Dipole 750MHz Calibration Report

FCC ID: R9C-X9006

RF Exposure Info

Download: PDF
FCCID_2238304

      SAR Reference Dipole Calibration Report

                                       Ref : ACR.269.2.13.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
       TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
               LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
   SATIMO COMOSAR REFERENCE DIPOLE
                      FREQUENCY: 750 MHZ
                      SERIAL NO.: SN 30/13 DIP0G750-259

                          Calibrated at SATIMO US
                  2105 Barrett Park Dr. - Kennesaw, GA 30144




                                      09/25/2013

    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
    to national metrology institutions.


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




                                  Name                                Function                        Date                    Signature
Prepared by :                Jérôme LUC                         Product Manager                    9/26/2013

Checked by :                 Jérôme LUC                         Product Manager                    9/26/2013

Approved by :          Kim RUTKOWSKI                            Quality Manager                    9/26/2013




                                                                        Customer Name
                                                                       Shenzhen Morlab
                                                                       Communications
                                       Distribution :
                                                                       Technology Co.,
                                                                             Ltd




   Issue                   Date                                                           Modifications
     A                  9/26/2013                  Initial release




                                                               Page: 2/10

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                               SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




                                                      TABLE OF CONTENTS



1    Introduction ................................................................................................................ 4
2    Device Under Test ..................................................................................................... 4
3    Product Description ................................................................................................... 4
    3.1      General Information _______________________________________________________ 4
4    Measurement Method ................................................................................................ 5
    4.1      Return Loss Requirements __________________________________________________ 5
    4.2      Mechanical Requirements ___________________________________________________ 5
5    Measurement Uncertainty .......................................................................................... 5
    5.1      Return Loss ______________________________________________________________ 5
    5.2      Dimension Measurement ___________________________________________________ 5
    5.3      Validation Measurement ____________________________________________________ 5
6    Calibration Measurement Results .............................................................................. 6
    6.1      Return Loss ______________________________________________________________ 6
    6.2      Mechanical Dimensions ____________________________________________________ 6
7    Validation measurement ............................................................................................ 7
    7.1      Measurement Condition ____________________________________________________ 7
    7.2      Head Liquid Measurement __________________________________________________ 7
    7.3      Measurement Result _______________________________________________________ 8
    7.4      Body Measurement Result __________________________________________________ 9
8    List of Equipment .................................................................................................... 10




                                                                    Page: 3/10

                    This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                    The information contained herein is to be used only for the purpose for which it is submitted and is not to
                    be released in whole or part without written approval of SATIMO.


                        SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




1          INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.



2          DEVICE UNDER TEST




                                                   Device Under Test
        Device Type                                       COMOSAR 750 MHz REFERENCE DIPOLE
        Manufacturer                                      Satimo
        Model                                             SID750
        Serial Number                                     SN 30/13 DIP0G750-259
        Product Condition (new / used)                    New

A yearly calibration interval is recommended.

3          PRODUCT DESCRIPTION

3.1     GENERAL INFORMATION
Satimo’s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEI/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                              Figure 1 – Satimo COMOSAR Validation Dipole



                                                             Page: 4/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                        SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




       4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

       4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of -20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

       4.2   MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEI/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5            MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

       5.1   RETURN LOSS
The following uncertainties apply to the return loss measurement:
                    Frequency band                                     Expanded Uncertainty on Return Loss
                      400-6000MHz                                                                0.1 dB

       5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:
                       Length (mm)                                         Expanded Uncertainty on Length
                            3 - 300                                                             0.05 mm


       5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEI/IEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                     Scan Volume                  Expanded Uncertainty
                                1g                                                     20.3 %
                               10 g                                                    20.1 %

                                                             Page: 5/10

              This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
              The information contained herein is to be used only for the purpose for which it is submitted and is not to
              be released in whole or part without written approval of SATIMO.


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




      6     CALIBRATION MEASUREMENT RESULTS

      6.1   RETURN LOSS




     Frequency (MHz)                                    Return Loss (dB)                                   Requirement (dB)
           750                                               -23.59                                              -20

      6.2   MECHANICAL DIMENSIONS
Frequency MHz                          L mm                                         h mm                                      d mm

                          required              measured                required             measured             required           measured

    300                 420.0 ±1 %.                                   250.0 ±1 %.                                6.35 ±1 %.
    450                 290.0 ±1 %.                                   166.7 ±1 %.                                6.35 ±1 %.
    750                 176.0 ±1 %.                PASS               100.0 ±1 %.               PASS             6.35 ±1 %.            PASS

    835                 161.0 ±1 %.                                    89.8 ±1 %.                                 3.6 ±1 %.
    900                 149.0 ±1 %.                                    83.3 ±1 %.                                 3.6 ±1 %.
    1450                 89.1 ±1 %.                                    51.7 ±1 %.                                 3.6 ±1 %.
    1500                 80.5 ±1 %.                                    50.0 ±1 %.                                 3.6 ±1 %.
    1640                 79.0 ±1 %.                                    45.7 ±1 %.                                 3.6 ±1 %.
    1750                 75.2 ±1 %.                                    42.9 ±1 %.                                 3.6 ±1 %.
    1800                 72.0 ±1 %.                                    41.7 ±1 %.                                 3.6 ±1 %.
    1900                 68.0 ±1 %.                                    39.5 ±1 %.                                 3.6 ±1 %.
    1950                 66.3 ±1 %.                                    38.5 ±1 %.                                 3.6 ±1 %.
    2000                 64.5 ±1 %.                                    37.5 ±1 %.                                 3.6 ±1 %.
    2100                 61.0 ±1 %.                                    35.7 ±1 %.                                 3.6 ±1 %.
    2300                 55.5 ±1 %.                                    32.6 ±1 %.                                 3.6 ±1 %.
    2450                 51.5 ±1 %.                                    30.4 ±1 %.                                 3.6 ±1 %.
    2600                 48.5 ±1 %.                                    28.8 ±1 %.                                 3.6 ±1 %.
    3000                 41.5 ±1 %.                                    25.0 ±1 %.                                 3.6 ±1 %.
    3500                 37.0±1 %.                                     26.4 ±1 %.                                 3.6 ±1 %.
    3700                 34.7±1 %.                                     26.4 ±1 %.                                 3.6 ±1 %.


                                                               Page: 6/10

                This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                The information contained herein is to be used only for the purpose for which it is submitted and is not to
                be released in whole or part without written approval of SATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




        7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEI/IEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance from the bottom surface of the phantom.

        7.1    MEASUREMENT CONDITION
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Head Liquid Values: eps’ : 42.4 sigma : 0.88
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               750 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %

        7.2    HEAD LIQUID MEASUREMENT
                          Frequency
                                                 Relative permittivity (r’)              Conductivity (σ) S/m
                             MHz
                                                 required           measured            required          measured
                              300               45.3 ±5 %                               0.87 ±5 %
                              450               43.5 ±5 %                               0.87 ±5 %
                              750               41.9 ±5 %              PASS             0.89 ±5 %            PASS
                              835               41.5 ±5 %                               0.90 ±5 %
                              900               41.5 ±5 %                               0.97 ±5 %
                             1450               40.5 ±5 %                               1.20 ±5 %
                             1500               40.4 ±5 %                               1.23 ±5 %
                             1640               40.2 ±5 %                               1.31 ±5 %
                             1750               40.1 ±5 %                               1.37 ±5 %
                             1800               40.0 ±5 %                               1.40 ±5 %
                             1900               40.0 ±5 %                               1.40 ±5 %
                             1950               40.0 ±5 %                               1.40 ±5 %
                             2000               40.0 ±5 %                               1.40 ±5 %
                             2100               39.8 ±5 %                               1.49 ±5 %
                             2300               39.5 ±5 %                               1.67 ±5 %
                             2450               39.2 ±5 %                               1.80 ±5 %
                             2600               39.0 ±5 %                               1.96 ±5 %
                             3000               38.5 ±5 %                               2.40 ±5 %
                             3500               37.9 ±5 %                               2.91 ±5 %

                                                                Page: 7/10

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                       SAR REFERENCE DIPOLE CALIBRATION REPORT                                                     Ref: ACR.269.2.13.SATU.A




      7.3   MEASUREMENT RESULT
The IEEE Std. 1528 and CEI/IEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

                      Frequency
                                                 1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                         MHz
                                             required           measured            required          measured
                          300                   2.85                                   1.94
                          450                   4.58                                   3.06
                          750                   8.49            8.51 (0.85)            5.55          5.55 (0.56)
                          835                   9.56                                   6.22
                          900                   10.9                                   6.99
                         1450                    29                                     16
                         1500                   30.5                                   16.8
                         1640                   34.2                                   18.4
                         1750                   36.4                                   19.3
                         1800                   38.4                                   20.1
                         1900                   39.7                                   20.5
                         1950                   40.5                                   20.9
                         2000                   41.1                                   21.1
                         2100                   43.6                                   21.9
                         2300                   48.7                                   23.3
                         2450                   52.4                                    24
                         2600                   55.3                                   24.6
                         3000                   63.8                                   25.7
                         3500                   67.1                                    25




                                                            Page: 8/10

             This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
             The information contained herein is to be used only for the purpose for which it is submitted and is not to
             be released in whole or part without written approval of SATIMO.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




        7.4    BODY MEASUREMENT RESULT
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAM71
Probe                                                                   SN 18/11 EPG122
Liquid                                                                  Body Liquid Values: eps’ : 55.7 sigma : 0.96
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=5mm
Frequency                                                               750 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21 °C
Lab Humidity                                                            45 %



                          Frequency
                                                     1 g SAR (W/kg/W)                      10 g SAR (W/kg/W)
                             MHz
                                                          measured                              measured
                              750                         8.81 (0.88)                           5.88 (0.59)




                                                                Page: 9/10

                 This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                 The information contained herein is to be used only for the purpose for which it is submitted and is not to
                 be released in whole or part without written approval of SATIMO.


                            SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.269.2.13.SATU.A




       8    LIST OF EQUIPMENT


                                                 Equipment Summary Sheet

   Equipment                 Manufacturer /                        Current                                               Next Calibration
                                            Identification No.
   Description                  Model                          Calibration Date                                                Date
                                                                                       Validated. No cal               Validated.      No cal
   SAM Phantom                       Satimo                SN-20/09-SAM71
                                                                                       required.                       required.
                                                                                       Validated. No cal               Validated.      No cal
COMOSAR Test Bench                 Version 3                          NA
                                                                                       required.                       required.
                             Rhode & Schwarz
  Network Analyzer                                              SN100132                         02/2013                        02/2016
                                  ZVA

      Calipers                       Carrera                   CALIPER-01                        12/2010                        12/2013

                                                                                        Characterized prior to Characterized prior to
  Reference Probe                    Satimo                EPG122 SN 18/11
                                                                                        test. No cal required. test. No cal required.

     Multimeter                 Keithley 2000                     1188656                        11/2010                        11/2013

  Signal Generator             Agilent E4438C                 MY49070581                         12/2010                        12/2013

                                                                                        Characterized prior to Characterized prior to
      Amplifier                  Aethercomm                        SN 046
                                                                                        test. No cal required. test. No cal required.

    Power Meter                   HP E4418A                    US38261498                        11/2010                        11/2013

   Power Sensor                 HP ECP-E26A                    US37181460                        11/2010                        11/2013

                                                                                        Characterized prior to Characterized prior to
 Directional Coupler           Narda 4216-20                       01386
                                                                                        test. No cal required. test. No cal required.
  Temperature and
                              Control Company                     11-661-9                        3/2012                        3/2014
  Humidity Sensor




                                                                Page: 10/10

                  This document shall not be reproduced, except in full or in part, without the written approval of SATIMO.
                  The information contained herein is to be used only for the purpose for which it is submitted and is not to
                  be released in whole or part without written approval of SATIMO.