      SAR Reference Dipole Calibration Report



                                       Ref : ACR.261.3.13.SATU.A




 SHENZHEN MORLAB COMMUNICATIONS
           TECHNOLOGY CO., LTD
  FL3, BUILDING A, FEIYANG SCIENCE PARK, NO.8
               LONGCHANG ROAD,
BLOCK 67, BAOAN DISTRICT, SHENZHEN, GUANGDONG
              PROVINCE, P.R. CHINA
   SATIMO COMOSAR REFERENCE DIPOLE
                      FREQUENCY: 835 MHZ
                      SERIAL NO.: SN 20/08 DIPCI9
                         Calibrated at SATIMO US
                  2105 Barrett Park Dr. — Kennesaw, GA 30144


                                     «a
                                     IACCIID"‘ID
                                     Caitemion CenT eazeo2


                                      09/25/2013


    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in SATIMO USA using the COMOSAR test bench. All calibrationresults are traceable
    to national metrology institutions


B{Ah 3   \               SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref. ACR 2613.13.SATUA




                                  Name                              Function                          Date                  Signature
Prepared by :               Jérome LUC                        Product Manager                    9/18/2013                   fi’
Checked by :                 Jérome LUC                        Product Manager                   |9/18/2013                  fi”
Approved by :          Kim RUTKOWSKI                           Quality Manager                   |9/18/2013                / ,, ast




                                                                       Customer Name
                                                                      Shenzhen Morlab
                                                                      Communications
                                      Distribution :                   ‘Technology Co.,
                                                                                 Ltd




   Issue                 Date                                                           Modifications
     A                 9/18/2013                  Initial release




                                                              Page: 2710

                This document shall not be reproduced, except in full or in part, without the writien approval ofSATTMO
                The information contained hereinis to be used onlyfor the purpasefor which it is submitted and is not to
                be released in whole orpart without written approval ofSATRO,


     (t.                     SAR REFERENCE DIPOLE CALIBRATION REPORT                                                 Ref: ACR.2613.13.8ATU A
SATIMO
    z


                                                    TABLE OF CONTENTS



1      IMETOGUCHOM.........sssccnnssnmmccnnmmmmnnnnnmmmmnnnnnnnnmmnnnnnnnmmnnnnnnnnnnnmnnnnnmnnnnnnnnmnnnnnnmmmmmin 4
2      DGVIC@ URGOL TOB.. oo iseccmccrmrmimcsmmnicsciccermemmetcresminereessmmmmennmminnersensemees 4
3      (FrOGUCLDESCHPHION.................... cualoonnommmammmenteiatbennmmramnemeting 4
     3.1     General Information                                                                                                          4
4      Measurement Method.
     4.1      Return Loss Requirements                                                                                                     5
     4.2      Mechanical Requirements                                                                                                      5
5.     MEBSUT@MENE UNCETEAIMTY ..............cssssssccnnnnnnmmncnnnmmmnnnnnnmmnnnnnnnmnnnnnnnmmnnnenemmmnnne 3
     5.1      Return Loss                                                                                                                  5
     5.2      Dimension Measurement                                                                                                        5
     5.3      Validation Measurement                                                                                                       5
6.     (CaHDIRIION MESSUIOIONEROMIS:...... c sssessissssrremscccsmereemcrccsmemmmensstemmsrmtimenccisccieine 6
     6.1      Return Loss
     6.2      Mechanical Dimensions                                                                                                       6
7      Validation measurement...                                                                                          7
     7.1      Measurement Condition                                                                                                        7
     7.2       Head Liquid Measurement                                                                                                     7
     73       Measurement Result                                                                                                           8
     7.4      Body Measurement Result                                                                                                     9
8      List of Equipment ...




                                                                 Page: 310

                    This document shallnot be reproduced, except infill or in part, without the writen approval ofSATIMO.
                    The information contained hereinis to be usedonlyfor the purposefor which it is submitted and is not to
                    be releasedin whole orpart without written approval ofSATBAO.


    i. \                SAR REFERENCE DIPOLE CALIBRATION REPORT                                                Rer Ack 261 5 13 SaTH A



1          INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin
C and CEV/IEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.



2          DEVICE UNDER TEST




                                                  Device Under Test
        Device Type                                      COMOSAR 835 MHz REFERENCE DIPOLE
        Manufacturer                                     Satimo
        Model                                            SID835
        Serial Number                                    SN 20/08 DIPCII
        Product Condition (new / used)                   Used

A yearly calibration interval is recommended.

3          PRODUCT DESCRIPTION

34      GENERAL INFORMATION

Satimo‘s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEUIEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                              Figure 1 — Satimo COMOSAR Validation Dipole



                                                           Page: 410

              This document shall not be reproduced, except in fill or in part, without the writen approval ofSATTMO
              The information containedherein is to be usedonlyfor the purposefor which i is submitied and is not to
              be released in whole or part without written approval ofSATIMO.


    (L..                   SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR.2613.13.8ATU A
SATIMO
    —

           4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CEI/IEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

           4.1    RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of —20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

           4.2    MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEV/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5                MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

           5.1   RETURNLOSS
The following uncertainties apply to the return loss measurement:
                        Frequency band                                   Expanded Uncertainty on Return Loss
                          400—6000MHz                                                              0.1 dB

           5.2   DIMENSION MEASUREMENT

The following uncertainties apply to the dimension measurements:
                          Length (mm)                                        Expanded Uncertainty on Length
                                3 — 300                                                          0.05 mm


           5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC ENS0361 and CEMIEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.

                           Scan Volume                                      Expanded Uncertainty
                                   1 g                                                  20.3 %
                                  10 g                                                  20.1 %

                                                               Page: 510
                  This document shall not be reproduced, except infill or in part, without the writeen approval of$ATIMO
                  The information containedhereini to be used onlyfor the purpose for which it is submitted and is not to
                  be releasedin whole orpart without written approval ofSATBAO


 f
sArImo
                          SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR 261.3.13.SATU A



       6     CALIBRATION MEASUREM                                  T RE          LTS

       6.1       RETURNLOSS




       Frequency (MHz)                                  Return Loss (dB)                                    Requirement (dB)
                 835                                         —25,29                                                     —20

       6.2       MECHANICAL DIMENSIONS
 Frequency MHz                          L mm                                        h mm                                      d mm

                          required              measured                required             measured             required           measured
      300                420.0 +1 %.                                  250.0 21 %.                                635 41 %

      450                290.0 +1 %.                                   166.7 +1 %                                6.35 41 %

      750                176.0 +1 %.                                   100.0 +1 %                                635 +1 %

      #as                161.0 +1 %.               Pass                89.8 +1 %                paSs              1611%                PAS$
      900                149.0 +1 %                                    83.3 +1 %.                                 3.641%

     1450                 89.1 +1 %                                    51.7 +1 %                                  3.611%

     1500                 80.5 21 %                                    50.0 41 %.                                 3.6 +1 %

     1640                 79.0 +1%                                     45.7 +1 %                                  3.641%

     1750                 75.2 +1 %                                    42.9 +1 %.                                 3.6+1%

     1800                 72.0 11 %                                    41.7 +1 %.                                 3.6+1%

     1900                 68.0 41 %                                    39.5 41 %.                                 3.6 11 %.

     1950                 66.3 +1 %                                    38.5 +1 %.                                 3.6 11 %.

     2000                 64.5 1 %.                                    37.5 41 %.                                 3.6 11 %.

     2100                 61.0 +1 %                                    35.7 41%                                   3611%

     2300                 55.5 41 %                                    32.6 +1 %.                                 3.6 +1 %

     2450                 51.5 41 %                                    30.4 41 %.                                 3.6 11 %

     2600                 48.5 11 %                                    28.8 +1 %                                  3611%

     3000                 41.5 +1 %                                    25.0 41 %                                  36 11%

     3500                 37.0+1 %                                     26.4 +1 %                                  3.6+1%

     3700                 34.721%                                      26.4 +1 %.                                 3.611%


                                                               Page: 6/10
                 This document shall not bereproduced, except in fill or in part, without the written approval of SATIMO
                 The information containedherein is to be used only for the purpose for whichit is submitied and is not to
                 bereleasedin whole or part without written approval of SATIMO


 ("‘ 4       \             SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR 261 3.13.SATU A



         7       VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEVIEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface ofthe dipole at the described distance from the bottom surface of the phantom.

         7.1      MEASUREMENT CONDITION
Software                                                                OPENSAR V4
Phantom                                                                 SN 20/09 SAMT7I
Probe                                                                   SN 18/11 EPG122
Liguid                                                                  Head Liquid Values: eps‘ 42.6 sigma 0.88
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=Smm
Frequency                                                               835 MHz
Input power                                                             20 dBm
Liquid Temperature                                                     21 °C
Lab Temperature                                                        21 °C
Lab Humidity                                                           45%

         7.2      HEAD LIQUID MEASUREMENT

                           hedm‘®                Relative permittivity (e,)              Conductivity (0) S/m
                                                 required          measured            required         measured
                               300              as3i5%                                D87 is%
                               450              43.5 15 %                             0.87 25 %

                               750              mosy%                                 o.89 i5 %
                               s3s              nse%                  pass            0.90 15 %            Pass
                               900              41.5 5 %                              0.97 15 %

                              1aso              a05 :5 %                              120i5%
                              1500              204 5%                                125%
                              1640              40.2 45 %                             131 15%

                              1750              a01 5%                                125%
                              1800              40.0 15 %                             1.40 15 %

                              1900              a0.0 25 %                             14025 %
                              1950              40.0 25 %                             1.40 15 %

                              2000              20015 %                               L40%
                              2100              39.8 15 %                             149 15 %

                              2300              so5s%                                 Ler#%
                              2aso              30245 %                               Leo:s%
                              2600              39.0 15 %                             1.96 15 %

                              3000              385 i5%                               2405%
                              3500              37.9 15 %                             2.91 15 %


                                                               Page: 710
                  This document shall not be reproduced, except in fill oin part, without the writeenapproval ofSATIMO.
                  The information containedhereinis to be used onlyfor the purposefor which it is submitted and is not to
                  be released in whole orpart without written approval ofSATBAO


        \                SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR 261.3.13.8ATU A


      7.3    MEASUREMENT RESULT
            Std. 1528 and CEVIEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation.  All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power

                        rauie‘®                    1 5 saR (W/ke/w)                     10 g sar (w/ke/w)
                                               required           measured           required          measured
                            300                  285                                    1.94

                            450                  4.58                                   3.06
                            750                  8.49                                   5.55
                            835                  9.56            9.71 (0.97)            6.22          6.27 (0.63)

                            900                  10.9                                   6.99
                            1450                  29                                     16
                            1500                 30.5                                   16.8
                            1640                 34.2                                   184

                            1750                 36.4                                   193
                            1800                 384                                    20.1
                            1900                 39.7                                   205

                            1950                 40.5                                   20.9
                           2000                  411                                    211

                           2100                  43.6                                   21.9

                           2300                  48.7                                   233

                           2450                  524                                     24
                           2600                  55.3                                   24.6

                           3000                  63.8                                   25.7
                           3500                  67.1                                    25




                                                             Page: 810
               This document shall not be reproduced, except in full or in part, without the writen approval of SATTMO
               The information containedherein is to be used only for the purpose for which it s submitied and is not to
               bereleased in whole or part without written approval ofSATDIO


 (( Ls                      SAR REFERENCE DIPOLE CALIBRATION REPORT                                                 Ref: ACK 261.3.13. 90 TU A
SATIMO

           74   BODY MEASUREMENT RESULT
Software                                                                OPENSAR V4
Phantom                                                                 SN 20(09 SAM7]
Probe                                                                   SN 18711 EPG122
Liquid                                                                  Body Liquid Values: eps‘          55.3 sigma : 0.96
Distance between dipole center and liquid                               15.0 mm
Area scan resolution                                                    dx=8mm/dy=8mm
Zoon Scan Resolution                                                    dx=8mm/dy=8m/dz=Smm
Frequency                                                               835 MHz
Input power                                                             20 dBm
Liquid Temperature                                                      21°C
Lab Temperature                                                         21°C
Lab Humidity                                                            45%



                           F
                               TV                     18 SAR (W/ke/w)                     10 g san (w/kg/w)
                                                          measured                             measured

                               835                       10.02 (1.00)                          6.56 (0.66)




                                                ;                                                 "a




                                                               Page: 910

                  This document shall not be reproduced, except in fillor in part, without the written approval of SATIMO
                  Theinformation contained herein is to beused only for thepurpose for which i is submitted and is not to
                  be released in whole or part without written approval of SATBMO


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACR 261 3,13.SATUA




    Cmpramopr1573
   SAM Phantom                      Satimo                SN—20/08—SAM71 gfi:fid No ce                                 gfz:‘_"         No cel
|COMOSAR Test Bench]               Version 3                        NA               :gfif;‘_” Nocel                 :fq'fizzd       No call
  Network Analyzer           RhOde8Schware                     grgoor32                        o2ro13                        o2ro16
      Calipers                      Carrera                   caLiPEeR—01                      1222010                       122013
  Reference Probe                    Satimo               EPG122 SN 1871 g:’afif'c‘:mzl‘:;" z:f"’,if;':ffezfll‘:;"
     Multimeter                 Keithley 2000                    1188655                       112010                        12013
  Signal Generator            Agilent E4438C                 MY49070581                        1212010                       122013

      Ampitier                   Aethercomm                      Sm oen               OMWetotmeinven(ostNeceredured.
    Power Meter                  HP E4418A                    ussezs1498                       112010                        12013
    Power Sensor               HP ECP—E26A                    ussr1814s0                        112010                       112013
 Directional Coupler           Narda 4216—20                      o1386               z:f’afig’gfggzl‘:;" g;‘:’“fifgfg:fl;’eé‘_’
  Lmfi:i’;‘;’:n:f             Control Company                    11486149                        32012                        aror4




                                                               Page: 10/10

                  This document shall not be reproduced, except infill or in part, without the writen approval of$ATTIMO
                  The information contained herein is to be used onlyfor the purposefor which it is submittedand is not to
                  be releasedin whole orpart without written approval ofSATBMO.