Test Report SAR Calibration Part 11

FCC ID: QISVOG-L0J

RF Exposure Info

Download: PDF
FCCID_4250777















             Justification of the extended calibration of Dipole D1750V2 SN:1123

 Per KDB 865664, we have measured the Impedance and Return Loss as below.
 1) The most recent return-loss result, measured at least annually, deviates by no more than 20%
    from the previous measurement.
 2) The most recent measurement of the real or imaginary parts of the impedance, measured at least
    annually is within 5 ohm from the previous measurement.


  Dipole 835 Head TST       Target Value             Measured Value                Difference

Impedance transformed to
                            50.6Ω-0.8j Ω             49.93Ω-1.83j Ω            R=-0.67Ω,X=-1.03j Ω
       feed point

      Return Loss             -40.2dB                  -39.17 dB                     -2.56%

  Dipole 835 Body TST       Target Value             Measured Value                Difference

Impedance transformed to
                            46.9Ω-1.0j Ω             46.05Ω+2.83j Ω            R=-0.85Ω,X=3.83j Ω
       feed point

      Return Loss             -29.5 dB                 -28.05 dB                     -4.92%

    Measured Date            2017-07-27                2018-07-17                     -------


              Impedance Test-Head                             Return Loss Test-Head




              Impedance Test-Body                             Return Loss Test-Body


                Justification of the extended calibration of Dipole D1900V2 SN:5d143

    Per KDB 865664, we have measured the Impedance and Return Loss as below.
    1) The most recent return-loss result, measured at least annually, deviates by no more than 20%
       from the previous measurement.
    2) The most recent measurement of the real or imaginary parts of the impedance, measured at least
       annually is within 5 ohm from the previous measurement.


 Dipole 1900 Head TST          Target Value          Measured Value                    Difference

Impedance transformed to
                               53.4Ω+4.3j Ω           49.75Ω+7.52j Ω               R=-3.65Ω,X=3.22Ω
       feed point

      Return Loss                -25.6dB                -25.32 dB                       -1.09%

 Dipole 1900 Body TST          Target Value          Measured Value                    Difference

Impedance transformed to
                               49.1Ω+5.9j Ω           46.18Ω+2.21j Ω               R=-2.92Ω,X=-3.69Ω
       feed point

      Return Loss                -24.4 dB                -24.5 dB                       0.41%

    Measured Date               2017-09-20             2018-09-20                        -------


                    Impedance Test-Head                             Return Loss Test-Head




                    Impedance Test-Body                             Return Loss Test-Body