2013SAR00099-rev2_part5_3.pdf

FCC ID: RAD357

RF Exposure Info

Download: PDF
FCCID_2058439

TMCEI

 Head TSL parameters at 5800 MHz
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity       Conductivity
      Nominal Head TSL parameters                                  22.0 °C           35.3             5.27 mho/m
      Measured Head TSL parameters                              (22.0 £ 0.2) °C   34.1 £6 %        5.11 mho/m + 6 %
      Head TSL temperature change during test                      <0.5 °C                                  ~~


 SAR result with Head TSL at 5800 MHz

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                        100 mW input power                  8.23 mW / g
      SAR for nominal Head TSL parameters                     normalized to 1W        81.6 mW / g + 19.9 %(k=2)


      SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                        100 mW input power                  2.35 mW / g
      SAR for nominal Head TSL parameters                     normalized to 1W        23.3 mW / g x 19.5 % (k=2)




 Certificate No: DSGHzV2—1040_Jun12                    Page 4 of 13


TMCEI

 Body TSL parameters at 5200 MHz
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Body TSL parameters                                  22.0 °C           49.0                5.30 mho/m
      Measured Body TSL parameters                              (22.0 £0.2) °C    47.0 £6 %           5.97 mho/m + 6 %
      Body TSL temperature change during test                      <0.5 °C            se«                      «n


 SAR result with Body TSL at 5200 MHz

      SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                        100 mW input power                   7.37 mW /g
      SAR for nominal Body TSL parameters                     normalized to 1W        73.1 mW / g # 19.9 % (k=2)


      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                        100 mW input power                   2.07 mW / g
      SAR for nominal Body TSL parameters                     normalized to 1W        20.5 mW / g # 19.5 % (k=2)



 Body TSL parameters at 5500 MHz
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Body TSL parameters                                  22.0 °C            48.6               5.65 mho/m
      Measured Body TSL parameters                              (22.0 £ 0.2) °C   46.5 2 6 %          5.76 mho/m + 6 %
      Body TSL temperature change during test                      <0.5 °C            w«                       se«


 SAR result with Body TSL at 5500 MHz

      SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
      SAR measured                                        100 mW input power                     7.87 mW / g
      SAR for nominal Body TSL parameters                     normalized to 1W        78.1 mW / g # 19.9 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                 condition
      SAR measured                                        100 mW input power                     2.19 mW /g
      SAR for nominal Body TSL parameters                     normalized to 1W        21.7 mW / g # 19.5 % (k=2)




 Certificate No: D5GHzV2—1040_Jun12                    Page 5 of 18


TMCEI

 Body TSL parameters at 5800 MHz
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity       Conductivity
      Nominal Body TSL parameters                                  22.0 °C           48.2             6.00 mho/m
      Measured Body TSL parameters                              (22.0 £ 0.2) °C   46.0 £6 %        6.16 mho/m + 6 %

      Body TSL temperature change during test                      <05°C              se«                   »e«*

 SAR result with Body TSL at 5800 MHz

      SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                        100 mW input power                  7.44 mW / g
      SAR for nominal Body TSL parameters                     normalized to 1W        73.8 mW / g £ 19.9 % (k=2)


      SAR averaged over 10 cm* (10 g) of Body TSL                 condition
      SAR measured                                        100 mW input power                  2.06 mW / g
      SAR for nominal Body TSL parameters                     normalized to 1W        20.4 mW / g # 19.5 % (k=2)




 Certificate No: D5GHzV2—1040_Jun12                    Page 6 of 13


TMCEI                                                                                      —

 Appendix

 Antenna Parameters with Head TSL at 5200 MHz

       Impedance, transformed to feed point     —                                    49.2 Q —7.1 jQ
       Return Loss                                                                      —22.8 dB


 Antenna Parameters with Head TSL at 5500 MHz

       Impedance, transformed to feed point                                          51.5 Q — 4.4 jQ
       Return Loss                                                                      —26.8 dB


 Antenna Parameters with Head TSL at 5800 MHz

       Impedance, transformed to feed point                                          55.3 Q — 2.7 jQ
       Return Loss                                                                      —24.9 dB

 Antenna Parameters with Body TSL at 5200 MHz

       Impedance, transformed to feed point                                          49.6 Q — 5.5 jQ
       Return Loss                                                                      —25.2 dB

 Antenna Parameters with Body TSL at 5500 MHz

       Impedance, transformed to feed point                                          52.5 Q — 8.2jQ
       Return Loss                                                                      —28.1 dB


 Antenna Parameters with Body TSL at 5800 MHz

       Impedance, transformed to feed point                                           56.6 Q — 1.3 jQ
       Return Loss                                                                      —24.0 dB



 General Antenna Parameters and Design

       Electrical Delay (one direction}                                                  1.202 ns                         —I


 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding lineis directly connected to the
 second arm of the dipole. The antennais therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole lengthis still
 according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections nearthe
 feedpoint may be damaged.

 Additional EUT Data

       Manufactured by                                                                   SPEAG
       Manufactured on                                                             December 30, 2005



 Certificate No: DSGHzV2—1040_Jun12                      Page 7 of 13


DASY5 Validation Report for Head TSL

                                                                                         Date: 19.06.2012

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 5GHz; Type: DSGHzV2; Serial: DSGHzV2 — SN: 1040

Communication System: CW; Frequency: 5200 MHz, Frequency: 5500 MHz, Frequency: 5800 MHz
Medium parameters used: f = 5200 MHz; 0 = 4.52 mho/m; &, = 35; p = 1000 kg/m‘ , Medium parameters
used: f = 5500 MHz; o = 4.8 mho/m; s = 34.6; p = 1000 kg/m3 , Medium parameters used: f = 5800 MHz; 0
=5.11 mho/m;& = 34.1; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN3503; ConvF(5.41, 5.41, 5.41); Calibrated: 30.12.2011, ConvF(4.91, 4.91,
       4.91); Calibrated: 30.12.2011, ConvF(4.81, 4.81, 4.81); Calibrated: 30.12.2011;

   *   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   *   Electronics: DAE4 Sn601; Calibrated: 04.07.2011

   &   Phantom: Flat Phantom 5.0 (front); Type: QDOOOPS0AA; Serial: 1001

   *   DASY52 52.8.1(838); SEMCAD X 14.6.5(6469)


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5200 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 65.507 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 30.371 mW/g
SAR(I g) = 8.2 mW/g; SAR(10 g) = 2.36 mWi/g
Maximum value of SAR (measured) = 19.0 mW/g

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5500 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 66.096 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 35.013 mW/g
SAR(I g) = 8.82 mW/g; SAR(10 g) = 2.52 mW/g
Maximum value of SAR (measured) = 21.2 mW/g

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5800 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurementgrid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 62.419 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 34.147 mW/g
SAR(I g) = 8.23 mW/g; SAR(10 g) = 2.35 mWi/g
Maximum value of SAR (measured) = 20.0 mW/g




Certificate No: DSGHzV2—1040_Jun12              Page 8 of 13


       —10.00




       —20.00




       —30.00




       ~40.00




       —50.00
                  0 dB = 20.0 mW/g = 26.02 dB mW/g




Certificate No: DSGHzV2—1040_Jun12           Page 9 of 13


TMC                                                                                       w




 Impedance Measurement Plot for Head TSL




                                                               19 Jun 2ei2 esi22:ss
          CHM sii     i uo rs         4149188 a   —74387
                                                   BR    0. 4.2075 pF      5 200.000 000 MHz

          x
                                                                                               CH1 Markers
          be?                                                                                   2 51.523 a
                                                                                                 ~4.9945 a
          cor                                                                                  5.50000 GHz
                                                                                                a95.916 a
                                                                                                 —2.7012 a
                                                                                               5.80000 GHz

          Av
          w

          H1d
           ch2


                                                                                               CH2 Markers
                                                                                               2—26.787 d8
          toe                                                                                  $.50000 GHz
                                                                                               ai—24.943 48
                                                                                                580000 GHz
          fv
          148

          H1d
                    START 5 000.000 000 MHz                         STOP 6 000.000 000 MHz




 Certificate No: D5GHzV2—1040_Jun12               Page 10 of 13


DASY5 Validation Report for Body TSL

                                                                                           Date: 18.06.2012

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 5GHz; Type: DSGHzV2; Serial: DSGHzV2 — SN: 1040

Communication System: CW; Frequency: 5200 MHz, Frequency: 5500 MHz, Freckuency: 5800 MHz
Medium parameters used: f = 5200 MHz; 0 = 5.37 mho/m; &, = 47; p = 1000 kg/m* , Medium parameters
used: £= 5500 MHz; 0 = 5.76 mho/m; &, = 46.5; p = 1000 kg/ml , Medium parameters used: £ = 5800 MHz;
0 = 6.16 mho/m; &, = 46; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2007)

DASY52 Configuration:

    «_   Probe: EX3DV4 — SN3503; ConvF(4.91, 4.91, 4.91); Calibrated: 30.12.2011, ConvF(4.43, 4.43,
         4.43); Calibrated: 30.12.2011, ConvF(4.38, 4.38, 4.38); Calibrated: 30.12.2011;

    *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *    Electronics: DAE4 Sn601; Calibrated: 04.07.2011

    &    Phantom: Flat Phantom 5.0 (back); Type: QDOOOP50AA; Serial: 1002

    *    DASY52 52.8.1(838); SEMCAD X 14.6.5(6469)


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5200 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 58.667 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 29.022 mW/g
SAR(I g) = 7.37 mW/g; SAR(10 g) = 2.07 mW/g
Maximum value of SAR (measured) = 17.2 mW/g

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5500 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 58.708 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 33.769 mW/g
SAR(I g) = 7.87 mW/g; SAR(10 g) = 2.19 mW/g
Maximum value of SAR (measured) = 19.0 mW/g

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5800 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurementgrid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 55.529 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 34.868 mW/g
SAR(I g) = 7.44 mW/g; SAR(10 g) = 2.06 mWi/g
Maximum value of SAR (measured)=18.1 mW/g




Certificate No: D5GHzV2—1040_Jun12               Page 11 of 13


       —20.00



       —30.00



       —40.00



       —50.00
                  0 dB = 18.1 mW/g=25.15 dB mW/g




Certificate No: DSGHzV2—1040_Jun12         Page 12 of 13


Impedance Measurement Plot for Body TSL




                                                                 18 Jun 2012 15:10:28
         CHI) sii     1 U FS           1149.613 0   —5.4824 n 5.5827 pF      5 200.000 000 MHz
                                                uescto~ ~—

                                                                                                 CHL Markers
         bel                                                                                      2 52.479 a
                                                                                                   —3,1523 a
         5e                                                                                      5.50000 6Hz
                                                                                                  3: 56,596   a
                                                                                                   82e¢A
                                                                                                 5.80000 CHz

         Av
         5°

         Hid
          cH2


                                                                                                 CH2 Markers
                                                                                                 2:—28.146 4B
          tor                                                                                     $.50000 6z
                                                                                                  :—23.996 aB
                                                                                                  5.80000 GHz
          Av
          169

          H1d
                    START 5 000.000 000 IMHz                           STOP 6 000.000 000 MHz




Certificate No: D5GHzV2—1040_Jun12                    Page 13 of 13


                                                                       No. 2013SAR00099
                                                                        Page 201 of 206


ANNEX I        DIPOLE QUALIFICATION FOR THE EXTENDED 3-YEAR

                CALIBRATION INTERVAL

I1 Dipole750
The information and documentation below are provided to qualify the extended 3-year
calibration interval of dipole.


I1.1 List of Equipment
 No.                   Name                               Type                  Serial Number
  01             Network analyzer                        E5071C                 MY46110673
  02                Power meter                          NRVD                      102083
  03                Power sensor                         NRV-Z5                    100542
  04             Signal Generator                        E4438C                 MY49070393
  05                  Amplifier                          60S1G4                   0331848
  06                E-field Probe                  SPEAG ES3DV3                     3149
  07                    DAE                            SPEAG DAE4                   771
  08            Dipole Validation Kit              SPEAG D750V3                     1045


I1.2 Results of Impedance, Return-loss and System validation
Dipole 750 - Head
                                                Year
                                                                    Deviation         Limit
                                        2011            2012
                    Real (Ω)            54.2             51.7         2.5Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           -2.3             0.9          3.2Ω       Deviation < 5Ω
       Return-loss (dB)                 -26.8           -26.7        0.1dB      Deviate < 0.2dB
  System               10g              1.4              1.43        2.14%      Deviation < 10%
 validation             1g              2.14             2.21        3.27%      Deviation < 10%


Dipole 750 - Body
                                                Year
                                                                    Deviation         Limit
                                        2011            2012
                    Real (Ω)            49.5             49.6         0.1Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           -4.1             0.6          4.7Ω       Deviation < 5Ω
       Return-loss (dB)                 -27.5           -27.6        0.1dB      Deviate < 0.2dB
  System               10g              1.45             1.47        1.38%      Deviation < 10%
 validation             1g              2.2              2.25        2.27%      Deviation < 10%


According to the above tables, it is not necessary to recalibration the dipoles in 2012.


                                                                       No. 2013SAR00099
                                                                        Page 202 of 206


I2 Dipole 835
The information and documentation below are provided to qualify the extended 3-year
calibration interval of dipole.


I2.1 List of Equipment
 No.                   Name                               Type                  Serial Number
  01             Network analyzer                        E5071C                 MY46110673
  02                Power meter                          NRVD                      102083
  03                Power sensor                         NRV-Z5                    100542
  04             Signal Generator                        E4438C                 MY49070393
  05                  Amplifier                          60S1G4                   0331848
  06                E-field Probe                  SPEAG EX3DV4                     3846
  07                    DAE                            SPEAG DAE4                   771
  08            Dipole Validation Kit              SPEAG D835V2                     443


I2.2 Results of Impedance, Return-loss and System validation
Dipole 835 - Head
                                                Year
                                                                    Deviation         Limit
                                        2012            2013
                    Real (Ω)            50.8             53.6         2.8Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           -6.7             -2.2         4.5Ω       Deviation < 5Ω
       Return-loss (dB)                 -23.5           -23.3        0.2dB      Deviate < 0.2dB
  System               10g              1.52             1.56        2.63%      Deviation < 10%
 validation             1g              2.33             2.38        2.15%      Deviation < 10%


Dipole 835 - Body
                                                Year
                                                                    Deviation         Limit
                                        2012            2013
                    Real (Ω)            46.8             49.1         2.3Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           -7.8             -3.7         4.1Ω       Deviation < 5Ω
       Return-loss (dB)                 -21.2           -21.4        -0.2dB     Deviate < 0.2dB
  System               10g              1.59             1.61        1.26%      Deviation < 10%
 validation             1g              2.42             2.42        0.00%      Deviation < 10%


According to the above tables, it is not necessary to recalibration the dipoles in 2013.


                                                                       No. 2013SAR00099
                                                                        Page 203 of 206


I3 Dipole 1750
The information and documentation below are provided to qualify the extended 3-year
calibration interval of dipole.


I3.1 List of Equipment
 No.                   Name                               Type                  Serial Number
  01             Network analyzer                        E5071C                 MY46110673
  02               Power meter                           NRVD                      102083
  03               Power sensor                          NRV-Z5                    100542
  04             Signal Generator                        E4438C                 MY49070393
  05                  Amplifier                          60S1G4                   0331848
  06               E-field Probe                   SPEAG EX3DV4                     3846
  07                    DAE                            SPEAG DAE4                   771
  08            Dipole Validation Kit              SPEAG D1750V2                    1003


I3.2 Results of Impedance, Return-loss and System validation
Dipole 1750 - Head
                                                Year
                                                                    Deviation         Limit
                                        2012            2013
                     Real (Ω)           50.3             51.5         1.2Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           1.3              3.7          2.4Ω       Deviation < 5Ω
       Return-loss (dB)                 -37.5           -37.3        0.2dB      Deviate < 0.2dB
  System               10g              4.77             4.91        2.94%      Deviation < 10%
 validation             1g              8.91             9.24        3.70%      Deviation < 10%


Dipole 1750 - Body
                                                Year
                                                                    Deviation         Limit
                                        2012            2013
                     Real (Ω)           45.7             48.3         2.6Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           1.2              2.9          1.7Ω       Deviation < 5Ω
       Return-loss (dB)                 -26.6           -26.5        0.1dB      Deviate < 0.2dB
  System               10g               5               5.01        0.20%      Deviation < 10%
 validation             1g              9.26             9.22        -0.43%     Deviation < 10%


According to the above tables, it is not necessary to recalibration the dipoles in 2013.



Document Created: 2013-08-22 10:28:32
Document Modified: 2013-08-22 10:28:32

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC