RF Exposure Info._20190121_v1 - PY7-54264H_TestRpt_RFExp_Appendix B

FCC ID: PY7-54264H

RF Exposure Info

Download: PDF
FCCID_4156281

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/28

#01_GSM850_GPRS (4 Tx slots)_Left Cheek_Ch251


Communication System: GSM850; Frequency: 848.8 MHz;Duty Cycle: 1:2.08
Medium: HSL_850_181128 Medium parameters used: f = 849 MHz; σ = 0.938 S/m; εr = 42.852; ρ
= 1000 kg/m3
                    23.7                      22.7

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.39, 6.39, 6.39) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.459 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 21.99 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 0.553 W/kg
SAR(1 g) = 0.427 W/kg; SAR(10 g) = 0.317 W/kg
Maximum value of SAR (measured) = 0.467 W/kg




               0 dB = 0.467 W/kg = -3.31 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/27

#02_GSM1900_GPRS (4 Tx slots)_Right Cheek_Ch810


Communication System: PCS; Frequency: 1909.8 MHz;Duty Cycle: 1:2.08
Medium: HSL_1900_181127 Medium parameters used: f = 1910 MHz; σ = 1.446 S/m; εr = 38.615;
ρ = 1000 kg/m3
                    23.1                      22.1

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(5.27, 5.27, 5.27) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.599 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 19.47 V/m; Power Drift = 0.13 dB
Peak SAR (extrapolated) = 0.830 W/kg
SAR(1 g) = 0.508 W/kg; SAR(10 g) = 0.301 W/kg
Maximum value of SAR (measured) = 0.596 W/kg




               0 dB = 0.596 W/kg = -2.25 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/28

#03_WCDMA V_RMC 12.2Kbps_Left Cheek_Ch4233


Communication System: WCDMA ; Frequency: 846.6 MHz;Duty Cycle: 1:1
Medium: HSL_850_181128 Medium parameters used: f = 847 MHz; σ = 0.936 S/m; εr = 42.878; ρ
= 1000 kg/m3
                    23.7                     22.7

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.39, 6.39, 6.39) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.489 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 22.75 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 0.593 W/kg
SAR(1 g) = 0.444 W/kg; SAR(10 g) = 0.326 W/kg
Maximum value of SAR (measured) = 0.489 W/kg




               0 dB = 0.489 W/kg = -3.11 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/28

#04_LTE Band 5_10M_QPSK_1_0_Left Cheek_Ch20525


Communication System: LTE; Frequency: 836.5 MHz;Duty Cycle: 1:1
Medium: HSL_850_181128 Medium parameters used: f = 836.5 MHz; σ = 0.926 S/m; εr = 42.998;
ρ = 1000 kg/m3
                    23.7                      22.7

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.39, 6.39, 6.39) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.456 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 22.48 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 0.534 W/kg
SAR(1 g) = 0.420 W/kg; SAR(10 g) = 0.312 W/kg
Maximum value of SAR (measured) = 0.460 W/kg




               0 dB = 0.460 W/kg = -3.37 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/28

#05_LTE Band 12_10M_QPSK_1_0_Left Cheek_Ch23095


Communication System: LTE; Frequency: 707.5 MHz;Duty Cycle: 1:1
Medium: HSL_750_181128 Medium parameters used : f = 707.5 MHz; σ = 0.852 S/m; εr = 43.631;
ρ = 1000 kg/m3
                    23.7                      22.7

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.56, 6.56, 6.56) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.229 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 16.56 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 0.263 W/kg
SAR(1 g) = 0.216 W/kg; SAR(10 g) = 0.168 W/kg
Maximum value of SAR (measured) = 0.234 W/kg




               0 dB = 0.234 W/kg = -6.31 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#06_WLAN2.4GHz_802.11b 1Mbps_Left Cheek_Ch1


Communication System: 802.11b; Frequency: 2412 MHz;Duty Cycle: 1:1
Medium: HSL_2450_181129 Medium parameters used: f = 2412 MHz; σ = 1.727 S/m; εr = 39.044;
ρ = 1000 kg/m3
                    23.2                        22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(4.69, 4.69, 4.69) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.683 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 19.03 V/m; Power Drift = 0.11 dB
Peak SAR (extrapolated) = 1.21 W/kg
SAR(1 g) = 0.505 W/kg; SAR(10 g) = 0.223 W/kg
Maximum value of SAR (measured) = 0.705 W/kg




               0 dB = 0.705 W/kg = -1.52 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/30


#07_WLANSGHz_802.11n—HT40 MCSO_Left Cheek_Ch46

Communication System: 802.11n; Frequency : 5230 MHz;Duty Cycle: 1:1.046
Medium: HSL_5G_181130 Medium parameters used: £= 5230 MHz; 0 = 4.53 S/m; &, = 37.079; p = 1000
kg/m3
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5S Configuration:
— Probe: EX3DV4 — SN3976; ConvF(5.56, 5.56, 5.56) ; Calibrated: 2018/1/23
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (101x81x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.08 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 13.40 V/m; Power Drift =—0.10 dB
Peak SAR (extrapolated) = 2.89 W/kg
SAR(1 g) = 0.591 W/kg; SAR(10 g) = 0.171 W/kg
Maximum value of SAR (measured) = 1.64 W/kg

        dB
         O


        —6.00


        —12.00


        —18.00


        —24.00
                       y
                      k>
        —30.00
                    0 dB = 1.64 W/kg = 2.15 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/30


#08_WLANSGHz_802.11n—HT40 MCSO_Left Cheek_Ch110



Communication System: 802.11n; Frequency: 5550 MHz;Duty Cycle: 1:1.046
Medium: HSL_5G_181130 Medium parameters used: £= 5550 MHz; 0 = 4.881 S/m; 8 = 36.662; p = 1000
kg/m3
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5S Configuration:
— Probe: EX3DV4 — SN3976; ConvF(4.97, 4.97, 4.97) ; Calibrated: 2018/1/23
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (101x81x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.987 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 9.971 V/m; Power Drift = 0.09 dB
Peak SAR (extrapolated) = 2.14 W/kg
SAR(1 g) = 0.413 W/kg; SAR(10 g) = 0.108 W/kg
Maximum value of SAR (measured) = 1.20 W/kg

        dB
         O


        —8.00


        —16.00


        —24.00


        —32.00
                       y
                      k
        —40.00
                    0 dB = 1.20 W/kg = 0.79 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/30


#09_WLANSGHz_802.111n—HT40 MCSO_Left Cheek_Ch159



Communication System: 802.11n; Frequency: 5795 MHz;Duty Cycle: 1:1.046
Medium: HSL_5G_181130 Medium parameters used : £= 5795 MHz; 0 = 5.162 S/m; g, = 36.395; p = 1000
kg/m3
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5S Configuration:
— Probe: EX3DV4 — SN3976; ConvF(5.04, 5.04, 5.04) ; Calibrated: 2018/1/23
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
— Phantom: SAM_Right; Type: SAM; Serial: TP:1479
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (101x81x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 1.61 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 10.60 V/m; Power Drift = 0.12 dB
Peak SAR (extrapolated) = 3.58 W/kg
SAR(1 g) =0.639 W/kg; SAR(10 g) = 0.168 W/kg
Maximum value of SAR (measured) = 1.85 W/kg

        dB
         O


        —6.00


        —12.00


        —18.00


        —24.00
                       y
                      k
        —30.00
                    0 dB = 1.85 W/kg = 2.67 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29


#10_Bluetooth_1Mbps_Left Cheek_Ch0



Communication System: Bluetooth; Frequency: 2402 MHz;Duty Cycle: 1:1.295
Medium: HSL_2450_181129 Medium parameters used : £= 2402 MHz; a= 1.718 $/m; s =39.079;
p = 1000 kg/m?
Ambicnt Temperature : 23.2 °C; Liquid Temperature ! 22.2 °C

DASY3 Configuration:
— Probe: ES3DV3 — SN3169; ConvF(4.69, 4.69, 4.69) ; Calibrated: 2018/5/28
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n853; Calibrated: 2018/7/24
— Phantom: SAM_Left; Type: QD00OP4OCD; Serial: TP:1431
— Measurement SW: DASY52, Version 52.10 (1),SEMCAD X Version 14.6.11 (7439)


Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.233 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value=9.321 V/m; Power Drift=0.07 dB
Peak SAR (extrapolated) =0.401 W/kg
SAR( g) = 0.168 W/kg; SAR(10 g) = 0.072 W/kg
Maximum value of SAR (measured)= 0.226 W/kg

   dB
    0


    —4.00


    —8.00

    —12.00


    —16.00


    —20.00
                 £;
               0 dB =0.226 W/ke =—6.46 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#11_GSM850_GPRS (4 Tx slots)_Back_10mm_Ch251


Communication System: GSM850; Frequency: 848.8 MHz;Duty Cycle: 1:2.08
Medium: MSL_850_181129 Medium parameters used: f = 849 MHz; σ = 1.002 S/m; εr = 57.558; ρ
= 1000 kg/m3
                    23.2                      22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.19, 6.19, 6.19) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.478 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 22.70 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 0.551 W/kg
SAR(1 g) = 0.448 W/kg; SAR(10 g) = 0.341 W/kg
Maximum value of SAR (measured) = 0.487 W/kg




               0 dB = 0.487 W/kg = -3.12 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/26

#12_GSM1900_GPRS (4 Tx slots)_Bottom Side_10mm_Ch810


Communication System: PCS; Frequency: 1909.8 MHz;Duty Cycle: 1:2.08
Medium: MSL_1900_181126 Medium parameters used: f = 1910 MHz; σ = 1.564 S/m; εr = 52.898;
ρ = 1000 kg/m3
                    23.5                      22.5

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(4.8, 4.8, 4.8) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (41x71x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.715 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 21.60 V/m; Power Drift = -0.15 dB
Peak SAR (extrapolated) = 0.977 W/kg
SAR(1 g) = 0.547 W/kg; SAR(10 g) = 0.290 W/kg
Maximum value of SAR (measured) = 0.684 W/kg




               0 dB = 0.684 W/kg = -1.65 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#13_WCDMA V_RMC 12.2Kbps_Back_10mm_Ch4233


Communication System: WCDMA; Frequency: 846.6 MHz;Duty Cycle: 1:1
Medium: MSL_850_181129 Medium parameters used: f = 847 MHz; σ = 1 S/m; εr = 57.576; ρ =
1000 kg/m3
                    23.2                     22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.19, 6.19, 6.19) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.549 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 24.39 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 0.642 W/kg
SAR(1 g) = 0.517 W/kg; SAR(10 g) = 0.393 W/kg
Maximum value of SAR (measured) = 0.569 W/kg




               0 dB = 0.569 W/kg = -2.45 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#14_LTE Band 5_10M_QPSK_1_0_Back_10mm_Ch20525


Communication System: LTE; Frequency: 836.5 MHz;Duty Cycle: 1:1
Medium: MSL_850_181129 Medium parameters used : f = 836.5 MHz; σ = 0.988 S/m; εr = 57.675;
ρ = 1000 kg/m3
                    23.2                      22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.19, 6.19, 6.19) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x101x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.662 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 27.04 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 0.762 W/kg
SAR(1 g) = 0.625 W/kg; SAR(10 g) = 0.478 W/kg
Maximum value of SAR (measured) = 0.684 W/kg




               0 dB = 0.684 W/kg = -1.65 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/28

#15_LTE Band 12_10M_QPSK_1_0_Back_10mm_Ch23095


Communication System: LTE; Frequency: 707.5 MHz;Duty Cycle: 1:1
Medium: MSL_750_181128 Medium parameters used : f = 707.5 MHz; σ = 0.935 S/m; εr = 54.655;
ρ = 1000 kg/m3
                    23.6                      22.6

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.3, 6.3, 6.3) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x101x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.309 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 18.91 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 0.350 W/kg
SAR(1 g) = 0.290 W/kg; SAR(10 g) = 0.229 W/kg
Maximum value of SAR (measured) = 0.315 W/kg




               0 dB = 0.315 W/kg = -5.02 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#16_WLAN2.4GHz_802.11b 1Mbps_Back_10mm_Ch1


Communication System: 802.11b; Frequency: 2412 MHz;Duty Cycle: 1:1.295
Medium: MSL_2450_181129 Medium parameters used: f = 2412 MHz; σ = 1.961 S/m; εr = 51.704;
ρ = 1000 kg/m3
                    23.8                        22.8

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(4.4, 4.4, 4.4) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (91x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.245 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 6.996 V/m; Power Drift = 0.07 dB
Peak SAR (extrapolated) = 0.365 W/kg
SAR(1 g) = 0.174 W/kg; SAR(10 g) = 0.081 W/kg
Maximum value of SAR (measured) = 0.228 W/kg




               0 dB = 0.228 W/kg = -6.42 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#17_Bluetooth_1Mbps_Back_10mm_Ch0


Communication System: Bluetooth ; Frequency: 2402 MHz;Duty Cycle: 1:1
Medium: MSL_2450_181129 Medium parameters used: f = 2402 MHz; σ = 1.946 S/m; εr = 51.78; ρ
= 1000 kg/m3
                    23.2                         22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(4.4, 4.4, 4.4) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.0506 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 3.329 V/m; Power Drift = 0.12 dB
Peak SAR (extrapolated) = 0.0810 W/kg
SAR(1 g) = 0.038 W/kg; SAR(10 g) = 0.017 W/kg
Maximum value of SAR (measured) = 0.0508 W/kg




               0 dB = 0.0508 W/kg = -12.94 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#18_GSM850_GPRS (4 Tx slots)_Back_15mm_Ch251


Communication System: GSM850; Frequency: 848.8 MHz;Duty Cycle: 1:2.08
Medium: MSL_850_181129 Medium parameters used: f = 849 MHz; σ = 1.002 S/m; εr = 57.558; ρ
= 1000 kg/m3
                    23.2                      22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.19, 6.19, 6.19) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.373 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 19.97 V/m; Power Drift = 0.17 dB
Peak SAR (extrapolated) = 0.594 W/kg
SAR(1 g) = 0.374 W/kg; SAR(10 g) = 0.259 W/kg
Maximum value of SAR (measured) = 0.380 W/kg




               0 dB = 0.380 W/kg = -4.20 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/26

#19_GSM1900_GPRS (4 Tx slots)_Front_15mm_Ch810


Communication System: PCS; Frequency: 1909.8 MHz;Duty Cycle: 1:2.08
Medium: MSL_1900_181126 Medium parameters used: f = 1910 MHz; σ = 1.564 S/m; εr = 52.898;
ρ = 1000 kg/m3
                    23.5                      22.5

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(4.8, 4.8, 4.8) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1431
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.289 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 13.59 V/m; Power Drift = -0.04 dB
Peak SAR (extrapolated) = 0.345 W/kg
SAR(1 g) = 0.237 W/kg; SAR(10 g) = 0.158 W/kg
Maximum value of SAR (measured) = 0.269 W/kg




               0 dB = 0.269 W/kg = -5.70 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#20_WCDMA V_RMC 12.2Kbps_Back_15mm_Ch4233


Communication System: WCDMA; Frequency: 846.6 MHz;Duty Cycle: 1:1
Medium: MSL_850_181129 Medium parameters used: f = 847 MHz; σ = 1 S/m; εr = 57.576; ρ =
1000 kg/m3
                    23.2                     22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.19, 6.19, 6.19) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.461 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 22.42 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 0.531 W/kg
SAR(1 g) = 0.429 W/kg; SAR(10 g) = 0.327 W/kg
Maximum value of SAR (measured) = 0.470 W/kg




               0 dB = 0.470 W/kg = -3.28 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#21_LTE Band 5_10M_QPSK_1_0_Back_15mm_Ch20525


Communication System: LTE; Frequency: 836.5 MHz;Duty Cycle: 1:1
Medium: MSL_850_181129 Medium parameters used : f = 836.5 MHz; σ = 0.988 S/m; εr = 57.675;
ρ = 1000 kg/m3
                    23.2                      22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.19, 6.19, 6.19) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.574 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 25.33 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 0.669 W/kg
SAR(1 g) = 0.543 W/kg; SAR(10 g) = 0.415 W/kg
Maximum value of SAR (measured) = 0.596 W/kg




               0 dB = 0.596 W/kg = -2.25 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/28

#22_LTE Band 12_10M_QPSK_1_0_Back_15mm_Ch23095


Communication System: LTE; Frequency: 707.5 MHz;Duty Cycle: 1:1
Medium: MSL_750_181128 Medium parameters used : f = 707.5 MHz; σ = 0.935 S/m; εr = 54.655;
ρ = 1000 kg/m3
                    23.6                      22.6

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(6.3, 6.3, 6.3) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (61x111x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 0.209 W/kg

Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 15.68 V/m; Power Drift = -0.00 dB
Peak SAR (extrapolated) = 0.240 W/kg
SAR(1 g) = 0.199 W/kg; SAR(10 g) = 0.156 W/kg
Maximum value of SAR (measured) = 0.214 W/kg




               0 dB = 0.214 W/kg = -6.70 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#23_WLAN2.4GHz_802.11b 1Mbps_Back_15mm_Ch1


Communication System: 802.11b; Frequency: 2412 MHz;Duty Cycle: 1:1
Medium: MSL_2450_181129 Medium parameters used: f = 2412 MHz; σ = 1.961 S/m; εr = 51.704;
ρ = 1000 kg/m3
                    23.8                        22.8

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(4.4, 4.4, 4.4) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (91x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.0979 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 5.236 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 0.147 W/kg
SAR(1 g) = 0.073 W/kg; SAR(10 g) = 0.036 W/kg
Maximum value of SAR (measured) = 0.0930 W/kg




               0 dB = 0.0930 W/kg = -10.32 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/30


#24_WLANSGHz_802.11n—HT40 MCSO_Back_15mm_Ch46

Communication System: 802.11n; Frequency : 5230 MHz;Duty Cycle: 1:1.046
Medium: MSL_5G_181130 Medium parameters used: f = 5230 MHz; 0 = 5.361 S/m; g, = 47.569; p = 1000
kg/m3
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5S Configuration:
— Probe: EX3DV4 — SN3976; ConvF(4.92, 4.92, 4.92) ; Calibrated: 2018/1/23
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
— Phantom: SAM_Left; Type: QDO0O0OP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (91x81x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.769 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 12.44 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 1.27 W/kg
SAR(1 g) = 0.368 W/kg; SAR(10 g) = 0.153 W/kg
Maximum value of SAR (measured) = 0.785 W/kg

        dB
         O


        —0.40


        —0.80


        —1.20


        —1.60


        —2.00         _
                    0 dB = 0.785 W/kg = —1.05 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/30


#25_WLANSGHz_802.11n—HT40 MCSO_Back_15mm_Ch110

Communication System: 802.11n; Frequency: 5550 MHz;Duty Cycle: 1:1.046
Medium: MSL_5G_181130 Medium parameters used: £ = 5550 MHz; 0 = 5.777 S/m; g, = 47.04; p = 1000
kg/m3
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5S Configuration:
— Probe: EX3DV4 — SN3976; ConvF(4.28, 4.28, 4.28) ; Calibrated: 2018/1/23
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
— Phantom: SAM_Left; Type: QDO0O0OP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (91x81x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.504 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 9.869 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 0.872 W/kg
SAR(1 g) = 0.235 W/kg; SAR(10 g) = 0.097 W/kg
Maximum value of SAR (measured) = 0.519 W/kg

        dB
         O


        —3.00


        —6.00


        —9.00


        —12.00


        —15.00        C‘
                    0 dB = 0.519 W/kg = —2.85 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/30


#26_WLANSGHz_802.11n—HT40 MCSO_Back_15mm_Ch159

Communication System: 802.11n; Frequency: 5795 MHz;Duty Cycle: 1:1.046
Medium: MSL_5G_181130 Medium parameters used: f = 5795 MHz; 0 = 6.116 S/m; &g, = 46.666; p = 1000
kg/m3
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5S Configuration:
— Probe: EX3DV4 — SN3976; ConvF(4.46, 4.46, 4.46) ; Calibrated: 2018/1/23
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
— Phantom: SAM_Left; Type: QDO0O0OP40CD; Serial: TP:1477
— Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (91x81x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 0.851 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 11.60 V/m; Power Drift =—0.11 dB
Peak SAR (extrapolated) = 1.56 W/kg
SAR(1 g) = 0.385 W/kg; SAR(10 g) = 0.156 W/kg
Maximum value of SAR (measured) = 0.865 W/kg

        dB
         O


        —3.00


        —6.00


        —9.00


        —12.00


        —15.00        _
                    0 dB = 0.865 W/kg = —0.63 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018/11/29

#27_Bluetooth_1Mbps_Back_15mm_Ch0


Communication System: Bluetooth; Frequency: 2402 MHz;Duty Cycle: 1:1.295
Medium: MSL_2450_181129 Medium parameters used: f = 2402 MHz; σ = 1.946 S/m; εr = 51.78; ρ
= 1000 kg/m3
                    23.2                        22.2

DASY5 Configuration:
- Probe: ES3DV3 - SN3169; ConvF(4.4, 4.4, 4.4) ; Calibrated: 2018/5/28
- Sensor-Surface: 3mm (Mechanical Surface Detection)
- Electronics: DAE4 Sn853; Calibrated: 2018/7/24
- Phantom: SAM-Right; Type: SAM; Serial: TP-1503
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 0.0196 W/kg

Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 1.934 V/m; Power Drift = 0.15 dB
Peak SAR (extrapolated) = 0.0290 W/kg
SAR(1 g) = 0.014 W/kg; SAR(10 g) = 0.00646 W/kg
Maximum value of SAR (measured) = 0.0196 W/kg




               0 dB = 0.0196 W/kg = -17.08 dBW/kg