COMOSAR E-FIELD PROBE CALIBRATION REPORT 6G

FCC ID: R9C-X9006

RF Exposure Info

Download: PDF
FCCID_2238449

                                       (((.
                                  SATIMO
                                  Tre microwavevsioncompany "




   COMOSAR E—Field Probe Calibration Report


                                                  Ref : ACR.269.1.13.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
       TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
               LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
SATIMO COMOSAR DOSIMETRIC E—FIELD PROBE
                      SERIAL NO.: SN 27/13 EPG193

                            Calibrated at SATIMO US
                  2105 Barrett Park Dr. — Kennesaw, GA 30144



                                     *a
                                     |ACCR!DI‘TID
                                     Caibraion Cent szzu t


                                      09/25/2013

    Summary:

    This document presents the method and results from an accredited COMOSAR Dosimetric E—Field
    Probe calibration performed in SATIMO USA using the CALISAR / CALIBAIR test bench, for use
    with a SATIMO COMOSAR system only.              All calibration results are traceable to national
    metrologyinstitutions


                        COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                   Ref: ACR.269.1.13.SATUA
sfifim\o



                                  Name                               Function                          Date                   Signature
Prepared by :                Jérome LUC                        Product Manager                    |9/26/2013
Checked by:                  Jérome LUC                        Product Manager                    9/26/2013

Approved by :          Kim RUTKOWSKI                           Quality Manager                    9/26/2013                 hus




                                                                        Customer Name
                                                                       Shenzhen Morlab
                                      Distribution :                   Communications
                                        istribution :                  Technology Co.,
                                                                                 Ltd




   Issue                    Date                                                         Modifications
     A                  9/26/2013                 Initial release




                                                               Page: 29

                This document shall not be reproduced, except infull or in part, without the written approval ofSATIMO:
                The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                be releasedin whole orpart without written approval ofSATIMO,


     ((L. \                 COMOSAR E—FIELD PROBE CALIBRATION REPORT                                               Ref: ACR.269.1.13.8ATUA



                                                   TABLE OF CONTENTS



1      Devi@ UNG@F TESt ...
2      Product DeSCriptiOn ...«
     2.1     General Information                                                                                                        4
3+     MEBSUIRI@NEM@TROU .... is ceusmmmunccnmennnmeineinciormeetieciieeccosteceentifrrreccieprnntecreprnirase 4
     3.1     Linearity                                                                                                                  4
     3.2     Sensitivity                                                                                                                5
     3.3     Lower Detection Limit                                                                                                      5
     3.4     Isotropy                                                                                                                   5
     3.5     Boundary Effect                                                                                                            5
4      MEASUTCMENEUNCETIAIMEY..............sssicsnssoncccmnmmmmmnmmmmmnnnnmmnmnnnnmnnnnnnnnnnmmnnmnmmm 5
5.     CaliDAtiOM MCASUT@MNE RESUIS................cssns0000000ee000000mneemmmenrmemmrvnnnnmmnenennmnmnence 6
     5.1     Sensitivity in air                                                                                                         6
     5.2     Linearity                                                                                                                  7
     5.3     Sensitivity in liquid                                                                                                      7
     5.4     Isotropy                                                                                                                   8
6      Li8t OF EQMIDMENE.........ccuossecrmcomemttinrmitentecasiitiserinstenienitineristenntemsernmnersitineee 9




                                                                Page: 39
                   This document shall not be reproduced, except infull or inpart, without the written approval ofSATIMO
                   The information contained herein is to be usedonsfor thepurposefor which it is submiteedand is not to
                   be released in whole or part without written approval ofSATO:


    {t . \                COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                    Ref: ACR.200.1 12. 8ATULA
SATIMO

1               DEVICE UNDER TEST

                                                      Device Under Test
      Device Type                                                    COMOSAR DOSIMETRIC E FIELD PROBE
      Manufacturer                                                   Satimo
      Model                                                          SSE2
      Serial Number                                                  SN 27/13 EPG193
      Product Condition (new / used)                                 New
      Frequency Range of Probe                                       0.7 GHz—6GHz
      Resistance of Three Dipoles at Connector                       Dipole 1: R1=0.205 MQ
                                                                     Dipole 2: R2=0.176 MQ
                                                                     Dipole 3: R3—0.214 MQ

A yearly calibration interval is recommended.

2               PRODUCT DESCRIPTION

2.1          GENERAL INFORMATION
Satimo‘s COMOSAR E field Probes are built in accordance to the IEEE 1528, OET 65 Bulletin C
and CEV/IEC 62209 standards.




                            Figure 1 —Satimo COMOSAR Dosimetric Efield Dipole

      Probe Length                                                   330 mm
      Length of Individual Dipoles                                   2 mm
      Maximum external diameter                                      8 mm
      Probe Tip External Diameter                                    2.5 mm
      Distance between dipoles / probe extremity                      1 mm


         3      MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C, CENELEC ENS0361 and CEI/IEC 62209 standards provide
recommended practices for the probe calibrations, including the performance characteristics of
interest and methods by which to assess their affect. All calibrations / measurements performed meet
the fore mentioned standards.

          3.1    LINEARITY
The evaluation of the linearity was done in free space using the waveguide, performing a power
sweep to cover the SAR range 0.01W/kg to 100W/kg.



                                                                 Page: 49

                  This document shall not be reproduced, except infull or in part, withoutthe writen approval ofSATIMO:
                  The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                  be releasedin whole orpart without written approval ofSATIMO.


 ((l. \                        COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                 Ref: ACR 269.1.13 SATU A
SATIMO

           3.2        SENSMVITY
The sensitivity factors of the three dipoles were determined using a two step calibration method (air
and tissue simulating liquid) using waveguides as outlined in the standards.

           3.3    LOWER DETECTION LIMIT
The lower detection limit was assessed using the same measurement set up as used for the linearity
measurement. The required lower detection limit is 10 mW/kg.

           3.4        ISOTROPY
The axial isotropy was evaluated by exposing the probe to a reference wave from a standard dipole
with the dipole mounted under the flat phantom in the test configuration suggested for system
validations and checks. The probe was rotated along its main axis from 0 — 360 degrees in 15 degree
steps. The hemispherical isotropy is determined by inserting the probe in a thin plastic box filled
with tissue—equivalentliquid, with the plastic box illuminated with the fields from a half wave dipole.
The dipole is rotated about its axis (0°%—180°) in 15° increments. At each step the probe is rotated
about its axis (0%—360°).

           3.5        BOUNDARY EFFECT
The boundary effect is defined as the deviation between the SAR measured data and the expected
exponential decay in the liquid when the probe is oriented normal to the interface. To evaluate this
effect, the liquid filled flat phantom is exposed to fields from either a reference dipole or waveguide.
With the probe normal to the phantom surface, the peak spatial average SAR is measured and
compared to the analytical value at the surface.

4                MEASUREMENT UNCERTAINTY

The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEVIEC
62209 standards were followed to generate the measurement uncertainty associated with an E—field
probe calibration using the waveguide technique. All uncertainties listed below represent an
expanded uncertainty expressed at approximately the 95%confidence level using a coverage factor
of k=2,traceable to the InternationallyAccepted Guides to Measurement Uncertainty.
 Uncertainty analysis ofthe probe calibration in waveguide
                                                    Uncertainty            Probability               e                      5       Standard
           ERROR SOURCES                             value (%)            Distribution            DW!9"                *        Uncertainty (%)
    Incident or forward power                          3.00%               Rectangular                   \/5     |      1            1.732%

    Reflected power                                    3.00%               Rectangular                   [       |      1            1.732%
    Liquid conductivity                                 5.00%              Rectangular                   [3      |      1            2.887%
 Liquid permittivity                                   4,00%              Rectangular                    [3      |      1            2309%

 Field homogeneity                                     3.00%              Rectangular                    ‘/5     ’      1            1.732%

 Field probe positioning                               5,00%               Rectangular                   \/_';   |      1            2.887%

 Field probe linearity                                 3.00%               Rectangular                   M       |      1            1.732%

                                                                    Page: 59

                      This document shall not be reproduced, except infull or in part, without the written approval ofSATIMO:
                      The informationcontainedherein is to be used ons for the purposefor which it is submitted and is not to
                      be released in whole orpart without written approval ofSATIMO


 ({I :       \                        COMOSAR E—FIELD PROBE CALIBRATION REPORT                                        Ref: ACR 269.1.13.SATUA
SATIMo

  Combined standard uncertainty                                                                                                   5.831%
         Expanded uncertainty                                                                                                          .
     95 % confidence level k = 2                                                                                                   12.0%




         5       CALIBRATION MEASUREMENT RESULTS


                                                          Calibration Parameters
   Liquid Temperature                                                       21 °C
   Lab Temperature                                                          21 °C
   Lab Humidity                                                             45 %


         5.1     SENSITIVITY
                     IN


                                             Normx dipole Normy dipole Normz dipol,e
                                             1 (wV/(V/im))| 2 (wV/(V/im))| 3 (V/(V/m))
                                                   0.50                 0.55                 0.52


                                             DCP dipole 1        DCP dipole 2            DCP dipole 3
                                                   (mV)                (mV)                 (mV)
                                                    90                      101              99


Calibration curves ei=f(V) (i=1,2,3) allow to obtain H—field value using the formula:




                                                   Calibration curves
                                    1003—
                                    §
                    E—Field (VAn)




                                                                                                          Dipole 1
                                    §




                                                                                                          Dgole 2
                                    8 §




                                      4.
                                       0&]   ooe   oor    oo   o8    ot0      of2   on    016 0;8
                                                               Votage (V)




                                                                     Page: 69

                  This document shall not be reproduced, except inill or i part, without the written approval of$ATIMO,
                  The information contained herem is to be used onlyfor the purposefor which it is submitied and is not to
                  be releasedin whole orpart without written approval of SATIMO.


 (( n        d                             COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                  Ref: ACR 269.1.13.SATUA
SATIN/ID

        5.2                      LINEARITY

                                                               Linearity
                               1.00—
                               076—
                               050
        Lined ity Ertor (0B)




                               025
                               0 90 fRitnbaed nlbe} fnd
                  $ 1%
                    050
                    q75—
                               1.00—
                                    0      100      200     300       400 5oo         600      700      s
                                                                  E—Field (V/m)
        Linea                           I+/1.54% (+/0 074B




        5.3                      SENSITIVITY IN LIQUID

  Liquid                                Frequency          Permittivity         Epsilon (S/m)             ConvE
                                        {MHz
                                         100MHz
  HLS200                                   5200                36.11                 4.81                  23.71
  BLS200                                  5200                 49.87                 4.99                  2439
  HLS400                                  5400                 36.61                 5.08                  26.58
  BLS400                                  5400                 49.09                 5.64                  27.58
  HLS600                                   5600                35.97                 5.37                  25.33
  BLS600                                   5600                48.64                 5.99                  26.30
  HLS800                                  5800                 35.33                 5.59                  25.64
  BLS800                                  5800                 47.76                 6.21                  26.47



LOWER DETECTION LIMIT: 7mW/kg




                                                                                 Page: 79

                                   This document shall not be reproduced, except infull or in part, without the writeen approval ofSATIMO
                                   The information contained herein is to be used onlfor the purposefor which it is submitted andis not to
                                   be releasedin whole orpart without written approval ofSATIMO.


 (( es \               COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                   Ref: ACR269 1. 13 8ATUA
SATIMo


       5.4    ISOTROPY
HL5400 MHz
— Axial isotropy:                              0.08 dB
— Hemisphericalisotropy:                       0.13 dB




    issh 4 & & t d incammemien‘t.




                                                             Page: 89

               This document shall not be reproduced, evcept infull or in part, without the written approval ofSATIMO
               The information contained herein is to be used onlyfor the purposefor which it is submittedand is not to
               be released in whole orpart without written approval ofSATIMO:


                             COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                  Ref: ACR.269.1 13 SATUA




            6   LIST OF EQUIPMENT



                                                  Equipment Summary Sheet


     Equipment                  Manufacturer /                      i         nhan               Current                 Next Calibration
    Description                        Model               |‘"%""**0990 M9—/ Copprofion Date                                       Date
     Fiat Phantom                      Satimo                SN—20/08—SAM71             Validated. No cal
                                                                                        Fequired.                      Validated
                                                                                                                       required.          No cal

COMOSAR Test Bench                   Version 3                           NA
                                                                                        Validated.
                                                                                        Fequired.
                                                                                                   No cal              Validated
                                                                                                                       required.
                                                                                                                                          No cal

   Network Analyzer            RNOG®&SChware                      g400132                        o2ro13                        o2ro16
                                          i                                             Characterized prior to |Characterized prior to
   Reference Probe                     Setima                 EP 94 SN 37/08            test. No calrequired. [test. No cal required.

      Multimeter                  Keithley 2000                    1188656                        1172010                      1172013

   Signal Generator              Agilent E4438C                 MY49070581                        1212010                      1222013

            i                                                                           Characterized prior to |Characterized prior to
       Amplifier                   Asthercomin                      SN 046              test. No calrequired. [test. No cal required

     Power Meter                   HP E4418A                    US38261498                        1172010                      1172013

    Power Sensor                  HP ECP—E26A                   US37181460                        1172010                      1172013
   us                                                                                   Characterized prior to |Characterized prior to
  prectionsi couple              Neadsit=0                              1908            test._ No cal required. |test._ No cal required.
                .                                  .                                    Validated. No cal              Validated. No cal
      Waveguide                  Mega Industries            069Y7—158—13—712 hlrag.                                    Fequired.
            .         m                            .                                    Validated. No cal              Validated. No cal
Waveguide Transition              Mega Industries           O69Y7—158—13—701 fequired.                                 Fequired.
        .           ww                            .                                     Validated. No cal              Validated. No cal
|Waveguide Termination|           Mega Industries           O69Y7—158—13—701 required                                  required.

[Temperature / Humidity
        Sensor                  Control Company                    11—661—9                       32012                        372014




                                                                   Page: 9.9

                    This document shall not be reproduced, except infull or in part, without the written approval ofSATIMO:
                    The information containedhereim is to be used onlyfor the purposefor which it is submitted and is not to
                    be releasedin whole orpart without written approval ofSATMO,