Test Report SAR Calibration Part 10

FCC ID: QISVOG-L0J

RF Exposure Info

Download: PDF
FCCID_4250750

Measurement Conditions
   DASY system configuration, as far as not iven on page 1.
     DASY Version                                              DASY5                              V52.10.1
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         2600 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.0               1.96 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   37.4 +# 6 %        2.01 mho/m + 6 %
     Head TSL temperature change during test                      <0.5 °C            =———                    =———


SAR result with Head TSL

     SAR averaged over 1 cm*° (1 g) of Head TSL                 Condition
     SAR measured                                        250 mW input power                     14.4 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        56.4 W/ikg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                     6.39 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        25.2 Wikg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity

     Nominal Body TSL parameters                                  22.0 °C           52.5               2.16 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   51.2 +6 %          2.18 mho/m + 6 %
     Body TSL temperature change during test                      «0.5 °C

SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        250 mW input power                     13.9 W/kg

     SAR for nominal Body TSL parameters                     normalized to 1W         55.1 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        250 mW input power                     6.17 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         24.5 Wikg + 16.5 % (k=2)



Certificate No: D2600V2—1032_Sep18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           49.9 0 — 5.0 |Q
      Return Loss                                                                       — 26.0 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           46.0 O — 3.5 |Q
      Return Loss                                                                      — 25.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.153 ns

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG

      Manufactured on                                                                March 03, 2009




Certificate No: D2600V2—1032_Sep18                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                       Date: 17.09.2018
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1032

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.01 S/m; & = 37.4; p = 1000 kg/m>
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    e     Probe: EX3DV4 — SN7349; ConvF(7.7, 7.7, 7.7) @ 2600 MHz; Calibrated: 30.12.2017

    e     Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    e     Electronics: DAE4 Sn601; Calibrated: 26.10.2017

    *     Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    e     DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 119.6 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 29.4 W/kg
SAR(1 g) = 14.4 W/kg; SAR(10 g) = 6.39 W/kg
Maximum value of SAR (measured) = 24.1 W/kg


          dB
        9
        — 5.20
          —10.40
          —15.60
          —20.80

          —26.00
                   0 dB = 24.1 W/kg = 13.82 dBW/kg




Certificate No: D26OO0V2—1032_Sep18              Page 5 of 8


Impedance Measurement Plot for Head TSL




File     YView   Channel       Sweep   Calibration   Trace   Scale   Marker       System   Window   Help

                                                                                                    2 600000 GHz       49.949 0)
                                                                                                         12.269 pF    —4.9893 0
                                                                                                    2600000 GHz      49.859 mU
                                                                                                                       —87. 723 ®




                 Ch1Avg= 20                                          *        |
        Ch1: Start 2.40000 GHza      ——                                                                              Stop 2.90000 GHz


   0.00
                                                                                                    2 §00000 GHz     —2§ 045 dB



   10.00

   15.00
   20.00


   5.00

   30.00

   35.00

       0.00      Ch 1 .    =   |20
        Ch1: Start 2.40000 GHza      ——                                                                              Stop 2.80000 GHz


  Status         __CH 1:                             C" 1—Port                    Avg=20 Delay                            ECL




Certificate No: D2600V2—1032_Sep18                                       Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                   Date: 17.09.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2600 MHz; Type: D2600V2; Serial: D2600V2 — SN:1032

Communication System: UID 0 — CW; Frequency: 2600 MHz
Medium parameters used: f = 2600 MHz; 0 = 2.18 S/m; & =51.2; p = 1000 kg/m>
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

       Probe: EX3DV4 — SN7349; ConvF(7.81, 7.81, 7.81) @ 2600 MHz; Calibrated: 30.12.2017

       Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 $n601; Calibrated: 26.10.2017

       Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

       DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 108.8 V/m; Power Drift = —0.08 dB
Peak SAR (extrapolated) = 29.0 W/kg
SAR(1 g) = 13.9 W/kg; SAR(10 g) = 6.17 W/kg
Maximum value of SAR (measured) = 23.1 W/kg


       dB


       —5.20
       —10.40

       —15.60

       —20.80

       —26.00
                  0 dB = 23.1 W/kg = 13.64 dBW/kg




Certificate No: D2600V2—1032_Sep18            Page 7 of 8


Impedance Measurement Plot for Body TSL




File     Yiew   Channel     Sweep     Calibration   Trace   Scale   Marker   System   Window   Help
                                                                                                    e           mm
                                                                                               2.600000 GHz       45.979 C
                                                                                                   17.570 pF     —3.4839 C
                                                                                               2.600000 GHz     55.394 mU
                                                                                                                  —197.91 *




                Ch1Avg= 20                                          _           seA
        Ch1: Start 2.40000 GHa      ———                                                                         Stop 2.80000 GHz



                                                                                                .§00000   GHz   —2$. 131 dB
  0.00

   5.00

   10.00

   15.00
   20.00

   25.00

   30.00

   35.00

       0.00     Ch 1 &vg=     120
        Ch1: Start 2.40000 GHza     ——                                                                          Stop 220000 GHz


  Status        CH 1:                               C 1.Por                  Avg=20 Delay




Certificate No: D2600V2—1032_Sep18                                      Page 8 of 8


Calibration
      .     Laboratory of                                           \\\\'////,
                                                                  *\\\/                                       S    Schweizerischer Kalibrierdienst
Schmid & Partner                                                 il_:\\/E!                                    c Service suisse d‘étalonnage
  Englneering AG                                                 2.z                                               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      27 ///\\\\\                                 S    swiss Calibration Service
                                                                    Trlilibe®®

Accredited by the Swiss Accreditation Service (SAS)                                                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       Huawei—SZ (Auden)                                                                      Certificate No: D5GHzV2—1155_Jun18

CALIBRATION CERTIFICATE
 Object                                D5GHzV2 — SN:1155


 Calibration procedure(s)              QA CAL—22.v3
                                       Calibration procedure for dipole validation kits between 3—6 GHz



 Calibration date:                     June 08, 2018



 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical forcalibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN:    104778             04—Apr—18   (No.   217—02672/02673)                   Apr—19
 Power sensor NRP—Z91                   SN:    103244             04—Apr—18   (No.   217—02672)                         Apr—19
 Power sensor NRP—Z91                   SN:    103245             04—Apr—18   (No.   217—026783)                        Apr—19
 Reference 20 dB Attenuator             SN:    5058 (20k)         04—Apr—18   (No.   217—02682)                         Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                             Apr—19
 Reference Probe EX3DV4                 SN: 3503                  30—Dec—17 (No. EX3—3503_Dec17)                        Dec—18
 DAE4                                   SN: 601                   26—0ct—17 (No. DAE4—601_Oct17)                        Oct—18


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN:    GB37480704         07—Oct—15   (in   house   check   Oct—16)             In   house   check:   Oct—18
 Power sensor HP 8481A                  SN:    US37202783         07—Oct—15   (in   house   check   Oct—16)             In   house   check:   Oct—18
 Power sensor HP 8481A                  SN:    MY41092317         07—Oct—15   (in   house   check   Oct—16)             In   house   check:   Oct—18
 RF generator R&S SMT—06                SN:    100972             15—Jun—15   (in   house   check   Oct—16)             In   house   check:   Oct—18
 Network Analyzer HP 8753E              SN:    US37390585         18—Oct—01   (in   house   check   Oct—17)             In   house   check:   Oct—18


                                        Name                                   Function                                 Signature
 Calibrated by:                         Manu Seitz                             Laboratory Technician                           32.



 Approved by:                           Katia Pokovic                          Technical Manager



                                                                                                                       Issued: June 15, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D5GHzV2—1155_Jun18                                    Page 1 of 16


    s      :                                                   yo
Callbratlon Laboratory of                                    s\‘@/”@           S   Schweizerischer Kalibrierdienst
Schmid & Partner                                            iIB\EE/ME@EE       e   Service suisse d‘étalonnage
   Engineering AG                                           se                     Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                {'/,,/fi\‘§§        S   swiss Calibration Service
                                                                 Whilabs

Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                         tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x,y,z
N/A                         not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
         Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
         MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    e    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         paralle! to the body axis.
    e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
         SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D5GHzV2—1155_Jun18                              Page 2 of 16


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY5                              V52.10.1
     Extrapolation                                   Advanced Extrapolation
     Phantom                                       Modular Flat Phantom V5.0

     Distance Dipole Center— TSL                               10 mm                             with Spacer
     Zoom Scan Resolution                        dx, dy = 4.0 mm, dz =1.4 mm         Graded Ratio = 1.4 (Z direction)
                                                       5250 MHz + 1 MHz
                                                       5400 MHz + 1 MHz
     Frequency                                         5600 MHz + 1 MHz
                                                       5750 MHz + 1 MHz
                                                       5850 MHz + 1 MHz


Head TSL parameters at 5250 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity

     Nominal Head TSL parameters                                  22.0 °C           35.9               4.71 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   36.2 26 %          4.57 mho/m + 6 %

     Head TSL temperature change during test                      x6.5°C


SAR result with Head TSL at 5250 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

     SAR measured                                        100 mW input power                     8.14 W/kg

     SAR for nominal Head TSL parameters                     normalized to 1W        81.4 Wikg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        100 mW input power                 2.35 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        23.5 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1155_Jun18                    Page 3 of 16


Head TSL parameters at 5400 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.8               4.86 mho/m

     Measured Head TSL parameters                              (22.0 £0.2) °C    35.9 +6 %          4.73 mho/m + 6 %
     Head TSL temperature change during test                      x45"°C


SAR result with Head TSL at 5400 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     8.33 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        83.3 W / kg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        100 mW input power                  2.39 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         23.9 W/kg + 19.5 % (k=2)


Head TSL parameters at 5600 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.5               5.07 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   35.7 x6 %          4.94 mho/m + 6 %

     Head TSL temperature change during test                      «05 °C


SAR result with Head TSL at 5600 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     8.52 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         85.2 Wikg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        100 mW input power                     2.43 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.3 Wikg x 19.5 % (k=2)




Certificate No: D5GHzV2—1155_Jun18                    Page 4 of 16


Head TSL parameters at 5750 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.4               5.22 mho/m
     Measured Head TSL parameters                              (22.0 £ 0.2) °C   35.4 +6 %          5.09 mho/m + 6 %

     Head TSL temperature change during test                      <0.5 °C


SAR result with Head TSL at 5750 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     7.85 W/kg

     SAR for nominal Head TSL parameters                     normalized to 1W        78.4 Wikg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        100 mW input power                  2.23 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         22.3 Wikg * 19.5 % (k=2)


Head TSL parameters at 5850 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Head TSL parameters                                  22.0 °C           35.2               5.32 mho/m
     Measured Head TSL parameters                              (22.0 £0.2) °C    35.3 +6 %          5.20 mho/m + 6 %
     Head TSL temperature change during test                      <60.5 °C           ~———                   we««


SAR result with Head TSL at 5850 MHz

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        100 mW input power                     8.50 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W         85.4 Wikg + 19.9 % (k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                 condition

     SAR measured                                        100 mW input power                  2.41 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         24.3 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1155_Jun18                    Page 5 of 16


Body TSL parameters at 5250 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           48.9               5.36 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   46.9 + 6 %         5.51 mho/m + 6 %

     Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL at 5250 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
     SAR measured                                        100 mW input power                     7.53 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         74.7 Wikg # 19.9 % (k=2)

     SAR averaged over 10 cm‘ (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                  211 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         20.9 W/kg + 19.5 % (k=2)


Body TSL parameters at 5400 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity

     Nominal Body TSL parameters                                  22.0 °C           48.7               5.53 mho/m
     Measured Body TSL parameters                              (22.0 £0.2) °C    46.7 x 6 %         5.71 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL at 5400 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        100 mW input power                  7.93 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         78.7 Wikg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                  2.22 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         22.0 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1155_Jun18                    Page 6 of 16


Body TSL parameters at 5600 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           48.5               5.77 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   46.3 + 6 %         5.99 mho/m + 6 %

     Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL at 5600 MHz

     SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
     SAR measured                                        100 mW input power                     8.02 W/kg

     SAR for nominal Body TSL parameters                     normalized to 1W        79.6 Wikg + 19.9 %(k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                  2.23 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         22.1 Wikg * 19.5 % (k=2)

Body TSL parameters at 5750 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity

     Nominal Body TSL parameters                                  22.0 °C           48.3               5.94 mho/m
     Measured Body TSL parameters                              (22.0 + 0.2) °C   46.0 x 6 %         6.19 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C            hn                     makh


SAR result with Body TSL at 5750 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
     SAR measured                                        100 mW input power                     7.39 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         73.3 Wikg + 19.9 %(k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

     SAR measured                                        100 mW input power                  2.06 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         20.4 Wikg + 19.5 % (k=2)




Certificate No: D5GHzV2—1155_Jun18                    Page 7 of 16


Body TSL parameters at 5850 MHz
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity       Conductivity
     Nominal Body TSL parameters                                  22.0 °C           48.1             6.06 mho/m
     Measured Body TSL parameters                              (22.0 £ 0.2) °C   45.9 26 %        6.33 mho/m + 6 %

     Body TSL temperature change during test                      x0.5 °C


SAR result with Body TSL at 5850 MHz

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition

     SAR measured                                        100 mW input power                  7.85 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W         78.1 Wikg + 19.9 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition
     SAR measured                                        100 mW input power                  2.18 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         21.7 Wikg £ 19.5 % (k=2)




Certificate No: D5GHzV2—1155_Jun18                    Page 8 of 16


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL at 5250 MHz

     Impedance, transformed to feed point                  55.3 Q — 8.1 jQ

     Return Loss                                             — 20.8 dB


Antenna Parameters with Head TSL at 5400 MHz

     Impedance, transformed to feed point                  49.5 Q — 5.4 jQ
     Return Loss                                             —25.3 dB


Antenna Parameters with Head TSL at 5600 MHz

     Impedance, transformed to feed point                  59.2 Q — 6.3 jQ
     Return Loss                                             —19.8 dB


Antenna Parameters with Head TSL at 5750 MHz

     Impedance, transformed to feed point                  57.3 Q — 2.8 jQ
     Return Loss                                             —22.8 dB


Antenna Parameters with Head TSL at 5850 MHz

     Impedance, transformed to feed point                  55.4 Q — 6.7 jQ
     Return Loss                                              —21.8 dB


Antenna Parameters with Body TSL at 5250 MHz

     Impedance, transformed to feed point                  53.4 Q — 5.5 jQ

     Return Loss                                              —24.1 dB


Antenna Parameters with Body TSL at 5400 MHz

     Impedance, transformed to feed point                  49.9 Q — 3.3 jQ

     Return Loss                                             — 29.6 dB


Antenna Parameters with Body TSL at 5600 MHz

     Impedance, transformed to feed point                  60.6 Q — 5.0 jQ
     Return Loss                                             —19.5 dB




Certificate No: D5GHzV2—1155_Jun18          Page 9 of 16



Document Created: 2019-03-11 15:38:11
Document Modified: 2019-03-11 15:38:11

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC