SAR Report 19.4 DAE4

FCC ID: QOQWT41

RF Exposure Info

Download: PDF
FCCID_2106366

Schmid & Partner Engineering AG                               s       p       e        a      g

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 44 245 9700, Fax +41 44 245 9779
info@speag.com, http://www.speag.com




                                        IMPORTANT NOTICE


USAGE OF THE DAE 4

The DAE unit is a delicate, high precision instrument and requires careful treatment by the user. There are no
serviceable parts inside the DAE. Special attention shall be given to the following points:

Battery Exchange: The battery cover of the DAE4 unit is closed using a screw, over tightening the screw may
cause the threads inside the DAE to wear out.

Shipping of the DAE: Before shipping the DAE to SPEAG for calibration, remove the batteries and pack the
DAE in an antistatic bag. This antistatic bag shall then be packed into a larger box or container which protects the
DAE from impacts during transportation. The package shall be marked to indicate that a fragile instrument is
inside.

E—Stop Failures: Touch detection may be malfunctioning due to broken magnets in the E—stop. Rough handling
of the E—stop may lead to damage of these magnets. Touch and collision errors are often caused by dust and dirt
accumulated in the Estop. To prevent Estop failure, the customer shall always mount the probe to the DAE
carefully and keep the DAE unit in a non—dusty environmentif not used for measurements.

Repair: Minor repairs are performed at no extra cost during the annual calibration. However, SPEAG reserves
the right to charge for any repair especially if rough unprofessional handling caused the defect.

DASY Configuration Files: Since the exact values of the DAE input resistances, as measured during the
calibration procedure of a DAE unit, are not used by the DASY software, a nominal value of 200 MOhm is given
in the corresponding configuration file.

Important Note:
Warranty and calibration is void if the DAE unit is disassembled partly or fully by the
Customer.


Important Note:
Never attempt to grease or oil the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by certified SPEAG personnel only and is part of the annual
calibration procedure.


Important Note:
To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
while turning the locking nut of the connector. The same care shall be used when
disconnecting the probe from the DAE.



Schmid & Partner Engineering


TN_BRO40315AD DAE4.doc                                                                                 11.12.2009


Calibration Laboratory of                                          l4
                                                                  w \\_// 7                                       Schweizerischer Kalibrierdienst
Schmid & Partner                                                ib%./{&                                           Service suisse d‘étalonnage
  Engineering AG                                                 ;///T\\ s                                        Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      +/’@\\\\\?                                      Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                           Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Client       CCS—TW (Auden)                                                                   Certificate No: DAE4—877_Mar13

CALIBRATION CERTIFICATE
 Object                               DAE4 — SD 000 DO4 BJ — SN: 877


 Calibration procedure(s)             QA CAL—06.v25
                                      Calibration procedure for the data acquisition electronics (DAE)




 Calibration date:                    March 12, 2013




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                     Cal Date (Certificate No.)                            Scheduled Calibration
 Keithley Multimeter Type 2001         SN: 0810278               02—Oct—12 (No:12728)                                  Oct—13


 Secondary Standards                    ID #                     Check Date (in house)                                 Scheduled Check
 Auto DAE Calibration Unit              SE UWS 053 AA 1001       07—Jan—13 (in house check)                            In house check: Jan—14
 Calibrator Box V2.1                    SE UMS 006 AA 1002       07—Jan—13 (in house check)                            In house check: Jan—14




                                        Name                                  Function                                  Signature
 Calibrated by:                        R.Mayoraz                              Technician                                  —g
                                                                                                                           PR ublzzzarcey

 Approved by:                          Fin Bomholt                             Deputy Technical Manager
                                                                                                                          iv. E& [Uuwr
                                                                                                                        Issued: March 12, 2013
 This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.




Certificate No: DAE4—877_Mar13                                        Page 1 of 5


    .       .                                                  y
Callbl_'atlon Laboratory of                                  S\\\\\t/_///,/’}    gNiSe    G   Schweizerischer Kalibrierdienst
Schmid & Partner                                            e rereal              ¢=          Service suisse d‘étalonnage
  Engineering AG                                            2‘3/%&;             y Efm 6   C   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 +/7/\\\\\\\3                 S   gwiss Calibration Service
                                                                 "thilals


Accredited by the Swiss Accreditation Service (SAS)                                       Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates


Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
   e  DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
     comparison with a calibrated instrument traceable to national standards. The figure given
     corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    e    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
                DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
                the nominal calibration voltage. Influence of offset voltage is included in this
                measurement.
                Common mode sensitivity: Influence of a positive or negative common mode voltage on
                the differential measurement.
                Channel separation: Influence of a voltage on the neighbor channels not subject to an
                input voltage.
                AD Converter Values with inputs shorted: Values on the internal AD converter
                corresponding to zero input voltage
                Input Offset Measurement: Output voltage and statistical results over a large number of
                zero voltage measurements.
                Input Offset Current: Typical value for information; Maximum channel input offset
                current, not considering the input resistance.
                Input resistance: Typical value for information: DAE input resistance at the connector,
                during internal auto—zeroing and during measurement.
                Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
                alarm signal is generated.
                Power consumption: Typical value for information. Supply currents in various operating
                modes.




Certificate No: DAE4—877_Mar13                                   Page 2 of 5


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          1LSB=      6.A1uV ,      full range =   —100...+300 mV
       Low Range:           1LSB =      61nV ,       full range =   ~1....... +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


     Calibration Factors             X                        ¥                         Z

     High Range             405.083 + 0.02% (k=2)   404.634 + 0.02% (k=2)   405.027 + 0.02% (k=2)
     Low Range              3.99415 + 1.55% (k=2)   3.98247 + 1.55% (k=2)   3.97249 + 1.55% (k=2)


Connector Angle

     Connector Angle to be used in DASY system                                       207 5°%1°




Certificate No: DAE4—877_Mar13                      Page 3 of 5


Appendix

1. DC Voltage Linearity
     High Range                                Reading (uV)         Difference (uV)            Error (%)
     Channel X       + Input                    199995.84                   0.25                0.00
     Channel X       + Input                     20005.32                   5.02                0.03
     Channel X       — Input                    —19997.62                   3.25                —0.02
     Channel Y       + Input                    199996.35                   0.78                0.00
     Channel Y       + Input                     20001.76                   1.50                0.01
     Channel Y       — Input                    —20000.68                   0.34                —0.00
     Channel Z       + Input                    199997.51                   1.67                0.00
     Channel Z       + Input                     20001.32                   112                 0.01
     Channel Z       — Input                    —20000.25                   0.82                —0.00


     Low Range                                 Reading (uV)         Difference (uV)            Error (%)
     Channel X       + Input                     2001.34                    0.75                0.04
     Channel X       + Input                      202.36                    1.26                0.62
     Channel X       — Input                     ~198.72                    0.08                —0.04
     Channel Y       + Input                     2000.65                    0.05                0.00
     Channel Y       + Input                      200.73                    —0.40               —0.20
     Channel Y       — Input                     —200.37                    ~1.50               0.76
     Channel Z       + Input                     2002.58                    2.02                0.10
     Channel Z       + Input                      200.28                    —0.70               —0.35

     Channel Z       — Input                     —199.94                    —0.98               0.49



2. Common mode sensitivity
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                        Common mode                       High Range                         Low Range
                        Input Voltage (mV)            Average Reading (uV)              Average Reading (uV)

    Channel X                           200                      14.89                        13.20
                                       — 200                    ~11.48                       —13.23
    Channel Y                           200                     —20.91                       —20.98
                                       — 200                     19.14                        18.93
    Channel Z                           200                      21.06                        20.36
                                       —200                     —21.86                       —22.39


3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                       Input Voltage (mV)         Channel X (uV)          Channel Y (uV)    Channel Z (uV)

    Channel X                    200                        —                  0.73              —3.72
    Channel Y                    200                   7.02                         —             2.20
    Channel Z                    200                   9.39                    417                    —


Certificate No: DAE4—877_Mar13                              Page 4 of 5


4. AD—Converter Values with inputs shorted
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                   High Range (LSB)                   Low Range (LSB)

    Channel X                                              15990                           16717

    Channel Y                                              15873                           16217

    Channel Z                                              15663                           14596



5. Input Offset Measurement
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
   Input 10MQ

                           Average (uV)        min. Offset (uV)      max. Offset (uV)      Sid. I?:‘\Il;ation

    Channel X                    0.72                —0.51                2.09                     0.49
    Channel Y                    —0.72               —1.97                0.48                     0.56

    Channel Z                    —0.85               —2.84                0.95                     0.70


6. Input Offset Current
    Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                 Zeroing (kOhm)                      Measuring (MOhm)
    Channel X                                             200                              200

    Channel Y                                             200                              200
    Channel Z                                             200                              200



8. Low Battery Alarm Voltage (Typical values for information)
    Typical values                                Alarm Level (VDC)
    Supply (+ Vec)                                                            +7.9

    Supply (— Vec)                                                            —1.6



9. Power Consumption (Typical values for information)
    Typical values                         Switched off (mA)        Stand by (mA)       Transmitting (mA)

    Supply (+ Vec)                                +0.01                  +6                      +14
    Supply (— Vec)                                —0.01                  —8                      —9




Certificate No: DAE4—877_Mar13                             Page 5 of 5