SAR Calibration

FCC ID: QDMPPF32

RF Exposure Info

Download: PDF
FCCID_4006988

                KCTL Inc.                                   Report No.:
             65, Sinwon-ro, Yeongtong-gu,
         Suwon-si, Gyeonggi-do, 16677, Korea               KR18-SPF0006
    Tel : 82-31-285-0894 Fax : 82-505-299-8311               Page (28) of (53)
                    www.kctl.co.kr




Appendix A. Calibration certificate
A.1 Probe Calibration certificate




            This test report shall not be reproduced, except in full, without the written approval.
KCTL-TIA002-004/1


                                                                                         — KCTL
Calibration Laboratory of                                                                       Schweizerischer Kallbrirdionst




                                                                                        w o w
Schmid & Partner                                                                                Service suisse détalonnage
  Engineering AG                                                                                Servizio svizzero dtaratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                    Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                        Acereditation No.: SCs 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multiateral Agroement for the recognition of calibration cortficates
Glossary:
TSL                         tissue simulating liquid
NORMxcy,z                   sensitivityin free space
ConvF                       sensitivity in TSL / NORMy,z
DCP                         diode compression point
CF                          crest factor (1/duty_cycle) of the RF signal
A, B,C, D                   modulation dependentlinearization parameters
Polarization                 rotation around probe axis
Polarization 8              8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                            Le.        0 is normal to probe axis
Connector Angle             information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak         Spatial—Averaged Specific
         Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
         Techniques®, June 2013
    b)   12C 622001 , ‘Measurement procedure for the assessmentof Specific Absorption Rate (SAR) from hand—
         held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)®, July 2016
    0    IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
         used in close proximity to the human body (frequency rangeof 30 MHz to 6 GHz)*, March 2010
   d     KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Methods Applied and Interpretation of Parameters:
         NORMxy,z: Assessed for E—field polarization 8 = 0 (f = 900 MHz in TEM—cell; {> 1800 MHz: R22 wevegulde}
         NORMxy,z are only intermediate values, ie., the uncertainties of NORMx,y,z does not affect the E*—field
         uncertainty inside TSL (see below ConvF).
         NORM()xy,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This linarization is
         implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency responseis included
         in the stated uncertainty of ConvF.
         DCPxy,z: DCP are numerical inearization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
         PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
         characteristics
         Axy.z; Buy,z: Coy.z; Dxy,z; VRxy,z: A, B, C, D are numerical inearization parameters assessed based on
         the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
         media. VR is the maximum calibration range expressed in RMS voltageacross the diode.
         Convand Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature Transfer
         Standard for f s 800 MHz) and inside waveguide using analytical field distributions based on power
         measurements for 1 > 800 MHz. The same setups are used for assessment of the parameters applied for
         boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
         used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
         to NORMxy,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
         ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to + 100
         MHz.
         Spherical isotropy (3D deviation from isotropy); in a field of low gradients realized using a flat phantom
         exposed by a patch antenna.
         Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
         (on probe axis). No tolerance required.
         Connector Angle: The angle is assessed using the information gained by determining the NORMx (no
         uncertainty required).


Certficate No: EX3—3028_Jan18                                  Page 2 of 11


EXGDVé — SN:3028                                                               January 23, 2018




                     Probe EX3DV4

                                        SN:3928

                                 Manufactured:             March 8, 2013
                                 Calibrated:               January 23, 2018




                                Calibrated for DASY/EASY Systems
                                   (Note: non—compatible with DASY2 system!)




Certficate No: EX3—3028_Jant8                     Page 3 of 11


                                                                                                                               KCTL
EXGDV4— SN:3028                                                                                                           January 23, 2018



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3928
Basic Calibration Parameters
                                               Sensor K                       Sensor Y                     SensorZ              Une (ke2)
 Norm (Vi(Vim))                                  0.49                           021                          055                £10.1%
 DCP (mV)"                                       se                             EFG                         100.6

Modulation Calibration Parameters
 up            Communication System Name                                  A            s           6          b         vr          Une®
                                                                         aB        aByjv                    as          my         (ke2)
 o             ow                                                x       oo         o0            10        on         fses        #25%
                                                                 y       o0          o0           10                   Tse8
                                                                 2       o0          o0           10                   1478

 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.


‘ The uncertainties of Norm XY.Z do not affect the E*feld uncertainty insde TSL (see Pages 5 and 6)
* Numericalinearization parameter: uncertainty not required.
 Uncertainty is determined using the max. deviation from linear response applying rectangular distrbution and is expressed forthe square of the
field value.




Certficate No: EX3—3028_.Jant8                                    Page 4 of 11


EXGDV4~ SN:3028                                                                                                          January 23, 2018




DASY/EASY — Parameters of Probe: EX3DV4 — SN:3928

Calibration Parameter Determined in Head Tissue Simulating Media
                       Relative         Conductivity                                                              Depth®           Une
     £(MHz)®_|_      Pormittivity"          (Sim)"         ConvFX        ConvEY        ConvEZ        Aipha®|       (mm)           (ke2)
        750              410                 0.89             9.79         a.79           9.70         0.44         0.80        £120%
        850              41.5                0.92             9.47         9.47           9.47         0.41         0.88        £12.0%

        900              41.5                0.97             9.28         9.28           9.28         0.37         0.97        £12.0%

       1750              40.1                1.37             8.34         8.34           8.34         0.29         0.80        £12.0%

       1900              40.0                140              8.09         8.09           8.00         0.34         0.80        £120%
       2300              39.5                1.67             7.68         7.68           7.68         0.37         0.80        £12.0%

       2450              39.2                1.80             7.28         7.28           7.28         0.34         0.94        £12.0%

       2600              39.0                196              717          77             717          0.35         084|        £120%
       5200              36.0                4.66             5.40         5.40           5.40         0.35         180         £181%
       5300              35.9                4.76             5.15         5.15           5.15         0.35         180         £131%
       5500              35.6                4.96             4.98         4.98           4.98         0.40         180         £13.1%
       5600              35.5                5.07             484          4.84           4.84         0.40         180         £13.1%
       5800              35.3                527              4.79         4.79           4.70         0.40         180         £181%

© Frequency valdty above 300 Mz of : 100 MHz only apples for DASY v4.4 and higher (see Page 2), es i is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvE: uncertainty at calloration frequency and the uncertaintyfor the indlcated frequency band. Frequency valdty
below 300 Mtiz is + 10, 25, 40, 50 and 70 MHz for Convassessments at 20, 64, 128, 150 and 220 MHz respectvely. Above 5 GHz frequency
yalidty can be extended to + 110 MHz.
" Atfrequencies below 3 GH, the validty ofssue parameters (e and a) can be relaxed to + 10% ifiquid compensation formula is applied to
measured SAR values. At requencies above 3 GHiz,the validty oftissue parameters (s and a)is restrcted to + 5%. The uncertaiy is the RSS of
the Convuncertainty foindicated target tssue parameters.
* AhaDapth are determined during calbration. SPEAG warrants thatthe remaining deviation due to the boundary effect ater compensation is
always less than : 19% for requencies below 3 GHiz and below * 2% for fequencies between 3—6 GHz at any distance larger than halfthe probe tip
ciameter from the boundary.




Certficate No: EX3—3928_Jan18                                    Page 5 of 11


                                                                                                                                KCTL
EXGDV4— SN:3028                                                                                                            January 23, 2018



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3928

Calibration Parameter Determined in Body Tissue Simulating Media
                Relative    Conductivity                                                                            Depth             Une
    £(MHz)_|_ Pormittivity"    (Sim)"    ConvFX ConvEY ConvFZ Aipha®                                                 |__(mm)         (ke2)

        750               588                 0.%6             9.70          9.70          9.70          0.42         0.89        £120%
        850               55.2                0.99             9.36          9.36          9.36          0.48         0.84        £120%
        00                55.0                1.05             9.26          9.26          9.26          0.44         089         £120%
       1750               53.4                149              7.85          785           7.85          o43          084         £120%
       1900               53.3                152              7.50          7.59          7.59          0.43         0.80        £120%
       2300               52.9                1.81             744           744           744           0.38         0.85        £12.0 %

       2450               52.7                1.95             7.38          7.38          7.38          0.37         0.85        £12.0 %

       2600               525                 2.16             6.87          6.87          6.87          0.25         1.02        £12.0 %

       5200               49.0                5.30             4.85          4.85          4.85          0.35         190         £131%
       5300               48.9                542              4.65          4.65          4.65          0.35         1.90        £13.1%

       5500               48.6                5.65             4.30          4.30          4.30          0.40         1.90        £13.1%
       se00               48.5                577              4.10          4.10          4.10          0.40         100         a18.1%
       5800               48.2                £.00             4.37          4.37          4.37          0.40         190         £131%


© Frequency valdty above 300 Mz of : 100 MHz only apples for DASY v4.4 and higher (see Page 2else ts resticted to + 50 MHz. The
uncertainty is the RSS o the Conv: uncertainty at calloration frequency and the uncertainty for the indicated frequency band. Frequency valdity
below 300 Mz is x 10, 25,40, 50 and 70 MHz for ConvE assessments at 30, 64, 128, 150 and 220 VHz respectively. Above 5 GHz frequency
validty can be extended to + 110 iz
* Atfrequencies below 3 GHz,the validty ofissue parameters (e and a) can be relaxed to : 10% ifiquid compensation formula is applied to
measured SAR values. At requencies above 3 GHz,the validty oftissue parameters (e and 0)is restrcted to + 5%. The uncertaints the RSS of
the Corv‘® uncertainty foindicated target issue parameters.
* AphaDepth are determined during caltration. SPEAG warrants thatthe remaining deviation due to the boundary effect aftr compensation is
always less than + 1% for frequencies below 3 GHiz and below + 2% for fequencies between .6 GHz at any ditance larger than halfthe probe tip
diameter rom the boundary:




Certficate No: EX3—3028_Jan18                                     Page 6 of 11


EXGDV4~ SN:3028                                                                                         January 23, 2018


                                            Frequency Response of E—Field
                                                (TEM—Cell:ifi110 EXX, Waveguide: R22)
         Frequency response (normalized)




                                           Uncertainty of Frequency Response of E—field: 2 6.3% (k=2)




Certficate No: EX3—3928_Jan18                                     Page 7 of 11


                                                                                               _ |KCTL
EXGDV4— SN:3028                                                                                                  January 23, 2018



                                           Receiving Pattern (¢), 8 = 0°

                       £=600 MHz,TEM                                                            £21800 MHz,R22




          we                                                      o                  i




                                r
                as                  >                    is                                as

                                        "—@s
            *                   *                    *            *                   #              xo




                                                                                                                     n
           To                   x                    ®            2                  To




                                                              £cicc              ioi
                     ‘ 4‘%0 I                  ® I         -g.fl              6
                                                                           RotP]
                                                                                                             &
                          10%%!11                        sc%’—u                           1e    Wz          sWz

                                         Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Certiicate No: EX3—3928_Jan18                                         Page 8 of 11


EXGDV4— SN:3928                                                                                           January 23, 2018




                                       Dynamic Range f(SARneaq)
                                                  (TEM cell , fou= 1900 MHz)




                  €    10°

                  F3
                  5
                  &
                  &
                  &




                             103           102          10             10       10‘           10     10
                                                                SAR {mWiem9]
                                                  C+]                                 Ce]
                                          not compensated                        compensated



                             2

                             1
                       @         .
                       &
                       wClkan
                           07               s
                       w

                                             T           V  4          I        t            t
                                 109        10@         101           100      101          10°    108
                                                                SAR {mWiomd]
                                                 C+]
                                           not compensated                     compensate
                                       Uncertainty of Linearity Assessment: £ 0.6% (k=2)




Certiicate No: EX3—3028_Jan18                                   Page 9 of 11


EXGDV4— SN:3028                                                                                   January 23, 2018


                                 Conversion Factor Assessment
                  1= 850 MHz.WGLS RO (H_convr)                           1= 1900 MHz.WGLS R22 (H_comr)
                                                                   s
        i
        .                                                          s[
        w5    B                                                 s
                  %
                                                               ¢s
                                                               2
                                                               a
        is
        io
        se

                       >           sfim       a                                a      sf
                      «3B                    wille                           wiBle
                                 Deviation from Isotropy in Liquid
                                                Error (6, 3), f= 900 MHz
                        jation




                             «10    —o8   ~o6   ~os   02    o0      o2     o4    o%   o8   10
                                 Uncertainty of Spherical Isotropy Assessment: £ 2.6%(k=2)




Certiicate No: EX3—3928_Jan18                          Page 10 of 11


                KCTL Inc.                                   Report No.:
             65, Sinwon-ro, Yeongtong-gu,
         Suwon-si, Gyeonggi-do, 16677, Korea               KR18-SPF0006
    Tel : 82-31-285-0894 Fax : 82-505-299-8311               Page (38) of (53)
                    www.kctl.co.kr




            This test report shall not be reproduced, except in full, without the written approval.
KCTL-TIA002-004/1


                KCTL Inc.                                   Report No.:
             65, Sinwon-ro, Yeongtong-gu,
         Suwon-si, Gyeonggi-do, 16677, Korea               KR18-SPF0006
    Tel : 82-31-285-0894 Fax : 82-505-299-8311               Page (39) of (53)
                    www.kctl.co.kr


A.2 Dipole Calibration certification
D2450V2




            This test report shall not be reproduced, except in full, without the written approval.
KCTL-TIA002-004/1


 Calibration Laboratory of                                                S   Schweizerischer Kalibrierdienst
 Schmid & Partner                                                         c   Service suisse d‘étalonnage
   Engineering AG                                                             Servizio svizzero di taratura
 Zeughausstrasse43, 8004 Zurich, Switzerland
                                                                          S   swiss Calibration Service

 Accredited by the Swiss Accrediation Service (SAS)                       Acereditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration cortificates
Glossary:
TSL                         tissue simulating liquid
ConvF                       sensitivity in TSL / NORM xy,z
N/A                         not applicable or not measured
Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques®, June 2013
    b) 1EC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
       (SAR) from hand—held and body—mounted devices used nextto the ear (frequency range of
       300 MHz to 6 GHz)®, July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
  * Antenna Parameters with TSL: The dipole is mounted with the spacerto position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
  * Feed Point Impedance and Returm Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurementat the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
  * Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
  * SAR measured: SAR measured at the stated antenna input power.
    *    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    *    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.
 The reported uncertainty of measurementis stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D2450V2—805_Jul18                           Page 2 of 8


                 flls cevemes reee cesmeee cee                                            |         ‘
             Gurnn we ogpevegagr ser rev in g en =—
        Mek l ue ol es se s l s 2e ns                                                        [

i—mms       m              mss               mss               mss                     mss             mss                mss                mss   ns




                  Measurement Conditions
                     DASY system configuration, as far as not given on page 1.
                        DASY Version                                                 DASY$                              V52.10.1
                        Extrapolation                                  Advanced Extrapolation
                        Phantom                                         Modular Flat Phantom
                        Distance Dipole Center— TSL                                  10 mm                             with Spacer
                       Zoom Scan Resolution                                dx, dy, dz =5 mm
                        Frequency                                         2450 MHz x 1 MHz



                  Head TSL parameters
                     The following parameters and calculations were     applied.
                                                                                     Temperature       Permittivity         Conductivity
                       Nominal Head TSL parameters                                      22.0 °C           392                1.80 mho/m
                       Measured Head TSL parameters                                 (22.0 £0.2) °C     37.8 £ 6 %         1.85 mho/m + 6 %

                       Head TSL temperature change during test                          <0.5 °C            m                       d


                  SAR result with Head TSL

                       SAR averaged over 1 m (1 g) of Head TSL                        Condition
                       SAR measured                                           250 mW input power                  13.1 Wig
                       SAR for nominal Head TSL parameters                         normalized to 1W        51.3 Wikg = 17.0 % (k=2)

                       SAR averaged over 10 m‘ (10 g) of Head TSL                     condition
                       SAR measured                                           250 mW input power                      6.10 Whg
                       SAR for nominal Head TSL parameters                         normalized to 1W        24.1 Wikg £16.5 % (k=2)

                  Body TSL parameters
                     The following parameters and calculations were applied.
                                                                                     Temperature       Permittivity         Conductivity
                       Nominal Body TSL parameters                                      22.0 °C           52.7               1.95 mho/m
                       Measured Body TSL parameters                                  (22.0 £0.2) °C    51.9 26 %          2.02 mho/m + 6 %
                        Body TSL temperature change during test                         <0.5 °C            ~=—                     «e


                  SAR result with Body TSL

                       SAR averaged over 1 cm* (1 g) of Body TSL                      Condition
                        SAR measured                                          250 mW input power                      12.9 Wig

                        SAR for nominal Body TSL parameters                        normalized to 1W         50.6 Wikg 17.0 % (k=2)

                        SAR averaged over 10 cm* (10 g) of Body TSL                    condition
                        SAR measured                                          250 mW input power                      6.03 Wy
                        SAR for nominal Body TSL parameters                        normalized to 1W         23.8 Wikg 2 16.5 % (k=2)




                  Certiicate No: D2450V2—895_Jul18                      Page 3 of 8


 Appendix (Additional assessments outside the scope of SCS 01 08)

Antenna Parameters with Head TSL

       Impedance, transformed to feed point                                           53.8 0+ 1.8 j0
       Return Loss                                                                       —27.9 dB


Antenna Parameters with Body TSL

       Impedance, transformed to feed point                                          49.2 0 + 5.0 jQ
       Retun Loss                                                                       —25.9 dB



General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.156 ns                         _
After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxia! cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—cirouited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
‘Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is stil
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                June 19, 2012




Certificate No: D2450V2—805_Jul18                        Page 4 of 8


 DASY5 Validation Report for Head TSL

                                                                                      Date: 24.07.2018
 Test Laboratory: SPEAG, Zurich, Switzerland

 DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:895

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.85 S/m; s, = 37.8; p = 1000 kg/m?
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:
    *    Probe: EX3DV4 — SN7349; ConvF(7.88, 7.88, 7.88) @ 2450 MHz:; Calibrated: 30.12.2017

    *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *    Electronics: DAE4 Sn601; Calibrated: 26.10.2017
    *    Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    «_   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 115.0 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 26.1 W/kg
SAR(I g) = 13.1 W/kg; SAR(IO g) = 6.1 W/kg
Maximum value of SAR (measured) = 21.4 W/kg


         dB
         0


         ~4.40


         —8.80


         113.20


         —17.60


         —22.00
                   0 dB =21.4 W/kg = 13.30 dBW/kg




Certiicate No: D2450V2—895_Jul18                Page 5 of 8


                                                                       KCTL
 Impedance Measurement Plot for Head TSL




                                                    2450000 GHz        53.813 O
                                                        115.00 pH      1.7703 0
                                                    2450000 GHz      40.493 mU
                                                                        23.925°




                   Ch tAug= 20
           Chi: Stat 225000 Oz   ——                                 Stop 265000 Gite
         0.00
          oo                                                        —2}.852 dB
          oo
          oo


          oo
           so
           co
           uo
           m      ch
           Cht: Sn




Certificate No: D2450V2—895_Jul18     Page 6 of 8


 DASY5 Validation Report for Body TSL

                                                                                      Date: 24.07.2018

Test Laboratory: SPEAG, Zurich, Switzerland
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:895

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 2.02 S/m; s, = 51.9; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    *    Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz:; Calibrated: 30.12.2017

    *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *    Electronics: DAE4 Sn601; Calibrated: 26.10.2017
    *    Phantom: Flat Phantom5.0 (back); Type: QD 000 P50 AA; Serial: 1002

    *«   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=Smm, dz=5mm
Reference Value = 108.0 V/m; Power Drift =—0.04 dB
Peak SAR (extrapolated) = 25.1 W/kg
SAR(I g) = 12.9 W/kg; SAR(1O g) = 6.03 W/kg
Maximum value of SAR (measured) = 20.9 W/kg




                            —12s0

                            —1800

                            2100
                                    0 dB = 20.9 W/kg = 13.20 dBW/kg




Certificate No: D2450V2—895_Jul8                  Page 7 of 8


                KCTL Inc.                                   Report No.:
             65, Sinwon-ro, Yeongtong-gu,
         Suwon-si, Gyeonggi-do, 16677, Korea               KR18-SPF0006
    Tel : 82-31-285-0894 Fax : 82-505-299-8311               Page (46) of (53)
                    www.kctl.co.kr




            This test report shall not be reproduced, except in full, without the written approval.
KCTL-TIA002-004/1