Test rpt SAR Part5

FCC ID: RFD-CSNGG

Test Report

Download: PDF
FCCID_4364133

                                           Validation Report                                                                          Kind of doc.:
                                                                                                                                      QM Template
                                       No. VAL_0946_EF 2019-03
EUROFINS PRODUCT SERVICE GmbH
Storkower Str. 38c, 15526 Reichenwalde, Germany
1       Customer
Eurofins Product Service GmbH


2       Object
Equipment Number                                       EF00946
Equipment Name:                                        System validation dipole
Equipment Type:                                        D750V3
Serial Number:                                         1125
Manufacturer:                                          Schmid & Partner Engineering AG


3       State of Measurement
Validation:
Performance Control:
Other:


4       Performance of Measurement
4.1        Generals
(e.g. object of validation such as specific setup, non-standard method or SW, specification of the requirements, test set-up configuration,
risk analysis etc.)

Dipol verification

4.2        Validation procedure / measurement
(e.g. comparison of results achieved with other methods, interlaboratory comparison, systematic assessment of factors influencing the
result, assessment of the uncertainty of the results based on scientific understanding of the theoretical principles of the method and
practical experience; criteria/requirements for approval/rejection etc.)

According KDB 865664 D01 SAR Measurement 100 MHz to 6 GHz v01r04 3.2.2 Dipole calibration

Limits for the verification:             return loss <20% to the original measurement or >20 dB minimum return-loss
                                         Impedance <5 Ω to the original measurement.


4.3        Used reference equipment
                                                                                                           Equipment
            Equipment name                           Equipment type                   Manufacturer                        Cal. Date       Cal. Due Date
                                                                                                            number
                                                                                  Hewlett-Packard
    RF Network analyzer                            8752 C                         Company Santa           EF00140      2018-07-25        2019-07-25
                                                                                  Clara

-     new acquired (incl. calibration)
-     new calibrated
-     check reference standard


4.4        Environmental conditions
Temperature:                                                                                     _23_°C + 2°C
Relative Air Humidity:                                                                           _50_ rH + 5%
Air Pressure:                                                                                    _1020_ hPa + 5%




                                                                                                                               Page 115 of 177
This report shall not be reproduced except in full, without the written approval of the laboratory.
version: 06                                              VAL_EF00946_2019-03_Validation report final                                      Page 1 / 5


<= eurofins
                                                    Product Service

                                                   Validation Report                                                               Kind of doc.;

                                      No. VAL_0946_EF 2019—03                                                                      QM Template

EUROFINS PRODUCT SERVICE GmbH
Storkower Str. 38¢, 15526 Reichenwalde, Germany


5        Results
5.1         General:
(e.q. measurement results, user instructions such as handling, transport, storage, preparation; checks to be made before the work
started; information about how to install (operations)—, to maintain—, to train and to use; safety measures etc.)


                                                     Original measurement                           Verification measurement       Margin
    Impedance,                                                        ;                                                :                     .
    transformend to feed point                           50.0 0 + 5.0 jQ                                50.15 0 + 3.38 jQ      0.15 0 — 1.62 jQ

    Return Loss                                               —26.1 dB                                     —29.72 dB               3.62 dB
    Tissue Validation                   sr                       55.5                                       54145                  244 %

    Tissue Validation                                                                                                                   3
                                gS                               0.96                                        0.986                 2.71 %
    System validation                                     8.52 Wikg (1g)                                 8.64Wi/kg (1g)            141 %
    Date:                                                   21.09.2017                                    13.03.2019



5.2         Measurement uncertainty
The reported expanded uncertainty of measurement is stated as the standard uncertainty multiplied by the coverage factor k=2, which
for a normal distribution corresponds to a coverage probability of approximately 95%.
+/— 2.5 %

5.3         Results of Validation
Validated
                                                                                              0o K




Not validated


6        Operator


Pudell
Name
                                                      P Frilf?
                                                      Signature


Place and Date of Verification:                       Reichenwalde, 14.03.2019




Attachment:
Impedance, Relurn Loss, System validierung




This report shall not be reproduced excent in full without the writen approval of the laboratory.
version: 06                                             VAL_EFOO946_2019—03_Validation report final                                     Page 2/ 5


                                           Validation Report                                               Kind of doc.:
                                                                                                           QM Template
                                       No. VAL_0946_EF 2019-03
EUROFINS PRODUCT SERVICE GmbH
Storkower Str. 38c, 15526 Reichenwalde, Germany




                                                                                                       Page 117 of 177
This report shall not be reproduced except in full, without the written approval of the laboratory.
version: 06                                              VAL_EF00946_2019-03_Validation report final           Page 3 / 5


                                           Validation Report                                               Kind of doc.:
                                                                                                           QM Template
                                       No. VAL_0946_EF 2019-03
EUROFINS PRODUCT SERVICE GmbH
Storkower Str. 38c, 15526 Reichenwalde, Germany




                                                                                                       Page 118 of 177
This report shall not be reproduced except in full, without the written approval of the laboratory.
version: 06                                              VAL_EF00946_2019-03_Validation report final           Page 4 / 5


                                           Validation Report                                               Kind of doc.:
                                                                                                           QM Template
                                       No. VAL_0946_EF 2019-03
EUROFINS PRODUCT SERVICE GmbH
Storkower Str. 38c, 15526 Reichenwalde, Germany




                                                                                                       Page 119 of 177
This report shall not be reproduced except in full, without the written approval of the laboratory.
version: 06                                              VAL_EF00946_2019-03_Validation report final           Page 5 / 5


Calibration
     .
            Laboratory of                                          SA/
                                                                       P                                    S      Schweizerischer Kalibrierdienst
                                                                   $
SChmI.d & Partner                                                 SB\\EHT{/&?                               c      Service suisse d‘étalonnage
  Englneermg AG                                                   gzy>e                                            Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       *///\\\ w                                S      Swiss Calibration Service
                                                                       Ahitule®

Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       Eurofins                                                                             Certificate No: D900V2—164_Sep18




 Object                                D900V2 — SN:164



 Calibration procedure(s)              QA CAL—O5.v10
                                       Calibration procedure for dipole validation kits above 700 MHz




 Calibration date:                     September 03, 2018



 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical forcalibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                      Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                            Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                            Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                            Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                            Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                       Dec—18
 DAE4                                   SN: 601                   26—Oct—17 (No. DAE4—601_Oct17)                       Oct—18

 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Power meter EPM—442A                   SN:    GB37480704         07—Oct—15   (in   house check   Oct—16)               In   house   check:   Oct—18
 Power sensor HP 8481A                  SN:    US37202783         07—Oct—15   (in   house check   Oct—16)               In   house   check:   Oct—18
 Power sensor HP 8481A                  SN:    MY41092317         07—Oct—15   (in   house check   Oct—16)               In   house   check:   Oct—18
 RF generator R&S SMT—06                SN:    100972             15—Jun—15   (in   house check   Oct—16)               In   house   check:   Oct—18
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—17)                     In house check: Oct—18


                                        Name                                      Function                               Signature
 Calibrated by:                         Michael Weber       _      8      ___ Laboratory Technician         _        B                ——        ————



 Approved by:                           Katia Pokovic                             Technical Manager                    W



                                                                                                                       Issued: September 4, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D900V2—164_Sep18                                       Page 1 of 8                                                                     7


Calibration
      .     Laboratory of                                     wl/7
                                                             Si\\/%            S   Schweizerischer Kalibrierdienst
                                                             $            Z
Schmid & Partner                                            ila\et—//mfi        c   Service suisse d‘étalonnage
  Engmeenng AG                                              to                     Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 {'/,,/Q\\\\F      s   Swiss Calibration Service
                                                                 Allals

Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
         MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook
Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
         Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
         SAR measured: SAR measured at the stated antenna input power.
         SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
         SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D900V2—164_Sep18                                 Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                              V52.10.1
      Extrapolation                                  Advanced Extrapolation
      Phantom                                         Modular Flat Phantom

      Distance Dipole Center— TSL                               15 mm                             with Spacer
     Zoom Scan Resolution                               dx, dy, dz =5 mm
      Frequency                                         900 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                   22.0 °C           41.5               0.97 mho/m

      Measured Head TSL parameters                              (22.0 + 0.2) °C   40.6 + 6 %         0.95 mho/m + 6 %
     Head TSL temperature change during test                       <0.5 °C


SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
     SAR measured                                        250 mW input power                  2.69 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         10.9 Wikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                  1.72 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         6.94 Wikg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                  22.0 °C           55.0               1.05 mho/m
      Measured Body TSL parameters                              (22.0 + 0.2) °C   54.7 x 6 %         1.02 mho/m + 6 %
      Body TSL temperature change during test                      <0.5 °C            ~———                    ~———


SAR result with Body TSL

      SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition

      SAR measured                                       250 mW input power                      2.73 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W          11.1 Wikg & 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                 condition

      SAR measured                                       250 mW input power                  1.78 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W         7.24 Wikg + 16.5 % (k=2)



Certificate No: D900V2—164_Sep18                       Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           49.9 Q — 4.7 jQ

      Return Loss                                                                       —26.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           45.6 Q — 7.6 jQ

      Return Loss                                                                       —20.8 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.407 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG

      Manufactured on                                                                 May 16, 2002




Certificate No: D900V2—164_Sep18                          Page 4 of 8                                                7


DASY5 Validation Report for Head TSL

                                                                                    Date: 03.09.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 900 MHz; Type: D900V2; Serial: D900V2 — SN:164

Communication System: UID 0 — CW; Frequency: 900 MHz
Medium parameters used: f = 900 MHz; 0 = 0.95 S/m; & = 40.6; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    e   Probe: EX3DV4 — SN7349; ConvF(9.71, 9.71, 9.71) @ 900 MHz; Calibrated: 30.12.2017

    e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    e   Phantom: Flat Phantom 4.9 (front); Type: QD OOL P49 AA; Serial: 1001

    e   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 65.40 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 4.07 W/kg
SAR(1 g) = 2.69 W/kg; SAR(10 g) = 1.72 W/kg
Maximum value of SAR (measured) = 3.60 W/kg


        dB


        —2.40

        ~4.80

        —1.20

        —9.60

        —12.00
                   0 dB = 3.60 W/kg = 5.56 dBW/kg




Certificate No: D900V2—164_Sep18               Page 5 of 8


Impedance Measurement Plot for Head TSL




 File        View   Channel   Sweep   Calibration   Trace Scale   Marker   System   Window   Help

                                                                                          900.000000 MHz       49.947 O
                                                                                                37.793 pF     —4.6791 Q
                                                                                                             46.768 mU
                                                                                                               ~87.970 °




                    Ch1Aug= 20
         Chi: Start 700.000 MHa   ———                                                                        Stop 1.10000 GHz


    0.00
   5.00
                                                                                          900.000000   MHz   —2§.601 dB
   0.00

        00

    10.00

    15.00
        0.00

        5.00

         00

         00
         0|       Ch1Augs
         Chi: Start 700.000 MHz   ——                                                                         Stop 1.10000 GHz


                    CH 1:                           C* 1—Port              Avg=20 Delay




Certificate No: D900V2—164_Sep18                                    Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                    Date: 03.09.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 900 MHz; Type: D900V2; Serial: D900V2 — SN:164

Communication System: UID 0 — CW; Frequency: 900 MHz
Medium parameters used: £ = 900 MHz; 0 = 1.02 S/m; & = 54.7; p = 1000 kg/m?‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

   e    Probe: EX3DV4 — SN7349; ConvF(9.83, 9.83, 9.83) @ 900 MHz; Calibrated: 30.12.2017

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 26.10.2017

        Phantom: Flat Phantom 4.9 (Back); Type: QD OOR P49 AA; Serial: 1005

        DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 63.60 V/m; Power Drift =—0.03 dB
Peak SAR (extrapolated) = 3.95 W/kg
SAR(I g) = 2.73 W/kg; SAR(10 g) = 1.78 W/kg
Maximum value of SAR (measured) = 3.57 W/kg




                   0 dB = 3.57 W/kg = 5.53 dBW/kg




Certificate No: D900V2—164_Sep18               Page 7 of 8


Impedance Measurement Plot for Body TSL




 Eile   View    Channel    Swesp   Calibration   Trace Scale   Marker   System   Window   Help



                                                                        SA§g0o0.000000 MHz
                                                                                       900.000000 MHz
                                                                                             23.327 pF
                                                                                                            45.596 0O
                                                                                                           —7.5809 C
                                                                                                          91.425 mU
                                                                                                            —115.62°




               Ch1Aug= 20                                      z.          .
        Ch1: Start 700,000 MHz   ——                                                                       Stop 1.10000 GHz


                                                                               > 1     900.000000   MHz   —211.779 dB




        00

    35.00
       o0       Ch1      =
       Chi: Start 700.000 MHz    ——                                                                       Stop 110000 GHz


   Status       CH1:      B1._______ JC‘iPot                            Avg=20 Delay




Certificate No: D900V2—164_Sep18                                 Page 8 of 8



Document Created: 2019-07-19 12:27:13
Document Modified: 2019-07-19 12:27:13

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC