RF Exposure SAR 3

FCC ID: RAD234

RF Exposure Info

Download: PDF
FCCID_1664135

TME                                                                                                   oo —          120C   —




 nefpaursttt‘~~ TVIRC
              &               3     se                ~

 Telecommunication Metrology Center of MIIT.

 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                        DASY5                                     V5.0

      Extrapolation                             Advanced Extrapolation

      Phantom                                   2mm Oval Phantom ELI4
      Distance Dipole Center — TSL                        15 mm                                  with Spacer

     Zoom Scan Resolution                         dx, dy, dz = 5 mm

      Frequency                                   835 MHz +1 MHz



 Head TSL parameters
   The following parameters and calculations were applied.
                                                          Temperature             Permittivity             Conductivity

      Nominal Head TSL parameters                           22.0°C                   41.5                  0.90 mho/m

     Measured Head TSL parameters                      (22.0 £0.2) °C              41.6 4 6 %          0.92mho/m     6%

     Head TSL temperature during test                  (21.7 20.2) °C                 oc                       «~




 SAR result with Head TSL

     SAR averaged over 1 C(1 g) of Head TSL                           Condition
      SAR measured                                                250 mW input power.                2.38 mW / g

     SAR normalized                                                normalized to 1W                  9.62 mW / g

     SAR for nominal Head TSL parameters ‘                         normalized to 1W         9.41 mW 1g + 17.0 % (k=2)

     SAR averaged over 10 Cn?" (10 g) of Head TSL                     Condition
     SAR measured                                                 250 mW input power                 1.54 mW /g

     SAR normalized                                                normalized to 1W                  6.16 mW !g

     SAR for nominal Head TSL parameters ‘                         normalized to 1W         6.12 mW ig £ 16.5 % (k=2)




        ‘ Correction to nominal TSL parameters according to d), chapter "SAR Sensitivities®

 Certificate No: D835V2—443_Feb10                   Page 3 of 9


TME                                                                                                  ~o—             —
                                                                                                                         w92




 naguercr<t~ TMVRC
              as              x    B                 N


 Telecommunication Metrology Center of MIIT


 Body TSL parameters
    The following parameters and calculations were applied.
                                                          Temperature             Permittivity        Conductivity

      Nominal Body TSL parameters                           22.0 °C                  55.2             0.97 mhoim

      Measured Body TSL parameters                       (22.0 £0.2) °C           54.5 2 6%         0.97mho/m + 6 %

      Body TSL temperature during test                   (21.920.2) °C                ——                  ~——




 SAR result with Body TSL

      SAR averaged over 1. Cm‘(1 g) of Body TSL                       Condition
      SAR measured                                              250 mW input power                241 mW /g

      SAR normalized                                              normalized to 1W                9.64 mW / g
      SAR for nominal Body TSL parameters *                       normalized to 1W          9.57 mW ig £ 17.0 % (k=2)


      SAR averaged over 10 cm°" (10 g) of Body TSL                    Condition
      SAR measured                                              250 mW input power                1.57 mW /g

      SAR normalized                                              normalized to 1W                6.28 mW / q

      SAR for nominal Body TSL parameters *                       normalized to 1W          6.24 mW /g + 16.5 % (k=2)




        * Correction to nominal TSL parameters according to d), chapter "SAR Sensitivities®

 Certificate No: D835V2—443_Feb10                   Page 4 of 9


TME

 TNLRMSMAb fiEfatB4 ut
 Telecommunication Metrology Center of MIIT
                                                               TME
  Appendix

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                      §3.70—3.7 |Q

      Return Loss                                                                   —25.90B



 Antenna Parameters with Body TSL

      impedance, transformed to feed point                                       49.40 — 5.1 j0

      Return Loss                                                                   —25.60B



 General Antenna Parameters and Design

      Electrical Delay (one direction)                                               1.387 ns                           :I


 Afterlong term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected
 to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near
 the feadpoint may be damaged.



 Additional EUT Data

      Manufactured by                                                                  SPEAG
      Manufactured on                                                            September 3, 2001




 Certificate No: D835V2—443_Feb10                    Page 5 of 9


TME

 Do+ c TUIRC
 Telecommunication Metrology Centerof MIIT


 Impedance Measurement Plot for Head TSL

        CHD san    i uons           4133.729 4.
                                              ~2.6096 0. 51.744 pF   #35,000 000 Mz
                                             onl

        a
                                                                                      OMA Rerkers
        vel                                                                            a sreed 4
        Con                                                                           sea.odo



        ty
        M




                                                                                      CH2 Markern

                                                                                      TebtbErcte




 Certificate No: D835V2—443_Feb10                  Page 7 of9


TME                                                                                   y



 netMebicta&+t~ TMIRC
 Telecommunication Metrology Center of MIIT

 DASY5 Validation Report for Head TSL

                                                                       Date/Time: 2010—2—26 14:31:40
 Test Laboratory: TMC, Beijing, China

 DUT: Dipole 835 MHz; Type: D835V2; Serial: SN: 443
 Communication System: CW Frequency: 835 MHz Duty Cyele: 1:1
 Medium: Head S35MHz
 Medium parameters used: f = 835 MHz; 0 = 0.92 mho/m; +. =41.6; p = 1000 kg/m‘
 Phantom section: Flat Section

 DASY5 Configuration:
    ®@  Probe: ESIDV3—SN3149; ConvF(6.56, 6.56, 6.56); Calibrated: 25.09.09
    ® Electroni¢s: DAE4 Sn771; Calibration: 19.11.09
     ®   Phantom: 2mm Oval Phantom EL14; Type: QDOVAOOIBB
     ®   Mcasurement SW: DASY5, ¥5.0 Build 119.9; Postprocessing SW: SEMCAD, V13.2 Build 87



 Pin=250mW; d=15mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=5mm, dy=5mm, dz=5mm
 Reference Value = 54.8 V/m; Power Drift = —0.037 dB
 Peak SAR (extrapolated) = 3. 11 W/kg
 SAR(L g) =2.38 mW/g; SAR(IO g) =1.54 mWl/g
 Maximum value of SAR (measured) = 2. 71 mW/g
          dB
         [ 0.000


           —2.24


           —4.48


           —6.72


           —0.96


           11.2
                      0 dB = 2. TimW/g




 Certificate No: D835V2—443_Feb10                Page 6 of 9


TME«
 netnestko TMIRC
 Telecommunication Metrology Center of MIIT



 DASYS Validation Report for Body TSL

                                                                        Date/Time: 2010—2—26 9:52:36
 Test Laboratory: TMC, Beijing, China
 DUT: Dipole 835 MHz; Type: D835V2; Serial: SN: 443
 Communication System: CW Frequency: 835 MHz Duty Cycle: 1:1
 Medium: Body 835MHz
 Medium parameters used: £ = 835 MHz; 0 =0.97 mho/m; £,= 54.5; p = 1000 kg/m‘
 Phantom section: Flat Section

 DASY5 Configuration:
    ®   Probe: ESSDV3 ~SN3149; ConyF(6.22, 6.22, 6.22); Calibrated: 25.09.09
    ®     Electronics: DAE4 Sn771; Calibration: 19.11.09
    ®     Phantom: 2mm Oval Phantom ELI4; Type: QDOVAOOIBB
     ©@   Mcasurement SWDASY3, V5.0 Build 119.9; Postprocessing SW: SEMCAD, ¥13.2 Build 87



 Pin=250mW; d=15mm/Zoom Scan (7x7x7)/Cube 0:
 Measurement grid: dx=bmm, dy=5mm,.            dz=5mm
 Reference Value = 54.0 ¥/m; Power Drift = —0.025 dB
 Peak SAR (extrapolated) = 3. 78 W/kg
 SAR(L g) =2.41 mW/g; SAR(IO g) = 1.57 mW/g
 Maximum value of SAR (measured) = 2.70 mW/g
            dB
          a 0.000


            214


            ~4.28


            —6.42


            —8.56


            10.7
                       0 dB = 2. TOm/g



 Certificate No: D835V2—443_Feb10                Page 8 of 9


TME                                                                                  exv e ev re narrvieivied
                                                                                        ~a~—
                                                                                                             1




 nhadurstsHtto TMIRC
 Telecommunication Metrology Center of MIIT



 Impedance Measurement Plot for Body TSL

        EHR sin    a uce           a149.4990 ~10684e 37.607 P       £35,000 000 nz
                                            9e
        3                                        t woascth
                                                 e        7    3A                       @4 Narkens
        ter                                r         aies Tnniks                         m §1.523 0
                                                                                        900.




                                                                                       Ch2 Narkers




Certificate No: D835V2—443_Feb10               Page 9 of9


                                          "#$!%&'%()*&&&%+
                     !                     ,-./!''0!#1!'23!

$
!"##$IJK$_&W.N+$*3N&@23)&.6$*+2)&X&53)+


TMEC
 DHegiMbet@&t~ TMRRC
 Telecommunication Metrology Center of MIIT




 Glossary:
 TSL                   tissue simulating liquid
 ConvF                 sensitivity in TSL / NORMx.y.z
 N/A                   not applicable or not measured


 Calibration is Performed According to the Following Standards:
 a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
    Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
      Measurement Techniques", December 2003
 b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held devices
    used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February 2005
 ©) Federal Communications Commission Office of Engineering & Technology (FCC OET),
    "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
    Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
    Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
    Supplement C (Edition 01—01) to Bulletin 65


 Additional Documentation:
 d) DASY System Handbook


 Methods Applied and Interpretation of Parameters:
 * Measurement Conditions: Further details are available from the Validation Report at the end of
    the certificate. All figures stated in the certificate are valid at the frequency indicated.
 * Antenna Parameters with TSL: The dipole is mounted with the spacerto position its feed point
    exactly below the center marking of the flat phantom section, with the arms oriented parallel to
    the body axis.
 «_    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
       positioned under the liquid filled phantom. The impedance stated is transformed from the
       measurement at the SMA connector to the feed point. The Return Loss ensures low reflected
       power. No uncertainty required.
 *     Electrical Delay: One—way delay between the SMA connector and the antenna feed point. No
       uncertainty required.
 *     SAR measured: SAR measured at the stated antenna input power.
 *     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
       connector.
 *     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
       nominal SAR result.




 Certificate No: D1900V2—541_Feb10          Page 2 of 9


TMC«                                                                                                   ~a—




 DeisMciestst&t+~TVR
     37   y fosy L. %

 Telecommunication Metrology Center of MIIT

 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                      DASYS                                       v5.0

      Extrapolation                              Advanced Extrapolation

      Phantom                                    2mm Oval Phantom ELH4

      Distance Dipole Center — TSL                      10 mm                                    with Spacer

      Zoom Sean Resolution                         dx, dy, dz = 5 mm

      Frequency                                    1900 MHz & 1 MHz




 Head TSL parameters
    The following parameters and calculations were applied.
                                                         Temperature              Permittivity             Conductivity
      Nominal Head TSL parameters                          22.0°C                    40.0                  1.40 mhotm

      Measured Head TSL parameters                      (22.0 £0.2) °C             39.6 16 %           1.40mhofm x 6 %

      Head TSL temperature during test                  (21.9 £0.2) °C                ——                       ——




 SAR result with Head TSL

      SAR averaged over 1 cim" (1 g) of Head TSL                      Condition
      SAR measured                                                250 mW input power                 9.91 mW/g

      SAR normalized                                               normalized to 1W                  39.6 mW /g

      SAR for nominal Head TSL parameters                          normalized to 1W         39.4 mW /g #17.0 % (k=2)

      SAR averaged over 10 C/7"(10 g) of Head TSL                    Condition
      SAR measured                                                250 mW input power                 5.06 mW / g
      SAR normalized                                               normalized to 1W                  20.2 mW !g

      SAR for nominal Head TSL parameters‘                     _ normalized to 1W           20.1 mW Ig & 16.5 % (k=2)




         ‘ Correction to nominal TSL parameters according to d), chapter "SAR Sensitivities®

 Certificate No: D1900V2—541_Feb10                  Page 3 of 9


TME                                                                                                 w




 Defaussittf&t~ TMRC
               s               a    3e x              N


 Telecommunication Metrology Center of MIIT



 Body TSL parameters
    The following parameters and calculations were applied.
                                                          Temperature            Permittivity        Conductivity

      Nominal Body TSL parameters                          22.0°C                   53.3             1.52 mholm

      Measured Body TSL parameters                     (22.0 £0.2) °C             52.5 26%         1.51 mhoim 2 6 %

      Body TSL temperature during test                 @18202)°C


 SA R result with Body TSL

      SAR averaged over 1 C/N" (1 g) of Body TSL                     Condition
      SAR measured                                               250 mW input power              10.4 mW /g

      SAR normalized                                             normalized to 1W                41.6 mW / g

      SAR for nominal Body TSL parameters *                      normalized to 1W          41.4 mW ig & 17.0 % (k=2)


      SAR averaged over 10 Cmm (10 g) of Body TSL                   Condition
      SAR measured                                               250 mW input power              §.24 mW /g

      SAR normalized                                             normalized to 1W                21.0 mW /g

      SAR for nominal Body TSL parameters "                      normalized to 1W          20.9 mW /g £ 16.5 % (k=2)




        * Correction to nominal TSL parameters according to d), chapter "SAR Sensitivities®
 Cert ificate No: D1900V2—541_Feb10                Page 4 of 9


T ME                                                                                         tm   fag—       c—    —      —




 nefiuessittit+MR
 Telecommunication MetrologyCenter of MIIT




  Appendix

 Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                       54.80 + 4.0 jQ

      Return Loss                  —                                               —23.70B



 Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                      47.90 + 7.1 j0

      Return Loss                                                                  —22.6dB



 General Antenna Parameters and Design

      Electrical Delay (one direction)                                              1.201 ns


 After longterm use with 100W radiated power, only a slight warmingof the dipale near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected
 to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals.
 No excessive force must be applied to the dipole arms, because they might bend or the saldered connections near
 the feedpoint may be damaged.



 Additional EUT Data

      Manufactured by                                                                 SPEAG

      Manufactured on                                                             Octaber 4, 2001




 Certificate No: D1900V2—541_Feb10                  Page 5 of 9


TM

DHKMHest&+~                                             TMUMK
Telecommunication Metrology Center of MIIT

DASYS5 Validation Report for Head TSL

                                                                      Date/Time: 2010—2—26 15:20:47
Test Laboratory: TMC, Beijing, China

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: SN: 541
Communication System: CW Frequency: 1900 MHz Duty Cycle: 1:1
Medium: Head 1900MHz
Medium parameters used: £ = 1900 MHz; 0= 1. 40 mho/m;                ¢,= 39.6; o = 1000 kg/m‘
Phantom section: Flat Section

DASYS5 Configuration:
        Probe: ES3DV3~SN3149; ConvF(6.03, 5.03, 5.03); Calibrated: 25.09.09
        Electronies: DAE4 Sn771; Calibration: 19.11.09
        Phantom: 2mm Oval Phantom EL14; Type: QDOVAOOIBB
        Measurement SW: DASYS, V5.0 Build 119.9; Postprocessing SW: SEMCAD, V13.2 Build 87



Pin=250mW; d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=omm,           dy=bmm, dz=5mm
Reference Value = 85.1 V/m; Power Drift = —0.057 dB
Peak SAR (extrapolated) = 18. 8 W/kg
SAR(L g) = 9.91 mW/g: SAR(IO g) = 5.05 mW/g
Maximum value of SAR (measured) = 11.5 mW/g
          d8
           0.000


          —3.46


          6.92


          —10.4


          ~13.8


          17.3
                     0 dB = 11. Smit/g




Certificate No: D1900V2—541_Feb10             Page 6 of 9


TME                                                                               - o v_ic. ..“,-..'._J
                                                                                      fago    ce — ~6   l>




 DERAUHS SBAkROHs t&4 C
              is            *    &      —A
                                                            I M,:‘
 Telecommunication Metrology Center of MIIT


 Impedance Measurement Plot for Head TSL

         EHD s     i urs         m 84609 a   $7676 0.   40343 pH   1 200.000 000 mz


                                                                                      Cit Mankens
                                                                                       u 46.250 6
                                                                                                @




                                                                                      CH2 Narkers
                                                                                      ‘Lessad




 Certificate No: D1900V2—541_Feb10            Page 7 of 9


TM         ' ‘
            "Alh,
                                                                                        ee. ce reve wrvvved
                                                                                          .    tg—   ce n ye   es   |

NeMsMsEst&t+                                               TMKC
Telecommunication MetrologyCenter of MIIT



DASYS5 Validation Report for Body TSL

                                                                        Date/Time: 2010—2—26 10:41:08
Test Laboratory: TMC, Beijing, China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: SN: 541
Communication System: CW Frequency: 1900 MHz Duty Cyele: 1:1
Medium: Body 1900MHz
Medium parameters used: f = 1900 MHz; o0= 1. 51 mho/m; £,=52.5; p =1000 kg/m‘
Phantom section: Flat Section

DASY5 Configuration:
   ®   Probe: ES3DV3— SN3149; ConvF(4.68, 4.68, 4.68); Calibrated: 25.09.09
    ®     Electronics: DAE4 Sn771; Calibration: 19.11.09
    ®@—   Phantom: 2mm Oval Phantom EL14; Type: QDOVAOOIBB
    ®     Measurement SW: DASYS, V5.0 Build 119.9; Postprocessing SW: SEMCAD, V13.2 Build 87




Pin=250mW; d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=omm,           dy=5mm,    dz=5mm
Reference Value = 80.2 V/m; Power Drift = —0.009 dB
Peak SAR (extrapolated) = 19.1 W/kg
SAR(I g) = 10.4 mW/g; SAR(IO g) = 5.24 mW/g
Maximum value of SAR (measured) = 12.0 mW/g
            dB
            0.000


            —3.46


            5.92


            —10.4


            —13.0


            17.3

                      0 dB = 12. OnW/g



Certificate No: D1900V2—541_Feb10               Page 8 of 9


TME

 meueguesitint— ME
 Telecommunication Metrology Center of MIIT




 Impedance Measurement Plot for Body TSL

        EB sn     a vors        2 47.564 o 7.0090 a 807.48 pH

                                                                CHL Narkere
                                                                 auuu:




        chr



                                                                Of2 Narkers
                                                                ar      gb
                                                                4        c




        o




 Certificate No: D1900V2—541_Feb10           Page 9 of 9


TMEI                                                                                                                   \
snn e n s




   Calibration Laboratory of                                          ‘Xw”"                                            Schweizerischer Kaltbrierdienst
   Schmid & Partner                                                  i               &                                 Service sulsse d‘étaionnage
      Engineering AG                                                 T taf                                             Servizio avizzero di taratura
   Zeughnusstrasse 43, 8004 Zurich, Switzerland                      1.4’,‘7“\"\\\“\3‘                                 Swiss Calibration Service


    Accredited by the Swiss Accreditation Service (SAS)                                               Accreditation No.: SCS 108
    The Swiss Accreditation Service is one ofthe signatories to the EA
    Multilateral Agreement for the recognition of calibration certificates




    Object


    Calibration procedura(s)




    Callbration dato:




    This callbration certificate documents the traceabilty to national standards, which realtze the physical units of moasuraments (S1).
    The measurements and the uncertainties wih confidence probability are given on the following pages and are part of the cortificate.

    All calibrations have been canduated in the closed Iabratory fecilty: environment temperature (22 3)°C and humnidity < 70%.


    Calibration Equipmont used (M&ATE oriical for callbration)


     Primary Standards                      ID #                         Cal Date (Gertificate No.)                        Schoduled Calibration
     Power mator EPM—442A                   GB37480704                   06—0ct—08 (No, 217—01086)                         Oct10
     Power sensor HP 8481A                  US37202703                   06—0ct—09 (No. 217—01086)                         Oct—10
     Referance 20 dB Atenuatar              h: soe (20g)                 30—Mar—10 (No. 217—01158)                          Mar11
     Type—N mismaich combination            SN: 5047.2 / 06327           30—Mar—10 (No. 217—01162)                          Mart1
     Reference Probe ESSDV3                 SN: 3205                     30—Apr—10 (No. ES$—3205_Apr10)                    Apr=11
     DAE4                                   SN: 601                      10—Jun—10 (No, DAEA—601_Jun10)                    Jun—1

     Secondary Standards                    iD #                         Check Date (in house)                             Scheduled Chack
     Power sensor HP B4B1A                  MY41000817                   18—0ct—02 (in house check Oct—09)                  in house check: Oct—11
     RAF generator R&S SMT—06               100008                       4—Aug—99 (in house chock Oct—09)                   In house chack: Oct—11
     Network Analyzer HP 8753E              US37300505 $4206             18—Oct—01 (in house check Oct—00)                  In house chack: Oct—10

                                            Name                                     Function                               Signature
     Calibrated by:                                                                      i       Techn


     Approved by:



                                                                                                                            Issued: Soptember29, 2010
     This callbration certificate shall not be reproduced except in full without written approval of the laboratory.


    Certificate No: D2450V2—853_Sap10                                       Page 1 of 9


TMC
                                                                               N‘




  &
 Callbration Laboratory of                                     se
                                                              sls              g_ Schweizerischer Kalibrierdienst
 Schmid & Partner                                            M                 c    Service sulsse d‘étalonnage
   Engineering AG                                            e                      Servizio svizzero di taratura
 Zeughausstrasse 43, 8004 Zurich, Switzerland                 @,‘If,\\\‘\\?'   S    swigs Callbration Service
                                                                 "tntal?


 Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
 The Swiss Accreditation Service Is one of the signatories to the EA
 Multilateral Agroement for the recognition of calibration certificates


 Glossary:
 TSL                         tissue simulating liquid
 ConvF                       sensitivity in TSL / NORM x,y,z
 N/A                         not applicable or not measured

 Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", December 2003
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) Federal Communications Commission Office of Engineering & Technology (FCC OET},
       "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
       Electromagnetic Fields; Additional Information for Evaluating Compliance of Motbile and
       Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
       Supplement C (Edition 01—01) to Bulletin 65

 Additional Documentation:
    d) DASY4/5 System Handbook

 Methods Applied and Interpretation of Parameters:
    * Measurement Conditions: Furtherdetails are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
          point exactly below the center marking of the flat phantom section, with the arms oriented
          parallel to the body axis.
     «_   Feed Point Impedance and Return Loss: These parameters are measured with the dipole
          positioned under the liquid filled phantom. The impedance stated is transformed from the
          measurement at the SMA connectorto the feed point. The Returm Loss ensures low
          reflected power. No uncertainty required.
     *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
          No uncertainty required.
     *    SAR measured: SAR measured at the stated antenna input power.
     «_   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
          connector.
     «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
          nominal SAR result.


 Certificate No: D2450V2—853_Sep10                               Page 2 of 0


TME                                                                                     P




 Measurement Conditions
    DASY system confiquration, as far as not given on page 1.
      DASY Version                                              DASYS                               v62.2
      Extrapolation                                   Advanced Extrapolation
      Phantom                                       Modular Flat Phantom V5.0
      Distance Dipole Center — TSL                              10 mm                            with Spacer
      Zoom Scan Resolution                              dx, dy, dz =5 mm
      Frequency                                         2450 MHz + 1 MHz



 Head TSL parameters
    The following paramaters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity
      Nominal Head TSL parameters                                  22.0°C            39.2              1.80 mho/m
      Measured Head TSL parameters                              (22.0 £ 0.2) °C   39.0 1 6 %        1.74 mho/m a 6 %
      Head TSL temperature during test                          (21.5 20.2) °C        se                     ~———



 SAR result with Head TSL

      SAR averaged over 1 em‘ (1 g) of Head TSL                  Condition

      SAR measured                                        250 mW input power                   13.1 mW /g
      SAR normalized                                          normalized to 1W                 §2.4 mW /g
     LEAH for nominal Head TSL parameters                     normalized to 1W        53.2 mW /g = 17.0 % (k=2)


      SAR averaged over 10 am* (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                   6.16 mW /g              |
      SAR normalized                                          normalized to 1W                 24.6 mW / q
      SAR for nominal Head TSL parameters                     normalized to 1W        24.8 mW 1g « 16.5 % (=2)




 Certificate No: D245OV2—853_Sep10                     Page 3 of 9


T ME
                                                                                        p;          Lan     ce ama




 Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity
      Nominal Body TSL parameters                                  22.0°C            52.7              1.95 mho/m
     7Measurad Body TSL parameters                              (22.0 + 0.2) °C   52.5 £ 6 %        1.95 mho/m £ 6 %
     jBndy TSL temperature during test                          (21.6 £0.2) °C


 SAR result with Body TSL

      SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
      SAR measured                                        250 mW input power                   12.9 mW /g
      SAR normalized                                          normalized to 1W                 §1.6 mW /g
      SAR for nominal Body TSL parameters                     normalized to 1W       51.5 mW / g + 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                        250 mW input power                   5.98 mW /g
      SAR normalized                                          normalized to 1W                 23.9 mW /g
      SAR for nominal Body TSL parameters                     normalized to 1W       23.9 mW / g + 16.5 % (k=2)




 Certificate No: D2450V2—859_Sep10                     Page 4 of 9


TME
 Appendix

 Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           54.6 0 + 2.8 O
       Retum Loss                                                                        —25.8 dB


 Antenna Parameters with Body TSL

       impedance, transformed to feed point                                           49.4 Q + 4.4 JC

       Return Loss                                                                       —27.1 dB


 General Antenna Parameters and Design

     Electrical Detay (one direction}                     f                               1.164 ns                        _]
 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding lina is directly connected to the
 second arm of the dipole. The antenna is therefore short—circuited for DC—signals.
 No excessive force must be applied to the dipole arms, because thay might bend or the soldered connections near the
 feedpoint may be damaged.


 Additional EUT Data

       Manufactured by                                                                    SPEAG
       Manufactured on                                                              November 10, 2009




 Cortificate No: D2450V2—853_Sep10                        Page 5 of 9


TME                                                                                       .
   DASYS Validation Report for Head TSL

                                                                                   Date/Time: 24.09.2010 14:10:17
   Test Laboratory: SPEAG, Zurich, Switzerland

   DUT: Dipole 2450 MHz;             Type: D2450V2; Serial: D2450V2 — SN:853
  Communication System: CW; Frequency: 2450 MHz;
                                                 Duty Cycle: 1: 1
  Medium: I      U12 BB
  Medium        eters used: f = 2450 MHz: 0 = 1.74 mho/m: &¢=39
                                                                ; p = 1000 kg/m’
  Phantom section: Flat Section
  Measurement Standard: DASYS5 (IE     IIBC/ANST C63.19—2007)
  DASY3 Configuration:

      *   Probe: ES3DV3 ~SN3205; ConvB(4.53, 4 53, 4.53); Calibrated:
                                                                      30.04.2010
      *   Sensor—Surface: 3mm(Mechanical Surface Detection)

      *   Electronics: DAEA Sn601; Calibrated:   10.06.2010

      *   Phantom: Flat Phantom 5.0 (front); Type: QDOUOPSOAA: Serial: 1001

      *   Measurement SW: DASY32, V52.2 Build.0, Version
                                                         52,2,0 (163)

      *   Postprocessing SW: SEMCADX, V14.2 Build
                                                  2, Version 14.2.2 (1685)



 Pin=250 mW/d=10mm, dist=3.0mm (ES—Probe)
                                          /Zoom Scan (7x7x7)/Cube 0: Measurement
 grid: dx=5mm, dy=Smm, dz=5mm
 Reference Value = 101.7 V/m; Power Drift = 0.028
                                                  dB
 Peak SAR (extrapolated) = 26.7 W/kg
 SAR(I g) = 13.1 mW/g; SAR(10 8) = 6.16 mWig
 Maximum value of SAR (measured) = 16.7 mW/g




                                    0 dB = 16. 7mWi/g

Certificate No: D2450v2—053_Sap10
                                                              Page 6 of 9


TME

  Impedance Measurement Plot for Head TSL



                                                              24 Sep 2010    duris:es
       CHA)   s    1 uore           3: 54.596 n   27832 0 100.90 pH         2450.000 000 MiHz
                                                    T s

                                                                                                CHL Markens
                                                                                                  4t §2.057 a
                                                                                                     17.99 a
                                                                                                 2.55000 GHz




      fvg
      16



       cH2 san    LoG        5 dB/REF —20 a8 _


                                                                                                CH2 Mankens
                                                                                                4i=44.291 dn
                                                                                                 2.55000 GHz



      fre
      18°




Certificate No: D2450V2—853_Sep10                     Page 7 of 9


TME                                                                                            —
   Validation Report for Body

                                                                                        Date/Time: 27.09.2010 13:39:49
   Test Laboratory: SPEAG, Zurich, Switzerland

   DUT: Dipole 2450 MHz; Type: D24SOV2; Serial
                                               : D2450V2 — SN:853
   Communication System: CW:; Frequency: 2450
                                              MHz; Duty Cycle: 1:1
   Medium: MSL U12 BB
   Medium parameters used: f = 2450 MHz: o = 1.95
                                                  mho/m; £, = 52.6; p = 1000 kg/m>
   Phantom section: Flat Section
   Measurement Standard: DASY3 (I               HIBEC/ANST C63.19—2007)
  DASY3 Configuration:

       *   Probe: ES3DV3 — SN3205; ConvF(4
                                           31, 4.31, 4.31); Calibrated:    30,04.2010
      *    Sensor—Surface: 3mm (Mechanical Surface
                                                   Detection)
      *    Electronics: DAE4 Sn601; Calibrated: 10.06.2010

      *    Phantom: Flat Phantom 5.0 (backy; Type: QDOOOPS5OAA;
                                                                Serial: 1002
      +/   Mcasurement SW: DASY52, V52.2 Build
                                               0, Version 52.2.0 (163)
      *    Postprocessing SW: SEMCAD X, V14.2 Build
                                                    2, Version 14,2.2 (1685)


  Pin=250 mW/d=10mm, dist=3.0mm (ES—Prob
                                         e)/Zoom Scan (7x7x7)/Cube 0: Measurement
 grid; dx=5mm, dy=5mm, dz=Smm
 Reference Value = 95.7 V/m; Power Drift=0.045
                                               dB
 Peak SAR (extrapolated) = 27 Wikg
 SAR(I g) =12.9 mW/g; SAR(10 g) = 5.98 mW/g
 Maximum value of SAR (measured) = 16.9 mW/g




                                     0 dB = 16.0mWig




Certificate No: D2450V2—853_Sep10
                                                             Page 8 of 9


TME                                                                                                        —
        Impedance Measurement Plot
                                   for Body TSL




                EHT sii       i uore                                         27 Sep 2010 oorgarse
                                                     #t49.499 0     43000 0.
                                                                    smy 284.04 pH       2 450.000 co8 mz


             9e%                                                                                               CH4 Nankers
             m                                   f              +                                               5 e0.o16 a
                                                                                                                   26447 a
                                                                                                               289000
                                             {              i
                                                            omcs                                                       Giz


            five
                2                                                               tf
                                                                        f
              :
                                                                       fus
            cn2 sit           ros        5.dB/REF —20 ds _
                                       I         |          |



           ca                                                                                              CH2 Mank ons
                                                                                                           s—a0.023 as
                                                                                                           255000 Oz


           is




Certificate No:     D2450V;     53_Sop10
                                                                      Page 5 of 9



Document Created: 3580-07-24 00:00:00
Document Modified: 3580-07-24 00:00:00

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC