SAR Report Appendix A

FCC ID: QYL8265FB1

RF Exposure Info

Download: PDF
FCCID_4427899

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/8/12


System Check_Head_2450MHz
DUT: D2450V2-736

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: HSL_2450_190812 Medium parameters used : f = 2450 MHz; σ = 1.773 S/m; εr = 38.581; ρ =
1000 kg/m3
Ambient Temperature:23.6 ℃; Liquid Temperature:22.6 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN7306;ConvF(7.48, 7.48, 7.48) @ 2450 MHz;Calibrated: 2019/7/22
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn495; Calibrated: 2019/5/21
- Phantom: ELI v4.0_Left; Type: QDOVA001BB; Serial: TP:1029
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 25.6 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 116.9 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 28.9 W/kg
SAR(1 g) = 13.8 W/kg; SAR(10 g) = 6.35 W/kg
Maximum value of SAR (measured) = 23.1 W/kg




                    0 dB = 23.1 W/kg = 13.64 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/8/14


System Check_Head_5250MHz
DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190814 Medium parameters used : f = 5250 MHz; σ = 4.74 S/m; εr = 36.153; ρ = 1000
kg/m3
Ambient Temperature:23.5 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN7306;ConvF(5.34, 5.34, 5.34) @ 5250 MHz;Calibrated: 2019/7/22
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn495; Calibrated: 2019/5/21
- Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1238
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.3 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 68.10 V/m; Power Drift = -0.08 dB
Peak SAR (extrapolated) = 30.9 W/kg
SAR(1 g) = 7.95 W/kg; SAR(10 g) = 2.28 W/kg
Maximum value of SAR (measured) = 19.8 W/kg




                    0 dB = 19.8 W/kg = 12.97 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/8/14


System Check_Head_5600MHz
DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190814 Medium parameters used: f = 5600 MHz; σ = 5.102 S/m; εr = 35.707; ρ = 1000
kg/m3
Ambient Temperature:23.5 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN7306;ConvF(4.79, 4.79, 4.79) @ 5600 MHz;Calibrated: 2019/7/22
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn495; Calibrated: 2019/5/21
- Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1238
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 22.6 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 64.53 V/m; Power Drift = 0.11 dB
Peak SAR (extrapolated) = 35.2 W/kg
SAR(1 g) = 8.33 W/kg; SAR(10 g) = 2.36 W/kg
Maximum value of SAR (measured) = 21.5 W/kg




                    0 dB = 21.5 W/kg = 13.32 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/8/14


System Check_Head_5750MHz
DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190814 Medium parameters used: f = 5750 MHz; σ = 5.252 S/m; εr = 35.556; ρ = 1000
kg/m3
Ambient Temperature:23.5 ℃; Liquid Temperature:22.5 ℃

DASY5 Configuration:
- Probe: EX3DV4 - SN7306;ConvF(4.93, 4.93, 4.93) @ 5750 MHz;Calibrated: 2019/7/22
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE3 Sn495; Calibrated: 2019/5/21
- Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1238
- Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.4 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 61.24 V/m; Power Drift = 0.13 dB
Peak SAR (extrapolated) = 33.9 W/kg
SAR(1 g) = 7.73 W/kg; SAR(10 g) = 2.19 W/kg
Maximum value of SAR (measured) = 20.2 W/kg




                    0 dB = 20.2 W/kg = 13.05 dBW/kg



Document Created: 2019-09-05 00:41:56
Document Modified: 2019-09-05 00:41:56