SAR test report 2 of 5

FCC ID: PY7-80422E

RF Exposure Info

Download: PDF
FCCID_4231096

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/27


System Check_Head_750MHz

DUT: D750V3—1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: HSL_750_190227 Medium parameters used: f = 750 MHz; 0 = 0.887 S/m; &, = 42.962; p = 1000
kg/m3
Ambient Temperature : 23.4 °C ; Liquid Temperature : 22.4 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(6.56, 6.56, 6.56) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Left; Type: QDOOOP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.38 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 52.70 V/m; Power Drift = —0.01 dB
Peak SAR (extrapolated) = 3.02 W/kg
SAR(1 g) =2.09 W/kg; SAR(10 g) = 1.41 W/kg
Maximum value of SAR (measured) = 2.42 W/kg

        dB
         0


        —1.80


        —3.60


        —5.40


        —1.20


        —9.00
                    0 dB = 2.42 W/kg = 3.84 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/25


System Check_Body_750MHz

DUT: D750V3—1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: MSL_750_190225 Medium parameters used: f = 750 MHz; 0 = 0.972 S/m; g, = 54.183; p = 1000
kg/m3
Ambient Temperature : 23.1 °C ; Liquid Temperature : 22.1 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(6.3, 6.3, 6.3) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.63 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 51.03 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 3.16 W/kg
SAR(1 g) =2.22 W/kg; SAR(10 g) = 1.51 W/kg
Maximum value of SAR (measured) = 2.56 W/kg

        dB
         0


        —1.87


        —3.74


        —5.60


        —1.A47


        —9.34
                    0 dB = 2.56 W/kg = 4.08 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/28


System Check_Head_835MHz

DUT: D835V2—499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL_835_190228 Medium parameters used: £= 835 MHz; 0 = 0.882 S/m; e, = 42.033; p = 1000
kg/m>
Ambient Temperature : 23.5 °C; Liquid Temperature : 22.5 °C

DASYS Configuration
— Probe: EX3DV4 — SN3931:ConvF(10.07, 10.07, 10.07) ;Calibrated: 2018/9/27
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1399; Calibrated: 2018/11/16
— Phantom: Twin—SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD; Serial: 1885
— Measurement SW: DASYS52, Version 52.10 (2), SEMCAD X Version 14.6.12 (7450)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.07 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurementgrid: dx=8mm, dy=&mm, dz=Smm
Reference Value = 60.89 V/m; Power Drift =—0.11 dB
Peak SAR (extrapolated) = 3.43 W/kg
SAR(1 g) = 2.29 W/kg; SAR(10 g) = 1.51 W/kg
Maximum value of SAR (measured) = 3.05 Whkg

        dB
        0




        —2.50



        —5.20




        —7.80




        —10.40

                        22

        —13.00
                        C                        &


                    0 dB = 3.05 W/kg = 4.84 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/26


System Check_Body_835MHz

DUT: D835V2—499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: MSL_8$50_190226 Medium parameters used: f = 835 MHz; 0 = 0.952 S/m; g, = 55.142; p = 1000
kg/m3
Ambient Temperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(6.19, 6.19, 6.19) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.75 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 56.59 V/m; Power Drift = —0.12 dB
Peak SAR (extrapolated) = 3.50 W/kg
SAR(1 g) =2.38 W/kg; SAR(10 g) = 1.57 W/kg
Maximum value of SAR (measured) = 2.77 W/kg

        dB
         0


        —2.08


        —4.16


        —6.241


        —8.32


        —10.40
                    0 dB = 2.77 W/kg = 4.42 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/27


System Check_Head_1750MHz

DUT: D1750V2—1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: HSL_1750_190227 Medium parameters used: f = 1750 MHz; 0 = 1.412 S/m; g, = 41.848; p = 1000
kg/m3
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(5.48, 5.48, 5.48) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.0 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 93.80 V/m; Power Drift = —0.10 dB
Peak SAR (extrapolated) = 16.7 W/kg
SAR(1 g) = 9.41 W/kg; SAR(10 g) = 5.02 W/kg
Maximum value of SAR (measured) = 11.7 W/kg

        dB
         0


        —3.52


        —£.04


        —10.57


        —14.09


        —17.61
                    0 dB = 11.7 W/kg = 10.68 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/25


System Check_Body_1750MHz

DUT: D1750V2—1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: MSL_1750_190225 Medium parameters used: f = 1750 MHz; 0 = 1.481 S/m; s = 55.17; p = 1000
kg/m3
Ambient Temperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(5.06, 5.06, 5.06) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Left; Type: QDOOOP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 11.7 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 90.73 V/m; Power Drift = 0.08 dB
Peak SAR (extrapolated) = 16.0 W/kg
SAR(1 g) =9.39 W/kg; SAR(10 g) = 5.1 W/kg
Maximum value of SAR (measured) = 11.6 W/kg

        dB
         0


        —3.20


        —6.40


        —9.60


        —12.80


        —16.00
                    0 dB = 11.6 W/kg = 10.64 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/27


System Check_Head_1900MHz

DUT: D1900V2—5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900_190227 Medium parameters used: f = 1900 MHz; 0 = 1.417 S/m; g, = 40.829; p = 1000
kg/m3
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(5.27, 5.27, 5.27) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 13.5 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 99.80 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 19.5 W/kg
SAR(1 g) =10.5 W/kg; SAR(10 g) = 5.42 W/kg
Maximum value of SAR (measured) = 13.1 W/kg

        dB
         0


        —3.88


        —1.16


        —11.64


        —15.52


        —19.40
                    0 dB = 13.1 W/ikg= 11.17 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/25


System Check_Body_1900MHz

DUT: D1900V2—5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900_190225 Medium parameters used: f = 1900 MHz; 0 = 1.512 S/m; s = 52.296; p =
1000 kg/m>
Ambient Temperature : 23.8 °C ; Liquid Temperature : 22.8 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(4.8, 4.8, 4.8) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Left; Type: QDOOOP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.0 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 92.32 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 16.8 W/kg
SAR(1 g) = 9.54 W/kg; SAR(10 g) = 5.02 W/kg
Maximum value of SAR (measured) = 12.1 W/kg

        dB
         0


        —3.40


        —6.80


        —10.20


        —13.60


        —17.00
                    0 dB = 12.1 W/ikg = 10.83 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/8


System Check_Head_2450MHz

DUT: D2450V2—736

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: HSL_2450_190308 Medium parameters used : £= 2450 MHz; 0 = 1.788 S/m; s = 39.537; p =
1000 kg/m>
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(4.69, 4.69, 4.69) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 18.2 W/kg

Pin=250mwW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 103.7 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 27.9 W/kg
SAR(1 g) =13.2 W/kg; SAR(10 g) = 6.07 W/kg
Maximum value of SAR (measured) = 17.2 W/kg

        dB
         0


        —4.00


        —8.00


        —12.00


        —16.00


        —20.00
                    0 dB = 17.2 W/kg = 12.36 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/7

System Check_Body_2450MHz

DUT: D2450V2-736

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: MSL_2450_190307 Medium parameters used: f = 2450 MHz; σ = 2.008 S/m; εr = 53.308;
ρ = 1000 kg/m3
                    23.5                     22.5

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(7.53, 7.53, 7.53) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 20.6 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.5 V/m; Power Drift = -0.06 dB
Peak SAR (extrapolated) = 25.2 W/kg
SAR(1 g) = 12.4 W/kg; SAR(10 g) = 5.73 W/kg
Maximum value of SAR (measured) = 20.6 W/kg




               0 dB = 20.6 W/kg = 13.14 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/3


System Check_Head_2600MHz

DUT: D2600V2—1008

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: HSL_2600_190303 Medium parameters used: f = 2600 MHz; 0 = 1.983 S/m; g, = 39.038; p = 1000
kg/m3
Ambient Temperature : 23.7 °C ; Liquid Temperature : 22.7 °C

DASYS Configuration
— Probe: EX3DV4 — SN3976;ConvF(7.54, 7.54, 7.54) ;Calibrated: 2019/1/29
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn778; Calibrated: 2018/5/25
— Phantom: SAM_Left; Type: QDO00O0P40CD ; Serial: S/N:1801
— Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 25.4 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 117.5 V/m; Power Drift = —0.08 dB
Peak SAR (extrapolated) = 31.1 W/kg
SAR(1 g) =14.7 W/kg; SAR(10 g) = 6.58 W/kg
Maximum value of SAR (measured) = 24.7 W/kg

        dB
         U


        —4.00


        —8.00


        —12.00


        —16.00

                       22

        20.00         &
                    0 dB = 24.7 W/ikg = 13.93 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/2/26


System Check_Body_2600MHz

DUT: D2600V2—1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: MSL_2600_190226 Medium parameters used: f = 2600 MHz; 0 = 2.207 S/m; 8. = 53.152; p =
1000 kg/m>
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(4.27, 4.27, 4.27) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Left; Type: QDOOOP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 20.4 W/kg

Pin=250mwW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 96.13 V/m; Power Drift = 0.09 dB
Peak SAR (extrapolated) = 35.3 W/kg
SAR(1 g) =15.2 W/kg; SAR(10 g) = 6.69 W/kg
Maximum value of SAR (measured) = 20.4 W/kg

        dB
         0


        —4.00


        —8.00


        —12.00


        —16.00


        —20.00        (
                    0 dB = 20.4 W/kg = 13.10 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/34


System Check_Head_5250MHz

DUT: DSGHzV2—1006

Communication System: CW ; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190304 Medium parameters used: £= 5250 MHz; a = 4.623 S/m; & = 36.874; p
=1000 kg/m>
Ambicnt Temperature : 23.5 °C; Liquid Temperature ! 22.5 °C

DASY3 Configuration:
— Probe: EX3DV4 — SN7515; ConvF(5.45, 5.45, 5.45) ; Calibrated: 2018/10/3
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 $n918; Calibrated: 2018/6/20
— Phantom: SAM—Right; Type: SAM; Serial: TP—1503
— Measurement SW: DASY52, Version 52.10 (1),SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.7 W/kzg

Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=tmm, dz=1.4mm
Reference Value=77.19 V/m; Power Drift=—0.12 dB
Peak SAR (extrapolated) =32.9 W/kg
SAR( g) = 8.3 W/kg; SAR(10 g) =2.38 W/kg
Maximum value of SAR (measured)= 20.6 W/kg

   dB
    0


    =7.01


    —14.02


    —21.04


    —20.05


    —35.06
               0 dB =20.6 W/kg = 13.14 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/4

System Check_Body_5250MHz

DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: MSL_5G_190304 Medium parameters used: f = 5250 MHz; σ = 5.483 S/m; εr = 47.303; ρ
= 1000 kg/m3
                    23.3                     22.3

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(4.96, 4.96, 4.96) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.8 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 65.50 V/m; Power Drift = 0.09 dB
Peak SAR (extrapolated) = 31.1 W/kg
SAR(1 g) = 8.02 W/kg; SAR(10 g) = 2.29 W/kg
Maximum value of SAR (measured) = 20.3 W/kg




               0 dB = 20.3 W/kg = 13.07 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3M


System Check_Head_5600MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190304 Medium parameters used: £= 5600 MHz; a = 4.981 S/m; e = 36.374; p
=1000 kg/m>
Ambicnt Temperature : 23.5 °C; Liquid Temperature ! 22.5 °C

DASY3 Configuration:
— Probe: EX3DV4 — SN7515; ConvF(4.83, 4.83, 4.83) ; Calibrated: 2018/10/3
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 $n918; Calibrated: 2018/6/20
— Phantom: SAM—Right; Type: SAM; Serial: TP—1503
— Measurement SW: DASY52, Version 52.10 (1),SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 21.0 W/kzg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=tmm, dz=1.4mm
Reference Value=72.63 V/m; Power Drift=—0.19 dB
Peak SAR (extrapolated) =34.5 W/kg
SAR( g) = 8.39 W/kg; SAR(10 g) = 2.43 W/kg
Maximum value of SAR (measured)= 21.4 W/kg

   dB



    —6.48


    "12.97


    ~19.45


    —25.94


    —32.42
               0 dB =21.4 W/kg =13.30 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/4

System Check_Body_5600MHz

DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: MSL_5G_190304 Medium parameters used: f = 5600 MHz; σ = 5.939 S/m; εr = 46.661; ρ
= 1000 kg/m3
                    23.3                     22.3

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(4.39, 4.39, 4.39) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 19.8 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.79 V/m; Power Drift = -0.12 dB
Peak SAR (extrapolated) = 32.5 W/kg
SAR(1 g) = 8.05 W/kg; SAR(10 g) = 2.22 W/kg
Maximum value of SAR (measured) = 19.8 W/kg




               0 dB = 19.8 W/kg = 12.97 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3M


System Check_Head_5750MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190304 Medium parameters used: £= 5750 MHz; a = 5.139 S/m; & = 36.165; p
=1000 kg/m>
Ambicnt Temperature : 23.5 °C; Liquid Temperature ! 22.5 °C

DASY3 Configuration:
— Probe: EX3DV4 — SN7515; ConvF(4.95, 4.95, 4.95) ; Calibrated: 2018/10/3
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 $n918; Calibrated: 2018/6/20
— Phantom: SAM—Right; Type: SAM; Serial: TP—1503
— Measurement SW: DASY52, Version 52.10 (1),SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 19.4 W/zg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=tmm, dz=1.4mm
Reference Value = 69.89 V/m; Power Drift=—0.15 dB
Peak SAR (extrapolated) =35.9 W/kg
SAR( g) = 7.91 W/kg: SAR(1O g) = 218 W/kg
Maximum value of SAR (measured)= 20.2 W/kg

   dB
    0


    —8.03


    —16.06


    —24.08


    3211


    ~40.14
               0 dB =20.2 W/kg =13.05 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/6

System Check_Head_5750MHz

DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190306 Medium parameters used: f = 5750 MHz; σ = 5.092 S/m; εr = 36.854; ρ
= 1000 kg/m3
                    23.7                     22.7

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(4.95, 4.95, 4.95) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 19.6 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 68.99 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 35.4 W/kg
SAR(1 g) = 8.02 W/kg; SAR(10 g) = 2.23 W/kg
Maximum value of SAR (measured) = 20.3 W/kg




               0 dB = 20.3 W/kg = 13.07 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/4

System Check_Body_5750MHz

DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: MSL_5G_190304 Medium parameters used: f = 5750 MHz; σ = 6.141 S/m; εr = 46.392; ρ
= 1000 kg/m3
                    23.3                     22.3

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(4.42, 4.42, 4.42) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 18.6 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 63.76 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 33.8 W/kg
SAR(1 g) = 7.43 W/kg; SAR(10 g) = 2.02 W/kg
Maximum value of SAR (measured) = 19.5 W/kg




               0 dB = 19.5 W/kg = 12.90 dBW/kg



Document Created: 2019-04-09 08:25:43