Appendix C Calibration certificate_part 1 of 6

FCC ID: QISCLT-L29

RF Exposure Info

Download: PDF
FCCID_4139959

                                       Report No: E5/2018/C0051




               Appendix C
             Calibration certificate
1. Dipole
D750V3-SN 1160(2016-6-22)
D835V2-SN 4d105(2016-12-08)
D1750V2-SN 1149(2016-6-23)
D1900V2-SN 5d028(2016-12-07)
D2450V2-SN 733(2016-12-07)
D2600V2-SN 1125(2016-6-22)
D5GHzV2-SN 1165(2016-12-13)
2. DAE
DAE4-SN 1428(2018-01-17)
DAE4-SN 896(2018-11-08)
2. Probe
EX3DV4-SN 3962(2018-01-11)
EX3DV4-SN 3789(2018-02-08)


Calibration Laboratory of                                         «7                                       g Schweizerischer Kalibrierdienst
Schmid & Partner                                                 iia\éfi{/fli%i                                Service suissed‘étaionnage
  Engineering AG                                                 *s                                        C servicio sviezero d terstura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      5ns                                      S       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       SGS—SZ (Auden)                                                                    Certificate No: D750V3—1160_Jun16

ICALIBRATION CERTIFICATE                                                                                                                            |
 Object                                D750V3 — SN:1160


 Calibration procedure(s)             QA CAL—05.v9
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    June 22, 2016


 This calibration certificate documents the traceabilty to national standards, which realizethe physical units of measurements (S1).
 The measurements and the uncertainties with confidence probabilty are given on the following pages and are part of the certificate.

 All calibrations have been conducted in the closed laboratory facifty: environment temperature (22 3)°C and humidity < 70%


 Callbration Equipment used (M&TE critical for callbration)


 Primary Standards                      1#                        Cal Date (Gertificate No.)                           Scheduled Calibration
 Power meter NRP                        SN: 104778                06—Aor—16 (No. 217—02288/02209)                      Aprt7
 Power sensor NRP—291                   SN: 108244                06—Aor—16 (No. 217—02288)                            Apr—t7
 Power sensor NRP—291                   SN: 103245                06—Aor—16 (No. 217—02289)                            Apr—t7
 Reference 20 dB Attenuator             SN:    5058 (20k)         05—Ror—16 (No. 217—02202)                            Apr—t7
 Type—N mismatch combination            SN:    5047.2 / 06327     05—Aor—16 (No. 217—02205)                            Apr—t7
 Reference Probe EX3DV4                 SN:    7349               15—Jun—16 (No. EX3—7349_Jun16)                       Jun—17
 DAE4                                   SN:    601                30—Dec—15 (No. DAE4—601_Dect5)                       Dec—16

 Secondary Standards                    iD #                      Check Date (in house)                                Scheduled Check
 Power meter EPM—442A                   SN:    GB37480704         07—Oct—15   (No. 217—02222)                          in   house   check Oct—16
 Power sensor HP 8481A                  SN:    US37202783         07—Oct—15   (No. 217—02222)                          in   house   check: Oct—16
 Power sensor HP 8481A                  SN:    MY41092317         07—Oct—15   (No. 217—02223)                          in   house   check Oct—16
 RF generator R&S SMT—06                SN:    100972             15—Jun—15   (in house check Jun—15)                  in   house   check: Oct—16
 Network Analyzer HP 8753E              SN:    US37390585         18—0ct—01   (in house check Oct—15)                  in   house   check Oct—16


                                        Name                                   Function                                 Signature
 Calibrated by:                        Jeton Kastrati                          Laboratory Technician           \            )
                                                                                                         csll w«

 Approved by:                           Katia Pokovie                          Technical Manager                   %f/f’


                                                                                                                       Issued: June27, 2016
 This calibration certificate shall not be reproduced except in full without writen approval of the laboratory.

Certificate No: D750V3—1160_Jun16                                     Page 1 of 8


Calibration
      4     Laboratory of                                                     «g   Schweizerischer Kalibrierdienst
Schmid & Partner                                                                   Service suissed‘étalonnage
  Engineering AG                                                              C    sevigesmmerssitersturs
Zeughausstrasse 43, 8004 Zurich, Switzerland                                  S    swiss Calibration Service

Aceredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)®,
       February 2005
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
         MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
   * Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
   * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
    *    SAR measured: SAR measured at the stated antenna input power.
    &    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
  The reported uncertainty of measurementis stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D750V3—1160_Jun16                               Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                              DASY3                              V52.8.8
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              15 mm                            with Spacer
     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                          750 MHz + 1 MHz



Head TSL parameters
   Thefollowing parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C          41.9               0.89 mho/m
     Measured Head TSL parameters                              (22.0 £0.2) °C   41.6 £ 6 %         0.91 mho/m   6%
     Head TSL temperature change during test                      <0.5 °C

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                    2.08 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W       8.17 Wikg 17.0 %(k=2)


     SAR averaged over 10 cm* (10 g) of Head TSL                condition
     SAR measured                                        250 mW input power                    1.36 Wkg
     SAR for nominal Head TSL parameters                     normalized to 1W       5.36 Wikg * 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C          55.5               0.96 mho/m
     Measured Body TSL parameters                              (22.0 £0.2) °C   55.4 26 %          1.00 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C           «—

SAR result with Body TSL

     SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
     SAR measured                                        250 mW input power                    2.21 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W       8.57 Wikg » 17.0 % (k=2)


     SAR averaged over 10 cm* (10 g) of Body TSL                condition
     SAR measured                                        250 mW input power                    145 Wikg
     SAR for nominal Body TSL parameters                     normalized to 1W        5.66 Wikg 2 16.5 % (k=2)




Certificate No: D750V3—1160_Jun16                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           54.8 0 — 1.6 io

      Return Loss                                                                       —26.3 dB


Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           50.2 Q — 3.8 i2
      Return Loss                                                                       —28.4 dB




General Antenna Parameters and Design

      Electrical Delay(one direction)                                                    1.040 ns

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of thedipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions* paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                      *       November 19, 2015




Certificate No: D750V3—1160_Jun16                         Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                      Date: 22.06.2016

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1160

Communication System: UID 0 — CW; Frequency: 750 MHz
Medium parameters used: f = 750 MHz; 0 = 0.91 S/m; & =41.6; p = 1000 kym"
Phantomsection: Flat Section
Measurement Standard: DASY5 (IEEEAEC/ANSI C63.19—2011)

DASY52 Configuration:

        Probe: EX3DV4 — SN7349; ConvF(10.17, 10.17, 10.17); Calibrated: 15.06.2016;

        Sensor—Surface: 1.4mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 30.12.2015

        Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

        DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 57.89 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 3.12 W/kg
SAR(I g) = 2.08 W/kg; S     AR(10 g) = 1.36 W/kg
Maximum value of SAR (measured) = 2.76 W/kg




        —2.00

        —4.00

        —6.00




        —10.00
                   0 dB = 2.76 W/kg =4.41 dBW/kg




Certificate No: D750V3—1160_Jun16                  Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                   22 Jun 2016   13:36:11
          CH® san     a uors          m 54.022 0   ~1,5723 0 134.97 pF           75e.000 000 MXz




          fvg
          16



          Hid


          cn2 si1     Los       _5 dB/REF —20_a8       f       m~26.300 aB       750,000 000 IMHz




          Ca




          igs           .      1       +       :       +       +         +       +        +
                        }      +       }       |       }       +         1       +        }
          Hie           }      }       +       +       +       L        —|       }        }

                    start 550.000 000 hz               —       —         StoP 950.000 000 mHz




Certificate No: D750V3—1160_Jun16                    Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                   Date: 22.06.2016
Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 — SN:1160

Communication System: UID 0 — CW; Frequency: 750 MHz                  .
Medium parameters used: = 750 MHz; 0 = 1 S/m; s = 55.4; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY3 (IEEEAEC/ANSI C63.19—2011)

DASY52 Configuration:

   «+   Probe: EX3DV4 — SN7349; ConyF(9.99, 9.99, 9.99); Calibrated: 15.06.2016;

   *    Sensor—Surface; 1.4mm (Mechanical Surface Detection)

   *    Electronics: DAE4 Sn601; Calibrated: 30.12.2015

   *«   Phantom: Flat Phantom 4.9L; Type: QDOOOPA49AA; Serial: 1001

   +    DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 56.66 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 3.29 W/kg
SAR(I g) =2.21 W/kg; SAR(10 g) = 1.45 W/kg
Maximum value of SAR (measured) =2.93 W/kg




        —2.00

        ~4.00

        —6.00

        —8.00

        —10.00
                   0 dB = 2.93 W/kg = 4.67 dBW/kg




Certificate No: D750V3—1160_Jun16              Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                22 Jun zeie 10:5e:20
          CH® sn       i uors          m 50164 a   —3.7969 n 55.090 pF      750.000 0g0 MHz




          Ay
          lEs



          Hid


          cn2 sii_     Los      .5 dB/REF —20_d8      .                     750.000 000 nHz




          Ca




                        L       |       1      1      2       L      1       .      }
          By
          165           +       /       +      —      I       +      4       +      +



          Hid           +       |       }      4      £F      L      4       ;      }
                     SthRT s50.000 000 MHz            —       —      stoP 950.000 con mz




Certificate No: D750V3—1160_Jun16                   Page 8 of 8


                                         In Collaboration with                        \1p1
                                                                                     wl l/1‘4,2                        HEBVLH
                                                                                                                          *



                 w ‘/"/"J, p__C_ §\\\/’3 A EiA
                 —V                      CALIBRATION LABORATORY                  MCNAS
                                                                                 2                *»

             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China   %,/@\§ v CAUBRAHON
             Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504            AululabsS® CNAS LO570
             E—mail: cttl@chinattl.com       Http://www.chinattl.cn

      Client               SGS(Boce)                                        Certificate No:            Z16—97239

CALIBRATION CERTIFICATE

Object                                    D835V2 — SN: 40105

Calibration Procedure(s)
                                          FD—Z11—003—01
                                          Calibration Procedures for dipole validation kits

Calibration date:                         December 8, 2016


This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘C and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #                Cal Date(Calibrated by, Certificate No.)            Scheduled Calibration
 Power Meter        NRP2           101919             27—Jun—16 (CTTL, No.J16X04777)                               Jun—17
 Power sensor NRP—Z91              101547             27—Jun—16 (CTTL, No.J16X04777)                               Jun—17
 Reference Probe EX3DV4           SN 7433             26—Sep—16(SPEAG,No.EX3—7433_Sep16)                           Sep—17
 DAE4                             SN 771              02—Feb—16(CTTL—SPEAG,No.Z16—97011)                           Feb—17


 Secondary Standards              ID #                Cal Date(Calibrated by, Certificate No.)            Scheduled Calibration
 Signal Generator E4438C          MY49071430          01—Feb—16 (CTTL, No.J16X00893)                               Jan—17
 Network Analyzer E5071C          MY46110673          26—Jan—16 (CTTL, No.J16X00894)                               Jan—17



                                   Name                          Function                                     Signature
Calibrated by:                   Zhao Jing                  SAR Test Engineer                              & é i/ |

Reviewed by:                     Qi Dianyuan                SAR Project Leader                             §2 g__ gor

Approved by:                     Lu Bingsong                Deputy Director of the laboratory

                                                                                         Issued: Decemb J
                                                                                                        11, 2016
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z16—97239                            Page 1 of 8


                                      In Collaboration with

           §7T7Jj a                  CALIBRATION LABORATORY
           e
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Glossary:
TSL                        tissue simulating liquid
Convr                      sensitivity in TSL / NORMx,y,z
N/A                        not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
  Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
   devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
    2005
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
    30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
e  Measurement Conditions: Further details are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
e     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
e     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
      SAR measured: SAR measured at the stated antenna input power.
      SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
e     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.



Certificate No: Z16—97239                               Page 2 of 8


                                     In Collaboration with

          @TTL a                     CALIBRATION LABORATORY
          y
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Measurement Conditions
   DASY system configuration, as far as notgiven on page 1.
     DASY Version                                         DASY52                                     52.8.8.1258

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                         15 mm                                    with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                       835 MHz + 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature            Permittivity           Conductivity

     Nominal Head TSL parameters                              22.0 °C                  41.5                0.90 mho/m

     Measured Head TSL parameters                        (22.0 + 0.2) °C             40.8 +6 %           0.91 mho/m +6 %

     Head TSL temperature change during test                  <1.0 °C                   ————                     _———
SAR result with Head TSL
     SAR averaged over 1 Cmm‘ (1 g) of Head TSL                         Condition
     SAR measured                                                 250 mW input power                   2.43 mW / g
     SAR for nominal Head TSL parameters                             normalized to 1W          9.59 mW /g + 20.8 % (k=2)
     SAR averaged over 10 cm (10 g) of Head TSL                         Condition

     SAR measured                                                 250 mW input power                   1.59 mW / g

     SAR for nominal Head TSL parameters                             normalized to 1W          6.29 mW /g + 20.4 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                             Temperature            Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  55.2                0.97 mho/m

     Measured Body TSL parameters                        (22.0 + 0.2) °C             54.7 £6 %           0.98 mho/m +6 %

     Body TSL temperature change during test                  <1.0 °C                                            ————
SAR result with Body TSL
     SAR averaged over 1 Cm" (1 g) of Body TSL                          Condition
     SAR measured                                                 250 mW input power                   2.44 mW / g

     SAR for nominal Body TSL parameters                             normalized to 1W          9.65 mW /g + 20.8 % (k=2)
     SAR averaged over 10 cm> (10 g) of Body TSL                        Condition

     SAR measured                                                 250 mW input power                   1.63 mW / g

     SAR for nominal Body TSL parameters                             normalized to 1W          6.46 mW /g + 20.4 % (k=2)
Certificate No: Z16—97239                              Page 3 of 8


          E7TTL $.P.g—2
          V
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


 Appendix

 Antenna Parameters with Head TSL


     Impedance, transformed to feed point                                   49.20— 3.41j0Q

     Return Loss                                                              — 29.10B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   45.80— 3.25j0

     Return Loss                                                              — 25.10B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.500 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                             SPEAG




Certificate No: Z16—97239                             Page 4 of 8


         ETTL $.2.g.2
          ‘/
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


DASYS5 Validation Report for Head TSL                                   Date: 12.08.2016
Test Laboratory: CTTL, Beijing, China
DUT : Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40105
   Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1: 1
   Medium parameters used: f= 835 MHz; 0 = 0.912 S/m; s, = 40.78; p = 1000 kg/m*
   Phantom section: Center Section
   Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(9.82, 9.82, 9.82); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 49.08V/m; Power Drift = 0.05 dB
        Peak SAR (extrapolated) = 3.62 W/kg
        SAR(I g) =2.43 W/kg; SAR(10 g) = 1.59 W/kg
        Maximum value of SAR (measured) = 3.08 W/kg

           dB
           0


           —2.09


           —4.18


           —6.26


           —8.35




                      0 dB = 3.08 W/kg = 4.89 dBW/kg




Certificate No: Z16—97239                              Page 5 of 8


            fi                         l;ColIabonmonewith 3


            ‘                         CALIBRATION LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079         Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com         Http://www.chinattl.cn




        Impedance Measurement Plot for Head TSL


      Tri s11 Log Mag 10.00dB/ Ref 0.000dsB [F1]
        900       1 ~g35. 000006 mhz —29. 063 ds
       40. 00


       30. 00
       20. 00
       10. 00
       0. 000 p
      —10. 00
      —20. 00                                                         4
      —30. 00
      —40. 00

      —50. 00                                                         s
    |PM@H s11 smith (R+jx) scale 1.000Uu [F1 del]
       »1   835.00000 mhz    49.239 o —3.4139 o    SSW’               \




                                                        IFBW 100Hz




Certificate No: Z16—97239                              Page 6 of 8


                              8      In Collaboration with

         "__—j"TTI, _
                                     CALIBRATION LABORATORY
          V
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http://www.chinattl.cn




DASYS Validation Report for Body TSL                                  Date: 12.08.2016
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40105
   Communication System: UID 0, CW; Frequency: 835 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 835 MHz; 0 = 0.983 S/m; g, = 54.74; p = 1000 kg/m3
  Phantom section: Right Section
  Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(9.5,9.5, 9.5); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAE4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 57.10 V/m; Power Drift = —0.02 dB
        Peak SAR (extrapolated) = 3.54 W/kg
        SAR(1 g) = 2.44 W/kg; SAR(10 g) = 1.63 W/kg
        Maximum value of SAR (measured) = 3.06 W/kg


           dB
            0


            —1.97


            —3.94


            —5.91


            —1.88

                          L* _———1
            —9.85

                       0 dB = 3.06 W/kg = 4.86 dBW/kg




Certificate No: Z16—97239                              Page 7 of 8


            iiEEEF!EF::;;E;EiE;/      ::Lllflllflfl'flflJfiE;UB()iz;NUI'Y
            m
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com        Http://www.chinattl.cn


        Impedance Measurement Plot for Body TSL


      Tri s11 Log Mag 10.00dB/ Rref 0.000ds [F1]

       990         1 ~g35. 00000 mhz —25. 066 d
       40. 00
       30. 00
       20. 00
       10. 00
       0. 000 p,
      —10. 00
      —20. 00
      —30. 00
      —40. 00

      —50. 00




                                              7
                                                                     Ah.
    P0 sii smith (r+jx) scale 1.000u [F1 del]
       »1   835.00000 mHz     45.756 o —3.2528 o   58.59
                                                                            _




Certificate No: Z16—97239                              Page 8 of 8


Calibration
      f     Laboratory of                                        um
                                                                 s3                                       g Schweizerischer Kalibrierdienst
Schmid & Partner                                                SBE/W/BE                                  @ Service suisse d‘étalonnage
  Engineering AG                                                 Tof                                        Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                     */,fi. w                                  S swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is oneof the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       SGS—SZ (Auden)                                                                   Certificate No: D1750V2—1149_Jun16

CALIBRATION CERTIFICATE                                                                                    $                                  I
 Object                               D1750V2 — SN:1149


 Calibration procedure(s)             QA CAL—05.v9
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                   June 23, 2016


This calibration certificate documents the traceabilty to national standards, which realize the physical units of measurements (S1).
The measurements and the uncertainties with confidence probabilty are given on the following pages and are part of the cortficate.

All callbrations have been conducted in the closed laboratory facifty: environment temperature (22 + 3)°C and humidity < 70%.


Calibration Equipment used (M&TE crtical for calibration)

 Primary Standards                     1D #                      Cal Date (Certificate No.)                       Scheduled Calibration
 Power meter NRP                       SN: 104778                06—Apr—16 (No. 217—02288/02289)                  Apr—17
 Power sensor NRP—Z91                  SN: 108244                06—Apr—16 (No. 217—02288)                        Apr—17
 Power sensor NRP—Z91                  SN: 103245                06—Apr—16 (No. 217—02289)                        Apr—17
 Reference 20 dB Attenuator            SN: 5058 (20k)            05—Apr—16 (No. 217—02202)                        Apr—17
Type—N mismatch combination            SN: 5047.2 / 06827        05—Apr—16 (No. 217—02205)                        Apr—17
 Reference Probe EX3DV4                SN: 7849                  15—Jun—16 (No. EX3—7349_Junt6)                   Jun—17
 DAE4                                  SN: 601                   30—Dec—15 (No. DAE4—601_Dect5)                    Dec—16


Secondary Standards                    1D #                      Check Date (in house)                            Scheduled Check
 Power meter EPM—442A                  SN: GB37480704            07—0ct—15 (No. 217—02222)                         In house check: Oct—16
 Power sensor HP 8481A                 SN: US37202783            07—0ct—15 (No. 217—02222)                        in house check: Oct—16
 Power sensor HP 8481A                 SN: MY41092317            07—0ct—15 (No. 217—02223)                        in house chack: Oct—16
 RF generator R&S SMT—06               SN: 100972                15—Jun—15 (in house check Jun—15)                in house check: Oct—16
 Network Analyzer HP 8753E             S Ussrsaoses              18—0ct—01 (in house check Oct—15)                in house check: Oct—16

                                       Name                                   Function                             Signature
Calibrated by:                        Michael Waber                           Laboratory Technician               %



                                      Katja Pokovic                          Technical Manager                   ////%/:
Approved by:



                                                                                                                  Issued: June 28, 2016
This callbration certificate shall not be reproduced exceptin full without written approval of the laboratory.

Certificate No: D1750V2—1149_Jun16                                   Page 1 of 8


Calibration Laboratory of                                                     g Schweizerischer Kalibrierdienst
Schmid & Partner                                                                Service suisse d‘étalonnage
  Engineering AG                                                              C   Servizio svizzero di taratura
Zeughausstrasse 48, 8004 Zurich, Switzerland
                                                                 mlutaied
                                                                              S swiss Calibration Service

Accredited by the Swiss Accreditation Service (GAS)                            Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM xy,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques®, June 2013
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear(frequency range of 300 MHz to 3 GHz)",
         February 2005
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    *    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    *    SAR measured: SAR measured at the stated antenna input power.
    e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurementis stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D1750V2—1149_Jun16                              Page 2 of 8


Measurement Conditions
    DASY system contfiguration, as far as not iven on page 1.
      DASY Version                                              DASY5                               V52.8.8
      Extrapolation                                   Advanced Extrapolation
      Phantom                                          Modular Flat Phantom
      Distance Dipole Center — TSL                               10 mm                            with Spacer
      Zoom Scan Resolution                               dx. dy, dz =5 mm
      Frequency                                         1750 MHz a 1 MHz



Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                  22.0 °C           40.1               1.37 mho/m
      Measured Head TSL parameters                              (22.0 £0.2) °C    39.8 2 6 %          1.36 mho/m + 6 %
      Head TSL temperature change during test                      <0.5 °C                                    ——

SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                     9.16 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W        36.7 Wikg # 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                     4.87 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W        19.5 Wikg 2 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature                            Conductivity
      Nominal Body TSL parameters                                  22.0 °C           53.4               1.49 mho/m
      Measured Body TSL parameters                              (22.0 £ 0.2) °C   53.8 £ 6 %         1.50 mho/m + 6 %
      Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL

      SAR averaged over 1 cm° (1 g) of Body TSL                  Condition
      SAR measured                                        250 mW input power                     9.27 Wikg
      SAR for nominal Body TSL parameters                     normalized to 1W        37.0 Wikg » 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                 condition

      SAR measured                                        250 mW input power                     4.93 Wikg
      SAR for nominal Body TSL parameters                     normalized to 1W        19.7 Wikg 2 16.5 % (k=2)



Certificate No: D1750V2—1149_Jun16                     Page 3 of 8



Document Created: 2019-01-23 13:50:02
Document Modified: 2019-01-23 13:50:02

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC