18.2_SAR Test Data Plots

FCC ID: QYLAP6255BC03

RF Exposure Info

Download: PDF
FCCID_4263636

                                                                           Date: 2019/3/1
WLAN 802.11b_Body_Edge1_CH 11_Main_5mm
Communication System: Wi-Fi; Frequency: 2462 MHz; Duty Cycle: 1:1
Medium parameters used: f = 2462 MHz; σ = 1.99 S/m; εr = 53.645; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.6°C; Liquid temperature: 22.2°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(7.59, 7.59, 7.59); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (71x101x1): Interpolated grid: dx=12 mm, dy=12 mm
Maximum value of SAR (interpolated) = 0.0223 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 1.299 V/m; Power Drift = 0.12 dB
Peak SAR (extrapolated) = 0.0220 W/kg
SAR(1 g) = 0.012 W/kg; SAR(10 g) = 0.00666 W/kg
Maximum value of SAR (measured) = 0.0162 W/kg




             0 dB = 0.0162 W/kg = -17.90 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11b_Body_Edge2_CH 11_Main_5mm
Communication System: Wi-Fi; Frequency: 2462 MHz; Duty Cycle: 1:1
Medium parameters used: f = 2462 MHz; σ = 1.99 S/m; εr = 53.645; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.6°C; Liquid temperature: 22.2°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(7.59, 7.59, 7.59); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (61x81x1): Interpolated grid: dx=12 mm, dy=12 mm
Maximum value of SAR (interpolated) = 0.129 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 3.572 V/m; Power Drift = 0.15 dB
Peak SAR (extrapolated) = 0.160 W/kg
SAR(1 g) = 0.091 W/kg; SAR(10 g) = 0.048 W/kg
Maximum value of SAR (measured) = 0.128 W/kg




             0 dB = 0.128 W/kg = -8.93 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11b_Body_Edge3_CH 11_Main_5mm
Communication System: Wi-Fi; Frequency: 2462 MHz; Duty Cycle: 1:1
Medium parameters used: f = 2462 MHz; σ = 1.99 S/m; εr = 53.645; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.6°C; Liquid temperature: 22.2°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(7.59, 7.59, 7.59); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (61x91x1): Interpolated grid: dx=12 mm, dy=12 mm
Maximum value of SAR (interpolated) = 0.147 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 3.947 V/m; Power Drift = 0.12 dB
Peak SAR (extrapolated) = 0.166 W/kg
SAR(1 g) = 0.097 W/kg; SAR(10 g) = 0.055 W/kg
Maximum value of SAR (measured) = 0.125 W/kg




             0 dB = 0.125 W/kg = -9.03 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11b_Body_Edge4_CH 11_Main_5mm
Communication System: Wi-Fi; Frequency: 2462 MHz; Duty Cycle: 1:1
Medium parameters used: f = 2462 MHz; σ = 1.99 S/m; εr = 53.645; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.6°C; Liquid temperature: 22.2°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(7.59, 7.59, 7.59); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x81x1): Interpolated grid: dx=12 mm, dy=12 mm
Maximum value of SAR (interpolated) = 0.0130 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 1.829 V/m; Power Drift = 0.14 dB
Peak SAR (extrapolated) = 0.0130 W/kg
SAR(1 g) = 0.00781 W/kg; SAR(10 g) = 0.00518 W/kg
Maximum value of SAR (measured) = 0.0102 W/kg




             0 dB = 0.0102 W/kg = -19.91 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11b_Body_Rear_CH 11_Main_5mm
Communication System: Wi-Fi; Frequency: 2462 MHz; Duty Cycle: 1:1
Medium parameters used: f = 2462 MHz; σ = 1.99 S/m; εr = 53.645; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.6°C; Liquid temperature: 22.2°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(7.59, 7.59, 7.59); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x71x1): Interpolated grid: dx=12 mm, dy=12 mm
Maximum value of SAR (interpolated) = 0.0825 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 2.571 V/m; Power Drift = 0.14 dB
Peak SAR (extrapolated) = 0.0970 W/kg
SAR(1 g) = 0.057 W/kg; SAR(10 g) = 0.036 W/kg
Maximum value of SAR (measured) = 0.0768 W/kg




             0 dB = 0.0768 W/kg = -11.15 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.3G_Body_Edge1_CH 64_Main_5mm
Communication System: WLAN 5G; Frequency: 5320 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5320 MHz; σ = 5.378 S/m; εr = 48.854; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.3°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.65, 4.65, 4.65); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (91x121x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.101 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 3.559 V/m; Power Drift = -0.12 dB
Peak SAR (extrapolated) = 0.384 W/kg
SAR(1 g) = 0.063 W/kg; SAR(10 g) = 0.026 W/kg
Maximum value of SAR (measured) = 0.0908 W/kg




             0 dB = 0.0908 W/kg = -10.42 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.3G_Body_Edge2_CH 64_Main_5mm
Communication System: WLAN 5G; Frequency: 5320 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5320 MHz; σ = 5.378 S/m; εr = 48.854; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.3°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.65, 4.65, 4.65); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (101x121x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.220 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 1.927 V/m; Power Drift = 0.16 dB
Peak SAR (extrapolated) = 0.270 W/kg
SAR(1 g) = 0.075 W/kg; SAR(10 g) = 0.031 W/kg
Maximum value of SAR (measured) = 0.151 W/kg




             0 dB = 0.151 W/kg = -8.21 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.3G_Body_Edge3_CH 64_Main_5mm
Communication System: WLAN 5G; Frequency: 5320 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5320 MHz; σ = 5.378 S/m; εr = 48.854; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.3°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.65, 4.65, 4.65); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (111x111x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0432 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 1.247 V/m; Power Drift = 0.15 dB
Peak SAR (extrapolated) = 0.238 W/kg
SAR(1 g) = 0.015 W/kg; SAR(10 g) = 0.00455 W/kg
Maximum value of SAR (measured) = 0.0323 W/kg




             0 dB = 0.0323 W/kg = -14.91 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.3G_Body_Edge4_CH 64_Main_5mm
Communication System: WLAN 5G; Frequency: 5320 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5320 MHz; σ = 5.378 S/m; εr = 48.854; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.3°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.65, 4.65, 4.65); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (91x81x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0377 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 1.146 V/m; Power Drift = 0.13 dB
Peak SAR (extrapolated) = 0.0350 W/kg
SAR(1 g) = 0.015 W/kg; SAR(10 g) = 0.012 W/kg
Maximum value of SAR (measured) = 0.0187 W/kg




             0 dB = 0.0187 W/kg = -17.28 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.3G_body_Rear_CH 64_Main_5mm
Communication System: WLAN 5G; Frequency: 5320 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5320 MHz; σ = 5.378 S/m; εr = 48.854; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.3°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.65, 4.65, 4.65); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x81x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0667 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.838 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 0.108 W/kg
SAR(1 g) = 0.036 W/kg; SAR(10 g) = 0.020 W/kg
Maximum value of SAR (measured) = 0.0608 W/kg




             0 dB = 0.0608 W/kg = -12.16 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.5G_Body_Edge1_CH 124_Main_5mm
Communication System: WLAN 5G; Frequency: 5620 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5620 MHz; σ = 5.859 S/m; εr = 47.904; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.4°C; Liquid temperature: 22.1°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.06, 4.06, 4.06); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (91x121x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.236 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 4.905 V/m; Power Drift = -0.18 dB
Peak SAR (extrapolated) = 0.387 W/kg
SAR(1 g) = 0.114 W/kg; SAR(10 g) = 0.043 W/kg
Maximum value of SAR (measured) = 0.245 W/kg




             0 dB = 0.245 W/kg = -6.11 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.5G_Body_Edge2_CH 124_Main_5mm
Communication System: WLAN 5G; Frequency: 5620 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5620 MHz; σ = 5.859 S/m; εr = 47.904; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.4°C; Liquid temperature: 22.1°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.06, 4.06, 4.06); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x91x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.193 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 3.376 V/m; Power Drift = 0.12 dB
Peak SAR (extrapolated) = 0.289 W/kg
SAR(1 g) = 0.095 W/kg; SAR(10 g) = 0.038 W/kg
Maximum value of SAR (measured) = 0.176 W/kg




             0 dB = 0.176 W/kg = -7.54 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.5G_Body_Edge3_CH 124_Main_5mm-0306
Communication System: WLAN 5G; Frequency: 5620 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5620 MHz; σ = 5.859 S/m; εr = 47.904; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.4°C; Liquid temperature: 22.1°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.06, 4.06, 4.06); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x91x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0283 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.311 V/m; Power Drift = 0.19 dB
Peak SAR (extrapolated) = 0.0860 W/kg
SAR(1 g) = 0.00534 W/kg; SAR(10 g) = 0.00198 W/kg
Maximum value of SAR (measured) = 0.0504 W/kg




             0 dB = 0.0504 W/kg = -12.98 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.5G_Body_Edge4_CH 124_Main_5mm
Communication System: WLAN 5G; Frequency: 5620 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5620 MHz; σ = 5.859 S/m; εr = 47.904; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.4°C; Liquid temperature: 22.1°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.06, 4.06, 4.06); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (101x101x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0522 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 1.650 V/m; Power Drift = 0.16 dB
Peak SAR (extrapolated) = 0.0680 W/kg
SAR(1 g) = 0.00471 W/kg; SAR(10 g) = 0.00117 W/kg
Maximum value of SAR (measured) = 0.0637 W/kg




             0 dB = 0.0637 W/kg = -11.96 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.5G_body_Rear_CH 124_Main_5mm
Communication System: WLAN 5G; Frequency: 5620 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5620 MHz; σ = 5.859 S/m; εr = 47.904; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 22.4°C; Liquid temperature: 22.1°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.06, 4.06, 4.06); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x81x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0887 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.943 V/m; Power Drift = 0.12 dB
Peak SAR (extrapolated) = 0.160 W/kg
SAR(1 g) = 0.042 W/kg; SAR(10 g) = 0.019 W/kg
Maximum value of SAR (measured) = 0.0768 W/kg




             0 dB = 0.0768 W/kg = -11.15 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.8G_Body_Edge1_CH 157_Main_5mm
Communication System: WLAN 5G; Frequency: 5785 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5785 MHz; σ = 6.128 S/m; εr = 47.382; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 21.6°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.38, 4.38, 4.38); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (91x111x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.187 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.675 V/m; Power Drift = -0.13 dB
Peak SAR (extrapolated) = 0.317 W/kg
SAR(1 g) = 0.068 W/kg; SAR(10 g) = 0.031 W/kg
Maximum value of SAR (measured) = 0.138 W/kg




             0 dB = 0.138 W/kg = -8.60 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.8G_Body_Edge2_CH 157_Main_5mm
Communication System: WLAN 5G; Frequency: 5785 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5785 MHz; σ = 6.128 S/m; εr = 47.382; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 21.6°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.38, 4.38, 4.38); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x91x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.106 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 3.127 V/m; Power Drift = 0.14 dB
Peak SAR (extrapolated) = 0.294 W/kg
SAR(1 g) = 0.063 W/kg; SAR(10 g) = 0.031 W/kg
Maximum value of SAR (measured) = 0.124 W/kg




             0 dB = 0.124 W/kg = -9.07 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.8G_Body_Edge3_CH 157_Main_5mm
Communication System: WLAN 5G; Frequency: 5785 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5785 MHz; σ = 6.128 S/m; εr = 47.382; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 21.6°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.38, 4.38, 4.38); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x91x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0499 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.561 V/m; Power Drift = 0.13 dB
Peak SAR (extrapolated) = 0.0860 W/kg
SAR(1 g) = 0.0027 W/kg; SAR(10 g) = 0.000818 W/kg
Maximum value of SAR (measured) = 0.0628 W/kg




             0 dB = 0.0628 W/kg = -12.02 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.8G_Body_Edge4_CH 157_Main_5mm
Communication System: WLAN 5G; Frequency: 5785 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5785 MHz; σ = 6.128 S/m; εr = 47.382; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 21.6°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.38, 4.38, 4.38); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (81x81x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0309 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 1.150 V/m; Power Drift = 0.18 dB
Peak SAR (extrapolated) = 0.0830 W/kg
SAR(1 g) = 0.021 W/kg; SAR(10 g) = 0.014 W/kg
Maximum value of SAR (measured) = 0.0458 W/kg




             0 dB = 0.0458 W/kg = -13.39 dBW/kg


                                                                           Date: 2019/3/1
WLAN 802.11a 5.8G_body_Rear_CH 157_Main_5mm
Communication System: WLAN 5G; Frequency: 5785 MHz; Duty Cycle: 1:1
Medium parameters used: f = 5785 MHz; σ = 6.128 S/m; εr = 47.382; ρ = 1000 kg/m3
Phantom section: Flat Section
Ambient temperature: 21.6°C; Liquid temperature: 21.9°C

DASY5 Configuration:
   Probe: EX3DV4 - SN3770; ConvF(4.38, 4.38, 4.38); Calibrated: 2018/4/25;
   Sensor-Surface: 2mm (Mechanical Surface Detection)
   Electronics: DAE4 Sn856; Calibrated: 2018/4/21
   Phantom: ELI
   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)
Area Scan (71x71x1): Interpolated grid: dx=10 mm, dy=10 mm
Maximum value of SAR (interpolated) = 0.0752 W/kg

Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=2mm
Reference Value = 2.268 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 0.177 W/kg
SAR(1 g) = 0.037 W/kg; SAR(10 g) = 0.018 W/kg
Maximum value of SAR (measured) = 0.0590 W/kg




             0 dB = 0.0590 W/kg = -12.29 dBW/kg