fig. 3. The plane passing through the two ear canals and M is defined
                                                       as the reference plane. The line NF (Neck-Front) perpendicular to the reference plane and
                                                       passing through the RE (or LE) is called the reference pivoting line. Line BM is perpendicular to
                                                       the NF line. With this definitions the test positions are given by


                                                       DASY Test-Report, Phoenix_6575_367                                                      Page 8 of 28

                                                       •   Cheek position (see Fig. 4):
                                                       Position the handset close to the surface of the phantom such that point A is on the (virtual)
                                                       extension of the line passing through points RE and LE on the phantom (see Fig. 3), such that the
                                                       plane defined by the vertical center line and the horizontal line of the phone is approximately
                                                       parallel to the sagittal plane of the phantom. Translate the handset towards the phantom along the
                                                       line passing through RE and LE until the handset touches the ear. While maintaining the handset
                                                       in this plane, rotate it around the LE-RE line until the vertical centerline is in the plane normal to
                                                       MB-NF including the line MB (called the reference plane). Rotate the phone around the vertical
                                                       centerline until the phone (horizontal line) is symmetrical with respect to the line NF. While
                                                       maintaining the vertical centerline in the reference plane, keeping point A on the line passing
                                                       through RE and LE, and maintaining the phone contact with the ear, rotate the handset about the
                                                       line NF until any point on the handset is in contact with a phantom point below the ear.




                                                                                           Fig. 4: The cheek position.
                                                       •   Tilted position (see Fig. 5):
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                           Fig. 5: The tilted position.


                                                       DASY Test-Report, Phoenix_6575_367                                                     Page 9 of 28

                                                       While maintaining the orientation of the phone retract the phone parallel to the reference plane
                                                       far enough to enable a rotation of the phone by 15°. Rotate the phone around the horizontal line
                                                       by 15°. While maintaining the orientation of the phone, move the phone parallel to the reference
                                                       plane until any part of the phone touches the head. In this position, point A will be located on the
                                                       line RE-LE.

                                                       3.2.3 Test to be Performed
                                                       The SAR test shall be performed with both phone positions described above, on the left and right
                                                       side of the phantom. The device shall be measured for all modes operating when the device is
                                                       next to the ear, even if the different modes operate in the same frequency band.
                                                       For devices with retractable antenna the SAR test shall be performed with the antenna fully
                                                       extended and fully retracted. Other factors that may affect the exposure shall also be tested. For
                                                       example, optional antennas or optional battery packs which may significantly change the
                                                       volume, lengths, flip open/closed, etc. of the device, or any other accessories which might have
                                                       the potential to considerably increase the peak spatial-average SAR value.
                                                       The SAR test shall be performed at the high, middle and low frequency channels of each
                                                       operating mode. If the SAR measured at the middle channel for each test configuration is at least
                                                       3.0 dB lower than the SAR limit, testing at the high and low channels is optional.

                                                       3.3 Body-worn and Other Configurations

                                                       3.3.1 Phantom Requirements
                                                       For body-worn and other configurations a flat phantom shall be used which is comprised of
                                                       material with electrical properties similar to the corresponding tissues.

                                                       3.3.2 Test Position
                                                       The body-worn configurations shall be tested with the supplied accessories (belt-clips, holsters,
                                                       etc.) attached to the device in normal use configuration. Devices with a headset output shall be
                                                       tested with a connected headset.

                                                       3.3.3 Test to be Performed
                                                       For purpose of determining test requirements, accessories may be divided into two categories:
                                                       those that do not contain metallic components and those that do. For multiple accessories that do
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                       not contain metallic components, the device may be tested only with that accessory which
                                                       provides the closest spacing to the body.
                                                       For multiple accessories that contain metallic components, the device must be tested with each
                                                       accessory that contains a unique metallic component. If multiple accessories share an identical
                                                       metallic component, only the accessory that provides the closest spacing to the body must be
                                                       tested.
                                                       If the manufacturer provides none body-worn accessories a separation distance of 1.5 cm
                                                       between the back of the device and the flat phantom is recommended. Other separation distances


                                                       DASY Test-Report, Phoenix_6575_367                                                  Page 10 of 28

                                                       may be used, but they shall not exceed 2.5 cm. In these cases, the device may use body-worn
                                                       accessories that provide a separation distance greater than that tested for the device provided
                                                       however that the accessory contains no metallic components.
                                                       For devices with retractable antenna the SAR test shall be performed with the antenna fully
                                                       extended and fully retracted. Other factors that may affect the exposure shall also be tested. For
                                                       example, optional antennas or optional battery packs which may significantly change the
                                                       volume, lengths, flip open/closed, etc. of the device, or any other accessories which might have
                                                       the potential to considerably increase the peak spatial-average SAR value.
                                                       The SAR test shall be performed at the high, middle and low frequency channels of each
                                                       operating mode. If the SAR measured at the middle channel for each test configuration is at least
                                                       3.0 dB lower than the SAR limit, testing at the high and low channels is optional.


                                                       4 The Measurement System
                                                       DASY is an abbreviation of „Dosimetric Assessment System“ and describes a system that is able
                                                       to determine the SAR distribution inside a phantom of a human being according to different
                                                       standards. The DASY4 system consists of the following items as shown in Fig: 6. Additional
                                                       Fig: 7 shows the equipment, similar to the installations in other laboratories [DASY 1995].


                                                          •   High precision robot with controller
                                                          •   Measurement server (for surveillance of the robot operation and signal filtering)
                                                          •   Data acquisition electronics DAE (for signal amplification and filtering)
                                                          •   Field probes calibrated for use in liquids
                                                          •   Electro-optical converter EOC (conversion from the optical into a digital signal)
                                                          •   Light beam (improving of the absolute probe positioning accuracy)
                                                          •   Two SAM phantoms filled with tissue simulating liquid
                                                          •   DASY4 software
                                                          •   SEMCAD
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH


                                                       DASY Test-Report, Phoenix_6575_367                                            Page 11 of 28




                                                                                Fig. 6: The DASY4 measurement system.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                       Fig. 7: The measurement set-up with two SAM phantoms containing tissue simulating liquid.


                                                       DASY Test-Report, Phoenix_6575_367                                                   Page 12 of 28

                                                       The mobile phone operating at the maximum power level is placed by a non metallic device
                                                       holder (delivered from Schmid & Partner) in the above described positions at a shell phantom of
                                                       a human being. The distribution of the electric field strength E is measured in the tissue simu-
                                                       lating liquid within the shell phantom. For this miniaturised field probes with high sensitivity and
                                                       low field disturbance are used. Afterwards the corresponding SAR values are calculated with the
                                                       known electrical conductivity σ and the mass density ρ of the tissue in the SEMCAD FDTD
                                                       software. The software is able to determine the averaged SAR values (averaging region 1 g or
                                                       10 g) for compliance testing.
                                                       The measurements are done by two scans: first a coarse scan determines the region of the
                                                       maximum SAR, afterwards the averaged SAR is measured in a second scan within the shape of a
                                                       cube. The measurement time takes about 20 minutes.

                                                       4.1 Phantom
                                                       For the measurements the Specific Anthropomorphic Mannequin (SAM Twin Phantom V4.0)
                                                       defined by the IEEE SCC-34/SC2 group and delivered by Schmid & Partner Engineering AG is
                                                       used. The phantom is a fibreglass shell integrated in a wooden table. The thickness of the
                                                       phantom amounts to 2 mm ± 0.2 mm. It enables the dosimetric evaluation of left and right hand
                                                       phone usage and includes an additional flat phantom part for the simplified performance check.
                                                       The phantom set-up includes a coverage (polyethylene), which prevents the evaporation of the
                                                       liquid. The details and the Certificate of conformity can be found in Fig. 10.

                                                       4.2 Probe
                                                       For the measurements the Dosimetric E-Field Probe ET3DV6 with following specifications is
                                                       used. It is manufactured and calibrated in accordance with FCC [FCC 2001] and IEEE
                                                       [IEEE 1528] recommendations annually by Schmid & Partner Engineering AG.
                                                          •   Dynamic range: 5µW/g to > 100mW/g
                                                          •   Tip diameter: 6.8 mm
                                                          •   Distance between probe tip and sensor center: 2.7 mm
                                                          •   Distance between sensor center and the inner phantom surface: 4 mm (repeatability better
                                                              than ± 1.0 mm)
                                                          •   Boundary effect compensation included within the SEMCAD software
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                          •   Probe linearity: ± 0.2 dB (30 MHz to 3 GHz)
                                                          •   Axial isotropy: ± 0.2 dB
                                                          •   Spherical isotropy: ± 0.4 dB
                                                          •   Calibration range: 835MHz/1900/2450 MHz for head & body simulating liquid
                                                          •   Angle between probe axis (evaluation axis) and surface normal line: less than 30°


                                                       DASY Test-Report, Phoenix_6575_367                                                Page 13 of 28

                                                       4.3 Measurement Procedure
                                                       The following steps are used for each test position:
                                                          •   Establish a carrier with the maximum output power by switching on the Thermal Imaging
                                                              System.
                                                          •   Measurement of the local E-field value at a fixed location (P1). This value serves as a
                                                              reference value for calculating a possible power drift.
                                                          •   Measurement of the SAR distribution with a grid spacing of 15 mm x 15 mm and a
                                                              constant distance to the inner surface of the phantom. Since the sensors can not directly
                                                              measure at the inner phantom surface, the values between the sensors and the inner
                                                              phantom surface are extrapolated. With this values the area of the maximum SAR is
                                                              calculated by a interpolation scheme (combination of a least-square fitted function and a
                                                              weighted average method). Additional all peaks within 2 dB of the maximum SAR are
                                                              searched.
                                                          •   Around this points, a cube of 30 mm x 30 mm x 30 mm is assessed by measuring
                                                              7 x 7 x 7 points whereby the first two measurement points are within the required 10 mm
                                                              of the surface. With these data, the peak spatial-average SAR value can be calculated
                                                              within the SEMCAD software.
                                                          •   The used extrapolation and interpolation routines are all based on the modified Quadratic
                                                              Shepard´s method [DASY4].
                                                          •   Repetition of the E-field measurement at the fixed location (P1) and repetition of the
                                                              whole procedure if the two results differ by more than ± 0.21dB.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH


                                                       DASY Test-Report, Phoenix_6575_367                                                     Page 14 of 28

                                                       4.4 Uncertainty Assessment
                                                       Table 2 includes the uncertainty budget suggested by the [IEEE 1528] and determined by
                                                       Schmid & Partner Engineering AG. The expanded uncertainty (K=2) is assessed to be ± 20.6%.

                                                                                             Uncertai    Probability                      Standard     vi² or
                                                       Error Sources                                                    Divisor    ci
                                                                                             nty Value   Distribution                    Uncertainty    veff

                                                       Measurement System
                                                       Probe calibration                     ± 4.8 %       Normal         1        1      ± 4.8 %       ∞
                                                       Axial isotropy                        ± 4.7 %     Rectangular      √3      0.7     ± 1.9 %       ∞

                                                       Hemispherical isotropy                ± 9.6 %     Rectangular      √3      0.7     ± 3.9 %       ∞

                                                       Boundary effects                      ± 1.0 %     Rectangular      √3       1      ± 0.6 %       ∞

                                                       Linearity                             ± 4.7 %     Rectangular      √3       1      ± 2.7 %       ∞

                                                       System detection limit                ± 1.0 %     Rectangular      √3       1      ± 0.6 %       ∞

                                                       Readout electronics                   ± 1.0 %       Normal         1        1      ± 1.0 %       ∞

                                                       Response time                         ± 0.8 %     Rectangular      √3       1      ± 0.5 %       ∞

                                                       Integration time                       ± 2.6%     Rectangular      √3       1      ± 1.5 %       ∞

                                                       RF ambient conditions                 ± 3.0 %     Rectangular      √3       1      ± 1.7 %       ∞

                                                       Probe positioner                      ± 0.4 %     Rectangular      √3       1      ± 0.2 %       ∞

                                                       Probe positioning                     ± 2.9 %     Rectangular      √3       1      ± 1.7 %       ∞

                                                       Algorithms for max SAR eval.          ± 1.0 %     Rectangular      √3       1      ± 0.6 %       ∞

                                                       Test Sample Related
                                                       Device positioning                    ± 2.9 %       Normal         1        1      ± 2.9 %      145
                                                       Device holder                         ± 3.6 %       Normal         1        1      ± 3.6 %        5
                                                       Power drift                           ± 5.0 %     Rectangular      √3       1      ± 2.9 %       ∞
                                                       Phantom and Set-up
                                                       Phantom uncertainty                   ± 4.0 %     Rectangular      √3       1      ± 2.3 %       ∞
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                       Liquid conductivity (target)          ± 5.0 %     Rectangular      √3      0.64    ± 1.8 %       ∞

                                                       Liquid conductivity (meas.)           ± 2.5 %       Normal         1       0.64    ± 1.6 %       ∞

                                                       Liquid permittivity (target)          ± 5.0 %     Rectangular      √3      0.6     ± 1.7 %       ∞

                                                       Liquid permittivity (meas.)           ± 2.5 %       Normal         1       0.6     ± 1.5 %       ∞

                                                       Combined Uncertainty                                                              ± 10.3 %

                                                                                      Table 2: Uncertainty budget of DASY4.


                                                       DASY Test-Report, Phoenix_6575_367                                                Page 15 of 28

                                                       5 SAR Results
                                                       The Tables below contain the measured SAR values averaged over a mass of 1 g.

                                                         Phantom                                SAR1g [W/kg] (Drift[dB])               Temperature
                                                       Configuration           Test      2458 MHz                 2474 MHz         Ambient      Liquid
                                                       (Liquid depth = 16.4   Position
                                                               cm)
                                                                                         26.5 dBm                  26.5 dBm           [° C]       [° C]



                                                                                 1       0.496 (0.1)              0.442 (0.15)     22.5/22.5 21.7/21.7
                                                              Flat               2       0.351 (-0.18)             0.368 (0.2)     22.6/22.7 21.7/21.7

                                                                                 3       0.401 (-0.04)                                22.8       21.7

                                                                                 4                                 0.177 (-0.1)       22.9       21.7

                                                       Table 3: Measured SAR results for the MSA EVOLUTION 5000 Transmitter.

                                                       The above mentioned power values are ERP values, they were measured on the same sample
                                                       which was prepared for the FCC approval. The values were delivered by Phoenix TEST-LAB
                                                       GmbH. The sample was carefully delivered by the manufacturer to the IMST GmbH to avoid
                                                       any damaging of the sample. After the SAR measurements the conducted power was measured
                                                       which is possible after open the transmitter housing. To avoid any damaging of the antenna
                                                       connection (the connector is not for customer use) conducted power was measured after the SAR
                                                       measurements and the values were within 5% of the values measured by Phoenix.
                                                       To control the output power stability during the SAR test the used DASY4 system calculates the
                                                       power drift by measuring the e-field at the same location at the beginning and at the end of the
                                                       measurement for each test position. These drift values can be found in the above tables labeled
                                                       as: (Drift[dB]). This ensures that the power drift during one measurement is within 5%.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH


                                                       DASY Test-Report, Phoenix_6575_367                                                        Page 16 of 28

                                                       6 Evaluation
                                                       In Fig. 8 the SAR results in body-worn configuration given in Table 3 are summarized and
                                                       compared to the limit.

                                                                                                           2458 MHz     2474 MHz

                                                                                    2
                                                                                                      FCC Guideline

                                                                                   1,5
                                                                   SAR 1g [W/kg]




                                                                                    1



                                                                                   0,5



                                                                                    0
                                                                                         Position 1       Position 2   Position 3   Position 4


                                                       Fig. 8: The measured body phantom SAR values for the MSA EVOLUTION 5000 Transmitter
                                                               in comparison to the FCC exposure limit.


                                                       The MSA EVOLUTION 5000 Transmitter (FCC ID: RPNE5KTX1004634) tested in
                                                       combination with the Evolution 5000 Thermal Imaging System is in compliance with the
                                                       Federal Communications Commission (FCC) Guidelines [FCC 2001] for uncontrolled
                                                       exposure.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH


                                                       DASY Test-Report, Phoenix_6575_367                                   Page 17 of 28

                                                       7 Appendix

                                                       7.1 Administrative Data
                                                       Date of validation:       2450 MHz, Body November 07, 2003
                                                       Date of measurement:      2450 MHz, Body: November 07, 2003
                                                       Data stored:              Phoenix_6575_367


                                                       7.2 Device under Test and Test Conditions
                                                       MTE:                     MSA EVOLUTION 5000 Transmitter, Production Line Unit;
                                                                                Evolution 5000 (Thermal Imaging System SN:A1-12009-008)
                                                       Date of receipt:         November 07, 2003
                                                       Equipment class:         Portable device
                                                       RF exposure environment: General Population/Uncontrolled
                                                       Power supply:            Internal Battery (Other batteries not available)
                                                       Antenna:                 Antenna Typ: Patch
                                                       Modulation:              FM
                                                       Crest Factor:            1
                                                       TX range:                2458 MHz & 2474 MHz
                                                       Used TX Channels:        low: 2458 MHz, high: 2474 MHz
                                                       Used Phantom:            SAM Twin Phantom V4.0, as defined by the IEEE SCC-34/SC2
                                                                                group and delivered by Schmid & Partner Engineering AG
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH


                                                       DASY Test-Report, Phoenix_6575_367                                                      Page 18 of 28

                                                       7.3 Tissue Recipes
                                                       The following recipes are provided in percentage by weight.
                                                       2450 MHz, Body:            29.68% Diethylenglykol-monobutylether
                                                                                  70.00% De-Ionized Water
                                                       7.4 Material Parameters

                                                       For the measurement of the following parameters the HP 85070B dielectric probe kit is used,
                                                       representing the open-ended coaxial probe measurement procedure.

                                                                                                                                         Temperature
                                                        Frequency                                     εr              σ [S/m]         Ambient         Liquid
                                                                                                                                         [° C]          [° C]


                                                        2450 MHz       Recommended Value        52.70 ± 2.67         1.95 ± 0.08     20.0 - 26.0           -
                                                          (Body)         Measured Value         50.70 ± 2.50         2.01 ± 0.20        22.5            217

                                                                            Table 4: Parameters of the tissue simulating liquids.


                                                       7.5 Simplified Performance Checking
                                                       The simplified performance check was realized using the dipole validation kit. The input power
                                                       of the dipole antenna was 250 mW (cw signal) and it was placed under the flat part of the SAM
                                                       phantom. The result is listed in the Table 5 and shown in Fig. 9. The target value was adopted
                                                       from the manufactures calibration certificate which is attached in the appendix. Table 6 includes
                                                       the uncertainty assessment for the system performance checking which was suggested by the
                                                       [IEEE 1528] and determined by Schmid & Partner Engineering AG. The expanded uncertainty
                                                       (K=2) is assessed to be ± 16.8%.

                                                                                                                                       Temperature
                                                         Frequency                       SAR1g [W/kg]          εr       σ [S/m]     Ambient         Liquid
                                                                                                                                       [° C]          [° C]


                                                         2450 MHz       Target Value          13.00          51.70        2.01      20.0 - 26.0        -
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                           (Body)      Measured Value         13.40          50.70        2.01         22.5           21.7

                                                                                   Table 5: Validation results, 2450 MHz.


                                                       DASY Test-Report, Phoenix_6575_367                                                     Page 19 of 28

                                                       Test Laboratory: IMST; File Name: 071103_1579.da4
                                                       DUT: Dipole 2450 MHz; Type: D2450V2; Serial: 709
                                                       Program: Unnamed Program
                                                       Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
                                                       Medium: Body 2450 MHz (σ = 2.01 mho/m, εr = 50.7, ρ = 1000 kg/m3)
                                                       Phantom section: Flat Section
                                                       DASY4 Configuration:
                                                       - Probe: ET3DV6 - SN1579; ConvF(4.4, 4.4, 4.4); Calibrated: 15.05.2003
                                                       - Sensor-Surface: 4mm (Mechanical And Optical Surface Detection)
                                                       - Electronics: DAE3 Sn335; Calibrated: 05.05.2003
                                                       - Phantom: SAM TP:1176; ;
                                                       - Measurement SW: DASY4, V4.1 Build 47; Postprocessing SW: SEMCAD, V1.6 Build 115
                                                       Unnamed procedure/Area Scan (5x5x1): Measurement grid: dx=20mm, dy=20mm
                                                       Reference Value = 92.7 V/m
                                                       Power Drift = -0.05 dB
                                                       Maximum value of SAR = 15.2 mW/g
                                                       Unnamed procedure/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                                       Peak SAR (extrapolated) = 27.5 W/kg
                                                       SAR(1 g) = 13.4 mW/g; SAR(10 g) = 6.21 mW/g
                                                       Reference Value = 92.7 V/m
                                                       Power Drift = -0.05 dB
                                                       Maximum value of SAR = 14.9 mW/g
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                Fig. 9:         Validation measurement 2450 MHz Head (November 07, 2003), coarse grid.
                                                                                     Ambient Temperature: 22.5° C, Liquid Temperature: 21.7° C.


                                                       DASY Test-Report, Phoenix_6575_367                                                    Page 20 of 28


                                                                                            Uncertai    Probability                      Standard     vi² or
                                                       Error Sources                                                   Divisor    ci
                                                                                            nty Value   Distribution                    Uncertainty    veff

                                                       Measurement System
                                                       Probe calibration                    ± 4.8 %       Normal         1        1      ± 4.8 %       ∞
                                                       Axial isotropy                       ± 4.7 %     Rectangular      √3       1      ± 2.7 %       ∞

                                                       Hemispherical isotropy                ±0%        Rectangular      √3       1       ±0%          ∞

                                                       Boundary effects                     ± 1.0 %     Rectangular      √3       1      ± 0.6 %       ∞

                                                       Linearity                            ± 4.7 %     Rectangular      √3       1      ± 2.7 %       ∞

                                                       System detection limit               ± 1.0 %     Rectangular      √3       1      ± 0.6 %       ∞

                                                       Readout electronics                  ± 1.0 %       Normal         1        1      ± 1.0 %       ∞

                                                       Response time                         ±0%        Rectangular      √3       1       ±0%          ∞

                                                       Integration time                       ± 0%      Rectangular      √3       1       ±0%          ∞

                                                       RF ambient conditions                ± 3.0 %     Rectangular      √3       1      ± 1.7 %       ∞

                                                       Probe positioner                     ± 0.4 %     Rectangular      √3       1      ± 0.2 %       ∞

                                                       Probe positioning                    ± 2.9 %     Rectangular      √3       1      ± 1.7 %       ∞

                                                       Algorithms for max SAR eval.         ± 1.0 %     Rectangular      √3       1      ± 0.6 %       ∞

                                                       Dipole
                                                       Dipole Axis to Liquid Distance       ± 2.0 %     Rectangular      1        1      ± 1.2 %       ∞
                                                       Input power and SAR drift mea.       ± 4.7 %     Rectangular      √3       1      ± 2.7 %       ∞
                                                       Phantom and Set-up
                                                       Phantom uncertainty                  ± 4.0 %     Rectangular      √3       1      ± 2.3 %       ∞

                                                       Liquid conductivity (target)         ± 5.0 %     Rectangular      √3      0.64    ± 1.8 %       ∞

                                                       Liquid conductivity (meas.)          ± 2.5 %       Normal         1       0.64    ± 1.6 %       ∞

                                                       Liquid permittivity (target)         ± 5.0 %     Rectangular      √3      0.6     ± 1.7 %       ∞

                                                       Liquid permittivity (meas.)          ± 2.5 %       Normal         1       0.6     ± 1.5 %       ∞
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                       Combined Uncertainty                                                              ± 8.4 %

                                                                          Table 6: Uncertainty budget for the system performance check.


                                                       DASY Test-Report, Phoenix_6575_367                                        Page 21 of 28

                                                       7.6 Environment
                                                       To comply with the required noise level (less than 12 mW/kg) periodically measurements
                                                       without a DUT were conducted.
                                                       Humidity:                37% ± 5 %

                                                       7.7 Test Equipment

                                                                                                         Serial       Last          Next
                                                       Test Equipment                      Model
                                                                                                        Number     Calibration   Calibration

                                                       DASY4 System
                                                       Software Version                     V4.0          N/A         N/A           N/A
                                                       Dosimetric E-Field Probe           ET3DV6          1579      05/2003       05/2004
                                                       Data Acquisition Electronics      DAE3 V1          565       04/2003       04/2004
                                                       Phantom                             SAM            1176        N/A           N/A
                                                       Performance Checking
                                                       System Validation Dipole, Body    D2450V2          709       07/2002       07/2004
                                                       Power Meter, Agilent               E4416A      GB41050414    11/2002       11/2003
                                                       Power Sensor, Agilent              E9301H      US40010212    12/2002       12/2003
                                                       RF Source (Network Analyzer)      HP8753D      3410A06555    11/2002       11/2003
                                                       RF Amplifier, Mini-Circuits        ZHL-42        D012296       N/A           N/A
                                                       Material Measurement
                                                       Network Analyzer                  HP8753D      3410A06555    11/2002       12/2003
                                                       Dielectric Probe Kit              HP85070B     US33020263      N/A           N/A

                                                                                        Table 7: Test equipment.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH


                                                       DASY Test-Report, Phoenix_6575_367                                           Page 22 of 28

                                                       7.8 Certificates of conformity
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                      Fig. 10: Certificate of conformity for the used SAM phantom


                                                       DASY Test-Report, Phoenix_6575_367                                       Page 23 of 28

                                                       7.9 Pictures of the device under test
                                                       Fig. 11 – 12 show the device under test.




                                                                                       Fig. 11: Side view of the device.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                  Fig. 12: Front and back view of the device.


                                                       DASY Test-Report, Phoenix_6575_367                                     Page 24 of 28

                                                       7.10 Test Positions for the Device under Test
                                                       Fig. 13 – Fig. 20 shown the test positions for the SAR measurements.




                                                                                            Fig. 13: Position 1.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                            Fig. 14: Position 1.


                                                       DASY Test-Report, Phoenix_6575_367                      Page 25 of 28




                                                                                        Fig. 15: Position 2.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                        Fig. 16: Position 2.


                                                       DASY Test-Report, Phoenix_6575_367                     Page 26 of 28




                                                                                     Fig. 17:   Position 3.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                     Fig. 18:   Position 3.


                                                       DASY Test-Report, Phoenix_6575_367                     Page 27 of 28




                                                                                     Fig. 19:   Position 4.
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                                     Fig. 20:   Position 4.


                                                       DASY Test-Report, Phoenix_6575_367                                            Page 28 of 28

                                                       7.11 Pictures to demonstrate the required liquid depth
                                                       Fig. 21 show the liquid depth in the used SAM phantom.




                                                                      Fig. 21:      Liquid depth for 2450 MHz Body measurements.
                                                       8 References

                                                       [DASY 1995]     Referenzliste des Herstellers, der Fa. Schmid & Partner Engineering AG, über
                                                                       installierte DASY-Systeme mit RX90 Robotern: Deutsche Telekom,
                                                                       Forschungs- und Technologiezentrum; Motorola Cellular - MRO; Motorola;
                                                                       Ericsson Mobile Communications AB; Nokia Mobile Phones LTD; IMST
                                                                       GmbH, 1995.
                                                       [FCC 2001]      Federal Communications Commission: Evaluating Compliance with FCC
                                                                       Guidelines for Human Exposure to Radiofrequency Electromagnetic Fields,
                                                                       Supplement C (Edition 01-01) to OET Bulletin 65 (Edition 97-01), FCC, 2001.
                                                       [IEEE 1999]     IEEE Std C95.1-1999: IEEE Standard for Safety Levels with Respect to
                                                                       Human Exposure to Radio Frequency Electromagnetic Fields, 3 kHz to
                                                                       300 GHz, Inst. of Electrical and Electronics Engineers, Inc., 1999.
                                                       [IEEE 1528]     IEEE P1528/D1.2: Recommended Practice for Determining the Peak Spatial-
dasy_report_fcc_supplement_c _3.3.doc/28.10..2003/CH




                                                                       Average Specific Absorption Rate (SAR) in the Human Head from Wireless
                                                                       Communications Devices: Measurement Techniques. April 21, 2003, Inst. of
                                                                       Electrical and Electronics Engineers.
                                                       [NIST 1994]     NIST: Guidelines for Evaluating and Expressing the Uncertainty of NIST
                                                                       Measurement Results, Technical Note 1297 (TN1297), United States
                                                                       Department of Commerce Technology Administration, National Institute of
                                                                       Standards and Technology, 1994.
                                                       [DASY4]         Schmid & Partner Engineering AG: DASY4 Manual. March 2003



Document Created: 2004-01-22 14:01:32
Document Modified: 2004-01-22 14:01:32

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC