2450 Calibration Report

FCC ID: R9C-X909

RF Exposure Info

Download: PDF
FCCID_1894330

  COMOSAR Dipole 2450 MHz                                             SQfi'iMD     A
          Calibration Report                                      R.,;:;;;:SaTax
                                                      Page: 1/6       Issue: B        Date: 2012/10/05




         DIPOLE 2450 MHZ CALIBRATION REPORT


Prepared By:                     LUC Jérdme, SATIMO

Project Description:            SAR TEST BENCH

Prepared For (End User):        Shenzhen Morlab Communication Technology




This document is issued by SATIMO, in confidence and is not to be reproduced in whole or in part
without the prior written permission. The information contained herein is to be used only for the
purpose for which it is submitted and is not to be released in whole or in part without the prior written
permission of SATIMO.


  COMOSAR Dipole 2450 MHz;                                              Goud
                                                                   S(A"I'IM
                                                                   nemnmme nsm
           Calibration Report                                   Ref ChzadTeSkTeA
                                                    Page: 276      Issue: B        Date: 2012/10/05



                                  DIPOLE 2450 MHz CALIBRATION REPORT


DATE: 19/02/2009

REFERENCE: SN 36/08 DIPJ103

OBJECT: COMOSAR IEEE REFERENCE DIPOLE

MANUFACTURER: SATIMO

SERIAL NUMBER: SN 36/08 DIPJ103

CUSTOMER: Shenzhen Morlab Communication Technology

CONTRACT: PF2130108b_SAR_Morlab

DATE OF CALIBRATION: 05/10/2012


WARRANTY:
      This Calibration certificate may not be reproduced other than in full. Calibration certificates
without signature and seal are not valid. This documentation contains property information which is
protected by copyright. All right are reserved. No part of this document may be photocopied,
reproduced without the prior written agreement of SATIMO. SATIMO shall not be liable for errors
contained herein or for incidental or consequentialin connection with the furnishing, performance or
use of this material. Warranty doesn‘t apply to Normal wear, Normal tear, Improper use, Improper
maintain, Improperinstallation.


  Date


                 COit. 8 , jor2                                            MO Bretagne
                                                                      ksg,“mpue Brest Iroise
                                                                              Zone du Vernis
  SAR TEAM %                                                          225 rue Pierre
                                                                              29200 BREST


COMOSAR Dipole 2450 MHz                                                (I!
                                                                     _E'._‘_\I_iMo

        Calibration Report                                       EteResbrageAten
                                                     Page: 376       Issue: B        Date: 2012/10/05


[                                     PRODUCT DESCRIPTION




                              Dimension: L=51.5 mm/ h=30.4mm / d=3.6 mm


[_                                 CALIBRATION TEST EQUIPMENT                                       _]

                  Type                   IDENTIFICATION               DATE OF CALIBRATIO
         Vector Network Analyzer     HP8753D (SN: 5410A08882)              09—12—2012

[                                   MEASUREMENT PROCEDURE
     We placed the dipole under the flat part of SAM phantom fill with 2450 MHz head and body liquid.


     COMOSAR Dipole 2450 MHz                                                                  (1,.\
                                                                                          SATIMO
                                                                                          sermmunnicurem .
                  Calibration Report                                                 Ref: CR—280—7—08—SATB—A
                                                                      Page: 4/6          Issue: B            Date: 2012/10/05




                Calibration was performed according to IEEE Std P1528—2003 and OET bulletin 65 Supplement C (Ed. 01—01)

[_                                          VSWR at 2450 MHz: —33.09 dB.


                                                         Frequency (GHz)
      l6 & 8 &R § & 3 4, o




                                                                                                                  2,55
              S11 (aB)


COMOSAR Dipole 2450 MHz                                                  (Ge., \
                                                                      SATIM
                                                                      nesnme mm mm ."
     Calibration Report                                           Ref: CR—280—7—08—SATB—A
                                                     Page: 5/6        Issue: B          Date: 2012/10/05


                      SAR MEASUREMENT EQUIPEMENT Head

    Voltmeter                 Keithley (2000, SN:1000572)             Date of calibration: 01—07—2012
 Signal generator     Rohde&Schwarz (SML_O3, SN:101868)               Date of calibration: 15—11—2011
 Power amplifier            Nuclétudes (ALB216, SN:10800)             Date of calibration: 24—10—2011
  Power meter          RohdeSchwarz (NRVD, SN:101066)                 Date of calibration: 04—07—2012
                              SATIMO Bretagne (SN:EP37)
      Probe                                                           Date of calibration: 04—10—2012
                               CF (39.563,33.614,37 .677)

                            SAR MEASUREMENT CONDITION

         Software                                           OpenSAR V3
         Phantom                          SATIMO Bretagne (SN: SN_20_07_SAM42)
                                         SATIMO Bretagne (Last Calibration: 05—10—12)
          Liquid
                                         Head Liquid Values: eps‘ : 39,20 sigma : 1,80
   Distance between the
  centerof the dipole and
                                                              10 mm
   the liquid (set with a
          spacer)
   Area scan resolution                                dx=8mm/dy=8mm
  Zoom scan resolution                             dx=8mm/dy=8m/dz=5mm
        Frequency                                            2450 MHz
       Input power                                            30 dBm
   Expanded uncertainty
                                                              8.09%
           (K=1)

                              SAR MEASUREMENT RESULT


                                        109                      19
                SAR measured            22,82 Wikg               50,45 Wikg
                Liquid : HL             —4,92%                   —3,72%
                Input power : 1W


                            |
                                              (
COMOSAR Dipole 2450 MHz     ‘                   [i_,
                                             saAarimo
   Calibration Report                    Ref: CR—280—7—08—SATB—A
                            |_Page: 66     _ | Issue: B   Date: 2012/10/05
                   SAR MEASUREMENT PLOTS_


COMOSAR Dipole 2450 MHz                                                    [d
                                                                      SATIMO
                                                                      nenmmenmmemen /—
     Calibration Report                                           Ref: CR—280—7—08—SATB—A
                                                     Page: 5/6        Issue: B        Date: 2012/10/05


                     SAR MEASUREMENT EQUIPEMENT Body

    Voltmeter                 Keithley (2000, SN:1000572)             Date of calibration: 01—07—2012
 Signal generator     Rohde&Schwarz (SML_O3, SN:101868)_|             Date of calibration: 15—11—2011
 Power amplifier          Nuclétudes (ALB216, SN:10800)               Date of calibration: 24—10—2011
  Power meter          Rohde&Schwarz (NRVD, SN:101066)                Date of calibration: 04—07—2012
                              SATIMO Bretagne (SN:EP37)
                                                                      Date of calibration: 04—10—2012
                               CF (39.772,33.946,37.835)

                          SAR MEASUREMENT CONDITION

         Software                                        OpenSAR V3
         Phantom                         SATIMO Bretagne (SN: SN_20_07_SAM42)
                                        SATIMO Bretagne (Last Calibration: 05—10—12)
          Liquid
                                         Body Liquid Values: eps‘ : 52,50 sigma     1,78
   Distance between the
  center of the dipole and
                                                             10 mm
    the liquid (set with a
          spacer)
   Area scan resolution                                dx=8mm/dy=8mm
  Zoom scan resolution                            dx=8mm/dy=8m/dz=5mm
        Frequency                                           2450 MHz
       Input power                                           30 ddm
   Expanded uncertainty
                                                             8.09%
           (K=1)

                              SAR MEASUREMENT RESULT                                             _I


                                        109                      19
                SAR measured            23,38 Wikg               53,59 Wikg
                Liquid : BL             ~2,58 %                  +2,27 %
                Input power : 1W


COMOSAR Dipole 2450 MHz                ([      i
   Calibration Report     p— ~~ RefcR2sorossateA _                  _
                                  | Issue: B       |   Date: 201271005