2013SAR00065_Rev02_part3

FCC ID: RAD330

RF Exposure Info

Download: PDF
FCCID_1995678

TME
Impedance Measurement Plot for Body TSL




                                                             9 May 2012 12:80i94
          CB] si      1 U re      1t46.815 6   ~783400   24.330 pF      835.000 000 Mz



                                                                                         Cha Markers
          ne 1                                                                            2 86745 6
          tor                                                                                27.27?
                                                                                         208.060  Mzo



          h
          158



          Hd


          on2


                                                                                         CH2 Mankens
          Cor
                                                                                         2e41.953  4e
                                                                                          200.000 Mitz



          By
          18°

          HId




Certificate No: DB5V2—443_May12                 Page 8 of 8


TME                                                                                                                   e
Calibration Laboratory of                                                                                         Schwelzerischer Kalibrierdenst
Schmid & Partner                                                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                                                  Servizio svizzero di taratura
Zoughausstrasse 43, 8004 Zurich, Switzerland                                                                      Swiss Callbration Service


 Accredited by the Swiss Accreditation Service (GAS)                                            Accreditation No.: SCS 108
 The Swiss Accreditation Service is one of thesignatories to the EA
 Multifateral Agreement forthe recognition of calibration cartificates
 Client     (TMC Beijfing                                                                       CertificateNo: D1750V2—1003_May12

[CALIBRATION CERTIFICATE
 Object                                D1750V2 — SN: 1003


 Celibration procedure(s)             Ga cat—osve
                                      Calibration procedure for dipole validation kits above 700 MHz


 Calibration date:                    May 08, 2012




 This calibration centficate documents the traceabilty to national standards, which realize the physical units of measurements (1)
 The measurements and the uncertainties with confidence probabilty are given on the following pages and are part of the certificate.

 All callbrations have bean conducted in the closed laboratory facility: environment temparature (22 > 3}°C and humidity < 70%.


 Calibration Equipment used (MBTE critical for calibration}


 Primary Standards                      10 &                      Cal Date (Cerificate No.)                           Scheduled Calibration
 Power meter EPM—442A                   GB37480704                06—Oct—11 (No. 217—01451)                           Oct—12
  Power sensor HP 8481A                 usar2027es                O5—Oct—11 (No. 217—01451)                           Oct—12
  Reterence 20 dB Attenuator            SN: 5058 (20k)            27—Mat—12 (No. 217—01830)                           Aprid
 Type—N mismatch combination            SN: 5047.2 / 06827        27—Mar—12 (No. 217—01533)                           Aprd
 Relerence Probe ES3DVS                 SN: 8205                  0—Dec—11 (No. ES3—8205_Dect1)                       Dec—12
 DaAE4                                  SN: 601                   04—Jul—11 (No. DAE4—601_Jult1)                      Jul—12


 Secondary Standards                    10                        Check Date (in house)                               Scheduled Check
  Power sensor HP 8481A                 MY41092317                18—Oct—02 (in house checkOct—11)                     in house check: Oct—13
  RF generator RAS SMT—06               100006                    04—Aug—29 (in house chack Oct—11)                    In house check: Oct—13
 Network Analyzer HP 8763E              USg7390505 $4206          18—Oct—01 (In house chack Oct—11)                    In house checic: Oct—12



                                        Name                                   Function                                Signature
 Callbrated by:                        Dimce lliew                            Lammmy Technician



 Approved by:                          Katja Pokovic                          Technical Manager



                                                                                                                      Issuad: May 8, 2012
 This callbration certificate shall not be reproduced except in full without written approval of the laboratory


 Certificate No: D1750V2—1008_May12                                   Page 1 of 8


TMCE

 Calibration Laboratory of                                                     .   Schwelzerischer Kalibrierdienst
 Schmid & Partner                                                              c   Service sulsse d‘stalonnage
   Engineering AG                                                                  Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                                  S   swiss Calibration Service

 Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multifateral Agreement for the recognition of callbration certificates

 Glossary:
 TSL                          tissue simulating liquid
 ConvF                        sensitivity in TSL / NORM x,y,z
 N/A                          not applicable or not measured

 Calibration is Performed According to the Following Standards:
     a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
        Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
        Communications Devices: Measurement Techniques®", December 2003
     b) 12C 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
          devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
          February 2005
     c) Federal Communications Commission Office of Engineering & Technology (FCC OET},
        "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
        Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
          Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
          Supplement C (Edition 01—01) to Bulletin 65

 Additional Documentation:
     d) DASY4/5 System Handbook

 Methods Applied and Interpretation of Parameters:
    * Measurement Conditions: Further details are available from the Validation Report at the end
          of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          parallel to the body axis.
     *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
     *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
          SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     +    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.



 Certificate No: D1750V2—1003_May12                              Page 2 of 8


TME«                                                                                         e
 Measurement Conditions
     DASY system configuration, as far as not given on page 1.
       DASY Version                                              DASY5                              V62.8.1
       Extrapolation                                  Advanced Extrapolation
       Phantom                                         Modular Flat Phantom
       Distance Dipole Center — TSL                              10 mm                            with Spacer

      Zoom Scan Resolution                               dx, dy, dz =5 mm
       Frequency                                         1750 MHz + 1 MHz



 Head TSL parameters
    The following parameters and calculations were applied.
                                                                 Temperature       Permittivity         Conductivity
      Nominal Head TSL parameters                                   22.0°C            40.1              1.37 mho/m
      Measured Head TSL parameters                               (22.0 2 0.2) °C   40.5 2 6 %         1.34 mho/m x 6 %
       Head TSL temperature change during test                      <0.5°C             «o                     «o

 SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
      SAR measured                                        250 mW input power                    8.91 mW /g
      SAR for nominal Head TSL paramaters                     normalized to 1W         36.2 mW /g x 17.0 % (k=2)


      SAR averaged over 10 em* (10 g) of Head TSL                  condition
       SAR measured                                       250 mW input power                    4.77 mW / q
       SAR for nomina! Head TSL parameters                    normalized to 1W         19.3 mW /g + 16.5 % (k=2)


 Body TSL parameters
    The following parameters and calculations were applied.
                                                                 Temperature       Permittivity         Conductivity
       Nominal Body TSL parameters                                  22.0 °C           53.4              1.49 mho/m
       Measured Body TSL parameters                              (22.0 x 0.2) °C   §2.0 16 %          1.46 mho/m £ 6 %
       Body TSL temperature change during test                      <0.5°C             se                     ssm

 SAR result with Body TSL

       SAR averaged over 1 cm° (1 g) of Body TSL                  Condition
       SAR measured                                       250 mW input power                    9.26 mW / g
       SAR for nominal Body TSL parameters                    normalized to 1W        37.4 mW / g « 17.0 % (k=2)


       SAR averaged over 10 em* (10 g) of Body TSL                 condition
       SAR measured                                       250 mW input power                    5.00 mW /g
       SAR for nominal Body TSL parameters                    normalized to 1W        20.1 mW / g x 16.5 % (k=2)




 Cortificate No: D1750V2—1003_May12                     Page 3 of 8


TMLI

 Appendix

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                            50.3 0 + 1.3 jQ
       Return Loss                                                                        —87.5 d8



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                            45.7 2 +1.2 j0
       Return Loss                                                                       ~26.6 4B




 General Antenna Parameters and Design

      | Electrical Delay (one direction)                                                  1.215 ns

 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxia!l cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—cirouited for DC—signals. On some of the dipales, small end caps
 are added to the dipole armsin order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessiveforce must be applied to the dipole arms, because they might bend or the soldered connections nearthe
 feedpoint may be damaged.


 Additional EUT Data

       Manufactured by                                                                     SPEAG
       Manufactured on                                                                  July 30, 2008




 Corlificate No: D1750V2—1003_May12                       Page 4 of 8


TMCE                                                                           h

 DASYS Validation Report for Head TSL

                                                                                    Date: 08.05.2012

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1003

 Communication System: CW; Frequency: 1750 MHz
 Medium parameters used: £ = 1750 MHz: 0 = 1.34 mho/m; £, =40.5; p = 1000 kgfm‘
 Phantom section; Flat Section
 Measurement Standard: DASY3 (IEEE/IEC/ANSI €63.19—2007)

 DASY52 Configuration:

    +   Probe: ES3DYV3 — SN3205; ConvF(5.22, 5.22, 5.22); Calibrated: 30.12.2011;

    +   Sensor—Surface: 3mm (Mechanical Surface Detection)

    +    Electronics: DAE4 Sn601; Calibrated: 04.07.201 1

    +    Phantom: Flat Phantom 5.0 (front); Type: QDOOOP30AA; Serial: 1001

    +    DASY52 52.8.1(838); SEMCAD X 14.6.5(6469)


 Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid; dx=5mm.,dy=5mm, dz=3mm
 Reference Value = 93.753 V/m; PowerDrift = 0.04 dB
 Peak SAR (extrapolated) = 15.720 mW/g
 SAR(I g) = 8.91 mW/g; SAR(IO g) = 4.77 mW/g
 Maximum value of SAR (measured) = 11.0 mW/g




                             zeo


                             snn


                             aaan



                                      0 dB = 11.0 mW/g= 20.83 dB mW/g




 Certificate No: D1750V2—1008_May12                 Page 5 of 8


TME«                                                                                     e
 Impedance Measurement Plot for Head TSL




                                                                   8 May 2012 10:09107
          CH]) sit     i U FS         1: 50.250 a   1.3184 6   419.90 pH      1 750,000 000 MHz




          Ay
          165


          Hid

           CH2




          Cor




          B
          159


          HId




 Certificate No: D1750V2—1003_May12                  Page 6 of 8


TMCE

 DASY5 Validation Report for Body TSL
                                                                                    Date: 08.05.2012

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1003

 Communication System: CW; Frequency: 1750 MHz
 Medium parameters used: f= 1750 MHz; 0 = 1.46 mho/m; &, = 52.9; p = 1000 kg/m"
 Phantomsection: Flat Section
 Measurement Standard: DASYS (IEEEAIEC/ANSI €63.19—2007)

 DASY52 Configuration:

    +    Probe: ES3DV3 — SN3205; ConyF(4.85, 4.85, 4.85); Calibrated: 30.12.2011;

    «+   Sensor—Surface: 3mm (Mechanical Surface Detection)

    +    Electronics: DAE4 Sn601; Calibrated: 04.07.201 1

    +    Phantom: Flat Phantom 5.0 (back); Type: QDOOOPSOAA; Serial: 1002

    +    DASY52 52.8.1(838); SEMCAD X 14.6.5(6469)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=3mm
 Reference Value = 93.095 V/m; Power Drift = 0.01 dB
 Peak SAR (extrapolated) = 15.760 mW/g
 SAR(I g) = 9.26 mW/g; SAR(1O g) = 5 mW/g
 Maximum value of SAR (measured) = 11.6 mW/g




                             ve.0


                             1600
                                      0 dB = 11.6 mW/g= 21.29 dB mW/g




 Cartificate No: D1750V2—1008_May12                 Page 7 of 8


TME«                                                                                       e
 Impedance Measurement Plot for Body TSL




                                                                     8 May 2012   1gi87ii0
                 s1t        1 uo Fe         11456908   14075 a   108.00 pH      1 750.000 008 MHz




          fvg
          16


          Hid

           CH2




          ty
          16°

          Hid

                       st       1 $50.a00   thiz                          StoP 2            iz




 Certificate No: D1750V2—1003_May12                     Page 8 of 8


TME                                                                                                                    e
 Calibration Laboratory of                                        wul,                                       g      SchweizerischerKalibrierdienst
       .                                                          i
 Schmid & Partner                                                w\_/&                                       c Service suissed‘étalonnage
    Engineering AG                                               en                                                 Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                    ns
                                                                 itant,                                             Swiss Callbration Service
                                                                     ltb

 Aceredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of callbration certificates

 client      (TMC Beijing                                                                        Certificate No: D1900V2—541_May12

 CALIBRATION CERTIFICATE
  Object                               D1900V2 — SN: 541


  Calibration procedure(s)             QA CAL—05.v8
                                       Calibrationprocedurefor dipole validation kitsabove 700 MHz


  Calibration date:                    May 09, 2012




  This callbration certificate documents the traceabilty to national standards, which realize the physical units of measurements (S1).
  The measurements and the uncartaintios with confidence probabilty are given on the following pages and are part of the cartificato.


  All calibrations have bean condusted in the closed laboratory facilty: environment temperature (22 « 3)°C and humidity < 70%.


  Calibration Equipment used (M&TE critical for calibration)

  Primary Standards                      ID #                      Cal Date (Gortiicate No.)                            Schedulad Calibration
  Power meter EPM—442A                   GB37480704                05—Oct—11 (No. 217—01451)                            Oct12
  Power sensorHP 8481A                   usarzor7es                05—O0ct—11 (No. 217—01451)                           Oct12
  Raference 20 dB Attenuator             N: sose (20k)             27—Mar12 (No. 217—01590)                             Apris
  Type—N mismatch combination            SN: 5047.2 / 0esa7        27—Mar12 (No. 217—01599)                             Apr18
  Raference Probe ESBDV3                 SN: 9205                  0—Dec—11 (No. ES3—8205_Dect1)                        Dec—12
  Dag4                                   SN: co1                   04—Jul—11 (No. DAE4—601_Jul11)                       dul—t2

  Secondary Standards                    ID #                      Check Date (in house)                                Schedulad Check
  Power sensor HP 8481A                  MY41002317                18—Oct—02 (in house check Oct—11)                    In house chack: Oct—13
  RF gensrator R&S SMT—06                100005                    04—Aug—98 (in house check Oct—11)                    In house chack: Oct—18
  Network Analyzer HP 8753E              Usereanses $4206          18—Oct—01 (inhouseccheck Oct—11)                     in house chack: Oct—12


                                         Name                                   Function                                 Signature
  Callbrated by:                        IsraeE—Naouq                            Laboratory Tevh_nibim



  Approved by:                          Katja Pokovie                           Technical Manager                   _fi“??4_

                                                                                                                        Issued: May 9, 2012
  This callbration cortificate shall not be repreduced excapt in full without written approval of the laboratory.


 Certificate No: D1900V2—541_May12                                     Page 1 of 8


TMCE
      i      4                                                 soltn,
 Calibration Laboratory of                                  plssoh           G    Schweizerischer Kalibrierdienst
 Schmid & Partner                                          ;ila‘\\e?//mfié    c   Service suisse d‘étalonnage
   Engineering AG                                          T * =g                Servizlo svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland               “,,,,fi\‘g        S   swiss Calibration Service
                                                               "ulal®


 Accredited by the Siwiss Accreditation Service (5AS)                        Accreditation No.: SCS 108
 The Swiss Accreditation Service is one of thesignatories to the EA
 Multilateral Agreementfor the recognitionof calibration certificates

 Glossary:
 TSL                          tissue simulating liquid
 ConvF                        sensitivity in TSL / NORM x,y,z
 N/A                          not applicable or not measured

 Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", December 2003
    b) 1EC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
          February 2005
     c) Federal Communications Commission Office of Engineering & Technology (FCC OET},
        "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
        Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
        Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions‘,
        Supplement C (Edition 01—01) to Bulletin 65

 Additional Documentation:
    d) DASY4/5 System Handbook

 Methods Applied and Interpretation of Parameters:
    * Measurement Conditions: Further details are available from the Validation Report at the end
       of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    «_ Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          parallel to the body axis.
     «_   Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
     «_   Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
          SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.



 Certificate No: D1900V2—641_May12                             Page 2 of 8


TMLI

 Measurement Conditions
     DASY system configuration, as far as not given on page 1.
       DASY Version                                              DaASYS                             V52.8.1
       Extrapolation                                  Advanced Extrapolation
       Phantom                                         Modular Flat Phantom
      Distance Dipole Center — TSL                               10 mm                            with Spacer
      Zoom Scan Resolution                               dx, dy, dz =5 mm
      Frequency                                          1900 MHz a 1 MHz



 Head TSL parameters
    The following parameters and calculations were applied.
                                                                 Temperature       Permittivity        Conductivity
       Nominal Head TSL parameters                                  22.0°C            40.0              1.40 mho/m
       Measured Head TSL parameters                              (22.0 2 0.2) °C   40.5 x 6 %         1.37 mho/m 2 6
       Head TSL temperature change during test                      <0.5°C             «——                    ———

 SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                   Condition
       SAR measured                                       250 mW input power                    9.62 mW / g
       SAR for nominal Head TSL parameters                    normalized to 1W         39.1 mW /g + 17.0 % (k=2)

       SAR averaged over 10 em* (10 g) of Head TSL                condition

       SAR measured                                       250 mW input power                    5.11 mW /g
       SAR for nominal Head TSL parameters                    normalized to 1W         20.6 mW /g = 16.5 % (k=2)


 Body TSL parameters
    The following parameters and calculations were applied.
                                                                 Temperature       Permittivity        Conductivity
       Nominal Body TSL parameters                                  22.0 °C           53.3              1.62 mho/m
       Measured Body TSL parameters                              (22.0 0.2) °C     §2.9 £ 6 %        1.52 mho/m a 6 %
       Body TSL temperature change during test                      <0.5°C             m                      mss

 SAR result with Body TSL

       SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
       SAR measured                                       250 mW input power                    10.0 mW /g

       SAR for nominal Body TSL parameters                    normalized to 1W        39.9 mW / g a 17.0 % (k=2)


       SAR averaged over 10 em> (10 g) of Body TSL                condition
       SAR measured                                       250 mW input power                    5.33 mW /g
       SAR for nominal Bady TSL parameters                    normalized to 1W        21.3 mW / g « 16.5 % (k=2)



 Cartificate No: D1900V2—541_May12                     Page 3 of 8


TMLI

 Appendix

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           §2.6 0 + 6.2 jQ
       Return Loss                                                                       —23.7 dB



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                           48.6 0 + 6.9 jQ
       Return Loss                                                                       —23.0 dB



 General Antenna Parameters and Design

     { Electrical Delay (one direction)                       ]                           1.197 ns

 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
 "Measurement Conditions® paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessive force must be applied to the dipols arms, because they might bend or the soldered connections near the
 feedpoint may be damaged.


 Additional EUT Data

       Manulactured by                                                                    SPEAG
       Manulactured on                                                                 July 26, 2001




 Certificate No: D1900V2—541_May12                        Page 4 of 8


TME                                                                           L
DASY5 Validation Report for Head TSL
                                                                                   Date: 09.05.2012
Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 541

Communication System: CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; a= 1.37 mho/m; &, = 40.5; p = 1000 kg m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/TEC/ANSI C63.19—2007)

DASY32 Configuration:

    s   Probe: ES3DV3 — SN3205; ConvF(5.01, 5.01, 5.01); Calibrated: 30.12.2011;

    «   Sensor—Surface: 3mm (Mechanical Surface Detection)

    «   Electronics: DAE4 Sn601; Calibrated: 04.07.201 1

    +   Phantom: Flat Phantom 5.0 (front); Type: QDOOOP50AA; Serial: 1001

   +    DASY52 52.8.1(838); SEMCAD X 14.6.5(6469)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 96.763 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 17.071 mW/g
SAR(I g) = 9.62 mW/g; SAR(10 g) = 5.11 mWi/g
Maximum value of SAR (measured) = 12.0 mW/g




        —3.60


        —1.20


        10.80


        "14.40


        18.00
                  0 dB = 12.0 mW/g =21.58 dB mW/g




Certificate No: D1900V2—541_May12               Page 5 of 8


TME                                                                                    e
 Impedance Measurement Plot for Head TSL




                                                              9 May 2012    ©9i¢4:17
          C sar       4 uce         11525920   61095 518.46 pH             1 900.000 000 Mz




          dvi
          ts°


          Hig




Certificate No: D1900V2—541_May12               Page 6 of 8


TMLI

 DASYS Validation Report for Body TSL

                                                                                    Date: 04.05.2012

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 541

 Communication System: CW; Frequency: 1900 MHz
 Medium parameters used: f = 1900 MHz; a= 1.52 mho/m:; & = 52.9; p = 1000 kgfm\
 Phantom section: Flat Section
 Measurement Standard: DASYS (IEEE/IEC/ANSI €63.19—2007)

 DASY32 Configuration:

         Probe: ES3DV3 — SN3205; ConvF(4.62, 4.62, 4.62); Calibrated: 30.12.2011;

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 Sn601; Calibrated: 04.07.201 1

         Phantom: Flat Phantom5.0 (back); Type: QDOOOP3OAA; Serial: 1002

         DASY52 52.8.1(838); SEMCAD X 14.6.5(6469)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value =95.165 V/m; Power Drift = —0.01 dB
 Peak SAR (extrapolated) = 17.442 mW/g
 SAR(I g) = 10 mW/g; SAR(10 g) = 5.33 mW/g
 Maximum value of SAR (measured) = 12.7 mW/g


        dB
        0


         —3.60


         —7.20


         10.80


         —14.40


         18.00
                   0 dB = 12.7 mW/g = 22.08 dB mW/g




 Cartificate No: D1900V2—541_May12              Page 7 of 8


TME«                                                                                      e
 Impedance Measurement Plot for Body TSL




                                                                4 May 2012.    irri4nd1
           CBI sn      i ue          1149.5720   699720   575.24 pH           1 $60.000 000 MHz




           Dei

           Cor




           fv
           is°

           Hid

           cH2




           Cor




           hy
           lEs


           H1d




 Cartificate No: D1900V2—541_May12                Page 8 of 8


TME                                                                                                                    2

 Calibration Laboratory of                                                                                         Schweizerischer Kalibrierdienst
 Schmid & Partner                                                                                                  Service sulsse d‘étalonnage
   Engineering AG                                                                                                  Servizio svizzero di taratura
Zoughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


 Accredited by the Swiss Accreditation Service (SAS)                                            Accreditation No.: SCS 108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of callbration certificates

 Ciient      TMC Beijing                                                                        Certificate No: D2450V2—853_May12

 CALIBRATION CERTIFICATE                                                                                                         §
 Object                                D2450V2 — SN: a53

  Callbration procedure(s)             QA CAL—05.v8
                                       Calibration procedure for dipole validation kitsabove 700 MHz


  Callbration date:                    May 02, 2012




 This callbration certificate documents the traceabilty to national standards, which reelize the physical units of measurements (81).
  The measurements and the uncertainties with confidenceprobabiity are given on the following pages and are part of the cortiicate.


  All eallbrations have been conducted in the closed laboratory facility; environment temperature (22 # 3)°C and humidity < 70%.


  Callbration Equipment used (METE critical for caltbration)


  Primary Standards                      ID #                     Cal Date (Certificate No.)                           Schadulad Calibration
  Power mater EPM—442A                  GB37480704                05—Oct—11 (No. 217—01451)                            Oct—12
  Power sensor HP 8481A                  US97202703               05—Oct—11 (No. 217—01451)                            Oct—12
  Reference 20 0B Attenuator            SN: 5058 (20k)            27—Mar—12 (No. 217—01530)                            Apr18
  Type—N mismatch combination           SN: 50472 / 06327         27—Mar—12 (No. 217—01533)                            Apr18
  Reference Probe ES3DV3                SN: 8205                  80:Dec—11 (No. ESS—8205_Dect1)                       Dec—12
  Daga                                  SN: s01                   O4—jul—11 (No. DAE4—601_Jul11)                       Jubt2

  Secondary Standards                    iD #                     Check Date (in house)                                Schadulad Check
  Power sensor HP 8481A                  MVA1092317                18—Oct—02 (in house check Oct—11}                    in house check: Oct—13
  AF generator R&S SMT—06                100005                   04—Aug—99 (in housecheck Oct—11)                      in house cheok: Oct—13
  Network Analyzer HP 8753E              US87390585 4206           18—Oct—01 (in house check Oot—11)                    in house cheok; Oct—12


                                         Name                                  Function                                 Signature
  Calibrated by:                        Israe E—Naouq                          Laboratory Technician                                 .
                                                                                                                       (.L‘".- ‘Cta L 'u‘\

  Approved by:                          Katja Pokovie                          Technical Minager                     ”%


                                                                                                                       Issued: May 2, 2012
  This callbration certificate shall not be reproduced except in full without witten approval of the laboratory.


 Crtificate No: D2450V2—853_May12 °                                   Page 1 of 8


TME«                                                                               ue
 Calibration Laboratory of                                                        Schweizerischer Kallbrierdienst
 Schmid & Partner                                                                 Service suisse u'élalormage

   Engineering AG                                                                 Servizio svizzero d taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                                     Swiss Callbration Service


 Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 108
 The Swiss Accreditation Serviceis one of the signatories to the EA
 Muftilatera! Agreement for the recognition of callbration certiticates

 Glossary:
 TSL                         tissue simulating liquid
 ConvF                       sensitivity in TSL / NORM xy,z
 N/A                          not applicable or not measured

 Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", December 2003
     b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
          devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
          February 2005
     c) Federal Communications Commission Office of Engineering & Technology (FCC OET),
        "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
        Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
        Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
          Supplement C (Edition 01—01) to Bulletin 65

 Additional Documentation:
    d) DASY4/5 System Handbook

 Methods Applied and Interpretation of Parameters:
     *    Measurement Conditions: Further details are available from the Validation Report at the end
          of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          paralle! to the body axis.
     *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connector to the feed point. The Return Loss ensures low
          reflected power. No uncertainty required.
     *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
     *    SAR measured: SAR measured at the stated antenna input power.
          SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.



 Certificate No: D2450V2—853_May12                               Page 2 of 8


TME«                                                                                        e
 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASYS                              V52.8.1
      Extrapolation                                   Advanced Extrapolation
      Phantom                                          Modular Flat Phantom
      Distance Dipole Center — TSL                               10 mm                           with Spacer
      Zoom Scan Resolution                              dx, dy, dz = 5 mm
      Frequency                                         2450 MHz + 1 MHz



 Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity
      Nominal Head TSL parameters                                  22.0 °C           39.2              1.80 mho/m
      Measured Head TSL parameters                              (22.0 £0.2) °C    39.6 26 %         1.81 mho/m « 6 %
      Head TSL temperature change during test                      <0.5°C             hss                    s«


 SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                   13.1 mW /g
      SAR for nominal Head TSL parameters                     normalized to 1 W       52.4 mW /g £ 17.0 % (k=2)


      SAR averaged over 10 cm* {10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                   6.09 mW /g
      SAR for nominal Head TSL parameters                     normalized to 1W        24.4 mW /g + 16.5 % (k=2)



 Body TSL parameters
    Thefollowing parameters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity

      Nominal Body TSL parameters                                  22.0 °C           52.7              1.95 mho/m
      Measured Body TSL parameters                              (22.0 £0.2) °C    52.4 + 6 %        1.98 mho/m + 6 %
       Body TSL temperature change during test                     <0.5°C             m

 SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
       SAR measured                                       250 mW input power                   12.7 mW /g
      SAR for nominal Body TSL parameters                     normalized to 1W        50.4 mW / g x 17.0 % (k=2)


       SAR averaged over 10 cm‘ (10 g) of Body TSL                condition
      SAR measured                                        250 mW input power                   5.92 mW /g
      SAR for nominal Body TSL parameters                     normalized to 1W        23.6 mW / g + 16.5 % (k=2)




 Cartificate No: D2450V2—053_May12                     Page 3 of 8


TME«                                                                                             e
 Appendix

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           53.7 0 +3.2 0
       Return Loss                                                                       ~26.4 dB



 Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                           49.9 Q + 4.8 Jn
       Return Loss                                                                       —26.4 dB



 General Antenna Parameters and Design

       Electrical Delay (one direction)                                                   1.1693 ns

 After long term use with 100W radiated power, only a stight warming of the dipole near thefeedpoint can be measured.

 The dipole is made of standard semirigid coaxia! cable. The center conductor of the feeding line is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
 are added to the dipole arms in arder to improve matching when loaded according to the position as explained in the
 *Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
 according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
 feedpoint may be damaged.


 Additional EUT Data

       Manufactured by                                                                    SPEAG
       Manufactured on                                                              November 10, 2009




 Certificate No: D2450V2—053_May12                        Page 4 of 5


TME«                                                                             L
 DASY5 Validation Report for Head TSL

                                                                                     Date: 02,05.2012

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 853

 Communication System: CW; Frequency: 2450 MHz
 Medium parameters used: f = 2450 MHz: 0 = 1.81 mho/m: &, = 39.6; p= 1000 kgf‘mj
 Phantom section: Flat Section
 Measurement Standard: DASY3 (IEEE/AIEC/ANSI C63.19—2007)

 DASY32 Configuration:

    «_   Probe: ES3DV3 — $SN3205; ConvF(4.45, 4.45, 4.45); Calibrated: 30.12.2011;

    «    Sensor—Surface: 3mm (Mechanical Surface Detection)

    +    Electronics: DAE4 Sn601; Calibrated: 04.07.2011

    «_   Phantom: Flat Phantom 5.0 (front); Type: QDOOOP5OAA; Serial: 1001

    «_   DASY32 52.8.1(838); SEMCAD X 14.6.5(6469)


 Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=3mm
 Reference Value = 100.0 V/m; Power Drift = 0.05 dB
 Peak SAR (extrapolated) = 26.785 mW/g
 SAR(I g) = 13.1 mW/g; SAR(10 g) = 6.09 mW/g
 Maximum value of SAR (measured) = 16.7 mW/g




         5.20


         10.40


         ~15.60


         —20.80


         —26.00
                   0 dB = 16.7 mW/g =24.45 dB mWi/g



 Certificate No: D2450V2—853_May12               Page 5 of 8


TME«                                                                                   e
 Impedance Measurement Plot for Head TSL



                                                                 2 May 2012 18110128
           CHM s1        i U Fs         i8a736a   3,.2309e   210.36 pH     2 450,000 008 MHz




           Del

           Ca




           &
           16


           H1d

           cha




           Avi
           16°

           Hd

                    st       250.000 600 MHz                         SToP 2 650.000 900 Mz




 Cartificate No: D2450V2—855_May12                 Page 6 of 8


TME«                                                                            e
 DASYS Validation Report for Body TSL

                                                                                    Date: 02.05.2012

 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2450 MHz; Type: D2450V2; Serial; D2450V2 — SN: 853

 Communication System: CW; Frequency: 2450 MHz
 Medium parameters used: f = 2450 MHz: a= 1.98 mho/m; s, = 52.4; p= 1000 kgfl'lfl‘
 Phantom section: Flat Section
 Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)

 DASY52 Configuration:

    «+   Probe: ES3DV3 — SN3205; ConvF(4.26, 4.26, 4.26); Calibrated: 30.12.2011;

    «    Sensor—Surface: 3mm (Mechanical Surface Detection)

    +    Electronics: DAE4 Sn601; Calibrated: 04.07.201 1

         Phantom: Flat Phantom 5.0 (back); Type: QDOOOPSOAA; Serial: 1002

    «—   DASY52 52.8.1(838); SEMCAD X 14.6.5(6469)


 Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 95.306 V/m; Power Drift = 0.00 dB
 Peak SAR (extrapolated) = 26.029 mW/g
 SAR(I g) = 12.7 mW/g; SAR(10 g) = 5.92 mW/g
 Maximum value of SAR (measured) = 16.8 mW/g




         ~4.80


         —9.60


         —14.40


         "19.20


         ~24.00
                   0 dB = 16.8 mW/g = 24.51 dB mW/g




 Cartificate No: D2450V2—853_May12               Page 7 of 8


TMEI                                                                                     h

 Impedance Measurement Plot for Body TSL




                                                                  2 May 2o12 aszooist
           CHD sin     a ues           1:49.896 6   4.8027 8. 11.99 pH       2 4se.000 ooo mHz


           Ded

           Ca




           ie
           16

           ma

           CHa2




           By
           159


           Hia

                     SthRt 2 250.000 000 Mz                           STuP 2 650.000 00 Miz




 Cartificate No: D2450V2—853_May12                   Page 8 of 8


                                                                       No. 2013SAR00065
                                                                        Page 151 of 151


ANNEX I        DIPOLE QUALIFICATION FOR THE EXTENDED 3-YEAR

                CALIBRATION INTERVAL

I1 Dipole 1900
The information and documentation below are provided to qualify the extended 3-year
calibration interval of dipole.


I1.1 List of Equipment
 No.                   Name                               Type                  Serial Number
  01             Network analyzer                        E5071C                 MY46110673
  02               Power meter                           NRVD                      102083
  03               Power sensor                          NRV-Z5                    100542
  04             Signal Generator                        E4438C                 MY49070393
  05                  Amplifier                          60S1G4                   0331848
  06               E-field Probe                   SPEAG EX3DV4                     3846
  07                    DAE                            SPEAG DAE4                   771
  08            Dipole Validation Kit              SPEAG D1900V2                    541


I1.2 Results of Impedance, Return-loss and System validation
Dipole 1900 - Head
                                                Year
                                                                    Deviation         Limit
                                        2012            2013
                     Real (Ω)           52.6             50.7        -1.9Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           6.2              2.5         -3.7Ω       Deviation < 5Ω
       Return-loss (dB)                 -23.7           -23.5        0.2dB      Deviate < 0.2dB
  System               10g              5.11             5.07        -0.78%     Deviation < 10%
 validation             1g              9.62             9.61        -0.10%     Deviation < 10%


Dipole 1900 - Body
                                                Year
                                                                    Deviation         Limit
                                        2012            2013
                     Real (Ω)           48.6             47.1        -1.5Ω       Deviation < 5Ω
 Impedance
                Imaginary (Ω)           6.9              3.3         -3.6Ω       Deviation < 5Ω
       Return-loss (dB)                 -23.0           -23.1        -0.1dB     Deviate < 0.2dB
  System               10g              5.33             5.46        2.44%      Deviation < 10%
 validation             1g               10              10.3        3.00%      Deviation < 10%


According to the above tables, it is not necessary to recalibration the dipoles in 2013.



Document Created: 2013-06-19 08:38:38
Document Modified: 2013-06-19 08:38:38

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC