SAR Report-8

FCC ID: R9C-R8006

RF Exposure Info

Download: PDF
FCCID_2237812

                                          (/r.\
                                   SATIMO
                                   Te micrmare visoncompany .A




      SAR Reference Dipole Calibration Report



                                          Ref: ACR.269.4.13.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
           TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
               LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
   SATIMO COMOSAR REFERENCE DIPOLE
                      FREQUENCY: 1900 MHZ
                      SERIAL NO.: SN 30/13 DIP1G900—261
                           Calibrated at SATIMO US
                  2105 Barrett Park Dr. — Kennesaw, GA 30144




                                        Caltrasion CERT w22402

                                         09/25/2013

    Summary:

    This documentpresents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
    tonational metrology institutions


 ((4.
sarimo
        a                SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 269.4.13.SATU A




                                  Name                              Function                          Date                 Signature
Prepared by :               Jérome LUC                        Product Manager                   9/26/2013                   fi’
Checked by :                 Jérome LUC                        Product Manager                  9/26/2013                    fi’

Approved by :          Kim RUTKOWSKI                           Quality Manager                  9/26/2013                 ... mathoti?




                                                                       Customer Name
                                                                      Shenzhen Morlab
                                           .        us                Communications
                                      Distribution :                  Technology Co.,
                                                                                Ltd




   Issue                 Date                                                          Modifications
     A                 9/26/2013                  Initial release




                                                             Page: 210

                This document shall not be reproduced, except infull oin part, without the writen approval ofSATTMO
                The information contained hereini to be used onlyfor the purposefor which it is submitted and is not to
                be releasedin whole orpart without written approval ofS1TIMO.


  (t. \                     SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR.2694.13.SATUA
SATIMO


                                                   TABLE OF CONTENTS



1      IMfOUUONON: u.cs mumtenemanncennumnneeuansununtendencrnemunnsecemen ce
2      D@ViC@ UMG@T TESt........ccssssssssccnnnnmmmmnnnmmmmmmmnnnnmnnnnnnmnnnnmnnnnnnnnnnnnenemnmsnnnmmnmmmn 4
3      PIOOUCHD@SETIPHGH ... . :s i. rmmmucms h crmncoenieetionemerrosriiensremateomerrenreeerneecenteivees 4
     3.1     General Information                                                                                                         4
&      MEASUIETI@NEMOIAOU ...:::,.....—sssmmssssscemcsmmmimmmmimnsnemmmnminanimimnemiiien B
     4.1     Return Loss Requirements                                                                                                    $
     4.2     Mechanical Requirements                                                                                                     5
3      MEASUIRiRi@RE UNICEAIANINY ....:.:—mmurenesncammmeiiicaiteientaraiercectetee
     5.1     Return Loss                                                                                                                 5
     5.2     Dimension Measurement                                                                                                       5
     5.3     Validation Measurement                                                                                                      5
6.     CalibAtiON MCASUTEMENE RESUIS...............sssmmssnmmemnnmmmmmmmnnnmmmmmnnnnnmmmnnsnmmmmmmms 6
     6.1     Return Loss
     6.2       Mechanical Dimensions                                                                                                     6
L.     NAHORHION IERSUERMIENLE . is s. iss srccomrsrmmmcmmmmmmemmmmnstimrrretiinmmnssitonmmmnsscemseresticseth 7
     7.1     Measurement Condition                                                                                                       7
     7.2     Head Liquid Measurement                                                                                                     7
     7.3     Measurement Result                                                                                                          8
     7.4     Body Measurement Result                                                                                                     9
$      LMSFOBRQUIPMENL .....Ssommummmmencemmenememmeannrnsinmemtnretenreisentiieacacters 10




                                                                Page: 310

                   This document shall not be reproduced, except infull or in part, without the written approval ofSATIMO.
                   The information contained herein is to be used onlyfor thepurposefor which it is submittedand is not to
                   be released in whole orpart without written approval ofSATTMO.


    {{ L:                  SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 269.4.13.SATU A



1              INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEV/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.



2              DEVICE UNDER TEST




                                                      Device Under Test
            Device Type                                      COMOSAR 1900 MHz REFERENCE DIPOLE
            Manufacturer                                     Satimo
            Model                                            $1D1900
            Serial Number                                    SN 30/13 DIP1G900—261
            Product Condition (new / used)                   New

A yearly calibration interval is recommended.

3              PRODUCT DESCRIPTION

3.1         GENERAL INFORMATION
Satimo‘s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEU/IEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.

                                      F




                                  Figure 1 —Satimo COMOSAR Validation Dipole



                                                               Page: 410

                  This document shall not be reproduced, except infull or in part, without the writen approval ofSATIMO.
                  The information containedherein is to be used onlyfor the purposefor which it is submitted andis not to
                  be released in whole or part without written approval ofSATTMO.


    {f Ls                   SAR REFERENCE DIPOLE CALIBRATION REPORT                                                 Ref: ACR 269 4.13 SATU A


            4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEV/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

            4.1    RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of —20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

            4.2    MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEVIEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5                 MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

            5.1   RETURNLOSS
The following uncertainties apply to the return loss measurement:
                         Frequency band                                  Expanded Uncertainty on Return Loss
                          400—6000MHz                                                              0.1 dB

            5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:
                           Length (mm)                                       Expanded Uncertainty on Length
                                3 —300                                                           0.05 mm


            5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC ENS0361 and CEMIEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                 Scan Volume                Expanded Uncertainty
                                    3                                                    20.3 %
                                   10 g                                                  20.1 %

                                                                Page: 510
                   This document shall not be reproduced, except infull or in part, without the writeen approval ofSATIMO
                   The information contained hereinis to be used only for thepurposefor which it is submittedandis not to
                   be releasedin whole or part without writeenapproval ofSATIMO


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 269.4.13.SATUA



      6     CALIBRATION MEASUREMENT RESULTS

      6.1       RETURNLOSS
                                                                Frequency, MR




      Frequency (MHz)                                   Return Loss (dB)                                  Requirement (dB)
                1900                                            —26.71                                               —20

      6.2       MECHANICAL DIMENSIONS
Frequency MH2                          L. mm                                       hmm                                      d mm
                          required              measured               required            measured             required           measured
     300                 420.0 £1%.                                  250.0 +1 %                                635 41 %.
     aso                 290.0 +1 %                                  166.7 +1 %                                635 41 %.
     750                 176.0 +1 %                                  100.0 +1 %                                6.35 +1 %.

    s35                  161.0£1%                                     s9.8 +1%                                  3611%
    200                  149.0 +1 %.                                  83.3+1%                                   36+1%
    1450                 8o.1 +1 %                                    51.7+41%                                  36+1%
    1500                  80.5 +1 %                                   50.0 +1 %.                                3611%

    1640                  79.0£1%                                     457 +1%                                   3611%
    1750                  752i1%                                      42.9 +1%.                                 36+1%
    1800                  72.0 +1 %                                   41.7 +1 %.                                3.611%

    1900                  68.0 +1 %                Pass               39.5 +1 %.               PASS             1621%                PaS$
    1950                  66.3 +1%                                    3a.5 +1 %.                                26:1%
    2000                  64.5 41 %                                   37.5 +1%                                  16+1%
    2100                  610 +1%                                     35.741%                                   16+:1%
    2300                  55.5 11 %                                   32641%                                    3611%
    2450                  515 41%                                     30.4 +1%                                  1611%
    2600                  a8.5 £1%.                                   20.8 +1%                                  36+1%
    3000                  ans +1%                                     25.0 +1%                                  36+1%
    3500                  37.041%                                     264 +1%                                   36+1%
    3700                  34.741%                                     26.4 +1 %.                                1611%

                                                              Page: 610
                 This documenshall not be reproduced, exceptin full or in part, withoutthe writen approval of SATIMO
                 The information contained herein is to be used only for thepurposefor whichit is submited and is not to
                 be releasedin whole or part without written approval of ATIMO


 (( Ls                    SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 269.4.13 SATU A


         7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEVIEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface ofthe dipole at the described distance from the bottom surface of the phantom.

         7.1     MEASUREMENT CONDITION
Software                                                              OPENSAR V4
Phantom                                                               SN 20/09 SAM71
Probe                                                                 SN 18/11 EPG122
Liguid                                                                Head Liguid Values: eps‘ : 39.8 sigma 1.43
Distance between dipole center and liquid                             10.0 mm
Area scan resolution                                                  dx=8mm/dy—=§mm
Zoon Scan Resolution                                                  dx=8mm/dy—=§m/dz=Smm
Frequency                                                              1900 MHz
Input power                                                           20 dBm
Liquid Temperature                                                    21 °C
Lab Temperature                                                       21 °C
Lab Humidity                                                          45 %

         7.2   HEAD LIQUID MEASUREMENT

                          ""Tg"                 Relative permittivity (e,‘)             Conductivity (0) $/m
                                                required          measured            required         measured

                              300              45.3 45 %                             0.87 15 %
                              450              43.5 15 %                             0.87 25 %
                              750              41945 %                               0.89 +5 %
                              835              41.5 25 %                             0.90 15 %
                              900              41.5 45 %                             0.97 15 %
                             1450              40.5 15 %                             1.20 45 %
                             1500              40.4 25 %                             12345%

                             1640              40.2 45 %                             13145 %
                             1750              40.1 25 %                             13745 %

                             1800              40.0 45 %                             1.40 45 %
                             1900              40.0 25 %             PASS            1.40 15 %            PASS
                             1950              40.0 15 %                             1.40 15 %
                             2000              40.0 25 %                             1.40 45 %

                             2100              39.8 15 %                             149 15 %
                             2300              39.5 15 %                             167 45%
                             2450              39.245 %                              1.80 25 %
                             2600              39.0 15 %                             19615 %
                             3000              38.5 45 %                             2.40 15 %
                             3500              37.945%                               29145 %

                                                              Page: 710
                 This document shall not he reproduced, except inul or in part, without the writen approval ofSATTMO
                 The information contained herein is to be used onlyfor thepurposefor which it is submittedand is not to
                 be releasedin whole or part without written approval ofSATIMO


 {f (.    \                SAR REFERENCE DIPOLE CALIBRATION REPORT                                                 Ref: ACR.2694.13 SATU A


         7.3      MEASUREMENT RESULT
The 1E         Std. 1528 and CEVIEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power

                           inrminl                   1 8 san (w/ke/w)                    10 g sar (w/ke/w)
                                                 required          measured           required         measured
                              300                  2.85                                  1.94

                              450                  458                                   3.06
                              750                  8.49                                  5.55
                              835                  956                                  6.22

                              900                  10.9                                 6.99
                              1450                  29                                    16
                              1500                 30.5                                  16.8
                              1640                 34.2                                  18.4

                              1750                 36.4                                  193
                              1800                 38.4                                  20.1
                              1900                 39.7           39.39 (3.94)           20.5         20.07 (2.01)

                              1950                 40.5                                  20.9
                              2000                 411                                   211

                              2100                 43.6                                  21.9

                              2300                 48.7                                  233

                              2450                 524                                    24
                              2600                 55.3                                  24.6
                              3000                 63.8                                  25.7

                              3500                 671                                    25




                                                                                                 \‘ UR




                                                               Page: 810
                  This document shall not be reproduced, except in full or in part, without the writen approval of ATIMO
                  Theinformation contained herein is to heused onlfor the purposefor whichit is submitted andis not to
                  be released in whole or part without written approval of S1TDIO


 (¢e. M                    SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref. ACR.269.4.13.SATUA
SATIMO

           7.4   BODY MEASUREMENT RESULT
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAMT7I
Probe                                                                   SN 18711 EPG122
Liquid                                                                  BodyLiquid Values: eps‘            52.5 sigma: 1.50
Distance between dipole center and liquid                               10.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=—Smm
Frequency                                                               1900 MHz
Inputpower                                                              20 dBm
Liquid Temperature                                                      21 °C
Lab Temperature                                                         21°C
Lab Humidity                                                            45%



                          F
                              ";':"1““               1 g sant (w/kg/w)                    10 g SAR (W/kg/w)
                                                          measured                              measured

                               1900                      42.33 (4.23)                          22.27 (2.23)




                                                                                                   \' UA —




                                                               Page 9710

                 This document shall not be reproduced, exceptinfull or in part, without the writen approval of ATIMO
                 Theinformation contained herein is to be used only for the purpose for which i is submitied and is not to
                 be released in whole or part without written approval of SATTMO


                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 269.4.13.SATU A




    Equipment               Manufacturer /                                                   Current                 Next Calibration
   Description                     Modet               [180006¢2000 NO—] Copiprotion Date                                   Date
                                                                        Validated. No cal                          Validated.       No call
   SAM Phantom                      Satimo               SN—20/08—SAM71 Fequired.                                  required.
                                                                                    Validated. No cal              Validated        No call
COMOSAR Test Bench                Version 3                        NA               Fequired                       required

  Network Analyzer          RN°C®;vi""wm                      sN100132                        o2r013                       o2r016
      Calipers                     Carrera                   CALIPER—01                       1212010                      1222013

                                                                                     Characterized prior to |Characterized prior to
  Reference Probe                  Satimo                EPG122 SN 18711             test. No calrequired. |test. No cal required

     Multimeter                Keithley 2000                    1188656                       1172010                      112013

  Signal Generator           Agilent E4438C                 MY49070581                        1272010                      1212013

          .                                                                         Characterized prior to |Characterized prior to
     Amplifier                  Asthercomm                      sn 046              test. No cal required. |test. No cal required.

    Power Meter                 HP E4418A                   US38261498                        1172010                      112013

   Power Sensor               HP ECP—E26A                    US37181460                       1172010                      1172013

   iands                                                                            Characterized prior to |Characterized prior to
 Directional Coupler          Narda 4216—20                      01386              test. No cal required. [test.. No cal required.

  Temperature and
  Humidity Sensor           Control Company                    11—661—9                       32012                        312014




                                                             Page: 10/10
                 This document shall not be reproduced, except infull or in part, without the writeen approval ofSATIMO:
                 The information containedherein is to be used onlyforthe purposefor which it is submitted andis not to
                 be releasedin whole orpart without written approval ofSATIMO.