RF Exposure Info._20190418_v1 - PY7-80422D_TestRpt_RFExp_Appendix A

FCC ID: PY7-80422D

RF Exposure Info

Download: PDF
FCCID_4255075

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/11

System Check_Head_750MHz

DUT: D750V3-1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: HSL_750_190311 Medium parameters used: f = 750 MHz; σ = 0.891 S/m; εr = 43.513; ρ
= 1000 kg/m3
Ambient Temperature:23.8 ℃; Liquid Temperature:22.8 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3124; ConvF(6.15, 6.15, 6.15) ; Calibrated: 2019/1/15
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn699; Calibrated: 2019/1/3
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1682
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.62 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 55.46 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 3.36 W/kg
SAR(1 g) = 2.27 W/kg; SAR(10 g) = 1.52 W/kg
Maximum value of SAR (measured) = 2.64 W/kg




               0 dB = 2.64 W/kg = 4.22 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/9

System Check_Body_750MHz

DUT: D750V3-1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: MSL_750_190309 Medium parameters used: f = 750 MHz; σ = 0.969 S/m; εr = 54.123; ρ
= 1000 kg/m3
Ambient Temperature:23.8 ℃; Liquid Temperature:22.8 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3124; ConvF(6.25, 6.25, 6.25) ; Calibrated: 2019/1/15
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn699; Calibrated: 2019/1/3
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1682
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.41 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 51.08 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 3.10 W/kg
SAR(1 g) = 2.1 W/kg; SAR(10 g) = 1.4 W/kg
Maximum value of SAR (measured) = 2.44 W/kg




               0 dB = 2.44 W/kg = 3.87 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/11

System Check_Head_835MHz

DUT: D835V2-499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL_850_190311 Medium parameters used: f = 835 MHz; σ = 0.9 S/m; εr = 42.398; ρ =
1000 kg/m3
Ambient Temperature:23.8 ℃; Liquid Temperature:22.8 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3124; ConvF(5.97, 5.97, 5.97) ; Calibrated: 2019/1/15
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn699; Calibrated: 2019/1/3
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1682
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.00 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.82 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 3.79 W/kg
SAR(1 g) = 2.53 W/kg; SAR(10 g) = 1.67 W/kg
Maximum value of SAR (measured) = 2.94 W/kg




               0 dB = 2.94 W/kg = 4.68 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/9

System Check_Body_835MHz

DUT: D835V2-499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: MSL_850_190309 Medium parameters used: f = 835 MHz; σ = 0.981 S/m; εr = 57.574; ρ
= 1000 kg/m3
Ambient Temperature:23.8 ℃; Liquid Temperature:22.8 ℃

DASY5 Configuration:
- Probe: ES3DV3 - SN3124; ConvF(6.07, 6.07, 6.07) ; Calibrated: 2019/1/15
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn699; Calibrated: 2019/1/3
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1682
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 3.15 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 60.48 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 3.92 W/kg
SAR(1 g) = 2.44 W/kg; SAR(10 g) = 1.59 W/kg
Maximum value of SAR (measured) = 3.15 W/kg




               0 dB = 3.15 W/kg = 4.98 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/6


System Check_Head_1750MHz

DUT: D1750V2—1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: HSL_1750_190306 Medium parameters used: f = 1750 MHz; 0 = 1.409 S/m; &, = 40.982; p = 1000
kg/m3
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(5.48, 5.48, 5.48) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Left; Type: QDOOOP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 11.9 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 93.80 V/m;, Power Drift = —0.10 dB
Peak SAR (extrapolated) = 16.6 W/kg
SAR(1 g) =9.38 W/kg; SAR(10 g) = 5.01 W/kg
Maximum value of SAR (measured) = 11.6 W/kg

        dB
         0


        —3.00


        —6.00


        —9.00


        —12.00


        —15.00
                    0 dB = 11.6 W/kg = 10.64 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/8

System Check_Body_1750MHz

DUT: D1750V2-1068

Communication System: CW ; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: MSL_1750_190308 Medium parameters used: f = 1750 MHz; σ = 1.484 S/m; εr = 55.128;
ρ = 1000 kg/m3
                    23.6                      22.6

DASY5 Configuration:
- Probe: EX3DV4 - SN3728; ConvF(7.83, 7.83, 7.83) ; Calibrated: 2019/1/15
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn316; Calibrated: 2019/1/3
- Phantom: SAM_Right; Type: SAM; Serial: TP:1681
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.4 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 92.22 V/m; Power Drift = -0.07 dB
Peak SAR (extrapolated) = 15.1 W/kg
SAR(1 g) = 8.75 W/kg; SAR(10 g) = 4.76 W/kg
Maximum value of SAR (measured) = 12.1 W/kg




               0 dB = 12.1 W/kg = 10.83 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/6


System Check_Head_1900MHz

DUT: D1900V2—5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900_190306 Medium parameters used: f = 1900 MHz; 0 = 1.458 S/m; g, = 41.063; p = 1000
kg/m3
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(5.27, 5.27, 5.27) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Left; Type: QDOOOP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 13.7 W/kg

Pin=250mwW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 99.09 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 20.0 W/kg
SAR(1 g) =10.7 W/kg; SAR(10 g) = 5.46 W/kg
Maximum value of SAR (measured) = 13.6 W/kg

        dB
         0


        —4.00


        —8.00


        —12.00


        —16.00


        —20.00
                    0 dB = 13.6 W/ikg = 11.34 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/8

System Check_Body_1900MHz

DUT: D1900V2-5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900_190308 Medium parameters used: f = 1900 MHz; σ = 1.567 S/m; εr = 53.492;
ρ = 1000 kg/m3
                    23.6                      22.6

DASY5 Configuration:
- Probe: EX3DV4 - SN3728; ConvF(7.54, 7.54, 7.54) ; Calibrated: 2019/1/15
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn316; Calibrated: 2019/1/3
- Phantom: SAM_Right; Type: SAM; Serial: TP:1681
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 14.3 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 93.95 V/m; Power Drift = 0.12 dB
Peak SAR (extrapolated) = 18.1 W/kg
SAR(1 g) = 10.2 W/kg; SAR(10 g) = 5.31 W/kg
Maximum value of SAR (measured) = 14.2 W/kg




               0 dB = 14.2 W/kg = 11.52 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/13

System Check_Head_2450MHz

DUT: D2450V2-736

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: HSL_2450_190313 Medium parameters used: f = 2450 MHz; σ = 1.746 S/m; εr = 38.118;
ρ = 1000 kg/m3
                    23.5                     22.5

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(7.42, 7.42, 7.42) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 22.0 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 115.8 V/m; Power Drift = -0.12 dB
Peak SAR (extrapolated) = 26.6 W/kg
SAR(1 g) = 13.2 W/kg; SAR(10 g) = 6.22 W/kg
Maximum value of SAR (measured) = 21.7 W/kg




               0 dB = 21.7 W/kg = 13.36 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/13

System Check_Body_2450MHz

DUT: D2450V2-736

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: MSL_2450_190313 Medium parameters used: f = 2450 MHz; σ = 2.007 S/m; εr = 52.975;
ρ = 1000 kg/m3
                    23.5                     22.5

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(7.53, 7.53, 7.53) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 20.6 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.5 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 25.2 W/kg
SAR(1 g) = 12.4 W/kg; SAR(10 g) = 5.73 W/kg
Maximum value of SAR (measured) = 20.6 W/kg




               0 dB = 20.6 W/kg = 13.14 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/7


System Check_Head_2600MHz

DUT: D2600V2—1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: HSL_2600_190307 Medium parameters used: f = 2600 MHz; 0 = 1.967 S/m; g, = 39.225; p = 1000
kg/m3
Ambient Temperature : 23.7 °C ; Liquid Temperature : 22.7 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(4.5, 4.5, 4.5) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1326; Calibrated: 2018/9/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=250mwW/Area Scan (81x81x1); Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 17.9 W/kg

Pin=250mwW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 97.71 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 28.8 W/kg
SAR(1 g) = 13.3 W/kg; SAR(10 g) = 6.01 W/kg
Maximum value of SAR (measured) = 17.8 W/kg

        dB
         0


        —4.64


        —9.29


        —13.93


        —18.58


        —23.22
                    0 dB = 17.8 W/kg = 12.50 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/27


System Check_Body_2600MHz

DUT: D2600V2—1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: MSL_2600_190327 Medium parameters used: £= 2600 MHz; a = 2.149 S/m; e = 51.589;
p = 1000 kg/m?
Ambicnt Temperature : 23.6 °C; Liquid Temperature ! 22.6 °C

DASY3 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(4.27, 4.27, 4.27) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n1326; Calibrated: 2018/9/18
— Phantom: SAM_Left; Type: QD00OP4OCD; Serial: TP:1431
— Measurement SW: DASY52, Version 52.10 (2).SEMCAD X Version 14.6.12 (7450)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 20.5 W/kzg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 89.68 V/m; Power Drift=—0.12 dB
Peak SAR (extrapolated) =31.1 W/kg
SAR( g) = 13.8 W/kg:; SAR(1O g) = 617 W/kg
Maximum value of SAR (measured)= 17.9 W/kg

 dB
[I


    ~4.73


    —9.46


    14.18


    "18.91


    —23.64       iss
               0 dB =17.9 W/kg =12.53 dBWkg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/9


System Check_Head_5250MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190309 Medium parameters used : £= 5250 MHz; a =4.517 S/m; e = 36.791; p
=1000 kg/m>
Ambicnt Temperature : 23.1 °C; Liquid Temperature ! 22.1 °C

DASY3 Configuration:
— Probe: EX3DV4 — SN7515; ConvF(5.45, 5.45, 5.45) ; Calibrated: 2018/10/3
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 $n918; Calibrated: 2018/6/20
— Phantom: SAM_Left; Type: QD00OP4OCD; Serial: TP:1431
— Measurement SW: DASY52, Version 52.10 (1),SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 17.4 W/zg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=tmm, dz=1.4mm
Reference Value=69.13 V/m; Power Drift=0.07 dB
Peak SAR (extrapolated) =29.5 W/kg
SAR( g) = 7.48 W/kg: SAR(O g) = 2.1 W/kg
Maximum value of SAR (measured)= 18.3 W/kg




    <soo         ~
               0 dB = 18.3 W/kg = 12.62 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/11


System Check_Body_5250MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: MSL_5G_190311 Medium parameters used: £= 5250 MHz; a = 5.45 8/m; & =47.192; p=
1000 kg/m*
Ambicnt Temperature : 23.7 °C; Liquid Temperature ! 22.7 °C

DASY3 Configuration:
— Probe: EX3DV4 — SN7515; ConvF(4.96, 4.96, 4.96) ; Calibrated: 2018/10/3
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 $n918; Calibrated: 2018/6/20
— Phantom: SAM—Right; Type: SAM; Serial: TP—1503
— Measurement SW: DASY52, Version 52.10 (1),SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.7 W/kzg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=tmm, dz=1.4mm
Reference Value = 65.50 V/m; Power Drift= 0.07 dB
Peak SAR (extrapolated) =30.9 W/kg
SAR( g) = 7.98 W/kg:; SAR(10 g) = 2.28 W/kg
Maximum value of SAR (measured)= 20.1 W/kg




                 &
               0 dB =20.1 W/kg =13.03 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/9


System Check_Head_5600MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190309 Medium parameters used: £= 5600 MHz; a = 4.853 S/m; & = 36.279; p
=1000 kg/m>
Ambicnt Temperature : 23.1 °C; Liquid Temperature ! 22.1 °C

DASY3 Configuration:
— Probe: EX3DV4 — SN7515; ConvF(4.83, 4.83, 4.83) ; Calibrated: 2018/10/3
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 $n918; Calibrated: 2018/6/20
— Phantom: SAM_Left; Type: QD00OP4OCD; Serial: TP:1431
— Measurement SW: DASY52, Version 52.10 (1),SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.3 W/kzg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=tmm, dz=1.4mm
Reference Value=72.30 V/m; Power Drift=0.05 dB
Peak SAR (extrapolated) =35.7 W/kg
SAR( g) = 8.37 W/kg:; SAR(1O g) = 2.32 W/kg
Maximum value of SAR (measured)= 21.1 W/kg




    <ensa        S
               0 dB =21.1 W/kg =13.24 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/11

System Check_Body_5600MHz

DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: MSL_5G_190311 Medium parameters used: f = 5600 MHz; σ = 5.905 S/m; εr = 46.55; ρ =
1000 kg/m3
                    23.7                     22.7

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(4.39, 4.39, 4.39) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 19.7 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 67.79 V/m; Power Drift = -0.15 dB
Peak SAR (extrapolated) = 32.3 W/kg
SAR(1 g) = 8.01 W/kg; SAR(10 g) = 2.21 W/kg
Maximum value of SAR (measured) = 19.7 W/kg




               0 dB = 19.7 W/kg = 12.94 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/9

System Check_Head_5750MHz

DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190309 Medium parameters used: f = 5750 MHz; σ = 5.015 S/m; εr = 36.059; ρ
= 1000 kg/m3
                    23.1                     22.1

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(4.95, 4.95, 4.95) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 21.6 W/kg

Pin=100mW/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 72.62 V/m; Power Drift = -0.05 dB
Peak SAR (extrapolated) = 37.9 W/kg
SAR(1 g) = 8.58 W/kg; SAR(10 g) = 2.44 W/kg
Maximum value of SAR (measured) = 22.1 W/kg




               0 dB = 22.1 W/kg = 13.44 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/3/11

System Check_Body_5750MHz

DUT: D5GHzV2-1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: MSL_5G_190311 Medium parameters used: f = 5750 MHz; σ = 6.105 S/m; εr = 46.281; ρ
= 1000 kg/m3
                    23.7                     22.7

DASY5 Configuration:
- Probe: EX3DV4 - SN7515; ConvF(4.42, 4.42, 4.42) ; Calibrated: 2018/10/3
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn918; Calibrated: 2018/6/20
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 18.5 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 63.76 V/m; Power Drift = 0.11 dB
Peak SAR (extrapolated) = 33.7 W/kg
SAR(1 g) = 7.39 W/kg; SAR(10 g) = 2 W/kg
Maximum value of SAR (measured) = 19.3 W/kg




               0 dB = 19.3 W/kg = 12.86 dBW/kg