DIPOLE 1750MHz CALIBRATION REPORT

FCC ID: R9C-X9006

RF Exposure Info

Download: PDF
FCCID_2238452

                                               ((.
                                    SATIMO
                                    The micrmare vision conpany..~




      SAR Reference Dipole Calibration Report


                                            Ref: ACR.269.3.13.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
           TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
               LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
   SATIMO COMOSAR REFERENCE DIPOLE
                       FREQUENCY: 1750 MHZ
                       SERIAL NO.: SN 30/13 DIP1G750—260

                            Calibrated at SATIMO US
                   2105 Barrett Park Dr. — Kennesaw, GA 30144


                                         «n
                                          ACCREDITED
                                         Calbanion CHAT eazue in


                                          09/25/2013

    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
    to national metrology institutions


   (L.. \                 SAR REFERENCE DIPOLE CALIBRATION REPORT                                                 Ref: ACR 260.3.13.SATU.A
BC-\TIN’ID



                                  Name                              Function                          Date                  Signature
Prepared by :                Jérome LUC                       Product Manager                    9/26/2013                   fi
Checked by :                 Jérome LUC                        Product Manager                   |9/26/2013                  fi/
Approved by:            Kim RUTKOWSKT                          Quality Manager                   9/26/2013                j..,, mathoush




                                                                       Customer Name
                                                                      Shenzhen Morlab
                                                    ;                 Communications
                                      Distribution:                   Technology Co.,
                                                                                Ltd




   Issue                    Date                                                       Modifications
     A                  9/26/2013                 Initial release




                                                             Page: 2710
                This document shall not be reprodced, except infull or in part, without the written approval ofSATIMO
                The information containedherein i to be used onlyfor the purposefor which it is submitted and is not to
                be releasedin whole orpart without written approval ofSATDAO,


  (6. \                     SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.260.3.13 SATU A
SKI'IN,ID


                                                   TABLE OF CONTENTS



1.     TMEEOUUOHON::smmmemsmemmrnicreynniiiitneeiteOittercemsverremmsemsnmmmmessnstmmeisriescereee 4
2.     D@ViC@ UMG@T TSb ...........csssssssmmmmmmmmmmmmmmmmnnnmnmmnsernnnmmmmmemmmssmmmmmmnmnnmmmmncen 4
3      PFOGUCKDESCHIPHON ... : omm merman efincermmnmunntemiientennrmitiremmetstteatiie in en
     3.1     General Information                                                                                                             4
4      Measurement MethOG ..................slllllll>
     4.1     Return Loss Requirements                                                                                                        3
     4.2     Mechanical Requirements                                                                                                         5
3.     MEBSUT@MNE UNC@TIIIEY ..................csnnnmeemmmmemnecenmmenreemmmmmnmeenmmmnssenmmmmsenmmme $
     5.1     Return Loss                                                                                                                     5
     5.2     Dimension Measurement
     5.3     Validation Measurement                                                                                                          5
6      CalibAtiOM MCASUT@MENE RESUIS..............ssssmssssmmmmmmeccnnmmmnnnnmmmmnnnnmmmnmmnmmnn 6
     6.1     Return Loss
     6.2      Mechanical Dimensions




                                                                                                                                            o
7      Validation measurement...             ............................................................................. mtead
     7.1     Measurement Condition




                                                                                                                                           sbow
     7.2     Head Liquid Measurement
     7.3     Measurement Result

                                                                                                                                           se
     7.4     Body Measurement Result
                                                                                                                                           w
§      LISEOFEQUIPMIGHE ... smm ommmmmumaniimmennnmmemetars




                                                               Page: 3710

                  This document shall not be reproduced, except infull or in part, without the writien approval ofSATIMO
                  The informationcontained hereim is to be used onlyfor thepurposefor which it is submittedandis not to
                  be released in whole orpart without written approval ofSATDO,


    ({ des \              SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACK.2693.13.SNTUA



1              INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEV/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.



2              DEVICE UNDER TEST




                                                     Device Under Test
         Device Type                                        COMOSAR 1750 MHz REFERENCE DIPOLE
         Manufacturer                                       Satimo
         Model                                              SID1750
         Serial Number                                      SN 30/13 DIP1G750—260
         Product Condition (new / used)                     New

A yearly calibration interval is recommended.

3              PRODUCT DESCRIPTION

3.1       GENERAL INFORMATION

Satimo‘s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEUIEC 62209 standards. The product is designed for use with the COMOSAR test bench
only

                                     pmmpminge~




                                Figure 1 — Satimo COMOSAR Validation Dipole




                                                              Page: 410

                This document shall not be reproduced, except infull or in part, without the written approval ofSATDO
                The information containedherein is to be usedonlyfor the purpose for which it is submitied and is not to
                be released in whole or part without written approval ofSTIO


    ((l s                    SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 269 3.13 SATU A


            4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

            4.1    RETURNLOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of—20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

            4.2    MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEVIEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


S                 MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

            5.1   RETURNLOSS
The following uncertainties apply to the return loss measurement:
                         Frequency band                                   Expanded Uncertainty on Return Loss
                           400—6000MHz                                                              0.1 dB


            5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:
                            Length (mm)                                       Expanded Uncertainty on Length
                                 3 — 300                                                          0.05 mm


            5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC ENS0361 and CEVIEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                            Scan Volume                                      Expanded Uncertainty
                                     Ig                                                   20.3 %
                                    10 g                                                  20.1 %

                                                                Page: 510

                   This document shall not be reproduced, except infull oin part, without the writeenapproval ofSATIMO
                   The information containedherein is to be used onlyfor the purposefor which it is submitted andis not to
                   be releasedin whole orpart without written approval ofSATIMO


 ((;   +                    SAR REFERENCE DIPOLE CALIBRATION REPORT                                                     Ref: AUK.209.3.13.SATUA
SATIMO

           6     CALIBRATION MEA                      IREMENT RESULTS

           6.1    RETURNLOSS




           Frequency (MHz)                                Return Loss (dB)                                     Requirement (dB)
                  1750                                             —22.80                                                  —20

           6.2    MECHANICAL DIMENSIONS
 Frequency MHz                           L. mm                                         h mm                                      dmm

                            required              measured                required             measured              required          measured
        300               420.0+1%.                                      250.0 +1 %                                 635 41%
        aso               290.0 21%.                                     166.7 +1 %                                 635 i1%.
        750               176.0 +1%                                      100.0 +1 %                                 635 41%
       #as                161.0+1%                                        89.8 +1%                                   36+1%
       900                149.0£1%                                        83.3 +1%                                   16+1%
       1450                89.1+1%                                        s17+1%                                     36+1%
       1500               80.5 +1 %                                       50.0 +1 %                                  26+1%
       1640                79.011%                                        45.7 41%                                   36+1%
       1750                75241%                    pass                 42.9 +1 %                PaS$              16+1%               PAS$
       1800                72.011%                                        41.741%                                    36+1%
       1900                68.0 +1 %                                      so5+1%                                     16+1%
       1950                66.3 41 %                                      38.5 +1 %.                                 16+1%
       2000               645 +1%                                         375i:1%                                    16+1%
       2100               s1.041%                                         35.7 +1%                                   26i1%
       2300                55.5 41 %                                      326i1%                                     36+1%
       2450                s1541%                                         30.4 +1%                                   16+1%
       2600               as.5 +1%                                        288 +1%                                    16+1%
       3000                415 +1%                                        25.0 +1 %                                  16+1%
       3500                37.0:1%                                        264 +1 %                                   16+1%
       3700                34.741%                                        264 £1%                                    16+1%

                                                                 Page: 6/10

                  This document shall not be reproduced, except in full or in part, without the writeen approval of ATIMO
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of SATTMO


  (6. \                   SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 2693.13.SATUA
&'HN’ID

         7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEV/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface ofthe dipole at the described distance from the bottom surface ofthe phantom.

         7.1     MEASUREMENT CONDITION
Software                                                               OPENSAR V4
Phantom                                                                SN 20/09 SAMT71
Probe                                                                  SN 1871 EPG122
Liguid                                                                 Head Liquid Values: eps_.39.9 sigma _1 37
Distance between dipole center and liquid                              10.0 mm
Area scan resolution                                                   dx=8mm/dy=8mm
Zoon Scan Resolution                                                   dx=8mm/dy=8m/dz=Smm
Frequency                                                              1750 MHz
Input power                                                            20 dBm
Liquid Temperature                                                     21 °C
Lab Temperature                                                        21 °C
Lab Humidity                                                           45%

         7.2     HEAD LIQUID                MEA               NT

                          is                      Relative permittivity (e,‘)           Conductivity (0) $/m
                                                  required         measured           required         measured

                              300                 45.3 25 %                          0.87 15 %
                              450                 43.5 15 %                          0.87 15 %
                              750                 41.9 15 %                          0.89 15 %

                              835                 41.5 15 %                          0.90 15 %
                              900                 415 15 %                           0.97 15 %
                             1450                 40.5 25 %                           1.20 45 %
                             1500                 40.4 25 %                           1.23 25 %
                             1640                 40.2 25 %                           13115 %
                             1750                 40.1 25 %           PASS            137 45%             PASS
                             1800                 40.0 15 %                           1.40 +5 %
                             1900                 40.0 15 %                           1.40 25 %

                             1950                 40.0 15 %                           1.40 15 %

                             2000                 40.0 15 %                           1.40 15 %
                             2100                 39.8 15 %                           1.49 15 %

                             2300                 39.5 15 %                           1.67 15 %
                             2450                 39.2 25 %                           1.80 45 %
                             2600                 39.0 15 %                          1.96 +5 %

                             3000                 38.5 15 %                          2.40 45 %
                             3500                 37.9 15 %                          2.91 15 %

                                                               Page: 710

                 This document shall not be reproduced, except infill or in part, without the written approval ofSATIMO
                 The information containedherein is to be used onlyfor thepurposefor which it is submitted and is not to
                 be releasedin whole orpart without written approval ofSATIMO


 {IV i                  SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 269.3.13.ATUA
SATIMO

       7.3   MEASUREMENT RESULT
The IEEE Std. 1528 and CEVIEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation.                All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

                       Eoi                        1 8 SaR (W/ke/w)                     10 g san (w/ke/w)
                                             required           measured            required         measured
                           300                  285                                   194

                           450                  458                                   3.06

                           750                  8.49                                  5.55
                           835                  9.56                                  6.22
                           900                  10.9                                  6.99

                          1450                   29                                     16
                          1500                  305                                   16.8
                          1640                  34.2                                  184
                          1750                  364            38.29 (3.83)            19.3         19.98 (2.00)
                          1800                  384                                   20.1
                          1900                  39.7                                  20.5

                          1950                  40.5                                  20.9
                          2000                  411                                   21.1

                          2100                  43.6                                  219

                          2300                  48.7                                  233
                          2450                  524                                    24
                          2600                  55.3                                  24.6

                          3000                  63.8                                  25.7
                          3500                  671                                    25




                                                                                               ta


                                                            Page: 810
              This document shall not be reproduced, except in full oin part, without the written approval o SATIMO
              The information containedhereinis to be used only for the purpose for which it is submittedand is not to
              be released in whole or part without written approval of ATTMO


                            SAR REFERENCE DIPOLECALIBRATION REPORT                                                  Ref: ACR.269.3.13.SATUA


           74   BODY MEASUREMENT RESULT
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM7I
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Body Liquid Values: eps‘          51.5 sigma        1.47
Distance between dipole center and liquid                               10.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=Smm
Frequency                                                               1750 MHz
Input power                                                             20 dBm
Liguid Temperature                                                      zi °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %



                           m““n':"(v                 1 g SAR (W/ke/w)                     10 g san (w/kg/w)
                                                          measured                             measured
                              1750                       40.12 (4.01)                          21.58 (2.16)




                                                               Page: 910
                  This document shall not bereproduced, except infull or in part, without the written approval of SATIMO:
                  Theinformation containedhereinis to be usedonly for the purposefor which it is submitted and is not to
                  be released in whole or part without written approval of S1TIMO


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref ACR 269 3.13.8ATU.A




          8   LIST OF EQUIPMENT


                                               Equipment Summary Sheet

    Equipment                Manufacturer /                                                   Current                 Next Calibration
   Description                      Modet               [19%49°=000 N91 pyypyojion Pate                                      Date
                                        .                                            Validated. No cal              Validated.      No call
   SAM Phantom                      Satimo                SN—20/08—SAM71             Fequired                       fequired.

                                      .                                              Validated. No cal              Validated       No cal
COMOSAR Test Bench                Version 3                         NA               fequired                       required.

  Network Analyzer          RhO4e8Schware                      gop132                          o22013                       oaror6
      Calipers                      Carrera                  CALIPER—01                        1212010                      1222013

                                        .                                            Characterized prior to |Characterized prior to
  Reference Probe                   Satimo               EPG122 SN 18/11             test. No cal required. |test. No cal required

     Multimeter                Keithley 2000                    1188656                        1172010                      112013

  Signal Generator            Agilent E4438C                 MY49070581                        1212010                      1212013

                                                                                     Characterized prior to |Characterized prior to
     Amplifier                  Aethercomm                       sn 46               test. No cal required. |test. No cal required.

    Power Meter                 HP E4418A                    US38261498                        1172010                      1172013

   Power Sensor               HP ECP—E26A                    US37181460                        1172010                      1172013
  omagl                                                                              Characterized prior to |Characterized prior to
 Dirsctional Coupier          Nerda 4216—20                      01386               test. No calrequired. [test. No cal required.
  Temperature and
  Humidity Sensor           Control Company                     11—661—9                       32012                        32014




                                                              Page: 10/10

                 This document shall not be reproduced, exceptinfull or in part, without the writen approval ofSATIMO
                 The information contained hereinis to be used onlyfor the purposefor which it is submitied and is not to
                 be releasedin whole orpart without written approval ofSATTMO