RF Exposure Info._20190712_v1 - PY7-11643I_TestRpt_RFExp_Appendix A

FCC ID: PY7-11643I

RF Exposure Info

Download: PDF
FCCID_4360527

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/14


System Check_Head_750MHz

DUT: D750V3—1012

Communication System: CW; Frequency: 750 MHz;Duty Cycle: 1:1
Medium: HSL_750_190614 Medium parameters used: £ = 750 MHz; 0 = 0.89 S/m; &, = 40.646; p = 1000
kg/m3
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3124; ConvF(6.15, 6.15, 6.15) ; Calibrated: 2019/1/15
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn316; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QD000P40CD ; Serial: TP:1684
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.61 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 55.40 V/m; Power Drift =—0.01 dB
Peak SAR (extrapolated) =3.35 W/kg
SAR(I g) = 2.26 W/kg; SAR(10 g) =1.52 W/kg
Maximum value of SAR (measured) = 2.63 W/kg

        dB
         0


        —1.96


        —3.91


        —5.87


        —£.82


        age           &
                    0 dB = 2.63 W/kg = 4.20 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/14


System Check_Head_835MHz

DUT: D835V2—499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL_8$50_190614 Medium parameters used: £= 835 MHz; 0 = 0.877 S/m; s, = 41.199; p = 1000
kg/m3
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3124; ConvF(5.97, 5.97, 5.97) ; Calibrated: 2019/1/15
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn316; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QD000P40CD; Serial: TP:1684
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 2.92 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 58.76 V/m; Power Drift = —0.06 dB
Peak SAR (extrapolated) = 3.68 W/kg
SAR(I g) = 2.46 W/kg; SAR(10 g) =1.62 W/kg
Maximum value of SAR (measured) = 2.86 W/kg

        dB
         0


        —2.08


        —4.16


        —6.24


        —8.32


        1o.an         &
                    0 dB = 2.86 W/kg = 4.56 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/13


System Check_Head_1750MHz

DUT: D1750V2—1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: HSL_1750_190613 Medium parameters used: £ = 1750 MHz; 0 = 1.365 S/m; &, = 40.966; p = 1000
kg/m3
Ambient Temperature : 23.9 °C ; Liquid Temperature : 22.9 °C

DASY5 Configuration
— Probe: EX3DV4 — SN7306;ConvF(8.69, 8.69, 8$.69) ;Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 $Sn699; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QDOO0OP40CD ; Serial: S/N:1801
— Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 14.7 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 106.0 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 16.9 W/kg
SAR(I g) =9.28 W/kg; SAR(10 g) = 4.97 W/kg
Maximum value of SAR (measured) = 13.9 W/kg

        dB
         0


        —3.59


        —7.18


        10.77


        —14.36


        angs          &
                    0 dB = 13.9 W/kg = 11.43 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/14


System Check_Head_1750MHz

DUT: D1750V2—1068

Communication System: CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: HSL_1750_190614 Medium parameters used: £= 1750 MHz; 0 = 1.373 S/m; & = 41.086; p = 1000
kg/m>
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASYS5 Configuration
— Probe: ES3DV3 — SN3170;ConvF(5.32, 5.32, 5.32) ;Calibrated: 2018/11/2
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n699; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QDO0O0OP40CD; Serial: S/N:1801
— Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 11.5 W/kg

Pin=250mW/Zoom Scan (Sx5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 92.59 V/m; Power Drift = 0.05dB
Peak SAR (extrapolated) = 16.3 W/kg
SAR(1 g) = 9.04 W/kg; SAR(10 g) = 4.8 W/kg
Maximum value of SAR (measured) = 11.3 W/kg

        dB
        0

        —3.54


        —£ .07

        —10.61


        —14.14


        —17.68
                    0 dB = 11.3 W/kg = 10.53 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/13


System Check_Head_1900MHz

DUT: D1900V2—50d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900_190613 Medium parameters used: £ = 1900 MHz; 0 = 1.407 S/m; &, = 40.285; p = 1000
kg/m3
Ambient Temperature : 23.9 °C ; Liquid Temperature : 22.9 °C

DASY5 Configuration
— Probe: EX3DV4 — SN7306;ConvF(8.26, 8.26, 8.26) ;Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAFE4 $Sn699; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QDOO0OP40CD ; Serial: S/N:1801
— Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11 (7439)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 17.1 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 111.9 V/m; Power Drift =—0.08 dB
Peak SAR (extrapolated) = 19.3 W/kg
SAR(I g) = 10.3 W/kg; SAR(10 g) = 5.29 W/kg
Maximum value of SAR (measured) = 16.0 W/kg

        dB
         0


        —3.83


        —7.66


        —11.48


        —15.31


        4g44          &
                    0 dB = 16.0 W/kg = 12.04 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/14


System Check_Head_1900MHz

DUT: D1900V2—5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900_190614 Medium parameters used: £= 1900 MHz; 0 = 1.417 S/m; &, = 40.406; p = 1000
kg/m>
Ambient Temperature : 23.6 °C ; Liquid Temperature : 22.6 °C

DASYS5 Configuration
— Probe: ES3DV3 — SN3169;ConvF(5.14, 5.14, 5.14) ;Calibrated: 2019/5/24
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n699; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QDO0O0OP40CD; Serial: S/N:1801
— Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.7 W/kg

Pin=250mW/Zoom Scan (Sx5x7)/Cube 0; Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 97.48 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 18.4 W/kg
SAR(1 g) = 10 W/kg; SAR(10 g) = 5.2 W/kg
Maximum value of SAR (measured) = 12.5 W/kg

        dB
        0

        —3.66


        —£.33

        —10.99


        —14.66


        —18.32
                    0 dB = 12.5 W/kg = 10.97 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/17


System Check_Head_2450MHz

DUT: D2450V2—736

Communication System: CW; Frequency : 2450 MHz;Duty Cycle: 1:1
Medium: HSL_2450_190617 Medium parameters used : £= 2450 MHz; 0 = 1.85 S/m; &, = 40.083; p = 1000 kg/m>
Ambient Temperature : 23.2 °C ; Liquid Temperature : 22.2 C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(7.43, 7.43, 7.43) ; Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn§53; Calibrated: 2018/7/24
— Phantom: SAM LEFT; Type: QDOO0OP4O0CD; Serial: TP:1718
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 19.7 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 110.7 V/m;, Power Drift = 0.05 dB
Peak SAR (extrapolated) = 25.8 W/kg
SAR(I g) = 12.9 W/kg; SAR(10 g) =6.07 W/kg
Maximum value of SAR (measured) = 21.3 W/kg

              dB
              0


               —4.56


               —9.12


               —13.68


               —18.24


               —22.80
                          0 dB = 19.7 W/kg = 12.94 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/15


System Check_Head_2600MHz

DUT: D2600V2—1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: HSL_2600_190615 Medium parameters used: £= 2600 MHz; 0 = 1.989 S/m; g, = 38.243; p = 1000
kg/m>
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5 Configuration
— Probe: ES3DV3 — SN3170;ConvF(4.43, 4.43, 4.43) ;Calibrated: 2018/11/2
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 $n699; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QDO0O0OP4O0CD; Serial: S/N:1801
— Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (81x81x1):; Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 18.6 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 98.77 V/m;, Power Drift = 0.02 dB
Peak SAR (extrapolated) = 30.0 W/kg
SAR(1 g) =13.9 W/kg; SAR(10 g) = 6.27 W/kg
Maximum value of SAR (measured) = 18.6 W/kg

        dB
        0


        —4.63


        —9.25


        —13.88


        —18.50


        —23.13        .
                    0 dB = 18.6 W/kg = 12.70 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/15


System Check_Head_2600MHz

DUT: D2600V2—1008

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: HSL_2600_190615 Medium parameters used: £f = 2600 MHz; 0 = 1.989 S/m; &, = 38.243; p = 1000
kg/m3
Ambient Temperature : 23.3 °C ; Liquid Temperature : 22.3 °C

DASY5 Configuration:
— Probe: ES3DV3 — SN3124; ConvF(4.37, 4.37, 4.37) ; Calibrated: 2019/1/15
— Sensor—Surface: 3mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn316; Calibrated: 2019/1/3
— Phantom: SAM_Left; Type: QD000P40CD ; Serial: TP:1684
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=250mW/Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 18.01 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0; Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 93.07 V/m; Power Drift = 0 dB
Peak SAR (extrapolated) = 29.286 W/kg
SAR(I g) =13.45 W/kg; SAR(10 g) = 6.042 W/kg
Maximum value of SAR (measured) = 18.01 W/kg

        dB
         0


        —3.80


        —7.60


        —11.40


        —15.20


        19.00         &
                    0 dB = 18.01 Whkg = 25.11 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/19


System Check_Head_5250MHz

DUT: DSGHzV2—1006

Communication System: CW ; Frequency: 5250 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190619 Medium parameters used : £= 5250 MHz; 0 = 4.601 S/m; &, = 35.923; p = 1000 kg/m>
Ambient Terperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(5.38, 5.38, 5.38) ; Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn§53; Calibrated: 2018/7/24
— Phantom: SAM LEFT; Type: QDOO0OP4O0CD; Serial: TP:1718
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 18.1 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 70.94 V/m; Power Drift =—0.19 dB
Peak SAR (extrapolated) =31.2 W/kg
SAR(I g) =7.63 W/kg; SAR(10 g) = 2.09 W/kg
Maximum value of SAR (measured) = 19.0 W/kg

              dB
              0


               —5.89


               —11.78


               —17.67


               —23.56


               —29.45
                          0 dB = 18.1 W/kg = 12.58 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/19


System Check_Head_5600MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5600 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190619 Medium parameters used: £= 5600 MHz; 0 = 4.934 S/m; s, = 35.463; p = 1000 kg/m>
Ambient Terperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(4.71, 4.71, 4.71) ; Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn§53; Calibrated: 2018/7/24
— Phantom: SAM LEFT; Type: QDOO0OP4O0CD; Serial: TP:1718
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.9 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value =72.51 V/m; Power Drift =—0.14 dB
Peak SAR (extrapolated) = 36.7 W/kg
SAR(1 g) = 8.25 W/kg; SAR(10 g) =2.22 W/kg
Maximum value of SAR (measured) =21.1 W/kg

              dB
              0


               —5.96


               —11.93


               —17.89


               —23.86


               —29.82
                          0 dB = 20.9 W/kg = 13.20 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2019/6/19


System Check_Head_5750MHz

DUT: DSGHzV2—1006

Communication System: CW; Frequency: 5750 MHz;Duty Cycle: 1:1
Medium: HSL_5G_190619 Medium parameters used: £= 5750 MHz; 0 = 5.095 S/m; s, = 35.231; p = 1000 kg/m>
Ambient Terperature : 23.5 °C ; Liquid Temperature : 22.5 °C

DASYS5 Configuration:
— Probe: EX3DV4 — SN7306; ConvF(5, 5, 5) ; Calibrated: 2018/7/26
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAF4 Sn§53; Calibrated: 2018/7/24
— Phantom: SAM LEFT; Type: QDOO0OP4O0CD; Serial: TP:1718
— Measurement SW: DASY52, Version 52.10 (2);SEMCAD X Version 14.6.12 (7470)


Pin=100mW/Area Scan (71x71x1): Interpolated grid: dx=1.000 mm, dy=1.000 mm
Maximum value of SAR (interpolated) = 20.2 W/kg

Pin=100mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 69.52 V/m; Power Drift =—0.17 dB
Peak SAR (extrapolated) = 34.7 W/kg
SAR(I g) =7.72 W/kg; SAR(10 g) =2.09 W/kg
Maximum value of SAR (measured) = 20.0 W/kg

              dB
              0


               —6.85


               —13.69


               —20.54


               —27.38


               —34.23
                          0 dB = 20.2 W/kg = 13.05 dBW/kg