Test Report_SAR_Appendix C Calibration certificate_Part 2 of 5

FCC ID: QISCLT-L29

RF Exposure Info

Download: PDF
FCCID_3794471

         i’Tfl:                     ;Collabomtionewith a

                                    CALIBRATION LABORATORY
          e
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Glossary:
TSL                       tissue simulating liquid
Convr                      sensitivity in TSL / NORMx,y,z
N/A                       not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
    devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
    2005
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
    30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
e  Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
e    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.


Certificate No: Z16—97240                              Page 2 of 8


                                    In Collaboration with

          a_‘/"/‘J
           imaggye
                   a
                   CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                           DASY52                                   52.8.8.1258

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                           10 mm                                  with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                       1900 MHz + 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                            Temperature             Permittivity           Conductivity

     Nominal Head TSL parameters                              22.0 °C                  40.0                1.40 mho/m

     Measured Head TSL parameters                        (22.0 + 0.2) °C             40.2 + 6 %          1.38 mho/m +6 %

     Head TSL temperature change during test                  <1.0 °C                   ————                     ————
SAR result with Head TSL
   SAR averaged over 1 C# (1 g) of Head TSL                             Condition
     SAR measured                                                   250 mW input power                 10.1 mW / g

     SAR for nominal Head TSL parameters                             normalized to 1W          40.7 mW ig + 20.8 % (k=2)
     SAR averaged over 10 cm (10 g) of Head TSL                         Condition

     SAR measured                                                   250 mW input power                 5.24 mW / g

     SAR for nominal Head TSL parameters                             normalized to 1W          21.1 mW /g + 20.4 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                            Temperature             Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  53.3                1.52 mho/m

     Measured Body TSL parameters                        (22.0 + 0.2) °C             54.3 + 6 %          1.51 mho/m +6 %

     Body TSL temperature change during test                  <1.0 °C
SAR result with Body TSL
     SAR averaged over 1 cm> (1 g) of Body TSL                          Condition

     SAR measured                                                   250 mW input power                 10.3 mW / g

     SAR for nominal Body TSL parameters                             normalized to 1W          41.6 mW ig + 20.8 % (k=2)

     SAR averaged over 10 cm> (10 g) of Body TSL                        Condition

     SAR measured                                                   250 mW input power                 5.32 mW / g
     SAR for nominal Body TSL parameters                             normalized to 1W          21.4 mW ig + 20.4 % (k=2)
Certificate No: Z16—97240                              Page 3 of 8


          E!T\fl: ;Collabonm'onewith a

          iiaagyye~                 CALBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


 Appendix

 Antenna Parameters with Head TSL


     Impedance, transformed to feed point                                   51.80+ 5.90j0

     Return Loss                                                              — 24.40B



 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   48.10+ 5.82j0

     Return Loss                                                              — 24.10B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.306 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                             SPEAG




Certificate No: Z16—97240                              Page 4 of 8


          mm&                       In Collaboration with

         w‘/"/"J, @9
                                    CALIBRATION LABORATORY
          apg»
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


DASY5 Validation Report for Head TSL                                     Date: 12.07.2016
Test Laboratory: CTTL, Beijing, China
DUT : Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50028
   Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 1900 MHz; 0 = 1.383 S/m; er = 40.16; p = 1000 kg/m3
   Phantom section: Center Section
   Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(7.98, 7.98, 7.98); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 102.5 V/m; Power Drift = 0.06 dB
        Peak SAR (extrapolated) = 18.7 W/kg
        SAR(1 g) =10.1 W/kg; SAR(10 g) = 5.24 W/kg
        Maximum value of SAR (measured) = 14.4 W/kg


           dB
            0


            —3.61


            —1.22


            —10.83


            —14.44

                          k
            —18.05

                       0 dB = 14.4 W/kg = 11.58 dBW/kg




Certificate No: Z16—97240                              Page 5 of 8


           b@                            in Collaboration with
          a_‘/"/‘]J, a
           V                             CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http:/www.chinattl.cn




        Impedance Measurement Plot for Head TSL


     Tri s11 Log Mag 10.00dB/ Rref 0.000dB [F1]
       50. 00
                  »1   1.9000000 GHz —24.374 dB
       40. 00

       30. 00

       20. 00

       10. 00

       0. 000 p




                                                                            P &A
      —10. 00

      —20. 00

      —30. 00

      —40. 00

      —50. 00                                                           a
    >WGi sil1 smith (R+jx) scale 1.000Uu [F1 Ddel]
      »1 1.9000000 GHz        51.759 o     5.9039 n;}f\




    Torempopmprremurceerorenronoommiraciirrarrrtaisoreonies




Certificate No: Z16—97240                                 Page 6 of 8


          mm &                      In Collaboration with

          aa_‘/"/"J, a
                                    CALIBRATION LABORATORY
            \pppemr~
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn




DASY5S Validation Report for Body TSL                                    Date: 12.07.2016
Test Laboratory: CTTL, Beijing, China
DUT : Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50028
   Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 1900 MHz; 0 = 1.506 S/m; &, = 54.26; p = 1000 kg/m*
   Phantom section: Right Section
   Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(7.7, 7.7, 7.7); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 99.69 V/m; Power Drift = —0.03 dB
        Peak SAR (extrapolated) = 18.8 W/kg
        SAR(I g) = 10.3 W/kg; SAR(10 g) = 5.32 W/kg
        Maximum value of SAR (measured) = 14.8 W/kg


           dB
            0


            —3.50


            —£.00


            —10.50


            —14.00

                         L.                       y            es           y
            —17.50

                      0 dB = 14.8 W/kg =11.70 dBW/kg




Certificate No: Z16—97240                             Page 7 of 8


            _ifi©                     in Collaboration with
                      a
            w ‘/"/"J, CALIBRATION
            -‘,                   LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http://www.chinattl.cn


        Impedance Measurement Plot for Body TSL


      Trl Ss11 Log Mag 10.00ds/ ref 0.000ds [F1]
        ‘9— 90     (<1~1,5000000 enz —24. 127 d
        40. 00
        30. 00
        20. 00
        10. 00
        0. 000 p                                                                       4
       —10. 00
      —20. 00
      —30. 00
       —40. 00

      —50. 00                                                         +
     PB si1 smith (R+jxX) scale 1.000U [F1 Ddel]
       »1   1.9000000 GoHz    48.141 o   5.8232 o   487.7



                                               /




                                                                              inould
                                                                            a e men




Certificate No: Z16—97240                               Page 8 of 8


                 A‘E                     In Collaboration with                 \&\\\\\
                                                                                   4/’/”           AB yA

                 a®‘/"/"J, s_
                           _ s\                                                                                       AFEL
                 ... cuunnuroct                                                iPcyirOVAS it




                                                                                            I,/"
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China %,/R\               v CALIBRATION
             Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504            "/,,,,,h,\“\‘        CNAS LO570
             E—mail: cttl@chinattl.com       Http://www.chinattl.cn
      Client              SGS(Boce)                                         Certificate No:        Z16—97242

CALIBRATION GERTIFICATE

Object                                     D2450V2 — SN: 733

Calibration Procedure(s)                   FD—211—003—01
                                           Calibration Procedures for dipole validation kits

Calibration date:                          December 7, 2016


This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #               Cal Date(Calibrated by, Certificate No.)             Scheduled Calibration
 Power Meter     NRP2             101919            27—Jun—16 (CTTL, No.J16X04777)                              Jun—17
 Power sensor       NRP—Z91       101547            27—Jun—16 (CTTL, No.J16X04777)                              Jun—17
 Reference Probe EX3DV4           SN 7433           26—Sep—16(SPEAG,No.EX3—7433_Sep16)                          Sep—17
 DAE4                             SN 771            02—Feb—16(CTTL—SPEAG,N0.Z16—97011)                          Feb—17


 Secondary Standards              ID #              Cal Date(Calibrated by, Certificate No.)          Scheduled Calibration
 Signal Generator E4438C          MY49071430        O01—Feb—16 (CTTL, No.J16X00893)                             Jan—17
 Network Analyzer E5071C          MY46110673         26—Jan—16 (CTTL, No.J16X00894)                             Jan—17



                                   Name                          Function                                     Signature
Calibrated by:                   Zhao Jing                 SAR Test Engineer                                a> é Z/ d

Reviewed by:                     Qi Dianyuan                SAR Project Leader                     M/

Approved by:                     Lu Bingsong                Deputy Director of the laboratory         f     ,2 ’%fi

                                                                                     Issued: Decembér 11, 2016
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.                   J



    Certificate No: Z16—97242                              Page 1 of 8


           m*                         in Collaboration with
           ex‘/"[‘]J, CALIB
            uiaaagee"
                      a     RATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Glossary:
TSL                       tissue simulating liquid
Convr                      sensitivity in TSL / NORMx,y,z
N/A                       not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
  Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) For hand—held
   devices used in close proximity to the ear (frequency range of 300MHz to 3GHz)", February
    2005
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
    communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz


Additional Documentation:
e) DASY4/5 System Handbook


Methods Applied and Interpretation of Parameters:
e  Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
e Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
e     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
e     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
e     SAR measured: SAR measured at the stated antenna input power.
e     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
e     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.



Certificate No: Z16—97242                            Page 2 of 8


                                      In Collaboration with

          ETTL a
          yar~~
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2 504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
     DASY Version                                        DASY52                                      52.8.8.1258

     Extrapolation                               Advanced Extrapolation

     Phantom                                     Triple Flat Phantom 5.1C

     Distance Dipole Center — TSL                        10 mm                                     with Spacer

     Zoom Scan Resolution                           dx, dy, dz = 5 mm

     Frequency                                      2450 MHz + 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                          Temperature               Permittivity           Conductivity

     Nominal Head TSL parameters                              22.0 °C                  39.2                1.80 mho/m

     Measured Head TSL parameters                        (22.0 + 0.2) °C             39.4 +6 %           1.81 mho/m +6 %

     Head TSL temperature change during test                  <1.0 °C                   —_———                    «w_=_~

SAR result with Head TSL
     SAR averaged over 1 C#2" (1 g) of Head TSL                         Condition
     SAR measured                                                  250 mW input power                  13.3 mW / g

     SAR for nominal Head TSL parameters                            normalized to 1W            53.1 mW /g + 20.8 % (k=2)

     SAR averaged over 10 Cm° (10 g) of Head TSL                        Condition
     SAR measured                                                  250 mW input power                  6.22 mW / g

     SAR for nominal Head TSL parameters                            normalized to 1W            24.9 mW ig + 20.4 % (k=2)

Body TSL parameters
   The following parameters and calculations were applied.
                                                          Temperature               Permittivity           Conductivity

     Nominal Body TSL parameters                              22.0 °C                  52.7                1.95 mho/m

     Measured Body TSL parameters                        (22.0 + 0.2) °C             53.1 +6 %           1.94 mho/m +6 %

     Body TSL temperature change during test                  <1.0 °C                   ————                     _———

SAR result with Body TSL
   SAR averaged over 1 Cm° (1 g) of Body TSL                            Condition
     SAR measured                                                  250 mW input power                  12.7 mW / g

     SAR for nominal Body TSL parameters                            normalized to 1W            51.0 mW /g + 20.8 % (k=2)

     SAR averaged over 10 cm~ (10 g) of Body TSL                        Condition
     SAR measured                                                  250 mW input power                  5.85 mW / g

     SAR for nominal Body TSL parameters                            normalized to 1W            23.5 mW /g + 20.4 % (k=2)


Certificate No: Z16—97242                            Page 3 of 8


          b~                            In Collaboration with

          '-'_-"TTL CALIBRATION
                    a
          e                     LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http://www.chinattl.cn


 Appendix

 Antenna Parameters with Head TSL


     Impedance, transformed to feed point                                   52.90+4.11j0
     Return Loss                                                             — 26.30B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                   49.70+ 5.900

     Return Loss                                                             — 24.6d0B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                         1.257 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.


 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data


     Manufactured by                                                             SPEAG




Certificate No: Z16—97242                              Page 4 of 8


         :-__!T\fl:                    ;Collaborationewith a

          m                           CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http://www.chinattl.cn


DASY5 Validation Report for Head TSL                                     Date: 11.07.2016
Test Laboratory: CTTL, Beijing, China
DUT : Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 733
   Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 2450 MHz; 0 = 1.809 S/m; er = 39.42; p = 1000 kg/m3
   Phantom section: Center Section
   Measurement Standard: DASY5 (IEEE/IEC/ANSI C€63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(7.45, 7.45, 7.45); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 106.8 V/m; Power Drift = 0.03 dB
        Peak SAR (extrapolated) =27.5 W/kg
        SAR(I g) = 13.3 W/kg; SAR(10 g) = 6.22 W/kg
        Maximum value of SAR (measured) = 20.4 W/kg


           dB
            0


            —4.34


            —8.69


            —13.03


            —17.38
                          2

                         .«
            —21.72

                       0 dB = 20.4 W/kg = 13.10 dBW/kg




Certificate No: Z16—97242                            Page 5 of 8


           «en TTL                       w
           zu*
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079         Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com         Http:/www.chinattl.cn




        Impedance Measurement Plot for Head TSL


     Tri s11 Log Mag 10.00dB/ ref 0.000dB [F1]
       50.00      »1   2.4500000 GHz —26.266 ds
       40. 00
       30. 00
       20. 00
       10. 00
       0. 000 p
      —10. 00
      —20. 00                                                         1
      —30. 00
      —40. 00

      —50. 00                                                        P
    > si1 smith (R+jX) scale 1.000U0 [F1 Ddel]
      »1   2.4500000 coHz     52.850 n   4.1123 o   267.1




Certificate No: Z16—97242                              Page 6 of 8


          im©                         In Collaboration with

         @‘/"/‘/, a
          y                           CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Http:/www.chinattl.cn


DASY5S Validation Report for Body TSL                                   Date: 12.07.2016
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 733
   Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 2450 MHz; 0 = 1.943 S/m; s = 53.12; p = 1000 kg/m*
   Phantom section: Right Section
   Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)
DASY5 Configuration:

            e    Probe: EX3DV4 — SN7433; ConvF(7.46, 7.46, 7.46); Calibrated: 9/26/2016;
            e    Sensor—Surface: 2mm (Mechanical Surface Detection)
            e    Electronics: DAF4 Sn771; Calibrated: 2/2/2016
            e    Phantom: Triple Flat Phantom 5.1C; Type: QD 000 P51 CA; Serial: 1161/1
            e    Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7372)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 98.60 V/m; Power Drift =—0.01 dB
        Peak SAR (extrapolated) = 26.0 W/kg
        SAR(I g) =12.7 W/kg; SAR(10 g) = 5.85 W/kg
        Maximum value of SAR (measured) = 19.2 W/kg


           dB
           0


            —4.34


            —8.68


            —13.01


            —17.35




                       0 dB = 19.2 W/kg = 12.83 dBW/kg




Certificate No: Z16—97242                            Page 7 of 8


            *s                          in Collsboration with
            uTIL _p_e_a
            oha  S e _2

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079         Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com         Http://www.chinattl.cn


        Impedance Measurement Plot for Body TSL


      Tr1 S11 Log Mag 10.00dB/ Rref 0.000ds [F1]
        "9—90 i ~~3.4500000 enz —24. 557 ds
       40. 00
       30. 00
       20. 00
       10. 00
       0. 000 p
      —10. 00                                                               a
      —20. 00
      —30. 00
      —40. 00

      —50. 00                                                         ic
    U s1i1 smith (R+jX) scale 1.000Uu [F1 Ddel]
       »1   2.4500000 GoHz   49.710 o   5.9037 o   383.5




Certificate No: Z16—97242                             Page 8 of 8


Calibration
       r        Laboratory of                                                                               g Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                            g Service suisse d‘étalonnage
  Engineering AG                                                                                              Servizio svizzero d taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                S swiss Calibration Service

Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is oneof the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates
Client       SGS—SZ (Auden)                                                                     Certiicate No: D2600V2—1125_Jun16

CALIBRATION CERTIFICATE                                                                                                                         l
 Object                                D2600V2 — SN: 1125


 Calibration procedure(s)              QA CAL—05.v9
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     June 22, 2016


 This calibration certificate documents the traceabilty to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probabiliy are given on the following pages and are part of the certificate.

 All calibrations have been conducted in the closed laboratory faciliy: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)

 Primary Standards                      i#                        Cal Date (Certificate No.)                        Scheduled Galibration
 Power meter NRP                        SN: 104778                06—Apr—16 (No. 217—02288/02289)                   Apri7
 Power sensor NRP—291                   SN: togase                06—Apr—16 (No. 217—02288)                         Aprt7
 Power sensor NRP—291                   SN: 103245                06—Apr—16 (No. 217—02289)                         Apr17
 Reference 20 dB Aftenuator             SNi: sose (20k)           05—Apr—16 (No. 217—02202)                         Apri7
 Type—N mismatch combination            SN: 5047.2 / 06827        O6—Apr—16 (No. 217—02205)                         Apri7
 Reference Probe EXSDV4                 SN: 7349                  15—Jun—16 (No. EX3—7349_Jun16)                    Jun—17
 Dagd                                   SN: 601                   0—Dec—15 (No. DAE4—601_Dect5)                     Dec—16
 Secondary Standards                    i9 #                      Check Date (in house)                             Scheduled Check
 Power meter EPM—442A                   SN: aBa74so704            07—0ct—15 (No. 217—02222)                         in house check: Oct—16
 Power sensor HP 8481A                  SN: Ussrasa7es            07—0ct—15 (No. 217—02222)                         in house check: Oct—16
 Power sensor HP 8481A                  SN: MY41002317            07—0ct—15 (No. 217—02223)                         in house chect
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Jun—15)                 in house check: Oct—16
 Network Analyzer HP 8759E              SN: ussrsaoses            18—0ct—01 (in house check Oct—15)                 in house check: Oct—16

                                        Name                                   Function                               igndjure
 Calibrated by:                         Claudio Leubler                        Laboratory Technician                      /G


                                        Katja Pokovie                          Technical Manager               //
 Approved by:



                                                                                                                    Issued: June 23, 2016
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.

Certificate No: D2600V2—1125_Jun16                                    Page 1 of 8


Calibration Laboratory of                                                        SchweizerischerKalibrierdienst
Schmid & Partner                                                                 Service suisse d‘étalonnage
  Engineering AG                                                                 Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                     Swiss Calibration Service

Aceredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates
Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    *    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
         SAR measured: SAR measured at the stated antenna input power.
         SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    «_   SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurementis stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2600V2—1125_Jun16                              Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not iven on page 1.
    DASY Version                                            DASYS                                  V52.8.8
     Extrapolation                                   Advanced Extrapolation
     Phantom                                          Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                               dx, dy, dz =5 mm
     Frequency                                         2600 MHz * 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           39.0               1.96 mho/m
     Measured Head TSL parameters                              (22.0 £0.2) °C    38.0 £ 6 %         2.04 mho/m x 6 %
     Head TSL temperature change during test                      <0.5 °C                                    «o

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                 14.5 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        56.6 Wikg * 17.0 % (k=2)

     SAR averaged over 10 m‘ (10 g) of Head TSL                  condition
     SAR measured                                        250 mW input power                     6.43 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W        25.4 Wikg 2 16.5 % (k=2)

Body TSL parameters
   Thefollowing parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Body TSL parameters                                  22.0 °C           52.5               2.16 mho/m
     Measured Body TSL parameters                              (22.0 £ 0.2) °C   52.226 %           2.21 mho/m £ 6 %
     Body TSL temperature change during test                      <0.5 °C            ~~—

SAR result with Body TSL

     SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
     SAR measured                                        250 mW input power                     13.7 Wig
     SAR for nominal Body TSL parameters                     normalized to 1W        54.2 Wikg 2 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Body TSL                 condition

     SAR measured                                        250 mW input power                 6.12 Whg
      SAR for nominal Body TSL parameters                    normalized to 1W        24.3 Wikg 2 16.5 %(k=2)



Certificate No: D2600V2—1125_Jun16                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           49.0 0 — 4.9 jQ
      Return Loss                                                                       —25.9 dB


Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           45.7 9 — 4.0 jQ
      Return Loss                                                                       —24.2 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                       ‘                           1.155 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are notaffected by this change. The overall dipole length is stil
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                               October 22, 2015




Certificate No: D2600V2—1125_Jun16                       Page 4 of 8



Document Created: 2018-03-25 09:39:14
Document Modified: 2018-03-25 09:39:14

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC