HAC_T-Coil Appendix B_PY7-21831A_Rev 01_02142018

FCC ID: PY7-21831A

Test Report

Download: PDF
FCCID_3782174

Calibration Laboratory of                                        F14079                                            Schweizerischer Kalibrierdienst
Schmid & Partner                                                 inyw<<*4                                  S
                                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                 i;m/a  *                                  C       Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      3/ m                                     S       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates


Client      Sporton (Auden)                                                                     Certificate No: AM1DV3—3130_Nov17

|CALIBRATION CERTIFICATE                                                                                                                             |
 Object                                AM1DV3 — SN: 3130


 Calibration procedure(s)              QA CAL—24.v4
                                       Calibration procedure for AM1D magnetic field probes and TMFS in the
                                       audio range

 Calibration date:                     November 21, 2017




 This calibration certificate documents the traceability to national standards, which realize the physicalunits of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               31—Aug—17 (No. 21092)                                 Aug—18
 Reference Probe AM1DV2                 SN: 1008                  30—Dec—16 (No. AM1D—1008_Dec16)                       Dec—17
 DAE4                                   SN: 781                   13—Jul—17 (No. DAE4—781_Jul17)                        Jul—18




 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 AMCC                                   SN: 1050                  01—Oct—13 (in house check Oct—17)                     Oct—19
 AMMI Audio Measuring Instrument        SN: 1062                  26—Sep—12 (in house check Oct—17)                     Oct—19




                                        Name                                   Function                                 Signature
 Calibrated by:                         Leif Klysner                           Laboratory Technician                  fi%‘/‘



 Approved by:                           KatJa Pokovic                          Technical Manager

                                                                                                                       ckzr
                                                                                                                        Issued: November21, 2017
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



Certificate No: AM1DV3—3130_Nov17                                     Page 1 of 3


[References

11       ANSI—C63.19—2007
         American National Standard for Methods of Measurement of Compatibility between Wireless
         Communications Devices and Hearing Aids.
[2]      ANS1—C63.19—2011
         American National Standard, Methods of Measurement of Compatibility between Wireless
         Communications Devices and Hearing Aids.
[31      DASY5 manual, Chapter: Hearing Aid Compatibility (HAC) T—Coil Extension

Description of the AM1D probe

      The AM1D Audio Magnetic Field Probe is a fully shielded magnetic field probe for the frequency
      range from 100 Hz to 20 kHz. The pickup coil is compliant with the dimensional requirements of
      [1+2]. The probe includes a symmetric low noise amplifier for the—signal available at the shielded
      3 pin connector at the side. Poweris supplied via the same connector (phantom power supply)
      and monitored via the LED near the connector. The 7 pin connector at the end of the probe does
      not carry any signals, but determines the angle of the sensor when mounted on the DAE. The
      probe supports mechanical detection of the surface.
      The single sensor in the probe is arranged in a tilt angle allowing measurement of 3 orthogonal
      field components when rotating the probe by 120° around its axis. It is aligned with the
      perpendicular component of the field, if the probe axis is tilted nominally 35.3° above the
      measurement plane, using the connector rotation and sensor—angle stated below.
      The probe is fully RF shielded when operated with the matching signal cable (shielded) and
      allows measurement of audio magnetic fields in the close vicinity of RF emitting wireless devices
      according to [1+2] without additional shielding.

Handling of the item

      The probe is manufactured from stainless steel. In order to maintain the performance and
      calibration of the probe, it must not be opened. The probe is designed for operation in air and
      shall not be exposed to humidity or liquids. For proper operation of the surface detection and
      emergency stop functions in a DASY system, the probe must be operated with the special probe
      cup provided (larger diameter).

Methods Applied and Interpretation of Parameters

      Coordinate System: The AM1D probe is mounted in the DASY system for operation with a HAC
      Test Arch phantom with AMCC Helmholtz calibration coil according to [3], with the tip pointing to
      "southwest" orientation.
      Functional Test: The functional test preceding calibration includes test of
      Noise level
      RF immunity (1kHz AM modulated signal). The shield of the probe cable must be well connected.
      Frequency response verification from 100 Hz to 10 kHz.
      Connector Rotation: The connector at the end of the probe does not carry any signals and is used
      for fixation to the DAE only. The probe is operated in the center of the AMCC Helmholtz coil using a
      1 kHz magnetic field signal. Its angle is determined from the two minima at nominally +120° and —
      120° rotation, so the sensor in the tip of the probe is aligned to the vertical plane in z—direction,
      corresponding to the field maximum in the AMCC Helmholtz calibration coil.
      Sensor Angle: The sensor tilting in the vertical plane from the ideal vertical direction is determined
      from the two minima at nominally +120° and —120°. DASY system uses this angle to align the
      sensor for radial measurements to the x and y axis in the horizontal plane.
Sensitivity: With the probe sensor aligned to the z—field in the AMCC, the output of the probe is
        compared to the magnetic field in the AMCC at 1 kHz. The field in the AMCC Helmholtz coil is
         given by the geometry and the current through the coil, which is monitored on the precision
         shunt resistor of the coil.

Certificate No: AM1DV3—3130_Nov17                   Page 2 of 3


AM1D probe identification and configuration data

    Item                       AM1DV3 Audio Magnetic 1D Field Probe
    Type No                    SP AM1 001 BA
    Serial No                  3130


    Overall length             296 mm
    Tip diameter               6.0 mm (at the tip)
    Sensor offset              3.0 mm (centre of sensor from tip)
    Internal Amplifier         20 dB


    Manufacturer / Origin      Schmid & Partner Engineering AG, Zurich, Switzerland
    Manufacturing date         July 09, 2012
    Last calibration date       November 16, 2016




Calibration data


Connector rotation angle       (in DASY system)       81.2 °              +/— 3.6 ° (k=2)


Sensor angle                   (in DASY system)       1.03 °              +/— 0.5 ° (Kk=2)


Sensitivity at 1 kHz           (in DASY system)       0.00743 V / (A/m)   +/— 2.2 % (k=2)




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 mulftiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.


Certificate No: AM1DV3—3130_Novi7                 Page 3 of 3


                 mm® in Collaboration with                                       §‘$\I_J/u/"”z,                      PBEnf
               m   ‘/"/ "/                                                      is            7                      EF4A
               Cmm                      “

                                                                                 7           y p" CALIBRATION
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China    %4@;&‘                              CNAS LO570
            Tel: +86—10—62304633—2218       Fax: +86—10—62304633—2209
            E—mail: cttl@chinattl.com       Http://www.chinattl.cn
      Client :        sporton                                                   Certificate No: Z17—97055




Object                                   DAE4 — SN: 854


Calibration Procedure(s)                  FF—211—002—01

                                          Calibration Procedure for the Data Acquisition Electronics
                                          (DAEx)

Calibration date:                         May 02, 2017

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)

Primary Standards               ID #           Cal Date(Calibrated by, Certificate No.)           Scheduled Calibration



Process Calibrator 753       1971018               27—June—16 (CTTL, No:J16X04778)                         June—17




                                Name                      Function                                   Signature
Calibrated by:                   Yu Zongying              SAR Test Engineer                           %

Reviewed by:                     Lin Hao                 SAR Test Engineer                            ‘fi@%

Approved by:                     Qi Dianyuan              SAR Project Leader                  _—%/
                                                                                             Issued: May 03, 2017
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




     Certificate No: Z17—97055                               Page 1 of 3


         _A®                       In Collaboration with
        =‘/"[‘]
        ~iigaage»""
                    a
                    CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2218      Fax: +86—10—62304633—2209
       E—mail: cttl@chinattl.com        Http:/www.chinattl.on



Glossary:
DAE                            data acquisition electronics
Connector angle                 information used in DASY system to align probe sensor X
                               to the robot coordinate system.


Methods Applied and Interpretation of Parameters:
e DC Voltage Measurement. Calibration Factor assessed for use in DASY
   system by comparison with a calibrated instrument traceable to national
   standards. The figure given corresponds to the full scale range of the
   voltmeter in the respective range.

e   Connector angle: The angle of the connector is assessed measuring the
    angle mechanically by a tool inserted. Uncertainty is not required.

e   The report provide only calibration results for DAE, it does not contain other
    performance test results.




Certificate No: Z17—97055                                  Page 2 of 3


          mm©                     in Collaboration with

        =‘7/"7‘J,
          aage>"
                  s p e a q      CALIBRATION LABORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2218      Fax: +86—10—62304633—2209
       E—mail: cttlt@chinattl.com     Http:/www.chinattl.on



DC Voltage Measurement
   A/D — Converter Resolution nominal
      High Range:     1LSB =      6.A1uV ,    full range =    —100...+300 mV
      Low Range:     1LSB =        61nvV ,    full range =    ~1.22222. +3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


     Calibration Factors                    X                              Y                   Z.
     High Range                  405.005 + 0.15% (k=2)         404.208 + 0.15% (k=2)   405.318 + 0.15% (k=2)

     Low Range                   3.96053 + 0.7% (k=2)          3.94760 + 0.7% (k=2)    3.96516 + 0.7% (k=2)


Connector Angle

    Connector Angle to be used in DASY system                                               325.5° +1 °




Certificate No: Z17—97055                                 Page 3 of 3



Document Created: 2018-03-16 04:38:16
Document Modified: 2018-03-16 04:38:16

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC