SAR report part 3

FCC ID: R38YL3636A

RF Exposure Info

Download: PDF
FCCID_3294414

Schmid & Partner Engineering AG                               s       p        e       a       a

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 44 245 9700, Fax +41 44 245 9779
info@speag.com, http://www speag.com




                                        IMPORTANT NOTICE


USAGE OF THE DAE 4
The DAE unit is a delicate, high precision instrument and requires careful treatment by the user. There are no
serviceable parts inside the DAE. Special attention shall be given to the following points:

Battery Exchange: The battery cover of the DAE4 unit is closed using a screw, over tightening the screw may
cause the threads inside the DAE to wear out.

Shipping of the DAE: Before shipping the DAE to SPEAG for calibration, remove the batteries and pack the
DAE in an antistatic bag. This antistatic bag shall then be packed into a larger box or container which protects the
DAE from impacts during transportation. The package shall be marked to indicate that a fragile instrument is
inside.

E—Stop Failures: Touch detection may be malfunctioning due to broken magnets in the E—stop. Rough handling
of the E—stop may lead to damage of these magnets. Touch and collision errors are often caused by dust and dirt
accumulated in the Estop. To prevent E—stop failure, the customer shall always mount the probe to the DAE
carefully and keep the DAE unit in a non—dusty environment if not used for measurements.

Repair: Minor repairs are performed at no extra cost during the annual calibration. However, SPEAG reserves
the right to charge for any repair especially if rough unprofessional handling caused the defect.

DASY Configuration Files: Since the exact values of the DAE input resistances, as measured during the
calibration procedure of a DAE unit, are not used by the DASY software, a nominal value of 200 MOhm is given
in the corresponding configuration file.

Important Note:
Warranty and calibration is void if the DAE unit is disassembled partly or fully by the
Customer.


Important Note:
Never attempt to grease or oil the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by certified SPEAG personnel only and is part of the annual
calibration procedure.


Important Note:
To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
while turning the locking nut of the connector. The same care shall be used when
disconnecting the probe from the DAE.



Schmid & Partner Engineering


TN_BRO40315AD DAE4.doc                                                                                 11.12.2009


Callblfatlon Laboratory of                                       \\\‘\QJ//’A}                                      Schweizerischer Kalibrierdienst
Schmid & Partner                                                 ifi(&é                                             Service suisse d‘étalonnage
  Engineering AG             .      .
                                                                 T _/ g                                            Servizio svizzero di taratura
                                                                  4 n‘                                                7     i    ;       i
Zeughausstrasse 43, 8004 Zurich, Switzerland                      e/I//\\\\\\“                                     Swiss Calibration Service
                                                                      Aidats®



Accredited by the Swiss Accreditation Service (SAS)                                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       Auden                                                                              Certificate No: DAE4—905_Jun16

|CALIBRATION CERTIFICATE                                                                                                                              |
 Object                               DAE4 — SD 000 D04 BK — SN: 905


 Calibration procedure(s)             QA CAL—06.v29
                                      Calibration procedure for the data acquisition electronics (DAE)




 Calibration date:                    June 22, 2016




 This calibration certificate documents the traceability to national standards, which realize the physicalunits of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               09—Sep—15 (No:17153)                                  Sep—16


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Auto DAE Calibration Unit              SE UWS 053 AA 1001        O5—Jan—16 (in house check)                            In house check: Jan—17
 Calibrator Box V2.1                    SE UMS 006 AA 1002        05—Jan—16 (in house check)                            In house check: Jan—17




                                        Name                                    Function                                 Signature
 Calibrated by:                         Dominique Steffen                       Technician                               @



 Approved by:                           Fin Bomholt                             Deputy Technical Manager                    .         \/ é QMbu/ ’d
                                                                                                                                124    >



                                                                                                                        Issued: June 22, 2016
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE4—905_Jun16                                        Page 1 of 5


Calibration Laboratory of                                                         Schweizerischer Kalibrierdienst
Schmid & Partner                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                      Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates



Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
   e DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
         comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    e    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
         e     DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
               the nominal calibration voltage. Influence of offset voltage is included in this
               measurement.
         e     Common mode sensitivity: Influence of a positive or negative common mode voltage on
               the differential measurement.
         e     Channel separation: Influence of a voltage on the neighbor channels not subject to an
               input voltage.
         e     AD Converter Values with inputs shorted: Values on the internal AD converter
               corresponding to zero input voltage
         e     Input Offset Measurement. Output voltage and statistical results over a large number of
               zero voltage measurements.
         e     Input Offset Current: Typical value for information; Maximum channel input offset
               current, not considering the input resistance.
         e     Input resistance: Typical value for information: DAE input resistance at the connector,
               during internal auto—zeroing and during measurement.
         e     Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
               alarm signal is generated.
         +     Power consumption: Typical value for information. Supply currents in various operating
               modes.




Certificate No: DAE4—905_Jun16                                   Page 2 of 5


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          1LSB =     6.AuV ,       full range =   —100...+300 mV
       Low Range:           1LSB=       61nV ,       full range =   ~1....... +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


     Calibration Factors              X                           Y                  Z
     High Range             404.683 + 0.02% (k=2)   405.232 £0.02% (k=2)    404.815 £ 0.02% (k=2)
      Low Range             3.98074 + 1.50% (k=2)   4.00183 + 1.50% (k=2)   3.99763 + 1.50% (k=2)


Connector Angle

     Connector Angle to be used in DASY system                                   260.0"41°




Certificate No: DAE4—905_Jun16                      Page 3 of 5


Appendix (Additional assessments outside the scope of SCS0108)
1. DC Voltage Linearity
     High Range                                Reading (uV)              Difference (uV)            Error (%)
     Channel X       + Input                    199992.43                        ~1.76               —0.00
     Channel X       + Input                     20001.63                        1.32                0.01
     Channel X       — Input                    —19999.90                        2.02                —0.01
     Channel Y       + Input                    199987.76                        —6.61               —0.00
     Channel Y       + Input                     19995.86                        —4.49               —0.02
     Channel Y       — Input                    —20000.23                        1.57                —0.01
     Channel Z       + Input                    199992.39                        2.00                —0.00
     Channel Z       + Input                     19995.03                        5.12                —0.03
     Channel 2       — Input                    —20000.48                        1.42                —0.01


     Low Range                                 Reading (1V)              Difference (uV)             Error (%)
     Channel X       + Input                     2000.08                         0.04                0.00
     Channel X       + Input                      200.55                         0.07                0.03
     Channel X       — Input                      —199.28                        —0.03               0.02
     Channel Y       + Input                     2000.27                         0.08                0.00
     Channel Y       + Input                      200.65                         —0.12               —0.06
     Channel Y       — Input                      —199.27                        —0.17               0.08
     Channel Z       + Input                     2000.30                         0.24                0.01
     Channel Z       + Input                      199.95                         —0.61               —0.31

     Channel Z       — Input                      —~199.92                       —0.62               0.31



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                        Common mode                         High Range                            Low Range
                        Input Voltage (mV)              Average Reading (uV)                 Average Reading (1V)

    Channel X                           200                            8.82                        7.40
                                       — 200                          ~7.07                        —8.74

    Channel Y                           200                            8.09                        7.82
                                       — 200                          —8.60                        —8.90
    Channel Z                           200                            1.50                        1.30
                                       — 200                          —2.84                        —2.92



3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                       Input Voltage (mV)          Channel X (uV)              Channel Y (1V)    Channel Z (uV)

     Channel X                   200                         —                      5.88              ~1.06
     Channel Y                   200                     9.49                            —            6.99

     Channel Z                   200                     9.56                       7.65                   —

Certificate No: DAE4—905_Jun16                                   Page 4 of 5


4. AD—Converter Values with inputs shorted
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                   High Range (LSB)                      Low Range (LSB)

    Channel X                                              15898                              16839

    Channel Y                                              16137                              14613

    Channel Z                                              16368                              15992



5. Input Offset Measurement
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
   Input 10M

                           Average (uV)        min. Offset (uV)         max. Offset (uv)      * [(’:“,’;a'm"
    Channel X                    —~4.09              —~4.85                  —2.84                    0.31

    Channel Y                    —0.23               —1.76                    0.56                    0.38

    Channel Z                    —0.45               ~1.77                    1.62                    0.54



6. Input Offset Current
    Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                 Zeroing (kOhm)                         Measuring (MOhm)

    Channel X                                             200                                 200
    Channel Y                                             200                                 200
    Channel Z                                             200                                 200



8. Low Battery Alarm Voltage (Typical values for information)
    Typical values                                Alarm Level (VDC)
    Supply (+ Vec)                                                               +7.9

    Supply (— Vec)                                                               —7.6



9. Power Consumption (Typical values for information)
    Typical values                         Switched off (mA)        Stand by (mA)          Transmitting (mA)
    Supply (+ Vec)                                +0.01                     +6                      +14

    Supply (— Vee)                                —0.01                     —8                      —9




Certificate No: DAE4—905_Jun16                            Page 5 of 5


Calibration Laboratory of                                           «l
                                                                    «um                                          Schweizorischer Kalibrierdienst
                                                                    s




                                                                                                         w o o
                                                                m



                                                                             as
Schmid & Partner                                                                                                 Service sulsse d‘étalonnage




                                                           E
  Engineering AG                                                    zn                                           Servizlo svizzero di taratura




                                                                w



                                                                             Py
Zoughausstrasse 43, 8004 Zurich, Switzerland                        Nt
                                                                    Yihtals®
                                                                                                                 Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                                       Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agroementfor the recognition of callbration certificates

Client      Sporton—SZ (Auden)                                                                Certificate No: EX3—3911_Sep16

CALIBRATION CERTIFICATE
Object                                EX3DV4 — SN:3911


Callbration procedure(s)              QA CAL—01.v9, QA CAL—14.v4, QA CAL—23.v5, QA CAL—25.v6
                                      Calibration procedure for dosimetric E—field probes


Callbration date:                     September 29, 2016

This calibration certficate documents the traceablity to national standards, which realize the physical units of measurements (S1):
The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certiicate.

All calibrations have been conducted in the closed laboratory facilty: environment temperature (22 + 3)°C and humidiy <70%.

Callbration Equipment used (M&TE criical for callbration)


 Primary Standards                   iD                                Gal Date (Gortificate No.)                   Scheduled Calibration
 Power meter NRP                     SN:    104778                     06—Apr—16 (No. 217—02288/02209)              Aprt7
 Power sensor NRP—291                SN:    103244                     06—Apr—16 (No. 217—02288)                    Apr—iT
 Power sensor NRP—291                SN     103245                     06—Apr—16 (No. 217—02289)                    Aprt?
 Reference 20 dB Attenuator          SN:    s527? (20)                 05—Apr—16 (No. 217—02203)                    Apr7
 Reference Probe ESSDV2              SN:    3013                       31—Dec—15 (No. ES3—3013_Dect5)               Dec—16
 DAE4                                SN:    660                        23—Dec—15 (No. DAE4—660_Dec15)               Dec—16


 Secondary Standards                 iD                                Check Date (in house)                        Scheduled Check
 Power meter E44198                  SN:    GB41203874                 06—Apr—16 (in house chack Jun—16)            in house check Jun—18
 Power sensor E4412A                 SN:    MY41498087                 06—Apr—16 (in house check Jun—16)            in house check: Jun—18
 Power sensor E4412A                 SN:    000110210                  06—Apr—16 (in house check Jun—16)            in house check Jun—18
 RF generator HP 8648G               ShN:   US3642U01700               04—Aug—99 (in house check Jun—16)            in house check Jun—18
 Network Analyzer HP 8753E           SN:    Us37300585                 18—0ct—01 (in house check Oct—15)            in house check: Oct—16




                                                                                                                    =7tp
                                       Name                                   Function                                Signature
 Callbrated by:                        Leif Klysner                           Laboratory Technician



 Approved by:                          Katia Pokovic                          Technical Manager
                                                                                                           xt
                                                                                                                      Issued: October 4, 2016
 This callbration certficate shall not be reproduced exceptin full without written approval of the laboratory.




Certificate No: EX3—3911_Sep16                                       Page 1 of 11


Calibration Laboratory of                                    «05
                                                             s\                              g Schwolzerischer Kalibrierdionst
          f
Schmid & Partner                                            i‘h\t;//m                        _ Service suisse d‘étalonnage
  Engineering AG                                            Ty                               «g Servirio sviezoro di taratura
Zeughausstrasse 43, 8004 Zurich, Switzeriand                 *Afi-\"‘?                             Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of callbration certificates

Glossary:
TSL                           tissue simulating liquid
NORMxy,z                      sensitivity in free space
ConvF                         sensitivity in TSL / NORMx.y,z
DCP                           diode compression point
CF                            crest factor(1/duty_cycle) of the RF signal
A, B,C, D                     modulation dependentlinearization parameters
Polarization ip               @p rotation around probe axis
Polarization 8                8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                              ie., 8 = 0 is normal to probe axis
Connector Angle               information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged Specific
              Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
              Techniques®, June 2013
     b)       IEC 62209—1, "Procedure to measurethe Specific Absorption Rate (SAR) for hand—held devices usedin close
              proximity to the ear (frequency range of 300 MHz to 3 GHz), February 2005
     0)       IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
              used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)®, March 2010
     d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
              NORMx.y,z: Assessed for E—field polarization 8 = 0 (f < 900 MHz in TEM—cell; f> 1800 MHz: R22 waveguide).
              NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E*—field
              uncertainty inside TSL (see below ConvF).
              NORM\(x,y,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This linearization is
              implementedin DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
              in the stated uncertainty of ConvF.
              DCPxy,z: DCP are numerical linearization parameters assessed based on the data of power sweep with CW
              signal (no uncertainty required). DCP does not depend on frequency nor media.
              PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
              characteristics
              AsyaBuy,z Cuy.z: Dxy,z; VRiy,z: A, B, C, D are numerical linearization parameters assessed based on
              the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
              media. VR is the maximum calibration range expressed in RMS voltage across thediode.
              ConvF and Boundary Effect Parameters: Assessedin flat phantom using E—field (or Temperature Transfer
              Standard for f < 800 MHz) and inside waveguide using analytical field distributions based on power
              measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
              boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
              used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
              to NORMx.y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
              ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to + 100
              MHz.
              Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
              exposed by a patch antenna.
              SensorOffset: The sensoroffset corresponds to theoffset of virtual measurement center from the probe tip
              (on probe axis). No tolerance required.
              Connector Angle: The angle is assessed using the information gained by determining the NORMx (no
              uncertainty required).


Certificate No: EX3—3911_Sep16                                  Page 2 of 11


EX3DV4 — SN:3911                                                                September 29, 2016




                      Probe

                                         SN:3911

                                  Manufactured:            September 4, 2012
                                  Calibrated:              September 29, 2016




                                 Calibrated for DASY/EASY Systems
                                    {Note: non—compatible with DASY2 system!)




Certificate No: EX3—3911_Sep16                    Page 3 of 11


EX3DV4— SN:3911                                                                                                          September 29, 2016




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3911
Basic Calibration Parameters
                                                Sensor X                         Sensor Y                    Sensor Z             Unc (k=2)
 Norm (uV/(V/im)*)*                                0.30                            0.33                         0.47               +10.1 %
 DCP (myv}*                                        101.9                          102.3                        100.2

Modulation Calibration Parameters
 UID     Communication System Name                                         A           B             C          D          VR         Unc®
                                                                           dB        dBvuV                     dB          mV         (k=2)
 0             CW                                                 xX       0.0            0.0       1.0        0.00       138.8       £3.5 %
                                                                  Y        0.0            0.0       1.0                   138.9
                                                                  Z        0.0            0.0       1.0                   138.4


 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.


* The uncertainties of Norm X,Y,Z do not affect the E*—field uncertainty inside TSL (see Pages 5 and 6).
® Numerical linearization parameter: uncertainty not required.
& Uncertainty is determined using the max. deviation from linear response applying rectangular distribution and is expressed for the square of the
field value.




Certificate No: EX3—3911_Sep16                                     Page 4 of 11


EX3DV4— SN:3911                                                                                                          September 29, 2016




DASY/EASY — Parameters of Probe: EX3DV4                                                                        SN:3911


Calibration Parameter Determined in Head Tissue Simulating Media
                       Relative          Conductivity                                                                Depth °           Unc
     f(MHz)°         Permittivity®          (S/im)"          ConvFX        ConvFY         ConvFZ        Aipha®        (mm)            (k=2)

        750               41.9                0.89             10.99         10.99          10.99         0.57         0.81         *# 12.0 %

        835               41.5                0.90             10.54         10.54          10.54         0.26         1.26         # 12.0 %

        900               41.5                0.97             10.05         10.05          10.05         0.38         0.93         £#12.0 %

       1750               40.1                1.37             8.88           8.88          8.88          0.31         0.93         + 12.0 %

       1900               40.0                1.40             8.50           8.50          8.50          0.40         0.80         #12.0 %

       2000               40.0                1.40             8.48           8.48          8.48          0.35         0.85         #* 12.0 %

       2300               39.5                1.67             7.93           7.93           7.93         0.36         0.80         + 12.0 %

       2450               39.2                1.80             7.43           7.A3          7.43          0.29         0.98         + 12.0 %

       2600               39.0                1.96             7.39           7.39           7.39         0.45         0.80         + 12.0 %

       5250               35.9                4.71             5.25           5.25           5.25         0.40          1.80        £# 13.1 %

       5600               35.5                5.07             4.49           4.49          4.49          0.50          1.80        + 13.1 %

       5750               35.4                5.22             4.75           4.75           4.75         0.50          1.80        *# 13.1 %



© Frequency validity above 300 MHz of £ 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is * 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to £ 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (s and 0) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: EX3—3911_Sep16                                     Page 5 of 11


EX3DV4— SN:3911                                                                                                          September 29, 2016




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3911


Calibration Parameter Determined in Body Tissue Simulating Media
                        Relative         Conductivity                                                                 Depth °          Unc
     f(MHz)°         Permittivity"®          (Sim}"          ConvF X       ConvF Y¥       ConvFZ        Alpha®         (mm)           (k=2)

        750               55.5                0.96             10.43         10.43          10.43         0.42         0.80         + 12.0 %

        835               55.2                0.97             10.19         10.19          10.19         0.20          1.33        + 12.0 %

        1750              53.4                1.49             8.46           8.46           8.46         0.42         0.80         £# 12.0 %

        1900              53.3                1.52             8.17           8.17           8.17         0.35         0.97         + 12.0 %

       2300               52.9                1.81             7.93           7.93           7.93         0.33         0.98         #12.0 %

       2450               52.7                1.95             7.66           7.66           7.66         0.43         0.80         + 12.0 %

       2600               52.5                216              7.38           7.38           7.38         0.33         0.80         +12.0 %
       5250               48.9                5.36             4.62           4.62           4.62         0.50          1.90        *13.1 %
        5600              48.5                5.77              3.78          3.78           3.78         0.60          1.90        £#13.1 %

       5750               48.3                5.94              3.95          3.95           3.95         0.60          1.90        £13.1 %



© Frequency validity above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to £ 110 MHz.
" At frequencies below 3 GHz, the validity of tissue parameters (e and 0) can be relaxed to + 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (s and 0) is restricted to + 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
8 Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: EX3—3911_Sep16                                     Page 6 of 11


EX3DV4— SN:3911                                                                                                     September 29, 2016


                                                        Frequency Response of E—Field
                                                            (TEM—Cell:if110 EXX, Waveguide: R22)

                                            Termmert
          Frequency response (normalized)




                                                       Uncertainty of Frequency Response of E—field: £ 6.3% (k=2)




Certificate No: EX3—3911_Sep16                                                Page 7 of 11


EX3DV4— SN:3911                                                                                    September 29, 2016




                                     Receiving Pattern (¢), 8 = 0°

                        £=600 MHz,TEM                                         f=1800 MHz,R22




           im                                        o



                 as .              vew#        is                       mss                 *         (ars


            ®             ®
                               o*                    #
           To:             x              ¥          2




                        |0‘FTMLZ                 Gfi%z                   1865.de                 2555!#)'11


                                   Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Certificale No: EX3—3911_Sep16                           Page 8 of 11


EX3DV4— SN:3911                                                                                  September 29, 2016




                                    Dynamic Range f(SARneaq)
                                             (TEM cell , fova= 1900 MHz)




                       105




                  3
                  ®
                   8
                  &
                  UJ
                  2 10
                  &




                       10°




                        10‘
                                        1          U             T       T            I
                              10       102         101          10       10‘          10     10°
                                             8           SAR (mWiom3]
                                             L4]
                                      not compensated                    compensated




                       a
                       &
                       5E
                       d


                                                                19°     10       1J0?      103
                                                         SAR {mW/om3]
                                       not compansated                  compensated

                                   Uncertainty of Linearity Assessment: £ 0.6% (k=2)




Certificate No: EX3—3911_Sep16                           Page 9 of 11


EXGDV4— SN:3911                                                                                                                                 September 29, 2016

                                              Conversion Factor Assessment
                  1= 835 MHz.WGLS R9 (H_conv)                                                      f= 1900 MHz,WGLS R22 (H_convr)
         «0                                                                     ‘


         as|                                                                    ‘         ®
         so: \                                                                            x

                  %
      §"t             "                                                                  gm|
                                                                                         ?
                                                                                                   >

       §.}
        16
                                                                                         5
              :                                *                                           ml
         18
         asl                                                  mss                            S
         pitoacimetrce on oi n                                sastoraca uce                  q cficfecitnpic ic                   ie is            .;      &
              iob         &               m        &0
                                               ainn)
                                                        &     0#      m         m              $       4C          aC        CA     C
                                                                                                                                  «finnt
                                                                                                                                                  %            in
                          81                                  _A                                                   al                            a)
                      ardhiea                               messued                                              anahiea                        meavees



                                          Deviation from Isotropy in Liquid
                                                               Error (¢, 9), £= 900 MHz
                              eviation:




                                     —10       —08      —06    .04        —02       00       02             04          06        08       10
                                          Uncertainty of Spherical Isotropy Assessment: # 2.6% (k=2)




Certificate No: EX3—3911_Sep16                                             Page 10 of 11


EX3DV4— SN:3911                                              September 29, 2016




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3911

Other Probe Parameters
 Sensor Arrangement                                                Triangular
 Connector Angle (°)                                                       80
 Mechanical Surface Detection Mode                                    enabled
 Optical Surface Detection Mode                                      disabled
 Probe Overall Length                                                 337 mm
 Probe Body Diameter                                                   10 mm
 Tip Length                                                             9 mm
 Tip Diameter                                                         2.5 mm
 Probe Tip to Sensor X Calibration Point                                1 mm
 Probe Tip to Sensor Y Calibration Point                                1 mm
 Probe Tip to Sensor Z Calibration Point                                1 mm
 Recommended Measurement Distance from Surface                        1.4 mm




Certificate No: EX3—3911_Sep16               Page 11 of 11



Document Created: 2017-12-23 07:27:54
Document Modified: 2017-12-23 07:27:54

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC