RF Exposure Appendix A

FCC ID: PXTIGM-02

RF Exposure Info

Download: PDF
FCCID_1872625

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2012/12/6


System Check_Body_835MHz_121206

DUT: D835V2—SN:499

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium : MSL_850_121206 Medium parameters used: {= 835 MHz; 0 = 0.976 mho/m; s, = 52.883; p =
1000 kg/m*
Ambient Temperature : 22.2 °C; Liquid Temperature : 21.2 °C

DASYS Configuration:
— Probe: ET3DV6 — SN1787; ConvF(6.08, 6.08, 6.08); Calibrated: 2012/5/29;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE3 $n495; Calibrated: 2012/4/23
— Phantom : ELI v4.0; Type: QDOVAOO1BB:; Serial: TP:1127
— Measurement SW: DASY52, Version 52.8 (3); SEMCAD X Version 14.6.5 (6469)


Configuration/Pin=250mW/Area Scan (61x61x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated) = 2.49 m W/g

Configuration/Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=Smm
Reference Value = 52.052 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 3.553 mW/g
SAR(I g) = 2.29 mW/g; SAR(10 g) = 1.44 mW/g
Maximum value of SAR (measured) = 2.51 mW/g

     dB
      0




      —2.15




      ~4.30




      6.45




      —8.60



      —10.75
                     &.
                      2¥
                                             &
                  0 dB = 2.51 mW/g = 7.99 dB mW/g


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2012/12/6


System Check_Body_1900MHz_121206

DUT: D1900V2—SN:5d041

Communication System: CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium : MSL_1900_121206 Medium parameters used: £=1900 MHz; 0 = 1.539 mho/m; &, = 54.55; p =
1000 kg/m*
Ambient Temperature : 22.6 °C; Liquid Temperature : 21.6 °C

DASYS Configuration:
— Probe: ET3DV6 — SN1787; ConvF(4.58, 4.58, 4.58); Calibrated: 2012/5/29;
— Sensor—Surface: 4mm (Mechanical Surface Detection)
— Electronics: DAE3 $n495; Calibrated: 2012/4/23
— Phantom : ELI v5.0 Left; Type: QDOVAOOZAA; Serial: TP:1131
— Measurement SW: DASY52, Version 52.8 (3); SEMCAD X Version 14.6.5 (6469)


Configuration/Pin=250mW/Area Scan (61x61x1): Measurement grid: dx=1 5mm, dy=1 5mm
Maximum value of SAR (interpolated) = 11.4 mW/g

Configuration/Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=Smm
Reference Value = 88.644 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 15.299 mW/g
SAR(I g) = 9.6 mW/g; SAR(10 g) = 5.16 mW/g
Maximum value of SAR (measured) = 11.0 mW/g

     dB
      0




      —3.54




      —7.08



      —10.61




      1415

                      2¥

      —17.69
                     K
                  0 dB = 11.0 mW/g = 20.83 dB mW/g



Document Created: 2019-09-09 17:46:02