SAR test report 4 of 4

FCC ID: PY7-61362R

RF Exposure Info

Download: PDF
FCCID_4249439

Calibration Laboratory of                                          3
                                                                    \\\\\'_//”/4                           S      Schweizerischer Kalibrierdienst
          :                                                       $              *
Schmid & Partner                                                 m                                         c      Service suisse d‘étalonnage
  Englneerlng AG                                                 ¥ /R >                                           Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      4,4[///-—\\\\\3                          S      Swiss Calibration Service
                                                                       thilabs

Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client        Sporton                                                                           Certificate No: D1750V2—1068_Nov18

ICALIBRATION CERTIFICATE                                                                                                                            |
 Object                               D1750V2 — SN:1068


 Calibration procedure(s)             QA CAL—O5.v10
                                      Calibration procedure for dipole validation kits above 700 MHz




 Calibration date:                    November 19, 2018



 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                      Apr—19
 Power sensor NRP—Z91                   SN: 103244                04—Apr—18 (No. 217—02672)                            Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02678)                            Apr—19
 Reference 20 dB Attenuator             SN: 5058 (20k)            04—Apr—18 (No. 217—02682)                            Apr—19
 Type—N mismatch combination            SN: 5047.2 / 06327        04—Apr—18 (No. 217—02683)                            Apr—19
 Reference Probe EX3DV4                 SN: 7349                  30—Dec—17 (No. EX3—7349_Dec17)                       Dec—18
 DAE4                                   SN: 601                   04—Oct—18 (No. DAE4—601_Oct18)                       Oct—19

 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—18)                    In house check: Oct—20
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—18)                    In house check: Oct—20
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—18)                    In house check: Oct—20
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—18)                    In house check: Oct—20
 Network Analyzer Agilent E8358A        SN: US41080477            31—Mar—14 (in house check Oct—18)                    In house check: Oct—19


                                        Name                                     Function                              Signature
 Calibrated by:                        Manu Seitz                                Laboratory Technician                             f



 Approved by:                          Katia Pokovic                             Technical Manager             // /g



                                                                                                                      Issued: November 20, 2018
 This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.


Certificate No: D1750V2—1068_Nov18                                     Page 1 of 8


Calibration Laboratory of                                                     S   Schweizerischer Kalibrierdienst
Schmid & Partner                                                                  Service suisse d'etalonnage
 Engineering AG                                                               C   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                  s   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.:   SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                         tissue simulating liquid
ConvF                       sensitivity in TSL / NORM x,y,z
N/A                         not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528-2013, "IEEE Recommended Practice for Determining the Peak Spatial-
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", June 2013
   b) IEC 62209-1, "Measurement procedure for the assessment of Specific Absorption Rate
      (SAR) from hand-held and body-mounted devices used next to the ear (frequency range of
      300 MHz to 6 GHz)", July 2016
   c) IEC 62209-2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
      communication devices used in close proximity to the human body (frequency range of 30
      MHz to 6 GHz)", March 2010
   d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
  e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
  • Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
  • Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
  • Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required .
  • Electrical Delay: One-way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
  • SAR measured: SAR measured at the stated antenna input power.
  • SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
  • SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.
 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.




Certificate No: D1750V2-1068_Nov18                              Page 2 of 8


Measurement Conditions
    DASY sys em con f'1guraf10n, as ar as not 1ven
                                              .    on page
      DASY Version                                              DASY5                               V52.10.2
      Extrapolation                                    Advanced Extrapolation
      Phantom                                          Modular Flat Phantom
      Distance Dipole Center • TSL                              10mm                              with Spacer
      Zoom Scan Resolution                               dx, dy, dz = 5 mm
      Frequency                                          1750 MHz± 1 MHz



Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity         Conductivity
      Nominal Head TSL parameters                                  22.0 °C           40.1               1.37 mho/m
      Measured Head TSL parameters                              (22.0 ± 0.2) °C   39.6 ±6 %          1.34 mho/m ± 6 %
      Head TSL temperature change during test                      < 0.5 °C           ----                     ----

SAR result with Head TSL

      SAR averaged over 1 cm3 (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                     9.17 W/kg
      SAR for nominal Head TSL parameters                     normalized to 1W        37.1 W/kg ± 17.0 % (k=2)


      SAR averaged over 1o cm3 (1 o g) of Head TSL               condition
      SAR measured                                        250 mW input power                     4.85 W/kg
      SAR for nominal Head TSL parameters                     normalized to 1W        19.5 W/kg ± 16.5 % (k=2)



Body TSL parameters
    The following parameters and calculations were applied
                                                                Temperature       Permittivity         Conductivity
      Nominal Body TSL parameters                                  22.0 °C           53.4               1.49 mho/m
      Measured Body TSL parameters                              (22.0 ± 0.2) °C   54.0±6 %           1.45 mho/m ±6 %
      Body TSL temperature change during test                      < 0.5 °C           ----                     ----
SAR result with Body TSL

      SAR averaged over 1 cm3 (1 g) of Body TSL                  Condition
      SAR measured                                        250 mW input power                     9.07 W/kg
      SAR for nominal Body TSL parameters                     normalized to 1W        37.0 W/kg ± 17.0 % (k=2)


      SAR averaged over 1 O cm 3 (1 O g) of Body TSL             condition
      SAR measured                                        250 mW input power                     4.84 W/kg

      SAR for nominal Body TSL parameters                     normalized to 1W        19.6 W/kg ± 16.5 % (k=2)




Certificate No: D1750V2-1068_Nov18                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           51.3 Q + 2.8 jQ
      Return Loss                                                                       - 30.3 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.1 Q + 3.6 jQ
      Return Loss                                                                       - 27.7 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.220 ns


After long term use with 1OOW radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short-circuited for DC-signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                June 15, 2010




Certificate No: D1750V2-1068_Nov18                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 19.11.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 - SN:1068

Communication System: UID O - CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; cr = 1.34 Sim; Er= 39 .6; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)

DASY52 Configuration:

    •   Probe: EX3DV4- SN7349; ConvF(8.5, 8.5, 8.5)@ 1750 MHz; Calibrated: 30.12.2017

    •   Sensor-Surface: 1.4mm (Mechanical Surface Detection)

    •   Electronics: DAE4 Sn601; Calibrated: 04.10.2018

    •   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

    •   DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)


Dipole Calibration for Head Tissue/Pin=250 mW, d=lOmm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value= 108.6 V/m; Power Drift= -0.07 dB
Peak SAR (extrapolated)= 16.9 W/kg
SAR(l g) = 9.17 W/kg; SAR(lO g) = 4.85 W/kg
Maximum value of SAR (measured)= 14.2 W/kg

        dB
        0


        -4.00


        -8.00


        -12.00


        -16.00


        -20.00
                  0 dB= 14.2 W/kg = 11.52 dBW/kg




Certificate No: 01750V2-1068_Nov18             Page 5 of 8


Impedance Measurement Plot for Head TSL




          File     Yiew   Channel     Sweep      Calibration   Trace   Scale   Marker   System   Window   Help

                                                                                                          1 750000 GHz       51253 Q
                                                                                                              257 01 pH      2.8259 C
                                                                                                          1 750000 GHz     30.518 mU
                                                                                                                              64 489 °




                          Ch1Augs 20
                  Chi: Start 155000 GHa     ——                                                                            Stop 1.95000 GHz

            rUL,‘é,U                                                                                        50000   GHz   —39 309 dB
            0.00

                 00

             10.00

             15.00
            20.00

            25.00
                  o0

                  10
                 0.00     Ch 1 Avg=   |20
                  Thi: Start 1 55000 oHz                                                                                  Stop 1.95000 GHz


            Status        CH 1:                                                         Avg=20 Delay                           LCL




Certificate No: D1750V2—1068_Nov18                                     Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                 Date: 19.11.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D17SOV2; Serial: D17SOV2 - SN:1068

Communication System: UID O - CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; cr = 1.45 Sim; er = 54; p = 1000 kg/m 3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)

DASY52 Configuration:

   •   Probe: EX3DV4- SN7349; ConvF(8.35, 8.35, 8.35)@ 1750 MHz; Calibrated: 30.12.2017

   •   Sensor-Surface: 1.4mm (Mechanical Surface Detection)

   •   Electronics: DAE4 Sn601; Calibrated: 04.10.2018

   •   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   •   DASY52 52.10.2(1495); SEMCAD X 14.6.12(7450)


Dipole Calibration for Body Tissue/Pin=250 mW, d=lOmm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value= 102.6 V/m; Power Drift= -0.07 dB
Peak SAR (extrapolated)= 16.0 W/kg
SAR(l g) = 9.07 W/kg; SAR(lO g) = 4.84 W/kg
Maximum value of SAR (measured)= 13.7 W/kg

                           dB
                            0


                            -4.00


                            -8.00


                            -12.00


                            -16.00


                            -20.00

                                     0 dB= 13.7 W/kg = 11.37 dBW/kg




Certificate No: 01750V2-1068_Nov18                 Page 7 of 8


Impedance Measurement Plot for Body TSL




          File     View    Channel     Sweep     Calibration   Trace       Scale Marker   System    Window   Help

                                                                                —*Lz                         1750000 GHz          48074 0
                                                                                     x*"A           #a              32543 pH      35783 0
                                                                                                    2.       1 250000 GHz       41 409 mU
                                                                                                                                    116.20 °




                          Ch1Aug= 20
                  Ch1: Start 1.55000 GHa    ——                                                                                 Stop 1.95000 GHz


             oo
            5.00
                                                                            _                                  s0000 Clz        __—2] 658 d6
            0.00

                 10
             10.00

             15.00
            20.00

                  00
            20.00

                 5.00
                  o         Chiug=
                   Ch1: Start 1.55000 GHz   ——                                                                                 Stop 1.95000 Ghs

                           CH 1:                               C* 1—Port                    Avg=20 Delay                            LCL




Certificate No: D1750V2—1068_Nov18                                     Page 8 of 8


                                    @
                  4me
                                                        £    :
                                                                                                                       PEA
                                                                                           ufhr
                                         in Collaborati      ith                         »lUl   n
                                           T     en                                    ~\“\\\-_L"'/,
                  a=‘/"/"J, s p_ e a g
                               NY _art‘*\
                  iz mronnsase iPeeC NAS 2t
             Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                                                                                       O
                                                                                       %,//—~Y v CALIBRATION




                                                                                 \\\
             Tel: +86—10—62304633—2079         Fax: +86—10—62304633—2504                Arlilubs®®                     CNAS LO570
             E—mail: cttl@chinattl.com         http:/www.chinattl.en

         Client            Sporton                                            Certificate No:          218—60324

CALIBRATION CERTIFICATE

Object                                     D1900V2 — SN: 54041

Calibration Procedure(s)                   FF—211—003—01

                                           Calibration Procedures for dipole validation kits

Calibration date:                          September 11, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S!). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(2243)‘C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                 ID #                 Cal Date(Calibrated by, Certificate No.)           Scheduled Calibration
 Power Meter        NRVD          102083              01—Nov—17 (CTTL, No.J17X08756)                               Oct—18
 Power sensor       NRV—Z5        100542              01—Nov—17 (CTTL, No.J17X08756)                               Oct—18
 Reference Probe EX3DV4           SN 7464             12—Sep—17(SPEAG,No.EX3—7464_Sep17)                           Sep—18
 DAE4                             SN 1524             13—Sep—17(SPEAG,No.DAE4—1524_Sep17)                          Sep—18


 Secondary Standards              ID #                Cal Date(Calibrated by, Certificate No.)            Scheduled Calibration
 Signal Generator E4438C          MY49071430           23—Jan—18 (CTTL, No.J18X00560)                              Jan—19
 NetworkAnalyzer E5071C           MY46110673           24—Jan—18 (CTTL, No.J18X00561)                              Jan—19



                                   Name                            Function                                   Signature
Calibrated by:                   Zhao Jing                   SAR Test Engineer                             \*\52‘ es

Reviewed by:                     Lin Jun                    SAR Test Engineer                                  —VFL},

Approved by:                     Qi Dianyuan                 SAR Project Leader                            ~C34 L.
                                                                                 Issued: September 15, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z18—60324                                Page 1 of 8


           _'\@’                    In Collaboration with
          _‘/"/‘J, a
           _\/                      CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      http:/www.chinattl.en


lossary:
TSL                       tissue simulating liquid
ConvF                     sensitivity in TSL / NORMx,y,z
N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
   Communications Devices: Measurement Techniques", June 2013
b) EC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
     SAR measured: SAR measured at the stated antenna input power.
     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.



   The reported uncertainty of measurement is stated as the standard uncertainty of
   Measurement multiplied by the coverage factor k=2, which for a normal distribution
   Corresponds to a coverage probability of approximately 95%.




Certificate No: Z18—60324                             Page 2 of 8


                                       In Coffeboretion with
                                       s     p       e         a g
                                       CALIBRATION LABORATOR'f

         Add: No.51 Xueyuan Road, Haidian District, Beijing, I 0019 1, China
         Tel: +86-10-62304633-2079      Fax: +86-10-62304633-2504
         E-mail: cttl@chinattl.com      http://www.chinattl. cn


Measurement Conditions
    DASY svst em confi1qurat"10n, as f ar as not q1ven
                                                  .    on paqe
      DASY Version                                          DASY52                                         52.10.1.1476

      Extrapolation                                Advanced Extrapolation

      Phantom                                      Triple Flat Phantom 5.1 C

      Distance Dipole Center - TSL                          10mm                                         with Spacer

      Zoom Scan Resolution                             dx, dy, dz= 5 mm

      Frequency                                        1900 MHz± 1 MHz


Head TSL parameters
   Thfll
    e o owmq                      i ta"ions were app r1ed .
          " paramet ers an d ca lcu
                                                               Temperature                Permittivity           Conductivity

      Nominal Head TSL parameters                                22.0    ·c                  40.0                1.40 mho/m

      Measured Head TSL parameters                         (22.0 ± 0.2)       0
                                                                                  c        40.4 ±6 %           1.44 mho/m ± 6 %

      Head TSL temperature change during test                    <1 .0   ·c                   ..---                    ----
SAR result with Head TSL
                         3
   SAR averaged over 1 cm (1             g) of Head TSL                       Condition

     SAR measured                                                      250 mW input power                    10.2 mW /g

     SAR for nominal Head TSL parameters                                 normalized to 1W             40.2 mW lg ± 18.8 % (k=2)

     SAR averaged over 10       cm 3   (10 g) of Head TSL                     Condition

     SAR measured                                                      250 mW input power                    5.35 mW/ g

     SAR for nominal Head TSL parameters                                 normalized to 1W             21 .2 mW /g ± 18.7 % (k=2)


Body TSL parameters
   Th e f o IIowmq
                . parameters an d ca Icu Iat1ons
                                            .    were aoo r1ed .
                                                               Temperature                Permittivity          Conductivity

     Nominal Body TSL parameters                                22.0     ·c                  53.3               1.52 mho/m

     Measured Body TSL parameters                          (22.0 ± 0.2)       0
                                                                                  c        53.3 ± 6 %         1.49 mho/m ± 6 %

     Body TSL temperature change during test                    <1.o     ·c                   ----                     ----
SAR resu It WI"th B 0 dIY TSL
                           3
   SAR averaged over 1 cm (1            g) of Body TSL                        Condition
     SAR measured                                                    250 mW input power                      9.94 mW/ g

     SAR for nominal Body TSL parameters                                 normalized to 1W             40.2 mW /g ± 18.8 % (k=2)

     SAR averaged over 10      cm 3    (10 g) of Body TSL                     Condition

     SAR measured                                                    250 mW input power                      5.35 mW/ g

     SAR for nominal Body TSL parameters                                 normalized to 1W             21.5 mW lg± 18.7 % (k=2)

Certificate No: Z 18-60324                               Page 3 of8


           _"\03                    in Collaboration with
          x/"/"J,
          _‘/
                  a
                  CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: ctti@chinattl.com       http:/www.chinattl.cn


  Appendix (Additional assessments outside the scope of CNAS LO570)

 Antenna Parameters with Head TSL


      Impedance, transformed to feed point                                  52.90+ 74310

      Return Loss                                                            —22.30B


 Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                  47.60+ 6.800

     Return Loss                                                             —22.70B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                        1.066 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data


     Manufactured by                                                            SPEAG




Certificate No: Z18—60324                             Page 4 of 8


           _A®                      in Collaboration with
          _‘/"/"J, a
          lizzyy,>~                 CALIERATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       http://www.chinattl.on


DASYS5 Validation Report for Head TSL                                   Date: 09.10.2018
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 504041
   Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
   Medium parameters used: £= 1900 MHz; 0 = 1.438 S/m; s = 40.37; p = 1000 kg/m3
   Phantom section: Center Section
DASYS Configuration:

            *    Probe: EX3DV4 — SN7464; ConvF(8.39, 8.39, 8.39) @ 1900 MHz; Calibrated:
                 9/12/2017
            e    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            e    Electronics: DAE4 $Sn1524; Calibrated: 9/13/2017
            &    Phantom: MFP_V5.1C ; Type: QD 000 P51CA; Serial: 1062
            e    Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11
                 (7439)

        System Performance Check/Zoom Scan (7x7x7) {7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 98.66 V/m; Power Drift =—0.02 dB
        Peak SAR (extrapolated) = 19.1 W/kg
        SAR(I g) = 10.2 W/kg; SAR(10 g) = 5.35 W/kg
        Maximum value of SAR (measured) = 15.7 W/kg




            —10.30


            —13.74


           1717
                         £—.

                      0 dB = 15.7 W/kg = 11.96 dBW/kg




Certificate No: Z18—60324                             Page 5 of 8


                                            In Collaboration with
                                            s     p       e         a g
                                        CALIBRATION LABORATORY

           Add: No.SI Xueyuan Road, 1-laidian District, Beijing, 10019 1, China
           Tel: +86-1 0-62304633-2079      Fax: +86- 10-62304633-2504
           E-mail: cttl@ chinattl.com      http://www.chinattl.cn




            Impedance Measurement Plot for Head TSL


         Trl Sll Log Mag 10. OOdsT Ref O. OOOdB [Fl]
          50. 00     >l   1.9000000 GHZ -22.250 dB
          40 . 00

          30.00
          20 . 00

          1 0 . 00

          0.000
         -1 0. 00

         -2 0. 00
         - 30. 00
          40. 00
         - 50. 00
     •11111  Sll sm1th (R+jx) scale 1.ooou [Fl oel]
         >l    1.9000000 GHZ     52.865 0     7.4264 0   622 . 0




    I1   Start 1.7 GHz                                             IFBW lOOHz     Stop 2.1 GHz   .j'


Certificate No: Z 18-60324                                    Page 6 of8


           ~                    ®     In CoRaboration with

          - TTL                       s      P e             a        g
           ...............           CALIBRATION LABORATORY

         Add: No.51 Xueyuan Road, Haidian District, Beijing, I 00191 , China
         Tel: +86-10-62304633-2079      Fax: +86-10-62304633-2504
         E-mail: cttl@chinattl.com      http://www.chinattl.cn




DASYS Validation Report for Body TSL                                      Date: 09.10.2018
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 - SN: 5d041
  Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1: 1
  Medium parameters used: f = 1900 MHz; cr = 1.493 S/m; €r = 53.34; p = 1000 kg/m3
  Phantom section: Right Section
DASY5 Configuration:

               •   Probe: EX3DV4- SN7464; ConvF(8.32, 8.32, 8.32) @ 1900 MHz; Calibrated:
                   9/12/2017
               •   Sensor-Surface: 1.4mm (Mechanical Surface Detection)
               •   Electronics: DAE4 Sn 1524; Calibrated: 9/13/2017
               •   Phantom: MFP_ V5.1C ; Type: QD 000 P51CA; Serial: 1062
               •   Measurement SW: DASY52, Version 52.10 (l); SEMCAD X Version 14.6.11
                   (7439))

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 94.03 V/m; Power Drift= -0.08 dB
        Peak SAR (extrapolated)= 17.5 W/kg
        SAR(l g) = 9.94 W/kg; SAR(lO g) = 5.35 W/kg
        Maximum value of SAR (measured)= 15.0 W/kg

            dB
               0


               -3.14


               -6.28


               -9.43


               -12.57


               -15.71
                          L.
                        0 dB= 15.0 W/kg = 11.76 dBW/kg




Certificate No: ZI 8-60324                              Page 7 of 8


                                             In Collaboration with
                                         s           p         e         a g
                                         CALIBRATION LABORATORY

           Add: No.SI Xueyuan Road, Haidian District, Beijing, 100191 , China
           Tel: +86-10-62304633-2079     Fax: +86-10-62304633-2504
           E-mail: cttl@chinattl. com     http://www.chinattl.cn


           Impedance Measurement Plot for Body TSL


     / Trl Sll Log Mag 1 0.00de/ Ref O.OOOdB- [ Fl]
          SO. OO >l      1.9000000 GHZ - 22.657           B
         40. 00

         30 . 00
         2 0 . 00

         10 . 00

         0. 000
         10.00

        -2 0 . 00
         30. 00
        -4 0. 00

        -5 0.00
     ~1111 Sll sm1th (R+jx) scale 1.ooou [Fl Del]
        >l    1 . 9000000 GHZ 47 . 621   Q     6 . 8033   Q   569. 8




     1 · Start 1.7 GHz                                                 IFBW 100 Hz   Stop 2.1 GHz . .-,




Certificate No: Zl 8-60324                                         Page 8 of8


                                                                                        ay lll“ 14                               *


                    43—© in Collaboration with                                       §i&‘\
                                                                                       \U M #EWUA
                    LL 1.B.t—i—A— PSCNAS::
                Add: No.51 Xueyuan Road, Haidian District, Beifing, 100191, China     :,/“[;\\“\\‘     v CNAS LO570
                                                                                                                             ErAWA
                                                                                                                            CALIBRATION
                Tel: +86—10—62304633—2512       Fax: +86—10—62304633—2504                 tlay

                E—mail: cttl@chinattl.com       Hitp://www.chinattl.en
          Client :       Sporton                                                    Certificate No: 218—60383

CALIBRATION CERTIFICATE

Object                                        DAE4 — SN: 1326

Calibration Procedure(s)                      FF—211—002—01

                                              Calibration Procedure for the Data Acquisition Electronics
                                              (DAEx)

Calibration date:                             September 18, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22#3)c and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                   ID #          Cal Date(Calibrated by, Certificate No.)           Scheduled Calibration



Process Calibrator 753          1971018                20—Jun—18 (CTTL, No.J18X05034)                            June—19




                                    Name                     Function                                   Signature
Calibrated by:                      Yu Zongying             SAR Test Engineer                                .     1{ 5
    ;           ;                                                                                     a it                 ~c~
Reviewed by:                        Lin Hao                 SAR Test Engineer                          ~ fi]’fi,%%          \ 4

Approved by:                        Qi Dianyuan             SAR Project Leader                           \“/’ /

                                                                                                 Issued: September 20, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




        Certificate No: Z18—60383                              Page 1 of 3


          WC( Collsboration with
          =‘/"[‘] a
          ies CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512     Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Hitp://www.chinattl.en

Glossary:
DAE                             data acquisition electronics
Connector angle                  information used in DASY system to align probe sensor X
                                to the robot coordinate system.


Methods Applied and Interpretation of Parameters:
e DC Voltage Measurement: Calibration Factor assessed for use in DASY
   system by comparison with a calibrated instrument traceable to national
   standards. The figure given corresponds to the full scale range of the
   voltmeter in the respective range.

«_   Connector angle: The angle of the connector is assessed measuring the
     angle mechanically by a tool inserted. Uncertainty is not required.

*    The report provide only calibration results for DAE, it does not contain other
     performance test results.




Certificate No: Z18—60383                              Page 2 of 3


           _A®                     In Collaboration with

          a="/"[‘], CALIBRATION
                    a
           uy                   LABORATORY

        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http:/www.chinattl.on



DC Voltage Measurement
    A/D — Converter Resolution nominal
        High Range:       1LSB =       6.1uV , full range =    —100... +300 mV
        Low Range:        1LSB =        61nV , full range =    AAuslll +3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors                    X                              Y                 Z
      High Range                  404.892 + 0.15% (k=2) 405.233 £0.15% (k=2)          404.616 + 0.15% (k=2)

      Low Range                   3.98868 + 0.7% (k=2)         3.99146 + 0.7% (k=2)   3.99194 + 0.7% (k=2)


Connector Angle

     Connector Angle to be used in DASY system                                             68041 °




Certificate No: Z18—60383                              Page 3 of 3


Calibration Laboratory of
Schmid & Partner                                                                                            s     Schweizerischer Kalibrierdienst
                                                                                                            C     Service suisse d'etalonnage
 Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                s     Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.:   $CS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      Sporton                                                                             Certificate No:   ES3-3169_May18

CALIBRATION CERTIFICATE                                                                                                                             I
Object                                ES3DV3 - SN:3169


Calibration procedure(s)              QA CAL-01 .v9, QA CAL-23.vS, QA CAL-25.v6
                                      Calibration procedure for dosimetric E-field probes



Calibration date:                     May 28, 2018

This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (SI).
The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature (22 ± 3)°C and humidity < 70%.


Calibration Equipment used (M&TE critical for cal ibration)



 Primary Standards                   ID                                 Cal Date (Certificate No.)                  Scheduled Calibration
 Power meter NRP                     SN: 104778                         04-Apr-18 (No. 2 17-02672/02673)            Apr-19
 Power sensor NRP-291                SN: 103244                         04-Apr-18 (No. 217-02672)                   Apr-19
 Power sensor NRP-291                SN: 103245                         04-Apr-18 (No. 217-02673)                   Apr-19
 Reference 20 dB Attenuator          SN: S5277 (20x)                   04-Apr-18 (No. 217-02682)                    Apr-19
 Reference Probe ES3DV2              SN: 3013                          30-Dec-17 (No. ES3-3013 Dec17)               Dec-1 8
 DAE4                                SN: 660                           2 1-Dec-17 {No. DAE4-660 Dec17)              Dec-18


 Secondary Standards                 ID                                Check Date {in house)                        Scheduled Check
 Power meter E4419B                  SN: GB41293874                    06-Apr-16 (in house check Jun-16)            In house check: Jun-18
 PowersensorE4412A                   SN : MY41498087                   06-Apr-16 (in house check Jun-16)            In house check: Jun-18
 PowersensorE4412A                   SN: 000110210                     06-Apr- 16 (in house check Jun-16)           In house check: Jun-1 8
 RF generator HP 8648C               SN: US3642U01700                  04-Aug-99 (in house check Jun-16)            In house check: Jun-18
 Network Analyzer HP 8753E           SN: US37390585                     18-0ct-01 (in house check Oct-17)           In house check: Oct-1 8

                                          Name                                Function
Calibrated by:                         Jeton Kastrati                         Laboratory Technician



Approved by:                           Katja Pokovic                          Technical Manager



                                                                                                                      Issued: May 29, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: ES3-3169_May18                                       Page 1 of 11


 Calibration Laboratory of
 Schmid & Partner                                                                         s    Schweizerischer Kalibrierdienst
                                                                                               Service suisse d'etalonnage
  Engineering AG                                                                          C
                                                                                              Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                              s   Swiss Calibration Service



 Accredited by the Swiss Accreditation Service {SAS)                                       Accreditation No.:   SCS 0108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
NORMx,y,z                    sensitivity in free space
ConvF                        sensitivity in TSL / NORMx,y,z
DCP                          diode compression point
CF                           crest factor (1/duty_cycle) of the RF signal
A, B, C, D                   modulation dependent linearization parameters
Polarization q>              q> rotation around probe axis
Polarization S               S rotation around an axis that is in the plane normal to probe axis (at measurement center),
                             i.e., S = O is normal to probe axis
Connector Angle              information used in DASY system to align probe sensor X to the robot coordinate system

Calibration is Performed According to the Following Standards:
     a)   IEEE Std 1528-2013, "IEEE Recommended Practice for Determining the Peak Spatial-Averaged Specific
          Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
          Techniques", June 2013
     b)   IEC 62209-1, ", "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from hand-
          held and body-mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)", July 2016
     c)   IEC 62209-2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
          used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March 2010
     d)   KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Methods Applied and Interpretation of Parameters:
    •     NORMx,y,z: Assessed for E-field polarization S = 0 (f s 900 MHz in TEM-cell; f > 1800 MHz: R22 waveguide).
           NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E 2-field
          uncertainty inside TSL (see below ConvF).
    •     NORM(f)x,y,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This linearization is
          implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
          in the stated uncertainty of ConvF.
    •     DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep with CW
          signal (no uncertainty required). DCP does not depend on frequency nor media.
    •     PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
          characteristics
    •     Ax,y,z; Bx,y,z; Cx,y,z; Dx,y,z; VRx,y,z: A, B, C, Dare numerical linearization parameters assessed based on
          the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
          media. VR is the maximum calibration range expressed in RMS voltage across the diode.
    •     ConvF and Boundary Effect Parameters: Assessed in flat phantom using E-field (or Temperature Transfer
          Standard for f s 800 MHz) and inside waveguide using analytical field distributions based on power
          measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
          boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
          used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
          to NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
          ConvF is used in DASY version 4.4 and higher which allows extending the validity from ± 50 MHz to± 100
          MHz.
    •     Spherical isotropy (30 deviation from isotropy) : in a field of low gradients realized using a flat phantom
          exposed by a patch antenna.
    •     Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
          (on probe axis). No tolerance required.
    •     Connector Angle: The angle is assessed using the information gained by determining the NORMx (no
          uncertainty required).



Certificate No: ES3-3169_May18                                  Page 2 of 11


ES3DV3-SN:3169                                                                  May 28, 2018




                      Probe ES3DV3

                                          SN:3169

                                  Manufactured:            October 8, 2008
                                  Calibrated:              May 28, 2018



                                 Calibrated for DASY/EASY Systems
                                    (Note: non-compatible with DASY2 system!)




Certificate No: ES3-3169_May18                    Page 3 of 11


ES3DV3-SN:3169                                                                                                                    May 28, 2018




DASY/EASY - Parameters of Probe: ES3DV3 - SN:3169
Basic Calibration Parameters
                                                SensorX                          SensorY                     SensorZ               Unc (k=2)
    Norm (uV/IV/m)"Y                               1.15                           1.15                          1.13               + 10.1 %
    DCP/mVl"                                       99.6                           99.3                          96.1

Modulation Calibration Parameters
    UID        Communication System Name                                    A            B           C           D         VR          Uncc
                                                                           dB        dBVµV                      dB         mV          (k=2)
    0          cw                                                 X        0.0          0.0         1.0        0.00       202.3       ±3.3 %
                                                                  y        0.0          0.0         1.0                   207.5
                                                                  z        0.0          0.0         1.0                   199.0



    The reported uncertainty of measurement is stated as the standard uncertainty of measurement
    multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
    probability of approximately 95%.


A The uncertainties of Norm X,Y,Z do not affect the E 2-field uncertainty inside TSL (see Pages 5 and 6).
8
  Numerical linearization parameter: uncertainty not required.
E Uncertainty is determined using the max. deviation from linear response applying rectangular distribution and is expressed for the square of the
field value.




Certificate No: ES3-3169_May18                                    Page 4 of 11


ES3DV3- SN:3169                                                                                                                May 28, 2018




DASY/EASY - Parameters of Probe: ES3DV3 - SN:3169

Calibration Parameter Determined in Head Tissue Simulating Media
                        Relative         Conductivity                                                               Depth u          Unc
      f !MHz\ c      PermittivitvF         /5/m\ F          ConvF X       ConvFY         ConvF Z       Aloha
                                                                                                                0
                                                                                                                     Imm)           (k=2)

        750               41.9                0.89             6.56          6.56          6.56          0.75        1.20         + 12.0 %

        835               41.5                0.90             6.39          6.39          6.39          0.80        1.13         + 12.0 %

        900               41.5                0.97             6.21          6.21          6.21          0.80        1.13         + 12.0 %

        1750              40.1                1.37            5.48           5.48          5.48          0.80        1.14         + 12.0 %

       1900               40.0                1.40            5.27           5.27          5.27         0.80         1.14         + 12.0 %

       2000               40.0                1.40            5.25           5.25          5.25         0.55         1.34         + 12.0 %

       2300               39.5                1.67            4.91           4.91          4.91         0.76         1.18         ± 12.0 %
       2450               39.2                1.80            4.69           4.69          4.69         0.76         1.25         + 12.0 %

       2600               39.0                1.96            4.50          4.50           4.50         0.70         1.30         ± 12.0 %


c Frequency validity above 300 MHz of± 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to± 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is± 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to ± 110 MHz.
FAt frequencies below 3 GHz, the validity of tissue parameters (E and o) can be relaxed to± 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and er) is restricted to± 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
G Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than ± 1% for frequencies below 3 GHz and below± 2% for frequencies between 3-6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: ES3-3169_May18                                   Page 5 of 11


ES3DV3- SN:3169                                                                                                                  May 28, 2018




DASY/EASY - Parameters of Probe: ES3DV3 - SN:3169

Calibration Parameter Determined in Body Tissue Simulating Media
                       Relative           Conductivity                                                                Depth"           Unc
      f /MHz) c      Permittivitv F         /5/m)F           ConvF X        ConvF Y       ConvF Z        Alpha 0       lmml            (k=2)

        750               55.5                0.96              6.30          6.30           6.30          0.33         1.78        + 12.0 %

        835               55.2                0.97              6.19          6.19           6.19         0.63          1.29        + 12.0 %

        1750              53.4                1.49              5.06          5.06           5.06         0.62          1.36        + 12.0 %

        1900              53.3                1.52              4.80          4.80           4.80         0.37          1.91        + 12.0 %

       2300               52.9                1.81              4.50          4.50           4.50         0.78          1.25        + 12.0 %

       2450               52.7                1.95             4.40           4.40           4.40         0.80          1.15        + 12.0 %

       2600               52.5                2.16             4.27           4.27           4.27         0.80          0.80        + 12.0 %


c Frequency validity above 300 MHz of± 100 MHz only applies for OASY v4.4 and higher (see Page 2), else it is restricted to± 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is± 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30. 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to ± 11 O MHz.
F At frequencies below 3 GHz, the validity of tissue parameters (c and cr) can be relaxed to± 10% if llquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (f: and c:r) is restricted to± 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
0
  Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than± 1% for frequencies below 3 GHz and below± 2% for frequencies between 3-6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: ES3-3169_May18                                    Page 6 of 11


ES3DV3— SN:3169                                                                                                               May 28, 2018

                                                          Frequency Response of E—Field
                                                               (TEM—Cell:ifi110 EXX, Waveguide: R22)
                                            5t
                                             ed
                                            =>
                                            1     TTTT
          Frequency response {normalized)




                                                                           i                                             T
                                                                         1000                                          3000
                                                                  —8]
                                                                  TEem
                                                         U ncertainty of Frequency Response of E—field: £ 6.3% (k=2)




Certificate No: ES3—3169_May18                                                  Page 7 of 11


ES3DV3— SN:3169                                                                                                                                 May 28, 2018




                                                 Receiving Pattern (¢), 8 = 0°

                                f=600 MHz,TEM                                                            f=1800 MHz,R22
                                             e                                                                      s


                                                                                                  1980          *       "%              —as




                   20                                          tsrer   f                   io       ®           *       C         ;   sg * on
                                                 |                                                          +




                         x5 .         Tsa            ‘ra        — us                              ms .          5       2e*            .815

                                             a                                                                  "#o 0
                   ®              &                        ®           e                    e               ®                 &                  6
                   Tot            X                        ¥           2                   Tot              X                 ¥                  2
      Error {dB]




                                                                  600 M~z                        1800 MHz                         2500 MHz


                                            Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Certificate No: ES3—3169_May18                                              Page 8 of 11


ES3DV3— SN:3169                                                                                                               May 28, 2018




                                          Dynamic Range f(SARneaq)
                                                        (TEM cell , feyai= 1900 MHz)



                         1064




                         10°




                   3
                   /     1044
                   o
                   E.
                   2
                   ®
                   5
                   [=3
                   £
                         103




                         1024



                                ~.            I                    T                T            juk—         T
                               108           102                   10              10           10‘          10         10°
                                                          _                   SAR [mW/cm3]
                                                        *
                                             not compensated                                     compensated




                               2


                               1
                         o
                         C
                         §     0

                         i
                             14


                             24—                    ‘         HP          i                1         ;         :
                                   10#        102                  1041              100       101       102       10
                                                                              SAR [mW/cm3]
                                                        _*
                                             not compensated                                   compensated


                                         Uncertainty of Linearity Assessment: + 0.6% (k=2)




Certificate No: ES3—3169_May18                                                Page 9 of 11


ESSDV3— SN:3169                                                                                                                                                                                    May 28, 2018



                                                         Conversion Factor Assessment
                           f= 835 MHz,WGLS R9 (H_convF)                                                                        f= 1900 MHz,WGLS R22 (H_convF)
                                                                                                              35 ;
         a0                                                                                                       I



                                                                                                                  f
                  —                                                                                               }
         30           \,                                                                                       s|

                                .                                                                           >..|
                           -‘                                                                                     I



                                 *                                                                          $"I                +4
                                                                                                            &0 |                    *
                                                                                                            % is                        ‘%


          o                                                                                                       |
                                                                                                              10 "                           °



                                                                                                                  1
         0s                                                                                                    5‘1

         is                               s                      f                 ruple aeguadht              §d     BPscny        uie Ti on              ie    ut aieg           ol           e reopitt
              0                      10              5      10                 x        35         a             o              5            i        15             x        25           20       3       40
                                                          z imm?                                                                                                     fmm]
                                     +                                         * }                                                           *                                             #
                                 arabtcal                                   measured                                                     anavice                                        measured



                                                     Deviation from Isotropy in Liquid
                                                                               Error (¢, 9), f = 900 MHz
                                         Deviation




                                               ~1.0      —08         —0.6      —0.4          —02       00      02               04               06             08          1.0
                                                     Uncertainty of Spherical Isotropy Assessment: £ 2.6% (k=2)




Certificate No: ES3—3169_May18                                                                Page 10 of 11


ES3DV3-SN:3169                                           May 28, 2018




DASY/EASY - Parameters of Probe: ES3DV3 - SN:3169

Other Probe Parameters
 Sensor Arrangement                                      Triangular
 Connector Angle (')                                          -24.7
 Mechanical Surface Detection Mode                         enabled
 Optical Surface Detection Mode                            disabled
 Probe Overall Length                                      337mm
 Probe Body Diameter                                         10mm
                                                               .


 Tip Length                                                 10 mm
 Tip Diameter                                                 4mm
 Probe Tip to Sensor X Calibration Point                      2mm
 Probe Tip to Sensor Y Calibration Point                      2mm
 Probe Tip to Sensor Z Calibration Point                      2mm
 Recommended Measurement Distance from Surface                3mm




Certificate No: ES3-3169_May18              Page11of11



Document Created: 2019-04-24 07:13:56
Document Modified: 2019-04-24 07:13:56

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC