Appendix C. Calibration Certificate_Part 15

FCC ID: QISELE-LX9M

RF Exposure Info

Download: PDF

DASY5 Validation Report for Head TSL

                                                                                  Date: 06.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole D5GHzV2; Type: D5GHzV2; Serial: D5GHzV2 — SN:1155
Communication System: UID 0 — CW; Frequency: 5250 MHz, Frequency: 5400 MHz, Frequency: 5600
MHz, Frequency: 5750 MHz, Frequency: 5850 MHz
Medium parameters used: f = 5250 MHz; 0 = 4.57 S/m; & = 36.2; p = 1000 kg/m*
Medium parameters used: f = 5400 MHz; 0 = 4.73 S/m; & = 35.9; p = 1000 kg/m
Medium parameters used: f = 5600 MHz; 0 = 4.94 S/m; & = 35.7; p = 1000 kg/m*
Medium parameters used: f = 5750 MHz; 0 = 5.09 S/m; & = 35.4; p = 1000 kg/m*
Medium parameters used: f = 5850 MHz; 0 = 5.2 S/m; & = 35.3; p = 1000 kg/m"
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2011)
DASY52 Configuration:

   e   Probe: EX3DV4 — SN3503; ConvF(5.51, 5.51, 5.51) @ 5250 MHz, ConvF(5.5, 5.5, 5.5) @ 5400
       MHz, ConvF(5.05, 5.05, 5.05) @ 5600 MHz, ConvF(4.98, 4.98, 4.98) @ 5750 MHz, ConvF(4.94,
       4.94, 4.94) @ 5850 MHz; Calibrated: 30.12.2017

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   &   Electronics: DAE4 Sn601 (5GHz); Calibrated: 26.10.2017

   e   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   e   DASY32 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5250 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 76.63 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) = 28.2 W/kg
SAR(1 g) = 8.14 W/kg; SAR(10 g) = 2.35 W/kg
Maximum value of SAR (measured) = 18.4 W/kg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5400 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 75.16 V/m; Power Drift = —0.02 dB
Peak SAR (extrapolated) = 31.4 W/kg
SAR(1 g) = 8.33 W/kg; SAR(10 g) = 2.39 W/kg
Maximum value of SAR (measured) = 19.4 W/kg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 75.94 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 32.5 W/kg
SAR(1 g) = 8.52 W/kg; SAR(10 g) = 2.43 W/kg
Maximum value of SAR (measured) = 20.1 W/kg


Certificate No: D5GHzV2—1155_Jun18            Page 11 of 16


Dipole Calibration for Head Tissue/Pin=100mwW, dist=10mm, £=5750 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 72.32 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 30.9 W/kg
SAR(I g) = 7.85 W/kg; SAR(10 g) = 2.23 W/kg
Maximum value of SAR (measured) = 18.7 W/kg

Dipole Calibration for Head Tissue/Pin=100mW, dist=10mm, £=5850 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 73.54 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 36.0 W/kg
SAR(1 g) = 8.5 W/kg; SAR(10 g) = 2.41 W/kg
Maximum value of SAR (measured) = 20.7 W/kg




                  0 dB = 18.4 W/kg = 12.65 dBW/kg




Certificate No: D5GHzV2—1155_Jun18           Page 12 of 16


Impedance Measurement Plot for Head TSL




                                                                   6 Jun 2018     14:02:18
         CHI s11     1 U FS          1:59.273 0   —8.0996 0   $.7428 pF          5 250.000 000 MHz
                                                  enfi
         3                                   <         ig um       P
                                                                       x                             CH1 Markers
         Del
                                                                                                      2 49.549 a
                                                                                                       —5.4258 a
         ter                                                                                         5.40008 GHz
                                                                                                      3: 59.207 a
                                                                                                        —6,.2039 a
                                                                                                     5.60000 GHz
                                                                                                      4: 97,252 a
                                                                                                        ~2.8379 a
         fivg                                                                                        5.75000 GHz
         16
                                                                                                      5: 55.438 a
                                                                                                        —6.0934 n
                                                                                                     5.85000 GHz
         H1d

          CH2




                                                                                                     CH2 Markers

               m                                                                                     2~—25,254 dB
         Cor                                                                                         5.40000 GHz
                                                                                                     3:—19.950 dB
                                                                                                      5.60000 GHz

                                                                                                     4:—22.785 dB
         fivg                                                                                         5.79000 GHz
         16
                                                                                                     St—21.762 dB
                                                                                                      5.85000 GHz
         H1d

                   START 5 0a0.000 00o MHz                                 STOP 6 000.000 aBD MHz




Certificate No: D5GHzV2—1155_Jun18                 Page 13 of 16


DASY5 Validation Report for Body TSL
                                                                                   Date: 08.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole D5GHzV2; Type: D5SGHzV2; Serial: D5GHzV2 — SN:1155
Communication System: UID 0 — CW; Frequency: 5250 MHz, Frequency: 5400 MHz, Frequency: 5600
MHz, Frequency: 5750 MHz, Frequency: 5850 MHz
Medium parameters used: £ = 5250 MHz; 0 = 5.51 S/m; & = 46.9; p = 1000 kg/m‘
Medium parameters used: f = 5400 MHz; 0 = 5.71 S/m; s = 46.7; p = 1000 kg/m‘
Medium parameters used: f = 5600 MHz; 0 = 5.99 S/m; &; = 46.3; p = 1000 kg/m3
Medium parameters used: f = 5750 MHz; 0 = 6.19 S/m; &= 46; p = 1000 kg/m?
Medium parameters used: f = 5850 MHz; 0 = 6.33 S/m; & = 45.9; p = 1000 kg/m?*
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)
DASY52 Configuration:

   e    Probe: EX3DV4 — SN3503; ConvF(5.26, 5.26, 5.26) @ 5250 MHz, ConvF(4.7, 4.7, 4.7) @ 5400
        MHz, ConvF(4.65, 4.65, 4.65) @ 5600 MHz, ConvF(4.57, 4.57, 4.57) @ 5750 MHz, ConvF(4.47,
        4.47, 4.47) @ 5850 MHz; Calibrated: 30.12.2017

   e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e    Electronics: DAE4 Sn601 (5GHz); Calibrated: 26.10.2017

   e    Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=100mwW, dist=10mm, £=5250 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.19 V/m; Power Drift = —0.06dB
Peak SAR (extrapolated) = 28.5 W/kg
SAR(I g) =7.53 W/kg; SAR(10 g) = 2.11 W/kg
Maximum value of SAR (measured) = 17.2 W/kg

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5400 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 68.41 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 31.3 W/kg
SAR(1 g) =7.93 W/kg; SAR(10 g) = 2.22 W/kg
Maximum value of SAR (measured) = 19.5 W/kg

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5600 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.66 V/m; Power Drift = —0.08 dB
Peak SAR (extrapolated) = 33.5 W/kg
SAR(I g) = 8.02 W/kg; SAR(10 g) = 2.23 W/kg
Maximum value of SAR (measured) = 19.2 W/kg


Certificate No: D5GHzV2—1155_Jun18            Page 14 of 16


Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5750 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 64.78 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) = 31.5 W/kg
SAR(I g) =7.39 W/kg; SAR(10 g) = 2.06 W/kg
Maximum value of SAR (measured) = 17.8 W/kg

Dipole Calibration for Body Tissue/Pin=100mW, dist=10mm, £=5850 MHz/Zoom Scan,
dist=1.4mm (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 65.90 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 34.8 W/kg
SAR(I g) = 7.85 W/kg; SAR(10 g) = 2.18 W/kg
Maximum value of SAR (measured) = 19.1 W/kg




       —12.00


       —18.00


       —24.00


       —30.00
                  0 dB = 17.2 W/kg = 12.36 dBW/kg




Certificate No: D5GHzV2—1155_Jun18           Page 15 of 16


Impedance Measurement Plot for Body TSL




                                                                         8 Jun 2018         17123101
         CHT sai     4 uors          5: 56.309 a     539 2 4.3502 pF                       5 850.000 gng MHz
                                                     *"~.
                                                            #   °+       .

                                                                                                               CH1 Markers
         *
                                                                     ‘           \

         ts"                                                                     *                              4t 52.424 a
                                                                                                                 ~5.9078 a
         tor                                                                                                   5.25000 6z
                                                                                                                2 49.993 a
                                                                                                                 ~5.9242 a
                                                                                                               5.40000 GHz
                                                                                                                3:60.596 a
                                                                  M     f                                        ~4.9625 a
         es                                                      —L + J                                        5.G0000 6Hz
                                                        f            §       *                                  41 5&.150 a
                                             —        ril e                                                       ~1.9899 A
                                                      4
                                                 . ueongult w2                                                 5.75000 GHz
         H1d
          cH2


                                                                                                               CH2 Markers

                                                                                                               1:—24.061 4B
         ter                                                                                                    5.25000 oh=
                                                                                                               2:—29.552 dB
                                                                                                                5.40000 GHz
                                                                                                                 —19.521 aB
         fvg                                                                                                     .€0000 GHz
         16
                                                                                                                 —22.289 aB
                                                                                                                5.75000 GHz
         H1d
                   START 5 000.000 ano MHz                                           STOP 6 000.000 000 MHz




Certificate No: D5GHzV2—1155_Jun18                  Page 16 of 16


Impedance Measurement Plot for Body TSL




File     Yiew   Channel     Sweep     Calibration   Trace   Scale   Marker   System   Window   Help
                                                                                                    e           mm
                                                                                               2.600000 GHz       45.979 C
                                                                                                   17.570 pF     —3.4839 C
                                                                                               2.600000 GHz     55.394 mU
                                                                                                                  —197.91 *




                Ch1Avg= 20                                          _           seA
        Ch1: Start 2.40000 GHa      ———                                                                         Stop 2.80000 GHz



                                                                                                .§00000   GHz   —2$. 131 dB
  0.00

   5.00

   10.00

   15.00
   20.00

   25.00

   30.00

   35.00

       0.00     Ch 1 &vg=     120
        Ch1: Start 2.40000 GHza     ——                                                                          Stop 220000 GHz


  Status        CH 1:                               C 1.Por                  Avg=20 Delay




Certificate No: D2600V2—1032_Sep18                                      Page 8 of 8


                Justification of the extended calibration of Dipole D835V2 SN:4d059

    Per KDB 865664, we have measured the Impedance and Return Loss as below.
    1) The most recent return-loss result, measured at least annually, deviates by no more than 20%
       from the previous measurement.
    2) The most recent measurement of the real or imaginary parts of the impedance, measured at least
       annually is within 5 ohm from the previous measurement.


  Dipole 835 Head TST          Target Value          Measured Value                    Difference

Impedance transformed to
                               52.0Ω-1.8j Ω           52.2Ω-2.83j Ω                R=0.2Ω,X=-1.03Ω
       feed point

      Return Loss                -31.4dB                -32.32 dB                       -2.92%

  Dipole 835 Body TST          Target Value          Measured Value                    Difference

Impedance transformed to
                               48.7Ω-4.2j Ω           48.07Ω-4.87j Ω               R=-0.63Ω,X=-0.67Ω
       feed point

      Return Loss                -27.1 dB               -29.58 dB                       -9.15%

    Measured Date               2016-04-20             2018-04-15                        -------


                    Impedance Test-Head                             Return Loss Test-Head




                    Impedance Test-Body                             Return Loss Test-Body


                Justification of the extended calibration of Dipole D1900V2 SN:5d143

    Per KDB 865664, we have measured the Impedance and Return Loss as below.
    1) The most recent return-loss result, measured at least annually, deviates by no more than 20%
       from the previous measurement.
    2) The most recent measurement of the real or imaginary parts of the impedance, measured at least
       annually is within 5 ohm from the previous measurement.


 Dipole 1900 Head TST          Target Value          Measured Value                    Difference

Impedance transformed to
                               53.4Ω+4.3j Ω           49.75Ω+7.52j Ω               R=-3.65Ω,X=3.22Ω
       feed point

      Return Loss                -25.6dB                -25.32 dB                       -1.09%

 Dipole 1900 Body TST          Target Value          Measured Value                    Difference

Impedance transformed to
                               49.1Ω+5.9j Ω           46.18Ω+2.21j Ω               R=-2.92Ω,X=-3.69Ω
       feed point

      Return Loss                -24.4 dB                -24.5 dB                       0.41%

    Measured Date               2017-09-20             2018-09-20                        -------


                    Impedance Test-Head                             Return Loss Test-Head




                    Impedance Test-Body                             Return Loss Test-Body